K Number
K190163

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-07-16

(166 days)

Product Code
Regulation Number
874.1820
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ALARA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.

Device Description

The ALARA Neuro Access Kit with Access Needle, Sheath and Extension Handle is a single use: sterile stimulating electrode is manufactured with the following materials: 316 and LVM Stainless Steel, IXEF 1022/9008 and Lustram ABS. The disposable ALARA Neuro Access Kit is an Access Needle insulated with a sheath (see figure 1). The sheath is full length except for the distal tip (either a diamond or bevel tip) which provides tissue stimulation. The needle utilizes an extension handle as an accessory (see figure 2) to provide the surgeon stability in placement of the access needle and by providing support by allowing for more working space during a surgical procedure. The device is kitted with a set of individually packaged and (single use) disposable Alligator Clips (Rhythmlink International Alligator Clip RLSP487) see figure 3.

AI/ML Overview

The provided document is a 510(k) summary for the ALARA Neuro Access Kit, which is a medical device for pedicle pilot hole preparation, locating, and identifying spinal roots/nerves by providing proximity feedback. The summary details performance testing conducted to demonstrate the device meets acceptance criteria.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
ALARA Cannula Handle Torque TestThe cannula handle meets all acceptance criteria since the torque required to cause the cannula to fail was over twice the required torque of 25 in-lbs.
ALARA Cannula Handle Pull TestThe cannula handle meets all acceptance criteria since the force required to cause the cannula to fail was over the required force of 60 lbf.
ALARA Stylet Handle Pull TestThe stylet handle meets all acceptance criteria since the force required to cause the stylet to fail was over the required force of 60 lbf.
ALARA Bevel Sheath Needle Functionality TestThe bevel sheath needle and extension handle meets all acceptance criteria since the samples were able to functionally perform the test without failure.
ALARA Diamond Sheath Needle Functionality TestThe diamond sheath needle and extension handle meets all acceptance criteria since the samples were able to functionally perform the test without failure.
Electrical Safety IEC 60601-1 Edition 3.1The limited evaluation and testing of the ALARA 11 Gauge Access Needle with Sheath found the sample(s) to be in compliance with the specified standards for: Leakage Current, Dielectric Voltage Withstand, Humidity Preconditioning, Enclosure Mechanical Strength, Drop Test, Ball Pressure, and Patient Leads and Cables.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific number of samples used for each bench test. It uses phrases like "the samples were able to functionally perform the test" or "the sample(s) to be in compliance," which implies that a sample size was used but not quantified.
The testing was retrospective bench testing performed on the device itself, simulating use cases in laboratory conditions. There is no information provided about the data provenance in terms of country of origin, as it is device bench testing rather than clinical data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This section is not applicable as the studies described are bench tests to assess physical and electrical properties of the device, not clinical performance studies requiring expert-established ground truth.

4. Adjudication Method for the Test Set

This section is not applicable as the studies described are bench tests and do not involve human interpretation or adjudication. The "ground truth" for these tests comes from engineering specifications and objective measurements (e.g., torque, force, electrical properties).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not performed. The provided document details bench testing for physical and electrical performance, not a clinical study involving human readers or AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. The ALARA Neuro Access Kit is a physical medical device, not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

7. The Type of Ground Truth Used

The "ground truth" for the bench tests is based on engineering specifications and established safety standards. For example, the torque required to cause failure must exceed 25 in-lbs, and the force required to cause failure must exceed 60 lbf. Electrical safety is verified against IEC 60601-1 Edition 3.1. These are objective, measurable criteria.

8. The Sample Size for the Training Set

Not applicable. This device is not an AI/ML product, so there is no "training set." The testing performed is to validate the physical device's performance against pre-defined specifications.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no training set for an AI/ML product, there is no ground truth to establish for such a set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

July 16, 2019

SurGenTec. LLC % Stephen Inglese Founder/CEO Quality Solutions and Support, LLC PO Box 8271 Holland, Michigan 49422

Re: K190163

Trade/Device Name: ALARA Neuro Access Kit Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: PDQ Dated: June 13, 2019 Received: June 17, 2019

Dear Stephen Inglese:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190163

Device Name ALARA Neuro Access Kit

Indications for Use (Describe)

The ALARA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the ALARA Neuro Access Kit is provided below.

