K Number
K190326
Date Cleared
2019-03-14

(29 days)

Product Code
Regulation Number
862.1770
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use only

The UREA test within the VITROS XT Chemistry Products UREA-CREA Slides quantitatively measures urea concentration, reported either as urea nitrogen or as urea (UREA), in serum, plasma, and urine using the VITROS XT 7600 Integrated System. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases

The CREA test within the VITROS XT Chemistry Products UREA-CREA Slides quantitatively measures creatinine (CREA) concentration in serum, plasma, and urine using the VITROS XT 7600 Integrated System. Creatinine measurements are used in the diagnosis and treatment of renal dialysis, and as a calculation basis for measuring other urine analytes.

Special conditions for use statement: For prescription use only.

Device Description

The new device, the VITROS XT Chemistry Products UREA-CREA Slide is a single device that contains both a UREA test and a CREA test multilayered, analytical element coated on a polyester support separated by a plastic barrier sealed within a single slide frame. In this format, individual reactions occur and test results are generated for each analyte independently of the other analyte.

To perform the UREA test, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Water and nonproteinaceous components then travel to the underlying reagent layer, where the urease reaction generates ammonia. The semipermeable membrane allows only ammonia to pass through to the colorforming layer, where it reacts with the indicator to form a dye. The reflection density of the dye is measured and is proportional to the concentration of urea in the sample.

To perform the CREA test, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Creatinine diffuses to the reagent layer, where it is hydrolyzed to creatine in the rate-determining step. The creatine is converted to sarcosine and urea by creatine amidinohydrolase. The sarcosine, in the presence of sarcosine oxidase, is oxidized to glycine, formaldehyde, and hydrogen peroxide. The final reaction involves the peroxidase-catalyzed oxidation of a leuco dye to produce a colored product. Following addition of the sample, the slide is incubated. During the initial reaction phase, endogenous creatine in the sample is oxidized. The resulting change in reflection density is measured at 2 time points. The difference in reflection density is proportional to the concentration of creatinine present in the sample.

AI/ML Overview

This document describes the analytical performance of the VITROS XT Chemistry Products UREA-CREA Slides for quantitatively measuring urea and creatinine concentrations. The information provided is for an in-vitro diagnostic device and does not involve AI assistance, human readers, or image analysis, thus many of the requested elements are not applicable.

1. Table of Acceptance Criteria and Reported Device Performance

The device performance is primarily assessed through method comparison (against predicate devices), precision, detection limits, and linearity. The "acceptance criteria" are implied by the measured performance demonstrating substantial equivalence to the predicate devices and meeting clinical laboratory standards for accuracy and precision.

Test ParameterAcceptance Criteria (Implied)Reported Device Performance
Method ComparisonHigh correlation (e.g., r > 0.975) and acceptable bias compared to predicate.UREA Serum: N=124, Slope=1.04, Intercept=0.00, Corr. Coeff.=0.999 (Test Range 3-106 mg/dL, Measuring Range 2.0-120.0 mg/dL) UREA Urine: N=128, Slope=1.05, Intercept=-13.21, Corr. Coeff.=0.999 (Test Range 105-2451 mg/dL, Measuring Range 67-2520 mg/dL) CREA Serum: N=130, Slope=1.00, Intercept=-0.01, Corr. Coeff.=1.000 (Test Range 0.20-13.49 mg/dL, Measuring Range 0.15-14.0 mg/dL) CREA Urine: N=116, Slope=1.01, Intercept=-0.93, Corr. Coeff.=0.998 (Test Range 13.0-336.6 mg/dL, Measuring Range 3.2-346.5 mg/dL)
PrecisionDemonstrated repeatability, within-day, and within-lab precision within acceptable CV% and SD limits for various concentration levels.Detailed tables provided showing SD and CV% for Repeatability, Within Day, and Within Lab precision for multiple pools/controls across Urea Serum, Urea Urine, CREA Serum, and CREA Urine. E.g., UREA Serum Pool 1: Repeatability SD 0.1, CV% 4.0; Within Lab SD 0.2, CV% 8.7. CREA Serum Pool 1: Repeatability SD 0.007, CV% 1.1; Within Lab SD 0.011, CV% 1.7. All measured values are generally low, indicating good precision.
Detection Limits (LoQ)Measured LoQ values no greater than claimed LoQ, with Total Error goal met.UREA Serum: LoQ 1.7 mg/dL (Claimed 2.0 mg/dL); Total Error goal ≤ 1.2 mg/dL UREA Urine: LoQ 41 mg/dL (Claimed 67 mg/dL); Total Error goal ≤ 21 mg/dL Urea N CREA Serum: LoQ 0.11 mg/dL (Claimed 0.15 mg/dL); Total Error goal ≤ 0.06 mg/dL CREA Urine: LoQ 2.3 mg/dL (Claimed 3.2 mg/dL); Total Error goal ≤ 1.2 mg/dL
LinearityLinear range supporting the claimed measuring range.UREA Serum: Linear Range 1.93-148.79 mg/dL (Claimed Measuring Range 2.0-120.0 mg/dL) UREA Urine: Linear Range 62.46-3198.85 mg/dL (Claimed Measuring Range 67-2520 mg/dL) CREA Serum: Linear Range 0.04-14.86 mg/dL (Claimed Measuring Range 0.15-14.0 mg/dL) CREA Urine: Linear Range 1.1-418.3 mg/dL (Claimed Measuring Range 3.2-346.5 mg/dL)

