K Number
K172612

Validate with FDA (Live)

Device Name
FiberTak DX
Manufacturer
Date Cleared
2017-09-28

(28 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FiberTak DX is intended to be used for suture or tissue fixation in the foot/ankle, knee, hand/wrist, elbow, and shoulder. Specific indications are listed below:
• Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament
Reconstruction
• Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps
Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or
Capsulolabral Reconstruction
• Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of
collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and
MCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstruction
and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty)
• Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair,
Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers,
Mid-foot reconstruction
• Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair,
Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band
Tenodesis

Device Description

The FiberTak DX is an "all-suture" soft-tissue fixation device. The anchor and connected sutures are impacted into a pilot hole. The sutures are then manually tensioned to set the anchor by "bulging" the suture sleeve within the pilot hole.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called "FiberTak DX." This document primarily focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a study proving the device meets specific performance acceptance criteria for an AI/algorithm-based diagnostic device.

Therefore, the requested information regarding acceptance criteria, study design for AI evaluation (sample sizes, expert involvement, ground truth, MRMC study, etc.), and training data for an algorithm cannot be extracted from the provided text.

The "Performance Data" section briefly mentions:

  • Pull-out testing: "demonstrated that the pull out displacements of the proposed FiberTak DX met the criteria established by the predicate device."
  • Bacterial endotoxin testing: "conducted to demonstrate that the device meets pyrogen limit specifications."

This indicates that a mechanical performance study (pull-out strength/displacement) and a biocompatibility/sterility-related test (bacterial endotoxin) were performed. However, the specific numerical acceptance criteria for these tests and the detailed results are not provided. The document states they "met the criteria established by the predicate device" and "meets pyrogen limit specifications," but does not give the values.

To directly answer your request based on the provided text, I can only indicate what information is not present:

  1. Table of acceptance criteria and reported device performance for an AI/algorithm: Not applicable as this is a physical medical device (suture anchor), not an AI/algorithm-based diagnostic device. The document mentions pull-out testing and bacterial endotoxin testing, but no specific numerical acceptance criteria or detailed performance values are given.
  2. Sample size for the test set and data provenance: Not applicable for an AI/algorithm. For the physical device tests (pull-out, endotoxin), sample sizes are not specified.
  3. Number of experts used to establish ground truth and their qualifications: Not applicable.
  4. Adjudication method for the test set: Not applicable.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable.
  6. Standalone (algorithm-only) performance: Not applicable.
  7. Type of ground truth used: Not applicable for an AI/algorithm. For the physical device, the "ground truth" would be established by validated mechanical testing methods (for pull-out) and chemical/biological testing methods (for endotoxin), not clinical experts.
  8. Sample size for the training set: Not applicable.
  9. How ground truth for the training set was established: Not applicable.

In summary, the provided document is a 510(k) clearance for a physical medical device (suture anchor), not an AI/algorithm diagnostic device. Therefore, the detailed acceptance criteria and study information typically associated with AI/algorithm validation, as requested in your prompt, are not present in this document.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows a partial view of the Department of Health & Human Services USA logo. The logo features the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern. To the right of the text are three stylized human profiles facing right, stacked one behind the other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 28, 2017

Arthrex, Inc. Ivette Galmez Senior Regulatory Affairs Specialist 1370 Creekside Boulevard Naples, Florida 34108

Re: K172612

Trade/Device Name: FiberTak DX Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: August 29, 2017 Received: August 31, 2017

Dear Ms. Galmez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Katherine D. Kavlock -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

