(109 days)
Not Found
No
The document describes a standard medical display monitor with features like high resolution, luminance, and DICOM support. There is no mention of AI or ML in the intended use, device description, or performance studies.
No
The device is a monitor intended for displaying diagnostic images, not for providing therapy or treatment.
Yes
The "Intended Use / Indications for Use" section explicitly states that the device is "intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc."
No
The device description explicitly lists hardware components such as an LCD Panel, motherboard, power cable, and data cable, in addition to software.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to detect diseases, conditions, or infections.
- Device Function: The R190/R191 LCD Monitor is a display device used to view digital images generated by imaging modalities like X-ray and MRI. It does not perform any tests on biological samples.
- Intended Use: The intended use clearly states it's for "displaying and viewing digital images for diagnosis." This is a function related to medical imaging, not in vitro testing.
The information provided describes a medical imaging display, which is a different category of medical device than an IVD.
N/A
Intended Use / Indications for Use
R190/R191 LCD Monitor is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis.
Product codes
PGY
Device Description
R190/R191 LCD Monitor is the display system with the resolution(1280*1024), high luminance (800cd/m2), and can present up to 1.07 billion colors(1024 grayscale), 1 DICOM look up table inside.
R190/R191 uses luminance stabilization mechanism to ensure brightness stability in the product service time.
The product is consisted of the following components:
-19inches, Mono-TFT LCD Panel
- M150M/P38 motherboard/FR-4
- R190/R191 LCD Monitor software
- Power cable
- Data Cable.
The LCD Monitor is designed, tested, and will be manufactured in accordance with both mandatory and voluntary standards.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray or MRI, etc.
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
trained medical practitioners
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Bench testing was conducted to demonstrate the R190/R191 meets all performance standards as follows:
- Measurement of the angular dependency of luminance response in horizontal, . vertical and diagonal directions
- Measurement of the luminance non-uniformity characteristics of the display screen as specified in TGI18 guideline.
- Measurement of the chromaticity non-uniformity characteristics of the display screen as specified in TG18 guideline.
- Measurement of small-spot contrast ratio.
- Measurement of temporal response
- . Performance data on luminance stability
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the R190/R191 The device complies with the IEC 60601-1:2012 standard for safety and the IEC 60601-1:2012 and EN 60601-1-2:2015 standard for EMC.
Animal and clinical study
The subject of this premarket submission,R190/R191, does not require animal or clinical studies to support substantial equivalence.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
510(k) Summary K183494
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: | Dec 3, 2018 |
---|---|
Submitter: | Nanjing Jusha Display Technology Co., Ltd |
Add: 8A, Block 1. Nanjing International Service Outsourcing | |
Mansion, No. 301, Hanzhongmen street, Nanjing, China. | |
Contact Person: | Zilong Liang |
Certification Manager | |
Nanjing Jusha Display Technology Co., Ltd | |
Tel: +86-25- 83305050 | |
Fax: +86-25-5878 3273 | |
Device Trade Name: | R190 LCD Monitor,R191 LCD Monitor |
Common/Usual Name: | 1MP Monochrome LCD Monitor |
Classification Name: Product | |
Code: | Display, Diagnostic Radiology 21CFR 892.2050 |
PGY | |
Predicate Device(s): | MDM 1900-1NR AND MDM 1900-1NR; K061303 |
Device Description: | R190/R191 LCD Monitor is the display system with the resolution |
(1280*1024), high luminance (800cd/m2), and can present up to 1.07 | |
billion colors(1024 grayscale), 1 DICOM look up table inside. | |
R190/R191 uses luminance stabilization mechanism to ensure | |
brightness stability in the product service time. |
The product is consisted of the following components:
-19inches, Mono-TFT LCD Panel
- M150M/P38 motherboard/FR-4
- R190/R191 LCD Monitor software
- Power cable
- Data Cable.
