K Number
K183027
Device Name
Steerable catheter control system
Manufacturer
Date Cleared
2019-09-06

(309 days)

Product Code
Regulation Number
870.5700
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Niobe System is intended to navigate compatible magnetic devices through tissue to designated target sites in the right and left heart and coronary vascular and peripheral vasculature by orienting the device tip in a desired direction. The Cardiodrive Catheter Advancement System (CAS) is intended to automatically advance and retract compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart when used in conjunction with a Stereotaxis MNS. The Cardiodrive system is not intended to advance the EP mapping and ablation catheters through the coronary vasculature or the coronary sinus. The Cardiodrive system is not intended to advance or retract non-compatible catheters and/or other non-compatible devices into the neurovasculature.
Device Description
The Stereotaxis Niobe Magnetic Navigation System (MNS) with Navigant Workstation (NWS) and Cardiodrive (a.k.a. the Niobe System) is an interventional workstation for the intravascular navigation of appropriately equipped, magnetically adapted, devices (e.g., catheters or guidewires) through tissue to designated target sites. The Niobe System uses computer-controlled permanent magnets for orientating the tip of a magnetic device and employs magnetic fields only to orient or steer the tip of a magnetic device and remotely advance and retract only compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart. The Niobe System incorporates software that determines the direction the magnetic field should be applied based on physician interaction with the user interface devices.
More Information

No
The description focuses on computer-controlled magnetic navigation based on physician input, with no mention of AI or ML algorithms for decision-making or learning.

No.
The device (Niobe System) is described as an "interventional workstation" that uses magnetic fields to navigate and orient the tip of compatible interventional devices (like catheters or guidewires) within the body and to advance/retract specific types of catheters. It is a navigational and advancement system used in conjunction with other devices, rather than directly providing therapy itself. The therapy (e.g., ablation) is performed by the connected catheters.

No.

The Niobe System is an interventional workstation used for the intravascular navigation and advancement/retraction of catheters, primarily for ablation procedures, not for collecting diagnostic information about a patient's condition. While it interacts with EP mapping catheters, its described function is to steer or advance them, not to perform the diagnostic mapping itself.

No

The device description explicitly states that the system includes "computer-controlled permanent magnets" and is an "interventional workstation," indicating significant hardware components beyond just software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostic devices are used to examine specimens (like blood, urine, or tissue) taken from the human body to provide information about a person's health. This information is used for diagnosis, monitoring, or screening.
  • Device Function: The Niobe System is described as an "interventional workstation for the intravascular navigation of appropriately equipped, magnetically adapted, devices (e.g., catheters or guidewires) through tissue to designated target sites." It uses magnetic fields to orient and steer devices inside the patient's body.
  • Lack of Specimen Analysis: The description does not mention the device analyzing any biological specimens. Its function is purely related to the physical manipulation and navigation of medical devices within the patient.

Therefore, the Niobe System falls under the category of a therapeutic or interventional medical device, not an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The Niobe System is intended to navigate compatible magnetic devices through tissue to designated target sites in the right and left heart and coronary vascular and peripheral vasculature by orienting the device tip in a desired direction.

The Cardiodrive Catheter Advancement System (CAS) is intended to automatically advance and retract compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart when used in conjunction with a Stereotaxis MNS.

The Cardiodrive system is not intended to advance the EP mapping and ablation catheters through the coronary vasculature or the coronary sinus.

The Cardiodrive system is not intended to advance or retract non-compatible catheters and/or other non-compatible devices into the neurovasculature.

Product codes (comma separated list FDA assigned to the subject device)

PJB, NDO

Device Description

The Stereotaxis Niobe Magnetic Navigation System (MNS) with Navigant Workstation (NWS) and Cardiodrive (a.k.a. the Niobe System) is an interventional workstation for the intravascular navigation of appropriately equipped, magnetically adapted, devices (e.g., catheters or guidewires) through tissue to designated target sites. The Niobe System uses computer-controlled permanent magnets for orientating the tip of a magnetic device and employs magnetic fields only to orient or steer the tip of a magnetic device and remotely advance and retract only compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart. The Niobe System incorporates software that determines the direction the magnetic field should be applied based on physician interaction with the user interface devices.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

right and left heart and coronary vascular and peripheral vasculature
patient's heart

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-clinical mechanical performance testing:

  • Magnetic field performance testing without the catheter connected uses a THM-7025 Hall Effect tesla meter to measure the applied field strength and accuracy in the cardinal direction and at 0.08T and 0.1T.
  • Manual Control Performance Testing: Deflection testing, accuracy testing with target phantom, and anatomical position testing with a plastic heart model were performed for the reference Niobe device.
  • Remote Control Performance Testing: Testing demonstrates equivalent mechanical performance between the subject device and reference device.
  • Safety Controls: Testing of four primary safety controls related to physical movement of magnet positioners, cover movement toward patient, continuous advancement of catheter, and motion e-stops was conducted on the proposed device. All passed.
  • Simulated-use testing in a bench anatomic model or animal model: Four non-GLP animal validation studies were conducted using iterations of software after the reference device was cleared for features such as Navigant 4.4 (features and ablation lesion assessment, sheath visualization, zone mapping), Navigant 5.0.1 (Auto NaviLine/single step NaviLine), Navigant 5.0.2 (NaviLine, Targeting, Point and line ablation), Navigant 5.0.3 (Targeting/Electrode Targeting, NaviLine, Point and Line Ablation).

