K Number
K183640
Date Cleared
2019-02-22

(58 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Cellvizio® 100 Series Systems with Confocal Miniprobes™ are confocal laser systems with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues, including, but not limited to, the identification of cells and their organization or architecture.

The AQ-Flex™ 19 Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e., gastrointestinal and respiratory tracts, accessed by an endoscope, or endoscopic accessories (e.g. aspiration needles used during procedures including EUS-FNA, EBUS-TBNA and TBNA needles).

Device Description

AQ-Flex™ 19 Confocal Miniprobes™ is used with Cellvizio® 100 Series system to provide imaging of anatomical tracts, i.e., gastrointestinal and respiratory tracts, accessed by an endoscope, or endoscopic accessories. It is designed to be used during transbronchial needle aspiration (TBNA), endoscopic ultrasound transbronchial needle aspiration (EBUS-TBNA) and endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) procedures.

There are no changes in design, materials, or function between the subject AQ-Flex™ 19 device and the cleared AQ-Flex™ 19 Confocal Miniprobe™ (K123676, K150831, and K172844), except for the length of the AQ-Flex 19 which has been reduced from 4 to 3 meters (AQ-Flex™ 19 Confocal Miniprobe™ has previously been cleared with a length of 4 meters).

A locking accessory already available in the package of the predicate device is used as an aid to secure placement inside the needle used during EUS-FNA procedures and can also be used during TBNA, and EBUS-TBNA procedures.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Cellvizio® 100 Series Confocal laser imaging systems and their Confocal Miniprobe™ AQ-Flex™ 19:

1. Table of Acceptance Criteria and Reported Device Performance

Test PerformedAcceptance CriteriaReported Device Performance
Biocompatibility (Cytotoxicity, Sensitization, Irritation or Intracutaneous reactivity & Systemic toxicity)All requirements metPass
Resistance to reprocessing methods: mechanical resistance, tensile strength assessment, functional testing, image quality etc.All requirements metPass
Efficacy of reprocessing methodsAll requirements metPass
Laser safetyAll requirements metPass
Imaging qualityAll requirements metPass
Compatibility of the AQ-Flex™ 19 Confocal Miniprobe™ with endoscopic accessories used during EUS-FNA procedures.All requirements metPass
Compatibility of the AQ-Flex™ 19 Confocal Miniprobe™ with endoscopic accessories used during (EBUS)-TBNA procedures: Mechanical tests to verify no creation of sharp edges on distal head.Verification of smooth distal surface and absence of spikes and edges on distal tip according to ISO 8600-1 standardPass (25/25)
Compatibility of the AQ-Flex™ 19 Confocal Miniprobe™ with endoscopic accessories used during (EBUS)-TBNA procedures: Confocal Miniprobe™ strength test by insertion/extraction and bending in (EBUS)-TBNA needles.Verification of mechanical resistance of the probe after extraction of AQ-Flex™ 19 from needle in EUS-TBNA and TBNA retroflex positionPass (25/25)
Protrusion test of the of the AQ-Flex™ 19 Confocal Miniprobe™ length when used with (This row is incomplete in the provided text, but the intention is clear from the acceptance criteria and result for securing position.)Securing position of AQ-Flex™ 19 distal tip with (The specific conditions for securing are not fully detailed in the provided text, but the general concept is present.)Pass (5/5)

2. Sample Sizes Used for the Test Set and Data Provenance

The document primarily focuses on bench testing and compatibility assessments, not clinical performance studies with human data in the test set.

  • Mechanical tests for sharp edges: 25 samples ("Pass (25/25)")
  • Mechanical strength test (insertion/extraction and bending): 25 samples ("Pass (25/25)")
  • Protrusion test: 5 samples ("Pass (5/5)")

The provenance of this data is not explicitly stated as country of origin, retrospective, or prospective. Given that these are primarily compatibility and mechanical tests, they are likely prospective bench tests conducted by the manufacturer.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not mention human experts being used to establish ground truth for the "test set" in the context of clinical performance. The "ground truth" for the mechanical and compatibility tests would primarily be adherence to engineering specifications and international standards (like ISO 8600-1), which are verified by testers, not expert clinicians establishing a diagnostic truth.

