K Number
K182966
Device Name
PLANET Onco Dose
Manufacturer
Date Cleared
2019-03-06

(131 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PLANET Onco Dose is a standalone software intended to be used with PET or SPECT hybrid imaging systems in order to manage, process, display and analyze nuclear medical medical image series, to assist in medical diagnosis, to assist in treatment analysis and in therapy response assessment, to assist in the contouring of region of interest for radiotherapy.

PLANET Onco Dose is dedicated to be used by qualified medical professionals in Molecular Imaging, and/or Medical Oncology.

The medical modalities of these medical imaging systems include CT. MRI, SPECT, PET. XA, RT Struct and RT Dose as supported by ACR/NEMA DICOM 3 standard format.

PLANET Onco Dose provides the user with the means to segment structures in medical image volumes by providing dedicated delineation, contouring and propagation tools for both tumors and normal tissues (i.e. Regions of Interest (ROI)).

PLANET Onco Dose provides tools to display, co-register (including deformable registration), compute Standardized Uptake Value (SUV) and import / export results (contours and dosimetries) to / from Treatment Planning Systems (TPS) and PACS devices for assessment and response of patients undergoing a course of oncology treatment.

PLANET Onco Dose provides the user with the means to assist in the assessment quantification of radiation doses received by tissues as a result of administering a radionuclide (e.g. Permanent Yttrium-90 microsphere implants).

PLANET Onco Dose provides tools for post-treatment absorbed dose calculation and evaluation on PET and SPECT images. The following functions are available to allow dose calculations for patients after they have received a treatment using permanent Yttrium-90 (Y90) microspheres:

  • 3D liver-lung shunt assessment;

  • Local Deposition Model;

  • Voxel S Value approach based on the schema in MIRD Pamphlet 17 [1]:

  • Dosimetry based on 90Y-microspheres-PET (or SPECT Bremsstrahlung) series;

  • Compatible with PET images acquired with another radioisotope instead of Y90 when Y90 acquisitions are not supported by the scanner (correction of branching ratio and decay parameters).

For Y90 microspheres. PLANET Onco Dose cannot be used to prescribe the radionuclide activity to be administered to the patient for the therapy. The user has to provide the parameters (e.g. activity) in order for PLANET Onco Dose to estimate the radiation doses that the tissues received as a result of the administration.

PLANET Onco Dose should only be used for the retrospective determination of dose and not for the case where is a need for retreatment using Y90 microspheres.

Device Description

PLANET Onco Dose is a software platform dedicated to medical diagnosis aid, therapy response assessment aid, contouring for radiotherapy and internal dosimetry computation, using molecular imaging modalities.

PLANET Onco Dose is a modular software suite composed of three elements:

  • PLANET: Core System: license controller, reviewing of multi-modal molecular image series -(PET/CT, SPECT/CT, PET/MRI, SPECT/MRI): fusion and registration;
  • -PLANET Onco: Oncology Module: contouring of region of interest, turnor segmentation, tumoral activity monitoring, therapy response assessment;
  • -PLANET Dose: Dosimetry Module: internal dosimetry computation for the Targeted Radionuclide Therapy (TRT).
AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical imaging software called PLANET Onco Dose (K182966). It aims to demonstrate substantial equivalence to a predicate device (Velocity K173636).

However, the document does not contain the specific details about acceptance criteria, the study design (e.g., sample size, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance), or the training set information (sample size, ground truth establishment) typically expected for a performance study proving a device meets acceptance criteria.

The "Performance Testing - Bench" section (Item 8) is very high-level and states: "PLANET Onco Dose was submitted to performance, functional and algorithmic testing, risk management assessment and validation activities under clinically representative conditions. The results of performance, functional and algorithmic testing, risk management assessment and validation activities under clinically representative conditions demonstrate the safety and effectiveness of PLANET Onco Dose." This is a general statement of compliance, not a detailed report of a study.

Therefore, I cannot fulfill most of the requested information based on the provided text.