Non-clinical electrical testing:

  • EMC and Electrical safety testing showing conformance with IEC 60601 were performed by TuV for the reference device.
  • EMC testing for the proposed device hardware demonstrating conformance with IEC 60601 was performed by TuV. All tests passed.
  • Electrical safety testing on the proposed device demonstrating compliance with IEC 60601, was performed by TuV.
  • The proposed system has been tested for compatibility with specific x-ray, ablation generators, mapping systems. In addition, electrical isolation and emissions testing have been performed by TuV. All tests passed.

In vivo testing:

  • Manipulation and Positioning: The VERSATILE Study (K141530) and in vivo animal testing (K071029) for the Cardiodrive system demonstrated the ability to manipulate compatible catheters to pre-specified cardiac locations.
  • Safety: Clinical data from studies sponsored by Stereotaxis (P050029 Helios II ablation catheter (ATTRAC, ATTRAC II and HEART), K071029 Cardiodrive with Ablation Catheter (ATTRAC II), K140804 V-CAS (VERSATILE)) were reviewed.
    • Sample size: Total of 511 patients from 4 studies.
    • Major complication rates:
      • ATTRAC: 7/182 (3.8%)
      • ATTRAC II: 1/80 (1.3%)
      • HEART Study: 7/129 (5.4%)
      • VERSATILE: 5/120 (4.2%)
      • Total: 20/511 (3.9%)
    • Reported major adverse events within 7 days post procedure: 1 cardiac tamponade related to right sided catheter, 1 cardiac tamponade related to transseptal puncture, 1 new focal wall abnormality, 1 change in LVEF (60% to 45-50%), 2 vena cava thrombi, 1 groin complication, 1 prolonged hospitalization for grogginess, 1 pseudoaneurysm, 1 bleeding, 1 anemia, 1 dementia, 1 pericardial effusion, 1 heart block requiring pacemaker, 2 pulmonary embolisms, 1 AV fistula, 2 arrhythmia recurrence.
  • Efficacy: Data from the same clinical studies as above were reviewed.
    • Sample size: Total of 511 patients from 4 studies.
    • Acute Success Rates (RMN):
      • ATTRAC: 175/182 (96.2%)
      • ATTRAC II: 71/75 (94.7%)
      • HEART Study: 108/121 (89.3%)
      • VERSATILE: 119/120 (99.2%)
      • Total: 473/498 (95.0%)
    • 90 day success Rates (RMN):
      • ATTRAC: 145/147 (98.6%)
      • ATTRAC II: 51/54 (94.4%)
      • HEART Study: 82/87 (94.3%)
      • VERSATILE: Not reported
      • Total: 278/288 (96.5%)
  • User assessment of device remote controls and safety features: Performance impressions were documented on physician feedback forms during evaluations for Navistar RMT, Celsius RMT, and Helios II.
  • Fluoroscopy times:
    • ATTRAC: 15.89+/-13.15 [1.05, 66.5] minutes
    • ATTRAC II: 10.64 +/-9.78 [0.63, 61.0] minutes
    • HEART Study: 16.91 +/- 9.92 [3.13, 44.18] minutes
    • VERSATILE: 11.0 +/-10.5 [0.03, 55.2] minutes

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found. Acute success rates (95.0%) and 90-day success rates (96.5%) are provided, along with major complication rates (3.9%).

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K140394

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K051760, K060967, K071029

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.5700 Steerable cardiac ablation catheter remote control system.