Device Common Name:Nerve Stimulator / Locator
Classification Name:Neurosurgical Nerve Locator
Device Proprietary Name:ALARA Neuro Access Kit
Submitter:SurGenTec, LLC7601 N Federal Hwy, Suite 150ABoca Raton, FL 33487561-990-7882
Contact:Stephen W IngleseCEO, FounderQuality Solutions and Support, LLCPhone: 561-251-0876Email: swi@qss-llc.com
Date Prepared:July 15, 2019
Classification Regulation:21 CFR § 874.1820, Class II
Panel:Neurology
Product Code:PDO
Predicate Device: -
510kProduct NameClearance Date
K171807ES2 Neuromonitoring AccessoryInstrumentsJuly 18,2017

Indication for Use:

The ALARA Neuro Access Kit is indicated for pedicle pilot hole preparation, locating, and identifying spinal roots / nerves by providing proximity feedback.

Device Description:

The ALARA Neuro Access Kit with Access Needle, Sheath and Extension Handle is a single use: sterile stimulating electrode is manufactured with the following materials: 316 and LVM Stainless Steel, IXEF 1022/9008 and Lustram ABS. The disposable ALARA Neuro Access Kit is an Access Needle insulated with a sheath (see figure 1). The sheath is full length except for the distal tip (either a diamond or bevel tip) which provides tissue stimulation. The needle

{4}------------------------------------------------

utilizes an extension handle as an accessory (see figure 2) to provide the surgeon stability in placement of the access needle and by providing support by allowing for more working space during a surgical procedure.

The device is kitted with a set of individually packaged and (single use) disposable Alligator Clips (Rhythmlink International Alligator Clip RLSP487) see figure 3. The device registered with Rhythmlink (FDA registered FEI: 3003831320).

Image /page/4/Picture/2 description: The image shows a black and blue tool on a white surface. The tool has a black handle with a blue top. The tool has a long, thin metal rod that extends from the handle. The tool is likely used for precision work.

Figure 1- ALARA Access Needle with Sheath

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a medical device on a white table. The device has a long, white handle with a silver circular piece in the middle. At the end of the handle, there is a black and blue component with a thin, black probe extending from it, and a black wire is connected to the device.

Figure 2- Extension Handle attached to the Access Needle

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains three separate photos. The first photo shows a medical device with two wires connected to it via alligator clips. The second photo shows a sealed package with the brand name "Rhythmlink" and other information such as the company address, phone number, and a statement that it is for single use only. The third photo shows a similar sealed package, but this time the contents are visible, revealing a black wire with an alligator clip attached to one end.

Figure 3- Access Needle with Extension Handle and Alligator Clip Attached / Packaged

{7}------------------------------------------------

Performance Testing:

Bench testing was performed with the device in simulated use cases to determine that the device met its performance specifications. Where applicable, functionality testing included materials that simulate relevant mechanical properties of bone.