Here's what can be inferred or stated about the acceptance criteria and the "study" based on the very limited information provided:

1. A table of acceptance criteria and the reported device performance:

The document does not explicitly state quantitative acceptance criteria or detailed reported device performance metrics. Instead, it relies on demonstrating substantial equivalence to a predicate device, focusing on similar technological characteristics and functionalities. The comparison table (pages 6-7) lists features and functionalities of PLANET Onco Dose and its predicate, implying that meeting the functionalities of the predicate is a key "performance" aspect.

Since no specific quantitative acceptance criteria or detailed performance data are provided, a table cannot be constructed. The overall "performance" claim is that the device "meets the requirements of the device, its user needs and intended use, which are demonstrated to be substantially equivalent to those of the predicate device."

2. Sample size used for the test set and the data provenance:

  • Sample size: Not specified.
  • Data provenance: Not specified. The document mentions "clinically representative conditions" for testing, but no details on the origin (e.g., country) or nature (retrospective/prospective) of the data are provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of experts: Not specified.
  • Qualifications of experts: Not specified.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Adjudication method: Not specified.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC study: Not mentioned or implied. The focus is on demonstrating substantial equivalence of the software's capabilities, not on human-in-the-loop performance improvement. The software is described as a "standalone software intended to be used with PET or SPECT hybrid imaging systems in order to manage, process, display and analyze nuclear medical medical image series, to assist in medical diagnosis, to assist in treatment analysis and in therapy response assessment, to assist in the contouring of region of interest for radiotherapy." It assists professionals but there's no mention of a study on how it improves human reader performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • The document states "PLANET Onco Dose is a standalone software". While it doesn't provide specific isolated algorithm performance metrics (e.g., sensitivity, specificity for a specific task), the "Performance Testing - Bench" section implies that the software's functional and algorithmic performance was evaluated independently to demonstrate its capabilities. However, specific metric-based results are not provided.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Type of ground truth: Not specified. The nature of the "performance, functional and algorithmic testing" and "validation activities" would typically involve some form of ground truth for evaluation, but the document does not elaborate on how this ground truth was established or what it comprised.

8. The sample size for the training set:

  • Training set sample size: Not specified. The document describes a premarket notification for a software, but does not indicate whether it is an AI/ML model that requires a discrete "training set." If it is, no details are provided.

9. How the ground truth for the training set was established:

  • Ground truth establishment: Not specified. (See point 8).

In summary, the provided FDA 510(k) clearance letter and summary primarily focus on establishing substantial equivalence based on intended use and technological characteristics compared to a predicate device. It lacks the detailed reporting of performance studies, including acceptance criteria, specific study design parameters (sample sizes, data provenance, expert involvement for ground truth, adjudication), and results, that would be present in a comprehensive study report or a different type of regulatory submission (e.g., De Novo or PMA for novel devices with specific performance claims).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.

March 6, 2019

DOSIsoft % Mr. Luc Diot Quality Director 45/47, Avenue Carnot Cachan, 94230 FRANCE

Re: K182966

Trade/Device Name: PLANET Onco Dose Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: February 8, 2019 Received: February 21, 2019

Dear Mr. Diot:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D.'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182966

Device Name PLANET Onco Dose

Indications for Use (Describe)

PLANET Onco Dose is a standalone software intended to be used with PET or SPECT hybrid imaging systems in order to manage, process, display and analyze nuclear medical medical image series, to assist in medical diagnosis, to assist in treatment analysis and in therapy response assessment, to assist in the contouring of region of interest for radiotherapy.

PLANET Onco Dose is dedicated to be used by qualified medical professionals in Molecular Imaging, and/or Medical Oncology.

The medical modalities of these medical imaging systems include CT. MRI, SPECT, PET. XA, RT Struct and RT Dose as supported by ACR/NEMA DICOM 3 standard format.

PLANET Onco Dose provides the user with the means to segment structures in medical image volumes by providing dedicated delineation, contouring and propagation tools for both tumors and normal tissues (i.e. Regions of Interest (ROI)).

