K Number
K190468
Device Name
Nihon Kohden SVM-7200 Series Vital Signs Monitor
Date Cleared
2019-05-28

(91 days)

Product Code
Regulation Number
870.2300
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The SVM-7200 series Vital Signs Monitor is intended for monitoring of physiologic parameters, including non-invasive blood pressure (systolic, and mean arterial pressure), pulse rate, non-invasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and temperature, and neonatal patients in hospital environments when used by clinical physicians or appropriate medical staff under the direction of physicians.
Device Description
The SVM-7200 series Vital Signs Monitor is intended to be used by clinicians, doctors, nurses, and medically qualified personnel for measuring noninvasive blood pressure (NIBP), noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), body temperature (TEMP) of one patient at a time. The monitors can be used in the specialized health care environment such as hospitals, clinics, special medical facilities, independent surgery centers, multitherapy facilities and wards. Adults, children and neonates can be monitored using this device. The SVM-7200 series monitors have an eight-inch TFT color display and two modes: spot check and continuous mode. The monitor can operate with either a wired or wireless LAN connection.
More Information

No
The summary describes a standard vital signs monitor and does not mention any AI or ML capabilities, image processing, or data training/testing sets typically associated with such technologies.

No
The device is described as a "Vital Signs Monitor" which is intended for "monitoring of physiologic parameters." It measures various vital signs but does not state any therapeutic function or intervention.

No

The device is a vital signs monitor that measures physiological parameters like blood pressure, pulse rate, SpO2, and temperature. While this information is used by clinicians, the device itself is for monitoring and does not provide a diagnosis.

No

The device description explicitly mentions an "eight-inch TFT color display" and the need for testing to compliance standards for "electrical and electromagnetic safety," indicating the presence of hardware components beyond just software.

Based on the provided information, the SVM-7200 series Vital Signs Monitor is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • IVD devices are used to examine specimens derived from the human body (like blood, urine, or tissue) to provide information for diagnosis, monitoring, or screening. The SVM-7200 series Vital Signs Monitor measures physiological parameters directly from the patient's body (blood pressure, pulse rate, oxygen saturation, temperature) using non-invasive methods.
  • The intended use and device description clearly state that it monitors physiological parameters of patients. There is no mention of analyzing biological specimens.

Therefore, the SVM-7200 series Vital Signs Monitor falls under the category of a patient monitoring device, not an IVD.

N/A

Intended Use / Indications for Use

The SVM-7200 series Vital Signs Monitor is intended for monitoring and alarming of physiologic parameters, including non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, non-invasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and temperature, on adult, pediatric, and neonatal patients in hospital environments when used by clinical physicians or appropriate medical staff under the direction of physicians.

Product codes

MWI, DXN, DQA, FLL

Device Description

The SVM-7200 series Vital Signs Monitor is intended to be used by clinicians, doctors, nurses, and medically qualified personnel for measuring noninvasive blood pressure (NIBP), noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), body temperature (TEMP) of one patient at a time. The monitors can be used in the specialized health care environment such as hospitals, clinics, special medical facilities, independent surgery centers, multitherapy facilities and wards. Adults, children and neonates can be monitored using this device.

The SVM-7200 series monitors have an eight-inch TFT color display and two modes: spot check and continuous mode. The monitor can operate with either a wired or wireless LAN connection. The models included are:

Model NameScreen Size (inch)Parameters
SVM-72308Masimo SpO2, NIBP, Temperature
SVM-72508Nellcor SpO2, NIBP, Temperature
SVM-72608Nihon Kohden SpO2, NIBP, Temperature

The thermometer options include the commercially available Covidien FILAC 3000 Thermometer or Exergen TAT-5000S Temporal Scanner.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

adult, pediatric, and neonatal patients

Intended User / Care Setting

clinical physicians or appropriate medical staff under the direction of physicians.
hospital environments

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance testing for the Nihon Kohden SVM-7200 series Vital Signs Monitor includes software verification and validation test which encompassed software unit testing, integration level testing, and system level testing. A system test was also performed per the software requirements specification. A summative usability validation has been conducted with the subject device in a simulated environment. Testing to compliance standards for electrical and electromagnetic safety was also performed. Traceability has been documented between all system specifications to validation test results. Test results were not provided.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K171751

Reference Device(s)