PSC Phibloding Ser we = (101) + + + + + +리크။ ၂၀၂ ခုနှစ်မျ(THT) FBBC ADS MAOF
11əyəldən bir baş bəy vilan bir və yerlər. Şərən və bir birinci və ya çəkili və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və birəri gübuləni ,əsrında ən ən ən ən bərəmizə əsi növlərinə əsasən və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və və:of nebud sinibules not suoitsegous guibulan this polision in sing toand and one to stamilies newud sint guidement should in noilement were andto noticellos s ,o) buodzer of beninger ton si nosnoge 10 toubros jou yem vonege aff້ອຍ ໄດ້ ເວີລ ແດ່ມ່ວນປະຊາ ສາດຫາກວາມນຸກອາ ວ່າ ຢູ່ກັບ ຂອງປີ ທຸກປະຊາທິປະສາທິປະຊາ ຂາຕັ้ายประเทศรอว่างเลข และแมนี bus ที่ไรอH ใน เทือกที่สุธยอOffice of Chief Information OfficesFood Starmber Drug AministrationPRASSISTIng@gda.hhs.goov
are and and en all and en and eventing of
O Over-The-Counter Use (2) CFR 80 (Subpart C)Type of Use (Select one or poth, as applicable)(0 กิธยุตินธ 108 Я 73 15 กิธ9) อยป กอเเชตการอยค 2
o
Posteriorrindən məbilər ildə 17 illər 17 illər 14 illər 14 ildən 11 ildən 11 ildən 1997 və 1
Repair, Halluxຕາຍປ່າງແປງເທດຈາກ 1001-bilM ,ຂາອາຣາສາຕາ ແດນຫຼາຍ ໄຮ່ເຮັງປະ ,ກວ່າວມາວມາຊາດວອາ ຂນງສໄປırısmasında sinfindən səhlidə Azərbaycan filmləri ildən çıxır. Bir və inzibati ərazisi ilə bir və bir Avov
อนทรายา 1 เรตอร์ เลขอเมริกา( visual in the suite the sed) visular the province in the program and
งการเจร รายไมว่เหตุเวอร์เทอะ ค.ศ. 25 เมริการ 2018 วันเทครี 2 เหตุร 8 ทิตร์ที่ 2 เหตุร 8 วันตรี 2 เรตุลา 2 รายชนิด 2uoilsunsunssan lerdslolusqs. Constructions Constructions of insquest
າເວນາລາວບາງ ເຕືອນຕຣອງໄປ ໂຮງອາໄປວົງ ໄຣເວລາ 10 ເຄາະນາໄປ ລາວການວນາງອີງ : ເບີດເວົ້າ 2011-01-11
ashuoda bar, wodle , trivitation , sen't , plants , and , and , and , and , and , and , and , and , aSpecific indications are listed below:
(ndications for Use (Describe)
FiberTak DXDevice Name
510(k) Number (if known)K172612
Form Approved: OMB Not 0910-0120Expiration Date: 06/30/2020See PAR Statement belowDEPATMENT OF HEALTH AND HUNAN SEPVICESndications for UseFood and Drug Administration

{3}------------------------------------------------

1. 510(k) Summary

Date PreparedSeptember 21, 2017
SubmitterArthrex Inc.1370 Creekside BoulevardNaples, FL 34108-1945
Contact PersonIvette GalmezSenior Regulatory Affairs Specialist1-239-643-5553, ext. 71263Ivette.galmez@arthrex.com
Name of DeviceFiberTak DX
Common NameSoft Tissue Fixation Device
Product CodeMBI
Classification Name21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener
Regulatory ClassII
Predicate DeviceK140476: Arthrex FiberTak Suture Anchor
Purpose of SubmissionThis Special 510(k) premarket notification is submitted to obtain clearance for a lineextension to the Arthrex FiberTak family cleared under K151230.
Device DescriptionThe FiberTak DX is an "all-suture" soft-tissue fixation device. The anchor andconnected sutures are impacted into a pilot hole. The sutures are then manuallytensioned to set the anchor by "bulging" the suture sleeve within the pilot hole.
Indications for UseThe FiberTak DX is intended to be used for suture or tissue fixation in the foot/ankle,knee, hand/wrist, elbow, and shoulder. Specific indications are listed below:• Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral LigamentReconstruction• Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, BicepsTenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift orCapsulolabral Reconstruction• Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction ofcollateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP andMCP joints for all digits, digital tendon transfers, Carpal Ligament Reconstructionand Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty)• Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair,Metatarsal Ligament Repair, Hallux Valgus reconstruction, digital tendon transfers,Mid-foot reconstruction• Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair,Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial BandTenodesis
Performance DataPull-out testing demonstrated that the pull out displacements of the proposedFiberTak DX met the criteria established by the predicate device.Bacterial endotoxin per EP 2.6.14 / USP <85> was conducted to demonstrate that thedevice meets pyrogen limit specifications.
ConclusionThe FiberTak DX is substantially equivalent to the predicate device in which the basicdesign features and intended uses are the same. Any differences between theproposed device and the predicate device are considered minor and do not raisequestions concerning safety or effectiveness.Based on the indications for use, technological characteristics, and the summary ofdata submitted, Arthrex Inc. has determined that the proposed device is substantiallyequivalent to the currently marketed predicate device.

N/A