K Number
K183611

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-05-22

(147 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Soniquence RF Generator is intended to provide high frequency energy for resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels during surgical procedures and is used to deliver RF energy via an assortment of surgical devices to cut and coagulate different kinds of tissue.

The Soniquence Electrodes are intended for use by a physician familiar with resection, dissection, incision, and hemostasis in soft tissue surgical procedures.

Device Description

The Soniquence RF Generators® are radiofrequency (RF) electrosurgical generators. It is a compact source of high-frequency RF energy employed for cutting and coagulation of soft tissue in a variety of surgical procedures. This action is achieved by front panel selection of waveforms and power levels. The subject device is designed to enable the performance of monopolar and bipolar functions for cutting, coagulation, and hemostasis in conjunction with separately-marketed electrosurgical accessories and ancillary equipment which have been developed for mutuallyexclusive use with the Soniquence® line of generators.

All selections are effected through push buttons and lamps, which give the operator feedback of status. The power level for each mode is indicated by front panel digital displays which also show the status of self-test and monitoring. The display is interlocked with controls to prevent operation when FAIL is displayed. The final output power control is made through foot and /or hand switches.

Soniquence Electrodes are used with the Soniquence RF Generator. The electrical power operating at radio frequency (RF) is transferred to tissue at the surgical site. The time-varying voltage produced by RF electrical power source yields a predetermined electrosurgical effect, such as tissue cutting or coagulation. The Soniquence family of electrodes is available in various shapes and sizes depending on the need of the surgeon. The devices are available in classic alloys (Tungsten, brass, stainless steel).

These devices are designed to comply with international safety standards including applicable IEC series electrical safety standards.

AI/ML Overview

This document, an FDA 510(k) submission summary for the Soniquence RF Generator and Soniquence Electrodes, does not contain information about a study proving the device meets acceptance criteria related to AI/algorithm performance on a test set, expert ground truth establishment, or clinical effectiveness studies with human readers.

The document states:

  • "This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence." (Page 11)
  • The non-clinical testing listed primarily focuses on electrical safety, EMC, Usability Engineering, sterilization, biocompatibility, and software verification and validation. It also includes "Energy output verification of each of the five modes and comparison to predicate" and "Thermal effect comparison for representative modes of the subject device and predicate on ex-vivo bovine tissue." (Page 11)

Therefore, I cannot provide the requested information regarding acceptance criteria and study details for AI/algorithm performance characteristics. The device appears to be a traditional electrosurgical unit, not an AI/algorithm-driven diagnostic or therapeutic device.

If you are looking for information about the performance of the Soniquence RF Generator and Soniquence Electrodes, the document primarily relies on substantial equivalence to predicate devices (Surgi-Max® Ultra (K170107), Bovie® IDS-310 Electrosurgical Generator (K134054), Elliquence Electrodes (K142410), and Cobbra RF™ Tissue Dissector (K160041)) based on technological characteristics and intended use.

Here's a summary of the performance testing mentioned, which is non-clinical and geared towards safety and functional equivalence, not AI performance characteristics.

Table of Acceptance Criteria and Reported Device Performance (Non-AI/Algorithm Related):

Acceptance Criteria CategorySpecific Tests/Performance Metrics Mentioned (Non-AI)Reported Device Performance
Electrical SafetyAAMI ANSI ES60601-1Successfully performed
IEC 60601-1-2 (EMC)Successfully performed
IEC 60601-2-2Successfully performed
UsabilityIEC 62366 (Usability Engineering)Successfully performed
SterilizationISO 11135 (EO sterilization)Successfully performed
ISO 11737-1 (Microbiological methods: population)Successfully performed
ISO 11737-2 (Microbiological methods: sterility tests)Successfully performed
ISO 11138-1 (Biological indicators)Successfully performed
BiocompatibilityISO 10993-1 (Risk management)Successfully performed
ISO 10993-5 (Cytotoxicity)Successfully performed
ISO 10993-7 (EO residuals)Successfully performed
ISO 10993-10 (Irritation/sensitization)Successfully performed
ISO 10993-11 (Systemic toxicity)Successfully performed
Shelf LifeASTM F1980-07 (Accelerated Aging of Barrier Systems)Successfully performed
SoftwareIEC 62304 / FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"Software verification and validation successfully performed
Functional PerformanceEnergy output verification of each of the five modesComparison to Predicate
Thermal effect comparison (ex-vivo bovine tissue)Comparison to Predicate