K080342

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

May 28, 2019

Nihon Kohden Corporation % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059

Re: K190468

Trade/Device Name: Nihon Kohden SVM-7200 series Vital Signs Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI, DXN, DQA, FLL Dated: April 25, 2019 Received: April 26, 2019

Dear Dave Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Matthew Hillebrenner
Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K190468

Device Name

Nihon Kohden SVM-7200 series Vital Signs Monitor

Indications for Use (Describe)

The SVM-7200 series Vital Signs Monitor is intended for monitoring of physiologic parameters, including non-invasive blood pressure (systolic, and mean arterial pressure), pulse rate, non-invasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and temperature, and neonatal patients in hospital environments when used by clinical physicians or appropriate medical staff under the direction of physicians.

Type of Use (Select one or both, as applicable)
XPrescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Image /page/3/Picture/1 description: The image contains the logo for NIHON KOHDEN. The logo consists of a blue square with a white abstract design on the left, followed by the company name "NIHON KOHDEN" in blue, sans-serif font. The text is bold and slightly blurred, giving it a soft appearance.

510(k) Summary K190468

| Sponsor: | Nihon Kohden Corporation
1-31-4 Nishiochiai, Shinjuku-Ku
Tokyo, Japan 161-8560 |
|----------|--------------------------------------------------------------------------------------|
| Contact: | Natalie J. Kennel
NJK & Associates |
| Address | 13721 Via Tres Vista
San Diego, CA 92129 |
| Phone | (858) 705-0350 |
| Fax | (858) 764-9739 |
| email | nkennel@njkconsulting.com |

  • May 22, 2019 Date Prepared:

DEVICE INFORMATION:

Proprietary Name:Nihon Kohden SVM-7200 series Vital Signs Monitor
Common Name:Vital Signs Monitor, Patient Monitor
Classification:Monitor, Physiological, Patient, (without arrhythmia detection or alarms) Class II
Product Codes:MWI, DXN, DQA, FLL
Regulations:21 CFR 870.2300
Predicate:Capsule Technologies SmartLinx Vitals Plus Patient Monitoring System (K171751)

4

Reference Device: Nihon Kohden Bedside Monitor, BSM-6000 series (K080342)

5

PRODUCT DESCRIPTION:

The SVM-7200 series Vital Signs Monitor is intended to be used by clinicians, doctors, nurses, and medically qualified personnel for measuring noninvasive blood pressure (NIBP), noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), body temperature (TEMP) of one patient at a time. The monitors can be used in the specialized health care environment such as hospitals, clinics, special medical facilities, independent surgery centers, multitherapy facilities and wards. Adults, children and neonates can be monitored using this device.

The SVM-7200 series monitors have an eight-inch TFT color display and two modes: spot check and continuous mode. The monitor can operate with either a wired or wireless LAN connection. The models included are:

| Model Name | Screen Size
(inch) | Parameters |
|------------|-----------------------|--------------------------------------|
| SVM-7230 | 8 | Masimo SpO2, NIBP, Temperature |
| SVM-7250 | 8 | Nellcor SpO2, NIBP, Temperature |
| SVM-7260 | 8 | Nihon Kohden SpO2, NIBP, Temperature |

The thermometer options include the commercially available Covidien FILAC 3000 Thermometer or Exergen TAT-5000S Temporal Scanner.

INDICATIONS FOR USE:

The SVM-7200 series Vital Signs Monitor is intended for monitoring and alarming of physiologic parameters, including non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, non-invasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and temperature, on adult, pediatric, and neonatal patients in hospital environments when used by clinical physicians or appropriate medical staff under the direction of physicians.

TECHNOLOGICAL CHARACTERISTICS - SUBSTANTIAL EQUIVALENCE:

Table 1 is a detailed comparison of the Nihon Kohden SVM-7200 Series Vital Signs Monitor and its predicate, SmartLinx Vitals Plus. The Nihon Kohden SVM-7200 Vital Signs Monitor is substantially equivalent to the SmartLinx Vitals Plus Patient Monitoring System, other than the minor differences described below. These differences do not raise questions regarding safety or efficacy.