(a)
Identification. A steerable cardiac ablation catheter remote control system is a prescription device that is external to the body and interacts with the manual handle of a steerable cardiac ablation catheter to remotely control the advancement, retraction, rotation, and deflection of a compatible, steerable ablation catheter used for the treatment of cardiac arrhythmias in the right side of the heart. The device allows reversion to manual control of the steerable cardiac ablation catheter without withdrawal of the catheter and interruption of the procedure.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Non-clinical mechanical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance testing must be performed:
(i) Mechanical performance of the system (without catheter connected);
(ii) Mechanical performance of the system with compatible catheters connected to verify that the system does not impact catheter function or performance. Assessments must include the following:
(A) Side-by-side remote control and manual comparisons of catheter manipulation (including all ranges of motion of catheter deflection and tip curl) for all compatible catheters; must include testing for worst-case conditions, and
(B) Evaluation of the accuracy and function of all device control safety features; and
(iii) Simulated-use testing in a bench anatomic model or animal model.
(2) Non-clinical electrical testing must include validation of electromagnetic compatibility (EMC), electrical safety, thermal safety, and electrical system performance. The following performance testing must be performed:
(i) Electrical performance of the system with compatible catheters connected to verify that the system does not impact catheter function or performance. Assessments must include the following:
(A) Side-by-side remote control and manual comparisons of catheter manipulation (including all ranges of motion of catheter deflection and tip curl) for all compatible catheters; must include testing for worst-case conditions, and
(B) Evaluation of the accuracy and function of all device control safety features; and
(ii) Electrical safety between the device and ablation catheter system and with other electrical equipment expected in the catheter lab or operating room.
(3) In vivo testing must demonstrate that the device performs as intended under anticipated conditions of use, including an assessment of the system impact on the functionality and performance of compatible catheters, and documentation of the adverse event profile associated with clinical use. Evidence must be submitted to address the following:
(i) Manipulation and Positioning: Ability to manipulate compatible catheters to pre-specified cardiac locations and confirm proper anatomic placement and tissue contact, in accordance with the system indications for use and the compatible catheter indications for use;
(ii) Safety: Assess device-related complication rate and major procedural complication rate (regardless of device relatedness) in comparison to literature and/or a manual comparison group for compatible ablation catheters to support the indications for use;
(iii) Efficacy: Assess ablation success in comparison to literature and/or a manual comparison group for compatible ablation catheters to support the indications for use; and
(iv) User assessment of device remote controls and safety features.
(4) Post-market surveillance (PMS) must be conducted and completed in accordance with FDA agreed upon PMS protocol.
(5) A training program must be included with sufficient educational elements that, upon completion of the training program, the clinician and supporting staff can:
(i) Identify the safe environments for device use,
(ii) Use all safety features of device, and
(iii) Operate the device in simulated or actual use environments representative of indicated environments and use for the indication of compatible catheters.
(6) Performance data must demonstrate the sterility of the sterile disposable components of the system.
(7) Performance data must support shelf life by demonstrating continued sterility of the device (of the sterile disposable components), package integrity, and device functionality over the requested shelf life.
(8) Labeling must include the following:
(i) Appropriate instructions, warnings, cautions, limitations, and information related to the intended patient population, compatible ablation catheters, and the device safeguards for the safe use of the device;
(ii) Specific instructions and the clinical training needed for the safe use of the device, which includes:
(A) Instructions on assembling the device in all available configurations, including installation and removal of compatible catheters;
(B) Instructions and explanation of all controls, inputs, and outputs;
(C) Instructions on all available modes or states of the device;
(D) Instructions on all safety features of the device; and
(E) Validated methods and instructions for reprocessing/disinfecting any reusable components;
(iii) A detailed summary of the mechanical compatibility testing including:
(A) A table with a complete list of compatible catheters tested (manufacturer trade name and model number), and
(B) A table with detailed test results, including type of test, acceptance criteria, and test results (
i.e., pass for meeting acceptance criteria);(iv) A detailed summary of the in vivo testing including:
(A) A table with a complete list of compatible catheters used during testing (manufacturer trade name and model number);
(B) Adverse events encountered pertinent to use of the device under use conditions;
(C) A detailed summary of the device- and procedure-related complications; and
(D) A summary of study outcomes and endpoints. Information pertinent to the fluoroscopy times/exposure for the procedure, patient, and operator fluoroscopic exposure;
(v) Other labeling items:
(A) A detailed summary of pertinent non-clinical testing information: EMC, mechanical, electrical, and sterilization of device and components;
(B) A detailed summary of the device technical parameters; and
(C) An expiration date/shelf life and storage conditions for the sterile accessories; and
(vi) When available, and according to the timeframe included in the PMS protocol agreed upon with FDA, provide a detailed summary of the PMS data including:
(A) Updates to the labeling to accurately reflect outcomes or necessary modifications based upon data collected during the PMS experience, and
(B) Inclusion of results and adverse events associated with utilization of the device during the PMS.

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

September 6, 2019

Stereotaxis. Inc. Kenneth Lock Sr. Director, Clinical Affairs 4320 Forest Park Avenue, Suite 100 St. Louis, Missouri 63108

Re: K183027

Trade/Device Name: Steerable catheter control system Regulation Number: 21 CFR 870.5700 Regulation Name: Steerable Cardiac Ablation Catheter Remote Control System Regulatory Class: Class II Product Code: PJB, NDO Dated: October 31, 2018 Received: November 1, 2018

Dear Kenneth Lock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

1

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark Fellman Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K183027

Device Name

Niobe® Magnetic Navigation System (MNS) with Navigation Workstation (NWS) and Cardiodrive®

Indications for Use (Describe)

The Niobe System is intended to navigate compatible magnetic devices through tissue to designated target sites in the right and left heart and coronary vascular and peripheral vasculature by orienting the device tip in a desired direction.

The Cardiodrive Catheter Advancement System (CAS) is intended to automatically advance and retract compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart when used in conjunction with a Stereotaxis MNS.

The Cardiodrive system is not intended to advance the EP mapping and ablation catheters through the coronary vasculature or the coronary sinus.

The Cardiodrive system is not intended to advance or retract non-compatible catheters and/or other non-compatible devices into the neurovasculature.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Image /page/3/Picture/1 description: The image shows the Stereotaxis logo. The logo consists of an orange abstract shape on the left and the word "STEREOTAXIS" in black font on the right. The abstract shape is made up of two curved shapes facing each other with a blue dot in the center.