2. Sample Size and Data Provenance

  • Method Comparison Test Set:
    • UREA Serum: 124 samples
    • UREA Urine: 128 samples
    • CREA Serum: 130 samples
    • CREA Urine: 116 samples
  • Precision Test Set: Not specified as a separate patient sample set, but rather "patient pools and quality control materials." Each precision study included "a minimum of 80 observations (2 replicates per run, 2 runs per day over 20 days)" for serum and urine pools/controls.
  • Detection Limits (LoQ) Test Set: 72 determinations for serum (UREA and CREA) and 64 for urine (CREA). 72 for urine (UREA).
  • Linearity Test Set: Twenty proportionally related admixtures of low and high test fluids, each tested in quadruplicate. Specific number of patient samples not stated, as it uses contrived samples.
  • Specificity (Interference) Test Set: Not explicitly stated how many patient samples were used, but rather "96 test substances" and "88 test substances" were spiked into serum samples, and "two substances" and "nineteen test substances" for urine.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the studies used standard clinical laboratory protocols (e.g., CLSI EP09-A3, EP05-A3, EP17-A2, EP06-A) which typically involve fresh or appropriately preserved clinical samples. The reference interval studies mention "an internal study of 3160 apparently healthy adults" and "an external study" (for UREA) or "an external study of apparently healthy adults (serum: 180 males and 180 females)" and "a separate external study" (for CREA).

3. Number of Experts and Qualifications for Ground Truth

  • This is an in-vitro diagnostic device (chemistry analyzer) measuring chemical concentrations. The "ground truth" is established by the reference methodology (predicate devices on a different system) and gravimetric/analytical preparation of calibrators and quality controls. It does not involve human expert interpretation of images or other subjective assessments. Therefore, the concept of "experts establishing ground truth" in the context of radiology or pathology (e.g., radiologists interpreting images) is not applicable here.

4. Adjudication Method for the Test Set

  • Not applicable. This is not a study involving human reader interpretation requiring adjudication. Performance is assessed analytically against quantitative measurements.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

  • No, an MRMC study was not done. This device is an automated in-vitro diagnostic assay, not an AI-assisted diagnostic tool for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, this study represents a standalone analytical performance evaluation of the device's ability to accurately measure urea and creatinine concentrations. "Algorithm" in this context refers to the chemical reactions and measurement principles of the assay, not a software algorithm that processes medical images or data for interpretation. The performance metrics (method comparison, precision, detection limits, linearity, specificity) directly quantify the device's analytical capabilities without human intervention in the measurement process itself.