6

SmartLinx Vitals PlusSVM-7200 Series Vital Signs Monitor
CharacteristicPredicate Device (K171751)Subject Device
Indications for
UseThe SmartLinx Vitals Plus Patient
Monitoring System is intended for
monitoring and alarming of physiologic
parameters, including non-invasive blood
pressure (systolic, diastolic, and mean
arterial pressure), pulse rate, functional
arterial oxygen saturation (SpO2), and
temperature, on adult, pediatric, and
neonatal patients in hospital environments
when used by clinical physicians or
appropriate medical staff under the
direction of physicians.The SVM-7200 series is intended for monitoring
and alarming of physiologic parameters,
including non-invasive blood pressure (systolic,
diastolic, and mean arterial pressure), pulse rate,
non-invasive functional oxygen saturation of
arteriolar hemoglobin (SpO2), and temperature,
on adult, pediatric, and neonatal patients in
hospital environments when used by clinical
physicians or appropriate medical staff under the
direction of physicians.
NIBP
MeasurementOscillometric design for Spot Check and
Intervals Mode measurements.Oscillometric design for Spot Check and
Intervals Mode measurements.
SpO2 MonitoringMasimo SpO2 for Spot Check
measurements and Continuous
Monitoring with the following features:
  1. Pulse tone pitch varying with oxygen
    saturation
  2. SpO2 alarm delay
  3. SpO2 sensor off alarm | Masimo SpO2, Nellcor SpO2, or Nihon Kohden
    SpO2 for Spot Check measurements and
    Continuous Monitoring with the following
    features:
  4. Pulse tone pitch varying with oxygen
    saturation
  5. SpO2 alarm delay
  6. SpO2 sensor off alarm |
    | Temperature
    Measurement | Two different thermometry options:
  7. Covidien FILAC 3000 thermometer
  8. Exergen TAT- 5000S temporal scanner | Two different thermometry options:
  9. Covidien FILAC 3000 thermometer
  10. Exergen TAT- 5000S temporal scanner |
    | Alarms | Configuration, annunciation, and
    acknowledgement of physiological and
    technical alarms for (Sys, Dia, MAP,
    Pulse Rate, SpO2, and Temp).
    Alarms are visibly annunciated and
    acknowledged on the screen.
    Alarms are audibly annunciated via
    speaker. | Configuration, annunciation, and
    acknowledgement of physiological and technical
    alarms for (Sys, Dia, MAP, Pulse Rate, SpO2,
    and Temp).
    Alarms are visibly annunciated and
    acknowledged on the screen.
    Alarms are audibly annunciated via speaker. |
    | Energy Source | Battery type: Lithium-Ion
    Power Supply: 100-240V AC, 50-60 Hz | Battery type: Lithium-Ion
    Power Supply: 100-240V AC, 50-60 Hz |

Table 1 Comparison of the Nihon Kohden SVM-7200 Series Vital Signs Monitor to its Predicate

7

These differences between the subject device and predicate are describe below:

  1. NIBP Measurement Ranges: The SVM-7200 has a higher end cut-off for systolic, diastolic and mean arterial pressure ranges for adults, pediatrics, and neonates as compared to the predicate. The subject device complies with the NIBP standard IEC 80601-2-30.

  2. SpO2 Monitoring: The SVM-7200 offers additional SpO2 probe options as compared to the predicate which only offers the Masimo SpO2 probe. The SVM-7200 can be used with its own Nihon Kohden SpO2 and Nellcor SpO2 pulse oximeter probes. All three SpO2 probe options were tested with the subject device which complies with the SpO2 standard IEC 80601-2-61.

  3. Operating Conditions: The SVM-7200 has a lower minimum operating temperature of 5°C as compared to 10°C of the predicate and reference devices. Also, the subject device has a lower maximum operating humidity of 85%, while the predicate and the reference device have a maximum humidity of 95%. This very minor difference should not affect the general functionality of the device and has no effect on safety or effectiveness. The subject device has been tested and complies to IEC 60601-1.

  4. Storage Conditions: The SVM-7200 has a higher maximum storage temperature of 65°C, as compared to 50°C of the predicate. Also, the subject device has a higher maximum storage humidity of 95% while the predicate has a maximum humidity of 90%. This very minor difference should not affect the general functionality of the device. The subject device has been tested and complies to IEC 60601-1.