  • ALARA Cannula Handle Torque Test The cannula handle meets all acceptance ● criteria since the torque required to cause the cannula to fail as over twice the required torque of 25 in-lbs.
  • ALARA Cannula Handle Pull Test The cannula handle meets all acceptance ● criteria since the force required to cause the cannula to fail as over the required force of 60 lbf.
  • . ALARA Stylet Handle Pull Test - The stylet handle meets all acceptance criteria since the force required to cause the stylet to fail as over the required force of 60 lbf.
  • ALARA Bevel Sheath Needle Functionality Test The bevel sheath needleand ● extension handle meets all acceptance criteria since the samples were able to functionally perform the test without failure.
  • . ALARA Diamond Sheath Needle Functionality Test - The diamond sheath needle and extension handle meets all acceptance criteria since the samples were able to functionally perform the test without failure.
  • Electrical Safety IEC 60601-1 Edition 3.1. with AAMI/CSA/EN Deviations: ● Leakage Current Test: (Clause 8.7), Dielectric Voltage Withstand: (Clause 8.8.3), Humidity Preconditioning Treatment: (Clause 5.7), Enclosure Mechanical Strength: (Clause 15.3), Drop Test: (Clause 15.3.4), Ball Pressure: (Clause 8.8.4.1), Patient Leads and Cables: (Clause 8.5.2.3). SUMMARY: The limited evaluation and testing of the ALARA 11 Gauge Access Needle with Sheath, found the sample(s) to be in compliance with the above specified standards.

Substantial Equivalence:

Based on similar intended use and technological characteristics, the ALARA Neuro Access Kit is Substantially equivalent to the predict device: K171807 - ES2 Neuromonitoring Accessory Instruments Table 1 demonstrates the technological features; both similar and different.

{8}------------------------------------------------

AdministrativeSubject DevicePredicate Device
CharacteristicSurGenTec, LLC –K190163Stryker -K171807
Device NameALARA Neuro Access KitES2 NeuromonitoringAccessory Instruments – 200,300 and 400 LITE Y-Needles
Regulation21 CFR § 874.1820, Class II21 CFR § 874.1820, Class II
Product CodePDQPDQ
Device Characteristic:
Indications For UseThe ALARA NeuroAccess Kit is indicated forpedicle pilot holepreparation, locating, andidentifyingspinal roots / nerves byproviding proximityfeedback.The ES2 Neuromonitoringinstruments (Awls, Taps,Screwdriver and LITe Y-NEEDLE 200, 300 and 400) canbe used by the surgeon to assistin location of the spinal nervesby providing proximityinformation before, during orafter bone preparation andplacement of bone screws inopen and percutaneousminimally invasive posteriorsurgical approaches of the non-cervical spine.
MaterialSurgical Stainless Steel,lxef® 1022/9008 andAcrylonitrile ButadieneStyrene (ABS)Surgical Stainless Steel andAcrylonitrile Butadiene Styrene(ABS)
IEC 60601 CompliantYesYes
Use of the NeedleNeedle depth stopNeedle depth stop
SterilitySterile (gamma)Sterile and Non SterileNon-sterile devices are providedwith validated steam sterilizationparameters to assure an SAL of10-6
Reusable /Single UseSingle Use200, 300, and 400 LITe Y Needle- Single UseAwls, Taps, and Screwdrivers -Reusable
Connection to aNeuromonitoring UnitWire and ClipWire and Clip
Extension HandleYesNo
Instrument Type(Description)ALARA Needle, HandleAwl, Taps, Screwdrivers,Guidewires (K-wires) and200,300, 400 LITE Y-Needles
Biocompatibility PatientContact DurationLimited patient durationcontact (≤24 hours)Limited patient duration contact(≤24 hours)
BiocompatibilityYesYes
Surgical ApproachOpen or Percutaneous/Minimally InvasiveOpen or Percutaneous/ MinimallyInvasive
Compatible with CommonNeuromonitoring Consoles& SoftwareYesYes
Connection toNeuromonitoring UnitClipClip or Probe (based onNeuromonitoring system)

Table 1: Device Comparison Table for Substantial Equivalence

{9}------------------------------------------------

*Bold text denotes a difference between the subject device and the predicate device.

Summary of Substantial Equivalence:

As described above, the difference in technological characteristics do not raise new types of safety and effectiveness issues. The testing data (biocompatibility, bench testing and electrical safety) provided in this 510(k) show equivalence to the predicate as cleared in K171807 and therefore the ALARA Neuro Access Kit can be found substantially equivalent to the predicate device.

§ 874.1820 Surgical nerve stimulator/locator.

(a)
Identification. A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.(b)
Classification. Class II.