7. The type of ground truth used

  • The ground truth is established through:
    • Reference Methods/Predicate Devices: For method comparison, the results generated by the predicate devices (VITROS BUN/UREA Slides and VITROS CREA Slides on the VITROS 5600 Integrated System) serve as the comparative ground truth.
    • Analytically Prepared Materials: For precision, detection limits, and linearity, the "ground truth" concentrations of control materials, patient pools, and serially diluted samples are established through precise gravimetric and volumetric preparations, often traceable to certified reference materials.
    • Clinical Laboratory Standards (CLSI Protocols): The studies adhered to widely accepted CLSI guidelines for analytical performance evaluation, implying that the methodologies used to establish "ground truth" for each parameter meet industry standards for accuracy and rigor in a clinical lab setting.

8. The Sample Size for the Training Set

  • This is not an AI/machine learning study that involves "training sets" in the conventional sense. The device's performance is based on the inherent chemical and optical principles of the dry-slide technology. There's no AI model being trained with data.

9. How the ground truth for the training set was established

  • Not applicable, as there is no AI training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 14, 2019

Ortho-Clinical Diagnostics, Inc. Darlene Phillips Manager, Regulatory Affairs 100 Indigo Creek Drive Rochester, NY 14626

Re: K190326

Trade/Device Name: VITROS XT Chemistry Products UREA-CREA Slides Regulation Number: 21 CFR 862.1770 Regulation Name: Urea nitrogen test system Regulatory Class: Class II Product Code: CDN, JFY Dated: February 15, 2019 Received: February 19, 2019

Dear Darlene Phillips:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm -S

Courtney H. Lias, Ph.D. for Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190326

Device Name VITROS XT Chemistry Products UREA-CREA Slides

Indications for Use (Describe)

For in vitro diagnostic use only

The UREA test within the VITROS XT Chemistry Products UREA-CREA Slides quantitatively measures urea concentration, reported either as urea nitrogen or as urea (UREA), in serum, plasma, and urine using the VITROS XT 7600 Integrated System. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases

The CREA test within the VITROS XT Chemistry Products UREA-CREA Slides quantitatively measures creatinine (CREA) concentration in serum, plasma, and urine using the VITROS XT 7600 Integrated System. Creatinine measurements are used in the diagnosis and treatment of renal dialysis, and as a calculation basis for measuring other urine analytes.

Special conditions for use statement: For prescription use only.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: _k190326

Submitter's Information

Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, New York 14626-5101 Phone: (585) 453-4253 Fax: (585) 453-3113

Contact Person:

Darlene J Phillips, RAC Manager, Regulatory Affairs

Date of Preparation: February 12, 2019

Device Proprietary Name(s):

VITROS XT Chemistry Products UREA-CREA Slides

Common Names:

Urea nitrogen assay Creatinine assay

Classification Names

TestProductCodeClassRegulation SectionPanel
UREACDNClass II21 CFR 862. 1770 Urea nitrogen testsystemClinicalChemistry
CREAJFYClass II21 CFR 862. 1225 Creatinine testsystem(75)

Predicate Device(s)

Predicate DevicesFDA 510(k) Number
VITROS Chemistry Products BUN/UREA SlidesK001885
VITROS Chemistry Products CREA SlidesK182063

{4}------------------------------------------------

Intended Use Statement(s) VITROS XT Chemistry Products UREA-CREA Slides

Rx Only For in vitro diagnostic use only

The UREA test within the VITROS XT Chemistry Products UREA-CREA Slides quantitatively measures urea concentration, reported either as urea nitrogen or as urea (UREA), in serum, plasma, and urine using VITROS XT 7600 Integrated Systems. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.

The CREA test within the VITROS XT Chemistry Products UREA-CREA Slides quantitatively measures creatinine (CREA) concentration in serum, plasma, and urine using VITROS XT 7600 Integrated Systems. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Device Description

The new device, the VITROS XT Chemistry Products UREA-CREA Slide is a single device that contains both a UREA test and a CREA test multilayered, analytical element coated on a polyester support separated by a plastic barrier sealed within a single slide frame. In this format, individual reactions occur and test results are generated for each analyte independently of the other analyte.

To perform the UREA test, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Water and nonproteinaceous components then travel to the underlying reagent layer, where the urease reaction generates ammonia. The semipermeable membrane allows only ammonia to pass through to the colorforming layer, where it reacts with the indicator to form a dye. The reflection density of the dye is measured and is proportional to the concentration of urea in the sample.

