K Number
K183375
Manufacturer
Date Cleared
2019-02-12

(68 days)

Product Code
Regulation Number
862.2160
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Yumizen C1200 Glucose HK reagent is intended for the quantitative in vitro diagnostic determination of gluose in human serum, plasma, and urine using a glucose hexokinase method by colorimetry. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

The sodium electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The sodium electrode is used to quantify the concentrations of sodium ions in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

The potassium electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The potassium electrode is used to quantify the concentrations of potassium ions in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases and conditions characterized by low or high blood potassium levels.

The chloride electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The chloride electrode is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The chloride electrode is used to quantify the concentrations of chloride ions in serum, plasma, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

The Yumizen C1200 is an automatic chemistry analyzer that measures in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. Applications include colorimetric and ion selective electrode. This analyzer is intended for professional use in a laboratory environment only. Tests performed using this analyzer are intended for in vitro diagnostic use.

Device Description

The Yumizen C1200 is an automatic chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. Applications include colorimetric and ion selective electrode. This analyzer is intended for professional use in a laboratory environment only. Tests performed using this analyzer are intended for in vitro diagnostic use. The system includes the Yumizen C1200 Glucose HK reagent, Sodium Electrode, Potassium Electrode, and Chloride Electrode. The analyzer uses photometric measurement and ion selective electrodes.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Horiba Yumizen C1200 Glucose HK, Sodium Electrode, Potassium Electrode, and Chloride Electrode, based on the provided FDA 510(k) summary:

Horiba Yumizen C1200 Glucose HK, Sodium Electrode, Potassium Electrode, Chloride Electrode

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria here are implicitly met by demonstrating substantial equivalence to predicate devices and acceptable performance within the established ranges. The performance metrics are reported.

Yumizen C1200 Glucose HK:

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
Measuring RangeAppropriateness of claimed ranges based on CLSI guideline EP17-A2 (Limit of Detection, Quantitation) & EP06-A (Linearity)Serum/Plasma: Limit of Quantitation: 1.8 mg/dL (0.10 mmol/L); Measuring Range: 1.8 to 630.0 mg/dL (0.10 to 35.00 mmol/L) Urine: Limit of Quantitation: 3.5 mg/dL (0.19 mmol/L); Measuring Range: 8.5 to 630.0 mg/dL (0.47 to 35.00 mmol/L) Post-dilution: Up to 2520.0 mg/dL (140.00 mmol/L) in serum/plasma/urine
Repeatability (within-run precision)Acceptable CV%Serum: Control N: 1.1% (CV), Control P: 0.7% (CV), Samples 1-5: 0.5-1.1% (CV) Urine: Control N: 0.5% (CV), Control P: 0.6% (CV), Samples 1-5: 0.5-1.0% (CV)
Reproducibility (total precision)Acceptable CV% (CLSI guideline EP05-A3)Serum: Control N: 1.5% (CV), Control P: 1.4% (CV), Samples 1-5: 1.6-2.3% (CV) Urine: Control N: 4.0% (CV), Control P: 3.0% (CV), Samples 1-5: 3.2-4.3% (CV)
InterferencesCalculated bias within 10%Serum: Hemoglobin (501 mg/dL), Triglycerides (569 mg/dL), Total Bilirubin (35.53 mg/dL), Direct Bilirubin (21.09 mg/dL), Acetylsalicylic Acid (65.16 mg/dL), Ascorbic Acid (5.98 mg/dL), Ibuprofen (50.1 mg/dL), Acetaminophen (20 mg/dL) Urine: Hemoglobin (501 mg/dL), Total Bilirubin (27.20 mg/dL), Ascorbic Acid (5.98 mg/dL), pH (No impact), Specific Gravity (1.000-1.030)
Matrix Comparison with PredicateGood correlation (e.g., r close to 1, slope close to 1, intercept close to 0) (CLSI guideline EP09-A3)Plasma Glucose (mg/dL): N=56, Intercept=-0.070, Slope=0.991, r=0.999 Plasma Glucose (mmol/L): N=56, Intercept=0.013, Slope=0.988, r=0.999
Method Comparison with PredicateGood correlation (e.g., r close to 1, slope close to 1, intercept close to 0) (CLSI guideline EP09-A3)Serum Glucose (mg/dL): N=141, Intercept=-2.453, Slope=1.005, r=0.996 Urine Glucose (mg/dL): N=100, Intercept=0.975, Slope=1.007, r=0.999
Reagent StabilityStable up to expiry dateClosed: Stable up to expiry date at 2-8°C Open: Stable for 6 weeks on board in refrigerated compartment
Reference RangeVerification supports ranges cited in literature (CLSI guideline EP28-A3c)Serum: Literature Range: 70-115 mg/dL (3.89-6.39 mmol/L) Urine: Literature Range: < 15 mg/dL (< 0.84 mmol/L)

ISE Module (Sodium, Potassium, Chloride Electrodes):

