K Number
K190476
Date Cleared
2019-03-15

(16 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vevo MD Imaging System is a general purpose imaging system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body.

Specific clinical applications and exam types include:

Abdominal
Dermatological
Musculoskeletal (conventional)
Musculoskeletal (superficial)
Neonatal Cephalic
Ophthalmic
Pediatric
Peripheral vessel
Small Organ (breast, thyroid, testicles, prostate)

Device Description

The Vevo MD Imaging System is a high frequency general purpose, software controlled, diagnostic imaging system used to acquire ad display high-resolution, real-time ultrasound data in 2D, Color Doppler, Power Doppler, PW Doppler and M-Mode. The Vevo MD Imaging System is comprised of transducers responsible for ultrasound signal generation and recording, and a main unit that controls the transducers, processes the acoustic data, and processes and displays images.

AI/ML Overview

This document, K190476, details the 510(k) premarket notification for the FUJIFILM SonoSite Vevo MD Imaging System.

Here's an analysis based on your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

This 510(k) summary primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting specific acceptance criteria and performance metrics from device testing. The document states that the device was evaluated for safety and effectiveness against its predicate. The "acceptance criteria" for a 510(k) typically involve demonstrating that the new device is as safe and effective as a legally marketed predicate device.

The study presented does not specify quantitative acceptance criteria (e.g., specific accuracy, sensitivity, specificity thresholds) or report numerical device performance against such metrics. Instead, it relies on demonstrating equivalence through:

  • Intended Use Alignment: Both the new device and the primary predicate (FUJIFILM SonoSite Vevo MD Imaging System K160674) share the same intended use.
  • Technological Characteristics Comparison: The document provides a table comparing features, modes of operation, acoustic output, DICOM capabilities, and channels with the predicate devices. The new device and primary predicate are highly similar, particularly for "Track 3" devices.
  • Compliance with Voluntary Standards: The device is designed to comply with several recognized standards related to biological evaluation, electrical safety, electromagnetic disturbances, ultrasonic medical diagnostic equipment, field characterization, software life cycle, and risk management. This compliance forms a key part of demonstrating safety.

Summary of "Acceptance Criteria" (Implicit) and Reported Performance:

CategoryImplicit Acceptance Criteria (for 510(k) equivalence)Reported Device Performance (as demonstrated for equivalence)
Intended UseIdentical or highly similar to predicate device(s).The Vevo MD Imaging System (this submission) has the same intended use as the predicate Vevo MD Imaging System (K160674): "Diagnostic ultrasound imaging or fluid flow analysis of the human body." Specific clinical applications for both include Abdominal, Dermatological, Musculoskeletal (conventional/superficial), Neonatal Cephalic, Ophthalmic, Pediatric, Peripheral vessel, and Small Organ.
Technological CharacteristicsNo significant differences in features or fundamental scientific technology compared to predicate that would raise new questions of safety or effectiveness.The Vevo MD Imaging System (this submission) is a "Track 3" device and employs the same fundamental scientific technology as its primary predicate (Vevo MD K160674). Key common features include Linear Array transducer type, 15-49MHz transducer frequency, MI/TI output display, 2D Mode, Color Doppler, M-Mode, Power Doppler, PW Doppler operation modes, DICOM 3.0 support, manual IMT measurement, 64 digital transmit and receive channels. The comparison shows substantial similarity.
Safety StandardsCompliance with relevant mandatory and voluntary medical device safety standards.Evaluated for electrical, thermal, mechanical, and EMC safety. Cleaning/disinfection and biocompatibility were also evaluated. Found to conform to applicable mandatory medical device safety standards. Specifically states compliance with AAMI/ANSI/ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62359, IEC 62304, ISO 14971, and NEMA UD 2. All patient contact materials are biocompatible and are either used in the predicate or meet ISO 10993.
Clinical PerformanceNo new clinical concerns or requirement for clinical studies if equivalence demonstrated through non-clinical means.Clinical studies were not required to support the determination of substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states: "The Vevo MD Imaging System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence." This means there was no clinical "test set" in the traditional sense of patient data used to evaluate performance metrics like sensitivity or specificity. The evaluation was primarily non-clinical, comparing technical specifications and adherence to standards against a predicate device. Therefore, no information on sample size or data provenance for a clinical test set is provided.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

As no clinical study was conducted for performance evaluation in this 510(k) submission, there was no "ground truth" to be established by experts for a test set of patient data.

