(98 days)
The DiaSorin Molecular Simplexa™ VZV Direct assay is intended for use on the LIAISON® MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA in cerebrospinal fluid (CSF) from patients signs and/or symptoms of meningitis and/or encephalitis. This test is intended as an aid in the diagnosis of VZV infections of the central nervous system (CNS).
Negative results do not preclude VZV infection and should not be used as the sole basis for treatment or other patient management decisions.
The assay is not intended for use as a donor screening test. The assay is for professional use only.
The Simplexa™ VZV Positive Control Pack is intended to be used as a control with the Simplexa™ VZV Direct kit.
This control is not intended for use with other assays or systems.
The Simplexa™ VZV Direct assay is a real-time polymerase chain reaction (PCR) system that enables the direct amplification and detection of VZV DNA from unprocessed cerebral spinal fluid (CSF) specimens without nucleic acid extraction. The system consists of the Simplexa™ VZV Direct assay, the LIAISON® MDX (with LIAISON® MDX Studio Software), the Direct Amplification Disc and associated accessories. In the Simplexa™ VZV Direct assay, fluorescent probes are used together with corresponding forward and reverse primers to amplify VZV and internal control targets. A well-conserved region of the VZV DNA polymerase gene is targeted to identify VZV DNA in the specimen. An internal control is used to detect PCR failure and/or inhibition.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Device Performance for Simplexa™ VZV Direct Assay
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list "acceptance criteria" as a set of predefined thresholds. Instead, it presents performance metrics from various studies. Based on the provided clinical agreement study, the implicit acceptance criteria for clinical performance would likely be a high Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA). For analytical performance, the LoD is a key metric.
| Metric (Implicit Acceptance Criteria) | Reported Device Performance (Simplexa™ VZV Direct) |
|---|---|
| Clinical Performance | |
| Positive Percent Agreement (PPA) | 100.0% (12/12) (95% CI: 75.7% to 100.0%) |
| Negative Percent Agreement (NPA) | 99.7% (623/625) (95% CI: 98.8% to 99.9%) |
| PPA (Contrived Samples) | 100.0% (120/120) (95% CI: 96.9-100.0%) |
| Analytical Performance | |
| Limit of Detection (LoD) VZV Strain 9939 | 2.03 TCID50/mL (1,614 copies/mL) |
| Limit of Detection (LoD) VZV Strain Ellen | 0.001 TCID50/mL (1,505 copies/mL) |
| Analytical Reactivity (VZV strains) | 100.0% agreement for 5 additional VZV strains |
| Cross-Reactivity (159 microorganisms) | 0.0% detection (no cross-reactivity observed) |
| Reproducibility (VZV) %CV | 0.5-4.6% |
| Interference | No interference observed |
| Inhibition by other microorganisms | No inhibition observed |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Clinical):
- Prospective Samples: 637 clinical samples.
- Contrived Samples: 240 contrived VZV positive samples.
- Data Provenance:
- Country of Origin: Not explicitly stated, but the submission is to the U.S. FDA by a company in Cypress, California, USA, and the testing was performed at "testing sites" and "DiaSorin Molecular, Cypress, CA", suggesting U.S.-based data.
- Retrospective or Prospective: The study included both:
- Prospective and prospectively banked collections from eight (8) collection sites (February 2018 to November 2018).
- Contrived samples were used to supplement low prevalence true positive clinical samples.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
Not applicable. This device is an in vitro diagnostic (IVD) based on molecular detection. Ground truth for diagnostic assays like this is typically established by comparative methods rather than expert human interpretation of images or other subjective data. No human "experts" were used to establish the ground truth in the sense of clinical decision-making from images.
4. Adjudication Method for the Test Set
Not applicable in the sense of expert human review (e.g., 2+1, 3+1). The ground truth for the clinical samples was established using a composite reference method:
- Two (2) validated real-time PCR assays.
- Followed by confirmation of positive PCR amplification products with bi-directional sequencing.
- Decision Rule: Samples were characterized as positive if one (1) or both PCR assays were positive AND confirmed by bi-directional sequencing. Samples were characterized as negative if both PCR assays were negative.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) designed for laboratory use to detect VZV DNA. Its performance is evaluated against reference methods, not against human readers (e.g., radiologists, pathologists). Therefore, improvement for human readers with AI assistance is not applicable in this context.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)
Yes, the primary performance studies presented focus on the standalone performance of the Simplexa™ VZV Direct assay. The "Clinical Agreement" section directly compares the Simplexa™ VZV Direct results to the composite reference method without explicit human intervention in the interpretation of the Simplexa™ VZV Direct results. The assay is an automated real-time PCR system.
7. Type of Ground Truth Used
The ground truth for the clinical agreement study was established using a composite reference method. This method involved:
- Two (2) validated real-time PCR assays.
- Confirmation of positive PCR amplification products with bi-directional sequencing.
8. Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of machine learning or AI models. This device is an in vitro diagnostic (IVD) based on real-time PCR technology, which relies on molecular biology principles (primers and probes) rather than machine learning models that require distinct training and test sets. The assay design and optimization would involve internal development and validation, but not typically a "training set" in the AI sense.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as a "training set" for an AI model is not described or relevant for this type of molecular diagnostic device. The analytical characteristics and design of the PCR assay (e.g., primer and probe specificity) are established through various analytical studies (e.g., analytical sensitivity, specificity, cross-reactivity) rather than a ground-truthed training set for machine learning.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 13, 2019
DiaSorin Molecular LLC Sharon Young Principal Regulatory Affairs Specialist 11331 Valley View Street Cypress, California 90630
Re: K190219
Trade/Device Name: Simplexa VZV Direct, Simplexa VZV Positive Control Pack Regulation Number: 21 CFR 866.3970 Regulation Name: Device to detect and identify microbial pathogen nucleic acids in cerebrospinal fluid Regulatory Class: Class II Product Code: PLO Dated: January 30, 2019 Received: February 4, 2019
Dear Sharon Young:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours.
for
Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K190219
Device Name
Simplexa VZV Direct, Simplexa VZV Positive Control Pack
Indications for Use (Describe)
The DiaSorin Molecular Simplexa™ VZV Direct assay is intended for use on the LIAISON® MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA in cerebrospinal fluid (CSF) from patients signs and/or symptoms of meningitis and/or encephalitis. This test is intended as an aid in the diagnosis of VZV infections of the central nervous system (CNS).
Negative results do not preclude VZV infection and should not be used as the sole basis for treatment or other patient management decisions.
The assay is not intended for use as a donor screening test. The assay is for professional use only.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA double helix in shades of green and blue on the left. To the right of the helix, the word "DiaSorin" is written in a bold, dark blue font. Below "DiaSorin", the word "Molecular" is written in a lighter green font.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 1 of 25
| Applicant | DiaSorin Molecular LLC.11331 Valley View StreetCypress, California 90630USA |
|---|---|
| Establishment Registration No. | 2023365 |
| Contact Person | Sharon YoungPrincipal Regulatory Affairs Specialisttel 562.240.6680fax 562.240.6529Sharon.Young@DiaSorin.com |
| Summary Date | May 3, 2019 |
| Proprietary Name | Simplexa™ VZV Direct and Simplexa™ VZV Positive Control Pack |
| US Product Codes/Names andRegulation Numbers | PLO / Device to detect and identify microbial pathogen nucleic acidsin cerebrospinal fluid / 21 CFR 866.3970OOI / Real Time Nucleic Acid Amplification System / 21 CFR 862.2570 |
| Classification | Class II |
Intended Use
Simplexa™ VZV Direct REF MOL3650
The DiaSorin Molecular Simplexa™ VZV Direct assay is intended for use on the LIAISON® MDX instrument for the qualitative detection of varicella-zoster virus (VZV) DNA in cerebrospinal fluid (CSF) from patients with signs and/or symptoms of meningitis and/or encephalitis. This test is intended as an aid in the diagnosis of VZV infections of the central nervous system (CNS).
