K Number
K190068

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-06-04

(140 days)

Product Code
Regulation Number
888.3560
Age Range
2 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  1. Painful and disabled knee joint resulting from osteoarthritis, or traumatic arthritis, or traumatic arthritis where one or more compartments are involved.

  2. Correction of varus, valgus, or posttraumatic deformity.

  3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous total joint replacement procedure.

Femoral components and tibial tray components with porous coatings are indicated for cemented biological fixation application. Non-coated (Interlok) femoral components and all polyethylene patellar components are indicated for cemented application only.

Device Description

The Vanguard XP Knee System is a total knee replacement system that consists of a femoral component composed of cobalt-chromium-molybdenum (Co-Cr-Mo), two styles of tibial trays manufactured out of Co-Cr-Mo (with locking bar), and dual bearings machined of Vivacit-E Vitamin-E Highly Crosslinked Polyethylene (VEHXPE). Biomet patellae can be used with the Vanguard XP Knee System. The subject Vanguard XP tibial components are available with a porous plasma spray (PPS®) of titanium alloy powder for uncemented or cemented fixation or with Biomet's Interlok coarse blasted finish for cemented fixation only. The Vanguard XP Knee System offers the flexibility to retain the ACL and PCL ligaments, retain just the PCL ligament and accommodate an intact, partially functioning PCL within a single system.

AI/ML Overview

The provided document, a 510(k) premarket notification for the Vanguard XP Knee System, primarily focuses on demonstrating substantial equivalence to predicate devices for a knee implant, specifically a modification to the tibial bearings. It does not contain information about acceptance criteria or a study that proves the device meets those criteria in the context of an Artificial Intelligence (AI) / Machine Learning (ML) enabled medical device.

The document discusses:

  • Device Identity: Vanguard XP Knee System, a total knee replacement system.
  • Modification: A modification to the tibial bearings within the system.
  • Purpose of Submission: To demonstrate that this modification does not adversely affect safety and efficacy and that the modified device remains substantially equivalent to predicate devices.
  • Testing: Engineering analysis and non-clinical tests (e.g., material characteristics, stress analysis, MRI compatibility, non-pyrogenicity).
  • Clinical Data: Explicitly states, "Clinical data was not deemed necessary for the subject device." This indicates that no human clinical study was conducted for this specific 510(k) submission, and certainly not a study related to AI/ML performance.

Therefore, I cannot fulfill the request to describe the acceptance criteria and study proving an AI/ML device meets them based on the provided text. The document pertains to a traditional orthopedic implant, not an AI/ML medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 4, 2019

Zimmer Inc. Charles Neitzel Senior Regulatory Affairs Specialist P.O. Box 708 Warsaw, Indiana 46581-0708

Re: K190068

Trade/Device Name: Vanguard XP Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, MBH, OIY, MBV Dated: May 7, 2019 Received: May 8, 2019

Dear Charles Neitzel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Raquel Peat, PhD, MPH, USPHS Director OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190068

Device Name Vanguard XP Knee System

Indications for Use (Describe)

  1. Painful and disabled knee joint resulting from osteoarthritis, or traumatic arthritis, or traumatic arthritis where one or more compartments are involved.

  2. Correction of varus, valgus, or posttraumatic deformity.

  3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous total joint replacement procedure.

Femoral components and tibial tray components with porous coatings are indicated for cemented biological fixation application. Non-coated (Interlok) femoral components and all polyethylene patellar components are indicated for cemented application only.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image features the logo for Zimmer Biomet. The logo consists of a blue circle with a stylized "Z" inside it, also in blue. Below the circle is the word "zimmer" in lowercase, using the same blue color as the "Z" and circle. The logo is simple and modern, with a focus on the company's name and a distinctive initial.

K190068 Page 1 of 4

P.O. Box 708 Warsaw, IN 46581-0708 574 267-6131

510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Vanguard XP Knee System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.

