(77 days)
The Additive Port Cap is indicated for use on the medication port of Baxter VIAFLEX, Baxter AVIVA, and Baxter ALL-IN-ONE EVA containers to provide both visual evidence that medication has been added and tamper evidence once the device is closed.
The Additive Port Cap (APC) is a polypropylene, single use device designed to snap over the outside of the medication port of compatible Baxter IV container (Baxter VIAFLEX, Baxter INTRAVIA, Baxter AVIVA, and Baxter ALL-IN-ONE EVA) after the addition of medication. Once closed, the device prevents the port from being accessed without causing visible damage to the IV container. The bright red coloration serves as a clear indicator that medication has been added. The APC device is non-fluid path and non-sterile. The APC device is marketed as a stand-alone device and packaged in bulk.
Here's an analysis of the provided text regarding the acceptance criteria and study for the "Additive Port Cap" (APC) device:
Important Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. It primarily details performance testing for engineering characteristics rather than clinical performance or AI/human reader studies. Therefore, many of the requested points related to AI, human readers, and ground truth for clinical outcomes will not be present in this type of document.
1. Table of Acceptance Criteria and Reported Device Performance
The document describes performance testing that was conducted, and states that "All pre-determined acceptance criteria were met." However, the specific quantitative acceptance criteria and the exact reported performance values are not explicitly provided in detail within this summary. Instead, it lists the types of tests performed.
Here's a table based on the information available, indicating where details are missing:
| Test Name | Acceptance Criteria (Not explicitly detailed) | Reported Device Performance (Not explicitly detailed) |
|---|---|---|
| Visual Inspection | No gross damage, adequate coverage of drug port, etc. | Met (Implied by statement: "All pre-determined acceptance criteria were met.") |
| Axial Detachment | Force required to remove the cap from IV container (specific force not given) | Met (Implied by statement: "All pre-determined acceptance criteria were met.") |
| Opening Force | Force required to open a properly installed cap (specific force not given) | Met (Implied by statement: "All pre-determined acceptance criteria were met." Also notes: "cannot be opened manually.") |
| Closing Force | Force required to properly install product (specific force not given) | Met (Implied by statement: "All pre-determined acceptance criteria were met.") |
| Dimensional Verification | Dimensional measurements of cap features (specific ranges not given) | Met (Implied by statement: "All pre-determined acceptance criteria were met.") |
| Leakage | Drug port freedom from leakage | Met (Implied by statement: "All pre-determined acceptance criteria were met.") |
| Packaging & Labeling | Integrity of packaging, legibility of label | Met (Implied by statement: "All pre-determined acceptance criteria were met.") |
| Transportation, Shelf Life | Withstand simulated conditions, maintain function after shelf life | Met (Implied by statement: "All pre-determined acceptance criteria were met.") |
| Product Validation (Human Factors & Usability) | Usability for intended purpose (specific metrics not given) | Met (Implied by statement: "All pre-determined acceptance criteria were met.") |
| Tamper Evidence | If opened, causes visible damage to bag/cap or tears bag. | Confirmed: "pulling on the cap in order to remove from the bag tears the bag" and "If a closed cap is opened (e.g. with a tool), it can cause damage to the bag or the cap making it unusable." |
Information Not Found / Not Applicable Given the Device Type:
The device, an "Additive Port Cap," is a physical medical device (plastic cap) designed for tamper evidence and visual indication. It is not an AI/software device or diagnostic imaging device. Therefore, many of the requested points related to AI/ML study design are not applicable to this submission.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact sample sizes used for each of the performance tests. It merely states that "Performance testing to demonstrate tamper evidence was conducted" and lists various tests.
- Data Provenance: Not applicable in the context of clinical data. The tests are engineering and benchtop tests of a physical product.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Not applicable. This is a physical device. "Ground truth" in the sense of expert medical interpretation for a diagnostic AI is not relevant here. The "ground truth" for the performance tests would be the measurement results from the engineering tests themselves (e.g., force required to open, presence of leakage).
4. Adjudication Method for the Test Set
- Not applicable. There's no clinical "test set" requiring adjudication by multiple experts. The tests are physical measurements and observations against engineering specifications.
5. If a Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. This type of study is typically done for diagnostic imaging devices or AI-assisted solutions to compare human performance with and without AI. It is not relevant for a physical device like an IV port cap.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not applicable. There is no algorithm or AI component in this device.
