(113 days)
Not Found
No
The summary describes a standard blood pressure monitoring system and does not mention any AI or ML capabilities.
No.
The device is a blood pressure monitoring system intended for measurement and monitoring, not for treating a condition or disease.
Yes
The device is intended to measure systolic and diastolic blood pressure and pulse rate, which are physiological parameters used to assess a person's health status and can indicate the presence of certain medical conditions, thus serving a diagnostic purpose.
No
The device description and performance studies clearly indicate the device includes hardware components (inflatable cuff, the device itself for operation) and underwent hardware-specific testing (electrical safety, EMC, shock and vibration, biocompatibility).
Based on the provided information, the FORA P100 Blood Pressure Monitoring System is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to measure blood pressure and pulse rate using a non-invasive technique on the upper arm. This is a physiological measurement taken directly from the body, not a test performed on a sample (like blood, urine, or tissue) outside of the body.
- Device Description: The description focuses on the device's function in assisting with blood pressure testing, recording, tracking, and monitoring.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples, reagents, or any of the typical components or processes associated with in vitro diagnostics.
IVD devices are specifically designed to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. The FORA P100 measures a physiological parameter directly from the patient.
N/A
Intended Use / Indications for Use
The FORA P100 Blood Pressure Monitoring System is intended to measure the systolic and diastolic blood pressure and pulse rate by using a non-invasive technique in which an inflatable cuff is wrapped on the upper arm. Do not use this system on babies, young children or persons who cannot express their consent.
Product codes
DXN
Device Description
FORA P100 Blood Pressure Monitoring System (Model: P100) can be operated by the device itself to assist in blood pressure testing, recording, tracking and monitoring in easy.
The TD-3130 in report is an internal model and FORA P100 is the marketing name for this system.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
upper arm
Indicated Patient Age Range
Do not use this system on babies, young children or persons who cannot express their consent.
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
The FORA P100 Blood Pressure Monitoring System was validated by the tests according to IEC 80601-2-30:2009 and met the requirements of ANSI/AAMI/ISO 81060-2:2009.
Software validation was performed to verify and validate the system works functionally.
Testing performed included electrical safety, EMC, and shock and vibration test. The proposed device met the requirements of IEC/EN 60601-1, IEC/EN 60601-1-2 and IEC/EN 60601-1-11.
Biocompatibility testing was also performed. The materials of cuff met the requirements of ISO 10993-5 and 10993-10.
The FORA P100 Blood Pressure Monitoring System has the same performance characteristics as the predicate device. Software verification and validation, and design validation confirmed that the performance, safety and effectiveness of the FORA P100 Blood Pressure Monitoring System are substantial equivalent to the predicate device.
Key Metrics
Not Found
Predicate Device(s)
FORA CARE BLOOD PRESSURE MONITORING SYSTEM K133588
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 12, 2019
TaiDoc Technology Corporation Sophia Wu Regulatory Affairs Specialist 6F, No. 127, Wugong 2nd Rd., Wugu District New Taipei City, 24888 TAIWAN
Re: K182934
Trade/Device Name: FORA P100 Blood Pressure Monitoring System Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: October 19, 2018 Received: October 22, 2018
Dear Sophia Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S5
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K182934
Device Name
FORA P100 Blood Pressure Monitoring System
Indications for Use (Describe)
The FORA P100 Blood Pressure Monitoring System is intended to measure the systolic and diastolic blood pressure and pulse rate by using a non-invasive technique in which an inflatable cuff is wrapped on the upper arm. Do not use this system on babies, young children or persons who cannot express their consent.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
K182934
Image /page/3/Picture/1 description: The image shows the logo for TaiDoc. The logo consists of the word "TaiDoc" in gray letters above three red hearts. Below the hearts is the website address "www.taidoc.com" in white letters on a gray background.
泰博科技股份有限公司 TaiDoc Technology Corp.
新北市24888五股區五工二路127號6樓 6F., No.127, Wugong 2nd Rd., W New Tainei City 7488
Tel : +886-2-6625-8188 Fax : +886-2-6625-0288
Section9 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The Assigned 510(k) number is: K182934
1. Submitter Information:
Company Name | TaiDoc Technology Corporation |
---|---|
Address | 6F, No. 127, Wugong 2nd Rd., Wugu District, |
New Taipei City, 24888, Taiwan | |
Establishment Registration No. | 3004145393 |
Date Prepared | October 19th, 2018 |
Contact Person | Sophia Wu |
Title | Vice President of Regulatory Affairs Department |
Phone | +886-2-6625-8188 |
Fax | +886-2-6625-0288 |
sophiawu@taidoc.com.tw / ra.cert@taidoc.com.tw |
2. Proposed Device Information:
Trade/Proprietary Name | FORA P100 Blood Pressure Monitoring System |
---|---|
Model Number | P100 |
Common Name | Blood Pressure Monitoring System |
Product Code | DXN |
Classification Panel | 74, Cardiovascular |
Classification | Class II |
Regulation citation | 21 CFR §870.1130, Non-invasive Blood Pressure |
Measurement System |
4
Image /page/4/Picture/1 description: The image shows the logo for TaiDoc. The word "TaiDoc" is written in gray letters at the top of the image. Below the word are three red hearts. At the bottom of the image, on a gray background, is the website address "www.taidoc.com".