I. SUBMITTER

Date Prepared:September 6, 2019
Submitter:Stereotaxis, Inc.
4320 Forest Park Avenue
St. Louis, MO 63108 USA
Contact Person:Kenneth H. Lock
Sr. Director, Clinical Affairs
Phone: (314) 678-6123
Fax: (763) 493-7849

II. DEVICE

| Name of Device: | Niobe® Magnetic Navigation System (MNS) with Navigant™
Workstation (NWS) and Cardiodrive® |
|-----------------------|----------------------------------------------------------------------------------------------|
| Common or Usual Name: | Steerable cardiac ablation catheter remote control system. |
| Classification Name: | Catheter remote control system |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 870.5700 |
| Product Code: | PJB |
| FDA Panel: | Cardiology |

III. PREDICATE DEVICE

Predicate510(k)RegulationDevice NameManufacturer
PrimaryK140394820.5700Amigo Remote Catheter SystemCatheter Robotics, Inc.
AdditionalK051760
K060967820.1290Niobe Magnetic Navigation system with Navigant Navigation WorkstationStereotaxis, Inc.
AdditionalK071029820.1290CardioDriveStereotaxis, Inc.

IV.DEVICE DESCRIPTION

The Stereotaxis Niobe Magnetic Navigation System (MNS) with Navigant Workstation (NWS) and Cardiodrive (a.k.a. the Niobe System) is an interventional workstation for the intravascular navigation of appropriately equipped, magnetically adapted, devices (e.g., catheters or guidewires) through tissue to designated target sites. The Niobe System uses computer-controlled permanent magnets for orientating the tip of a magnetic device and employs magnetic fields only to orient or steer the tip of a magnetic device and remotely advance and retract only compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart. The Niobe System incorporates software that determines the direction the magnetic field should be applied based on physician interaction with the user interface devices.

4

Image /page/4/Picture/1 description: The image shows the logo for Stereotaxis. The logo consists of an orange abstract shape on the left, resembling two curved leaves with a blue dot in the center. To the right of the shape is the word "STEREOTAXIS" in a simple, sans-serif font, with each letter in black.

V. INDICATIONS FOR USE

The Niobe System is intended to navigate compatible magnetic devices through tissue to designated target sites in the right and left heart and coronary vasculature, neurovascular and peripheral vasculature by orienting the device tip in a desired direction.

The Cardiodrive Catheter Advancement System (CAS) is intended to automatically advance and retract compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart when used in conjunction with a Stereotaxis MNS.

The Cardiodrive system is not intended to advance the EP mapping and ablation catheters through the coronary vasculature or the coronary sinus.

The Cardiodrive system is not intended to advance or retract non-compatible catheters and/or other noncompatible devices into the neurovasculature.

VI.COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The Niobe System has the same primary technological characteristics as the predicate device, including, but not limited to the following:

Primary Predicate Device:

  • . Remotely controlling PMA class III ablation catheter
    Additional Reference Devices

  • . Primary Design Elements (magnets, positioners, software)

  • . Principle of Operation (magnetic navigation)

  • Performance Parameters (field modes and positions) .

  • User Interface Devices (controllers, keyboard, display monitors) .

  • Companion Devices and Systems (digital fluoroscopy systems,) .

Any minor differences in technological characteristics between the proposed and reference Niobe System have demonstrated continued performance and do not raise new issues of safety or effectiveness.

VII. NON-CLINICAL STUDIES

The Niobe System underwent verification and validation activities to ensure that the device continues to meet established performance criteria. Evaluations included:

A. Biocompatibility Testing

The Niobe System does not contain patient-contacting components. Biocompatibility testing was not required to support of this submission.

B. Electromagnetic Compatibility (EMC)

Stereotaxis conducted appropriate Electromagnetic compatibility testing (EMC) in accordance with IEC 60601-1 and collateral standards (i.e., ISO 60601-1-2) to ensure that the Niobe System device does not interfere with other medical equipment and is immune to interference from other equipment.

2

5

Image /page/5/Picture/1 description: The image shows the logo for Stereotaxis. The logo consists of an orange abstract symbol on the left, resembling two curved shapes with a blue dot in the center. To the right of the symbol is the word "STEREOTAXIS" in a simple, sans-serif font, with each letter in black.

C. Software Verification and Validation

Stereotaxis performed software verification and validation activities in accordance with ISO 62304 and ISO 14971 for software utilized by the Niobe System device to ensure that the software is safe and appropriate risk control measures are effective.

D. Performance Testing

Where appropriate, system verification and validation activities were conducted to assess the performance of the device following software and/or hardware modifications. The testing demonstrated that the Niobe System met pre-determined requirements and continued to perform as intended.