4. Adjudication Method for the Test Set

Not applicable, as no clinical test set for performance evaluation was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an imaging system, not an AI-based diagnostic aid that would involve human readers improving with AI assistance. Clinical studies were not performed for this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is an imaging system, not a standalone algorithm. Clinical studies were not performed for this submission.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable, as no clinical study for performance evaluation was conducted where ground truth would be established. The "ground truth" for this submission revolves around the technical specifications and safety profile of the device aligning with regulated standards and the predicate device.

8. The sample size for the training set

Not applicable. As this 510(k) is for an ultrasound imaging system and not for an AI/ML-driven device or an algorithm requiring a training set, no information about a training set size is provided.

9. How the ground truth for the training set was established

Not applicable, as there was no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

FUJIFILM SonoSite, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW BUFFALO MN 55313

March 15, 2019

Re: K190476

Trade/Device Name: FUJIFILM SonoSite Vevo MD Imaging System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: February 25, 2019 Received: February 27, 2019

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara

For

Thalia Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190476

Device Name FUJIFILM SonoSite Vevo MD Imaging System

Indications for Use (Describe)

The Vevo MD Imaging System is a general purpose imaging system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body.

Specific clinical applications and exam types include:

Abdominal Dermatological Musculoskeletal (conventional) Musculoskeletal (superficial) Neonatal Cephalic Ophthalmic Pediatric Peripheral vessel Small Organ (breast, thyroid, testicles, prostate)

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

Table 1.3.1: Diagnostic Ultrasound Indications for Use Form - FUJIFILM SonoSite Vevo MD Imaging System

System:Vevo MD Imaging System
Transducer:UHF22, UHF48, UHF70
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation
Clinical Application2DMPWDCWDColorDoppler(CD)PowerDoppler(CPD)Combined(Spec.)Other(Spec.)
OphthalmicN1,2,3
Fetal
AbdominalPPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2,3
Intra-operative (Abdominalorgans and vascular)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Small Organ (breast,thyroid, testicles, prostate)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Neonatal CephalicNNNNN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel. (Convent.)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Musculo-skel. (Superfic.)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Intra-luminal
Other (spec.)
Cardiac Adult
Cardiac Pediatric
Trans-esophageal (card.)
Other (spec.)
Peripheral vesselPPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Other (Dermatology)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2

N= New Indicated Imaging Examination P=Previously Indicated Imaging Examination

Additional Comments/Other (Spec.):

I: The Vevo MD Imaging System uses ultra high frequency (UHF) series transducers that limit the imaging of abdominal organs or non-superficial musculoskeletal structure may be limited to neonatal and small pediative is limited to the maximum imaging depth of the transducer as indicated in the specifications for each transducer.

2: Includes imaging to assist in the placement of needler or other anatomical structures and imaging guidance for peripheral nerve block procedures.

3: Abdominal and Ophthalmic clinical applications are indications for use for the UHF22 transducer only

Page 1 of 4

VOL 002 Section 1 General Information Section 003 Indications for Use

{4}------------------------------------------------

System:Vevo MD Imaging System
Transducer:UHF22
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of OperationCombined(Spec.)Other(Spec.)
B(2D)MPWDCWDColorDoppler(CD)PowerDoppler(CPD)
OphthalmicN1,2,3
Fetal
AbdominalPPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2,3
Intra-operative (Abdominalorgans and vascular)1,2,3
Intra-operative (Neuro.)
Laparoscopic
PediatricPPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Small Organ (breast,thyroid, testicles, prostate)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Neonatal CephalicNNNNN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel. (Convent.)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Musculo-skel. (Superfic.)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Intra-luminal
Other (spec.)
Cardiac Adult
Cardiac Pediatric
Trans-esophageal (card.)
Other (spec.)
Peripheral vesselPPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Other (Dermatology)NNNNN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2

N= New Indicated Imaging Examination P=Previously Indicated Imaging Examination

Additional Comments/Other (Spec.):

I: The Vevo MD Imaging System uses ultra high frequency (UHF) series transducers that limit the imaging depth. For this reason imaging of abdominal organs or non-superficial musculoskeletal structure may be limited to neonatal and small imaging is limited to the maximum imaging depth of the transducer as indicated in the specifications for each transducer.

