K Number
K182965
Manufacturer
Date Cleared
2019-03-29

(155 days)

Product Code
Regulation Number
886.4300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BL-Cart™ IOL Delivery Cartridge is a single-use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through a surgical procedure. The cartridge is intended to be used in conjunction with Alcon Monarch® III IOL Delivery System injector. The BL-Cart™ IOL Delivery Cartidge is only for the insertion of Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the approved IOL labeling.

Device Description

BL-Cart™ IOL Delivery Cartridge (Model: Type D) is an intraocular lens (IOL) delivery cartridge solution capable of supplying the ophthalmologist surgeon with a cartridge pre-coated with lubricious coating for use during a typical cataract surgery. The BL-Cart™ cartridge is used to deliver an approved IOL lens. The BL-Cart™ IOL Delivery Cartridge and referenced predicate device, Monarch® III D cartridge, utilize the same lens folding system design in which IOLs are folded and/or delivered from an injector system and the resulting optical and physical properties of the IOL remain unchanged as a result of folding/delivery. The BL-Cart™ IOL Delivery Cartridge comes with one pre-coated cartridge, packaged in a Tyvek pouch, placed in a 10 unit box as intended for marketing, and sterilized by ethylene oxide (EtO). The BL-Cart™ IOL Delivery Cartridge and the referenced legally marketed predicate device share the same principal method to implant an intraocular lens into the eye.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for the BL-Cart™ IOL Delivery Cartridge. This device is classified as a Class I medical device (Intraocular Lens Guide). Class I devices generally have lower regulatory burden and often rely on substantial equivalence to a predicate device rather than extensive clinical studies.

Here's an analysis of the acceptance criteria and supporting study information:

Acceptance Criteria and Reported Device Performance

Given the nature of a Class I device and the 510(k) submission, explicit "acceptance criteria" for performance metrics like sensitivity or specificity are not typically defined or required in the same way as for higher-risk devices or software. Instead, acceptance is based on demonstrating substantial equivalence to a legally marketed predicate device. This involves showing that the new device is as safe and effective as the predicate, often through shared technological characteristics, intended use, and performance testing that confirms it meets established standards or does not raise new safety/effectiveness concerns.

For this device, the primary performance goals are related to its ability to deliver an IOL without damage to the IOL or the cartridge, and to maintain the IOL's properties.

Acceptance Criterion (Implicit)Reported Device Performance (as demonstrated by non-clinical tests)
BiocompatibilityPassed cytotoxicity, sensitization, ocular irritation, acute systemic toxicity, and material-mediated pyrogenicity tests in accordance with ISO 10993-5, 10, and 11.
IOL Delivery Success (without affecting IOL functionality)All lenses were successfully delivered through the BL-Cart™ IOL Delivery Cartridge according to loading and delivery instructions.
IOL Integrity Post-Delivery (optical, dimensional, surface)IOLs evaluated for optical properties, dimensional properties, and overall surface/bulk homogeneity before and after manipulation using the BL-Cart™ showed passing results and were within dimensional specifications. Lenses observed for possible damages or scratches using a microscope showed passing results.
Cartridge Integrity Post-Delivery (surface, bulk homogeneity)BL-Cart™ showed no damage after lens delivery.
IOL Opening TimeEvaluated this property after folding. (Specific results not given, but implied to be acceptable as part of overall positive conclusion).
Sterilization (SAL 1x10-6)Achieved with Ethylene Oxide (EtO) sterilization.
Shelf-LifeTesting was performed (specific results not detailed, but implied to be acceptable).