Since some of the features are not exactly the same between the subject device and the predicate, a reference device, the Nihon Kohden Bedside Monitor, BSM-6000 series (K080342) is utilized to provide technical information to help address the safety and effectiveness of the different characteristics. The methods used to test the SVM-7200 series Vital Signs Monitor are according to the same standards as the cleared reference device BSM-6000 and have been deemed acceptable. Also, the BSM-6000 series incorporates the NIBP module. SpO2 module and firmware of the same design as the SVM-7200 series which controls the operation of NIBP (including measurement ranges) and SpO2 monitoring. Additionally, the storage conditions of the BSM-6000 series bedside monitor are the same as for the SVM-7200 series Vital Signs Monitor and the method used to test the maximum storage temperature and humidity is the same.

PERFORMANCE TESTING

Performance testing for the Nihon Kohden SVM-7200 series Vital Signs Monitor includes software verification and validation test which encompassed software unit testing, integration level testing, and system level testing. A system test was also performed per the software requirements specification. A summative usability validation has been conducted with the subject device in a simulated environment. Testing to compliance standards for electrical and electromagnetic safety was also performed.

8

Traceability has been documented between all system specifications to validation test results. Table 2 lists the performance testing to compliance standards for the VSM-7200 series Vital Signs Monitor.

| Performance

CharacteristicsStandard NumberStandards Title
GeneralIEC 60601-1:2005 +
CORR.1:2006 + CORR.
2:2007 + A1:2012 (with
CAN/CSA C22.2 No.60601-
1:14)Medical electrical equipment - Part 1: General
requirements for basic safety and essential
performance
EMCIEC 60601-1-2:2014Medical Electrical Equipment - Part 1-2: General
Requirements for Basic Safety and Essential
Performance - Collateral Standard: Electromagnetic
Compatibility - Requirements and Tests
UsabilityIEC 60601-1-6:2010 + Amd
1:2013Medical Electrical Equipment - Part 1-6: General
Requirements for Basic Safety and Essential
Performance - Collateral Standard: Usability
IEC 62366:2007 +
Amendment 1:2014Medical devices -- Application of usability
engineering to medical devices
AlarmsIEC 60601-1-8:2006 &
A1:2012Medical electrical equipment - Part 1-8: General
requirements for basic safety and essential
performance - collateral standard: General
requirements, tests and guidance for alarm systems
in medical electrical equipment and medical
electrical systems
Patient
MonitoringIEC 60601-2-49:2011Medical electrical equipment - Part 2-49: Particular
requirements for the basic safety and essential
performance of multifunction patient monitoring
equipment
NIBPIEC 80601-2-
30:2009+A1:2013Medical electrical equipment - Part 2-30: Particular
requirements for the basic safety and essential
performance of automated non-invasive
sphygmomanometers
TempISO 80601-2-56:2009Medical Electrical Equipment - Part 2-56: Particular
Requirements for the Basic Safety and Essential
Performance of Clinical Thermometers for Body
Temperature Measurement
SpO2ISO 80601-2-61:2011Medical electrical equipment - Part 2-61: Particular
requirements for the basic safety and essential
performance of pulse oximeter equipment
SoftwareIEC 62304:2006 +
Amd1:2015Medical Device Software - Software Life-cycle
Processes
Wireless and Co-
ExistenceAAMI TIR69: 2017Technical Information Report Risk management of
radio-frequency wireless coexistence for medical
devices and systems.
Performance
CharacteristicsStandard NumberStandards Title
Wireless and Co-
ExistenceANSI IEEE C63.27-2017American National Standard for Evaluation of
Wireless Coexistence
ANSI IEEE C63.4-2003American National Standard for Methods of
Measurement of Radio-Noise Emissions from Low
Voltage Electrical and Electronic Equipment in the
Range of 9 kHz to 40 GHz.

Table 2 Performance Testing to Standards

9

CLINICAL ACCURACY VALIDATION

The Nihon Kohden SVM-7200 series Vital Signs Monitor incorporates clinical subsystems from Masimo, Nellcor and Nihon Kohden for SpO2 monitoring, from Covidien and Exergen for temperature measurement, and from Nihon Kohden for NIBP monitoring. New clinical accuracy validation testing was not necessary as these subsystems along with their accessories have been previously validated and cleared. This 510(k) leverages the previous clinical accuracy data from the predicates, as shown in Table 3, for all of the clinical functionalities in the subject device.