The LCD Monitor is designed, tested, and will be manufactured in
accordance with both mandatory and voluntary standards. |
| Intended Use: | 1. IEC 60601-1:2012,Medical electrical equipment-Part 1:General requirements for basic safety and essential performance
2. IEC 60601-1-2:2014,EN 60601-1-2:2015,Medical electrical equipment-Part 1-2:General requirements for basic safety and essential performance-Collateral Standard:Electromagnetic disturbances-Requirements and tests |
| Intended Use: | R190/R191 LCD Monitor is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis. |
| Technology: | R190/R191 LCD Monitor is the display system with the resolution monitor (1 megapixels) with electronic capabilities for evaluation of high resolution medical images,the resolution $(1280*1024)$ , high luminance (800cd/m²), and can present up to 1.07 billion colors(1024 grayscale), 1 DICOM look up table inside. R190/R191 uses luminance stabilization mechanism to ensure brightness stability in the product service time. |
| Determination of Substantial Equivalence: | Summary of Non-Clinical Tests:
The LCD Monitor complies with voluntary standards as following:
- IEC 60601-1:2012,Medical electrical equipment-Part 1:General requirements for basic safety and essential performance
- IEC 60601-1-2:2014,EN 60601-1-2:2015,Medical electrical equipment-Part 1-2:General requirements for basic safety and essential performance-Collateral Standard:Electromagnetic disturbances-Requirements and tests
R190/R191 is substantially equivalent to MDM 1900-1NR.
R190/R191 employs the maximum resolution values same as that of MDM 1900-1NR. Comparison table of the principal characteristics of 2 devices is shown in the Attachment 1.
Attachment 1
The following quality assurance measures were applied to the development of the system:
• Risk Analysis
• Requirements Reviews
• Design Reviews
• Raw materials verification |
| | Integration testing (System verification) Final acceptance testing (Validation) Performance testing (Verification) Safety testing (Verification)
Summary of Clinical Tests:
The subject of this premarket submission, LCD Monitor, did not require clinical studies to support substantial equivalence.
The proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the subject device is determined as safe and effectiveness. |
| Conclusion: | Nanjing Jusha Display Technology Co., Ltd Considers the R190/R191 LCD Monitor to be as safe, as effective, and performance is substantially equivalent to the predicate device(s). |
1
2
12.1 Product Comparison
This comparison identifies the similarities and differences of the proposed R190/R191 LCD Monitor device to the legally marketed predicate CHILIN MDM 1900-1NR LCD Monitor device to which substantial equivalency is claimed.
Attributes | Predicate Device | Proposed Device | Proposed Device | Discussion of Differences |
---|---|---|---|---|
Product | CHILIN MDM | |||
1900-1NR | R190 LCD Monitor | R191 LCD Monitor | - | |
510(k) Number | K061303 | K183494 | K183494 | - |
Display Performance/Specifications | ||||
Screen technology | 19" Monochrome TFT LCD Panel | 19" Monochrome TFT LCD Panel | 19" Monochrome TFT LCD Panel | Same |
Viewing angle (H, V) | Horizontal 170°,Vertical 17 0° | Horizontal 176°,Vertical 176° | Horizontal 176°,Vertical 176° | - |
Attribute | ||||
S | Predicate Device | Proposed Device | Proposed Device | Discussion of |
Differences | ||||
Product | CHILIN MDM | |||
1900-1NR | R190 LCD Monitor | R191 LCD Monitor | ||
510(k) | ||||
Number | K061303 | K183494 | K183494 | |
Resolutio | ||||
n | 1024 x 1280 | 1280*1024 | 1280*1024 | The difference |
between the two | ||||
devices is due to | ||||
the different | ||||
viewing directions | ||||
and does not affect | ||||
performance. | ||||
Display | ||||
area | 376.32 (H) x | |||
301.056(V) mm | 374.7840 (H) x | |||
299.8272(V) mm | 374.7840 (H) x | |||
299.8272(V) mm | Same | |||
Contrast | ||||
Ratio | 800:1 | 1000:1 | 1000:1 | - |
Response | ||||
Time | 18ms | 16ms | 16ms | - |
Scanning | ||||
frequenc | ||||
y (H; V) | 75 | |||
kHz;63.4~79.78Hz | 52 | 52~76 | ||
kHz;59~61Hz | This difference | |||
between the two | ||||
device is caused by | ||||
