(121 days)
The Delivery Catheter is indicated for the introduction of various types of catheters and pacing or defibrillator leads.
The Delivery Catheter is a single-use percutaneous catheter intended to introduce various types of catheters and pacing or defibrillator leads. The Delivery Catheter is packaged with a dilator for introduction into the vasculature. Proximally, the Delivery Catheter is equipped with a hemostatic valve, and the distal soft, rounded, radiopaque tip facilitates imaging under fluoroscopy. The Delivery Catheter is designed to be slittable, thereby allowing its removal after device placement. A variety of curves and lengths are available to accommodate various anatomies and different lead locations. The Delivery Catheter has an inner diameter of 6.5F, an outer diameter of 8F, and the dilator is compatible with a 0.035" guidewire
This document is a 510(k) summary for a medical device called "Delivery Catheter". This type of document is typically used for medical devices that are being cleared for marketing by demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness de novo. Therefore, it focuses on comparing the new device to existing ones and confirming that it performs as intended, rather than establishing new acceptance criteria and conducting studies to meet those.
Based on the provided text, here's what can be extracted and what cannot:
1. A table of acceptance criteria and the reported device performance:
The document does not explicitly state quantitative "acceptance criteria" in the format typically seen for a new medical device's performance study. Instead, it describes various tests performed and states that the device "passed" these tests or "meets intended product specifications." The "reported device performance" is a qualitative statement of success in meeting these testing criteria.
| Test / Performance Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Biocompatibility | Compliance with ISO 10993-1 | Passed |
| Bench testing (dimensional evaluation, tensile testing, etc.) | Meet intended product specifications | Confirmed to meet intended product specifications |
| Simulated Use testing | Perform as intended in a simulated environment | Confirmed to perform as intended in a simulated environment |
| In Vivo Testing | Perform as intended in an in vivo model | Confirmed to perform as intended in an in vivo model |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The document does not provide any specific sample sizes for the test sets used in the biocompatibility, bench testing, simulated use testing, or in vivo testing.
The data provenance (country of origin, retrospective/prospective) is not mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable or not provided. For this type of submission (510(k) for a delivery catheter), the "ground truth" isn't typically established by expert consensus on diagnoses or interpretations of images, but rather by objective physical and biological measurements during performance testing. The document does not mention any experts involved in reviewing test results in a "ground truth" capacity.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable or not provided. Adjudication methods are typically used in clinical studies where interpretation of data (e.g., medical images, patient outcomes) by multiple experts is required to establish a gold standard. The tests mentioned (biocompatibility, bench, simulated use, in vivo) involve objective measurements or observations of the device's physical and biological behavior, not expert interpretation.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned and is not applicable to this type of device submission. MRMC studies are typically for AI/imaging devices where the performance of human readers, with and without AI assistance, is evaluated using a set of cases. This device is a physical delivery catheter.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
A standalone algorithm performance study is not applicable as this is a physical medical device, not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the performance tests mentioned (biocompatibility, bench, simulated use, in vivo), the "ground truth" would be established by:
- Biocompatibility: Standardized tests for cytotoxity, sensitization, irritation, material-mediated pyrogenicity, according to ISO 10993-1. The "ground truth" is adherence to the scientific principles and pass/fail criteria of these established international standards.
- Bench testing: Engineering specifications and objective measurements (e.g., dimensional tolerances, tensile strength, material properties). The "ground truth" is whether the measured values fall within the predefined acceptable ranges of the product specifications.
- Simulated Use testing: Observation of the device's functional performance in a simulated environment (e.g., successful navigation, deployment, or removal without failure). The "ground truth" is direct observation of successful operation per design intent.
- In Vivo testing: Observation of the device's functional performance and interaction with biological tissue in an animal model (e.g., successful delivery, absence of adverse tissue reaction beyond expected). The "ground truth" is direct observation of satisfactory performance in a living system.
8. The sample size for the training set:
This information is not applicable and not provided. This is not an AI/ML device that requires a training set.
9. How the ground truth for the training set was established:
This information is not applicable and not provided. This is not an AI/ML device that requires a training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 28, 2019 CenterPoint Systems Marybeth Gamber Vice President, Regulatory & Quality 3338 Parkway Blvd West Valley City, Utah 84119
Re: K190475
Trade/Device Name: Delivery Catheter SSPC1, Delivery Catheter SSPC2, Delivery Catheter SSPC3, Delivery Catheter SSPC4 Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: June 4, 2019 Received: June 5, 2019
Dear Marybeth Gamber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jessica Paulsen Implantable Electrophysiology Devices Team Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K190475
Device Name Delivery Catheter
Indications for Use (Describe)
The Delivery Catheter is indicated for the introduction of various types of catheters and pacing or defibrillator leads.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
5 510(K) SUMMARY
5.1 Submitter
| Name | CenterPoint Systems |
|---|---|
| Address | 3338 Parkway BlvdWest Valley City UT |
| Phone | 877-848-0828 |
| Contact Person: | Marybeth Gamber, Vice President Regulatory Affairs & Quality Assurance |
| Date Prepared: | February 26, 2019 |
5.2 Device
| Name of Device: | Delivery Catheter |
|---|---|
| Common or Usual Name | Delivery Catheter |
| Classification Name: | Catheter, Percutaneous |
| Regulatory Class: | Class II per 21 CFR 870.1250 |
| Product Code: | DQY |
5.3 Predicate Device
Predicate Name and 510(k) Number: Medtronic C315 Delivery Catheter, K101885
This predicate has not been subject to a design-related recall.