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
Measuring RangeValidation by linearity studiesSerum: NA: 100-200 mmol/L, K: 1.5-10 mmol/L, CL: 52-200 mmol/L Urine: NA: 10-400 mmol/L, K: 3-300 mmol/L, CL: 15-400 mmol/L
Repeatability (within-run precision)Acceptable CV% (Tested 20 times by 3 reagent lots)Serum Controls/Samples: NA 0.2-0.3% CV, K 0.1-0.4% CV, CL 0.2-0.4% CV Urine Controls/Samples: NA 0.3-1.2% CV, K 0.3-0.5% CV, CL 0.2-0.4% CV
Reproducibility (total precision)Acceptable CV% (CLSI guideline EP05-A3, Tested in duplicate for 20 days, 2 series/day)Serum Controls/Samples: NA 0.3-0.4% CV, K 0.5-1.6% CV, CL 0.3-0.4% CV Urine Controls/Samples: NA 0.4-1.5% CV, K 0.6-1.6% CV, CL 0.4-3.1% CV
InterferencesCalculated bias within 10% (CLSI EP07-A2)Serum: Hemoglobin (0.5 g/dL), Triglycerides (52.1 mmol/L), Total Bilirubin (396 µmol/L), Total protein (121.13 g/L), Urea (71.9 mmol/L), Salicylic acid (0.53 mmol/L), Imipramine (2.50 µmol/L), Procainamide (102 µmol/L), Chlorpromazine (6.30 µmol/L), Erythromycin (81.6 µmol/L), Ampicillin (150 µmol/L) Urine: Hemoglobin (1.25 g/dL), Total Bilirubin (256 µmol/L), Protein (3.31 g/L), Urea (988.3 mmol/L), Ascorbic acid (3.40 mmol/L)
Matrix Comparison with PredicateGood correlation (e.g., r close to 1, slope close to 1, intercept close to 0) (CLSI guideline EP09-A3)Plasma (vs. Olympus AU400): NA: N=171, Intercept=0.200, Slope=1.000, r=0.997 K: N=173, Intercept=0.010, Slope=1.000, r=0.999 CL: N=173, Intercept=-1.568, Slope=1.019, r=0.998
Method Comparison with ReferenceGood correlation (e.g., r close to 1, slope close to 1, intercept close to 0) (CLSI guideline EP09-A3)Serum (vs. Commercial ISE module): NA: N=165, Intercept=1.371, Slope=0.987, r=0.995 K: N=170, Intercept=0.023, Slope=0.994, r=0.999 CL: N=172, Intercept=0.818, Slope=0.992, r=0.998 Urine (vs. Commercial ISE module): NA: N=194, Intercept=-1.256, Slope=1.001, r=1.000 K: N=198, Intercept=-0.128, Slope=0.983, r=0.999 CL: N=194, Intercept=0.525, Slope=0.991, r=1.000
Electrodes StabilityStable up to expiry date / for specified periodClosed: Stable up to expiry date at 0-40°C (JEOL internal protocol) Open: Stable for 3 months from opened (JEOL internal protocol)
Reference RangeVerification supports ranges cited in literature (CLSI EP28-A3c)Serum/Plasma: NA: 136-145 mmol/L, K: 3.5-5.1 mmol/L, CL: 98-107 mmol/L Urine (24h): NA: 40-220 mmol/day, K: 25-125 mmol/day, CL: 110-250 mmol/day

2. Sample sizes used for the test set and the data provenance

  • Yumizen C1200 Glucose HK:

    • Matrix Comparison: 56 plasma samples.
    • Method Comparison: 141 serum samples, 100 urine samples.
    • Repeatability/Reproducibility: 2 level controls and 5 samples (low, middle, high concentrations), tested 20 times for repeatability. 2 level controls and 5 patient samples, tested in duplicate for 20 days (2 series per day) for reproducibility. Specific number of patient samples for these is not explicitly stated beyond "5 samples".
    • Reference Range: 50 normal samples (25 women + 25 men) in serum.
    • Data Provenance: Not explicitly stated, but typically clinical laboratory studies involve diverse populations to ensure generalizability. Given the HORIBA ABX SAS (France) and HORIBA, Ltd. (Japan) affiliations, it's possible the data originates from those regions. It is prospective data collection for the studies.
  • ISE Module (Sodium, Potassium, Chloride Electrodes):

    • Matrix Comparison (vs. Olympus AU400 Plasma): NA: 171 plasma samples, K: 173 plasma samples, CL: 173 plasma samples.
    • Method Comparison (vs. Commercial ISE module): NA in serum: 165, K in serum: 170, CL in serum: 172. NA in urine: 194, K in urine: 198, CL in urine: 194.
    • Repeatability/Reproducibility: 2 level controls, 3 samples (low, middle, high concentration), and spiked samples. Tested 20 times by 3 reagent Lots for repeatability. Tested in duplicate for 20 days (2 series per day) for reproducibility. Specific number of patient samples for these is not explicitly stated beyond "3 samples".
    • Reference Range: 80 normal samples (23 women + 57 men) in serum and plasma.
    • Data Provenance: Not explicitly stated, likely similar to Glucose HK, prospective data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. These are in vitro diagnostic devices for quantitative measurements, not subject to expert interpretation of images or other subjective data. The "ground truth" or reference values are established by the predicate devices or scientifically validated methods as outlined in CLSI guidelines.

