(245 days)
A Patient examination glove is a disposable device intended for medical purposes that worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs as listed below.
PowderFree White Blue Sterilized Nitrile Copolymer Examination Gloves Tested for Use with Chemotherapy Drugs.
This document is an FDA 510(k) clearance letter for "PowderFree White Blue Sterilized Nitrile Copolymer Examination Gloves Tested for use with chemotherapy drugs." It outlines the device name, indications for use, and a table detailing the breakthrough detection times for various chemotherapy drugs when used with white and blue versions of these gloves.
The primary "acceptance criterion" for these gloves, beyond their general use as patient examination gloves, relates to their performance against chemotherapy drugs. The study aims to demonstrate that the gloves provide a sufficient barrier to these hazardous substances.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state pre-defined acceptance criteria for "Breakthrough Detection Time" in terms of what would be considered "acceptable." Instead, it presents the results of testing against these drugs. The implicit acceptance is that the gloves demonstrably offer a barrier, especially for drugs with breakthrough times greater than 240 minutes, which is often a standard benchmark for extended protection in such contexts. The warning about Carmustine and Thio-Tepa highlights drugs that do not meet a high standard of protection (implicitly, a breakthrough time significantly less than 240 minutes).
However, based on the provided data, we can infer a performance comparison:
| Chemotherapy Drugs | Implicit Acceptance Criteria (Inferred from common safety standards/other drugs' performance) | White Glove Performance (Breakthrough Detection Time in Minutes) | Blue Glove Performance (Breakthrough Detection Time in Minutes) |
|---|---|---|---|
| Carmustine (BCNU) 3.3 mg/ml | Ideally >240 min (or a clearly stated acceptable minimum for safe handling) | 7.1 min | 23.4 min |
| Cisplatin 1.0 mg/ml | Ideally >240 min | >240 min | >240 min |
| Cyclophosphamide 20mg/ml | Ideally >240 min | >240 min | >240 min |
| Dacarbazine (DTIC) 10.0 mg/ml | Ideally >240 min | >240 min | >240 min |
| Doxorubicin HCl 2.0 mg/ml | Ideally >240 min | >240 min | >240 min |
| Etoposide 20.0 mg/ml | Ideally >240 min | >240 min | >240 min |
| Fluorouracil 50.0 mg/ml | Ideally >240 min | >240 min | >240 min |
| Ifosfamide 50.0 mg/ml | Ideally >240 min | >240 min | >240 min |
| Methotrexate 25 mg/ml | Ideally >240 min | >240 min | >240 min |
| Mitomycin C 0.5 mg/ml | Ideally >240 min | >240 min | >240 min |
| Mitoxantrone 2.0 mg/ml | Ideally >240 min | >240 min | >240 min |
| Paclitaxel (Taxol) 6.0 mg/ml | Ideally >240 min | >240 min | >240 min |
| Thio-Tepa 10.0 mg/ml | Ideally >240 min (or a clearly stated acceptable minimum for safe handling) | 78.6 min | 39.3 min |
| Vincristine Sulfate 1.0 mg/ml | Ideally >240 min | >240 min | >240 min |
Note: The "Implicit Acceptance Criteria" column is an inference based on the reported ">240 min" for most drugs, and the explicit warning for those below this threshold. A formal acceptance criterion would typically be a pre-defined quantifiable value.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: The document does not specify the sample size (number of gloves or repetitions per drug) used for the permeation testing.
- Data Provenance: The document does not explicitly state the country of origin of the data or whether the study was retrospective or prospective. It is a submission by "Smart Glove Corporation Sdn. Bhd." from Malaysia, suggesting the testing likely originates from or was commissioned by this Malaysian company.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This point is not applicable to the document. The "ground truth" here is the chemical permeation performance measured through standardized laboratory testing, not medical expert opinion or consensus.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This point is not applicable. Adjudication methods like 2+1 or 3+1 typically refer to the process of resolving discrepancies in expert interpretations of data (e.g., medical images). Here, the testing involves objective chemical analysis, not subjective human interpretation requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This point is not applicable. An MRMC study is relevant for evaluating the impact of AI on human performance in diagnostic tasks. This document concerns the physical barrier properties of gloves against chemical permeation, which does not involve "human readers" or "AI assistance."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This point is not applicable. "Standalone performance" refers to the performance of an algorithm without human intervention, usually in the context of AI/ML devices. This document refers to the physical testing of gloves.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth used is objective laboratory measurement of chemical permeation/breakthrough time as determined by a standardized testing method (likely ASTM F739 or similar, although not explicitly stated). This is a direct measurement of the glove's barrier integrity against specific chemotherapy drugs.
8. The sample size for the training set
This point is not applicable. There is no "training set" in the context of physical product testing like glove permeation. Training sets are relevant for machine learning models.
9. How the ground truth for the training set was established
This point is not applicable as there is no training set involved in this type of device submission.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services-USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
July 26, 2019
Smart Glove Corporation Sdn. Bhd. Khon Foo Managing Director Lot 6487, Batu 5 3/4, Sementa, Jalan Kapar Klang, 42100 My
Re: K183280
Trade/Device Name: PowderFree White Blue Sterilized Nitrile Copolymer Examination Gloves Tested for use with chemotherapy drugs
Regulation Number: 21 CFR 880.6250 Regulation Name: Patient examination glove Regulatory Class: Class I Product Code: LZA Dated: June 18, 2019 Received: June 24, 2019
Dear Khon Foo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
PowderFree White Blue Sterilized Nitrile Copolymer Examination Gloves Tested for Use with Chemotherapy Drugs.
Indications for Use (Describe)
A Patient examination glove is a disposable device intended for medical purposes that worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs as listed below.
Chemotherapy Drug permeation and concentration ( Breakthrough Detection Time) in Minutes. The following chemicals have been tested with these gloves.
| Breakthrough Detection Time in Minutes | ||
|---|---|---|
| Chemotherapy Drugs: | White Glove | Blue Glove |
| 1. Carmustine ( BCNU)3.3 mg/ml(3,300 ppm) | 7.1 min | 23.4 min |
| 2. Cisplatin 1.0 mg/ml (1,000ppm) | >240 min | >240 min |
| 3. Cyclophosphamide (Cytoxan) 20mg/ml(20,000ppm) | >240 min | >240 min |
| 4. Dacarbazine (DTIC)10.0 mg/ml(10,000ppm) | >240 min | >240 min |
| 5. Doxorubicin Hydrochloride 2.0 mg/ml (2,000ppm) | >240 min | >240 min |
| 6. Etoposide (Toposar) 20.0 mg/ml (20,000ppm) | >240 min | >240 min |
| 7. Fluorouracil 50.0 mg/ml (50,000ppm) | >240 min | >240 min |
| 8. Ifosfamide 50.0 mg/ml (50,000ppm) | >240 min | >240 min |
| 9. Methotrexate 25 mg/ml (25,000ppm) | >240 min | >240 min |
| 10.Mitomycin C 0.5 mg/ml (500ppm) | >240 min | >240 min |
| 11. Mitoxantrone 2.0 mg/ml (2,000ppm) | >240 min | >240 min |
| 12. Paclitaxel (Taxol) 6.0 mg/ml (6,000ppm) | >240 min | >240 min |
| 13. Thio-Tepa 10.0 mg/ml (10,000ppm) | 78.6 min | 39.3 min |
| 14.Vincristine Sulfate 1.0 mg/ml (1,000ppm) | >240 min | >240 min |
WARNING: Do not use with Carmustine or Thio-Tepa.
Please note that Carmustine and Thio-Tepa have low permeation time of less than 240 minutes.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.