K Number
K183230
Date Cleared
2019-08-15

(268 days)

Product Code
Regulation Number
864.7470
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OneDraw™ A1C Test System, which consists of the OneDraw Blood Collection Device and the OneDraw A1C Test, is intended to collect capillary blood from the upper arm of individuals 18 years of age or older onto filter (matrix) paper within the collection device by a healthcare professional. Samples are delivered to the laboratory for the quantitative measurement of HbA1c for monitoring the long-term control of blood sugar (glucose) in people with diabetes. Testing performed on samples collected with this device should not be used to diagnose or screen for diabetes. The OneDraw A1C Test System should not be used with neonates.

Device Description

The OneDraw™ A1C Test System includes the OneDraw Blood Collection Device and the OneDraw A 1C Test. The OneDraw Blood Collection Device is a single-use, sterile, capillary blood specimen collection device. The OneDraw Blood Collection Device includes a transport sleeve, accessories, and instructions (OneDraw Blood Collection Device Instructions for Use (IFU)) which are needed to collect, package, and mail the sample to the designated certified clinical laboratory for HbA1c testing, using the OneDraw A1C Test.

The OneDraw Blood Collection Device incorporates lancets to make incisions in the skin and a vacuum to draw blood at the surface of the skin through channels to deposit the blood onto collection and stabilization matrices (matrix strips). The matrix strips are contained within a cartridge which is removed from the device after the draw is complete. The cartridge is then inserted into the transport sleeve which encloses and protects the sample during shipping to the clinical laboratory.

Once the transport sleeve containing the sample is received by the clinical laboratory, one of the dry blood sample matrices is removed. The matrix is then eluted in Beckman Hemolyzing Reagent (BHR) in 2 mL tubes or 2.2 mL deepwell plates using an orbital shaker. Next, the sample is diluted in BHR to its final concentration and tested using FDA-cleared Beckman Coulter AU480 Chemistry Analyzer and A 1 c reagents, including calibrators, (K 120199) per the OneDraw A 1 C Test IFU.

AI/ML Overview

The document describes the OneDraw™ A1C Test System, which includes the OneDraw Blood Collection Device and the OneDraw A1C Test. The device is intended to collect capillary blood from the upper arm for quantitative measurement of HbA1c in people with diabetes. The testing is not for diagnosis or screening of diabetes and should not be used with neonates. The following information outlines the acceptance criteria and the studies performed to demonstrate the device meets these criteria.