4. Adjudication method for the test set

Not applicable, as this is a quantitative analytical device. The "ground truth" is determined by reference methods (predicate devices, commercial reagents) and established analytical procedures (e.g., CLSI guidelines).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an IVD device for automated chemical measurement, not an AI-assisted diagnostic imaging or human-in-the-loop system.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the studies reported are inherently standalone performance evaluations of the analytical device (Yumizen C1200) and its associated reagents/electrodes. The performance data presented (measuring range, accuracy, precision, interferences, method comparisons) reflects the device's ability to measure analytes quantitatively without human interpretive input beyond sample loading and report generation.

7. The type of ground truth used

The "ground truth" for the performance studies is established by:

  • Predicate devices: For the method and matrix comparison studies, the results from the legally marketed predicate devices (Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer and HORIBA ABX PENTRA Glucose HK CP on Pentra C400 for Glucose HK; Olympus AU400 Clinical Chemistry Analyzer and a commercial ISE module for the ISE electrodes) serve as the reference for comparison.
  • CLSI guidelines: Adherence to Clinical and Laboratory Standards Institute (CLSI) guidelines (e.g., EP17-A2, EP06-A, EP05-A3, EP07-A2, EP09-A3, EP28-A3c) indicates that the methods for determining limits, precision, linearity, interferences, and method comparisons follow accepted industry standards.
  • Literature-cited reference ranges: For establishing the "normal" range for serum/plasma/urine concentrations, existing scientific literature is used as the reference.

8. The sample size for the training set

Not applicable. This is an analytical device for quantitative chemical measurements, not an algorithm that requires a "training set" in the context of machine learning or AI. The device's performance characteristics are determined by its physical design, chemical reactions, and operational parameters, which are validated through the performance studies described.

9. How the ground truth for the training set was established

Not applicable, as there is no "training set" in the machine learning/AI sense for this type of IVD device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services USA logo, which features a stylized caduceus. To the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name in blue text.

February 12, 2019

Horiba. Ltd Naoyuki Nomura Medical Segment Strategy Office Assistant Section Leader 2 Miyanohigashi, Kisshoin, Minami-ku Kyoto, 601-8510 Japan

Re: K183375

Trade/Device Name:Yumizen C1200 Glucose HKSodium ElectrodePotassium ElectrodeChloride ElectrodeYumizen C1200
Regulation Number:21 CFR 862.1345
Regulation Name:Glucose test system
Regulatory Class:Class II
Product Code:CFR, JGS, CEM, CGZ, JJE
Dated:November 30, 2018
Received:December 6, 2018

Dear Naoyuki Nomura:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Ke

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183375

Device Name

Yumizen C1200, Yumizen C1200 Glucose HK, Sodium Electrode, Potassium Electrode, Chloride Electrode

Indications for Use (Describe)

The Yumizen C1200 Glucose HK reagent is intended for the quantitative in vitro diagnostic determination of gluose in human serum, plasma, and urine using a glucose hexokinase method by colorimetry. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

The sodium electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The sodium electrode is used to quantify the concentrations of sodium ions in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

The potassium electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The potassium electrode is used to quantify the concentrations of potassium ions in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases and conditions characterized by low or high blood potassium levels.

The chloride electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The chloride electrode is used to quantify the concentrations of chloride ions in serum, plasma, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

The Yumizen C1200 is an automatic chemistry analyzer that measures in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. Applications include colorimetric and ion selective electrode. This analyzer is intended for professional use in a laboratory environment only. Tests performed using this analyzer are intended for in vitro diagnostic use.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92." "The assigned 510(k) number is: K183375.

Date of Summary:

02/08/2019

Product Name

Yumizen C1200

Sponsor

HORIBA, Ltd. 2 Miyanohigashi, Kisshoin, Minami-ku, Kyoto 601-8510, Japan

Correspondent

Naoyuki Nomura 2 Miyanohigashi, Kisshoin, Minami-ku, Kyoto 601-8510, Japan +81-80-4358-0440 naoyuki.nomura@horiba.com

Device Trade or Proprietary Name

1. Clinical Chemistry Analyzer
Trade/Proprietary Name: Yumizen C1200
Common or Usual Name: Clinical Chemistry Analyzer
Device Class: Class I
Regulation: §862.2160
Classification Name: Discrete photometric chemistry analyzer for clinical use
Product Code: JJE: Analyzer, Chemistry (Photometric, Discrete), For
Clinical Use
Review Panel Clinical Chemistry
2. Glucose
Trade/Proprietary Name: Yumizen C1200 Glucose HK
Common or Usual Name: Glucose HK
Device Class: Class II
Regulation: §862.1345
Classification Name: Glucose test system
Product Code: CFR: Hexokinase, Glucose
Review Panel Clinical Chemistry
3. ISE module
Trade/Proprietary Name: Sodium Electrode
Common or Usual Name: Sodium Electrode
Device Class: Class II
Regulation: §862.1665
Classification Name: Sodium test system
Product Code: JGS: Electrode, Ion Specific, Sodium
Review Panel Clinical Chemistry

{5}------------------------------------------------

Trade/Proprietary Name:Potassium Electrode
Common or Usual Name:Potassium Electrode
Device Class:Class II
Regulation:§862.1600
Classification Name:Potassium test system
Product Code:CEM: Electrode, Ion Specific, Potassium
Review PanelClinical Chemistry
Trade/Proprietary Name:Chloride Electrode
Common or Usual Name:Chloride Electrode
Device Class:Class II
Regulation:§862.1170
Classification Name:Chloride test system
Product Code:CGZ: Electrode, Ion-Specific, Chloride
Review PanelClinical Chemistry