the different no | ||||
display area | ||||
defined by | ||||
different | ||||
manufacturers , | ||||
nothing to do with | ||||
the display | ||||
function | ||||
Recomme | ||||
nded | ||||
Luminanc | ||||
e | 1000cd/m2 | 800cd/m2 | 800cd/m2 | This item is the |
maximum | ||||
brightness of the | ||||
LCD panel. After | ||||
the curve is | ||||
corrected, the | ||||
difference is not | ||||
large and does not | ||||
affect the use. | ||||
Pixel | ||||
Pitch | 0.294x0.294 mm | 0.2928x0.2928mm | 0.2928x0.2928mm | - |
Backlight | LED | LED | LED | Same. |
Attribute | Predicate Device | Proposed Device | Proposed Device | Discussion of |
Differences | ||||
Product | CHILIN MDM | |||
1900-1NR | R190 LCD Monitor | R191 LCD Monitor | ||
510(k) | ||||
Number | K061303 | K183494 | K183494 | |
DICOM | ||||
LUT | 12-bit : 4096 | 10-bit:1024 | 10-bit:1024 | Both devices use |
grayscale extension | ||||
technology, and the | ||||
final display effect | ||||
is not much | ||||
different, which | ||||
does not affect the | ||||
use of the product. | ||||
Luminance | ||||
calibration | Built in calibration | |||
sensor provided | Built in calibration | |||
sensor provided | Built in calibration | |||
sensor provided | Same | |||
Video Signal Input | ||||
Input | ||||
signals | DVI standard | |||
D-Sub | DVI-D | |||
DisplayPort | ||||
HDMI | ||||
DVI-A | DVI-D | |||
DisplayPort | ||||
HDMI | ||||
SOG | ||||
Input | ||||
terminational | DVI-D (dual link) x | |||
1, | ||||
D-Sub x 1 | DVI-D (dual link) x 1, | |||
DisplayPort x 1 | ||||
HDMI x1 | ||||
DVI-A x 1 | DVI-D (dual link) x | |||
1, DisplayPort x 1 | ||||
HDMI x1 | ||||
SOG x 1 | - | |||
Display | ||||
controller | Off the shelf | Off the shelf | Off the shelf | Same |
Power Related Specification | ||||
Power | ||||
Requirement | AC 100~240V | AC 100~240V | AC 100~240V | Same |
Power | ||||
Consumption/Save | ||||
Mode | 47W/less than 0.5W | 25W/less than 0.5W | 25W/less than 0.5W | - |
Attributes | Predicate Device | Proposed Device | Proposed Device | Discussion of |
Differences | ||||
Product | CHILIN MDM | |||
1900-1NR | R190 LCD Monitor | R191 LCD Monitor | ||
510(k) | ||||
Number | K061303 | K183494 | K183494 | |
Power | ||||
Management | DVI DMPM | |||
DisplayPort | DVI DMPM | |||
DisplayPort | DVI DMPM | |||
DisplayPort | - | |||
Miscellaneous Features/Specifications | ||||
Service | ||||
interface | RS232 | USB | USB | This interface is |
only used to | ||||
communicate with | ||||
the host, does not | ||||
affect the | ||||
performance of the | ||||
display. | ||||
Dimensions w/o stand | ||||
(W x H x D) | Without stand: | |||
344mmx418mmx82.5mm | Without stand: | |||
419mm x343mm | ||||
x71mm | Without stand: | |||
419mm x343mm | ||||
x71mm | Different housing | |||
design due to the | ||||
different panel size. | ||||
With stand: | ||||
344mmx433mmx227 mm | With stand: | |||
419mm x379mm | ||||
x198mm | With stand: | |||
419mm x379mm | ||||
x198mm |
3
4
5
6
Attributes | Predicate Device | Proposed Device | Proposed Device | Discussion of Differences |
---|---|---|---|---|
Product | CHILIN MDM | |||
1900-1NR | R190 LCD Monitor | R191 LCD Monitor | ||
510(k) | ||||
Number | K061303 | K183494 | K183494 | |
Indication for use | MDM 1900-1NR | |||
Monitor is intended | ||||
to be used in | ||||
displaying and | ||||
viewing digital | ||||
images for diagnosis | ||||
of X-ray or MRI, | ||||
etc. by trained | ||||
medical | ||||
practitioners. The | ||||
device does not | ||||
support the display | ||||
of mammography | ||||
images for | ||||
diagnosis. | R190 LCD Monitor is | |||
intended to be used in | ||||
displaying and | ||||
viewing digital | ||||
images for diagnosis | ||||
of X-ray or MRI, etc. | ||||
by trained medical | ||||
practitioners. The | ||||
device does not | ||||
support the display of | ||||
mammography | ||||
images for diagnosis. | R191 LCD Monitor | |||
is intended to be | ||||
used in displaying | ||||
and viewing digital | ||||
images for | ||||
diagnosis of X-ray | ||||
or MRI, etc. by | ||||
trained medical | ||||
practitioners. The | ||||
device does not | ||||
support the display | ||||
of mammography | ||||
images for | ||||
diagnosis. | Same | |||