1. Table of Acceptance Criteria & Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implicit from Study Outcomes)Reported Device Performance
Precision (Assay)Repeatability (within-run and within-day) %CV expected to be low for HbA1c measurements across relevant ranges.Repeatability (within-run and within-day):
Sample 1 (5.10% HbA1c)-Within-run SD: 0.073, %CV: 1.44%; Within-day SD: 0.030, %CV: 0.58%; Total SD: 0.119, %CV: 2.33%
Sample 2 (6.46% HbA1c)-Within-run SD: 0.091, %CV: 1.40%; Within-day SD: 0.033, %CV: 0.51%; Total SD: 0.135, %CV: 2.09%
Sample 3 (7.87% HbA1c)-Within-run SD: 0.082, %CV: 1.05%; Within-day SD: 0.025, %CV: 0.32%; Total SD: 0.114, %CV: 1.45%
Sample 4 (11.44% HbA1c)-Within-run SD: 0.119, %CV: 1.04%; Within-day SD: 0.071, %CV: 0.62%; Total SD: 0.161, %CV: 1.40%
Precision (Device)Low %CV for lot-to-lot and operator-to-operator variability to ensure reproducibility.Lot-to-Lot Analysis: Average CV: 1.6% (range: 0.0% - 3.6%). • < 6% HbA1c: Total %CV 1.99% (95% CI: 1.57, 2.42) • ≥ 6% HbA1c: Total %CV 1.10% (95% CI: 0.74, 1.45) Operator-to-Operator: Average CV: 1.5% (range: 0.1% - 4.8%). • < 6% HbA1c: Total %CV 2.24% (95% CI: 1.76, 2.72) • ≥ 6% HbA1c: Total %CV 1.24% (95% CI: 0.84, 1.64)
Method ComparisonStrong correlation and agreement with standard venipuncture methods.Passing-Bablok Regression: • Slope: 1.00 (95% lower bound: 0.97, 95% upper bound: 1.03) • Intercept: -0.11 • Pearson correlation coefficient (R): 0.9907 (95% lower bound: 0.9864, 95% upper bound: 0.9937)
LinearityEstablished measuring range with a strong linear fit.Measuring range: 4.70-14.3 % HbA1c (DCCT/NGSP). Linear fit: Y = 1.01x + 0.06; R=0.99.
InterferenceNo significant interference (±10% bias) from common exogenous and endogenous substances, including specific Hb variants.No Interference Observed (< ±10% bias) from: Acetaminophen (20 mg/dL), Acetylsalicylic acid (65 mg/dL), Glyburide (0.2 mg/dL), Ibuprofen (50 mg/dL), Metformin (4.0 mg/dL), L-ascorbic acid (3.0 mg/dL), Triglycerides (3400 mg/dL), Bilirubin (conjugated 33.2mg/dL, unconjugated 30mg/dL), Rheumatoid factor (600 IU/mL). Hemoglobin Variants: No significant interference for HbA2 (≤5.8%), HbC (≤40.1%), HbD (≤41.2%), HbE (≤22%), HbS (≤34.8%). Note: HbF levels > 8.5% show significant negative bias and should not be used.
Limits of DetectionMatch the established linearity range.Claimed measuring range: 4.70%-14.3% HbA1c.
Product StabilityOneDraw Blood Collection Device stable for 12 months. Samples stable for 21 days at room temperature, robust to temperature excursions.Device: 12-month expiry date confirmed. Sample storage and shipping: Stable for up to 21 days at room temperature. Extreme temperature excursions do not cause significant difference in HbA1c measurement. Matrices and transport sleeves withstand stressed shipping/storage conditions.
Flex StudiesAcceptable performance across variations in blood volume, hematocrit, and comparability of matrix strips and collection methods.Blood Volume: 52.5 µL and 90 µL volumes were within ±10% relative bias compared to 75 µL samples. Hematocrit: Varying hematocrit levels do not interfere. Matrix Strip Comparability: No significant difference in %HbA1c results between the two matrix strips in the same cartridge. Collection Method Comparability: HbA1c results from dried whole blood spotted on matrix strips and capillary blood collected with OneDraw device are comparable to standard venipuncture (venous whole blood).

2. Sample Size for Test Set and Data Provenance

  • Precision (Assay - Repeatability): 80 data points were collected. The source of the blood samples is not explicitly mentioned as a specific country of origin, but the testing was performed in the context of an FDA submission, implying a US-based or international study adhering to US regulatory standards. It is a prospective study as samples were specifically analyzed for the purpose of the study.
  • Precision (Device - Lot-to-Lot Analysis): 23 participants. The study involved multiple collection sites, suggesting prospective data collection.
  • Precision (Device - Operator-to-Operator): 25 participants. The study involved multiple collection sites, suggesting prospective data collection.
  • Method Comparison: 107 participants. Blood collections were conducted at two different clinical sites. This implies prospective collection for the study.
  • Linearity, Interference, Limits of Detection, Product Stability, Flex Studies: The sample sizes (e.g., number of replicates, levels tested, specific blood specimens) were described within each section (e.g., "at least 20 days, two runs per day" for assay precision, "two (2) HbA1c levels... six (6) blood volumes per level... 6 matrix strips per volume" for blood volume flex study). Data provenance is prospective testing conducted for the device's validation.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

  • The document does not mention the use of "experts" in the traditional sense (e.g., radiologists, pathologists) to establish ground truth for this in vitro diagnostic device.
  • Instead, the ground truth for HbA1c measurements is established by comparison to a "standard venipuncture (tested using Beckman's NGSP-certified method on the Beckman Coulter AU480 Analyzer)" (for method comparison) and other established laboratory methods and guidelines (e.g., CLSI guidelines). The accuracy of these reference methods is implicitly accepted as the ground truth.
  • The qualifications of the personnel operating the reference methods are not explicitly stated but are assumed to be trained laboratory professionals.

