K Number
K183117

Validate with FDA (Live)

Date Cleared
2018-11-15

(6 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PathLoc-L MIS Spinal System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine. The PathLoc-L MIS Spinal System can be used in an open approach and a percutaneous approach.

The PathLoc-L MIS Spinal System is intended for the following indications:

• Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)

  • · Spondylolisthesis
  • · Trauma (i.e., fracture or dislocation)
  • · Spinal stenosis
  • · Curvatures (i.e., scoliosis, kyphosis, lordosis)
  • Tumor
  • Pseudarthrosis
  • · Failed previous fusion
Device Description

PathLoc-L MIS Spinal System consists of poly screws, straight rods, curved rods and set screw components that can be used via percutaneous surgical approach. The components are available in a variety of diameters and lengths in order to accommodate patient anatomy and are made of titanium alloy (ASTM F136).

AI/ML Overview

The provided text is a 510(k) Summary for the PathLoc-L MIS Spinal System. It describes a medical device, which is a pedicle screw spinal system used to stabilize spinal segments. The submission is for adding new components and changing part numbers to an already cleared device (K161766).

Based on the document, this is not a study that proves a device meets acceptance criteria for an AI/CADe system. This is a submission for a physical medical device (spinal implant), and the "performance testing" section refers to mechanical tests conducted on the physical components of the spinal system, not the performance of an AI algorithm.

Therefore, most of the requested information regarding AI/CADe systems (e.g., sample sizes for test/training sets, ground truth establishment, MRMC studies, standalone algorithm performance, expert qualifications, adjudication methods) is not applicable to this document.

However, I can extract the relevant "acceptance criteria" and "device performance" in the context of this physical device's mechanical testing:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Test)Reported Device Performance (as stated in the document)
ASTM F1717 Static Compression Test"The PathLoc-L MIS Spinal System was tested according to ASTM F1717... specifically, static compression test" (Implicitly, the device met the requirements of this standard as it was found "as safe, as effective and perform as well as the predicate device".)
ASTM F1717 Static Tension Test"The PathLoc-L MIS Spinal System was tested according to ASTM F1717... specifically, static tension test"
ASTM F1717 Static Torsion Test"The PathLoc-L MIS Spinal System was tested according to ASTM F1717... specifically, static torsion test"
ASTM F1717 Dynamic Fatigue Test"The PathLoc-L MIS Spinal System was tested according to ASTM F1717... and dynamic fatigue test"
ASTM F543 Axial Pullout Test"The PathLoc-L MIS Spinal System was tested... under ASTM F543, axial pullout test"

Note on "Acceptance Criteria" and "Reported Performance": The document states that the performance testing was conducted according to specific ASTM standards. For a 510(k) submission, meeting these recognized standards for mechanical properties means the device is considered as safe and effective as a predicate device that also meets these standards. The document doesn't provide specific numerical results of these tests, but rather states that the tests were performed and the device (including the new components) performs as well as the predicate.

Given this is a physical implant device, the following points are not directly extractable or applicable as they pertain to AI/CADe systems:

  • 2. Sample size used for the test set and the data provenance: Not applicable. This is not an AI/CADe system.
  • 3. Number of experts used to establish the ground truth... and the qualifications of those experts: Not applicable. Ground truth for an AI system is not relevant here.
  • 4. Adjudication method: Not applicable.
  • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  • 7. The type of ground truth used: Not applicable.
  • 8. The sample size for the training set: Not applicable.
  • 9. How the ground truth for the training set was established: Not applicable.

In summary, the provided document details the regulatory submission for a physical spinal implant system, focusing on its mechanical properties and equivalence to a predicate device, rather than the performance of a diagnostic AI algorithm.

{0}------------------------------------------------

November 15, 2018

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym with the full agency name on the right. The FDA part of the logo is in blue, with the acronym in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" written in a sans-serif font.