Neqative results do not prection and should not be used as the sole basis for treatment or other patient management decisions.
The assay is not intended for use as a donor screening test. The assay is for professional use only.
Simplexa™ VZV Positive Control Pack REF MOL3660
The Simplexa™ VZV Positive Control Pack is intended to be used as a control with the Simplexa™ VZV Direct kit.
This control is not intended for use with other assays or systems.
Device Description
The Simplexa™ VZV Direct assay is a real-time polymerase chain reaction (PCR) system that enables the direct amplification and detection of VZV DNA from unprocessed cerebral spinal fluid (CSF) specimens without nucleic acid extraction. The system consists of the Simplexa™ VZV Direct assay, the LIAISON® MDX (with LIAISON® MDX Studio Software), the Direct Amplification Disc and associated accessories. In the Simplexa™ VZV Direct assay, fluorescent probes are used together with corresponding forward and reverse primers to amplify VZV and internal control targets. A well-conserved region of the VZV DNA polymerase gene is targeted to identify VZV DNA in the specimen. An internal control is used to detect PCR failure and/or inhibition.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix on the left, colored in a gradient from green to blue. To the right of the helix are the words "DiaSorin" in a dark blue, sans-serif font, with the word "Molecular" underneath in a lighter green color. The logo is clean and modern, suggesting a focus on molecular diagnostics or related fields.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 2 of 25
Simplexa™ VZV Direct REF MOL3650
| Component Name | REF | ECSYMBOLONLABEL | AbbreviatedName | CapColor | Numberof Vials | ReactionsperVial/Kit | Volumeper Vial | |
|---|---|---|---|---|---|---|---|---|
| Simplexa™ VZVDirect Reaction Mix | MOL3651 | REAG | C | RM | Purple | 24 | 1/24 | 50 μL |
Simplexa™ VZV Direct Components and Descriptions
| Kit Component | Contents | |||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Simplexa™ VZV Direct Reaction Mix (RM) | DNA polymerase, buffer, dNTPs, template DNA (Internal Control), dye-labeled fluorescent probes and primers specific for detection of VZV and for the DNA Internal Control.Target Probe Fluorophore (Dye) Excitation (nm) Emission (nm) Targeted Gene VZV FAM 495 520 VZV DNA polymerase Internal Control DNA Q670 644 670 DNA (IC) | |||||||||||||||
| Simplexa™ VZV Direct Kit Barcode Card | Assay specific parameters and lot information. |
Simplexa™ VZV Positive Control Pack REF MOL3660 Component and Description
| Component Name | REF | Description | CapColor | Numberof Vials | ReactionsperVial/Kit | Volumeper Vial |
|---|---|---|---|---|---|---|
| Simplexa™ VZV DirectPositive Control | MOL3661 | Inactivatedvaricella zostervirus | Red | 10 | 1/10 | 50 μL |
Materials Supplied Separately
Direct Amplification Disc Kit (REF MOL1455) Direct Amplification Discs for use on the LIAISON® MDX
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a stylized DNA helix in green and blue on the left, with the words "DiaSorin" in blue on the top right. Below "DiaSorin" is the word "Molecular" in green.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 3 of 25
| Comparison to Predicate Device | ||
|---|---|---|
| Comparison toPredicate Device | Predicate Device:BioFire® FilmArray®Meningitis/Encephalitis (ME) PanelK160462 | Candidate Device:Simplexa™ VZV Direct and Simplexa™VZV Positive Control Pack |
| Product Code | PLO | Same |
| RegulationNumber | 21 CFR 866.3970 – Device to detectand identify microbial pathogen nucleicacids in cerebrospinal fluid. | Same |
| OrganismDetected | varicella zoster virus | Same |
| Measurand | DNA from varicella zoster virus | Same |
| Intended Use | The FilmArray Meningitis/Encephalitis(ME) Panel is a qualitative multiplexednucleic acid-based in vitro diagnostictest intended for use with FilmArrayand FilmArray 2.0 systems. TheFilmArray ME Panel is capable ofsimultaneous detection andidentification of multiple bacterial, viral,and yeast nucleic acids directly fromcerebrospinal fluid (CSF) specimensobtained via lumbar puncture fromindividuals with signs and/or symptomsof meningitis and/or encephalitis. Thefollowing organisms are identified usingthe FilmArray ME Panel: Bacteria:Escherichia coli K1 Haemophilusinfluenzae Listeria monocytogenesNeisseria meningitidis (encapsulated)Streptococcus agalactiaeStreptococcus pneumoniae Viruses:Cytomegalovirus Enterovirus Herpessimplex virus 1 Herpes simplex virus 2Human herpesvirus 6 Humanparechovirus Varicella zoster virusYeast: Cryptococcus neoformans/gattiiThe FilmArray ME Panel is indicated asan aid in the diagnosis of specificagents of meningitis and/orencephalitis and results are meant tobe used in conjunction with otherclinical, epidemiological, and laboratorydata. Results from the FilmArray MEPanel are not intended to be used asthe sole basis for diagnosis, treatment, | Simplexa™ VZV Direct REF MOL3650The DiaSorin Molecular Simplexa™ VZVDirect assay is intended for use on theLIAISON® MDX instrument for thequalitative detection of varicella-zostervirus (VZV) DNA in cerebrospinal fluid(CSF) from patients with signs and/orsymptoms of meningitis and/orencephalitis. This test is intended as anaid in the diagnosis of VZV infections ofthe central nervous system (CNS).Negative results do not preclude VZVinfection and should not be used as thesole basis for treatment or other patientmanagement decisions.The assay is not intended for use as adonor screening test. The assay is forprofessional use only.Simplexa™ VZV Positive Control PackREF MOL3660The Simplexa™ VZV Positive ControlPack is intended to be used as a controlwith the Simplexa™ VZV Direct kit.This control is not intended for use withother assays or systems. |
| Comparison toPredicate Device | Predicate Device:BioFire® FilmArray®Meningitis/Encephalitis (ME) PanelK160462 | Candidate Device:Simplexa™ VZV Direct and Simplexa™VZV Positive Control Pack |
| decisions. Positive results do not ruleout co-infection with organisms notincluded in the FilmArray ME Panel.The agent detected may not be thedefinite cause of the disease. Negativeresults do not preclude central nervoussystem (CNS) infection. Not all agentsof CNS infection are detected by thistest and sensitivity in clinical use maydiffer from that described in thepackage insert. The FilmArray MEPanel is not intended for testing ofspecimens collected from indwellingCNS medical devices. The FilmArrayME Panel is intended to be used inconjunction with standard of careculture for organism recovery,serotyping, and antimicrobialsusceptibility testing. | ||
| AutomatedSystem (Sampleto Answer) | Yes | Yes |
| Instrumentation | FilmArray or FilmArray 2.0 system | LIAISON® MDX |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix in green and blue on the left, with the words "DiaSorin" in blue on the top right and "Molecular" in green below it. The logo is clean and modern, with a focus on the company's expertise in molecular diagnostics.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 4 of 25
CLINICAL AGREEMENT
A total of six hundred thirty-seven (637) clinical samples, meeting inclusion criteria were obtained from prospective and prospectively banked collections from eight (8) collection sites from February 2018 to November 2018. The study was conducted using remnant cerebrospinal fluid (CSF) samples from human patients suspected of VZV infection of the central nervous system (CNS).
Due to the low prevalence of VZV infection in CSF samples, two hundred forty (240) contrived samples were included to supplement the number of positive samples were contrived across the clinical range of the Simplexa™ VZV Direct assay and thirty percent (30%) of the samples were contrived at approximately 2X LoD. The samples were stored at 2-8 °C for up to seven (7) days post collection and at <-70 ℃ thereafter. All Simplexa™ VZV Direct testing was performed at the testing sites.