4. Adjudication Method for the Test Set

  • Adjudication methods (like 2+1, 3+1) are typically used in image-based diagnostic studies where human interpretation of medical images can vary.
  • For this in vitro diagnostic device, which provides quantitative measurements, there is no "adjudication" in the sense of reconciling differing expert opinions. The performance is assessed by comparing the device's quantitative results against established reference methods or accepted criteria through statistical analysis.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted.
  • MRMC studies are relevant for evaluating the impact of an AI system on human reader performance, typically in diagnostic imaging. This device is a blood collection and testing system for HbA1c, not an AI diagnostic imaging tool or a system designed to assist human readers in interpretation.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Yes, the performance studies described are essentially standalone performance evaluations of the OneDraw™ A1C Test System.
  • The device functions as a blood collection and testing system, and its output (HbA1c levels) is directly measured and compared to reference methods. There is no "human-in-the-loop" interpretative step by a medical professional whose diagnostic accuracy is being augmented or tested. The healthcare professional collects the sample, but the analysis is done by the device system and associated laboratory methods.

7. Type of Ground Truth Used

  • The ground truth used is primarily based on reference laboratory results obtained from "standard venipuncture (tested using Beckman's NGSP-certified method on the Beckman Coulter AU480 Analyzer)" for HbA1c measurement.
  • Other ground truth validations include established ranges for linearity, known concentrations for interference testing, and performance against recognized CLSI guidelines for precision, stability, and limits. This is essentially measurement against accepted reference methods and established statistical criteria.

8. Sample Size for the Training Set

  • The document does not describe explicit "training sets" in the context of machine learning or AI models.
  • This device is an in vitro diagnostic system for quantitative measurement, not a machine learning algorithm that requires a training set in the typical sense.
  • The "development" or "internal validation" data—distinct from the performance data presented for regulatory submission—is not detailed in this summary.

9. How the Ground Truth for the Training Set Was Established

  • As concluded in point 8, the document does not describe a "training set" for a machine learning model for which ground truth would need to be established. Therefore, this question is not applicable based on the provided information.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two distinct elements: the Department of Health and Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health and Human Services logo features a stylized depiction of a human figure, while the FDA portion includes the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 15, 2019

Drawbridge Health, Inc. Annie Wright Associate Director, Regulatory Affairs & Quality Assurance 11535 Sorrento Valley Road, Suite 407 San Diego, CA 92121

Re: K183230

Trade/Device Name: OneDraw™ A1C Test System Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated hemoglobin assay Regulatory Class: Class II Product Code: LCP, PRJ Dated: July 15, 2019 Received: July 15, 2019

Dear Annie Wright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183230

Device Name OneDraw™ A1C Test System

Indications for Use (Describe)

The OneDraw™ A1C Test System, which consists of the OneDraw Blood Collection Device and the OneDraw A1C Test, is intended to collect capillary blood from the upper arm of individuals 18 years of age or older onto filter (matrix) paper within the collection device by a healthcare professional. Samples are delivered to the laboratory for the quantitative measurement of HbA1c for monitoring the long-term control of blood sugar (glucose) in people with diabetes. Testing performed on samples collected with this device should not be used to diagnose or screen for diabetes. The OneDraw A1C Test System should not be used with neonates.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Drawbridge Health. The logo consists of a stylized, abstract shape in shades of blue and red on the left, followed by the company name "Drawbridge" in a dark gray sans-serif font. Below the company name is the word "HEALTH" in a smaller, also dark gray, sans-serif font.

510(k) SUMMARY

OneDraw™ A1C Test System

Applicant Name:Drawbridge Health, Inc.11535 Sorrento Valley Road, Suite 407San Diego, CA 92121 USA
Company Contact:Annie WrightAssociate Director, Regulatory Affairs & Quality AssurancePhone: (408) 421-1510Email: awright@dbhealth.com
Date Prepared:July 25, 2019