7. The Type of Ground Truth Used
- Engineering Specifications and Physical Measurements: The ground truth for this device's performance would be derived from the mechanical properties, dimensional tolerances, and functional performance (e.g., ability to indicate tamper, withstand certain forces, prevent leakage) as defined by its design and intended use. This is established via physical testing.
8. The Sample Size for the Training Set
- Not applicable. This is not an AI/ML device; therefore, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
- Not applicable. There is no AI/ML device or training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 30, 2019
International Medical Industries, Inc. Peter Lehel Senior Manager Engineering 2981 Gateway Drive Pompano Beach, Florida 33069
Re: K190305
Trade/Device Name: Additive Cap Regulation Number: 21 CFR 880.5025 Regulation Name: IV Container Regulatory Class: Class II Product Code: KPE Dated: February 7, 2019 Received: February 12, 2019
Dear Peter Lehel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K190305
Device Name Additive Port Cap
Indications for Use (Describe)
The Additive Port Cap is indicated for use on the medication port of Baxter VIAFLEX, Baxter AVIVA, and Baxter ALL-IN-ONE EVA containers to provide both visual evidence that medication has been added and tamper evidence once the device is closed.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| □ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for IMI. The logo consists of a blue square with three curved white lines inside, followed by the letters "IMI" in black. There is a registered trademark symbol to the upper right of the last "I" in the logo.
K190305 510(k) Summary
| Date Summary Prepared: | February 7, 2019 | |
|---|---|---|
| Manufacturing Company: | International Medical Industries, Inc. | |
| Manufacturing Address: | 2981 Gateway Drive, Pompano Beach, Florida 33069 | |
| Corresponding Official: | Peter Lehel | |
| Senior Manager Engineering | ||
| Telephone Number: | 954.917.9570 x 227 | |
| Fax Number: | 954.917.9244 | |
| Email Address: | plehel@imiweb.com | |
| Trade Name: | Additive Port Cap | |
| Device Common Name: | Additive Cap | |
| Regulation Number: | 21 CFR 880.5025 | |
| Regulation Name: | I.V. Container | |
| Product Class: | II | |
| Product Panel: | General Hospital | |
| Product Code: | KPE | |
| Predicate Device: | Baxter Additive Cap for Viaflex Containers | |
| K111217 cleared June 14, 2011 |
5.1 Device Description
The Additive Port Cap (APC) is a polypropylene, single use device designed to snap over the outside of the medication port of compatible Baxter IV container (Baxter VIAFLEX, Baxter INTRAVIA, Baxter AVIVA, and Baxter ALL-IN-ONE EVA) after the addition of medication. Once closed, the device prevents the port from being accessed without causing visible damage to the IV container. The bright red coloration serves as a clear indicator that medication has been added. The APC device is non-fluid path and non-sterile. The APC device is marketed as a stand-alone device and packaged in bulk.
5.2 Indications for Use
| Device:Additive Port Cap (APC) | Predicate:Baxter Additive Cap for ViaflexContainers K111217 |
|---|---|
| The Additive Port Cap is indicated for useon the medication port of BaxterVIAFLEX, Baxter INTRAVIA, BaxterAVIVA, and Baxter ALL-IN-ONE EVAcontainers to provide both visual evidencethat medication has been added andtamper evidence once the device isclosed. | Additive Cap is indicated for use on themedication port of VIAFLEX andAVIVA containers to provide visualevidence that medication has beenadded. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the IMI logo. The logo consists of a blue square with three curved lines inside, followed by the letters "IMI" in black. The "I" on the right has a registered trademark symbol next to it. The logo is simple and modern.
5.3 Technological Characteristics and Basis for Substantial Equivalence
The APC is substantially equivalent to the predicate device, Baxter Additive Cap for VIAFLEX and AVIVA Containers (K111217). The devices are both indicated as non-fluid path caps which provide visual evidence that medication has been added to the specified IV containers. Both devices are injection molded polypropylene and are provided non-sterile.