To perform the CREA test, a drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Creatinine diffuses to the reagent layer, where it is hydrolyzed to creatine in the rate-determining step. The creatine is converted to sarcosine and urea by creatine amidinohydrolase. The sarcosine, in the presence of sarcosine oxidase, is oxidized to glycine, formaldehyde, and hydrogen peroxide. The final reaction involves the peroxidase-catalyzed oxidation of a leuco dye to produce a colored product. Following addition of the sample, the slide is incubated. During the initial reaction phase, endogenous creatine in the sample is oxidized. The resulting change in reflection density is measured at 2 time points. The difference in reflection density is proportional to the concentration of creatinine present in the sample.

{5}------------------------------------------------

Comparison to Predicate Devices

The following tables show similarities and differences between the new and predicate devices. devices:
Summary of the technological characteristics of the device compared to the

T

DeviceCharacteristicNew DeviceVITROS XT UREA-CREA Slide(New)Predícate DevicesVITROS BUN/UREA Slide[K001885]VITROS CREA Slide [K182063](Current)
Intended UseSame for each individual testFor in vitro diagnostic use only.The UREA test within the VITROSXT Chemistry Products UREA-CREA Slides quantitativelymeasures urea concentration,reported either as urea nitrogenor as urea (UREA), in serum,plasma, and urine. The CREAtest within the VITROS XTChemistry Products UREA-CREA Slides quantitativelymeasures creatinine (CREA)concentration in serum, plasma,and urine.For in vitro diagnostic use only.VITROS Chemistry ProductsBUN/UREA Slides quantitativelymeasure urea concentration,reported either as urea nitrogen(BUN) or as urea (UREA), inserum, plasma, and urine.VITROS Chemistry Products CREASlides quantitatively measurecreatinine (CREA) concentrationin serum, plasma, and urine
DeviceDescriptionNo ChangeMultilayered, analytical elementcoated on a polyester support
Basic PrincipleNo ChangeBUN/UREAColorimetricCREAColorimetricTwo point rate
Incubation timeand temperatureNo ChangeApproximately 5 minutes37°C (98.6° F)
Sample typeNo ChangeSerum, plasma, and urine
Summary of the technological characteristics of the device compared to thepredicate device
DeviceCharacteristicNew DeviceVITROS XT UREA-CREA Slide(New)Predícate DevicesVITROS BUN/UREA Slide[K001885]VITROS CREA Slide [K182063](Current)
Amount of SlideReactiveIngredients percm² (test)The composition of the analyticalelement of each test within theVITROS XT Slide will remainthe same as that used in eachpredicate deviceBUN/UREAUrease (jack bean) 1.2 U and N-propyl-4-(2,6-dinitro-4-chlorobenzyl)-quinoloniumethane sulfonate (ammoniaindicator)0.26 mg.CREACreatinine amidohydrolase(Flavobacterium sp.,) 0.20 U;creatine amidinohydrolase(Alcaligenes sp.) 3.6 U; sarcosineoxidase (Bacillus sp.) 0.55 U;peroxidase (horseradish root) 1.6U and 2-(3,5-dimethoxy-4-hydroxyphenyl)-4,5-bis(4-dimethylaminophenyl) imidazole(leuco dye) 32 μg.
Assay RangeNo ChangeBUN/UREA (mg/dL urea N)Serum/plasma 2.0-120.0Urine 67–2520CREA (mg/dL)Serum/plasma 0.15-14.0Urine 3.2-346.5
CalibratorsSameVITROS Chemistry ProductsCalibrator Kit 1
ControlsSameVITROS Chemistry ProductsPerformance Verifier I and II
Differences
InstrumentationVITROS XT 7600 IntegratedSystemVITROS 250/350, 5,1 FS and 4600Chemistry SystemsVITROS 5600 and XT 7600Integrated Systems
Sample volumeUREA 4.3 µLCREA 3.2 µLBUN/UREA 5.5 µLCREA 6 µL

{6}------------------------------------------------

4

{7}------------------------------------------------

Non-Clinical Testing Analytical Performance

Method Comparison

Method Comparison testing followed CLSI Protocol EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples. Serum and urine samples were evaluated on the VITROS XT Chemistry Products UREA-CREA Slides using the VITROS XT 7600 Integrated System and on VITROS Chemistry Products BUN/UREA Slides and VITROS Chemistry Products CREA Slides using the VITROS 5600 Integrated System. The correlation between the predicate and the new tests within the VITROS XT UREA-CREA Slides on the VITROS XT 7600 Integrated System is summarized below.