E. Special Controls Testing Summary

Special ControlsHow Special Control Has Been Met
1) Non-clinical mechanical performance
testing must demonstrate that the device
performs as intended under anticipated
conditions of use. The following
performance testing must be performed:
i. Mechanical performance of the system
(without catheter connected);Using the proposed device, Magnetic field performance testing
without the catheter connected uses a THM-7025 Hall Effect tesla
meter to measure the applied field strength and accuracy in the
cardinal direction and at 0.08T and 0.1T (which represent worst
case positions).
ii. Mechanical performance of the system
with compatible catheters connected to
verify that the system does not impact
catheter function or performance.
Assessments must include the
following:
(A) Side-by-side remote control and
manual comparisons of catheter
manipulation (including all ranges
of motion of catheter deflection and
tip curl) for all compatible
catheters; must include testing for
worst-case conditions, andManual Control Performance Testing
Performance testing data for the reference Niobe device included
deflection testing, accuracy testing with target phantom and
anatomical position testing with a plastic heart model. The results
of this testing provide mechanical performance data for the
reference device.
Remote Control Performance Testing
The current version (proposed device) of the Niobe MNS
Substantial Equivalence Application Testing demonstrates
equivalent mechanical performance between the subject device
and reference device.
(B) Evaluation of the accuracy and
function of all device control safety
features; andTesting of the four primary safety controls related to the physical
motion of the system was conducted on the proposed device.
These concern physical movement of the magnet positioners,
movement of the covers toward the patient, continuous
advancement of the catheter, and motion e-stops. All four safety
controls passed the test.
iii. Simulated-use testing in a bench
anatomic model or animal model.Four non-GLP animal validation studies were conducted and
tested the following using iterations of software after the
reference device was cleared:
Navigant 4.4 features and ablation lesion assessment
• Sheath visualization
• Zone mapping
Navigant 5.0.1 Validation Study
Special ControlsHow Special Control Has Been Met
• Auto NaviLine ./ single step NaviLine
Navigant 5.0.2 Validation Study
• NaviLine
• Targeting
• Point and line ablation
Navigant 5.0.3 Validation Study
• Targeting / Electrode Targeting
• NaviLine
• Point and Line Ablation
2. Non-clinical electrical testing must
include validation of electromagnetic
compatibility (EMC), electrical safety,
thermal safety, and electrical system
performance. The following performance
testing must be performed:
i. Electrical performance of the systemManual Control Performance Testing
with compatible catheters connected
to verify that the system does not
impact catheter function orEMC and Electrical safety testing showing conformance with IEC
60601 were performed by TuV for the reference device.
performance. Assessments must
include the following:
A. Side-by-side remote control and
manual comparisons of catheter
manipulation (including all
ranges of motion of catheter
deflection and tip curl) for all
compatible catheters; must
include testing for worst-case
conditions, andRemote Control Performance Testing
EMC testing for the proposed device hardware demonstrating
conformance with IEC 60601 was performed by TuV.
All tests passed.
B. Evaluation of the accuracy and
function of all device control
safety features; andElectrical safety testing on the proposed device demonstrating
compliance with IEC 60601, was performed by TuV.
ii. Electrical safety between the device
and ablation catheter system and with
other electrical equipment expected in
the catheter lab or operating room.The proposed system has been tested for compatibility with
specific x-ray, ablation generators, mapping systems. In addition,
electrical isolation and emissions testing have been performed by
TuV.
All tests passed.

6

Image /page/6/Picture/1 description: The image shows the logo for Stereotaxis. The logo features an abstract orange graphic on the left, resembling two curved shapes with a blue dot in the center. To the right of the graphic is the word "STEREOTAXIS" in a simple, sans-serif font.

7

Image /page/7/Picture/1 description: The image shows the Stereotaxis logo. The logo consists of an orange abstract shape on the left, resembling two leaves facing each other with a blue dot in the center. To the right of the shape is the word "STEREOTAXIS" in a simple, sans-serif font.

Special ControlsHow Special Control Has Been Met
3. In vivo testing must demonstrate that the device performs as intended under anticipated conditions of use, including an assessment of the system impact on the functionality and performance of compatible catheters, and documentation of the adverse event profile associated with clinical use. Evidence must be submitted to address the following:
i. Manipulation and Positioning: Ability to manipulate compatible catheters to pre-specified cardiac locations and conform proper anatomic placement and tissue contact, in accordance with the system indications for use and compatible catheter indications for use;The VERSATILE Study used to support K141530 as well as in vivo animal testing provided in K071029 for the Cardiodrive system, demonstrated the ability to manipulate compatible catheters to pre-specified cardiac locations.
ii. Safety: Assess device-related complication rate and major procedural complication rate (regardless of device relatedness) in comparison to literature and/or manual comparison group for compatible ablation catheters to support the indications for use;Clinical data to support the safety of the Niobe Magnetic Navigation System, which includes Niobe, Navigant Software, the Cardiodrive catheter advancement system and a compatible ablation catheter, was reviewed.
The following is a summary of data from clinical studies sponsored by Stereotaxis that have been submitted to the FDA in three separate submissions:
• P050029 Helios II ablation catheter (ATTRAC, ATTRAC II and HEART)
K071029 Cardiodrive with Ablation Catheter (ATTRAC II)
K140804 V-CAS (VERSATILE)
A total of 511patients from 4 studies were enrolled using the Niobe Magnetic Navigation System, which included the Niobe MNS, Cardiodrive, a Magnetic Ablation Catheter, and the Navigant Software.
The following table describes the major complication rates reported in these trials.
StudyRMN Patients7 Day Major
Complication
ATTRAC1827/182 (3.8%)
ATTRAC II801/80 (1.3%)
HEART Study1297/129 (5.4%)
VERSATILE1205/120 (4.2%)
Total51120/511 (3.9%)
Special ControlsHow Special Control Has Been Met
iii.
Efficacy: Assess ablation success
in comparison to literature and/or a
manual comparison group for
compatible ablation catheters to
support the indications for use; andData to support the ablation success of the Niobe Magnetic
Navigation System which includes Niobe, Navigant Software, the
Cardiodrive catheter advancement system and a compatible
ablation catheter was reviewed.
The following is a summary of data from clinical studies
sponsored by Stereotaxis that have been submitted to the FDA in
three separate submissions:
P050029 Helios II ablation catheter (ATTRAC, ATTRAC II and HEART study) K071029 Cardiodrive with Ablation Catheter (ATTRAC) K140804 V-CAS (VERSATILE) A total of 511 patients from 4 studies were enrolled using the
Niobe Magnetic Navigation System, which included the Niobe
MNS, Cardiodrive, a Magnetic Ablation Catheter, and the
Navigant software. The ATTRAC study series and the HEART
Study used the Helios ablation catheter, and the Versatile study
used the BWI RMT Thermocool catheter. The following table
describes the acute and 90-day success rates reported in these
trials.
StudyRMN
PatientsAcute
Success
RMN90 day
success
RMN
ATTRAC182175/182
(96.2%)145/147
(98.6%)
ATTRAC II8071/75
(94.7%)51/54
(94.4%)
HEART Study129108/121
(89.3)82/87
(94.3%)
VERSATILE120119/120
(99.2%)Not reported
Total511473/498
(95.0%)278/288
(96.5%)
iv.
User assessment of device remote
controls and safety features.The Navigant system was used to steer the catheter during the
evaluations performed for Navistar RMT, Celsius RMT and
Helios II. Performance impressions were documented on
physician feedback forms which are included in the reports.
4. Post-market surveillance (PMS) must be
conducted and completed in accordance
with FDA agreed upon PMS protocol.Given the amount of clinical data from submitted studies and peer
reviewed publications, a post market surveillance study is not
warranted.
The following summarizes the overwhelming evidence available
to support the safety and efficacy of the MNS system:
Four studies submitted to the FDA demonstrated a major
adverse event rate of 3.9%. Data from these four studies
                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                 |                             |                          |