Since the requested information pertains to an AI/algorithm-driven device, and this document describes a traditional electrosurgical unit without AI components, the following points cannot be answered from the provided text:

  • Sample size used for the test set and data provenance: No test set (in the AI sense) is mentioned.
  • Number of experts used to establish ground truth & qualifications: Not applicable.
  • Adjudication method: Not applicable.
  • Multi-Reader Multi-Case (MRMC) comparative effectiveness study: Not applicable.
  • Standalone (algorithm only) performance: Not applicable.
  • Type of ground truth used: Not applicable.
  • Sample size for the training set: Not applicable.
  • How ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal for the Department of Health & Human Services - USA. To the right of the seal, there is a blue square with the letters "FDA" in white. To the right of the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

May 22, 2019

Soniquence, LLC Suzanne Lucas Regulatory Affairs 2473 Grand Avenue Baldwin, New York 11510

Re: K183611

Trade/Device Name: Soniquence RF Generator and Soniquence Electrodes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: March 22, 2019 Received: March 25, 2019

Dear Suzanne Lucas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jennifer R. Stevenson Acting Division Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K183611

Device Name

Soniquence RF Generator and Soniquence Electrodes

Indications for Use (Describe)

The Soniquence RF Generator is intended to provide high frequency energy for resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels during surgical procedures and is used to deliver RF energy via an assortment of surgical devices to cut and coagulate different kinds of tissue.

The Soniquence Electrodes are intended for use by a physician familiar with resection, incision, and hemostasis in soft tissue surgical procedures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the first page of a document, as indicated by the text "Page 1 of 9" in the upper right corner. The document number, "K183611", is printed at the top center of the page. The page features a blue graphic on the left and the word "Soniquence" in blue on the right.

Exhibit 4. 510(k) SUMMARY

(As required by 21 CFR 807.92(a))

Date Prepared

May 22, 2019

Submitter's Information (807.92(a)(1))

Company Name and Address:

Soniquence, LLC 2473 Grand Avenue Baldwin, NY 11510 Phone: (516) 654-4000 www.soniquence.com

Establishment Registration for Soniquence, LLC will be completed once this 510(k) has been FDA cleared.

Contact Information:

Suzanne Lucas - Regulatory Affairs Phone: (516) 654-8017 Fax: (516) 654-8000 Email: slucas@soniquence.com

Paul D. Buhrke IV- QA/RA Manager Phone: (516) 277-9010 Fax: (516) 277-9011 Email: pbuhrke@soniquence.com

{4}------------------------------------------------

Device Information (807.92(a)(2))

Trade Name: Soniquence RF Generator and Soniquence Electrodes

Common/Usual Name

Electrosurgical unit and accessories, electrosurgical cutting and coagulation device and accessories

Classification Name and Requlation

Electrosurgical Cutting and Coagulation Device and Accessories; 21 CFR 878.4400

Classification Panel

General and Plastic Surgery Device

Device Class/Product Code

FDA Classification: Class II FDA Product Code: GEI

Predicate Devices (807.92(a)(3))

  • Surgi-Max® Ultra (K170107) o
  • Bovie® IDS-310 Electrosurgical Generator (K134054) ●
  • Elliquence Electrodes (K142410) o
  • Cobbra RF™ Tissue Dissector (K160041) o

Device Description (807.92(a)(4))

The Soniquence RF Generators® are radiofrequency (RF) electrosurgical generators. It is a compact source of high-frequency RF energy employed for cutting and coagulation of soft tissue in a variety of surgical procedures. This action is achieved by front panel selection of waveforms and power levels. The subject device is designed to enable the performance of monopolar and bipolar functions for cutting, coagulation, and hemostasis in conjunction with separately-marketed electrosurgical accessories and ancillary equipment which have been developed for mutuallyexclusive use with the Soniquence® line of generators.