Summary of Method Comparison Weighted Deming Regression Analysis Data Units mg/dL

TestNSlopeInterceptCorr.Coeff.Test RangeMeasuringrange
UREA Serum1241.040.000.9993 - 1062.0 - 120.0
UREA Urine1281.05-13.210.999105 - 245167 - 2520
CREA Serum1301.00-0.011.0000.20 - 13.490.15 - 14.0
CREA Urine1161.01-0.930.99813.0 - 336.63.2 - 346.5

Precision

Precision was evaluated with patient pools and quality control materials following CLSI Protocol EP05-A3, Evaluation of Precision Performance of Quantitative Methods; Approved Guideline-Third Edition, using the VITROS XT Chemistry Products UREA-CREA Slides on the VITROS XT 7600 Integrated System. The test included a minimum of 80 observations (2 replicates per run, 2 runs per day over 20 days) for serum and urine. The long term precision analysis is summarized below.

The data presented are a representation of test performance and are provided as a guideline. Variables such as sample handling and storage, reagent handling and storage, laboratory environment, and system maintenance can affect reproducibility of test results.

Repeatability (formerly called within-run precision) was determined using two replicates per run. Within Day precision was determined using two runs per day with two replications per run. Within Lab precision was determined using a single lot of slides and a single calibration.

UREA nitrogen SerumUnits (mg/dL urea N)
MeanRepeatabilityWithin DayWithin LabNo.No.
Fluid IdConc.SDCV%SDCV%SDCV%ofObs.ofDays
Pool 130.14.00.28.00.28.78020
Pool 2100.11.00.21.80.22.18020
Native Pool110.11.20.22.10.22.18020
Control 1170.31.60.31.90.31.98020
Control 2510.71.30.71.30.81.58020
Pool 31071.11.01.21.21.41.38020

{8}------------------------------------------------

UREA nitrogen UrineUnits (mg/dL urea N)
RepeatabilityWithin DayWithin LabNo.No.
Fluid IdMeanConc.SDCV%SDCV%SDCV%ofObs.ofDays
Pool 1842.42.86.07.17.28.58020
Native Pool2933.81.35.41.86.82.38020
Control 14044.11.07.11.89.12.38020
Control 26837.61.17.61.112.21.88020
Pool 2145314.01.019.51.319.51.38020
Pool 3233121.40.928.21.230.61.38020
CREA SerumUnits (mg/dL)
Fluid IdMeanRepeatabilityWithin DayWithin LabNo.No.
Conc.SDCV%SDCV%SDCV%ofObs.ofDays
Pool 10.660.0071.10.0081.20.0111.78020
Control 10.850.0101.20.0121.40.0141.68020
Native Pool0.860.0121.40.0121.40.0141.68020
Control 25.410.0300.60.0400.70.0841.68020
Pool 29.410.0680.70.0720.80.1631.78020
Pool 312.620.0900.70.1070.80.2201.78020
CREA UrineUnits (mg/dL)
MeanRepeatabilityWithin DayWithin LabNo.No.
Conc.SDCV%SDCV%SDCV%ofObs.ofDays
Native Pool39.00.260.70.371.00.521.38020
Control 158.50.671.10.901.51.142.08020
Control 2137.51.471.12.201.63.152.38020
Pool 2239.42.130.92.270.93.131.38020
Pool 3317.22.000.64.131.35.441.78020

Detection Limits

Detection capability studies for the UREA and CREA tests within the VITROS XT UREA-CREA Slides were evaluated according to CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures.

The Limit of Quantitation (LoQ) for the UREA test within the VITROS XT UREA-CREA Slides is 2.0 mg/dL for serum/plasma and 67 mg/dL for urine. The total number of LoQ determinations was 72. The Total Error goal used to accept the LoQ was ≤ 1.2 mg/dL for serum and ≤ 21 mg/dL Urea N for urine.

