K Number
K183205
Device Name
PureBMC SupraPhysiologic Concentrating System 30 mL System, PureBMC SupraPhysiologic Concentrating System 60 mL System, PureBMC SupraPhysiologic Concentrating System 120 mL System
Manufacturer
Date Cleared
2019-05-30

(192 days)

Product Code
Regulation Number
862.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The PureBMC SupraPhysiologic Concentrating System is intended to be used in a clinical laboratory or intraoperatively at the point of care for the safe and rapid preparation of platelet concentrate from a small sample of bone marrow aspirate. The safety and effectiveness of this device for in vivo indications for use has not been established.
Device Description
Not Found
More Information

Not Found

Not Found

No
The summary does not mention AI, ML, image processing, or any other technology typically associated with AI/ML in medical devices. The description focuses on a system for concentrating bone marrow aspirate.

No
The device is intended for the preparation of platelet concentrate from bone marrow aspirate, which is a preparatory step, not a direct therapeutic intervention. The text explicitly states, "The safety and effectiveness of this device for in vivo indications for use has not been established," further indicating it's not a therapeutic device itself.

No
Explanation: The device is intended for the "preparation of platelet concentrate from a small sample of bone marrow aspirate," which is a preparative function, not a diagnostic one.

No

The description clearly indicates a "System" intended for "preparation of platelet concentrate from a small sample of bone marrow aspirate," which strongly suggests a physical device or kit for processing biological samples, not a software-only solution.

Based on the provided information, the PureBMC SupraPhysiologic Concentrating System is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is for the "preparation of platelet concentrate from a small sample of bone marrow aspirate." This is a process performed on a biological sample, but the device itself is not performing a diagnostic test on that sample to determine a disease or condition. It's a processing device.
  • Location of Use: While it can be used in a "clinical laboratory," it can also be used "intraoperatively at the point of care." IVDs are typically used to perform tests on samples, often in a laboratory setting, to provide diagnostic information.
  • Lack of Diagnostic Claims: The description does not mention any diagnostic claims or the measurement of any analytes to diagnose, monitor, or screen for a disease or condition.
  • Focus on Preparation: The core function is the "preparation" of a concentrate, not the analysis of the sample for diagnostic purposes.

Therefore, the device falls outside the typical definition and function of an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The PureBMC SupraPhysiologic Concentrating System is intended to be used in a clinical laboratory or intraoperatively at the point of care for the safe and rapid preparation of platelet concentrate from a small sample of bone marrow aspirate. The safety and effectiveness of this device for in vivo indications for use has not been established.

Product codes

JQC

Device Description

Not Found

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

clinical laboratory or intraoperatively at the point of care

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

Not Found

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 862.2050 General purpose laboratory equipment labeled or promoted for a specific medical use.

(a)
Identification. General purpose laboratory equipment labeled or promoted for a specific medical use is a device that is intended to prepare or examine specimens from the human body and that is labeled or promoted for a specific medical use.(b)
Classification. Class I (general controls). The device is identified in paragraph (a) of this section and is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The device is also exempt from the current good manufacturing practice requirements of the quality system regulation in part 820 of this chapter, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.

0

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 30, 2019

EmCyte Corporation Patrick Pennie Chairman & CEO 4331 Veronica S. Shoemaker Blvd, Suite 4 Fort Myers, Florida 33916

Re: K183205

Trade/Device Name: PureBMC SupraPhysiologic Concentrating System 30 mL System. PureBMC SupraPhysiologic Concentrating System 60 mL System. PureBMC SupraPhysiologic Concentrating System 120 mL System Regulation Number: 21 CFR 862.2050 Regulation Name: General purpose laboratory equipment labeled or promoted for a specific medical use Regulatory Class: Class I Product Code: JQC Dated: October 29, 2018 Received: November 19, 2018

Dear Patrick Pennie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

The OHT7: Office of In Vitro Diagnostics and Radiological Health has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that

1

such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the device's labeling:

The safety and effectiveness of this device for in vivo indications for use has not been established.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and

2

Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For, Timothy Stenzel, M.D., Ph.D. Director OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

3

Indications for Use

510(k) Number (if known) K183205

Device Name

PureBMC SupraPhysiologic Concentrating System

Indications for Use (Describe)

The PureBMC SupraPhysiologic Concentrating System is intended to be used in a clinical laboratory or intraoperatively at the point of care for the safe and rapid preparation of platelet concentrate from a small sample of bone marrow aspirate. The safety and effectiveness of this device for in vivo indications for use has not been established.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."