DEVICE IDENTIFICATION

Device Name: OneDraw™ A1C Test System, which consists of the following:
Trade or Proprietary Names:OneDraw™ A1C Test
Device Classification Name:Glycosylated hemoglobin assay
Product Codes:LCP
Regulatory Class:Class II
Classification Regulation:21 CFR 864.7470
Panel:Hematology (81)
Trade or Proprietary Names:OneDraw™ Blood Collection Device
Device Classification Name:Blood specimen collection device
Product Codes:PRJ
Regulatory Class:Class II
Classification Regulation:21 CFR 862.1675
Panel:Chemistry (75)
Predicate Device:Home Access Health Corp., Home Access® A1C Test andHome Access® Collection Cassette (K141944)

INTENDED USE

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Drawbridge Health. The logo consists of a blue and red abstract shape on the left, and the words "Drawbridge" and "Health" on the right. The word "Drawbridge" is in a larger, darker font than the word "Health", which is located underneath the word "Drawbridge".

The OneDraw™ A1C Test System, which consists of the OneDraw Blood Collection Device and the OneDraw A1C Test, is intended to collect capillary blood from the upper arm of individuals 18 years of age or older onto filter (matrix) paper within the collection device by a healthcare professional. Samples are delivered to the laboratory for the quantitative measurement of HbA1c for monitoring the long-term control of blood sugar (glucose) in people with diabetes. Testing performed on samples collected with this device should not be used to diagnose or screen for diabetes. The OneDraw A1C Test System should not be used with neonates.

DEVICE DESCRIPTION

The OneDraw™ A1C Test System includes the OneDraw Blood Collection Device and the OneDraw A 1C Test. The OneDraw Blood Collection Device is a single-use, sterile, capillary blood specimen collection device. The OneDraw Blood Collection Device includes a transport sleeve, accessories, and instructions (OneDraw Blood Collection Device Instructions for Use (IFU)) which are needed to collect, package, and mail the sample to the designated certified clinical laboratory for HbA1c testing, using the OneDraw A1C Test.

The OneDraw Blood Collection Device incorporates lancets to make incisions in the skin and a vacuum to draw blood at the surface of the skin through channels to deposit the blood onto collection and stabilization matrices (matrix strips). The matrix strips are contained within a cartridge which is removed from the device after the draw is complete. The cartridge is then inserted into the transport sleeve which encloses and protects the sample during shipping to the clinical laboratory.

Once the transport sleeve containing the sample is received by the clinical laboratory, one of the dry blood sample matrices is removed. The matrix is then eluted in Beckman Hemolyzing Reagent (BHR) in 2 mL tubes or 2.2 mL deepwell plates using an orbital shaker. Next, the sample is diluted in BHR to its final concentration and tested using FDA-cleared Beckman Coulter AU480 Chemistry Analyzer and A 1 c reagents, including calibrators, (K 120199) per the OneDraw A 1 C Test IFU.

TECHNOLOGICAL CHARACTERISTICS

The OneDraw A1C Test System has technological characteristics that are substantially equivalent to the predicate device as identified in the table below. Both the subject device and the predicate device provide a method to collect a capillary blood sample to be analyzed in a clinical laboratory for determination of Hemoglobin A1c (HbA1c) using a previously cleared reagents and analysis system.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for Drawbridge Health. The logo consists of a stylized, colorful icon resembling a curved shape with gradients of blue, red, and orange. To the right of the icon is the text "Drawbridge" in a gray, sans-serif font, with the word "HEALTH" in smaller letters underneath.