  1. Includes imaging to assist in the placement of needles and cather anatomical structures and imaging guidance for peripheral nerve block procedures.

3: Abdominal and Ophthalmic clinical applications are indications for use for the UHF22 transducer only

{5}------------------------------------------------

System:Vevo MD Imaging System
Transducer:UHF48
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of OperationOther(Spec.)
B(2D)MPWDCWDColorDoppler(CD)PowerDoppler(CPD)Combined(Spec.)
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominalorgans and vascular)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Small Organ (breast,thyroid, testicles, prostate)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Neonatal CephalicNNNNN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel. (Convent.)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Musculo-skel. (Superfic.)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Intra-luminal
Other (spec.)
Cardiac Adult
Cardiac Pediatric
Trans-esophageal (card.)
Other (spec.)
Peripheral vesselPPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2
Other (Dermatology)PPNPN2D+CD, 2D+M, 2D+CPD,2D+PWD1,2

N= New Indicated Imaging Examination P=Previously Indicated Imaging Examination

Additional Comments/Other (Spec.):

19 101 820 800000000 1 100 16

1: The Vevo MD Imaging System uses ultra high frequency (UHF) series transducers that limit the imaging of abdominal organs or non-superficial musculoskelal structure may be limited to neonatal and small pediatric patients and all imaging is fimited to the maximum imaging depth of the transducer as indicated in the specifications for each transducer.

2: Includes imaging to assist in the placement of needler or other anatomical structures and imaging guidance for peripheral nerve block procedures.

FINAL TRANCE

{6}------------------------------------------------

System:Vevo MD Imaging System
Transducer:UHF70
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical Application
B (2D)MPWDCWDColor Doppler (CD)Power Doppler (CPD)Combined (Spec.)Other (Spec.)
Ophthalmic
Fetal
Abdominal
Intra-operative (Abdominal organs and vascular)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPNPN2D+CD, 2D+M, 2D+CPD, 2D+PWD1,2
Small Organ (breast, thyroid, testicles, prostate)PPNPN2D+CD, 2D+M, 2D+CPD, 2D+PWD1,2
Neonatal CephalicNNNNN2D+CD, 2D+M, 2D+CPD, 2D+PWD1,2
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skel. (Convent.)PPNPN2D+CD, 2D+M, 2D+CPD, 2D+PWD1,2
Musculo-skel. (Superfic.)PPNPN2D+CD, 2D+M, 2D+CPD, 2D+PWD1,2
Intra-luminal
Other (spec.)
Cardiac Adult
Cardiac Pediatric
Trans-esophageal (card.)
Other (spec.)
Peripheral vesselPPNPN2D+CD, 2D+M, 2D+CPD, 2D+PWD1,2
Other (Dermatology)PPNPN2D+CD, 2D+M, 2D+CPD, 2D+PWD1,2

N= New Indicated Imaging Examination P=Previously Indicated Imaging Examination

Additional Comments/Other (Spec.):

I The Vevo MD Imaging System uses ultra high frequency (UHF) series transducers that limit the imaging depth. For this reason imaging of abdominal organs or non-superficial musculoskeletal structure may be limited to neonatal and small pediatric patients and all imaging is limited to the maximum imaging depth of the transducer as indicated in the specifications for each transducer.

2: Includes imaging to assist in the placement of needles and cather anatomical structures and imaging guidance for peripheral nerve block procedures.

{7}------------------------------------------------

SonoSite

510(K) Summary-K190476

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.