PLANET Onco Dose provides tools to display, co-register (including deformable registration), compute Standardized Uptake Value (SUV) and import / export results (contours and dosimetries) to / from Treatment Planning Systems (TPS) and PACS devices for assessment and response of patients undergoing a course of oncology treatment.

PLANET Onco Dose provides the user with the means to assist in the assessment quantification of radiation doses received by tissues as a result of administering a radionuclide (e.g. Permanent Yttrium-90 microsphere implants).

PLANET Onco Dose provides tools for post-treatment absorbed dose calculation and evaluation on PET and SPECT images. The following functions are available to allow dose calculations for patients after they have received a treatment using permanent Yttrium-90 (Y90) microspheres:

  • 3D liver-lung shunt assessment;

  • Local Deposition Model;

  • Voxel S Value approach based on the schema in MIRD Pamphlet 17 [1]:

  • Dosimetry based on 90Y-microspheres-PET (or SPECT Bremsstrahlung) series;

  • Compatible with PET images acquired with another radioisotope instead of Y90 when Y90 acquisitions are not supported by the scanner (correction of branching ratio and decay parameters).

For Y90 microspheres. PLANET Onco Dose cannot be used to prescribe the radionuclide activity to be administered to the patient for the therapy. The user has to provide the parameters (e.g. activity) in order for PLANET Onco Dose to estimate the radiation doses that the tissues received as a result of the administration.

PLANET Onco Dose should only be used for the retrospective determination of dose and not for the case where is a need for retreatment using Y90 microspheres.

[1] Bolch WE et al. MIRD pamphlet no. 17: the dosimetry of nonuniform activity distributions-Radionuclide S values at the voxel level. J Nucl Med, 1999, 40:11S-36S

{3}------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "DOSIsoft" with a butterfly image in between "DOSI" and "soft". The word "DOSI" is in a bold, serif font, and the word "soft" is in a thinner, sans-serif font. The butterfly image is blue and has red arrows pointing to it from the words on either side.

Image /page/4/Picture/1 description: The image is a logo with a blue circle and the letters "iSO" in the center. The word "CERTIFIED" is written in blue around the top of the circle. There are two red arrows on either side of the circle, pointing towards the center. The number "73485" is written in blue around the bottom of the circle.

Premarket Notification 510(k) Summary [as required by 21 CFR 807.92]

510(k) Summary

K182966

    1. Date the Summary was prepared: 2018-10-17

2. Submitter:

Company Name:DOSIsoft SA
Address:45-47, Avenue Carnot94230 CACHAN - France
Phone No.:+33 1 41 24 26 26
Fax No.:+33 1 41 24 26 28
Contact Name:Mr. Marc USZYNSKI

3. Device Information:

Proprietary Name:PLANET Onco Dose
Trade Names:PLANET Onco, PLANET Dose, PLANET Onco Dose
Common Name:Medical Imaging Software
Classification Name:System, Image Processing, Radiological(21 CFR 892.2050, Class II, Product Code LLZ)

4. Predicate Device:

K173636 - Velocity - Varian Medical Systems, Inc.

Executive Summary

5. Product Description:

PLANET Onco Dose is a software platform dedicated to medical diagnosis aid, therapy response assessment aid, contouring for radiotherapy and internal dosimetry computation, using molecular imaging modalities.

PLANET Onco Dose is a modular software suite composed of three elements:

  • PLANET: Core System: license controller, reviewing of multi-modal molecular image series -(PET/CT, SPECT/CT, PET/MRI, SPECT/MRI): fusion and registration;
  • -PLANET Onco: Oncology Module: contouring of region of interest, turnor segmentation, tumoral activity monitoring, therapy response assessment;
  • -PLANET Dose: Dosimetry Module: internal dosimetry computation for the Targeted Radionuclide Therapy (TRT).