No reference predicates were used in this submission.
5.4 Device Description
The Delivery Catheter is a single-use percutaneous catheter intended to introduce various types of catheters and pacing or defibrillator leads.
The Delivery Catheter is packaged with a dilator for introduction into the vasculature. Proximally, the Delivery Catheter is equipped with a hemostatic valve, and the distal soft, rounded, radiopaque tip facilitates imaging under fluoroscopy. The Delivery Catheter is designed to be slittable, thereby allowing its removal after device placement. A variety of curves and lengths are available to accommodate various anatomies and different lead locations. The Delivery Catheter has an inner diameter of 6.5F, an outer diameter of 8F, and the dilator is compatible with a 0.035" guidewire
510(k) Summary
{4}------------------------------------------------
Indications for Use ર.5
The Delivery Catheter is indicated for the introduction of various types of catheters and pacing or defibrillator leads.
Comparison of Technological Characteristics with the Predicate Device 5.6
The Proposed Device and Predicate Device are similar in indications for use, intended use, technological characteristics, and principles of operation.
The differences between the Proposed Device and the Predicate Device are minor, thus it was concluded that the Proposed Device is substantially equivalent to the Predicate Device. In accordance with 21CFR807.92(a)(6) a summary of how the technological characteristics of the Proposed Device compares to the Predicate Device is provided below.
| Feature | Delivery Catheter(proposed device) | Medtronic C315 DeliveryCatheter (K101885) |
|---|---|---|
| Intended Use | Percutaneous catheter for the deliveryof catheters and leads | Same |
| Indications for Use | The Delivery Catheter is indicated forthe introduction of various types ofcatheters and pacing or defibrillatorleads. | Same |
| Device Class | Class II | Same |
| Product Code | DQY, 21 CFR 870.1250 | Same |
| Prescription device | Yes | Same |
| Catheter Type | Percutaneous Catheter | Same |
| Guidewire compatibility | 0.035" | Same |
| Catheter Outer Diameter | 8.0F | 7.0F Substantially equivalent |
| Catheter Inner Diameter | 6.5F | 5.4F Substantially equivalent |
| Catheter Length | 30 - 40cm | 20 – 43cm Substantiallyequivalent |
| Hydrophilic Liner | Yes | Same |
| Radiopaque Catheter | Yes | Same |
| Distal Tip | ||
| Hemostasis valve | Yes | Same |
| Components Provided | Catheter, Dilator | Same |
| Multiple Distal End | Yes | Same |
| Shapes Available | ||
| Sterility | Provided Sterile | Same |
| Number of uses | Single patient use | Same |
510(k) Summary
{5}------------------------------------------------
| Feature | Delivery Catheter(proposed device) | Medtronic C315 DeliveryCatheter (K101885) |
|---|---|---|
| Principals of Operation | After venous access is gained, thecatheter and dilator are advanced over aguidewire to the desired location. Thedilator is removed and a catheter or leadis placed through the Delivery Catheter.The Delivery Catheter may be removedby slitting. | Same |
The Delivery Catheter is used for the same intended use in the same anatomical location using the same principles of operation as the predicate device. Therefore, the Delivery Catheter can be considered substantially equivalent to the predicate device.
5.7 Performance Data
All necessary performance testing has been conducted on the Delivery Catheter to assure substantial equivalence to the predicate devices and to demonstrate the device performs as intended. All testing was performed on test units representative of finished devices.
The device passed the following tests, which were conducted in accordance with noted standards:
| Test | Consensus Standard/FDA Guidance/Description |
|---|---|
| Biocompatibility | FDA Final Guidance Document, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (June 2016) |
| Bench testing, including dimensional evaluation, tensile testing, etc | Confirm that the device meets intended product specifications |
| Simulated Use testing | Confirm that the device will perform as intended in a simulated environment |
| In Vivo Testing | Confirm the device will perform as intended in an in vivo model. |
{6}------------------------------------------------
5.8 Conclusions
Upon reviewing the information provided in this submission and comparing the intended use, principle of operation and overall technological characteristics, the Delivery Catheter is substantially equivalent to existing legally marketed devices.
510(k) Summary
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).