(56 days)
The PathLoc-L MIS Spinal System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The PathLoc-L MIS Spinal System can be used in an open approach and a percutaneous approach.
The PathLoc-L MIS Spinal System is intended for the following indications:
· Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
- · Spondylolisthesis
- · Trauma (i.e., fracture or dislocation)
- · Spinal stenosis
- · Curvatures (i.e., scoliosis, kyphosis, lordosis)
- · Tumor
- Pseudarthrosis
- · Failed previous fusion
PathLoc-L MIS System consists of cannulated poly screws, straight rods, curved rods and set screw components that can be used via percutaneous surgical approach. The components are available in a variety of diameters and lengths in order to accommodate patient anatomy and are made from titanium alloy (ASTM F136). The implants will be provided non-sterile.
The provided text is an FDA 510(k) Premarket Notification from L&K BIOMED Company, Limited for their Pathloc-L MIS Spinal System. It is an approval letter and a summary of the device and its intended use, as well as a comparison to a predicate device and a statement on performance testing.
However, this document does not contain information about an AI/ML-based device that would require acceptance criteria for algorithmic performance, expert adjudication, or MRMC studies. The device described is a physical medical implant (pedicle screw spinal system) for surgical use. The "performance testing" mentioned in the document refers to mechanical testing (pullout testing per ASTM F543-13) of the physical screws, not the performance of an AI algorithm.
Therefore, the requested information regarding acceptance criteria, study details for AI performance, sample sizes for AI test/training sets, ground truth establishment, expert qualifications, and MRMC studies cannot be extracted from this document as it pertains to a mechanical medical device, not a software or AI-driven diagnostic/therapeutic device.
If the request was based on a misunderstanding of the document's content, please provide a different document that describes an AI/ML medical device for analysis.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, representing the department's mission to protect the health of all Americans and provide essential human services.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2016
L&K BIOMED Company, Limited Ms. Yerim An Regulatory Affairs Specialist #201, 202 16-25, Dongbaekjungang-ro 16 Beon-gil Giheung-gu, Yongin-si, Gyeonggi-do 446-916 KOREA
Re: K161766
Trade/Device Name: Pathloc-L MIS Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI Dated: June 20, 2016 Received: June 27, 2016
Dear Ms. An:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Vincent J. Devlin -S
for
- Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K161766
Device Name PathLoc-L MIS Spinal System
Indications for Use (Describe)
The PathLoc-L MIS Spinal System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The PathLoc-L MIS Spinal System can be used in an open approach and a percutaneous approach.
The PathLoc-L MIS Spinal System is intended for the following indications:
· Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
- · Spondylolisthesis
- · Trauma (i.e., fracture or dislocation)
- · Spinal stenosis
- · Curvatures (i.e., scoliosis, kyphosis, lordosis)
- · Tumor
- Pseudarthrosis
- · Failed previous fusion
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):
- Submitter: L&K BIOMED Company, Limited 1. #201, 202 16-25, Dongbaekjungang-ro 16 beon-qil Giheung-gu, Yongin-si, Gyeonggi-do, 446-916, Korea Phone. 82-2-6717-1985 e-mail: yerim2706@gmail.com Contact Person: Yerim An Date prepared Aug, 12, 2016
Device Identification 2.
| Trade Name | PathLoc-L MIS Spinal System |
|---|---|
| Regulatory Class | Class III |
| Regulation Name/Common Name | Pedicle Screw Spinal System |
| Panel | Orthopaedic |
| Product Codes | NKB, MNH, MNI |
3. Purpose of 510(k)
L&K BIOMED Co. Ltd., hereby submits this 510(k) proposing the addition of new screw types to the PathLoc-L MIS Spinal System.
Predicate Device 4.
- Primary Predicate Device: K120140 LnK MIS Spinal System
Description of the Device 5.
PathLoc-L MIS System consists of cannulated poly screws, straight rods, curved rods and set screw components that can be used via percutaneous surgical approach. The components are available in a variety of diameters and lengths in order to accommodate patient anatomy and are made from titanium alloy (ASTM F136). The implants will be provided non-sterile.
Indication for Use 6.
The PathLoc-L MIS Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the
{4}------------------------------------------------
thoracic, lumbar and sacral spine. The PathLoc-L MIS Spinal System can be used in an open approach and a percutaneous approach.
The PathLoc-L MIS Spinal System is intended for the following indications:
- · Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)
- · Spondylolisthesis
- · Trauma (i.e., fracture or dislocation)
- · Spinal stenosis
- · Curvatures (i.e., scoliosis, kyphosis, lordosis)
- Tumor
- Pseudarthrosis
- · Failed previous fusion
Comparison of the technological characteristics of the subject and 7. predicate devices
The PathLoc-L MIS Spinal System is considered substantially equivalent to the legally marketed LnK MIS Spinal System (K120140). The systems have similar design, material, scientific technology, and indications for use. The only difference is the addition of different screw types which have been evaluated via mechanical testing.
8. Performance Testing
Pullout testing per ASTM F543-13 was performed on the subject screws and compared to a legally marketed predicate device. Results of the testing demonstrate substantially equivalent performance.
Conclusion 9.
The PathLoc-L MIS Spinal System is substantially equivalent to a legally marketed predicate.