Study Information

1. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: The document does not explicitly state a numerical "sample size" for the non-clinical tests in terms of how many IOLs were tested or how many cartridges were used. It states, "All lenses were delivered through the BL-Cart™ IOL Delivery Cartridge," and that testing involved "Lenstec's hydrophilic Softec 1 IOLs of low to high diopters and IOL models validated for use with this device." This implies a sufficient number of tests to cover the range of intended use.
  • Data Provenance: The data is from non-clinical (bench) testing performed by the manufacturer, AST Products, Inc. There is no mention of country of origin for the data beyond being conducted by the submitting company in the USA. The data is prospective in the sense that it was generated specifically for this submission to demonstrate substantial equivalence.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This device is an IOL delivery cartridge, not a diagnostic or AI-driven device requiring expert interpretation or ground truth establishment in a clinical sense. The "ground truth" for its performance is determined by established engineering and material science standards (e.g., ISO 11979-3 for IOL properties, ISO 10993 for biocompatibility) and direct physical observation/measurement (e.g., micro-inspection for damage, dimensional checks). No human experts are described as establishing "ground truth" for the test set observations.

3. Adjudication method for the test set:

  • Not applicable. As there are no human-interpreted "cases" or "ground truth" to adjudicate, no adjudication method is mentioned or required. Performance is determined by objective measurements and adherence to standards.

4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. An MRMC study is relevant for diagnostic AI devices where multiple readers interpret cases with and without AI assistance. This device is a manual surgical tool (cartridge) and does not involve AI or human readers for diagnostic purposes.

5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is purely mechanical/physical and does not involve any algorithms or AI components.

6. The type of ground truth used:

  • The "ground truth" for this device's performance is based on International Standards (e.g., ISO 11979-3 for IOL mechanical properties, ISO 10993 for biocompatibility), engineering specifications (dimensional, optical properties of IOLs post-delivery), and physical observation (absence of damage to IOL or cartridge). The performance is assessed against these objective benchmarks rather than a clinical outcome or expert consensus on clinical findings.

7. The sample size for the training set:

  • Not applicable. As this device does not involve AI or machine learning algorithms, there is no "training set."

8. How the ground truth for the training set was established:

  • Not applicable. No training set exists for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

March 29, 2019

AST Products, Inc. % Dr. David Lim VP. R&D and Regulatory Affairs 9 Linnell Circle Billerica. MA 01821

Re: K182965

Trade/Device Name: BL-Cart™ IOL Delivery Cartridge Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I Product Code: MSS Dated: February 14, 2019 Received: February 19, 2019

Dear Dr. Lim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours.

Jennifer N. Brown -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182965

Device Name BL-Cart™ IOL Delivery Cartridge

Indications for Use (Describe)

The BL-Cart™ IOL Delivery Cartridge is a single-use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through a surgical procedure. The cartridge is intended to be used in conjunction with Alcon Monarch® III IOL Delivery System injector.

The BL-Cart™ IOL Delivery Cartidge is only for the insertion of Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the approved IOL labeling.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

As required by the Safe Medical Devices Act (SMDA) of 1990 and in accordance with 21 CFR § 807.92, 510(k) summary is provided.

1. Submitter

AST Products, Inc. 9 Linnell Circle Billerica, MA 01821 USA Tel: (978) 667-4500 Fax: (978) 667-9778

2. Official Correspondence/Contact Person

William Lee. Ph.D. VP, R&D and Regulatory Affairs AST Products, Inc. 9 Linnell Circle Billerica, MA 01821 USA Tel: (978) 667-4500 Fax: (978) 667-9778 Email: wlee@astp.com

3. 510(k) Preparer

Regulatory Doctor™ RCTC, LLC 3955 Riner Road, Box 148 Riner, VA 24149 Tel: (800) 321-8567 Email: info@regulatorydoctor.com

4. Product Name

  • Trade Name: BL-Cart™ IOL Delivery Cartridge (Model: Type D)
  • Common Name: Folders and Injectors, Intraocular Lens (IOL) |

5. Device Classification

  • Classification Name: Class I (21 CFR § 886.4300 Intraocular lens guide)
  • Classification Panel: Ophthalmic (86)
  • Product Code: MSS

6. Legally Marketed Predicates

  • K063155: Monarch® III IOL Delivery System (D Cartridge)

7. Device Description

BL-Cart™ IOL Delivery Cartridge (Model: Type D) is an intraocular lens (IOL) delivery

{4}------------------------------------------------

cartridge solution capable of supplying the ophthalmologist surgeon with a cartridge precoated with lubricious coating for use during a typical cataract surgery. The BL-Cart™ cartridge is used to deliver an approved IOL lens.