L&K Biomed Co., Ltd. Ms. Jihyeon Seo Regulatory Affairs Associate #201. #202 16-25. Dongbaekjungang-ro 16 beon-gil. Giheung-gu, Yongin-si, Gyeonggi-do, 17015 KOREA

Re: K183117

Trade/Device Name: PathLoc-L MIS Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: November 8, 2018 Received: November 9, 2018

Dear Ms. Seo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Colin O'neill -S

for MNM

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K183117

Device Name PathLoc-L MIS Spinal System

The PathLoc-L MIS Spinal System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine. The PathLoc-L MIS Spinal System can be used in an open approach and a percutaneous approach.

The PathLoc-L MIS Spinal System is intended for the following indications:

• Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)

  • · Spondylolisthesis
  • · Trauma (i.e., fracture or dislocation)
  • · Spinal stenosis
  • · Curvatures (i.e., scoliosis, kyphosis, lordosis)
  • Tumor
  • Pseudarthrosis
  • · Failed previous fusion

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):

  • Submitter: L&K BIOMED Co., Ltd. 1. #201, #202 16-25, Dongbaekjungang-ro 16 beon-qil Giheung-gu, Yongin-si, Gyeonggi-do, 17015, Korea Phone. 82-2-6717-1983 e-mail: kate.seo@lnkbiomed.com Contact Person: Jihyeon Seo Date prepared November 8th, 2018

Device Identification 2.

Trade NamePathLoc-L MIS Spinal System
Regulatory ClassClass II
Regulation Name/Common NamePedicle Screw Spinal System
Classification NameThoracolumbosacral pedicle screwsystem (21 CFR 888.3070)
PanelOrthopedic
Product CodeNKB

3. Purpose of 510(k)

L&K BIOMED Co. Ltd., is submitting this Special 510(k) to add new components and to change part number of two set screws of the existing 510(k)-cleared device (K161766)

The name of the legally marketed (unmodified) device 4.

K161766 PathLoc-L MIS Spinal System (primary predicate)

Description of the Device 5.

PathLoc-L MIS Spinal System consists of poly screws, straight rods, curved rods and set screw components that can be used via percutaneous surgical approach. The components are available in a variety of diameters and lengths in order to accommodate patient anatomy and are made of titanium alloy (ASTM F136).

{4}------------------------------------------------

Indication for Use 6.

The PathLoc-L MIS Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The PathLoc-L MIS Spinal System can be used in an open approach and a percutaneous approach.

The PathLoc-L MIS Spinal System is intended for the following indications:

  • Degenerative disc disease (defined as back pain of discogenic origin with • degeneration of the disc confirmed by history and radiographic studies)
  • Spondylolisthesis
  • . Trauma (i.e., fracture or dislocation)
  • Spinal stenosis
  • Curvatures (i.e., scoliosis, kyphosis, lordosis) .
  • . Tumor
  • . Pseudarthrosis
  • . Failed previous fusion

7. Comparison of the technological characteristics of the subject and predicate devices

The new components of the PathLoc-L MIS Spinal System are considered substantially equivalent to the referenced device (K161766). The systems have the same design, material, scientific technology, and indications for use. The only difference is that there are additional components which use different sterilization method (gamma sterilization) and changed catalog number of two set screws of the existing device.

Performance Testing 8.

Mechanical performance of additional components are the same with the legally marketed device (unmodified) PathLoc-L MIS Spinal System (K161766). Therefore, we substitute mechanical test data of additional components of PathLoc-L MIS Spinal System with the one of unmodified device (K161766).

{5}------------------------------------------------

A rationale was provided as to why the gamma sterilization does not affect the performance of the device. They are the same product in all aspect, except sterilization method.

The PathLoc-L MIS Spinal System was tested according to ASTM F1717, specifically, static compression test, static tension test, static torsion test and dynamic fatigue test, and under ASTM F543, axial pullout test.

Conclusion 9.

The additional components of PathLoc-L MIS Spinal System are as safe, as effective and perform as well as the predicate device. Therefore, the additional components of PathLoc-L MIS Spinal System are substantially equivalent to the predicate device (K161766).

N/A