The Limit of Quantitation (LoQ) for the CREA test within the VITROS XT UREA-CREA Slides is 0.15 mg/dL for serum/plasma and 3.2 mg/dL for urine. The total number of LoQ

{9}------------------------------------------------

determinations was 72 for serum and 64 for urine. The Total Error goal used to accept the LoQ was ≤ 0.06 mg/dL for serum and ≤ 1.2 mg/dL for urine.

The results of this analysis are summarized below:
UREA (mg/dL urea N)CREA (mg/dL)
SerumUrineSerumUrine
LOB1.5230.051.1
LOD1.7270.071.3
LOQ1.7410.112.3
Claimed LOQ2.0670.153.2

Linearity

Linearity studies were performed according to CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approved Guideline (2003). VITROS XT UREA-CREA Slides were tested on the VITROS XT 7600 Integrated System. A series of twenty proportionally related admixtures of low and high test fluids were tested with to verify linearity of the UREA and CREA serum and urine tests; each sample was tested in quadruplicate. The linearity studies support the claimed measuring ranges for the individual tests within the VITROS XT UREA-CREA Slides.

AssayFluidMeasuring Range(mg/dL)Linear Range(mg/dL)
UREA nitrogenSerum/plasma2.0-120.01.93-148.79
Urine*67 - 252062.46-3198.85
CREASerum/plasma0.15-14.00.04-14.86
Urine*3.2-346.51.1-418.3

* After multiplying by 21x dilution factor

Serum and plasma samples with values greater than the measuring range may be diluted up to a maximum dilution of 1 part sample with 1 part diluent. A 7% BSA solution was found to be an acceptable diluent for serum and plasma specimens assaved using VITROS XT Chemistry Products UREA-CREA Slides.

Urine samples with values greater than the measuring range may be diluted up to a maximum dilution of 1 part sample with 1 part diluent. Isotonic saline and reagent-grade water were both found to be acceptable diluents for UREA urine specimens assayed using the VITROS XT Chemistry Products UREA-CREA Slides. Reagent-grade water was found to be an acceptable diluent for CREA urine specimens assayed using the VITROS XT Chemistry Products UREA-CREA Slides.

Expected Values

The expected values of the UREA and CREA tests within the VITROS XT UREA-CREA Slides are the not changed from those of the predicate assays. Each laboratory should confirm the validity of these intervals for the population it serves.

{10}------------------------------------------------

UREA Reference Interval

The serum reference interval is the central 95% of results from an internal study of 3160 apparently healthy adults from a working population (612 females and 2548 males).

The urine reference interval is based on an external study.

UREA Reference Interval
Serum
Male9-20 mg/dL urea N
Female7-17 mg/dL urea N
Urine
24-hour12-20 g urea N /day*
  • Urea nitrogen concentration (mg/dL) x 24-hour volume (dL) = mg/day.

CREA Reference Interval

The serum reference intervals are the central 95% of results from an external study of apparently healthy adults (serum: 180 males and 180 females).

The urine reference intervals are based on a separate external study.

CREA Reference Interval
Serum
Male0.66–1.25 mg/dL
Female0.52–1.04 mg/dL
Urine
Male1000–2000 mg/day*
Female800–1800 mg/day*
  • Creatinine concentration (mg/dL) x 24-hour volume (dL) = mg/day.

Specificity

The UREA and CREA tests within the VITROS XT Chemistry Products UREA-CREA Slide were screened for interfering substances following CLSI document EP07-A3, Interference Testing in Clinical Chemistry. The supplemental tables in CLSI document EP37 were referenced for recommended testing concentrations for analytes and endogenous substances that may interfere in clinical chemistry measurement procedures.

Point estimates of the effects of potential interferents on VITROS XT UREA-CREA Slides were made using the paired difference method. Dose-response analysis was conducted to characterize the degree of interference for each substance, and expected bias was reported at the lowest test level which did not meet acceptance criteria for bias as shown in the product claims.

Known Interferences

Ammonium ions may cause an increase in measured UREA value equivalent to the specimen's nitrogen content.

{11}------------------------------------------------

Serum and Plasma

The substances listed in the table, when tested at the concentrations indicated, caused the bias shown. The bias is an estimate of the maximum bias observed.

