K Number
K190487
Date Cleared
2020-02-18

(355 days)

Product Code
Regulation Number
868.1890
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The InPen dose calculator, a component of the InPen app, is indicated for the management of diabetes by people with diabetes age 12 and older by calculating an insulin dose or carbohydrate intake based on user entered data. The device is indicated for use with NovoLog® or Humalog® U-100 insulin.

For an insulin dose based on amount of carbohydrates, a healthcare provide patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the software prior to use.

For an insulin dose based on fixed/variable meal sizes, a healthcare professional must provide patient-specific fixed doses/ meal sizes to be programmed into the software prior to use.

Device Description

The InPen app is designed to manage the wireless transfer of insulin dose data from the InPen, log insulin dose data, and provide a dose calculator to aid mealtime insulin dose calculations. The insulin dose calculations provided by the app are meant for patients undergoing multiple daily injection (MDI) therapy. The InPen app is not intended to serve as an accessory to an insulin pump.

AI/ML Overview

The provided text describes the InPen Dose Calculator, outlining its indications for use and
comparing it to a predicate device. However, it does not contain specific acceptance criteria, a
detailed study that proves the device meets those criteria, or the requested specific performance
metrics like sensitivity, specificity, or accuracy
.

Here's an attempt to answer the questions based only on the provided text, highlighting
where information is missing:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state numerical acceptance criteria or present a table of device
performance against such criteria. It generally states that the device "satisfies
all functional performance and safety requirements, meets its intended use, and is safe for the
intended user population."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data,
retrospective or prospective)

The text mentions a "summative evaluation" where "patients with sufficient diabetes knowledge
completed self-training and then completed a series of critical tasks." However, it does not
specify the sample size of this test set, the country of origin of the data, or whether it was
retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of
those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document. The "ground truth" for the dose calculations
would theoretically be the correct insulin dose based on provided parameters, but the process of
establishing this for the test set is not detailed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe any adjudication method for the test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was
the effect size of how much human readers improve with AI vs without AI assistance

The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study.
The InPen Dose Calculator is an algorithm to calculate insulin doses, not designed for human
"readers" to interpret medical images or data. Therefore, the concept of human readers improving
with AI assistance in this context does not apply in the manner typically associated with MRMC
studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The "summative evaluation" involved patients using the device, but the core function is an algorithm
calculating doses based on user input. The document states that the "dose calculator uses the
standard approach using healthcare provider specified insulin-to-carbohydrate ratio and insulin
sensitivity factors for making calculations." This implies a standalone algorithmic function based
on pre-programmed parameters. The "Clinical Evidence" section focuses on usability and safety with
human users
, rather than solely on the algorithm's performance in isolation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the dose calculations, the "ground truth" is implied to be the standard approach using
healthcare provider specified insulin-to-carbohydrate ratio and insulin sensitivity factors for
making calculations.
The text also mentions: "the calculator includes a consideration for insulin
on-board based on the published study by Mudaliar et al (1999) for the duration of insulin action."
This suggests calculations are validated against established medical formulas and literature.

8. The sample size for the training set

The document does not provide any information about a training set or its sample size. The device
appears to be a rule-based calculator rather than a machine learning model that would typically
require a training set.

9. How the ground truth for the training set was established

As no training set is mentioned or implied for this rule-based dose calculator, this question is not
applicable based on the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 18, 2020

Companion Medical, Inc. Jasper Benke Vice President, RA/QA/CA 11011 Via Frontera, Suite D San Diego, CA 92127

Re: K190487

Trade/Device Name: InPen Dose Calculator Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: Class II Product Code: NDC Dated: January 16, 2020 Received: January 17, 2020

Dear Jasper Benke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190487

Device Name InPen Dose Calculator

Indications for Use (Describe)

The InPen dose calculator, a component of the InPen app, is indicated for the management of diabetes by people with diabetes age 12 and older by calculating an insulin dose or carbohydrate intake based on user entered data. The device is indicated for use with NovoLog® or Humalog® U-100 insulin.

For an insulin dose based on amount of carbohydrates, a healthcare provide patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the software prior to use.

For an insulin dose based on fixed/variable meal sizes, a healthcare professional must provide patient-specific fixed doses/ meal sizes to be programmed into the software prior to use.

Type of Use ( Select one or both, as applicable )
------------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Companion Medical, Inc.

510(k) SUMMARY - K190487 InPen Dose Calculator

l. SUBMITTER

Address:Companion Medical, Inc.
11011 Via Frontera, Suite D
San Diego, California 92127
Phone:(858) 522-0252
Contact:Mr. Jasper Benke
Date Prepared:February 18, 2020

II. DEVICE

Name of Device: InPen Dose Calculator Common Name: Insulin Dose Calculator Classification Name: Predictive pulmonary-function value calculator (21 CFR 868.1890); Class II Product Codes: NDC

III. PREDICATE DEVICES

InPen System (K160629)

This predicate has not been subject to a design-related recall. No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The InPen app is designed to manage the wireless transfer of insulin dose data from the InPen, log insulin dose data, and provide a dose calculator to aid mealtime insulin dose calculations. The insulin dose calculations provided by the app are meant for patients undergoing multiple daily injection (MDI) therapy. The InPen app is not intended to serve as an accessory to an insulin pump.