All selections are effected through push buttons and lamps, which give the operator feedback of status. The power level for each mode is indicated by front panel digital displays which also show the status of self-test and monitoring. The display is interlocked with controls to prevent operation when FAIL is displayed. The final output power control is made through foot and /or hand switches.

Soniquence Electrodes are used with the Soniquence RF Generator. The electrical power operating at radio frequency (RF) is transferred to tissue at the surgical site. The time-varying voltage produced by RF electrical power source yields a predetermined electrosurgical effect, such as tissue cutting or coagulation. The

{5}------------------------------------------------

Soniquence family of electrodes is available in various shapes and sizes depending on the need of the surgeon. The devices are available in classic alloys (Tungsten, brass, stainless steel).

These devices are designed to comply with international safety standards including applicable IEC series electrical safety standards.

Intended Use:

The Soniquence RF Generator is intended to provide high frequency energy for resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels during surgical procedures and is used to deliver RF energy via an assortment of surgical devices to cut and coagulate different kinds of tissue.

The Soniquence Electrodes are intended for use by a physician familiar with resection, dissection, incision, and hemostasis in soft tissue surgical procedures

Substantial Equivalence Comparison (807.92(a)(6))

The Soniquence RF Generator is substantially equivalent in both intended use and technological characteristics to the Surgi-Max® Ultra (K170107), as well as the Bovie® IDS-310 Electrosurgical Generator (K134054). An extensive comparison chart concerning key characteristics is provided on the following pages (Table 1.1).

Table 1.1 Substantial Equivalence Comparison Chart (RF Generator)

PredicateinformationSoniquence RF GeneratorPROPOSED DEVICESurgi-Max® Ultra(K170107)PREDICATEBovie® IDS-310ElectrosurgicalGenerator (K134054)PREDICATEPrinciple ofOperationIdentical to predicatesConversion of electricalenergy to high-frequencyRF energy(monopolar and bipolar)Conversion of electricalenergy to high-frequency RF energy(monopolar and bipolar)
ManufacturerSoniquence, LLCelliquence, LLCBovie Medical Corp.Modes ofoperationIdentical to elliquenceK170107CutBlendHemoBipolarBipolar Turbo9 Monopolar and 4bipolar modes
ProductcodeGEIGEIGEIOutputfrequencyIdentical to elliquenceK1701074 MHz (monopolar)1.7 MHz (bipolar)350 - 490 kHz(monopolar)490 kHz (bipolar)
RegulationnumberIdentical to predicates21 CFR 878.440021 CFR 878.4400Maximumpower outputIdentical to elliquenceK170107170W300W
RegulationnameIdentical to predicatesElectrosurgical Cutting andCoagulation Device andaccessoriesElectrosurgical Cuttingand Coagulation Deviceand accessoriesVoltage(peak-to-peak)Identical to elliquenceK1701071,200V8,000V
Indicationsfor useThe Soniquence RFGenerator is intended toprovide high frequency energyfor resection, ablation, andcoagulation of soft tissues andhemostasis of blood vesselsduring surgical proceduresand is used to deliver RFenergy via an assortment ofsurgical devices to cut andOrthopedic, arthroscopic,spinal, and neurologicalFor resection, ablation, andcoagulation of soft tissuesand hemostasis of bloodvessels in orthopedic,arthroscopic, spinal andneurological procedures. Forsoft tissue resection andablation during arthroscopicsurgical procedures of knee,The IDS-310 HighFrequency ElectrosurgicalGenerator is used todeliver RF energy via anassortment of surgicaldevices to cut andcoagulate different kindsof tissue.Duty CycleIdentical to elliquenceK17010710 sec On / 30 sec OFF10 sec On / 30 sec OFF
coagulate different kinds oftissue.shoulder, ankle, elbow, hipand wrist.CuttingSnoring, Submucosal palatalshrinkage, traditionaluvulopalatoplasty (RAUP),myringotomy with effectivehemorrhage control, epistaxistreatment, and turbinateshrinkage, skin incisions,biopsy, cysts, abscesses,tumors, cosmetic repairs,development of skin flaps,skin tags, Blepharoplasty.Blended Cutting andCoagulationSnoring, Submucosal palatalshrinkage, traditionaluvulopalatoplasty (RAUP),myringotomy with effectivehemorrhage control, epistaxistreatment, and turbinateshrinkage, skin tags,papilloma Keloids, Keratosis,Verrucae, Basal CellCarcinoma, Nevi, Fistulas,Epithelioma, CosmeticRepairs, Cysts, Abscesses,Development of skin flaps.Hemostasis andNonablative CoagulationControl of bleeding, Epilation,TelangiectasiaBipolarPinpoint, PreciseCoagulation, PinpointHemostasis, in any field (wetor dry), snoring, submucosalpalatal shrinkage, traditionaluvulopalatoplasty (RAUP),myringotomy with effectivehemorrhage control, epistaxistreatment, and turbinateshrinkage.Supply inputpowerIdentical to elliquenceK170107100 VAC - 240 VAC,50/60 Hz100 VAC - 240 VAC,50/60 Hz
Prescriptiondevice(21 CFR801.109)Identical to predicatesYesYesPoweractivationcontrolIdentical to predicatesFootswitch and/orfingerswitchFootswitch and/orfingerswitch
Electricalsafety andEMC testingstandardsIdentical to elliquenceK170107AAMI ANSI ES60601-1IEC 60601-1-2IEC 60601-2-2IEC 60601-1IEC 60601-1-2IEC 60601-2-2
Applied parttypeIdentical to elliquenceK170107CFCF