8

Image /page/8/Picture/1 description: The image shows the logo for Stereotaxis. The logo consists of an orange abstract shape on the left, resembling two curved leaves with a blue dot in the center. To the right of the shape is the word "STEREOTAXIS" in a simple, sans-serif font.

9

Image /page/9/Picture/1 description: The image shows the logo for Stereotaxis. The logo consists of an abstract orange symbol on the left, resembling two leaves with a blue dot in the center. To the right of the symbol is the word "STEREOTAXIS" in a simple, sans-serif font.

Special ControlsHow Special Control Has Been Met
and PMA approval in P050029.
Acute success and 90 success rates were 95.0% and 96.5% respectively in those studies More than 8,000 patients were reported in the literature using the MNS System with a major complication rate of 0.72% compared to a manual rate of 2.1% Acute success rates and long-term success rates were similar in both the MNS and manual groups Stereotaxis proposes to continue monitoring the safety and efficacy of the Niobe System through the literature and Post Marketing Surveillance program.
5. A training program must be included with
sufficient educational elements that, upon
completion of the training program, the
clinical and supporting staff can:
i. Identify the safe environments for
device use,
ii. Use all safety features of the device,
and
iii. Operate the device in simulated or
actual use environments
representative of indicated
environments and use for the
indication of compatible catheters.Representatives from the company train the physician and staff in the use of the Niobe Magnetic Navigation system.
The Niobe ES User's Manual provides detailed operating
instructions on the system and the navigation software.
This information is reviewed with the physician and staff during simulation sessions and phantom training with the actual system
6. Performance data must demonstrate the
sterility of the sterile disposable
components of the systemThe Niobe System is not provided in sterile form nor is it required to be sterilized prior to use, therefore, sterilization information is not applicable. The Niobe System does not include sterile disposable components as part of the system.
Cardiodrive includes a single-use disposable (QuikCAS) to interface with the compatible ablation catheter. The sterile components underwent sterilization testing. All sterilization testing resulted in a PASS.
7. Performance data must support shelf life
by demonstrating continued sterility of the
device (of the sterile disposable
components), package integrity, and
device functionality over the requested shelf life.The Niobe System is not provided in sterile form nor is it required to be sterilized prior to use, therefore, shelf life information,
including package integrity and function of the device over the stated shelf life are not applicable.
Cardiodrive includes a single-use disposable (QuikCAS) to interface with the compatible ablation catheter. The sterile components underwent shelf-life /packaging and sterilization testing. All sterilization and packaging testing resulted in a PASS, and each component was validated for a shelf life of 3 years.
Special ControlsHow Special Control Has Been Met
8. Labeling must include:
i. Appropriate instructions, warnings, cautions, limitations, and information related to the intended patient population, compatible ablation catheters, and the device safeguards for the device;
ii. Specific instructions and clinical training needed for the safe use of the device, which includes:Niobe ES User's Manual includes:
• Compatible Catheters
• Indications for Use
• Warnings
• Safety Controls
The Niobe System does not require assembly of the device prior to use. The Niobe System is considered permanent equipment installed in a Stereotaxis Magnetic Navigation Lab. Stereotaxis performs all installation activities for each Niobe System; therefore, assembly and installation instructions to the physician user are not required.
A. Instructions on assembling the device in all available configurations, including installation and removal of compatible catheters;With regard to use of magnetically-compatible devices with the Niobe System, assembly and installation of these devices to/with the Niobe System is not required. Magnetically-compatible catheters interface with the Cardiodrive CAS device which advances and retracts the device and instructions related to the catheter-CAS interface are provided in the Cardiodrive instructions for use. The Niobe System only applies a magnetic field to orient the distal tip of a magnetically-compatible device; therefore, the manipulation of the distal end of the device is by indirect (non-physical) means for which assembly or installation of other devices is not required.
B. Instructions on all available modes or states of the devices;
C. Instructions and explanation of all controls, inputs, and outputs
D. Instructions on all safety features of the device and
E. Validated methods and instructions for reprocessing/disinfecting any reusable componentsThe Niobe System User Guide includes an explanation and instructions for the following, controls, modes, and states of the device:
• Procedure room components
• System positions
• Tableside Magnet Controller
• System power up
• Cover Force Sensor
• Navigation Position Assistance