The BL-Cart™ IOL Delivery Cartridge and referenced predicate device. Monarch® III D cartridge, utilize the same lens folding system design in which IOLs are folded and/or delivered from an injector system and the resulting optical and physical properties of the IOL remain unchanged as a result of folding/delivery.

The BL-Cart™ IOL Delivery Cartridge comes with one pre-coated cartridge, packaged in a Tyvek pouch, placed in a 10 unit box as intended for marketing, and sterilized by ethylene oxide (EtO).

The BL-Cart™ IOL Delivery Cartridge and the referenced legally marketed predicate device share the same principal method to implant an intraocular lens into the eve.

Note: LubriMATRIX™ is a proprietary treatment composed of a hydrophilic polymer and a lubricious polymer engrafted onto the surface of the cartridge via a chemical polymerization process. Refer to MAF-1963 for further information. LubriMATRIX™ is applied to AST Products' current line of IOL Delivery Systems, lioli™ IOL Delivery System (K142056) and pioli™ IOL Delivery System (K172228).

8. Intended Use and Indications for Use

The BL-Cart™ IOL Delivery Cartridge is a single use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through a surgical procedure. The cartridge is intended to be used in conjunction with Alcon Monarch® III IOL Delivery System injector.

The BL-Cart™ IOL Delivery Cartridge is only for the insertion of Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the IOL approved labeling.

The BL-Cart™ IOL Delivery Cartridge and the referenced legally marketed predicate device share the same principal folding method to implant an intraocular lens into the eye and similar intended use and indications for use for lens insertion into the eye, and thus does not raise any new questions or concerns regarding safety and effectiveness. Refer to Table 1.

{5}------------------------------------------------

TradeNameManufacturerRegulationNo.ProductCodeRegulationDescription510(k) No.Intended Use &Indications for Use
BL-Cart™IOL DeliveryCartridge(Model:Type D)AST Products,Inc.886.4300MSSIntraocularLens GuideNew deviceThe BL-Cart™ IOL Delivery Cartridgeis a single use, sterile device intendedto insert an approved single-piecefoldable acrylic intraocular lens (IOL)into the human eye through a surgicalprocedure. The cartridge is intendedto be used in conjunction with AlconMonarch® III IOL Delivery Systeminjector.The BL-Cart™ IOL Delivery Cartridgeis only for the insertion of LenstecSoftec I IOL and IOL models validatedfor use with this device as indicated inthe IOL approved labeling.
Monarch® IIIIOL DeliverySystem (DCartridge)Alcon Research,Ltd.886.4300MSSIntraocularLens GuideK063155The MONARCH® III IOL DeliverySystem is used for implantation ofALCON® qualified ACRYSOF®intraocular lenses into the eyefollowing cataract removal.

Table 1. Device Comparisons – Intended Use and Indications for Use

{6}------------------------------------------------

9. Technological Characteristics

The BL-Cart™ IOL Delivery Cartridge is an intraocular lens delivery cartridge used for progressive folding in combination with a reusable handpiece to deliver an intraocular lens into the eye for replacement of the human crystalline lens during cataract surgery. The device consists of a lubricious coated cartridge in preparation for intraocular lens loading and lens folding.

The BL-Cart™ IOL Delivery Cartridge shares similar technological characteristics for lens insertion into the eye with the predicate device referenced in this 510(k) premarket notification application. Refer to Tables 2 and 3.