It is possible that other interfering substances may be encountered. These results are representative; however, your results may differ somewhat due to test-to-test variation. The degree of interference at concentrations other than those listed might not be predictable.

UREA Test

InterferentInterferentConcentrationUrea nitrogenConcentrationBias
Total protein12 g/dL37 mg/dL-4.3 mg/dL

Ninety-six (96) test substances, when tested at the concentrations indicated, as well as serum pH at 6.8 and 8.8, were found not to interfere, bias < 2.0 mg/dL at 9 mg/dL urea N and bias < 4.0 mg/dL at 40 mg/dL urea N. with the UREA test within VITROS XT UREA-CREA Slides in serum.

CREA Test

  • Creatine: At a creatinine concentration of 1.5 mg/dL (133 umol/L), creatine greater ● than 8 mg/dL (707 umol/L) will be flagged with a DP code (because highly elevated creatine concentrations may cause excessive background density). For unflagged samples, residual bias because of creatine will be less than 0.15 mg/dL (13 umol/L). At a creatinine concentration of 14 mg/dL (1237 umol/L), creatine greater than 1 mg/dL (88 umol/L) will be flagged with a DP code. Residual bias for unflagged samples will be less than 2%. Refer to "Sample Dilution" for dilution instructions.
  • . Proline: Patients receiving hyperalimentation fluids containing proline may show an increase of 0.2 mg/dL (18 umol/L). Do not collect specimens from intravenous fluid lines contaminated with hyperalimentation fluid.
InterferentInterferentConcentrationCreatinineConcentrationBias
Conv. UnitsConv. Units(mg/dL)Conv. Units(mg/dL)
Bilirubin, conjugated58 mg/dL1.6-0.16
Dipyrone(Metamizole)*9 mg/dL1.6-0.20
14 mg/dL4.9-0.73
Ethamsylate4.1 mg/dL1.7-0.18
Glutathione69 mg/dL1.6-0.23
N-Ethyl glycine**0.40 mg/dL1.70.21
Proline18 mg/dL1.70.24
Tolazamide5.0 mg/dL1.6-0.24
4.6-0.55
Total protein15 g/dL1.70.26
12 g/dL5.00.53

米 Dipyrone at 9 mg/dL is equivalent to 6X the equivalent of a 1000 mg oral dose, or 1.5X the equivalent of a 1000 mg intravenous dose.

{12}------------------------------------------------

** N-ethyl glycine is a metabolite of lidocaine and may be present at high levels in patients on long-term lidocaine therapy.

Eighty-eight (88) test substances, when tested at the concentrations indicated, as well as serum pH at 6.8 and 8.8. were found not to interfere, bias < 0.13 mg/dL at 1.5 mg/dL and bias < 0.44 mg/dL at 5 mg/dL, with the CREA test within VITROS XT UREA-CREA Slides in serum.

Urine

UREA Test

Two (2) substances were found to interfere with the UREA test in VITROS XT UREA-CREA Slides in urine. Both of these substances are urine preservatives containing boric acid, and the use of urine specimens preserved with boric acid is not recommended for use in the UREA test within VITROS XT UREA-CREA Slides.

Nineteen (19) test substances were found not to interfere with the UREA test in urine, bias < 90 mg/dL at 900 mg/dL urea N and bias < 140 mg/dL at 1400 mg/dL urea N.

CREA Test

Twenty-three (23) test substances were found not to interfere, bias < 5.8 mg/dL at 60 mg/dL and bias < 17 mg/dL at 200 mg/dL, with the CREA test in urine.

Other Limitations

Certain drugs and clinical conditions are known to alter urea nitrogen and creatinine concentrations in vivo. For additional information, refer to one of the published summaries.

Conclusion

The conclusions drawn from the nonclinical tests (discussed above) demonstrate the VITROS XT Chemistry Products UREA-CREA Slides for use on the VITROS XT 7600 Integrated System are as safe, effective, and perform as well as the predicate devices. The information submitted in the premarket notification is complete and supports a substantial equivalence decision.

§ 862.1770 Urea nitrogen test system.

(a)
Identification. A urea nitrogen test system is a device intended to measure urea nitrogen (an end-product of nitrogen metabolism) in whole blood, serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.(b)
Classification. Class II.