{6}------------------------------------------------

{7}------------------------------------------------

Table 1.2 – Substantial equivalence comparison table (Electrodes)

PredicateInformationSoniquence ElectrodesPROPOSED DEVICEElliquence Electrodes(K142410)PREDICATECobbra™ RF TissueDissector (K160041)PREDICATE
Intended UseThe SoniquenceElectrodes are intended foruse by a physician familiarwith resection, dissection,incision, and hemostasis insoft tissue surgicalprocedures.The ElliquenceElectrodes are intendedfor use by a physicianfamiliar with resection,dissection, incision, andhemostasis in softtissue surgicalprocedures. The typesIntended for use by aphysician familiar withresection, dissection,incision, and hemostasis insoft tissue surgicalprocedures. Generalsurgery, laparoscopic,procedures, endoscopicprocedures, open
of surgery intended are:General surgery,Laparaoscopicprocedures, Endoscopicprocedures, Openabdominal, Orthopediccoagulation,Thorascopiccoagulation,Neurosurgicalcoagulation,Gynecologicalcoagulation, (except foruse in femalesterilization), Ear, Nose,Throat coagulation.abdominal, orthopediccoagulation, thorascopiccoagulation, neurosurgicalcoagulation, gynecologicalcoagulation, and ENTcoagulation are the typesof intended surgeries.
Product CodeIdentical to predicatesGEIGEI
Device ClassificationIdentical to predicatesClass IIClass II
FunctionsIdentical to predicatesCutting/coagulationCutting/coagulation
Energy type/operating principleIdentical to predicatesMonopolarMonopolar
Design SpecificationsIdentical to predicatesIntegrated sterilecomponents; handle,cable, connector, shaft,monopolar electrodeIntegrated sterilecomponents; handle,cable, connector, shaft,monopolar electrode
Shaft formsCobbra models identical toCobbra predicateAll others are identical toElliquence ElectrodespredicateStraight, Bayonet,angled, and curvedStraight
Shaft malleabilityCobbra models identical toCobbra predicateAll others are identical toElliquence ElectrodespredicateFixed and malleablemodelsFixed models only
Shaft lengthsIdentical to predicate1.9 - 60 cmelectrode shaft: 12.7 cmhandpiece shaft: 18.55 cm
Shaft diameterCobbra models identical toCobbra predicateAll others are identical toElliquence Electrodespredicate1/16" and 3/32"electrode shaft: 6.0 mmhandpiece shaft: 21.7 mm
Tip ConfigurationsCobbra models identical toCobbra predicateAll others are identical toElliquence ElectrodespredicateLoop (round, diamond),ball, and blade (needle,fine wired, spatula)Spatula
Material Composition
HandleCobbra models identical toCobbra predicateAll others are identical toElliquence ElectrodespredicateMolded PlasticMolded plastic
ShaftSurgical StainlessSteel/BrassSurgical stainless steel
ElectrodeSurgical Stainless Steel/Tungsten/BrassSurgical stainless steel /PFA insulation
Patient contactingmaterialsCobbra models identical toCobbra predicateAll others are identical toElliquence ElectrodespredicateMedical grade stainlesssteel, brass, tungsten,Loctite, PFA, and PTFEMedical grade stainlesssteel and PFA
CoatingIdentical to predicatesNoneNone
BiocompatibilityIdentical to predicatesISO 10993ISO 10993
Sterilization MethodsIdentical to predicatesGamma or EthyleneOxideGamma or Ethylene Oxide
Expiration DatingIdentical to predicatesYesYes
PackagingCobbra models identical toCobbra predicateAll others are identical toElliquence ElectrodespredicateSterile, Blister TraySterile, Peel PouchSterile, Blister Tray
Activation MethodIdentical to predicatesFingerswitch/FootswitchFingerswitch/Footswitch
Manual ControlsIdentical to predicatesNone, all power settingsestablished ongeneratorNone, all power settingsestablished on generator
Internal Memory/circuitry for generatorrecognitionIdentical to predicatesNoneNone
Min., Max., Defaultoutput Values(Wattage)Identical to predicates0-1700-170
Patient LeakagecurrentIdentical to predicatesType CFType CF
Cable Length (m)Cobbra models identical toCobbra predicateAll others are identical toElliquence Electrodespredicate33
Temperature ProbeIdentical to predicatesNoNo
Cooling FunctionIdentical to predicatesNoneNone

