K Number
K190297
Device Name
Celerity HP Incubator
Manufacturer
Date Cleared
2019-05-14

(92 days)

Product Code
Regulation Number
880.2800
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Use the Celerity HP Incubator to incubate and automatically read Celerity 20 HP Biological Indicators at 57 °C for a fluorescent result within 20 minutes.
Device Description
Celerity HP Incubator (Incubator) is an incubator/reader designed for use specifically with the Celerity 20 HP Biological Indicator (BI). The incubator/reader provides a constant temperature range to allow for activation and outgrowth of Geobacillus stearothermophilus which is detected by an enzyme that cleaves the a substrate leading to production of a fluorescent moiety. The presence of an increasing fluorescence signal due to increasing concentrations of this fluorescent moiety in the BI is detected by the incubator/reader and indicates the presence of viable test microorganisms.
More Information

No
The description focuses on standard biological indicator incubation and fluorescence detection, with no mention of AI/ML for analysis or interpretation.

No
The device is an incubator/reader for biological indicators used to detect the presence of microorganisms, not for treating any medical condition or disease.

No

This device is an incubator/reader for biological indicators, used to test the effectiveness of sterilization processes, not to diagnose a medical condition in a patient.

No

The device description clearly states it is an "incubator/reader" and describes hardware components like a temperature control system and a fluorescence detection system. The performance studies also include electrical safety and electromagnetic compatibility testing, which are relevant to hardware devices.

Based on the provided information, the Celerity HP Incubator is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • Intended Use: The intended use is to incubate and read biological indicators used for sterilization monitoring. This is a quality control process for sterilization equipment, not a diagnostic test performed on a human sample to diagnose a disease or condition.
  • Device Description: The device description confirms its function is to detect the presence of viable microorganisms in a biological indicator, which is a test of the sterilization process, not a test of a patient sample.
  • Lack of IVD Characteristics: The information provided does not mention any of the typical characteristics of an IVD device, such as:
    • Testing human specimens (blood, urine, tissue, etc.)
    • Providing information for the diagnosis, treatment, or prevention of disease
    • Being used in a clinical laboratory setting for patient testing

The device is clearly designed for use in a sterilization monitoring context, which falls outside the scope of IVD regulation.

N/A

Intended Use / Indications for Use

Use the Celerity HP Incubate and automatically read Celerity 20 HP Biological Indicators at 57 °C for a fluorescent result within 20 minutes.

Product codes (comma separated list FDA assigned to the subject device)

FRC

Device Description

Celerity HP Incubator (Incubator) is an incubator/reader designed for use specifically with the Celerity 20 HP Biological Indicator (BI). The incubator/reader provides a constant temperature range to allow for activation and outgrowth of Geobacillus stearothermophilus which is detected by an enzyme that cleaves the a substrate leading to production of a fluorescent moiety. The presence of an increasing fluorescence signal due to increasing concentrations of this fluorescent moiety in the BI is detected by the incubator/reader and indicates the presence of viable test microorganisms.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance testing was performed to verify that the subject device will meet the acceptance criteria of the performance test shown in Table 5-3 below.

  • Electrical Safety - General: Conform with UL 61010-1 3rd Edition (PASS)
  • Electrical Safety - Heating Equipment: Conform with UL 61010-2-010:2015 Ed.3 (PASS)
  • Electromagnetic Compatability: Conform with IEC 61326-1:2012 and FCC 47 CFR Part 15 (PASS)
  • Alarm, LED and Print function Test: Demonstrate proper function of alarms, LED and print outputs (PASS)

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K171587

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K173670

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

May 14, 2019

Steris Corporation Anthony Piotrkowski Director, Regulatory Affairs 5960 Heisley Rd Mentor, Ohio 44060

Re: K190297

Trade/Device Name: Celerity HP Incubator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: April 16, 2019 Received: April 17, 2019

Dear Anthony Piotrkowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

1

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Krause, PhD Acting Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

2

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K190297

Device Name Celerity HP Incubator

Indications for Use (Describe)

Use the Celerity HP Incubate and automatically read Celerity 20 HP Biological Indicators at 57 °C for a fluorescent result within 20 minutes.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

3

STERIS®

510(k) Summary For Celerity HP Incubator K190297

Sponsor Facility

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

Manufacturing Facility

STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896

Contact: Anthony Piotrkowski

Telephone: (440) 392-7437 Fax No: e-mail: tony_piotrkowski@steris.com

February 8, 2019 Submission Date:

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

4

Device Name 1.