{7}------------------------------------------------

TradeNameDesignOperating PrincipleBiocomp.SterilizationMethodSettingUsedPatientContactRecommendedfor Use WithHumanFactorsPerformanceCriteria Met?
BL-Cart™IOLDeliveryCartridge(Model:Type D)A sterile,single- usedisposablecoatedcartridgeThe one-piecedisposable cartridgeis loaded into areusable handpieceinjection system andthe IOL is pushedthrough the cartridgeinto the eyeYesETO sterilizationSAL level 1x10-6Hospital/SurgerycenterDistal endof cartridgeto placelens in eyeFDA-approvedviscoelasticFor use bydoctorsonlyYes
Monarch®III IOLDeliverySystem (DCartridge)Monarch® IIIreusablehandpiece andsingle-use,sterile coatedcartridgeThe one-piecedisposable cartridgeis loaded into areusable handpieceinjection system andthe IOL is pushedthrough the cartridgeinto the eyeYesETO sterilizationHospital/SurgerycenterDistal endof cartridgeto placelens in eyeFDA-approvedviscoelasticFor use bydoctorsonlyYes

Table 2. Device Comparisons – Technological Characteristics

Table 3. Device Comparisons – Materials

Trade NameDevice MaterialsStandards Met
BL-Cart™ IOL Delivery Cartridge (Model: Type D)Cartridge - Polypropylene (RJ880MO-12)ISO 11979-3:2012ISO10993-5: 2009ISO 10993-10: 2010
Monarch® III IOL Delivery System (D Cartridge)Cartridge- Polypropylene with a polyvinylpyrrolidone (PVP) coating on the inner lumenISO 11979-3:2012ISO10993-5: 2009ISO 10993-10: 2010

{8}------------------------------------------------

10. Performance Data (Nonclinical and/or Clinical)

  • Non-clinical Tests:
    The BL-Cart™ IOL Delivery Cartridge contacts the patient's eye directly for a limited (≤24hr) time. The device also has limited contact with the IOL. Based on the contact duration, cytotoxicity, sensitization and ocular irritation tests were performed. In addition, acute systemic toxicity and material-mediated pyrogenicity were also performed. The biocompatibility testing was performed in accordance with International Standard Organization (ISO) Biocompatibility evaluation of medical devices-parts 5. 10 and 11. The biocompatibility testing was found acceptable.

All lenses were delivered through the BL-Cart™ IOL Delivery Cartridge according to the loading and delivery instructions.

The IOLs were evaluated for optical properties, dimensional properties and overall surface and bulk homogeneity before and after being surgically manipulated using the BL-Cart™ IOL Delivery Cartridge, as well as lens opening time after folding. The BL-Cart™ was also evaluated for its cartridge performance, such as overall cartridge surface and bulk homogeneity.

IOL optical properties and overall surface and bulk homogeneity inspection were conducted in accordance with ISO 11979-3:2012, Ophthalmic implants – Intraocular lenses - Part 3: Mechanical properties and test methods.

After delivery, lenses were observed for possible damages or scratches using a microscope and showed passing results, and were within dimensional specifications. BL-Cart™ also showed no damage after lens delivery.

The resulting data from simulated surgical manipulation of BL-Cart™ IOL Delivery Cartridge (Model: Type D) to deliver intraocular lenses showed that the device can successfully deliver Lenstec's hydrophilic Softec 1 IOLs of low to high diopters and IOL models validated for use with this device as indicated in the IOL approved labeling without affecting the functionality of the lens.

  • Clinical Tests: Not required for this device.

11. Conclusion

The resulting data from simulated surgical manipulation of BL-Cart™ IOL Delivery Cartridge to deliver intraocular lenses showed that BL-Cart™ (Model: Type D) can successfully deliver Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the IOL approved labeling without affecting the functionality of the lens.

Based on the assessment of these findings, along with the results of Biocompatibility, Sterilization and Shelf-Life testing, we conclude that the BL-Cart™ IOL Delivery Cartridge is substantially equivalent to the referenced legally marketed predicate device provided in this 510(k) premarket notification submission.

§ 886.4300 Intraocular lens guide.

(a)
Identification. An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.(b)
Classification. Class I (general controls). Except when used as folders or injectors for soft or foldable intraocular lenses, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 886.9.