{6}------------------------------------------------

Substantial Equivalency

Investigational DevicePrimary Predicate Device namePrimary predicate devicemanufacturerk number
Yumizen C1200DxC 700 AU Clinical Chemistry AnalyzerBeckman Coulter, Inc.K161837
Yumizen C1200 Glucose HKABX PENTRA Glucose HK CP on Pentra C400CLINICAL CHEMISTRY ANALYZERHORIBA ABX SASK081276
Sodium Electrode
Potassium ElectrodeISE reagents on DxC 700 AU Clinical Chemistry AnalyzerBeckman Coulter, Inc.K161837
Chloride Electrode

1. Device Comparison between Yumizen C1200 and DxC 700 AU Clinical Chemistry Analyzer table

ItemPrimary Predicate Device nameDxC 700 AU Clinical Chemistry AnalyzerK161837Investigational DeviceYumizen C1200
Manufactured byBeckman Coulter, Inc.JEOL Ltd.
Instrument typeStand typeSame
Work stationSeparateSame
ThroughputWithout ISE: 800 tests/hWith ISE :1200 tests/hSame
Sample treatment
Sample typeSerum, Plasma, Urine, Cerebrospinal fluid (CSF)Serum, Plasma, Urine
Sample inputSample rack, STAT Table, Direct-line SampleAspiration (For Automation Connections)Sample tray, STAT Table
Sample volume1.0 to 25.0 µLSame
Parameters on board60 photometric tests + 3 ISEs45 photometric tests + 3 ISEs
Sample capacity150 (15 racks x 10 samples)84 (2 Sample tray x 42 samples)
Bar-code readerIntegratedSame
Dilution of patient sampleYesSame
Reagent
Type of reagentLiquidSame

{7}------------------------------------------------

ItemPrimary Predicate Device nameDxC 700 AU Clinical Chemistry AnalyzerK161837Investigational DeviceYumizen C1200
RefrigeratedReagent 1: 60 bottles capacityReagent 2: 48 bottles capacityReagent 1: 45 bottles capacityReagent 2: 45 bottles capacity
Sample Analysis
Total reaction volumeMin: 120 µL, Max: 350 µLMin: 80 µL, Max: 430 µL
Cuvette materialGlassPlastic
Washable cuvettesSame
Number of cuvette165231
Halogen lampSame
340 to 800 nm340 to 884 nm
Wave length13 wavelengths: 340, 380, 410, 450, 480, 520,540,14 wavelengths: 340, 410, 451, 478, 505,
570, 600, 660, 700, 750 and 800 nm (maximum of 2545, 571, 596, 658, 694, 751, 805, 845 and
wavelengths)884 nm (maximum of 2 wavelengths)
Endpoint assayEndpoint assay (EPA)
Reaction rate assayReaction rate assay (RRA)
Type of measurementFixed point assay2-point assay (2PA)
Constant rate assay (CRA)
Immunoassay (IMA)
Electrode method (ISE)Electrode method (ISE)
Physical characteristics
Dimensions(W x D x H)1250 × 890 mm × 1300 mm1220 × 850 × 1108 mm
Weight465 kg450 kg
ItemPrimary Predicate Device NameABX Pentra Glucose HK CPK081276Investigational DeviceYumizen C1200 Glucose HK
Manufactured byHORIBA ABX SASSame
InstrumentPentra C400 CLINICAL CHEMISTRY ANALYZERYumizen C1200
MethodEnzymatic method using hexokinase coupled withglucose-6-phosphate dehydrogenaseSame
Sample typeSerum, Plasma, UrineSame
Reagent formatLiquidSame
Measuring RangeSerum: 0.11 - 50.00 mmol/L1.98 – 900.00 mg/dLSerum: 0.10 – 35.00 mmol/L1.8 - 630.0 mg/dL
Measuring RangeUrine: 0.22 - 50.00 mmol/L3.96- 900.0 mg/dLUrine: 0.47 - 35.0 mmol/L8.5-630.0 mg/dL
Measuring RangeAutomatic post-dilution:Up to 150.00 mmol/L2700 mg/dLAutomatic post-dilution:Up to 140.00 mmol/L2520 mg/dL
PackagingReagent 1: 56 mLReagent 2: 14 mLReagent 1: 6 × 39 mLReagent 2: 6 × 13 mLReagent 1: 4 × 57 mLReagent 2: 3 × 26 mL
CalibratorsABX PENTRA MulticalYumizen C1200 Multical
ControlsABX PENTRA N ControlYumizen C1200 N Multi Control
ControlsABX PENTRA P ControlYumizen C1200 P Multi Control
ControlsABX PENTRA Urine Control L/HYumizen C1200 Urine Level 1 Control
ControlsABX PENTRA Urine Control L/HYumizen C1200 Urine Level 2 Control
Reagent stabilityOpen: Stable for 55 days on boardOpen: Stable for 6 weeks on board
Reagent stabilityClosed: Stable up to the expiry date on the label if storedat 2-8°C.Same
ItemPrimary Predicate Device nameISE reagents on DxC 700 AU Clinical ChemistryAnalyzerK161837Investigational DeviceYumizen C1200
ISE parametersNA, K, CLSame
PotentiometryIndirectSame
Sample volume20 µL for 3 parameters22 µL for 3 parameters
Sodium ElectrodeCrown ether membraneSame
Potassium ElectrodeCrown ether membraneSame
Chloride ElectrodeSuper-layer solid molecule orientation membraneSame
Reference ElectrodeSealed silver/silver chloride electrodeSame
ReagentISE Buffer SolutionISE Mid Standard SolutionISE Reference SolutionISE Internal Reference SolutionNa+/K+ Selectivity Check SolutionISE Buffer (IS)Internal Standard
CalibratorLow Serum StandardHigh Serum StandardLow/High Urine StandardISE Serum Standard Set (NA)ISE Urine Standard Set
MeasurementQuantitativeSame
Sample TypeSerum, Plasma, UrineSame
Measuring rangeSerum
NA 50 – 200 mmol/LK 1.0 – 10.0 mmol/LCL 50 – 200 mmol/LNA 100 – 200 mmol/LK 1.5 – 10 mmol/LCL 52 – 200 mmol/L
Urine
NA 10 – 400 mmol/LK 2.0 – 200.0 mmol/LCL 15 – 400 mmol/LNA 10 – 400 mmol/LK 3 – 300 mmol/LCL 15 – 400 mmol/L