Trade Name:Celerity HP Incubator
Common/usual Name:Incubator/Reader (accessory to Biological Indicator)
Device Classification:Class II
Classification Name:Indicator, Biological Sterilization Process
(21 CFR 880.2800, FRC)

2. Predicate Device

K171587 - VERIFY® Incubator for Assert™ |VH2O2 Self-Contained Biological Indicators (name changed to Celerity HP Incubator after clearance)

3. Reference Device

K173670 - Celerity 20 Steam Incubator

4. Description of Device

Celerity HP Incubator (Incubator) is an incubator/reader designed for use specifically with the Celerity 20 HP Biological Indicator (BI). The incubator/reader provides a constant temperature range to allow for activation and outgrowth of Geobacillus stearothermophilus which is detected by an enzyme that cleaves the a substrate leading to production of a fluorescent moiety. The presence of an increasing fluorescence signal due to increasing concentrations of this fluorescent moiety in the BI is detected by the incubator/reader and indicates the presence of viable test microorganisms.

ട. Intended Use/ Indications for Use

Use the Celerity HP Incubator to incubate and automatically read Celerity 20 HP Biological Indicators at 57 ℃ for a fluorescent result within 20 minutes.

6. Summary of Technical Characteristics

Comparisons of technical characteristics versus the predicate and reference devices are summarized in Tables 5-1 and 5-2 respectively.

5

Table 5-1 Summary of Incubator Physical Description and Technological Properties vs Predicate

| Feature | VERIFY Incubator for Assert
VH2O2 SCBI
(predicate K171587) | Celerity HP Incubator
(modified) | Comparison |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Intended Use | Use the Celerity HP Incubator to
incubate and automatically read
VERIFY Assert VH2O2 Self-
Contained Biological Indicators at
57 °C for a fluorescent result
within 20 minutes. | Use the Celerity HP Incubator to
incubate and automatically read
Celerity HP Biological Indicators
at 57 °C for a fluorescent result
within 20 minutes. | Same
indications
for use – the
name of the
incubator
has changed. |
| Basis of
Readout | Photodiode detects fluorescence
produced by enzymatic activity
that results from growing
biological indicator organisms | Photodiode detects fluorescence
produced by enzymatic activity
that results from growing
biological indicator organisms | Same |
| Incubation
Temperature
Range | 55 - 60 °C | 55 - 60 °C | Same |
| Voltage
Range | 100-240 VAC with 12 VDC
conversion. | 100-240 VAC with 12 VDC
conversion. | Same |
| Test capacity | 8 wells | 8 wells | Same |
| Calibration | Factory calibration – no calibration
by customer | Factory calibration – no calibration
by customer | Same |
| Incubation
Time | 20 minutes | 20 minutes | Same |
| Fluorescence
Detection | UV LEDs are used to excite the
fluorescent molecule produced by
enzyme cleavage of the
fluorogenic substrate contained in
the SCBI's media. The emitted
light is detected by a photodiode. | UV LEDs are used to excite the
fluorescent molecule produced by
enzyme cleavage of the
fluorogenic substrate contained in
the SCBI's media. The emitted
light is detected by a photodiode. | Same |
| Compliance | Electrical Safety and EMC Testing
• IEC 61010-1 (2010) Third Ed
• IEC 61010-2-010 (2013) Third
Ed
Electromagnetic compatibility
• USA Title 47, Code of Federal
Regulations (2007) for:
o Radiated Emissions (FCC
Part 15, Subpart B, Class A)
o Conducted Emissions (FCC