Simplexa™ VZV Direct results were compared to results from a composite reference method. The composite reference method consisted of two (2) validated real-time PCR assays followed by confirmation of positive PCR amplification products with bi-directional sequencing. Samples were characterized as positive if one (1) or both PCR assays were positive and confirmed by bi-directional sequencing. Samples were characterized as negative if both PCR assays were negative. Samples were tested on Simplexa™ VZV Direct at the collection sites and the composite reference method was performed at DiaSorin Molecular, Cypress, CA.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue, with the word "Molecular" underneath in green.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 5 of 25
Table 1 shows the Clinical Agreement of the Simplexa™ VZV Direct as compared to the composite reference method of two (2) validated real-time PCR assays followed by confirmation of positive PCR amplification products with bi-directional sequencing.
| Table 1. Prospective Clinical Agreement Results for Simplexa™ VZV Direct vs Composite |
|---|
| Reference Method |
| Simplexa™ VZV Direct | Composite PCR/Sequencing | ||
|---|---|---|---|
| Results | Detected | Not Detected | Total |
| Detected | 12 | 2a | 14 |
| Not Detected | 0 | 623 | 623 |
| Total | 12 | 625 | 637 |
| PPA100.0% (12/12)95% CI: 75.7% to100.0% | NPA99.7% (623/625)95% CI: 98.8% to99.9% |
3 One (1) of two (2) discordant results were positive with an alternate NAAT used by collection site during routine testing.
Due to the low prevalence of VZV infection in CSF samples, contrived positive samples were tested to supplement the number of positive samples tested. For each of the two strains, VZV 9939 60 samples were contrived across the clinical range of the Simplexa VZV Direct assay with 60% of the samples contrived as low positive samples close to the LoD of the test. Samples were tested at four clinical sites in a blinded manner, randomized with VZV negative samples were stored at 2-8 °C for up to 7 days post collection and at <-70 °C thereafter. Simplexa VZV Direct test results for contrived samples were compared to the same Composite Reference Method described for the prospective clinical study above. The contrived sample study demonstrated a positive percent agreement of 100% (120/120) with a 95% Confidence Interval of 96.9-100.0%.
REPRODUCIBILITY
Reproducibility for the Simplexa™ VZV Direct assay was evaluated. Three (3) investigative sites assessed the device's inter-site, inter-day and inter/intra-assay reproducibility. Each of the laboratories tested Simplexa™ VZV Direct Positive Control, No Template Control (Synthetic CSF), and four (4) contrived samples in negative matrix. Two (2) strains of VZV were used in the study, 9939 and Ellen. The four (4) contrived samples consisted of a low positive (LP) and medium positive (MP) for each VZV strain. The assays were performed in triplicate on five (5) different days. Each site had two (2) operators who each assayed the entire sample panel and Positive Control once per day, for a total of two (2) sets of data per day on a total of six (6) LIAISON® MDX instruments. The combined results for all sites are presented in Tables 2 and 3. The results show the reproducibility of the Simplexa™ VZV Direct % CV to range between 0.5-4.6.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA helix in shades of green and blue on the left. To the right of the helix, the words "DiaSorin" are written in a dark blue, sans-serif font. Below "DiaSorin", the word "Molecular" is written in a lighter green, sans-serif font.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 6 of 25
| Site 1 | Site 2 | Site 3 | Total %AgreementwithExpectedResults | 95% CI | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| SampleVZV Strain(FAM) | %AgreementwithExpectedResults | Avg.Ct | Total%CV | %AgreementwithExpectedResults | Avg.Ct | Total%CV | %AgreementwithExpectedResults | Avg.Ct | Total%CV | ||
| 9939 -LP | 100.0%(30/30) | 35.8 | 4.6 | 100.0%(30/30) | 35.7 | 3.6 | 100.0%(30/30) | 36.1 | 4.2 | 100.0%(90/90) | 95.9% to100.0% |
| 9939 -MP | 100.0%(30/30) | 35.3 | 2.2 | 100.0%(30/30) | 34.6 | 2.4 | 100.0%(30/30) | 35.0 | 2.3 | 100.0%(90/90) | 95.9% to100.0% |
| Ellen -LP | 100.0%(30/30) | 36.7 | 2.6 | 100.0%(30/30) | 36.9 | 2.2 | 100.0%(30/30) | 36.9 | 1.6 | 100.0%(90/90) | 95.9% to100.0% |
| Ellen -MP | 100.0%(30/30) | 35.6 | 1.6 | 100.0%(30/30) | 35.6 | 1.0 | 100.0%(30/30) | 35.6 | 1.1 | 100.0%(90/90) | 95.9% to100.0% |
| PositiveControl | 100.0%(30/30) | 30.3 | 1.1 | 100.0%(30/30) | 29.8 | 1.2 | 100.0%(30/30) | 30.7 | 0.9 | 100.0%(90/90) | 95.9% to100.0% |
| NTC | 100.0%(30/30) | 0.0 | N/A | 100.0%(30/30) | 0.0 | N/A | 100.0%(30/30) | 0.0 | N/A | 100.0%(90/90) | 95.9% to100.0% |
| TotalAgreement | 100.0% (180/180)95% CI: 97.9% to 100.0% | 100.0% (180/180)95% CI: 97.9% to 100.0% | 100.0% (180/180)95% CI: 97.9% to 100.0% | 100.0% (540/540)95% CI: 99.3% to100.0% |
Table 2. Simplexa™ VZV Direct Reproducibility - VZV (FAM)
ANALYTICAL SENSITIVITY/LIMIT OF DETECTION
The Limit of Detection (LoD) was determined for the Simplexa™ VZV Direct assay using quantified stocks of two (2) VZV strains (Ellen and 9939) serially diluted into negative human CSF matrix. The LoD was determined to be the lowest concentration that could be detected positive > 95% of the time.
| Table 3. Simplexa™ VZV Direct Limit of Detection | ||||||
|---|---|---|---|---|---|---|
| -- | -------------------------------------------------- | -- | -- | -- | -- | -- |
| VZV strain | Concentration | |
|---|---|---|
| TCID50/mL | copies/mL | |
| 9939 | 2.03 TCID50/mL | 1,614 |
| Ellen | 0.001 TCID50/mL | 1,505 |
ANALYTICAL REACTIVITY/CROSS REACTIVITY
Analytical Reactivity
The analytical reactivity of the Simplexa™ VZV Direct assay was evaluated using different strains of VZV that were not used in the determination of the limit of detection (LoD) for the assay. Quantified viral material was spiked into negative CSF using a single dilution and assayed in triplicate. The Simplexa™ VZV Direct assay was able to detect other strains of VZV at 2X LoD. The results are presented in Table 4. In addition, in silico BLAST analysis indicated that the assay should detect at least one hundred seventy-eight (178) additional VZV strains.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA double helix in shades of green and blue on the left. To the right of the DNA graphic are the words "DiaSorin" in a dark blue, sans-serif font, stacked above the word "Molecular" in a lighter green, sans-serif font.