4. Adjudication Method for the Test Set

Not applicable as the reported tests are primarily objective mechanical and compatibility tests against predefined criteria and standards, not subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of an MRMC comparative effectiveness study in this document. The device is a confocal laser imaging system for visualizing tissue microstructure, and the submission is for an expanded indication of use for an existing component (AQ-Flex™ 19 Confocal Miniprobe™) in specific endoscopic procedures. There is no AI component or human-in-the-loop performance evaluation described.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is an imaging tool that provides visual information to a human operator, not an AI algorithm performing a diagnostic task independently.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the tests reported is primarily:

  • Adherence to engineering specifications and standards (e.g., ISO 8600-1 for smooth distal surface, absence of spikes/edges).
  • Verification of mechanical integrity and functionality after stress tests.
  • Compliance with safety standards (e.g., laser safety).
  • Internal validation of performance characteristics (e.g., image quality, resistance to reprocessing).

There is no mention of clinical ground truth such as pathology or outcomes data in this submission for the expanded indication.

8. The Sample Size for the Training Set

Not applicable. This document describes the testing and validation for a hardware device and its accessories, not the development or training of an algorithm or AI system.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for an algorithm described in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Mauna Kea Technologies % Michael Daniel President Daniel & Daniel Consulting 340 Jones Lane Gardnerville, Nevada 89460

February 22, 2019

Re: K183640

Trade/Device Name: Cellvizio 100 Series Confocal Laser Imaging systems and their Confocal Miniprobes Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OWN, GCJ Dated: December 21, 2018 Received: December 26, 2018

Dear Michael Daniel:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

ForNeil ROgden -SDigitally signed byNeil R Ogden -SDate: 2019.02.2209:27:22 -05'00'
Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director
Division of Surgical Devices
Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183640

Device Name

Cellvizio® 100 Series with Confocal Miniprobe™ AQ-Flex™ 19

Indications for Use (Describe)

Cellvizio® 100 Series System with Confocal Miniprobes™ are confocal laser systems with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues including, but not limited to, the identification of cells and vessels and their organization or architecture.

The AQ-Flex™ 19 Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e., gastrointestinal and respiratory tracts, accessed by an endoscopic accessories (e.g aspiration needles used during procedures including EUS-FNA, EBUS-TBNA and TBNA needles).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. Premarket Notification 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number: K183640

Applicant Information:

Date Prepared: December 21, 2018

Manufacturer Contact:

Aline Criton
Name:Mauna Kea Technologies
Address:9 rue d'Enghien
F-75010 Paris, France
Phone:+33 1 48 24 03 45
Fax:+33 1 48 24 12 18

Submission Contact:

Michael A Daniel, Consultant

madaniel@clinregconsult.com Phone Number: (415) 407-0223 (775) 392-2970 Office: Facsimile Number: (610) 545-0799

Device Information:

Device Trade Name:Cellvizio® 100 Series Confocal laser imaging systems and theConfocal Miniprobe™ AQ-Flex™ 19
Common Name:Endoscope and Accessories
Classification Name(s):Confocal Optical Imaging
Product Code/ Reg.:OWN / GCJ 21 CFR 876.1500
Classification:Class II

Predicate Devices:

  • Previously cleared versions of Cellvizio® 100 Series Confocal laser imaging systems . and the AQ-Flex™ 19 Confocal Miniprobe™ (K123676, K150831, and K172844).

Reference Devices:

  • Previously cleared versions of Cellvizio® 100 Series Confocal laser imaging systems . and the AlveoFlexTM Confocal MiniprobeTM (K111047, K150831, and K172844).

{4}------------------------------------------------

Device Description:

AQ-Flex™ 19 Confocal Miniprobes™ is used with Cellvizio® 100 Series system to provide imaging of anatomical tracts, i.e., gastrointestinal and respiratory tracts, accessed by an endoscope, or endoscopic accessories. It is designed to be used during transbronchial needle aspiration (TBNA), endoscopic ultrasound transbronchial needle aspiration (EBUS-TBNA) and endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) procedures.

There are no changes in design, materials, or function between the subject AQ-Flex™ 19 device and the cleared AQ-Flex™ 19 Confocal Miniprobe™ (K123676, K150831, and K172844), except for the length of the AQ-Flex 19 which has been reduced from 4 to 3 meters (AQ-Flex™ 19 Confocal Miniprobe™ has previously been cleared with a length of 4 meters).

A locking accessory already available in the package of the predicate device is used as an aid to secure placement inside the needle used during EUS-FNA procedures and can also be used during TBNA, and EBUS-TBNA procedures.

Intended Use:

Cellvizio® 100 Series Systems with Confocal Miniprobes™ are confocal laser systems with fiber optic probes that are intended to allow imaging of the internal microstructure of tissues, including, but not limited to, the identification of cells and their organization or architecture.