Part 15, Subpart B, Class A)
• IEC 61326:2013 -
• EN 55011:2009, Inc. A1:2010
• EN 61000-3-2:2006, Inc.
A1:2009 and A2:2009
EN 61000-3-3:2013 | Electrical Safety and EMC Testing
• IEC 61010-1 (2010) Third Ed
• IEC 61010-2-010 (2013) Third
Ed
Electromagnetic compatibility
• USA Title 47, Code of Federal
Regulations (2007) for:
o Radiated Emissions (FCC
Part 15, Subpart B, Class A)
o Conducted Emissions (FCC
Part 15, Subpart B, Class A)
• IEC 61326:2013 -
• EN 55011:2009, Inc. A1:2010
• EN 61000-3-2:2006, Inc.
A1:2009 and A2:2009
EN 61000-3-3:2013 | Same |
| Indication of
Results | Positive - audible alarm, visual
LED lights and screen
Negative - no alarm, visual
indication with LED lights and
LCD screen
User must acknowledge results | Positive - audible alarm, visual
LED lights and screen
Negative - no alarm, visual
indication with LED lights and
LCD screen
User must acknowledge results | Same |
| Feature | VERIFY Incubator for Assert
VH2O2 SCBI
(predicate K171587) | Celerity HP Incubator
(modified) | Comparison |
| System
Operation | The reader/incubator wells are
arranged in 2 banks of 4 wells and
preset to 57℃. The measurement
of fluorescence is initiated by
placement of a VERIFY Assert
SCBI into any of the incubation
wells and pressing the adjacent
"ACTION" button.
When an SCBI is placed into a
well, the auto-reader detects its
presence. Upon pressing the button
associated with that well, a
blinking yellow light indicates that
incubation is in process and the
read initiated.
A "positive" reading is interpreted
as an indication of a potential
sterilization cycle failure. A
"positive" finding is indicated to
the user by red light on the front
panel adjacent to the well, by an
audible alarm, and by text
displayed on the LCD screen. The
alarm must be muted by the
operator when a positive result is
obtained. The LCD screen
provides instructions for the user to
turn off the alarm for that specific
SCBI.
Should another BI become
"positive", the alarm will sound
again and the above actions are
repeated.
If the result is not positive at the
end of the incubation time, the
result is negative. Negative results
are identified by a green light on
the front panel adjacent to the well
with the "negative" BI and by text
on the LCD. | The reader/incubator wells are
arranged in 2 banks of 4 wells and
preset to 57℃. The measurement
of fluorescence is initiated by
placement of a VERIFY Assert
SCBI into any of the incubation
wells and pressing the adjacent
"ACTION" button.
When an SCBI is placed into a
well, the auto-reader detects its
presence. Upon pressing the button
associated with that well, a
blinking yellow light indicates that
incubation is in process and the
read initiated.
A "positive" reading is interpreted
as an indication of a potential
sterilization cycle failure. A
"positive" finding is indicated to
the user by red light on the front
panel adjacent to the well, by an
audible alarm, and by text
displayed on the LCD screen. The
alarm must be muted by the
operator when a positive result is
obtained. The LCD screen
provides instructions for the user to
turn off the alarm for that specific
SCBI.
Should another BI become
"positive", the alarm will sound
again and the above actions are
repeated.
If the result is not positive at the
end of the incubation time, the
result is negative. Negative results
are identified by a green light on
the front panel adjacent to the well
with the "negative" BI and by text
on the LCD. | Same |