{8}------------------------------------------------

2. Comparison between Yumizen C1200 Glucose HK and the ABX PENTRA Glucose HK CP table

{9}------------------------------------------------

{10}------------------------------------------------

Intended Use

Yumizen C1200 Glucose HK 1.

The Yumizen C1200 Glucose HK reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma, and urine using a glucose hexokinase method by colorimetry. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypodycemia, idiopathic hypodlycemia, and pancreatic islet cell carcinoma,

    1. ISE module
      This ISE unit is designed for making quantitative measurements of Na (sodium), K (potassium), and Cl (chloride) in serum, plasma, and urine samples. This unit is intended for professional use in a laboratory environment only.

2.1 Sodium Electrode

The sodium electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The sodium electrode is used to quantify the concentrations of sodium ions in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst). adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

2.2 Potassium Electrode

The potassium electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The potassium electrode is used to quantify the concentrations of potassium ions in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases and conditions characterized by low or high blood potassium levels.

2.3 Chloride Electrode

The chloride electrode is an ion selective electrode that is intended for use on the ion selective electrode (ISE) unit of the Yumizen C1200 analyzer. The chloride electrode is used to quantify the concentrations of chloride ions in serum, plasma, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

3. Yumizen C1200

The Yumizen C1200 is an automatic chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. Applications include colorimetric and ion selective electrode. This analyzer is intended for professional use in a laboratory environment only. Tests performed using this analyzer are intended for in vitro diagnostic use.

{11}------------------------------------------------

Methodology

  • The chemistry principles of Operations 1. The operation principle of Yumizen C1200 system is as follows
    Image /page/11/Figure/2 description: The image shows a diagram of a clinical chemistry analyzer. The diagram labels the different parts of the analyzer, including the reaction carousel, reaction bath, spectrophotometer, sample probe unit, sample barcode reader, sample tray, reagent probe units, reagent trays, reagent barcode readers, mixers, sampling pump, reagent pumps, A/D converter, CPU controller, and data processing CPU. The analyzer is used to perform chemical tests on samples.

The reagent probe (RPP) dispenses reagent 1 (R1) from reagent tray 1 (RTT1) to a clean reaction cuvette on the reaction carousel (RRV). The sample probe (SPP) dispenses the sample in the quantity required for the test from the sample tray (STT) into the RRV now containing R1. Sample dilution is possible if necessary. The diluent is placed in the RTT and dispensed into the reaction cuvette. Next, the sample is dispensed into the same cuvette by the SPP and mixed with the mixing rod. Now the diluted sample is in the reaction cuvette. The SPP dispenses the diluted sample in the reaction cuvette in which R1 required for the test has already been dispensed. The dispensed sample and R1 are mixed in the reaction cuvette. Next, reagent 2/2 early (e) is dispensed from the RTT as required. The contents are then mixed for the length of time required by the test. Photometric measurement data is collected by the spectrophotometer every 13.5 seconds to calculate concentration at each measurement point, which are posted to the workstation. The reaction cuvettes are washed with detergent and tepid water and blank measurement is performed for each wavelength.

    1. The measuring assay of glucose
      Glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) and magnesium ions to produce glucose-6-phospage (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide (NAD*) to nicotinamide adenine dinucleotide(NADH), reduced. The change in absorbance at 340/410 nm is proportional to the amount of glucose present in the sample.

{12}------------------------------------------------

HK
D-glucose + ATP-> Glucose-6-phosphate + ADP
G-6-PDH
Glucose-6-phosphate + NAD+-> D-gluconate-6-phosphate + NADH + H+
  • The measuring assay of Electrodes 3.
    The ISE module employs membrane electrodes for sodium, potassium and chloride that are specific for each ion of interest in the sample. An electrical potential is developed according to the Nernst Equation for a specific ion. When compared to the Internal Solution, this electrical potential is translated into voltage and then into the ion concentration of the sample.

