K Number
K190070
Date Cleared
2019-04-11

(86 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sonopet iQ Ultrasonic Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue and hard tissue is desirable, including Neurosurgery, Gastrointestinal and affiliated Organ surgery, Urological surgery, Plastic and Reconstructive surgery, General surgery, Orthopedic surgery, Gynecological surgery, Thoracic surgery, Laparoscopic surgery, and Thoracoscopic surgery.

CONTRAINDICATION: This ultrasonic surgical aspirator device is not indicated for and should not be used for the fragmentation, emulsification, and aspiration of uterine fibroids.

Device Description

The Sonopet iQ Ultrasonic Aspirator System is an ultrasonically vibrating surgical device which, in combination with irrigation and aspiration, fragments, emulsifies and removes unwanted tissue. It allows the selective resection of target tissue while preserving vessels, ducts and other delicate structures. The system consists of a console which provides control and power functions, a surgical handpiece to provide ultrasonic mechanical energy (25kHz), fifteen (15) titanium tips with irrigation sleeves, and an Irrigation Suction Cassette. Users may choose from three different styles of footswitches. An optional hand controller that allows adjustment of the console settings is also available.

Control of the system functional outputs (ultrasonic energy, irrigation, and aspiration) is achieved via the console. The console interfaces with the irrigation cassette to control the levels of aspiration and irrigation at the surgical site. It also controls the energy delivered to the connected handpiece to control the motion of the tip. The settings on the console are adjusted directly on the console or via the optional connected hand controller. Activation of the system is controlled by pressing the treadle of a wired or wireless foot pedal.

This system is for professional use only within a professional healthcare environment.

AI/ML Overview

The provided document K190070 describes the Sonopet iQ Ultrasonic Aspirator System, a surgical ultrasonic aspirator. The submission focuses on demonstrating substantial equivalence to a predicate device, the Sonopet Model UST-2001 Ultrasonic Surgical Aspirator (K010309).

This document does not contain acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of an AI/ML device. The device described is a surgical instrument and not an AI/ML diagnostic or assistive technology.

Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor details about sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone performance for an AI/ML device, as this information is not relevant to the described medical device and is not present in the provided text.

The document discusses nonclinical testing to support the submission, which includes:

  • Reprocessing, sterilization, and package integrity testing
  • Biocompatibility testing
  • Software and wireless technology testing
  • EMC and Electrical Safety testing
  • Bench testing
  • Simulated Use testing
  • Human Factors and Usability testing

It explicitly states: "No clinical testing was required to support this submission." This further confirms that the provided text does not describe the kind of study (e.g., diagnostic accuracy, comparative effectiveness with human readers) that would be relevant to an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase and black. The word is centered and takes up most of the frame. The background is white.

K190070

510(k) Summary

807.92(a)(1) – Submitter Information
510(k) Owner/Submitter:Stryker Instruments4100 E. Milham AvenueKalamazoo, MI 49001(p) 269-323-7700
Contact Person:Julia L. HelgesonPrincipal RA Specialist
Registration Number:1811755
Date Summary Prepared:April 09, 2019
807.92(a)(1) – Submitter Information
510(k) Owner/Submitter:Stryker Instruments4100 E. Milham AvenueKalamazoo, MI 49001(p) 269-323-7700
Contact Person:Julia L. HelgesonPrincipal RA Specialist
Registration Number:1811755
807.92(a)(2) – Name of Device
Trade Name(s):Sonopet iQ Ultrasonic Aspirator System
Common Name:Surgical Ultrasonic Aspirator
Classification Name:Aspirator, Surgical Ultrasonic
Classification Panel:General & Plastic Surgery
Device Class:Unclassified
Product Code(s):LFL

Page 1/4

{1}------------------------------------------------

807.92(a)(3) – Legally marketed device(s) to which equivalence is claimed Sonopet Model UST-2001 Ultrasonic Surgical Aspirator – K010309

807.92(a)(4) – Device Description

The Sonopet iQ Ultrasonic Aspirator System is an ultrasonically vibrating surgical device which, in combination with irrigation and aspiration, fragments, emulsifies and removes unwanted tissue. It allows the selective resection of target tissue while preserving vessels, ducts and other delicate structures. The system consists of a console which provides control and power functions, a surgical handpiece to provide ultrasonic mechanical energy (25kHz), fifteen (15) titanium tips with irrigation sleeves, and an Irrigation Suction Cassette. Users may choose from three different styles of footswitches. An optional hand controller that allows adjustment of the console settings is also available.

Control of the system functional outputs (ultrasonic energy, irrigation, and aspiration) is achieved via the console. The console interfaces with the irrigation cassette to control the levels of aspiration and irrigation at the surgical site. It also controls the energy delivered to the connected handpiece to control the motion of the tip. The settings on the console are adjusted directly on the console or via the optional connected hand controller. Activation of the system is controlled by pressing the treadle of a wired or wireless foot pedal.

This system is for professional use only within a professional healthcare environment.

807.92(a)(5) – Intended Use of the Device
Indications for Use:The Sonopet iQ Ultrasonic Aspirator System is indicated for use insurgical procedures where fragmentation, emulsification andaspiration of soft tissue and hard tissue is desirable, includingNeurosurgery, Gastrointestinal and affiliated Organ surgery,Urological surgery, Plastic and Reconstructive surgery, Generalsurgery, Orthopedic surgery, Gynecological surgery, Thoracicsurgery, Laparoscopic surgery, and Thoracoscopic surgery.
CONTRAINDICATION: This ultrasonic surgical aspirator device is notindicated for and should not be used for the fragmentation,emulsification, and aspiration of uterine fibroids.