• Software Basic Information
• Cardiodrive CAS user interface
• Activation codes
The Niobe System is a medical device that does not have any reusable components that reprocessing or disinfection prior to use. The magnet pods are covered by drapes during the procedure and may be wiped clean using hospital-grade EPA-registered germicide solutions following each procedure. Cleaning instructions are provided in the "Cleaning the Niobe System" section of the Niobe System User Guide).
Special ControlsHow Special Control Has Been Met
iii.
A detailed summary of the
mechanical compatibility testing
including:
A. A table with a complete list of
compatible catheters tested
(manufacturer trade name and
model number), and- Biosense Webster Navistar RMT
  • Biosense Webster Navistar RMT Thermocool
  • Biosense Webster Celsius RMT
  • Biosense Webster Celsius RMT Thermocool
  • Stereotaxis Helios II |
    | B. A table with detailed test results,
    including type of test, acceptance
    criteria, and test results (i.e., pass
    for meeting acceptance criteria); | Mechanical performance testing of compatible catheter is
    included in the Niobe Users Guide. |
    | iv.
    A detailed summary of the in vivo
    testing including:
    A. A table with a complete list of
    compatible catheters used during
    testing (manufacturer trade name
    and model number);
    B. Adverse events encountered
    pertinent to use of the device
    under use conditions; | - Biosense Webster Navistar RMT
  • Biosense Webster Navistar RMT Thermocool
  • Biosense Webster Celsius RMT
  • Biosense Webster Celsius RMT Thermocool
  • Stereotaxis Helios II
    Data from four clinical studies involving 511 patients who
    underwent catheter ablation using the Stereotaxis Magnetic
    Navigation System (MNS) are summarized. Collectively, the data
    in these studies demonstrates the safety and effectiveness of the
    Magnetic Navigation System for catheter ablation.
    Study Design:
    All four studies were prospective in nature and included safety
    endpoints. Data from all four studies were used to support
    regulatory approvals. Evaluation of 7 day major adverse events
    for safety is reported.
    Study Results:
    Adverse Events:
    The overall 7 day major complication rate for all four studies was
    20/511 (3.9%) Major adverse events that occurred within 7 days
    post procedure included:
  • 1 cardiac tamponade related to right sided
    catheter
  • 1 cardiac tamponade related to the transseptal
    puncture
  • 1 new focal wall abnormality
  • 1 change in LVEF (60% to 45-50%)
  • 2 vena cava thrombi
  • 1 groin complication |
    | Special Controls | How Special Control Has Been Met |
    | | 1 prolonged hospitalization for grogginess 1 pseudoaneurysm 1 bleeding 1 anemia 1 dementia 1 pericardial effusion 1 heart block requiring pacemaker 2 pulmonary embolisms 1 AV fistula 2 arrhythmia recurrence (per protocol requirement) |
    | | Conclusion
    The data in this summary support the reasonable assurance of safety and effectiveness of the Stereotaxis Magnetic Navigation System for cardiac ablation procedures. |
    | | These data have been summarized in the Niobe ES User's Guide. |
    | C. A detailed summary of device-
    and procedure-related
    complications; and | The following listing of the procedure related complications have
    been reported in clinical trials submitted to the agency:
    Two (2) patients out of 177 (1.1%) had an acute major adverse
    event and 5/182 (2.7%) had a major adverse event within 7 days
    of ablation in the ATTRAC Study. There was 1 death reported
    in the study secondary to respiratory failure, unrelated to the
    device. No cardiac tamponades or perforations were reported.
    One patient experienced a groin complication, two patients
    experienced vena cava thrombi, one experienced chest soreness,
    and one had a prolonged hospitalization due to being groggy.
    There were no device related adverse events. |
    | | In the ATTRAC II Study, 1/80 (1.3%) patient had a major
    adverse event which was a cardiac tamponade due to a right
    sided non-study catheter. There were no deaths reported in this
    study. There were no device related adverse events. |
    | | In the HEART study, the overall acute major adverse event rate
    was 5.4% for the magnetic arm. No deaths or cardiac
    tamponades were reported. There was 1 pericardial effusion, 1
    heart block requiring pacemaker, 2 pulmonary embolisms, 1 AV
    fistula, and 2 arrhythmia recurrences (considered a major
    adverse event per the protocol) in the magnetic arm of the study.
    All were reported as device or procedure related adverse events. |
    | | The overall major adverse event rate in the VERSATILE Study
    was 5/120 (4.2%) in this study. There was one cardiac
    tamponade, one pseudo aneurysm, one bleeding, one patient
    presented with anemia and dementia. The cardiac tamponade
    event started as a pericardial effusion and progressed into |

10

Image /page/10/Picture/1 description: The image shows the logo for Stereotaxis. The logo consists of an abstract orange graphic on the left, resembling two curved shapes with a blue dot in the center. To the right of the graphic is the word "STEREOTAXIS" in a simple, sans-serif font.