AttributePredicate DeviceHome Access® A1C Test (K141944)Candidate DeviceOneDraw™ A1C Test System
Similarities and Differences
Intended UseThe Home Access® A1C Test is an in vitro test methodfor quantitative measurement of Hemoglobin A1c usingcapillary blood collected from the fingertip, collectedonto filter paper via the Home Access collectioncassette. The Home Access A1C Test is formeasurement of HbA1c on blood specimens which canbe collected at the patient' home or in a healthcareprofessional setting and delivered to the laboratory bymail. Measurements obtained through this method canbe used for long-term control of blood sugar (glucose) inpeople with diabetes.This test is not to be used to diagnose or screen for diabeteNot for use with neonates.The OneDraw™ A1C Test System, which consistsof the OneDraw Blood Collection Device and theOneDraw A1C Test, is intended to collect capillaryblood from the upper arm of individuals 18 years ofage or older onto filter (matrix) paper within thecollection device by a healthcare professional.Samples are delivered to the laboratory for thequantitative measurement of HbA1c for monitoringthe long-term control of blood sugar (glucose) inpeople with diabetes. Testing performed on samplescollected with this device should not be used todiagnose or screen for diabetes. The OneDraw A1CTest System should not be used with neonates.
Sample typesDry capillary bloodSame
StandardizationTraceable to the International Federation of ClinicalChemistry and Laboratory Medicine (IFCC) andDiabetes Control and Complications Trial (DCTT)reference method. Certified via the NationalGlycohemoglobin Standardization Program (NGSP).Same
Kit ComponentsBlood sample collection cassette containing filterpaper for specimen collection Sample pouch with desiccant for specimenpackaging 2 sterile safety lancets Gauze Pad 2 Bandages Instructions for Use/Things You Should KnowAbout A1C Prepaid Return Mailer for specimen mailing Patient Info Card for specimen labeling and consent Outer PackagingBlood sample collection cartridge containingfilter paper for specimen collection Sample pouch containing desiccant for specimenpackaging 2 sterile lancets within the assembly Alcohol prep pad Gauze pad Bandage Instructions for Use Return Mailer (Chipboard) for specimen mailing Outer Carton
Sample PreparationFinger stick blood collected on a filter paper within thecassetteBlood drawn from upper arm and is collected on afilter paper within the cartridge
Location of collectionSample collected at home or HCP settingHealthcare Professional (HCP) setting
Location of analysisLaboratorySame
DistributionPrescription and Over the CounterPrescription
AnalysisMail to laboratorySame
ReportMailed to userMailed to healthcare professional
Measuring Range4.5-14.5% HbA1c4.7-14.3% HbA1c
Puncture siteFingertipUpper arm
Mechanism of blood drawFingerstickCollection device (lancet-stick and vacuum)
Collection methodBlood is collected using the fingerstick methodThe blood is drawn into the collection device bypuncture of capillaries located near the surface of theskin using lancets or similar small sharp objects

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Drawbridge Health. The logo consists of a stylized, abstract shape on the left, with the company name "Drawbridge" and the word "HEALTH" stacked below it on the right. The abstract shape is a curved, teardrop-like form with a gradient of colors, transitioning from blue at the top to red at the bottom. The text is in a dark gray color.

PERFORMANCE TESTING

The following standards/guidance documents were used in the design and testing of OneDraw™ A1C Test System:

Designation/RevisionTitle
CLSI C44-A, 2002Harmonization of Glycohemoglobin Measurements; Approved Guideline
CLSI EP05-A3, 2014Evaluation of Precision Performance of Quantitative Measurement Methods; ApprovedGuideline - Third Edition
CLSI-EP06-A, 2003Evaluation of the Linearity of Quantitative Measurement Procedures: A StatisticalApproach; Approved Guideline
CLSI-EP07-A2, 2005Interference Testing in Clinical Chemistry; Approved Guideline – Second Edition
CLSI-EP09-A3, 2013Method Comparison and Bias Estimation using Patient Samples; Approved Guideline –Third Edition
CLSI-EP17-A2, 2012Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures;Approved Guideline-Second Edition
CLSI- EP25-A, 2009Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
ANSI/AAMI/ISO 11137-1:2006/(R)2015; A1:2013Sterilization of Health Care Products-Radiation - Part 1: Requirements for development,validation and routine control of a sterilization process for medical devices
ANSI/AAMI/ISO 11137-2:2013Sterilization of Health Care Products-Radiation - Part 2: Establishing the sterilizationdose
ANSI/AAMI/ISO 11137-3:2017Sterilization of Health Care Products-Radiation - Part 3: Guidance on dosimetricaspects.
ISO 10993-1:2009Biological Evaluation of Medical Devices-Part 1: Evaluation and testing within a riskmanagement process
ISO 10993-3:2014Biological Evaluation of Medical Devices-Part 3: Tests for genotoxicity,carcinogenicity, and reproductive toxicity
ISO 10993-4:2002,A1:2006Biological Evaluation of Medical Devices-Part 5: Selection of tests for interactions withblood
ISO 10993-5, 2009Biological Evaluation of Medical Devices-Part 5: Test for in vitro cytotoxicity
ISO 10993-10, 2010Biological Evaluation of Medical Devices-Part 10: Test for Irritation and SkinSensitization
ISO 10993-11, 2006Biological Evaluation of Medical Devices-Part 11: Test for Systemic Toxicity
ASTM F1980-16Standard Guide for Accelerated Aging of Sterile Medical Device Packages
ASTM-4169-16Standard Practice for Performance Testing of Shipping Containers and System
Pouch Leak test perASTM F2096-11Standard Test Method for Detecting Gross Leaks in Medical Packaging by InternalPressurization (Bubble Test)
Pouch Seal Strength testper ASTM F88/F88M-15Standard Test Method for Seal Strength of Flexible Barrier Materials