Attributes | Predicate Device | Proposed Device | Proposed Device | Discussion of Differences |
Product | CHILIN MDM | |||
1900-1NR | R190 LCD Monitor | R191 LCD Monitor | ||
510(k) Number | K061303 | K183494 | K183494 | |
Applicable standard | 1 IEC 60601-1Medical equipment medical electrical equipment - Part 1: General requirements for basic safety and essential performance 2005+CORR.1(2006)+CORR.2(2007) | |||
2 IEC 60601-1-2 Edition 3:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests. | 1.IEC 60601-1:2012,Medical electrical equipment-Part 1:General requirements for basic safety and essential performance | |||
2.IEC 60601-1-2:2014,EN 60601-1-2:2015,Medical electrical equipment-Part 1-2:General requirements for basic safety and essential performance-Collateral Standard:Electromagnetic disturbances-Requirements and tests | 1.IEC 60601-1:2012,Medical electrical equipment-Part 1:General requirements for basic safety and essential performance | |||
2.IEC 60601-1-2:2014,EN 60601-1-2:2015,Medical electrical equipment-Part 1-2:General requirements for basic safety and essential performance-Collateral Standard:Electromagnetic disturbances-Requirements and tests | - |
7
8
PERFORMANCE DATA:
The following performance data were provided in support of the substantial equivalence determination.
Bench testing:
Bench testing was conducted to demonstrate the R190/R191 meets all performance standards as follows:
- Measurement of the angular dependency of luminance response in horizontal, . vertical and diagonal directions
- Measurement of the luminance non-uniformity characteristics of the display screen as specified in TGI18 guideline.
- Measurement of the chromaticity non-uniformity characteristics of the display screen as specified in TG18 guideline.
- Measurement of small-spot contrast ratio.
- Measurement of temporal response
- . Performance data on luminance stability
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the R190/R191 The device complies with the IEC 60601-1:2012 standard for safety and the IEC 60601-1:2012 and EN 60601-1-2:2015 standard for EMC.
Animal and clinical study
The subject of this premarket submission,R190/R191, does not require animal or clinical studies to support substantial equivalence.
CONCLUSIONS
R190/R191 Medical Monitor is substantially equivalent to the predicate device with respect to technical characteristics, performance, application and intended use. The non-clinical data support the safety of the device should perform as intended in the specified use conditions. Nanjing Jusha Display Technology Co., Ltd considers the R190/R191 Medical Monitor does not raise any new issues of safety or effectiveness.
9
Image /page/9/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
Nanjing Jusha Display Technology Co., Ltd. % Zilong Liang Certification Manager Suite A, 8/F, Bldg 1, No. 301, Hanzhongmen St. Nanjing, Jiangsu 210036 CHINA
April 5, 2019
Re: K183494
Trade/Device Name: R190/R191 LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: PGY Dated: January 11, 2019 Received: January 11, 2019
Dear Zilong Liang:
We have reviewed your Section 510(k) premarket notification of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the devices, subject to the general controls provisions of the Act. Although this letter refers to your products as devices, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your devices are classified (see above) into either class II (Special Controls) or class III (PMA), they may be subject to additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
10
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
11
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name
R190/R191 LCD Monitor
Indications for Use (Describe)
R190/R191 LCD Monitor is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRJ, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
Confidential and Privileged. This document contains confidential and privileged trade secrets and other information of Nanjing Jusha Display Technology Co., Ltd and as such may not be disclosed to others not employed by Nanjing Jusha Display Technology Co., Ltd. All rights reserved.
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.