Indications for Use:

The AQ-Flex™ 19 Confocal Miniprobe™ is intended to allow imaging of anatomical tracts, i.e., gastrointestinal and respiratory tracts, accessed by an endoscope, or endoscopic accessories (e.g. aspiration needles used during procedures including EUS-FNA, EBUS-TBNA and TBNA needles).

Comparison to Predicate Device:

The Cellvizio® 100 Series systems and the AQ-Flex™ 19 Confocal Miniprobe™ remain exactly the same devices in terms of design (except for the AQ-Flex™ 19 Confocal Miniprobe™ reduced from 4 to 3 meters), performance and general intended use (allow imaging of the internal microstructure) as the previously cleared devices (K123676, K150831, and K172844). The indication for use is being expanded to include selected needle-based endoscopic procedures for TBNA and EBUS-TBNA procedures.

Verification and validation testing has shown that the AQ-Flex™ 19 Confocal Miniprobes™ are compatible with endoscopes or endoscopic accessories designed and commonly used in these procedures (such as endoscopic needles).

Testing Completed:

The table below summarizes the verification and validation testing performed on predicate device (AQ-Flex™ 19 Confocal Miniprobe™) and the reference device (AlveoFlex™ Confocal Miniprobe ™) that have been provided in previous submissions K111047, K123676, K150831, and K172844 respectively.

{5}------------------------------------------------

Test performedAcceptance CriteriaTest Results
Biocompatibility (Cytotoxicity, Sensitization, Irritationor Intracutaneous reactivity & Systemic toxicity)All requirements metPass
Resistance to reprocessing methods: mechanicalresistance, tensile strength assessment, functionaltesting, image quality etc.All requirements metPass
Efficacy of reprocessing methodsAll requirements metPass
Laser safetyAll requirements metPass
Imaging qualityAll requirements metPass
Compatibility of the AQ-Flex™ 19 ConfocalMiniprobe™ with endoscopic accessories used duringEUS-FNA procedures.All requirements metPass

The potential risks associated with the use of AQ-Flex™ 19 Confocal Miniprobes™ during EBUS-TBNA and TBNA procedures have been assessed. Additional testing with endoscopic accessories used during EBUS-TBNA and TBNA was performed to validate that the AQ-Flex 19 ™ Confocal Miniprobes™ can be used during EBUS-TBNA and TBNA procedures without change of safety, performance and increase of residual risks. The same test protocols as the ones used when previously testing the AQ-Flex™ 19 Confocal Miniprobes™ for EUS-FNA in the gastro-intestinal tract have been used to assess the risks with EBUS-TBNA and TBNA accessories.

The results from these tests, summarized in the table below, demonstrate that the use of AQ-Flex ™ 19 Confocal Miniprobes™ during EBUS-TBNA and TBNA procedures does not change performance or safety and effectiveness, and do not change the global residual risk.

Test performedAcceptance CriteriaResult
Compatibility of the AQ-FlexTM 19 ConfocalMiniprobeTM with endoscopic accessories usedduring (EBUS)-TBNA procedures:Mechanical tests to verify no creation of sharpedges on distal head.Verification of smoothdistal surface and absenceof spikes and edges ondistal tip according to ISO8600-1 standardPass(25/25)
Compatibility of the AQ-FlexTM 19 ConfocalMiniprobeTM with endoscopic accessories usedduring (EBUS)-TBNA procedures: ConfocalMiniprobe TM strength test byinsertion/extraction and bending in (EBUS)-TBNA needles.Verification of mechanicalresistance of the probe afterextraction of AQ-FlexTM 19from needle in EUS-TBNAand TBNA retroflexpositionPass(25/25)
Protrusion test of the of the AQ-FlexTM 19Confocal MiniprobeTM length when used withSecuring position of AQ-FlexTM 19 distal tip withPass (5/5)

{6}------------------------------------------------

Summary:

The AQ-Flex™ 19 Confocal Miniprobe™, when used as part of the Cellvizio® 100 Series, has been shown to be substantially equivalent to the cleared predicate device and can be used as intended to image the internal microstructure of tissues in a variety of anatomical locations. The previously cleared Confocal Miniprobe™ has been verified to be compatible with endoscopes and endoscopic accessories designed to be used in these applications.

Based upon the Intended Use, Indications for Use, product technical information, performances, safety and effectiveness, and biocompatibility provided in this submission and in the submissions of the previously cleared AQ-Flex™ 19 Confocal Miniprobe™ (K123676, K150831) used during EBUS-TBNA and TBNA procedures, the subject device is substantially equivalent to the predicate device. The subject device can safely and effectively be used to visualize and image the internal microstructure of tissues during endoscopic and during EBUS-TBNA and TBNA procedures.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.