510(k) Summary Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 7 of 25
| VZV Strain | Agreement withExpected Results(#Detected/#Total) |
|---|---|
| VZV Strain 82 | 3/3 |
| VZV Strain 275 | 3/3 |
| VZV Strain 1700 | 3/3 |
| VZV Isolate A | 3/3 |
| VZV Isolate B | 3/3 |
Table 4. Simplexa™ VZV Direct Analytical Reactivity
Cross-Reactivity (Analytical Specificity)
The Simplexa™ VZV Direct assay's analytical specificity was evaluated by testing the ability of the assay to exclusively identify VZV virus with no cross-reactivity to organisms that are closely related, or cause similar clinical symptoms or may be present in CSF. One hundred and fifty-nine (159) microorganisms were spiked into negative CSF and assayed in triplicate. Some organisms with low concentrations (<1 x 10° CFU/mL for bacteria, fungi, cells or parasites; <1 x 105 IU/mL or PFU/mL or TCIDs/mL for viruses) were additionally tested in silico. For organisms not available to be tested, in silico analysis was performed. No cross-reactivity was observed. The results are presented in Table 5.
| Table 5. Simplexa™ VZV Direct Cross-Reactivity (Analytical Specificity) | |||
|---|---|---|---|
| No. | Organism | Tested Concentration | Agreement with ExpectedResults: % Detection(# Detected/#Tested) |
|---|---|---|---|
| 1 | Adenovirus A12** | NA | NA |
| 2 | Adenovirus B35** | NA | NA |
| 3 | Adenovirus B7A | 1 x 105 IU/mL | 0.0% (0/3) |
| 4 | Adenovirus C1 | 1 x 105 IU/mL | 0.0% (0/3) |
| 5 | Adenovirus C2 | 1 x 105 IU/mL | 0.0% (0/3) |
| 6 | Adenovirus D20 | 1 x 105 IU/mL | 0.0% (0/3) |
| 7 | Adenovirus E4** | NA | NA |
| 8 | Adenovirus F41** | NA | NA |
| 9 | Aspergillus fumigatus | 1 x 106 CFU/mL | 0.0% (0/3) |
| 10 | Bacillus cereus | 1 x 106 CFU/mL | 0.0% (0/3) |
| 11 | Bacillus subtilis | 1 x 106 CFU/mL | 0.0% (0/3) |
| 12 | BK virus | 1 x 105 IU/mL | 0.0% (0/3) |
| 13 | Candida albicans | 1 x 106 CFU/mL | 0.0% (0/3) |
| No. | Organism | Tested Concentration | Agreement with ExpectedResults: % Detection(# Detected/#Tested) |
| 14 | Candida krusei | 1 x 106 CFU/mL | 0.0% (0/3) |
| 15 | Candida parapsilosis | 1 x 106 CFU/mL | 0.0% (0/3) |
| 16 | Candida tropicalis | 1 x 106 CFU/mL | 0.0% (0/3) |
| 17 | Citrobacter freundii | 1 x 106 CFU/mL | 0.0% (0/3) |
| 18 | Citrobacter koseri | 1 x 106 CFU/mL | 0.0% (0/3) |
| 19 | Coronavirus 229E* | 1 x 104 IU/mL* | 0.0% (0/3) |
| 20 | Coronavirus NL63* | 1 x 104 IU/mL* | 0.0% (0/3) |
| 21 | Coronavirus OC43 | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 22 | Corynebacterium striatum** | NA | NA |
| 23 | Corynebacteriumurealyticum** | NA | NA |
| 24 | Coxsackievirus A10** | NA | NA |
| 25 | Coxsackievirus A16 | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 26 | Coxsackievirus A17** | NA | NA |
| 27 | Coxsackievirus A21* | 1 x 104 TCID50/mL* | 0.0% (0/3) |
| 28 | Coxsackievirus A24** | NA | NA |
| 29 | Coxsackievirus A6** | NA | NA |
| 30 | Coxsackievirus A9 | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 31 | Coxsackievirus B1 | 1 x 105 IU/mL | 0.0% (0/3) |
| 32 | Coxsackievirus B2 | 1 x 105 IU/mL | 0.0% (0/3) |
| 33 | Coxsackievirus B3 | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 34 | Coxsackievirus B4 | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 35 | Coxsackievirus B5* | 1 x 104 TCID50/mL* | 0.0% (0/3) |
| 36 | Cronobacter sakazakii | 1 x 106 CFU/mL | 0.0% (0/3) |
| 37 | Cryptococcus albidus** | NA | NA |
| 38 | Cryptococcus amylolentus** | NA | NA |
| 39 | Cryptococcus gattii** | NA | NA |
| 40 | Cryptococcus laurentii** | NA | NA |
| No. | Organism | Tested Concentration | Agreement with ExpectedResults: % Detection(# Detected/#Tested) |
| 41 | Cryptococcus neoformans | 1 x 106 CFU/mL | 0.0% (0/3) |
| 42 | Cryptococcus uniguttulatus** | NA | NA |
| 43 | Cytomegalovirus (AD169Strain) | 1 x 105 IU/mL | 0.0% (0/3) |
| 44 | Dengue virus (Type 1)* | 1 x 104 IU/mL* | 0.0% (0/3) |
| 45 | Dengue virus (Type 2) | 1 x 105 IU/mL | 0.0% (0/3) |
| 46 | Echovirus 1 | 1 x 105 IU/mL | 0.0% (0/3) |
| 47 | Echovirus 11 | 1 x 105 IU/mL | 0.0% (0/3) |
| 48 | Echovirus 18** | NA | NA |
| 49 | Echovirus 4 | 1 x 105 IU/mL | 0.0% (0/3) |
| 50 | Echovirus 6 | 1 x 105 IU/mL | 0.0% (0/3) |
| 51 | Echovirus 7 | 1 x 105 IU/mL | 0.0% (0/3) |
| 52 | Echovirus 9 | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 53 | Encephalomyocarditis virus | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 54 | Enterobacter aerogenes | 1 x 106 CFU/mL | 0.0% (0/3) |
| 55 | Enterobacter cloacae | 1 x 106 CFU/mL | 0.0% (0/3) |
| 56 | Enterovirus 68** | NA | NA |
| 57 | Enterovirus 70 | 1 x 105 IU/mL | 0.0% (0/3) |
| 58 | Enterovirus 71 | 1 x 105 IU/mL | 0.0% (0/3) |
| 59 | Epstein-Barr virus (B95-8Strain) | 1 x 105 copies/mL | 0.0% (0/3) |
| 60 | Escherichia coli (O157:H7) | 1 x 106 CFU/mL | 0.0% (0/3) |
| 61 | Escherichia coli K1** | NA | NA |
| 62 | Escherichia fergusonii** | NA | NA |
| 63 | Escherichia hermanii** | NA | NA |
| 64 | Escherichia vulneris** | NA | NA |
| 65 | Filobasidium capsuligenum** | NA | NA |
| 66 | Haemophilus ducreyi | 1 x 106 CFU/mL | 0.0% (0/3) |
| 67 | Haemophilus haemolyticus** | NA | NA |
| No. | Organism | Tested Concentration | Agreement with ExpectedResults: % Detection(# Detected/#Tested) |
| 68 | Haemophilus influenza TypeB | 1 x 106 CFU/mL | 0.0% (0/3) |
| 69 | Haemophilus influenzae TypeA | 1 x 106 CFU/mL | 0.0% (0/3) |
| 70 | Haemophilusparahaemolyticus** | NA | NA |
| 71 | Haemophilus parainfluenzae | 1 x 106 CFU/mL | 0.0% (0/3) |
| 72 | Hepatitis A virus | 1 x 105 IU/mL | 0.0% (0/3) |
| 73 | Hepatitis B virus | 1 x 105 IU/mL | 0.0% (0/3) |
| 74 | Hepatitis C virus | 1 x 105 IU/mL | 0.0% (0/3) |
| 75 | Hepatitis D virus** | NA | NA |
| 76 | HHV-6A | 1 x 105 copies/mL | 0.0% (0/3) |
| 77 | HHV-6B | 1 x 105 copies/mL | 0.0% (0/3) |
| 78 | HHV-7SB | 1 x 105 IU/mL | 0.0% (0/3) |
| 79 | HHV-8 | 1 x 105 copies/mL | 0.0% (0/3) |
| 80 | HIV-1 IIIB | 1 x 105 IU/mL | 0.0% (0/3) |
| 81 | HIV-2 NIHZ | 1 x 105 IU/mL | 0.0% (0/3) |
| 82 | HPeV-1** | NA | NA |
| 83 | HPeV-2** | NA | NA |
| 84 | HPeV-3 | 1 x 105 IU/mL | 0.0% (0/3) |
| 85 | HPeV-4** | NA | NA |
| 86 | HPeV-5** | NA | NA |