6. Intended Use and User profiles:

PLANET Onco Dose is a standalone software intended to be used with PET or SPECT hybrid imaging systems in order to manage, process, display and analyze nuclear medicine and other digital mage series, to assist in medical diagnosis, to assist in treatment analysis and in therapy response assessment, to assist in the contouring of region of interest for radiotherapy.

PLANET Onco Dose is dedicated to be used by qualified medical professionals in Molecular Imaging, and/or Medical Oncology.

The medical modalities of these medical imaging systems include CT, MRI, SPECT, PET, XA, RT Struct and RT Dose as supported by ACR/NEMA DICOM 3 standard format.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "DOSIsoft" in a stylized font. The words "DOSI" and "soft" are in black, while the "I" is replaced with a blue butterfly-like shape. The text is horizontally aligned and the font is sans-serif.

Image /page/5/Picture/1 description: The image is a blue circular logo with the text "CERTIFIED" at the top. Inside the circle is the text "iSO" in a sans-serif font. Below the text "iSO" is the number "73485" with a bullet point on either side. There are two red arrows on either side of the text "iSO".

PLANET Onco Dose provides the user with the means to segment structures in medical image volumes by providing dedicated delineation, contouring and propagation tools for both tumors and normal tissues (i.e. Regions of Interest (ROI)).

PLANET Onco Dose provides to display, co-register (including deformable registration), compute Standardized Uptake Value (SUV) and import / export results (contours and dosimetries) to / from Treatment Planning Systems (TPS) and PACS devices for assessment and response of patients undergoing a course of oncology treatment.

PLANET Onco Dose provides the user with the means to assessment quantification and verification of radiation doses received by tissues as a result of administering a radionuclide (e.g. Permanent Yttrium-90 microsphere implants).

PLANET Onco Dose provides tools for post-treatment absorbed dose calculation and evaluation on PET and SPECT images. The following functions are available to allow dose calculations for patients after they have received a treatment using permanent Yttrium-90 (Y90) microspheres:

  • ー 3D liver-lung shunt assessment;
  • । Local Deposition Model;
  • Voxel S Value approach based on the schema in MIRD Pamphlet 17 [1]; ।
  • Dosimetry based on 90Y-microspheres-PET (or SPECT Bremsstrahlung) series; ।
  • Compatible with PET images acquired with another radioisotope instead of Y90 when Y90 acquisitions are not supported by the scanner (correction of branching ratio and decay parameters).

For Y90 microspheres, PLANET Onco Dose cannot be used to prescribe the radionuclide activity to be administered to the patient for the therapy. The user has to provide the radionuclide parameters (e.g. activity) in order for PLANET Onco Dose to estimate the radiation doses that the tissues received as a result of the administration.

PLANET Onco Dose should only be used for the retrospective determination of dose and not for the case where there is a need for retreatment using Y90 microspheres.

[1] Bolch WE et al. MIRD pamphlet no. 17: the dosimetry of nonuniform activity distributions-Radionuclide S values at the voxel level. J Nucl Med, 1999, 40:11S-36S

7. Technological Characteristics:

4230 Cachan - France

The table below compares the technological characteristics and the features of PLANET Onco Dose and its predicate Velocity K173636.