1) Submitter's name, address, telephone number, contact person;

FUJIFILM SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904

Corresponding Official:Anoush Frankian
Sr. Manager, Regulatory Affairs
E-mail:anoush.frankian@fujifilm.com
Telephone:(425) 951-6824
Facsimile:(425) 491-8356
Date prepared:February 11, 2019

2) Name of the device. including the trade or proprietary name if applicable. the common or usual name, and the classification name, if known:

Common/ Usual Name

Diagnostic Ultrasound System with Accessories

Proprietary Name

FUJIFILM SonoSite Vevo MD Imaging System

Classification Names

NameFR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System892.1550IYN
Diagnostic Ultrasound Transducer892.1570ITX
Ultrasonic Pulsed Echo Imaging System892.1560IYO

3) Identification of the predicate or legally marketed device:

FUJIFILM SonoSite Vevo MD Imaging System (K160674) (Primary Predicate) SonoSite X-Porte ultrasound system (K171437) (Reference Predicate)

4) Device Description:

The Vevo MD Imaging System is a high frequency general purpose, software controlled, diagnostic imaging system used to acquire ad display high-resolution, real-time ultrasound data in 2D, Color Doppler, Power Doppler, PW Doppler and M-Mode. The Vevo MD Imaging System is comprised of

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a red square with a white triangle cut out of it, which is placed above the "I" in Fujifilm. Below the logo, in a smaller font, is the text "Value from Innovation."

Image /page/8/Picture/1 description: The image shows the word "SonoSite" in a bold, sans-serif font. The letters are all black and evenly spaced. The dot above the "i" in "Site" is a perfect circle.

transducers responsible for ultrasound signal generation and recording, and a main unit that controls the transducers, processes the acoustic data, and processes and displays images.

5) Intended Use:

The Vevo MD Imaging System is a general purpose imaging system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body.

Specific clinical applications and exam types include:

Abdominal Dermatological Musculoskeletal (conventional) Musculoskeletal (superficial) Neonatal Cephalic Ophthalmic

Pediatric

Peripheral vessel Small Organ (breast, thyroid, testicles, prostate)

6) Technological Characteristics:

The Vevo MD Imaging System subject of this submission and its primary predicate Vevo MD Imaging System (K160674) and reference predicate X-Porte (K171437) are all Track 3 devices that employ the same fundamental scientific technology. A comparison table is provided below.

FeatureSonoSite VevoMDUltrasound System(this submission)SonoSiteVevoMDUltrasoundSystem(K160674)SonoSite X-PorteUltrasound System(K171437)
Intended UseDiagnosticultrasound imagingor fluid flow analysisof the human bodyDiagnosticultrasoundimaging or fluidflow analysis ofthe human bodyDiagnostic ultrasoundimaging or fluid flowanalysis of the humanbody
TransducerTypesAvailableLinear ArrayLinear ArrayLinear ArrayCurved Linear ArrayIntracavitaryPhased Array
TransducerCenterFrequency15-49MHz15-49MHz1.0 - 15.0 MHz
FeatureSonoSite VevoMDUltrasound System(this submission)SonoSiteVevoMDUltrasoundSystem(K160674)SonoSite X-PorteUltrasound System(K171437)
AcousticOutputDisplay &FDA LimitsMI Output DisplayTI Output DisplayMI OutputDisplayTI Output DisplayDisplay Feature forHigher OutputsMI Output DisplayTI Output Display
Modes ofOperation2D Mode, ColorDoppler, M-Mode,Power Doppler, PWDopplerB-mode (2-DGrayscaleImagingColor DopplerCombinationModesM-modeB-mode GrayscaleImagingTissue HarmonicImagingM-modeColor M-ModeColor Power DopplerZoomCombination ModesPulsed Wave (PW)DopplerContinuous Wave(CW) DopplerSonoHD2 NoiseReductionSonoMB/MBe ImageCompoundingSteered CW DopplerVelocity ColorDopplerTissue DopplerImaging (TDI)
FeatureSonoSite VevoMDUltrasound System(this submission)SonoSiteVevoMDUltrasoundSystem(K160674)SonoSite X-PorteUltrasound System(K171437)
DICOMDICOM 3.0 Store,Print, ModalityWorklist, PerformProcedure Step(PPS), StorageCommitmentDICOM 3.0 Store,Print, ModalityWorklist, PerformProcedure Step(PPS), StorageCommitmentDICOM 3.0 Store,Print, ModalityWorklist, PerformProcedure Step (PPS),Storage Commitment
IMTMeasurementManualMeasurementavailable on theultrasound systemitself.ManualMeasurementavailable on theultrasound systemitself.Not available
#TransmitChannels64 digital channels64 digitalchannels128 digital channels
#ReceiveChannels64 digital channels64 digitalchannels64 digital channels(128 digital channelsusing SyntheticAperture)
510(k) TrackTrack 3Track 3Track 3