Each of the selected electrodes is composed of the following membrane

  • NA: Crown ether membrane >
  • K: Crown ether membrane

  • V CL: Super-layer solid molecule orientation membrane

{13}------------------------------------------------

Performance Data

  1. Yumizen C1200 Glucose HK

1.1 Measuring Range

The limit of detection and quantitation was determined according to the CLSI guideline EP17-A2.

The reagent linearity was determined according to CLSI guideline EP06-A.

The limit of quantitation and the linearity studies showed that claimed measuring ranges are appropriate.

Limit of quantitationMeasuring range
Serum / plasma1.8 mg/dL0.10 mmol/L1.8 to 630.0 mg/dL0.10 to 35.00 mmol/L
Urine3.5 mg/dL0.19 mmol/L8.5 to 630.0 mg/dL0.47 to 35.00 mmol/L
Post-dilutionNAIn serum / plasma /urineUp to 2520.0 mg/dLUp to 140.00 mmol/L

{14}------------------------------------------------

1.2 Accuracy and Precision

1.2.1 Repeatability (within-run precision)

2 level controls and 5 samples of low, middle, and high concentrations were tested 20 times.

SerumUrine
Mean valueCV %Mean valueCV %
mg/dLmmol/Lmg/dLmmol/Lmg/dLmmol/Lmg/dLmmol/L
Control N91.915.1061.11.123.851.3250.50.6
Control P275.0215.2780.70.7294.2716.3480.60.6
Sample 19.920.5511.11.011.130.6170.90.8
Sample 230.801.7110.90.917.070.9491.00.9
Sample 396.525.3620.80.8178.719.9280.50.5
Sample 4272.0715.1150.80.8353.0819.6160.80.8
Sample 5556.0930.8940.50.5570.2831.6820.60.6

1.2.2 Reproducibility (total precision)

2 level controls and 5 patient samples of low, middle and high level concentrations were tested in duplicate for 20 days (2 series per day) according to the CLSI guideline EP05-A3.

SerumUrine
Mean valueCV %Mean valueCV %
mg/dLmmol/Lmg/dLmmol/Lmg/dLmmol/Lmg/dLmmol/L
Control N92.705.1501.51.525.041.3914.04.0
Control P255.2414.1801.41.4279.6515.5363.03.1
Sample 112.650.7032.32.211.870.6603.63.6
Sample 230.631.7021.81.816.640.9243.43.4
Sample 3103.565.7531.71.7174.809.7113.23.2
Sample 4259.2914.4051.61.6324.7018.0393.33.3
Sample 5571.8731.7701.61.6537.4729.8594.34.3

{15}------------------------------------------------

1.3 Interferences

The Interferences were determined according to the CLSI guideline EP07-A2.

The calculated bias was within the 10% acceptance citteria on serum and urine samples for up to the following concentrations.

SerumUrine
Hemoglobin501 mg/dL (290 µmol/L)Hemoglobin501 mg/dL (290 µmol/L)
Triglycerides569 mg/dL (6.50 mmol/L)Total Bilirubin27.20 mg/dL (465 µmol/L)
Total Bilirubin35.53 mg/dL (607 µmol/L)Ascorbic Acid5.98 mg/dL (340 µmol/L)
Direct Bilirubin21.09 mg/dL (360 µmol/L)Modification of the pH of the samplesNo impact on glucose assay measurement
Acetylsalicylic Acid65.16 mg/dL (3.62 mmol/L)Specific gravity1.000 - 1.030
Ascorbic Acid5.98 mg/dL (340 µmol/L)
Ibuprofen50.1 mg/dL (2.43 mmol/L)
Acetaminophen20 mg/dL (1324 µmol/L)

1.4 Matrix comparison on predicate device

This protocol was according to the CLSI guideline EP09-A3. 56 plasma samples were evaluated on Yumizen C1200 using Yumizen C1200 Glucose HK reagent and with the Pentra C400 Clinical Chemistry Analyzer with ABX PENTRA Glucose HK reagent (Predicate device). The equation for the regression line using Passing Bablok was obtained.

Passing BablokNInterceptSlopeCorrelation - r
Glucose in plasma (mg/dL)56-0.0700.9910.999
Glucose in plasma (mmol/L)560.0130.9880.999

{16}------------------------------------------------

1.5 Method comparison with a predicate device

Yumizen C1200 was compared with Pentra C400 Clinical Chemistry Analyzer. 140 (in urine) patient samples were correlated with a commercial reagent taken as reference according to the CLSI quideline EP09-A3. The equation for the regression line using Passing Bablok was obtained.

Passing BablokNInterceptSlopeCorrelation - r
Glucose in serum (mg/dL)141-2.4531.0050.996
Glucose in urine (mg/dL)1000.9751.0070.999

1.6 Reagent Stability

1.6.1 Closed stability

The closed stability was determined according to the CLSI guideline EP25-A. Stable up to the expiry date on the label if stored at 2 - 8 degree C.

1.6.2 Open stability

Stable up to the expiry date on the label if stored at the refrigerated compartment is stable for 6 weeks.

1.7 Reference Range

The Reference Range was determined according to the CLSI guideline EP28-A3c.