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Drawbridge Health. The logo consists of a stylized, abstract shape on the left, resembling a rounded, elongated teardrop or a stylized ribbon. The shape transitions in color from a light blue at the top to a reddish-orange at the bottom. To the right of the shape is the text "Drawbridge" in a dark gray sans-serif font, with the word "HEALTH" in a smaller font size directly below it.

1. Precision

A. Precision (Assay) - Repeatability

Repeatability (within-run and within-day) precision was determined according to CLSI Guideline EP05-A3. Blood samples spotted onto matrix strips and tested using different Beckman reagent lot combinations and one Beckman analyzer were analyzed for at least 20 days, two runs per day. A run was defined as running each sample in duplicate.

MeanRepeatability(within-run)Repeatability(within-day)Total
Sample%HbA1cSD%CVSD%CVSD%CV
15.100.0731.44%0.0300.58%0.1192.33%
26.460.0911.40%0.0330.51%0.1352.09%
37.870.0821.05%0.0250.32%0.1141.45%
411.440.1191.04%0.0710.62%0.1611.40%

Table 1a: Repeatability Precision (n=80*)

*total of 81 data points was collected, an additional data point for Sample 1

B. Precision (Device) - Lot-to-Lot Analysis

A lot-to-lot device precision study was conducted to demonstrate the reproducibility of collecting samples using different lots of manufactured OneDraw Blood Collection Devices across collection sites in accordance with CLSI Guideline EP05-A3. Figure 1a shows high agreement between results from the same participant, same operator, but different device lots. The range of %HbA1c was 4.84% to 7.55%. The average CV is 1.6% (range: 0.0% - 3.6%).

Image /page/7/Figure/10 description: The image is a scatter plot showing %HbA1c values on the y-axis and SID ordered by %HbA1c on the x-axis. There are three data series plotted: Blue, Green, and Orange. The %HbA1c values range from approximately 5.0 to 7.5, with the data points generally increasing from left to right.

Figure 1a: %HbA1c results for samples collected from two or three device lots across 23 participants

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Drawbridge Health. The logo consists of a stylized, abstract shape in blue and red on the left, followed by the text "Drawbridge" in gray. Below "Drawbridge" is the word "HEALTH" in smaller, gray letters. The overall design is clean and modern.

%HbA1cgroupMean%HbA1cWithin Lot %CV(95% CI)Between-Lot %CV(95%CI)Total %CV(95%CI)
< 65.231.95% (1.53, 2.36)0.37% (0.29, 0.44)1.99% (1.57, 2.42)
≥ 66.581.04% (0.70, 1.34)0.36% (0.24, 0.48)1.10% (0.74, 1.45)

Table 1b: Precision, Within and Between Lots and Total %CV

C. Precision (Device) - Operator-to-Operator

An inter-operator precision study was conducted to demonstrate the reproducibility of collecting samples by different operators across collection sites using the same lot of manufactured OneDraw Blood Collection Devices in accordance with CLSI Guideline EP05-A3. Figure 1b shows high agreement between results from the same participant, same lot of devices, by different operators. The range of %HbA1c was 4.66% - 14.38%. The average CV is 1.5% (range: 0.1% - 4.8%).

Image /page/8/Figure/6 description: The image is a scatter plot showing %HbA1c values for different SIDs, ordered by %HbA1c. The y-axis represents %HbA1c, ranging from 4 to 14, while the x-axis represents SIDs. There are four categories of data points, CF, GH, LS, and MM, each represented by a different color. The %HbA1c values generally increase from left to right along the x-axis.

