K Number
K183395

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-04-17

(131 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arthrex 2.5 mm Tenodesis Screw is intended to be used for suture or tissue fixation in the foot, ankle, knee, hand, wrist, elbow, and shoulder. Specific indications are listed below:
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

Device Description

The Arthrex 2.5 mm Tenodesis Screw is a fully threaded, cannulated screw with a rounded head. The screw will be offered in a 2.5 mm diameter and 6 mm length. The screw is manufactured from Polyetheretherketone (PEEK). The screw is sold as sterile, single-use.

AI/ML Overview

The provided text describes the regulatory filing for a medical device, the Arthrex 2.5 mm Tenodesis Screw, and its substantial equivalence to predicate devices. It does not contain information about an AI/ML-based device, nor does it detail acceptance criteria or studies related to AI/ML performance.

Therefore, I cannot provide the requested information regarding:

  • A table of acceptance criteria and reported device performance for an AI/ML device.
  • Sample sizes, data provenance, number/qualifications of experts, or adjudication methods for an AI/ML device test set.
  • Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done for an AI/ML device.
  • Whether standalone (algorithm-only) performance was done for an AI/ML device.
  • The type of ground truth used for an AI/ML device.
  • Sample size for an AI/ML device training set.
  • How ground truth for an AI/ML device training set was established.

The document discusses performance data for the Arthrex 2.5 mm Tenodesis Screw in terms of mechanical testing (pull-out and failure torque/insertion torque), sterilization (ethylene oxide with a SAL of 10⁻⁶), shelf life (5 years for PEEK material), and biocompatibility testing (Bacterial Endotoxins Test, Cytotoxicity, Sensitization, Irritation, Genotoxicity, Systemic Toxicity, Subchronic/Subacute Toxicity, Implantation, and Material Characterization). The acceptance criteria for these tests are implicitly that the new device performs "statistically equivalent" to the predicate device for mechanical properties and meets established standards (e.g., ISO 10993-1:2018 for biocompatibility, ANSI/AAMI ST72:2011/(R)2016 for sterility).

In summary, the provided document is for a traditional medical device (a mechanical screw) and does not involve AI/ML. Thus, the specific questions about AI/ML device evaluation metrics cannot be answered from this text.

{0}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K183395

Device Name Arthrex 2.5mm Tenodesis Screw

Indications for Use (Describe)

The Arthrex 2.5 mm Tenodesis Screw is intended to be used for suture or tissue fixation in the foot, ankle, knee, hand, wrist, elbow, and shoulder. Specific indications are listed below:

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction

Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendon transfers

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers, Mid-foot reconstruction

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{1}------------------------------------------------

April 17, 2019

Image /page/1/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of human services. To the right of the symbol is a blue square containing the acronym "FDA" in white letters. Next to the blue square, the words "U.S. FOOD & DRUG" are written in blue, with the word "ADMINISTRATION" appearing below in a smaller font size.

Arthrex Inc. Rebecca Homan Regulatory Affairs Associate 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K183395

Trade/Device Name: Arthrex 2.5 mm Tenodesis Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC, MBI Dated: December 5, 2018 Received: December 7, 2018

Dear Rebecca Homan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{2}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Summary or 510(k) Statement

Date PreparedJanuary 8, 2019
SubmitterArthrex Inc.1370 Creekside BoulevardNaples, FL 34108-1945
Contact PersonRebecca R. HomanRegulatory Affairs Associate1-239-643-5553, ext. 73429rebecca.homan@arthrex.com
Name of DeviceArthrex 2.5 mm Tenodesis Screw
Common NameScrew, fixation, boneFastener, Fixation, Nondegradable, Soft Tissue
Product CodeHWC; MBI
Classification Name21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener
Regulatory ClassII
Predicate DeviceK063479: Arthrex 2.5 mm PushLockK172612: Arthrex FiberTak DX (Reference)
Purpose of SubmissionThis Traditional 510(k) premarket notification is submitted to obtain clearance forthe Arthrex 2.5 mm Tenodesis Screw.
Device DescriptionThe Arthrex 2.5 mm Tenodesis Screw is a fully threaded, cannulated screw with arounded head. The screw will be offered in a 2.5 mm diameter and 6 mm length.The screw is manufactured from Polyetheretherketone (PEEK). The screw is soldas sterile, single-use.
Indications for UseThe Arthrex 2.5 mm Tenodesis Screw is intended to be used for suture or tissuefixation in the foot, ankle, knee, hand, wrist, elbow, and shoulder. Specificindications are listed below:Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral LigamentReconstructionShoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, BicepsTenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift orCapsulolabral ReconstructionHand/Wrist: Scapholunate Ligament Reconstruction, Carpal LigamentReconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexorand Extensor Tendons at the PIP, DIP and MCP joints for all digits, digital tendontransfersFoot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair,Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendontransfers, Mid-foot reconstructionKnee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair,Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial BandTenodesis
TechnologicalComparisonThe Arthrex 2.5 mm Tenodesis Screw is substantially equivalent to the predicatedevices in which the basic design features, intended use, indications for use,material, biocompatibility, method of sterilization, packaging, and shelf-life are
identical.
The Arthrex 2.5 mm Tenodesis Screw is a line extension to the predicate devices,which include minor dimensional modifications with no change to intended useor function. Any differences between the Arthrex 2.5 mm Tenodesis Screw andthe predicate devices are considered minor and do not raise different questionsof safety or effectiveness.
Performance DataPull-out and failure torque/insertion torque testing were conducted to
demonstrate that the Arthrex 2.5 mm Tenodesis Screw performs statisticallyequivalent to the predicate device cleared under K063479.
The Arthrex 2.5 mm Tenodesis Screw is sterilized by ethylene oxide (EO). Thesterility assurance level (SAL) is 10⁻⁶. This is identical to the predicate devicescleared under K063479 and K172612.
The shelf life of the Arthrex 2.5 mm Tenodesis Screw is five (5) years for the
Polyetheretherketone (PEEK) material, which is identical to the predicate devices
cleared under K063479 and K172612.
Bacterial Endotoxins Test (BET) was performed on the Arthrex 2.5 mm TenodesisScrew utilizing the Kinetic Chromogenic Method in accordance with ANSI/AAMIST72:2011/(R)2016, USP <161>, USP <85>, EP 2.6.14. Testing was performed incompliance with US FDA good manufacturing practice (GMP) regulations 21 CFRParts 210, 211 and 820. The testing conducted demonstrates that the Arthrex2.5 mm Tenodesis Screw meets pyrogen limit specifications.
Cytotoxicity, Sensitization, Irritaion, Genotoxicity, Systemic Toxicity,
Subchronic/Subacute Toxicity, Implantation and Material Characterization testingwas conducted on the Arthrex 2.5 mm Tenodesis Screw in accordance with ISO10993-1:2018.
ConclusionThe Arthrex 2.5 mm Tenodesis Screw is substantially equivalent to the predicatedevices in which the basic design features and intended uses are the same. Anydifferences between the proposed device and the predicate devices are
considered minor and do not raise different questions of safety or effectiveness.
Based on the indications for use, technological characteristics, and the summary
of data submitted, Arthrex Inc. has determined that the Arthrex 2.5 mm
Tenodesis Screw is substantially equivalent to the currently marketed predicatedevice.

{4}------------------------------------------------

N/A