50 normal samples (25 women + 25 men) in serum were assayed with the method in evaluation. Each sample was assayed in duplicates. The mean of the duplicate results for each subject was compared against reference ranges cited in literature. The verification studied support in the following reference ranges which were established through literature.

Serum: 70 to 115 mg/dL (3.89 to 6.39 mmol/L)

Urine: < 15 mg/dL (< 0.84 mmol/L)

Reference

Thomas L. Clinical Laboratory Diagnostics. 1st ed. Frankfurt: TH-Books Verlagsgesellschaft: 1998. p. 131-7, 1368.

{17}------------------------------------------------

2 ISE module on Yumizen C1200

2.1 Measuring Range

The measuring range of the Sodium, Potassium and Chloride Electrodes on Yumizen C1200 was validated by the linearity studies.

ParametersMeasuring range
NA in serum (mmol/L)100 to 200
K in serum (mmol/L)1.5 to 10
CL in serum (mmol/L)52 to 200
NA in urine (mmol/L)10 to 400
K in urine (mmol/L)3 to 300
CL in urine (mmol/L)15 to 400

{18}------------------------------------------------

2.2 Accuracy and Precision

2.2.1 Repeatability (within-run precision)

The Repeatability was determined by analyzing 2 level controls, 3 samples of low, middle and high level concentration and the spiked samples were tested 20 times by 3 reagent Lots.

Control in serumControl NControl PControl N2Control P2
Lot1Lot2Lot3Lot1Lot2Lot3Lot1Lot2Lot3Lot1Lot2Lot3
NAMean value (mmol/L)133.80134.68134.55134.55135.37135.84122.13123.10123.25154.39155.18154.66
CV (%)0.20.20.30.20.20.30.20.20.20.20.20.2
KMean value (mmol/L)4.2964.3364.3336.6436.6896.7003.8253.8653.8666.5366.5676.553
CV (%)0.10.20.30.20.30.30.30.30.40.20.30.3
CLMean value (mmol/L)114.04115.47114.24112.75114.50113.4591.2191.7091.64112.01112.17111.72
CV (%)0.20.30.40.30.20.30.20.30.30.30.30.3
Patient sample in serumLow sampleMiddle sampleHigh sampleSpiked sample
Lot1Lot2Lot3Lot1Lot2Lot3Lot1Lot2Lot3Lot1Lot2Lot3
NAMean value (mmol/L)125.85126.81126.74138.64139.61139.18148.72149.51148.84165.20165.83164.90
CV (%)0.20.30.20.20.30.30.30.30.30.20.30.3
KMean value (mmol/L)2.7732.7792.7844.0414.0674.0655.8475.8965.8996.6886.7426.733
CV (%)0.20.30.30.30.30.20.20.20.30.30.30.3
CLMean value (mmol/L)90.6190.8990.46101.81102.29101.71112.79113.09112.46130.18130.35129.65
CV (%)0.20.30.20.20.40.30.40.40.30.30.30.3

{19}------------------------------------------------

Control in urineControl NControl P
Lot1Lot2Lot3Lot1Lot2Lot3
NAMean value(mmol/L)85.7586.6886.65178.97179.04178.09
CV (%)0.30.30.30.30.30.4
KMean value(mmol/L)31.10231.16031.10483.43783.45383.115
CV (%)0.50.50.40.40.40.4
CLMean value(mmol/L)62.0462.4866.09163.60162.67171.54
CV (%)0.40.30.40.20.20.3
Patient sample
Low sampleMiddle sampleHigh sampleSpiked sample
in urineLot1Lot2Lot3Lot1Lot2Lot3Lot1Lot2Lot3Lot1Lot2Lot3
Mean value(mmol/L)27.6928.4428.36103.54104.24103.53188.09187.72186.54336.99334.35331.45
NACV (%)1.00.81.20.20.20.30.20.30.30.30.30.3
KMean value(mmol/L)29.37329.30929.17668.76568.92268.365118.831118.845118.050220.783221.125218.953
CV (%)0.50.30.30.40.30.50.40.30.40.50.50.5
CLMean value(mmol/L)24.1325.7930.78102.07104.55106.76199.91199.77190.33322.88318.66292.00

2.2.2 Reproducibility (total precision)

The Reproducibility was determined according to the CLSI guideline EP05-A3.

2 level controls, 3 samples of low, middle and high concentrations and the spiked samples were tested in duplicate for 20 days (2 series per day).

{20}------------------------------------------------

NA in serumK in serumCL in serum
serumMean value(mmol/L)CV (%)Mean value(mmol/L)CV (%)Mean value(mmol/L)CV (%)
Control N134.090.44.2960.5113.490.4
Control P134.220.36.5990.6111.340.4
Control N2122.410.43.8140.790.770.4
Control P2154.020.36.5140.5111.510.4
Low sample126.470.32.8001.690.250.3
Middle sample139.090.34.0580.6101.590.3
High sample148.820.35.8610.6112.360.3
Spiked sample164.530.46.7060.8129.460.4
NA in urineK in urineCL in urine
urineMean value(mmol/L)CV (%)Mean value(mmol/L)CV (%)Mean value(mmol/L)CV (%)
Control N85.310.731.1310.760.401.8
Control P179.900.484.2690.6162.601.2
Low sample27.931.529.3881.624.293.1
Middle sample103.290.568.6380.7100.790.7
High sample187.180.4117.7061.1199.130.4
Spiked sample335.210.7222.1081.3323.050.6

{21}------------------------------------------------

2.3 Interferences

This protocol was determined according to the CLSI EP07-A2.