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

Non-Clinical Testing (807.92(b)(1))

The Soniquence RF Generator and Soniquence Electrodes will be manufactured in accordance with the design control requirements of 21 CFR 820.30. Appropriate nonclinical verification and validation activities were planned and conducted in order to address identified risks and ensure the safety and effectiveness of the device. The following tests were successfully performed:

  • AAMI ANSI ES60601-1 Medical Electrical Equipment, Part I: General 0 requirements for basic safety and essential performance
  • . IEC 60601-1-2 - Medical Electrical Equipment, Part I-2: General requirements for basic safety and essential performance. Collateral Standard : Electromagnetic Compatibility
  • . IEC 60601-2-2 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • IEC 62366 Medical devices Application of usability engineering to medical ● devices.
  • ISO 11135 Sterilization of health-care products ethylene oxide requirements ● for the development, validation and routine control of a sterilization process for medical devices
  • ISO 11737-1 Sterilization of health care products Microbiological methods o Part 1: Determination of a population of microorganisms on products
  • 0 ISO 11737-2:2009 - Sterilization of medical devices -- Microbiological methods --Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process /

{11}------------------------------------------------

  • o ISO 11138-1 - Sterilization of health care products - Biological indicators - Part 1: General requirements
  • ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and 0 testing within a risk management process
  • ISO 10993-5 - Biological evaluation of medical devices - Part 5: Tests for in vitro cvtotoxicity
  • . ISO 10993-7 - Biological evaluation of medical devices - part 7: ethylene oxide sterilization residuals
  • ISO 10993-10 Biological evaluation of medical devices Part 10: Tests for 0 irritation and skin sensitization
  • ISO 10993-11 Biological evaluation of medical devices Part 11: Tests for 0 systemic toxicity
  • ASTM F1980-07 - Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
  • 0 Software verification and validation in accordance with IEC 62304 and the FDA Final Guidance document titled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
  • o Energy output verification of each of the five modes and comparison to predicate.
  • . Thermal effect comparison for representative modes of the subject device and predicate on ex-vivo bovine tissue.

Technological Characteristics

The devices are substantially equivalent to the predicate devices based on a comparison of physical and performance characteristics.

Clinical Testing (807.92(b)(2))

This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

Conclusion (807.92(b)(3))

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the similarity to the predicate devices in terms of technology, performance, and indications for use, Soniquence, LLC concludes that the Soniquence RF Generator and Soniquence Electrodes are substantially equivalent to the predicate devices as described above.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.