PLANET Onco Dose (This submission)Velocity K173636 (Predicate)
Indications for Use / Intended Use
PLANET Onco Dose is a standalone softwareintended to be used with PET or SPECT hybridimaging systems in order to manage, process,display and analyze nuclear medicine and otherdigital medical image series, to assist in medicaldiagnosis, to assist in treatment analysis and intherapy response assessment, to assist in thecontouring of region of interest for radiotherapy.PLANET Onco Dose is dedicated to be used byqualified medical professionals in Molecular Imaging,and/or Medical Oncology.Velocity is a software package that provides thephysicians a means for comparison of medical dataincluding imaging data that is DICOM compliant.It allows the display, annotation, volume operation,volume rendering, registration and fusion of medicalimages as an aid during use by diagnostic radiology,oncology, radiation therapy planning and othermedical specialties.Velocity is not intended for mammography.
Device Type
Standalone softwareStandalone software
Users
Qualified medical professionals in MolecularImaging, and/or Medical OncologyPhysicians
Cybersecurity
DOSISOFT SA45/47, avenue Carnot+33 (0)1 41 24 26 26info@dosisoft.comPage 2/5
PLANET Onco Dose (This submission)Velocity K173636 (Predicate)
Limits software access to trusted users onlyLimits software access to trusted users only
WorkflowWorkflow
Oncology workflow automationOncology workflow automation
Receive, store and retrieve digital DICOM medicalimages from a variety of sources and modalitiesPET/SPECT/CT/MRIReceive, store and retrieve digital DICOM medicalimages from a variety of sources and modalitiesPET/SPECT/CT/MRI
Digital medical images patient's structured databaseOrganize data into a logical overview of a patient'simaging history
Image and registered image data export as DICOMseriesImage and registered image data export as DICOMseries
Structure Import, Save & Export as DICOM RTObjectStructure Import, Save & Export as DICOM RTObject
Dose data Import, Save & Export as DICOM RTObjectDose data Import, Save & Export as DICOM RTObject
Reports can be exported as DICOM filesReports can be exported as print sessions
Image Management ToolsImage Management Tools
Support of 4D seriesSupport of 4D series with breathing cine mode
Advanced visualization and navigation featuresAdvanced visualization and navigation features
Management of multi-time point imageseriesManagement of multi-time point imageseries
Manual registration methods including landmarksManual registration method
Automatic rigid image registration and fusion ofmedical series from multiple modalitiesAutomatic rigid image registration and fusion ofmedical series from multiple modalities
Automatic deformable image registration and fusionof medical images from multiple modalitiesAutomatic deformable image registration and fusionof medical images from multiple modalities
Combines images from different modalities forenhanced volume definitionCombines images from different modalities forenhanced volume definition
Visualization ToolsVisualization Tools
General image viewer with view layout selection andtoolbarsGeneral image viewer with view layout selection andtoolbars
2D visualization (Multi-planar)2D visualization (Multi-planar)
3D volume rendering3D volume rendering
Image fusionAnatomical and functional seriesImage fusionAnatomical and functional series
Image analysis toolsImage analysis tools
General Contouring FeaturesGeneral Contouring Features
Multi-modality manual contouring toolsMulti-modality manual contouring tools
Contour transformation toolsContour transformation tools
Semi-automatic advanced contouring toolsContouring with automated scripting
Automated structure of interest segmentation toolsAtlas auto-segmentation
Propagates contours from one time point to anotherPropagates contours from one time point to another
Contour transformation (interpolation,margining) between modalities-
Functional Segmentation Tools-
PLANET Onco Dose (This submission)Velocity K173636 (Predicate)
Assessment Aid and Patient Follow-up Tools
Visualization and analytical monitoring of diseaseprogression or response to treatment or therapyusing multiexam comparison and correlationVisualization and analytical monitoring of diseaseprogression or response to treatment or therapyusing multiexam comparison and correlation
Quantification of metabolic activity and volumemeasurement of segmented tumors on PET modalityVolume operationsBiological effective dose (BED) scaling
Measure Standard Uptake Value - SUVMeasure Standard Uptake Value - SUV
Advanced Tumor SUV Statistics-
Patient's follow up analysis with volumetric graphsPatient's follow up analysis with volumetric graphs
Allows generation of RECIST 1.1 & PERCIST 1.0format reports.-
Dose Distribution Import and Estimation Tools
The dose distribution can be imported from anysystem such as radiation therapy planning systemsusing DICOM RT Dose formatThe dose distribution can be imported from anysystem such as radiation therapy planning systemsusing DICOM RT Dose format
Intended for post-treatment absorbed dosecalculation and evaluationIntended for post-treatment absorbed dosecalculation and evaluation
Compatible with Y90 PET and SPECT image typesCompatible with Y90 PET and SPECT image types
Use the Local Deposition Model for dose calculationUse the Local Deposition Model for dose calculation
Use the Voxel S Value approach based on theschema in MIRD Pamphlet 17 ([1])-
Dosimetry Analysis Tools
Display of the dose distribution in patient tissuesDisplay of the dose distribution in patient tissues
Dose Volume Histogram statistics displayDose Volume Histogram statistics display
Table of dose to organsTable of dose to organs
Simultaneous display of two dose distributions andcomparison tools-
Addition of multiple treatment stages: internaldosimetry and/or external beam radiotherapyAddition of multiple treatment stages: internaldosimetry and/or external beam radiotherapy
Operating System
Linux operating systemMicrosoft Windows and Apple macOS operatingsystems