11

Image /page/11/Picture/1 description: The image shows the logo for Stereotaxis. The logo consists of an orange abstract symbol on the left, resembling two curved shapes with a blue dot in the center. To the right of the symbol is the word "STEREOTAXIS" in a simple, sans-serif font.

12

Image /page/12/Picture/1 description: The image shows the logo for Stereotaxis. The logo consists of an orange abstract shape on the left, resembling two curved leaves with a blue dot in the center. To the right of the shape is the word "STEREOTAXIS" in a simple, sans-serif font.

13

Image /page/13/Picture/1 description: The image shows the Stereotaxis logo. The logo consists of an abstract orange graphic on the left, resembling two curved shapes with a blue dot in the center. To the right of the graphic is the word "STEREOTAXIS" in a simple, sans-serif font.

Special ControlsHow Special Control Has Been Met
cardiac tamponade due to the transseptal puncture. The DSM
adjudicated these events to be possibly and probably related tothe procedure, respectively. There were no device related
adverse events.
D. A summary of study outcomes
and endpoints. Information
pertinent to the fluoroscopy
times/exposure for the procedure,
patient, and operator
fluoroscopic exposure;The following table summarizes the acute and long term success
rates as well as fluoroscopy times reported in the clinical trials.
Study RMN
Patients Acute
Success
RMN 90 day success
RMN Mean
Fluoroscopy
time
Mean +/-S.D
[range] ATTRAC 182 175/182
(96.2%) 145/147
(98.6%) 15.89+/-13.15
[1.05, 66.5] ATTRAC II 80 71/75
(94.7%) 51/54 (94.4%) 10.64 +/-9.78
[0.63,61.0] HEART
Study 129 108/121
(89.3) 82/87 (94.3%) 16.91 +/- 9.92
[3.13,44.18 VERSATILE 120 119/120
(99.2%) Not reported 11.0 +/-10.5
[0.03,55.2] Total 511 473/498
(95.0%) 278/288
(96.5%)
V. Other labeling items:
A. A detailed summary of pertinent
non-clinical testing information:
EMC, mechanical, electrical, and
sterilization of device and
components;These data have been summarized in Niobe ES User's Guide.
EMC: The Niobe System User Manual includes a summary of
Electromagnetic Compatibility (EMC) testing under the section
titled "Electromagnetic Compatibility Information".
Electrical: In EMC section - Immunity (include ESD, etc.)
Mechanical: Non-clinical performance testing of compatible
catheters is included and summarized in the Niobe ES User's
Guide
Sterilization: The Niobe System is not provided in sterile form
nor is it required to be sterilized prior to use, therefore,
sterilization information is not applicable.
B. A detailed summary of the
device technical parameters; andThe Niobe system generates a directional 0.08T or 0.1T magnetic
field within the patient's heart. The navigation volume is 6 inches
in diameter centered at X-ray isocenter.
C. An expiration date/shelf life and
storage conditions for the sterile
accessories; andThe Niobe System is not provided in sterile form nor is it required
to be sterilized prior to use, therefore, expiration date, shelf life,
and storage conditions are not applicable.

14

Image /page/14/Picture/1 description: The image shows the Stereotaxis logo. The logo consists of an orange abstract shape on the left, resembling two curved leaves with a blue dot in the center. To the right of the shape is the word "STEREOTAXIS" in a simple, sans-serif font, with each letter in black.

Special ControlsHow Special Control Has Been Met
vi. When available, and according to the timeframe included in the PMS protocol agreed upon with FDA, provide a detailed summary of the PMS data including:
A. Updates to the labeling to accurately reflect outcomes or necessary modifications based upon data collected during the PMS experience, and
B. Inclusion of results and adverse events associated with utilization of the device during the PMS.Relevant warnings based on complaints and clinical studies have been included in The Niobe User's Guide.

VIII. ANIMAL STUDIES

Animal studies were not required to support the modifications to the Niobe System however; non-GLP animal studies for validation of specific versions of software features are included in the special controls.

IX.CLINICAL STUDIES

Clinical studies were not required to support the modifications to the Niobe System however, human clinical data from other regulatory submissions are included in the special controls.

X. CONCLUSION

Based on the information presented in this Premarket Notification, Stereotaxis concludes that the Niobe System is substantially equivalent to the predicate and reference devices based on the same intended use and same primary technological characteristics. Any minor differences in technological characteristics between the proposed and reference Niobe System have demonstrated continued performance and do not raise new issues of safety or effectiveness.