The calculated bias was within the 10% acceptance criteria on Sodium, Potassium and urine samples for up to the following concentrations.

NA, K, CL in serumNA, K, CL in urine
Hemoglobin0.5 g/dLHemoglobin1.25 g/dL
Triglycerides52.1 mmol/LTotal Bilirubin256 µmol/L
Total Bilirubin396 µmol/LProtein3.31 g/L
Total protein121.13 g/LUrea988.3 mmol/L
Urea71.9 mmol/LAscorbic acid3.40 mmol/L
Salicylic acid0.53 mmol/L
Imipramine2.50 µmol/L
Procainamide102 µmol/L
Chlorpromazine6.30 µmol/L
Erythromycin81.6 µmol/L
Ampicillin150 µmol/L
  • 2.4 Matrix comparison on predicate device
    This protocol was according to the CLSI guideline EP09-A3.

Plasma samples were evaluated on Yumizen C1200 and Olympus AU400 Clinical Chemistry Analyzer. The equation for the regression line using Passing Bablok was obtained.

Passing BablokNMin (mmol/L)Max (mmol/L)InterceptSlopeCorrelation - r
NA171102.9188.90.2001.0000.997
K1731.769.530.0101.0000.999
CL17353.1178.4-1.5681.0190.998

{22}------------------------------------------------

2.5 Method comparison with a reference device

Yumizen C1200 was compared with Olympus AU400 Clinical Chemistry Analyzer.

The patient samples were correlated with a commercial ISE module taken as reference according to the CLSI guideline EP09-A3.

The equation for the regression line using Passing Bablok was obtained.

Passing BablokNInterceptSlopeCorrelation - r
NA in serum1651.3710.9870.995
K in serum1700.0230.9940.999
CL in serum1720.8180.9920.998
NA in urine194-1.2561.0011.000
K in urine198-0.1280.9830.999
CL in urine1940.5250.9911.000

2.6 Electrodes Stability

The protocol of this study was followed by JEOL internal protocol.

The evaluation was performed using the same parameter checked in manufacturing of the electrodes.

2.6.1 Closed stability

Stable up to the expiry date on the label if stored at 0-40 degree C.

  • 2.6.2 Open stability
    Stable for 3 months from opened.

{23}------------------------------------------------

2.7 Reference Range

The Reference Range was determined according to the CLSI EP28-A3c.

80 normal samples (23 women + 57 men) in serum and plasma were assayed with the method in evaluation. Each sample was assayed in duplicates. The mean of the duplicate results for each subject was compared against reference ranges cited from literature. The verification studies supported in the following reference ranges which were established through literature. NA in serum and plasma: 136 to 145 mmol/L1

K in serum and plasma: 3.5 to 5.1 mmol/L1

(The reference range of K in plasma is about 0.5 mmol/L lower than in serum)2

CL in serum and plasma: 98 to 107 mmol/L1

The expected values based on 24 hours urine out-put were cited from the literature.

NA in urine: 40 to 220 mmol/day1

K in urine: 25 to 125 mmol/day1

CL in urine: 110 to 250 mmol/day1

Reference:

  1. Tietz NW, Clinical Guide for Laboratory Tests 3rd edition. WB Saunders Company, Philadelphia, PA, pp. 610-611 (1995)

  2. Walker HK,et al. Clinical Methods: The History, Physical, and Laboratory Examinations, 3rd edition, Boston: Butterworths, pp 884-887(1990)

{24}------------------------------------------------

Non clinical tests

The non clinical studies tests conclude that the safety and effectiveness of the devices are not compromised. Yumizen C1200 meets the following.

ISO 14971: 2007Medical devices - Application of risk management to medical devices
IEC 61326-2-6:2012Electrical equipment for measurement, control and laboratory use — EMC requirements —Part 2-6: Particular requirements — In vitro diagnostic (IVD) medical equipment
IEC 61326-1: 2012Electrical equipment for measurement, control and laboratory use - EMC requirements - Part1: General requirements
IEC 61010-1: 2010Safety requirements for electrical equipment for measurement, control, and laboratory use —Part 1 General requirements
IEC 61010-2-010: 2014Safety requirements for electrical equipment for measurement, control and laboratory use -Part 2-010: Particular requirements for laboratory equipment for the heating of materials
IEC 61010-2-101: 2015Safety requirements for electrical equipment for measurement, control, and laboratory use —Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
IEC 62304: 2006Medical device software — software life-cycle processes
IEC 60825-1 2014Safety of laser products - Part 1: Equipment classification and requirements

Conclusion

The data presented demonstrate that the Yumizen C1200 glucose, sodium, and chloride assays are substantially equivalent to the predicate devices.

§ 862.2160 Discrete photometric chemistry analyzer for clinical use.

(a)
Identification. A discrete photometric chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. Different models of the device incorporate various instrumentation such as micro analysis apparatus, double beam, single, or dual channel photometers, and bichromatic 2-wavelength photometers. Some models of the device may include reagent-containing components that may also serve as reaction units.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.