| 87 | HPeV-6** | NA | NA |
| 88 | HSV-1 (MacIntyre) | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 89 | HSV-2 (G) | 1 x 105 IU/mL | 0.0% (0/3) |
| 90 | Human Rhinovirus A16 | 1 x 105 IU/mL | 0.0% (0/3) |
| 91 | Human Rhinovirus B3** | NA | NA |
| 92 | Human Rhinovirus B83** | NA | NA |
| 93 | Influenza A/California/7/2009 | 1 x 105 IU/mL | 0.0% (0/3) |
| 94 | Influenza A/H1N1 | 1 x 105 TCID50/mL | 0.0% (0/3) |
| No. | Organism | Tested Concentration | Agreement with ExpectedResults: % Detection(# Detected/#Tested) |
| 95 | Influenza B/Florida/02/2006 | 1 x 105 IU/mL | 0.0% (0/3) |
| 96 | JC virus (MAD-4 strain) | 1 x 105 IU/mL | 0.0% (0/3) |
| 97 | Klebsiella pneumoniae | 1 x 106 CFU/mL | 0.0% (0/3) |
| 98 | La Crosse encephalitis virus | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 99 | Listeria innocua** | NA | NA |
| 100 | Listeria ivanovii** | NA | NA |
| 101 | Listeria monocytogenes | 1 x 106 CFU/mL | 0.0% (0/3) |
| 102 | Measles virus | 1 x 105 IU/mL | 0.0% (0/3) |
| 103 | Morganella morganii** | NA | NA |
| 104 | Mumps virus | 1 x 105 IU/mL | 0.0% (0/3) |
| 105 | Mycobacterium tuberculosisgenomic DNA | 1.6 x 106 copies/mL | 0.0% (0/3) |
| 106 | Naegleria fowleri* | 1.8 x 105 cells/mL* | 0.0% (0/3) |
| 107 | Neisseria gonorrhoeae | 1 x 106 CFU/mL | 0.0% (0/3) |
| 108 | Neisseria lactamica** | NA | NA |
| 109 | Neisseria meningitidis(serogroup A) | 1 x 106 CFU/mL | 0.0% (0/3) |
| 110 | Neisseria meningitidis(unencapsulated)** | NA | NA |
| 111 | Neisseria mucosa | 1 x 106 CFU/mL | 0.0% (0/3) |
| 112 | Neisseria sicca** | NA | NA |
| 113 | Pantoea agglomerans** | NA | NA |
| 114 | Parainfluenza Type 1 | 1 x 105 IU/mL | 0.0% (0/3) |
| 115 | Parainfluenza Type 2 | 1 x 105 IU/mL | 0.0% (0/3) |
| 116 | Parainfluenza Type 3 | 1 x 105 IU/mL | 0.0% (0/3) |
| 117 | Parainfluenza Type 4 | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 118 | Parvovirus B19 | 1 x 105 IU/mL | 0.0% (0/3) |
| 119 | Poliovirus (Type 3) | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 120 | Propionibacterium acnes | 1 x 106 CFU/mL | 0.0% (0/3) |
| 121 | Proteus mirabilis Z050 | 1 x 106 CFU/mL | 0.0% (0/3) |
| No. | Organism | Tested Concentration | Agreement with ExpectedResults: % Detection(# Detected/#Tested) |
| 122 | Pseudomonas aeruginosa | 1 x 106 CFU/mL | 0.0% (0/3) |
| 123 | Rabies virus | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 124 | Respiratory syncytial virus | 1 x 105 IU/mL | 0.0% (0/3) |
| 125 | Rhinovirus 1A | 1 x 105 IU/mL | 0.0% (0/3) |
| 126 | Rotavirus (Type Wa)* | 1 x 104 IU/mL* | 0.0% (0/3) |
| 127 | Rubella virus | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 128 | Salmonella bongori** | NA | NA |
| 129 | Salmonella enterica** | NA | NA |
| 130 | Serratia marcescens | 1 x 106 CFU/mL | 0.0% (0/3) |
| 131 | Shigella boydii** | NA | NA |
| 132 | Shigella flexneri | 1 x 106 CFU/mL | 0.0% (0/3) |
| 133 | Shigella sonnei** | NA | NA |
| 134 | Simian Virus type 40 | 1 x 105 IU/mL | 0.0% (0/3) |
| 135 | St. Louis encephalitis virus | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 136 | Staphylococcus aureus(MRSA), COL | 1 x 106 CFU/mL | 0.0% (0/3) |
| 137 | Staphylococcus capitis ** | NA | NA |
| 138 | Staphylococcus epidermidis(MRSE) | 1 x 106 CFU/mL | 0.0% (0/3) |
| 139 | Staphylococcushaemolyticus** | NA | NA |
| 140 | Staphylococcus hominis ** | NA | NA |
| 141 | Staphylococcuslugdunensis ** | NA | NA |
| 142 | Staphylococcussaprophyticus | 1 x 106 CFU/mL | 0.0% (0/3) |
| 143 | Streptococcus agalactiae | 1 x 106 CFU/mL | 0.0% (0/3) |
| 144 | Streptococcus anginosus ** | NA | NA |
| 145 | Streptococcus bovis ** | NA | NA |
| 146 | Streptococcus dysgalactiae | 1 x 106 CFU/mL | 0.0% (0/3) |
| 147 | Streptococcus intermedius | 1 x 106 CFU/mL | 0.0% (0/3) |
| 148 | Streptococcus mutans | 1 x 106 CFU/mL | 0.0% (0/3) |
| No. | Organism | Tested Concentration | Agreement with ExpectedResults: % Detection(# Detected/#Tested) |
| 149 | Streptococcus oralis** | NA | NA |
| 150 | Streptococcus pneumoniae | 1 x 106 CFU/mL | 0.0% (0/3) |
| 151 | Streptococcuspseudopneumoniae** | NA | NA |
| 152 | Streptococcus pyogenesZ018 | 1 x 106 CFU/mL | 0.0% (0/3) |
| 153 | Streptococcus salivarius | 1 x 106 CFU/mL | 0.0% (0/3) |
| 154 | Streptococcus sanguinis** | NA | NA |
| 155 | Toxoplasma gondii | 1 x 106 tachyzoites/mL | 0.0% (0/3) |
| 156 | Treponema pallidum** | NA | NA |
| 157 | Tropheryma whipplei** | NA | NA |
| 158 | West Nile virus | 1 x 105 TCID50/mL | 0.0% (0/3) |
| 159 | White Blood Cells (HumanGenomic DNA) | 1 x 106 cells/mL | 0.0% (0/3) |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a green and blue DNA helix on the left, with the words "DiaSorin" in blue and "Molecular" in green on the right. The logo is simple and modern, and it is likely used to represent the company's focus on molecular diagnostics.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 8 of 25
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue, with "Molecular" underneath in a lighter green color. The logo is clean and modern, suggesting a focus on biotechnology and molecular diagnostics.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 9 of 25
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue, with "Molecular" underneath in a lighter green color. The logo is clean and modern, suggesting a focus on biotechnology and molecular diagnostics.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 10 of 25
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue on the top line and "Molecular" in light green on the bottom line.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 11 of 25
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue, with "Molecular" underneath in green.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 12 of 25
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in a dark blue, sans-serif font, stacked above the word "Molecular" in a lighter green, sans-serif font.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 13 of 25
- Additional testing was performed in silico due to low concentration of stock and also found not to be cross reactive with VZV.