807.92(a)(6) Summary of the technological characteristics of the device compared to the predicate The subject and predicate devices are ultrasonic surgical instruments which use electrical energy in the ultrasonic frequency range to drive a tip for the fragmentation and emulsification of soft and hard tissue. The systems use irrigation at the surgical site to emulsify the fragmented tissue. The systems also use aspiration to remove the fragmented and emulsified tissue from the surgical site.

The following primary characteristics are identical for the subject and predicate devices:

{2}------------------------------------------------

CharacteristicSubject DevicePredicate Device
Energy SourceAC Powered from ElectricalMains through detachablecordAC Powered from ElectricalMains through detachablecord
Operating Principle/Functional OutputsMechanical Ultrasonic EnergyIrrigationAspirationMechanical Ultrasonic EnergyIrrigationAspiration
Tissue Fragmentation MethodUltrasonic EnergyUltrasonic Energy
Ultrasonic Energy ControlPiezo-electric, sinusoidal, non-continuousPiezo-electric, sinusoidal, non-continuous
Irrigation ControlForced, via peristaltic pumpForced, via peristaltic pump
Aspiration ControlVacuumVacuum
System Activation ControlMethodFoot PedalFoot Pedal
Terminal Sterilization MethodsEthylene oxideEthylene oxide
Reprocessing Methods (SterileReusable)Manual or automated(mechanical) cleaning withMoist Heat SterilizationManual or automated(mechanical) cleaning withMoist Heat Sterilization
Reprocessing Methods (Non-sterile Reusable)Clean and DisinfectClean and Disinfect
Electrical ClassificationClass IIClass II
MR SafetyMR UnsafeMR Unsafe

The following technological characteristics of the subject and predicate device systems differ. These differences and changes offer new or altered features over those of the predicate device, but do not impact the intended use, indications for use, operating principle, or energy source.

CharacteristicSubject DevicePredicate Device
Power Delivery OptionsConstant, VariableConstant
Aspiration Flow ControlVacuum pump, combinationof solenoid and pinch valvesVacuum pump, combinationof needle and pinch valves
Aspiration Potential (vacuumpressure)75.5kPa maximum85kPa maximum
Frequency25kHz25kHz and 34kHz
Accessory Connections/Device RecognitionKeyed connectors, illuminatedports for status, RFIDrecognition, communicationbus recognitionKeyed connectors, recognitionby discrete component
Handpiece to Tip AssemblyIn sterile field, direct torquewith wrench (no need to restin sterilization tray)In sterile field, torque withwrench, requires sterilizationtray to stabilize handpiece
Primary User InterfaceGraphic User Interface (GUI),illuminated portsPush button, LED indicators
Foot Pedal OptionsWired or wirelessBasic or advanced featuresWiredBasic
Tip Lengths11 – 20cm9.7 – 20cm

{3}------------------------------------------------

Tip Distal GeometriesStraight, Knife, Claw, BarracudaStraight, Knife, Claw, Barracuda, Curette
Operating Conditions10 – 27°C10% - 85% RH70kPa – 106kPa10 – 40°C10% - 85% RH70kPa – 106kPa

807.92(b)(1) - Nonclinical Testing to Support Submission

A suite of performance testing was executed to demonstrate substantial equivalence with the predicate devices. Testing included, but was not limited to:

  • Reprocessing, sterilization, and package integrity testing per FDA Guidance documents and recognized standards
  • Biocompatibility testing per FDA Guidance documents and recognized standards
  • Software and wireless technology testing per FDA Guidance documents and recognized ● standards
  • EMC and Electrical Safety testing per FDA recognized standards
  • Bench testing per various internal protocols ●
  • Simulated Use testing per various internal protocols
  • Human Factors and Usability testing per FDA Guidance documents and recognized standards

807.92(b)(2) - Clinical Testing

No clinical testing was required to support this submission.

807.92(b)(3) - Conclusions Drawn from Testing Performed

Results of performance testing were determined successful for all protocols and demonstrate that the functionality, integrity, and safety and effectiveness of the Sonopet iQ Ultrasonic Aspirator System are sufficient for their intended use. Verification test results for the Sonopet iQ Ultrasonic Aspirator System confirm differences in technology raise no new issues of safety and effectiveness when compared to the predicate devices.

Conclusion/ Substantial Equivalence (SE) Rationale:

The subject device, in comparison with the legally marketed predicate, has the same intended use, indications for use, operating principles, energy source, and functional outputs. Performance testing and risk analysis demonstrate that the device is as safe and effective as the predicate device and does not raise different significant questions of safety and effectiveness.

{4}------------------------------------------------

Please wait...

If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.

You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.

For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.

Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 11, 2019

Stryker Corporation Julia Helgeson Principal RA Specialist 4100 E. Milham Ave. Kalamazoo, Michigan 49001

Re: K190070

Trade/Device Name: Sonopet iQ Ultrasonic Aspirator System Regulatory Class: Unclassified Product Code: LFL Dated: January 11, 2019 Received: January 15, 2019

Dear Julia Helgeson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{6}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

N/A