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a small red square above the letter "I". Below the logo, the words "Value from Innovation" are written in a smaller, black font. The logo is simple and modern.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the Fujifilm logo. The logo consists of the word "FUJIFILM" in bold, black letters. The "I" in Fujifilm has a red square on the top right corner. Below the word "FUJIFILM" is the tagline "Value from Innovation".

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a red accent on the "I" in Fujifilm. Below the logo is the text "Value from Innovation" in black.

7) Determination of Substantial Equivalence:

Summary of Non-Clinical Tests:

The Vevo MD Imaging System has been evaluated for electrical, thermal, mechanical and EMC safety. Additionally, cleaning/disinfection, biocompatibility, and acoustic output have been evaluated, and the device has been found to conform to applicable mandatory medical device safety standards. Assurance of quality was established by employing the following elements of product development: Design Phase Reviews, Risk Assessment, Requirements Development, System and Software Verification, Hardware Verification, Safety Compliance Verification, Clinical Validation. All patient contact materials are biocompatible and are materials that are already used in the predicate device or meet 10993.

The Vevo MD Imaging System is designed to comply with the following voluntary standards.

Reference No.RecognitionNo.Title
AAMI/ANSI/ISO10993-12-220Biological evaluation of medical devices - Part 1:Evaluation and testing within a risk managementprocess
IEC 60601-119-4Medical electrical equipment - Part 1: Generalrequirements for basic safety and essentialperformance
IEC 60601-1-219-8Medical Electrical Equipment - Part 1-2: GeneralRequirements For Basic Safety And EssentialPerformance - Collateral Standard: ElectromagneticDisturbances - Requirements And Tests (Edition 4)
IEC 60601-2-3712-293Particular Requirements for the basic safety andessential performance of ultrasonic medicaldiagnostic and monitoring equipment (Edition 2.1)
IEC 6235912-316Ultrasonic - Field Characterization - Test Methods ForThe Determination Of Thermal And MechanicalIndices Related To Medical Diagnostic UltrasonicFields
IEC 6230413-79Medical Device Software - Software Life CycleProcesses
ISO 149715-40Medical devices - Application of risk management tomedical devices
NEMA UD 212-105Acoustic Output Measurement Standard forDiagnostic Ultrasound Equipment

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a red square on top of the letter 'I'. Below the logo is the text 'Value from Innovation' in black font. The logo is simple and modern.

Image /page/12/Picture/1 description: The image shows the word "SonoSite" in a bold, sans-serif font. The letters are black against a white background. The word is horizontally oriented and centered in the image.

Summary of Clinical Tests:

The Vevo MD Imaging System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence.

8) Conclusion:

Intended uses and other key features are consistent with traditional clinical practice and FDA guidance. The Vevo MD device and predicate both conform to applicable electromedical device safety standards with compliance verified through independent evaluation. The Vevo MD device and predicate both meet FDA requirements for Track 3 devices, share indications for use, have biosafety equivalence and are manufactured using the same ISO 13485 quality system. FUJIFILM SonoSite, Inc. believes that the Vevo MD Ultrasound System is substantially equivalent with regard to safety and effectiveness to the predicate device(s).

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.