Figure 1b: %HbA1c results for samples collected by two or three operators across 25 participants

%HbA1cgroupMean%HbA1cWithin Operator%CV (95% CI)Between-Operator %CV(95%CI)Total %CV(95%CI)
<65.211.97% (1.55, 2.38)1.06% (0.84, 1.30)2.24% (1.76, 2.72)
≥ 67.741.23% (0.83, 1.62)0.16% (0.11, 0.21)1.24% (0.84, 1.64)

Table 1c: Precision, Within and Between Operators and Total %CV

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the logo for Drawbridge Health. The logo consists of a stylized, abstract shape on the left, resembling a twisted ribbon or a stylized 'D'. The shape transitions in color from blue at the top to red/orange at the bottom. To the right of the shape is the text "Drawbridge" in a sans-serif font, with the word "HEALTH" in smaller letters underneath.

2. Method Comparison

The method comparison study was conducted to demonstrate the equivalence of HbA1c measurement using blood samples collected and tested using the OneDraw A1C Test System as compared to results using blood samples collected using standard venipuncture (tested using Beckman's NGSP-certified method on the Beckman Coulter AU480 Analyzer). Blood collections were conducted at two different clinical sites and sample analysis was conducted at the designated test laboratory.

Estimate95% lower bound95% upper bound
Slope1.000.971.03
Intercept-0.11
Pearson correlation coefficient (R)0.99070.98640.9937
Number of participants107
Range tested (OneDraw A1C TestSystem results, %HbA1c)4.70 – 14.3
Range tested (venipuncture,%HbA1c)4.76 - 14.7

Table 2: Regression analysis results (Passing-Bablok)

3. Linearity

The linearity of the OneDraw A1C Test is established as 4.70-14.3 % HbA1c (DCCT/NGSP) and in accordance with CLSI-EP6-A, determined the following linear fit, Y = 1.01x + 0.06; R=0.99.

4. Interference

Interference was assessed following EP07-A2. Studies were performed to assess common or known exogenous and endogenous substances that could interfere with the OneDraw A1C Test. The exogenous substances were Acetaminophen, Acetylsalicylic Acid, Ibuprofen, L-ascorbic Acid, Metformin, Glyburide; the endogenous substances were Rheumatoid Factors, Triglycerides, Bilirubin (conjugated and unconjugated); and Hemoglobin Variants (HbC, HbD, HbE, HbF, and HbS).

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Drawbridge Health. The logo consists of a stylized, abstract shape in blue and pink on the left, followed by the words "Drawbridge" in gray. Below "Drawbridge" is the word "HEALTH" in smaller, gray letters. The logo is clean and modern, with a focus on the company name.

Interference was calculated as the difference between (a) measurement of blood specimens with interferent as compared to (b) measurement of control group specimens without interferent. Interference is claimed when the difference is equal to or exceeds ±10% bias from the following substances:

SubstanceSubstancedescriptionHighestconcentration testedInterference observedWith highest testconcentration
AcetaminophenAnalgesic20 mg/dLNo
Acetylsalicylic acidAnalgesic65 mg/dLNo
GlyburideAntidiabetic0.2 mg/dLNo
IbuprofenAnti-inflammatory,analgesic50 mg/dLNo
MetforminAntidiabetic4.0 mg/dLNo
L-ascorbic acidAntioxidant3.0 mg/dLNo
TriglyceridesEndogenous3400 mg/dLNo
Bilirubin - conjugatedEndogenous33.2mg/dLNo
Bilirubin - unconjugatedEndogenous30mg/dLNo
Rheumatoid factorEndogenous600 IU/mLNo

Whole blood containing hemoglobin variants A2 (≤5.8%), C (≤40.1%), D (≤41.2%), E (≤22%), S (≤34.8%), or low F (≤ 8.5%) do not show significant interference (over 10% absolute bias) when spotted on matrix strips were compared to the reference laboratory liquid blood values. The OneDraw A1C Test is suitable for adoption in the laboratory with the appropriate disclosure in the labeling that Hemoglobin Variant HbF (elevated) has been known to interfere with test results. The following limitation is presented in the OneDraw A1C Test labeling (OneDraw A1C Test IFU): "Hemoglobinopathies may interfere with glycated hemoglobin analysis. Studies show that there is no significant interference for HbA2 (≤5.8%), HbC (≤ 40.1%), HbD (≤ 41.2%), HbE (≤ 22%), and HbS (≤ 34.8%). Samples containing HbF levels > 8.5% show a significant negative bias with the OneDraw A1C Test. Do not use this test if the patient has this variant."