股份有限公司
新北市24888万股幅于
Tel : +886-2-6625-8188
Fax : +886-2-6625-0288
3. Predicate Device:
Trade/Proprietary Name | FORA CARE BLOOD PRESSURE MONITORING SYSTEM |
---|---|
Common Name | Blood Pressure Monitoring System |
Submitter | TaiDoc Technology Corporation |
510(k) Number | K133588 |
4. Intended Use
The FORA P100 Blood Pressure Monitoring System is intended to be used to measure the systolic and diastolic blood pressure and pulse rate by using a non-invasive technique in which an inflatable cuff is wrapped on the upper arm. Do not use this system on babies, young children or persons who cannot express their consent.
5. Device Description:
FORA P100 Blood Pressure Monitoring System (Model: P100) can be operated by the device itself to assist in blood pressure testing, recording, tracking and monitoring in easy.
The TD-3130 in report is an internal model and FORA P100 is the marketing name for this system.
6. Test Principle:
Blood pressure is measured non-invasively at the arm based on oscil-lometric method.
For people of common arrhythmia, such as atrial or ventricular pre-mature beats or atrial fibrillation, we recommend to use auscultatory mode. The reading obtained by single and average mode which use oscillometic method is for reference only and should be discussed with the healthcare professionals.
5
Image /page/5/Picture/1 description: The image shows the logo for TaiDoc. The logo consists of the word "TaiDoc" in gray, with three red hearts below it. Below the hearts is the website address "www.taidoc.com" in white text on a gray background.
泰博科技股份有限公司 TaiDoc Technology Corp. 新北市24888五股區五工二路127號6樓 6F., No.127, Wugong 2nd Rd., Wugu Dist. New Taipei City 24888, Taiwan
Tel : +886-2-6625-8188 Fax : +886-2-6625-0288
7. Substantial Equivalence Information:
Similarities | Modifications | |
---|---|---|
• | Operating principle | • Meter appearance |
• | Detection method / scientific technology | • Power source |
• | Employs the same test strip | • User interface |
• | Narrower intended use | • Speaking function |
Predicate device name | FORA CARE BLOOD PRESSURE MONITORING SYSTEM |
---|---|
Predicate K number | K133588 |
Comparison with | |
predicate | The modified FORA P100 Blood Pressure Monitoring System |
has the following similarities to the predicate device: | |
same intended use, same operating principle, same fundamental scientific technology, incorporate the same basic circuit design, incorporate the same materials, manufactured by the same process. | |
The modifications encompass: | |
Meter memory Speaking function Modification of the devices physical appearance Labeling change due to the above modifications |
The FORA P100 Blood Pressure Monitoring System and the predicate device both use the oscillometric method within the software algorithm to determine the systolic, diastolic blood pressure and pulse rate with an inflatable arm cuff. The major difference between the two devices is physical appearance of device. The minor changes are software.
6
Image /page/6/Picture/1 description: The image shows the logo for TaiDoc. The logo consists of the word "TaiDoc" in gray letters above three red hearts. Below the hearts is the website address "www.taidoc.com" in white letters on a gray background.
8. Performance Characteristics:
The FORA P100 Blood Pressure Monitoring System was validated by the tests according to IEC 80601-2-30:2009 and met the requirements of ANSI/AAMI/ISO 81060-2:2009.
Software validation was performed to verify and validate the system works functionally.
Testing performed included electrical safety, EMC, and shock and vibration test. The proposed device met the requirements of IEC/EN 60601-1, IEC/EN 60601-1-2 and IEC/EN 60601-1-11.
Biocompatibility testing was also performed. The materials of cuff met the requirements of ISO 10993-5 and 10993-10.
The FORA P100 Blood Pressure Monitoring System has the same performance characteristics as the predicate device. Software verification and validation, and design validation confirmed that the performance, safety and effectiveness of the FORA P100 Blood Pressure Monitoring System are substantial equivalent to the predicate device.
9. Conclusion:
Based on the information provided in this submission, the FORA P100 Blood Pressure Monitoring System is substantially equivalent to the predicate FORA CARE BLOOD PRESSURE MONITORING SYSTEM.