** Tested in silico due to unavailability of the organism.
INTERFERENCE
The performance of the Simplexa™ VZV Direct assay was evaluated with potentially interfering substances that may be present in CSF samples at the concentrations indicated in the table below. A total of sixteen (16) potentially interfering substances were tested in a low positive VZV sample at approximately 2 times LoD (based on ≥ 95% detection rate) in CSF matrix and assayed in triplicate. No interference was observed at the interferent concentrations indicated. The results are presented in Table 6.
| No. | Potential Interferent | VZV Strain | InterferentConcentration | Agreement withExpected Results(#Detected/#Total) |
|---|---|---|---|---|
| 1 | Acyclovir | Ellen | 9.4 mg/mL | 100.0% (3/3) |
| 9939 | 9.4 mg/mL | 100.0% (3/3) | ||
| 2 | Albumin | Ellen | 50 mg/mL | 100.0% (3/3) |
| 9939 | 50 mg/mL | 100.0% (3/3) | ||
| 3 | Bilirubin | Ellen | 0.0125 mg/mL | 100.0% (3/3) |
| 9939 | 0.0125 mg/mL | 100.0% (3/3) | ||
| 4 | Casein | Ellen | 9.0 mg/mL | 100.0% (3/3) |
| 9939 | 9.0 mg/mL | 100.0% (3/3) | ||
| No. | Potential Interferent | VZV Strain | InterferentConcentration | Agreement withExpected Results(#Detected/#Total) |
| 5 | Foscarnet | Ellen | 0.6 mg/mL | 100.0% (3/3) |
| 9939 | 0.6 mg/mL | 100.0% (3/3) | ||
| 6 | Gamma globulin | Ellen | 10.4 mg/mL | 100.0% (3/3) |
| 9939 | 10.4 mg/mL | 100.0% (3/3) | ||
| 7 | Glucose | Ellen | 11 mg/mL | 100.0% (3/3) |
| 9939 | 11 mg/mL | 100.0% (3/3) | ||
| 8 | Hemoglobin | Ellen | 3.5 mg/mL | 100.0% (3/3) |
| 9939 | 3.5 mg/mL | 100.0% (3/3) | ||
| 9 | Human Genomic DNA | Ellen | 72 µg/mL | 100.0% (3/3) |
| 9939 | 72 µg/mL | 100.0% (3/3) | ||
| 10 | Immunoglobulin | Ellen | 10 mg/mL | 100.0% (3/3) |
| 9939 | 10 mg/mL | 100.0% (3/3) | ||
| 11 | Lactate | Ellen | 2.2 mg/mL | 100.0% (3/3) |
| 9939 | 2.2 mg/mL | 100.0% (3/3) | ||
| 12 | Topical Antiseptic | Ellen | 5% (v/v) | 100.0% (3/3) |
| 9939 | 5% (v/v) | 100.0% (3/3) | ||
| 13 | Trans-Isolate Medium | Ellen | 50% (v/v) | 100.0% (3/3) |
| 9939 | 50% (v/v) | 100.0% (3/3) | ||
| 14 | UTM | Ellen | 50% (v/v) | 100.0% (3/3) |
| 9939 | 50% (v/v) | 100.0% (3/3) | ||
| 15 | White blood cells | Ellen | 2x107 WBC/mL | 100.0% (3/3) |
| 9939 | 2x107 WBC/mL | 100.0% (3/3) | ||
| 16 | Whole Blood in EDTA | Ellen | 10% (v/v) | 100.0% (3/3) |
| 9939 | 10% (v/v) | 100.0% (3/3) | ||
| No. | VZV Strain | Organism | Tested Concentration | Agreement withExpected Results:(# Detected/#Tested) |
| 1 | NA | Adenovirus A12** | NA | NA |
| 2 | NA | Adenovirus B35** | NA | NA |
| 3 | Ellen | Adenovirus B7A | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 4 | Ellen | Adenovirus C1 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 5 | Ellen | Adenovirus C2 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 88.9% (8/9) | ||
| 6 | Ellen | Adenovirus D20 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 7 | NA | Adenovirus E4** | NA | NA |
| 8 | NA | Adenovirus F41** | NA | NA |
| 9 | Ellen | Aspergillus fumigatus | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 10 | Ellen | Bacillus cereus | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 11 | Ellen | Bacillus subtilis | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 12 | Ellen | BK virus | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 13 | Ellen | Candida albicans | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 14 | Ellen | Candida krusei | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 15 | Ellen | Candida parapsilosis | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| No. | VZV Strain | Organism | Tested Concentration | Agreement withExpected Results:(# Detected/#Tested) |
| 16 | Ellen | Candida tropicalis | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 17 | Ellen | Citrobacter freundii | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 18 | Ellen | Citrobacter koseri | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 19 | Ellen | Coronavirus 229E | 1 x 104 IU/mL* | 100.0% (3/3) |
| 9939 | 1 x 104 IU/mL* | 100.0% (3/3) | ||
| 20 | Ellen | Coronavirus NL63 | 1 x 104 IU/mL* | 100.0% (3/3) |
| 9939 | 1 x 104 IU/mL* | 100.0% (3/3) | ||
| 21 | Ellen | Coronavirus OC43 | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 22 | NA | Corynebacteriumstriatum** | NA | NA |
| 23 | NA | Corynebacteriumurealyticum** | NA | NA |
| 24 | NA | Coxsackievirus A10** | NA | NA |
| 25 | Ellen | Coxsackievirus A16 | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 26 | NA | Coxsackievirus A17** | NA | NA |
| 27 | Ellen | Coxsackievirus A21 | 1 x 104 TCID50/mL* | 100.0% (3/3) |
| 9939 | 1 x 104 TCID50/mL* | 100.0% (3/3) | ||
| 28 | NA | Coxsackievirus A24** | NA | NA |
| 29 | NA | Coxsackievirus A6** | NA | NA |
| 30 | Ellen | Coxsackievirus A9 | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 31 | Ellen | Coxsackievirus B1 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 32 | Ellen | Coxsackievirus B2 | 1 x 105 IU/mL | 100.0% (3/3) |
| No. | VZV Strain | Organism | Tested Concentration | Agreement withExpected Results:(# Detected/#Tested) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 33 | Ellen | Coxsackievirus B3 | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 34 | Ellen | Coxsackievirus B4 | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 35 | Ellen | Coxsackievirus B5 | 1 x 104 TCID50/mL* | 100.0% (3/3) |
| 9939 | 1 x 104 TCID50/mL* | 100.0% (3/3) | ||
| 36 | Ellen | Cronobacter sakazakii | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 37 | NA | Cryptococcus albidus** | NA | NA |
| 38 | NA | Cryptococcusamylolentus** | NA | NA |
| 39 | NA | Cryptococcus gattii** | NA | NA |
| 40 | NA | Cryptococcus laurentii** | NA | NA |
| 41 | Ellen | Cryptococcusneoformans | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 42 | NA | Cryptococcusuniguttulatus** | NA | NA |
| 43 | Ellen | Cytomegalovirus(AD169 Strain) | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 44 | Ellen | Dengue virus (Type 1) | 1 x 104 UI/mL* | 100.0% (3/3) |
| 9939 | 1 x 104 IU/mL* | 100.0% (3/3) | ||
| 45 | Ellen | Dengue virus (Type 2) | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 46 | Ellen | Echovirus 1 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 47 | Ellen | Echovirus 11 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 48 | NA | Echovirus 18** | NA | NA |
| No. | VZV Strain | Organism | Tested Concentration | Agreement withExpected Results:(# Detected/#Tested) |
| 49 | Ellen | Echovirus 4 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 50 | Ellen | Echovirus 6 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 51 | Ellen | Echovirus 7 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 52 | Ellen | Echovirus 9 | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 53 | Ellen | Encephalomyocarditisvirus | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 54 | Ellen | Enterobacter aerogenes | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 55 | Ellen | Enterobacter cloacae | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 56 | NA | Enterovirus 68** | NA | NA |
| 57 | Ellen | Enterovirus 70 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 58 | Ellen | Enterovirus 71 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 59 | Ellen | Epstein-Barr virus (B95-8 Strain) | 1 x 105 copies/mL | 100.0% (3/3) |
| 9939 | 1 x 105 copies/mL | 100.0% (3/3) | ||
| 60 | Ellen | Escherichia coli(O157:H7) | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 61 | NA | Escherichia coli K1** | NA | NA |
| 62 | NA | Escherichia fergusonii** | NA | NA |
| 63 | NA | Escherichia hermanii** | NA | NA |
| 64 | NA | Escherichia vulneris ** | NA | NA |
| 65 | NA | Filobasidium | NA | NA |
| No. | VZV Strain | Organism | Tested Concentration | Agreement withExpected Results:(# Detected/#Tested) |
| capsuligenum** | ||||
| 66 | Ellen | Haemophilus ducreyi | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 67 | NA | Haemophilushaemolyticus** | NA | NA |
| 68 | Ellen | Haemophilus influenzaeType A | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 69 | Ellen | Haemophilus influenzaeType B | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 70 | NA | Haemophilusparahaemolyticus** | NA | NA |