5. Limits of Detection

The claimed measuring range is 4.70%-14.3% for the OneDraw A1C Test, which is based on linearity testing.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Drawbridge Health. The logo consists of a blue and pink abstract shape on the left, followed by the words "Drawbridge" in gray, with the word "HEALTH" in smaller gray letters underneath. The abstract shape is a curved, teardrop-like form with a gradient from blue at the top to pink at the bottom.

6. Product Stability

In accordance with EP25-A and EP09-A3, the difference in test results using aged OneDraw Blood Collection Devices was evaluated. Four levels of HbA1c covering the measuring range were evaluated. The OneDraw Blood Collection Device has an expiry date 12 months after the date of manufacturing.

Sample storage and shipping stability studies were conducted using whole blood spotted on matrix strips. Based on these studies, patient samples are stable for up to 21 days at room temperature. The results also demonstrate that extreme temperature excursions of whole blood spotted onto the matrix strips do not cause a significant difference in the measurement of HbA1c using the OneDraw A1C Test. Overall these studies demonstrate the robustness of the matrices and transport sleeves to ensure that they are able to withstand the stressed conditions experienced during shipping and storage of the sample prior to HbA1c testing.

7. Flex Studies

Flex studies were conducted to assess any potential pre-analytical error that could be obtained from blood samples by the testing laboratory. These studies evaluated blood sample acceptability in terms of:

  • (a) the minimum and maximum blood volume collected onto the device matrix strips
  • (b) potential interference of hematocrit with the OneDraw A1C Test
  • (c) the equivalency between the two matrix strips from the same cartridge for the OneDraw A1C Test
  • (d) a comparison of HbA1c results for the following:
    • Venous whole blood collected in EDTA tubes with venous whole blood spotted on the matrix O strips
    • Venous whole blood collected in EDTA tubes with capillary blood collected using OneDraw O Blood Collection Device
    • Venous whole blood spotted on the OneDraw matrix strips with capillary blood collected o using OneDraw Blood Collection Device

Results of Flex Studies

  • For minimum and maximum blood sample acceptability, two (2) HbA1c levels were tested with (a) six (6) blood volumes per level. There were 6 matrix strips per volume of blood, 72 matrix strips

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Drawbridge Health. The logo consists of a stylized, abstract shape on the left, with the company name "Drawbridge" in a sans-serif font to the right. Below "Drawbridge" is the word "HEALTH" in a smaller font size. The abstract shape is a gradient of blue and pink.

in total. This study demonstrated that spotted volumes ranging between 52.5 µL and 90 µL were within ±10% relative bias when compared to the corresponding 75 µL spotted samples.

  • The hematocrit interference study used two (2) HbA1c levels tested with the OneDraw A1C Test, (b) six matrix strips per level: six each of control, or "normal" hematocrit, high hematocrit, and low hematocrit specimens. The study demonstrated that varying hematocrit levels do not interfere with the OneDraw A1C Test.
  • The two matrix strips contained within the same cartridge were analyzed and compared. The data (c) demonstrates that there is not a significant difference in %HbA1c results between the two matrices.
  • (d) HbA1c results obtained from dried whole blood spotted on matrix strips and from capillary blood collected with the OneDraw Blood Collection Device are comparable to those collected via the standard phlebotomy technique (venous whole blood through venipuncture).

CONCLUSION

The information and data in this 510(k) application demonstrate that the samples collected and tested using the OneDraw A1C Test System result in accurate and precise results that correlate well with current cleared methods. Based on the criteria stated above, this device is found to be substantially equivalent to the predicate device.

§ 864.7470 Glycosylated hemoglobin assay.

(a)
Identification. A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.(b)
Classification. Class II (performance standards).