www.dosisoft.com

RCS Créteil 443 195 433

+33 (0)1 41 24 26 28 |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "DOSI" in bold, black font, followed by a blue butterfly outline, and the word "soft" in a thin, black font. The butterfly is positioned between the two words, creating a visual break. The overall design is simple and clean, with a focus on the text and the butterfly symbol.

Image /page/6/Picture/1 description: The image is a logo for ISO certification. The logo is a blue circle with the letters "iSO" in white in the center. There are two red arrows on either side of the letters. The number "73485" is printed in black at the bottom of the circle. The word "CERTIFIED" is printed in white at the top of the circle.

DOSISOFT SA 45/47, avenue Carnot 94230 Cachan - France

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "DOSI" in bold, black font, followed by a blue butterfly outline. The word "soft" is written in a thin, black font to the right of the butterfly. The image appears to be a logo or brand name.

Image /page/7/Picture/1 description: The image is a logo with the letters "iSO" in a white circle with a light blue background. The word "CERTIFIED" is written in blue letters above the circle, and the number "73485" is written in blue letters below the circle. There are two small red arrows on either side of the circle, pointing towards the letters "iSO".

8. Performance Testing - Bench:

PLANET Onco Dose was designed and documented in accordance with the recommandations of FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" for software devices identified as Major Level of Concern.

PLANET Onco Dose was submitted to performance, functional and algorithmic testing, risk management assessment and validation activities under clinically representative conditions. The results of performance, functional and algorithmic testing, risk management assessment and validation activities under clinically representative conditions demonstrate the safety and effectiveness of PLANET Onco Dose.

PLANET Onco Dose meets the requirements of the device, its user needs and intended use, which are demonstrated to be substantially equivalent to those of the predicate device.

DOSISOFT SA 45/47, avenue Carnot 94230 Cachan - France

+33 (0)1 41 24 26 26 |
+33 (0)1 41 24 26 28 |

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "DOSI" in a bold, serif font, followed by a blue butterfly outline, and the word "soft" in a thin, sans-serif font. The butterfly is positioned between the two words, creating a logo-like appearance. The overall design is clean and modern.

Image /page/8/Picture/1 description: The image is a circular logo with the text "CERTIFIED" at the top and ".73485." at the bottom. The letters "iSO" are in the center of the circle in a sans-serif font. There are two arrows on either side of the letters "iSO" pointing towards the center. The logo is light blue and white.

9. Conclusions

The substantial equivalence discussion sustains the claim that PLANET Onco Dose intended use, clinical and technical (principles of operation, functionalities and critical performances) characteristics are the same as for the predicate device Velocity legally marketed in the U.S. (K173636).

The differences on some functional features do not significantly affect, or may not significantly affect, safety or effectiveness.

The results of verification and validation activities demonstrate the safety and effectiveness of PLANET Onco Dose.

PLANET Onco Dose meets the requirements of the device, its user needs and intended use, which are demonstrated to be substantially equivalent to those of the predicate device.

In summary, DOSIsoft therefore considers that, in its opinion, PLANET Onco Dose is substantially equivalent to and is at least as safe and effective as the predicate device.

DOSISOFT SA 45/47, avenue Carnot 94230 Cachan - France

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).