| 71 | Ellen | Haemophilusparainfluenzae | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 72 | Ellen | Hepatitis A virus | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 73 | Ellen | Hepatitis B virus | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 74 | Ellen | Hepatitis C virus | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 75 | NA | Hepatitis D virus** | NA | NA |
| 76 | Ellen | HHV-6A | 1 x 105 copies/mL | 100.0% (3/3) |
| 9939 | 1 x 105 copies/mL | 100.0% (3/3) | ||
| 77 | Ellen | HHV-6B | 1 x 105 copies/mL | 100.0% (3/3) |
| 9939 | 1 x 105 copies/mL | 100.0% (3/3) | ||
| 78 | Ellen | HHV-7 SB | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 88.9% (8/9) | ||
| 79 | Ellen | HHV-8 | 1 x 105 copies/mL | 100.0% (3/3) |
| 9939 | 1 x 105 copies/mL | 100.0% (3/3) | ||
| 80 | Ellen | HIV-1 IIIB | 1 x 105 IU/mL | 100.0% (3/3) |
| No. | VZV Strain | Organism | Tested Concentration | Agreement withExpected Results:(# Detected/#Tested) |
| 9939 | 1 x 105 IU/mL | 88.9% (8/9) | ||
| 81 | Ellen | HIV-2 NIHZ | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 82 | NA | HPeV-1** | NA | NA |
| 83 | NA | HPeV-2** | NA | NA |
| Ellen | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 84 | 9939 | HPeV-3 | 1 x 105 IU/mL | 100.0% (3/3) |
| 85 | NA | HPeV-4** | NA | NA |
| 86 | NA | HPeV-5** | NA | NA |
| 87 | NA | HPeV-6** | NA | NA |
| Ellen | HSV-1 (MacIntyre) | 1 x 105 TCID50/mL | 100.0% (3/3) | |
| 88 | 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | |
| Ellen | HSV-2 (G) | 1 x 105 IU/mL | 100.0% (3/3) | |
| 89 | 9939 | 1 x 105 IU/mL | 100.0% (3/3) | |
| Ellen | Human Rhinovirus A16 | 1 x 105 IU/mL | 100.0% (3/3) | |
| 90 | 9939 | 1 x 105 IU/mL | 100.0% (3/3) | |
| 91 | NA | Human Rhinovirus B3** | NA | NA |
| 92 | NA | Human Rhinovirus B83** | NA | NA |
| Ellen | InfluenzaA/California/7/2009 | 1 x 105 IU/mL | 100.0% (3/3) | |
| 93 | 9939 | 1 x 105 IU/mL | 100.0% (3/3) | |
| Ellen | Influenza A/H1N1 | 1 x 105 TCID50/mL | 100.0% (3/3) | |
| 94 | 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | |
| Ellen | InfluenzaB/Florida/02/2006 | 1 x 105 IU/mL | 100.0% (3/3) | |
| 95 | 9939 | 1 x 105 IU/mL | 100.0% (3/3) | |
| Ellen | JC virus (MAD-4 strain) | 1 x 105 IU/mL | 100.0% (3/3) | |
| 96 | 9939 | 1 x 105 IU/mL | 100.0% (3/3) | |
| Ellen | Klebsiella pneumoniae | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 97 | 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | |
| No. | VZV Strain | Organism | Tested Concentration | Agreement withExpected Results:(# Detected/#Tested) |
| 98 | Ellen | La Crosse encephalitisvirus | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 99 | NA | Listeria innocua** | NA | NA |
| 100 | NA | Listeria ivanovii** | NA | NA |
| 101 | Ellen | Listeria monocytogenes | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 102 | Ellen | Measles virus | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 103 | NA | Morganella morganii** | NA | NA |
| 104 | Ellen | Mumps virus | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 105 | Ellen | Mycobacteriumtuberculosis genomicDNA | 1.6 x 106 copies/mL | 100.0% (3/3) |
| 9939 | 1.6 x 106 copies/mL | 100.0% (3/3) | ||
| 106 | Ellen | Naegleria fowleri | 1.78 x 105 cells/mL* | 100.0% (3/3) |
| 9939 | 1.78 x 105 cells/mL* | 100.0% (3/3) | ||
| 107 | Ellen | Neisseria gonorrhoeae | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 108 | NA | Neisseria lactamica** | NA | NA |
| 109 | Ellen | Neisseria meningitidis(serogroup A) | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 110 | NA | Neisseria meningitidis(unencapsulated)** | NA | NA |
| 111 | Ellen | Neisseria mucosa | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 112 | NA | Neisseria sicca** | NA | NA |
| 113 | NA | Pantoea agglomerans** | NA | NA |
| 114 | Ellen | Parainfluenza Type 1 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| No. | VZV Strain | Organism | Tested Concentration | Agreement withExpected Results:(# Detected/#Tested) |
| 115 | Ellen | Parainfluenza Type 2 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 116 | Ellen | Parainfluenza Type 3 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 117 | Ellen | Parainfluenza Type 4 | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 118 | Ellen | Parvovirus B19 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 119 | Ellen | Poliovirus (Type 3) | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 120 | Ellen | Propionibacterium acnes | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 121 | Ellen | Proteus mirabilis Z050 | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 122 | Ellen | Pseudomonasaeruginosa | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | 1 x 106 CFU/mL | 100.0% (3/3) | ||
| 123 | Ellen | Rabies virus | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 124 | Ellen | Respiratory syncytialvirus | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 125 | Ellen | Rhinovirus 1A | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | 1 x 105 IU/mL | 100.0% (3/3) | ||
| 126 | Ellen | Rotavirus (Type Wa) | 1 x 104 IU/mL* | 100.0% (3/3) |
| 9939 | 1 x 104 IU/mL* | 100.0% (3/3) | ||
| 127 | Ellen | Rubella virus | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | 1 x 105 TCID50/mL | 100.0% (3/3) | ||
| 128 | NA | Salmonella bongori** | NA | NA |
| 129 | NA | Salmonella enterica** | NA | NA |
| No. | VZV Strain | Organism | Tested Concentration | Agreement withExpected Results:(# Detected/#Tested) |
| 130 | Ellen | Serratia marcescens | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Serratia marcescens | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 131 | NA | Shigella boydii** | NA | NA |
| 132 | Ellen | Shigella flexneri | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Shigella flexneri | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 133 | NA | Shigella sonnei** | NA | NA |
| 134 | Ellen | Simian Virus type 40 | 1 x 105 IU/mL | 100.0% (3/3) |
| 9939 | Simian Virus type 40 | 1 x 105 IU/mL | 100.0% (3/3) | |
| 135 | Ellen | St. Louis encephalitis virus | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | St. Louis encephalitis virus | 1 x 105 TCID50/mL | 100.0% (3/3) | |
| 136 | Ellen | Staphylococcus aureus (MRSA), COL | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Staphylococcus aureus (MRSA), COL | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 137 | NA | Staphylococcus capitis** | NA | NA |
| 138 | Ellen | Staphylococcus epidermidis (MRSE) | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Staphylococcus epidermidis (MRSE) | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 139 | NA | Staphylococcus haemolyticus** | NA | NA |
| 140 | NA | Staphylococcus hominis** | NA | NA |
| 141 | NA | Staphylococcus lugdunensis** | NA | NA |
| 142 | Ellen | Staphylococcus saprophyticus | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Staphylococcus saprophyticus | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 143 | Ellen | Streptococcus agalactiae | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Streptococcus agalactiae | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 144 | NA | Streptococcus anginosus** | NA | NA |
| 145 | NA | Streptococcus bovis** | NA | NA |
| 146 | Ellen | Streptococcus dysgalactiae | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Streptococcus dysgalactiae | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 147 | Ellen | Streptococcus | 1 x 106 CFU/mL | 100.0% (3/3) |
| No. | VZV Strain | Organism | Tested Concentration | Agreement withExpected Results:(# Detected/#Tested) |
| 148 | 9939 | intermedius | 1 x 106 CFU/mL | 100.0% (3/3) |
| 149 | Ellen | Streptococcus mutans | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Streptococcus mutans | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 150 | NA | Streptococcus oralis ** | NA | NA |
| 151 | Ellen | Streptococcus pneumoniae | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Streptococcus pneumoniae | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 152 | NA | Streptococcus pseudopneumoniae ** | NA | NA |