| Parameter | Proposed Device | Predicate Device |
|---|---|---|
| Image | Image: Proposed Device Images: Injection port fit. and Installed. | Image: Predicate Device Images: Injection port fit. and Installed. |
| Proprietary DeviceName | Additive Port Cap (APC) | Additive Cap for Viaflex Containers |
| Company Name | International Medical Industries, Inc. | Baxter Healthcare Corporation |
| 510(k) # | K190305 | K111217 |
| Product Code | KPE | KPE |
| Regulation No. | 21 CFR 880.5025 | 21 CFR 880.5025 |
| Classification | Class II | Class II |
| Sterility | Non-Sterile | Non-Sterile |
| Number of Uses | Single Use | Single Use |
| Cap Material | Polypropylene | Polypropylene |
| BiocompatibilityContact andDuration | Device does not contact the patient's bodydirectly or indirectly. | Device does not contact the patient's bodydirectly or indirectly. |
Table 5.3.1 - Comparison Between APC and Predicate Devices
The label modification is to expand the indications for use statement by adding compatibility with the Baxter INTRAVIVA and Baxter ALL-IN-ONE EVA containers which are functionally equivalent to the VIAFLEX and AVIVA. The second label modification is to add tamper evidence to the indications for use. Performance testing to demonstrate tamper evidence was conducted. Once the Additive Port Cap is attached and closed, it cannot be opened manually. If a closed cap is opened (e.g. with a tool), it can cause damage to the bag or the cap making it unusable. Additionally, performance testing demonstrated that pulling on the cap in order to remove from the bag tears the bag.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for IMI. The logo consists of a blue square with white curved lines inside, followed by the letters "IMI" in black. There is a registered trademark symbol to the right of the last "I".
| Feature of the Device | Proposed Device | Predicate Device | Discussion / Comment |
|---|---|---|---|
| Compatibility with Baxterfluid containers. | Baxter VIAFLEX,Baxter INTRAVIA,Baxter AVIVA, andBaxter ALL-IN-ONEEVA containers | VIAFLEX andAVIVA containers | The medication port areequivalent for all thecontainers. |
| Tamper Evidence | Includes tamperevidence | Does not includetamper evidence. | Performance testing wasconducted to support claimof tamper evidence. |
Table 5.3.2 - Differences Between APC and Predicate Devices
The safety and effectiveness of the APC are adequately supported by the material information, comparison of design characteristics with the predicate device, testing rationale and data provided within this premarket notification. This testing supports the substantial equivalence of the APC to the predicate and shows that no different questions of safety and effectiveness have been introduced with this device.
5.4 Performance Data
Design verification and validation were performed to ensure that the APC device meets its performance specifications and demonstrates substantial equivalence to the predicate device. There are no known performance standards for this device. A summary of the performance testing conducted is presented below. All pre-determined acceptance criteria were met. The data demonstrates that the proposed cap device is substantially equivalent to the predicate device.
| Test Name | Test Description |
|---|---|
| Visual Inspection | Visual Inspection to ensure no gross damage, coverageof drug port etc. |
| Axial Detachment | Force required to remove the additive port cap from IVcontainer |
| Opening Force | Force required to open a properly installed cap |
| Closing Force | Force required to properly install product on injectionport of IV container |
| Dimensional Verification | Dimensional measurements of cap features |
| Leakage | Drug port freedom from leakage |
Table 5.4.1 - Summary of Performance Testing Conducted on APC Device
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for IMI. The logo consists of a blue square with white curved lines inside, followed by the letters "IMI" in black. There is a registered trademark symbol next to the last "I".
| Packaging and Labeling | Visual inspections to verify integrity of the packaging toprotect the device and of the label to be legible. |
|---|---|
| Transportation, Shelf Life | Simulated transportation conditions and shelf lifefollowed by design verification testing. |
| Product Validation | Human factors and Usability |
5.5 Biocompatibility Testing
The device does not contact the patient's body either directly. In accordance to FDA Guidance Document "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process; Guidance for Industry and Food and Drug Administration Staff" issued June 16, 2016, no biocompatibility testing is required for the APC device.
Conclusion
Review of the performance test data as well as comparison of the device classification, indications for use, operating principle, technological characteristics, sterility, and biocompatibility demonstrate that the subject device, Additive Port Cap (APC) is substantially equivalent to the predicate cap Baxter Additive Cap for VIAFLEX and AVIVA Baxter containers K111217. Any differences between the subject and the predicate devices do not raise any different questions of safety or effectiveness.
§ 880.5025 I.V. container.
(a)
Identification. An I.V. container is a container made of plastic or glass used to hold a fluid mixture to be administered to a patient through an intravascular administration set.(b)
Classification. Class II (performance standards).