Sponsor:Zimmer, Inc.P.O. Box 708Warsaw, IN 46581-0708
Contact Person:Charles NeitzelSenior Specialist, Regulatory AffairsTelephone: (262) 716-3243Fax: (574) 372-4605
Date:January 11, 2019
Trade Name:Vanguard XP Knee System
Common Name:Knee Bearing
Classification Namesand References:JWH – Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer (21CFR 888.3560)MBH - Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Uncemented, Porous Coated,Polymer/Metal/Polymer (21 CFR 888.3565)OIY - Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer +Additive/Metal/Polymer + Additive (21 CFR888.3560)MBV - Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uhmwpe, Pegged, Cemented,Polymer/Metal/Polymer (21 CFR 888.3560)
Classification Panel:Orthopedics/87
Predicate Device(s):Design Predicates:Vanguard XP Knee System, manufactured byBiomet, K122160, cleared on 03/20/2013Vanguard XP Knee System, manufactured byBiomet, K132873, cleared on 12/11/2013Vanguard XP Femoral Component (GenII)/Vanguard XP Femoral Trial, manufactured byBiomet, K141407, cleared on 10/27/2014

{4}------------------------------------------------

  • Vanguard XP Knee System, manufactured by Biomet, K153657, cleared 05/10/2016
    Material Predicate:

  • Persona® Personalized Knee System, . manufactured by Zimmer, K121771, cleared 11/07/2012

  • . Persona® Personalized Knee System, manufactured by Zimmer, K150090, cleared 05/20/2015

The purpose of this subject 510(k) is to introduce a modification to the tibial bearings within the Vanguard XP Knee System. There are no changes to the other components within the system.

The Vanguard XP Knee System is a total knee replacement system that consists of a femoral component composed of cobalt-chromium-molybdenum (Co-Cr-Mo), two styles of tibial trays manufactured out of Co-Cr-Mo (with locking bar), and dual bearings machined of Vivacit-E Vitamin-E Highly Crosslinked Polyethylene (VEHXPE). Biomet patellae can be used with the Vanguard XP Knee System. The subject Vanguard XP tibial components are available with a porous plasma spray (PPS®) of titanium alloy powder for uncemented or cemented fixation or with Biomet's Interlok coarse blasted finish for cemented fixation only. The Vanguard XP Knee System offers the flexibility to retain the ACL and PCL ligaments, retain just the PCL ligament and accommodate an intact, partially functioning PCL within a single system.

  1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, or traumatic arthritis where one or more compartments are involved.

  2. Correction of varus, valgus, or posttraumatic deformity.

  3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous total joint replacement procedure.

Femoral components and tibial tray components with porous coatings are indicated for cemented and uncemented biological fixation application. Non-coated (Interlok) femoral components, tibial tray components

Purpose and Device Description:

Intended Use / Indications for Use:

{5}------------------------------------------------

and all polyethylene patellar components are indicated for cemented application only.

Summary of Technological Characteristics:

The rationale for substantial Identical equivalence is based on consideration of the following characteristics:

  • Intended Use: Identical to the predicates
  • · Indications for Use: Similar to the Vanguard XP Knee System predicates
  • · Materials: Identical to the Persona Personalized Knee System predicates
  • · Design Features: Similar to the Vanguard XP Knee System predicates
  • · Sterilization: Identical to the Persona Personalized Knee System predicates

Summary of Performance Data (Nonclinical and/or Clinical)

● Non-Clinical Tests:

Engineering analysis was conducted to demonstrate that the modifications did not adversely affect safety and efficacy, and to demonstrate substantial equivalence to the predicate components. The test reports are listed below:

  • o Vivacit-E material characteristics presented in MAF-1868
  • o Material and dimensional comparison of Vivacit-E to E1 tibial bearings
  • o Evaluation of stress in manufacturing slots of Vivacit-E tibial bearings using finite element analysis
  • o Laser etching of Vanguard XP Vivacit-E material
  • o Magnetic resonance imaging (MRI) compatibility
  • o Testing to establish product non-pyrogenicity

Clinical Tests: ●

  • Clinical data was not deemed necessary for the o subject device.

{6}------------------------------------------------

Substantial Equivalence Conclusion

The subject device has the same intended use and similar indications for use as the Vanguard XP Knee System predicate devices. The subject device has similar technological characteristics to the predicates, and the performance data and analyses demonstrate that:

  • any differences do not raise new questions of safety . and effectiveness; and
  • the proposed device is at least as safe and effective as the legally marketed predicate devices.

§ 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.