| 153 | Ellen | Streptococcus pyogenesZ018 | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Streptococcus pyogenesZ018 | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 154 | Ellen | Streptococcus salivarius | 1 x 106 CFU/mL | 100.0% (3/3) |
| 9939 | Streptococcus salivarius | 1 x 106 CFU/mL | 100.0% (3/3) | |
| 155 | NA | Streptococcus sanguinis ** | NA | NA |
| 156 | Ellen | Toxoplasma gondii | 1 x 106 Tachyzoites/mL | 100.0% (3/3) |
| 9939 | Toxoplasma gondii | 1 x 106 Tachyzoites/mL | 100.0% (3/3) | |
| 157 | NA | Treponema pallidum ** | NA | NA |
| 158 | NA | Tropheryma whipplei ** | NA | NA |
| 159 | Ellen | West Nile virus | 1 x 105 TCID50/mL | 100.0% (3/3) |
| 9939 | West Nile virus | 1 x 105 TCID50/mL | 100.0% (3/3) | |
| 160 | Ellen | White blood cells(Human genomic DNA) | 1 x 106 cells/mL | 100.0% (3/3) |
| 9939 | White blood cells(Human genomic DNA) | 1 x 106 cells/mL | 100.0% (3/3) |
Table 6. Simplexa™ VZV Direct Interference
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue, with "Molecular" underneath in green. The logo is clean and modern, suggesting a focus on biotechnology and molecular diagnostics.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 14 of 25
INHIBITION BY OTHER MICROORGANISMS
The Simplexa™ VZV Direct assay was evaluated by testing the ability to identify VZV virus when other potentially inhibitory organisms are present. The panel of one hundred and sixty (160) potentially inhibitory organisms was individually spiked into a pool with a low concentration at approximately 2 times LoD (based on ≥ 95% detection rate) in CSF. No inhibition by other organisms was observed for VZV at the concentrations indicated in Table 7.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue, with "Molecular" underneath in green. The logo is clean and modern, suggesting a company focused on molecular diagnostics or biotechnology.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 15 of 25
Table 7. Simplexa™ VZV Direct Interference
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a DNA helix graphic on the left, with the text "DiaSorin" in blue to the right of the helix. Below "DiaSorin" is the word "Molecular" in green.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 16 of 25
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a DNA helix graphic on the left, with the text "DiaSorin" in blue to the right of the helix. Below "DiaSorin" is the word "Molecular" in green.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 17 of 25
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image contains the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue above the word "Molecular" in green.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 18 of 25
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the logo for DiaSorin Molecular. The word "DiaSorin" is written in dark blue, and the word "Molecular" is written in green. To the left of the words is a green and blue DNA helix.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 19 of 25
{22}------------------------------------------------
Image /page/22/Picture/0 description: The image contains the logo for DiaSorin Molecular. The word "DiaSorin" is written in a dark blue font, and the word "Molecular" is written in a light green font. To the left of the text is a green and blue DNA helix.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 20 of 25
{23}------------------------------------------------
Image /page/23/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a DNA helix graphic on the left, with the text "DiaSorin" in blue to the right of the helix. Below "DiaSorin" is the word "Molecular" in green.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 21 of 25
{24}------------------------------------------------
Image /page/24/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a DNA helix graphic on the left, with the text "DiaSorin" in blue to the right of the helix. Below "DiaSorin" is the word "Molecular" in green.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 22 of 25
{25}------------------------------------------------
Image /page/25/Picture/0 description: The image contains the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue above the word "Molecular" in green.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 23 of 25
{26}------------------------------------------------
Image /page/26/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in a dark blue, sans-serif font, with the word "Molecular" underneath in a lighter green color.
510(k) Summary
Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 24 of 25
- Any organisms with low titer stock were analyzed in silico by BLAST (Basic Local Alignment Search Tool, NCBI, NIH). ** Tested in silico due to unavailability of the organism.
CARRY-OVER CONTAMINATION
An amplification carry-over for the Simplexa™ assays has been assessed. The study was designed by alternately placing high positive and negative samples on each disc. No evidence of carry-over contamination was observed.
EXPECTED VALUES
The prevalence of VZV as determined by the Simplexa™ VZV Direct assay in a multi-site clinical study with prospectively collected specimens is shown in Table 8 below. The expected values for VZV per site varied between 0% and 7.4%. The overall prevalence for all sites was 2.2% (fourteen of six hundred thirty-seven (14/637).
{27}------------------------------------------------
Image /page/27/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue, with the word "Molecular" underneath in light green. The logo is clean and modern, suggesting a company focused on molecular diagnostics or related fields.
510(k) Summary Simplexa™ VZV Direct Catalog No. MOL3650 Simplexa™ VZV Positive Control Pack Catalog No. MOL3660 May 3, 2019 Page 25 of 25
Table 8. Prospective VZV prevalence as determined by Simplexa™ VZV Direct
| Site ID | Total | Simplexa™ VZV DirectPositive Samples | VZV Prevalence |
|---|---|---|---|
| 1 | 199 | 3 | 1.5% |
| 3 | 181 | 2 | 1.1% |
| 6 | 27 | 2 | 7.4% |
| 7 | 71 | 0 | 0% |
| 8 | 20 | 0 | 0% |
| 10 | 139 | 7 | 5% |
| All | 637 | 14 | 2.2% |
§ 866.3970 Device to detect and identify microbial pathogen nucleic acids in cerebrospinal fluid.
(a)
Identification. A device to detect and identify microbial pathogen nucleic acids in cerebrospinal fluid is a qualitative in vitro device intended for the detection and identification of microbial-associated nucleic acid sequences from patients suspected of meningitis or encephalitis. A device to detect and identify microbial pathogen nucleic acids in cerebrospinal fluid is intended to aid in the diagnosis of meningitis or encephalitis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including primer/probe sequence, design, and rationale for sequence selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical studies: Analytical sensitivity (limit of detection), inclusivity, reproducibility, interference, cross reactivity, and specimen stability.
(3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted comparator methods.
(4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardware-based devices that incorporate software.
(5) The Intended Use statement in the device labeling must include a statement that the device is intended to be used in conjunction with standard of care culture.
(6) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(7) The device labeling must include a limitation stating that the negative results do not preclude the possibility of central nervous system infection.
(8) The device labeling must include a limitation stating that device results are not intended to be used as the sole basis for diagnosis, treatment, or other patient management decisions.
(9) The device labeling must include a limitation stating that positive results do not mean that the organism detected is infectious or is the causative agent for clinical symptoms.
(10) As part of the risk management activities performed as part of your 21 CFR 820.30 design controls, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.