6

Table 5-2 Summary of Incubator Physical Description and Technological Properties vs. Reference Device

| Feature | Celerity Steam Incubator
(Reference K173670) | Celerity HP Incubator
(modified) | Comparison |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Intended Use | Use the Celerity 20 Steam
Incubator to incubate and
automatically read Celerity 20
Steam Biological Indicators at 57
°C for a fluorescent result within
20 minutes. | Use the Celerity HP Incubator to
incubate and automatically read
Celerity HP Biological Indicators
at 57 °C for a fluorescent result
within 20 minutes. | Same
operating
temperature
and read
time |
| Feature | Celerity Steam Incubator
(Reference K173670) | Celerity HP Incubator
(modified) | Comparison |
| Basis of
Readout | Photodiode detects fluorescence
produced by enzymatic activity
that results from growing
biological indicator organisms | Photodiode detects fluorescence
produced by enzymatic activity
that results from growing
biological indicator organisms | Same |
| Incubation
Temperature
Range | 55 - 60 °C | 55 - 60 °C | Same |
| Voltage
Range | 90 to 264 VAC with 12 VDC
conversion. | 100-240 VAC with 12 VDC
conversion. | Proposed
within range
of reference |
| Test capacity | 8 wells | 8 wells | Same |
| Calibration | Factory calibration – no calibration
by customer | Factory calibration – no calibration
by customer | Same |
| Incubation
Time | 20 minutes | 20 minutes | Same |
| Fluorescence
Detection | UV LEDs are used to excite the
fluorescent molecule produced by
enzyme cleavage of the
fluorogenic substrate contained in
the SCBI's media. The emitted
light is detected by a photodiode. | UV LEDs are used to excite the
fluorescent molecule produced by
enzyme cleavage of the
fluorogenic substrate contained in
the SCBI's media. The emitted
light is detected by a photodiode. | Same |
| Compliance | Electrical Safety and EMC Testing
IEC 61010-1 (2010) Third Ed IEC 61010-2-010 (2013) Third Ed Electromagnetic compatibility USA Title 47, Code of Federal Regulations (2007) for: Radiated Emissions (FCC Part 15, Subpart B, Class A) Conducted Emissions (FCC Part 15, Subpart B, Class A) IEC 61326:2013 - EN 55011:2009, Inc. A1:2010 EN 61000-3-2:2006, Inc. A1:2009 and A2:2009 EN 61000-3-3:2013 | Electrical Safety and EMC Testing
IEC 61010-1 (2010) Third Ed IEC 61010-2-010 (2013) Third Ed Electromagnetic compatibility USA Title 47, Code of Federal Regulations (2007) for: Radiated Emissions (FCC Part 15, Subpart B, Class A) Conducted Emissions (FCC Part 15, Subpart B, Class A) IEC 61326:2013 - EN 55011:2009, Inc. A1:2010 EN 61000-3-2:2006, Inc. A1:2009 and A2:2009 EN 61000-3-3:2013 | Same |
| Indication of
Results | Positive - audible alarm, visual
LED lights and screen
Negative - no alarm, visual
indication with LED lights and
LCD screen
User must acknowledge results | Positive – audible alarm, visual
LED lights and screen
Negative - no alarm, visual
indication with LED lights and
LCD screen
User must acknowledge results | Same |
| Feature | Celerity Steam Incubator
(Reference K173670) | Celerity HP Incubator
(modified) | Comparison |
| System
Operation | The reader/incubator wells are
arranged in 2 banks of 4 wells and
preset to 57℃. The measurement
of fluorescence is initiated by
placement of a VERIFY Assert
SCBI into any of the incubation
wells and pressing the adjacent
"ACTION" button.
When an SCBI is placed into a
well, the auto-reader detects its
presence. Upon pressing the button
associated with that well, a
blinking yellow light indicates that
incubation is in process and the
read initiated.
A "positive" reading is interpreted
as an indication of a potential
sterilization cycle failure. A
"positive" finding is indicated to
the user by red light on the front
panel adjacent to the well, by an
audible alarm, and by text
displayed on the LCD screen. The
alarm must be muted by the
operator when a positive result is
obtained. The LCD screen
provides instructions for the user to
turn off the alarm for that specific
SCBI.
Should another BI become
"positive", the alarm will sound
again and the above actions are
repeated.
If the result is not positive at the
end of the incubation time, the
result is negative. Negative results
are identified by a green light on
the front panel adjacent to the well
with the "negative" BI and by text
on the LCD. | The reader/incubator wells are
arranged in 2 banks of 4 wells and
preset to 57℃. The measurement
of fluorescence is initiated by
placement of a VERIFY Assert
SCBI into any of the incubation
wells and pressing the adjacent
"ACTION" button.
When an SCBI is placed into a
well, the auto-reader detects its
presence. Upon pressing the button
associated with that well, a
blinking yellow light indicates that
incubation is in process and the
read initiated.
A "positive" reading is interpreted
as an indication of a potential
sterilization cycle failure. A
"positive" finding is indicated to
the user by red light on the front
panel adjacent to the well, by an
audible alarm, and by text
displayed on the LCD screen. The
alarm must be muted by the
operator when a positive result is
obtained. The LCD screen
provides instructions for the user to
turn off the alarm for that specific
SCBI.
Should another BI become
"positive", the alarm will sound
again and the above actions are
repeated.
If the result is not positive at the
end of the incubation time, the
result is negative. Negative results
are identified by a green light on
the front panel adjacent to the well
with the "negative" BI and by text
on the LCD. | Same |

7

8

7. Summary of Nonclinical Tests

Performance testing was performed to verify that the subject device will meet the acceptance criteria of the performance test shown in Table 5-3 below.

9

TestAcceptance CriteriaConclusion
Electrical Safety -
GeneralConform with UL 61010-1 3rd EditionPASS
Electrical Safety –
Heating EquipmentConform with UL 61010-2-010:2015 Ed.3PASS
Electromagnetic
CompatabilityConform with IEC 61326-1:2012 and FCC 47
CFR Part 15PASS
Alarm, LED and
Print function TestDemonstrate proper function of alarms, LED and
print outputsPASS

Table 5-3. Summary of Non-clinical Testing

8. Conclusion

The Celerity HP Incubator has met the established performance criteria. Based on the intended use, technological characteristics and non-clinical test performed, the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device, VERIFY® Incubator for Assert™ VH2O2 Self-Contained Biological Indicators, K171587, Class II (21 CFR 880.2800, Product code FRC).