V. INDICATIONS FOR USE

The InPen dose calculator, a component of the InPen app, is indicated for the management of diabetes by people with diabetes age 12 and older by calculating an insulin dose or carbohydrate intake based on user entered data. The device is indicated for use with NovoLog® or Humalog® U-100 insulin.

{4}------------------------------------------------

For an insulin dose based on amount of carbohydrates, a healthcare professional must provide patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the software prior to use.

For an insulin dose based on fixed/variable meal sizes, a healthcare professional must provide patient-specific fixed doses/meal sizes to be programmed into the software prior to use.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The InPen dose calculator is substantially equivalent to other legally marketed dose calculators. Specifically, the InPen dose calculator is substantially equivalent to the InPen dose calculator (K160629) cleared on July 26, 2016. The InPen dose calculator has the same intended use and indications, technological characteristics, and principles of operation as the previously cleared predicate device. A substantial equivalence chart of the similarities and differences between the InPen dose calculator and the predicate device is shown in Table 1. The minor differences in technological characteristics do not change the intended use or raise new questions of safety or effectiveness.

AttributeSubject Device(K190487)Predicate Device(K160629)
Classification(Class II - NDC - 21 CFR §868.1890)Same
Indications For UseThe InPen dose calculator, acomponent of the InPen app, isindicated for the management ofdiabetes by people with diabetesage 12 and older by calculating aninsulin dose or carbohydrate intakebased on user entered data.The InPen dose calculator, acomponent of the InPen app, isindicated for the management ofdiabetes by people with diabetesage 12 and older by calculatingan insulin dose or carbohydrateintake based on user entereddata. Prior to use, a healthcareprofessional must provide thepatient-specific target bloodglucose, insulin-to- carbohydrateratio, and insulin sensitivityparameters to be programmedinto the software.
For an insulin dose based on amountof carbohydrates, a healthcareprofessional must provide patient-specific target blood glucose,insulin-to-carbohydrate ratio, andinsulin sensitivity parameters to beprogrammed into the software priorto use.
For an insulin dose based onfixed/variable meal sizes, ahealthcare professional mustprovide patient-specific fixeddoses/meal sizes to be programmedinto the software prior to use.

{5}------------------------------------------------

Prescription UseYesSame
User GroupDiabetes patients treated with multiple daily insulin injection (MDI) therapySame
Communication with insulin pumpsNoSame
Software Level of ConcernMajorSame
Wireless ConnectivityBluetooth Low Energy (BLE)Same
Control or affect blood glucose measurementsNoSame
Control or affect insulin deliveryNoSame
Reports, graphs, and Electronic Log BookYesSame
Meal Size EntryPre-specified meal/dose sizes (fixed or variable)Grams of carbohydrates
Insulin Dose CalculatorCalculates insulin doses for meals and corrections while accounting for insulin on boardSame
Carbohydrate CalculatorCalculates carbohydrate intake based on user-entered dataSame
Manual Dose EntryYesSame
InPen Dose EntryYesSame
Tracking of residual bolus insulin to mitigate stackingYesSame
Operating platformAndroid and iOS platformsSame
UI StandardsAndroid and iOS standardsSame

VII. Performance Data

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.

Clinical Evidence

Companion Medical has demonstrated the InPen dose calculator is appropriate for its intended use through the use of hazard analysis according ISO 14971. Verification and validation of the

{6}------------------------------------------------

user interface was completed through a comprehensive usability evaluation. In the summative evaluation, patients with sufficient diabetes knowledge completed self-training and then completed a series of critical tasks using the InPen dose calculator. The summative evaluation demonstrated that after self-training, patients were able to use the InPen dose calculator without making critical errors that could lead to a hazard. No new use-related hazards were identified during the study. The InPen dose calculator satisfies all functional performance and safety requirements, meets its intended use, and is safe for the intended user population. Substantial equivalence was based in part on the evaluation.

The dose calculator uses the standard approach using healthcare provider specified insulin-tocarbohydrate ratio and insulin sensitivity factors for making calculations. In addition, the calculator includes a consideration for insulin on-board based on the published study by Mudaliar et al (1999) for the duration of insulin action.

VIII. CONCLUSIONS

The subject device is substantially equivalent to the predicate devices, as demonstrated by performance data. It has the same intended use/indications for use, and substantially equivalent technological characteristics and principles of operation.

§ 868.1890 Predictive pulmonary-function value calculator.

(a)
Identification. A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.(b)
Classification. Class II (performance standards).