10

| Patient
Monitoring

SubsystemData Collection MethodSubsystem Cleared in 510(k)SVM7200 Model
MethodComponent510(k)Device Name
ManufacturerSVM-
7230SVM-
7250SVM-
7260
Oxygen
Saturation
(SpO2)
MasimoMasimoBoard,
AlgorithmK053269
K080342Masimo Set Rad-8 Pulse Oximeter
Masimo
BSM-6000
Nihon KohdenX
Sensor cable
(connection
cord)K080342
K151080BSM-6000 (for JL-632P)
Nihon Kohden
CSM-1901(for JL-631P)
Nihon KohdenX
ProbesK040214
K060143
K992340SpO2 probe
MasimoX
Oxygen
Saturation
(SpO2)
NellcorNellcorBoard,
AlgorithmK060576
K080342N-600X
Nellcor
BSM-6000
Nihon KohdenX
Sensor cable
(connection
cord)K151080CSM-1901
Nihon KohdenX
Patient
Monitoring
SubsystemData Collection MethodSubsystem Cleared in 510(k)SVM7200 Model
MethodComponent510(k)Device Name
ManufacturerSVM-
7230SVM-
7250SVM-
7260
Oxygen
Saturation
(SpO2)
Nihon Kohden
Nihon KohdenProbesK052186
K012891SpO2 probe
NellcorX
Nihon
KohdenBoard,
AlgorithmK080342BSM-6000 Nihon KohdenX
Sensor cables
(connection
cord)K082785BSM-9100
Nihon KohdenX
K163459GZ-140P
Nihon KohdenX
Finger probesK043517ZS-940P
Nihon KohdenX
Multisite
probesK043517ZS-940P
Nihon KohdenX
Disposable
probesK120888PSG-1100
Nihon KohdenX
K151080CSM-1901
Nihon KohdenX
K163459GZ-140P
Nihon KohdenX
Patient
Monitoring
SubsystemData Collection MethodSubsystem Cleared in 510(k)SVM7200 Model
MethodComponent510(k)Device Name
ManufacturerSVM-
7230SVM-
7250SVM-
7260
Non-invasive
Blood Pressure
monitoring
(NIBP)
(inflation
mode)Nihon
Kohden
iNIBP
(inflation)Board,
AlgorithmK163459GZ-140P
Nihon KohdenXXX
Sensor cables
(air hoses)K082785BSM-9100
Nihon KohdenXXX
Cuffs -
Reusable and
disposableK163459GZ-140P
Nihon KohdenXXX
Non-invasive
Blood Pressure
monitoring
(NIBP)
(deflation
mode)Nihon
Kohden
NIBP
(deflation)Board,
AlgorithmK914092
K011918
K032749
K080342BSM-2101
BSM-2302
OPV-1510
BSM-6100
Nihon KohdenXXX
Sensor cables
(air hoses)K151080CSM-1901
Nihon KohdenXXX
K082785BSM-9100
Nihon KohdenXXX
Cuffs -
Reusable and
disposableK163459GZ-140P
Nihon KohdenXXX
Data Collection MethodSubsystem Cleared in 510(k)SVM7200 Model
Patient
Monitoring
SubsystemMethodComponent510(k)Device Name
ManufacturerSVM-
7230SVM-
7250SVM-
7260
Body
Temperature
(TEMP)CovidienBoard.
algorithmK003313FILAC FAS TEMP
CovidienXXX
Thermometer
probeK003313FILAC FAS TEMP
CovidienXXX
AccessoriesK003313FILAC FAS TEMP
CovidienXXX
Body
Temperature
(TEMP)ExergenThermometerK011291
K171751Temporal Scanner
Exergen
SmartLinx Vital Plus Patient Monitoring
Capsule TechnologiesXXX

Table 3 SVM-7200 Series Clinical Subsystems in SVM-7200 Series with Corresponding Subsystems in Predicates

11

12

13

14

CONCLUSION

Based on a review of indications for use, technological characteristics, safety and performance testing, the Nihon Kohden SVM-7200 series Vital Signs Monitor is as safe and effective as the SmartLinx Vitals Plus Patient Monitoring System for the intended use described above. Furthermore, the minor differences do not raise new concerns of safety and effectiveness. Therefore, the data provided shows that the subject device is substantially equivalent to the predicate.