(51 days)
The ARK EDDP Assay is an immunoassay intended for the qualitative and/or semiquantitative determination of EDDP in human urine at cutoff concentrations of 100 ng/mL and 300 ng/mL. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.
The semiquantitative mode is for the purpose of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method, such as Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS), or (2) permitting laboratories to establish quality control procedures.
The ARK EDDP Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed positive analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.
The ARK EDDP Assay is a homogeneous enzyme immunoassay technique used for the analysis of EDDP in human urine. The assay is based on competition between EDDP in the specimen and EDDP labeled with recombinant glucose-6-phosphate dehydrogenase (rG6PDH) for antibody binding sites. As the latter binds antibody, enzyme activity decreases. In the presence of EDDP from the specimen, enzyme activity increases and is directly related to the EDDP concentration. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically. Endogenous G6PDH does not interfere because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
The ARK EDDP Assay consists of reagents R1 anti-EDDP rabbit antibody with substrate and R2 EDDP derivative labeled with bacterial recombinant G6PDH enzyme.
Here's a breakdown of the acceptance criteria and study details for the ARK EDDP Assay, based on the provided FDA 510(k) summary:
This document describes a medical device, the ARK EDDP Assay, which is an immunoassay for the qualitative and/or semi-quantitative determination of EDDP (a methadone metabolite) in human urine. The study presented is a non-clinical performance evaluation, as indicated by "Brief Description of Nonclinical and Clinical Data" and the nature of the tests (Precision, Analytical Recovery, Analytical Specificity, Interference, etc.). No clinical study involving human patients or human readers in an MRMC setting is described.
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly state "acceptance criteria" as a single, consolidated table with specific pass/fail values for each performance characteristic. Instead, performance is evaluated against expected behaviors for an in vitro diagnostic (IVD) device of this type. The inferred acceptance criteria are that the device should demonstrate acceptable precision, accurate analytical recovery, appropriate cross-reactivity and interference profiles, and good agreement with a gold standard (GC/MS) in method comparison for qualitative and semi-quantitative results.
| Performance Characteristic | Inferred Acceptance Criteria (Implicit) | Reported Device Performance and Discussion |
|---|---|---|
| Precision | Consistent qualitative and semi-quantitative results, particularly around the cutoff concentrations. | 100 ng/mL Cutoff: |
| - Qualitative (100 ng/mL) | All samples ≤ 75 ng/mL should be negative; all samples ≥ 125 ng/mL should be positive. Samples at 100 ng/mL (cutoff) may show mixed results. | At 0.0, 25.0, 50.0, and 75.0 ng/mL, all 160 results were Negative. At 125.0, 150.0, 175.0, and 200.0 ng/mL, all 160 results were Positive. At the 100.0 ng/mL cutoff, 123 were Negative and 37 were Positive, which is expected for a cutoff concentration. |
| - Semiquantitative (100 ng/mL) | Mean measured concentrations should be close to theoretical values. Expected mixed qualitative results at cutoff. | Mean concentrations were close to theoretical (e.g., 98.1 ng/mL for 100 ng/mL theoretical). At 100.0 ng/mL cutoff, 114 were Negative and 46 were Positive, consistent with a cutoff. |
| - Qualitative (300 ng/mL) | All samples ≤ 225 ng/mL should be negative; all samples ≥ 375 ng/mL should be positive. Samples at 300 ng/mL (cutoff) may show mixed results. | At 0.0, 75.0, 150.0, and 225.0 ng/mL, all 160 results were Negative. At 375.0, 450.0, 525.0, and 600.0 ng/mL, all 160 results were Positive. At the 300.0 ng/mL cutoff, 57 were Negative and 103 were Positive, which is expected. |
| - Semiquantitative (300 ng/mL) | Mean measured concentrations should be close to theoretical values. Expected mixed qualitative results at cutoff. | Mean concentrations were close to theoretical (e.g., 298.8 ng/mL for 300 ng/mL theoretical). At 300.0 ng/mL cutoff, 85 were Negative and 75 were Positive, consistent with a cutoff. |
| Analytical Recovery | Measured concentration should be within an acceptable percentage range of the theoretical concentration. | Recovery percentages ranged from 94.6% to 107.9% across EDDP concentrations from 50.0 to 1000.0 ng/mL, indicating good analytical recovery. |
| Analytical Specificity | Minimal to no cross-reactivity with structurally related and unrelated compounds, ensuring specificity for EDDP. | Structurally Related Compounds: EDDP showed 100% cross-reactivity (as expected). Methadone and EMDP showed very low cross-reactivity (<0.005% to <0.03%). Several other structurally related compounds (Chlorpromazine, Diphenhydramine, Methylphenidate, Doxylamine) showed low cross-reactivity (<0.1% to <0.3%). This indicates good specificity for EDDP while acknowledging minor cross-reactivity with very high concentrations of certain related compounds. |
| Interference | Endogenous substances, specific gravity variations, and pH variations should not cause false positive or false negative results. | Structurally Unrelated Compounds: Numerous compounds were tested at high concentrations (e.g., 100,000 ng/mL, 500,000 ng/mL). For both 100 ng/mL and 300 ng/mL cutoffs, none of the listed substances yielded a false result (i.e., they were negative at -25% cutoff EDDP concentration and positive at +25% cutoff EDDP concentration), indicating no significant interference from these compounds even at very high concentrations. |
| Endogenous Substances: High concentrations of common endogenous substances (e.g., Acetone, Ascorbic Acid, Bilirubin, Creatinine, Ethanol, Glucose, Hemoglobin, etc.) were tested. No interference was observed for either the 100 ng/mL or 300 ng/mL cutoff, meaning they did not cause false results. | ||
| Specific Gravity: Urine samples across the physiologic range (1.002 to 1.030) did not show interference for either cutoff, meaning results remained correct (negative at -25% cutoff EDDP, positive at +25% cutoff EDDP). | ||
| pH: Urine samples across the physiologic pH range (3.0 to 11.0) did not show interference for either cutoff, meaning results remained correct (negative at -25% cutoff EDDP, positive at +25% cutoff EDDP). | ||
| Method Comparison | High concordance with the GC/MS gold standard, particularly for samples well above or below the cutoff. Minor discrepancies near the cutoff are acceptable. | 100 ng/mL Cutoff: 104 out of 109 samples showed perfect agreement with GC/MS (40 Low Negative, 60 High Positive). 5 samples were Near Cutoff Negative by GC/MS but Negative by immunoassay (expected). 4 samples were Near Cutoff Positive by GC/MS but Positive by immunoassay (expected). This shows high concordance. |
| 300 ng/mL Cutoff: 104 out of 109 samples showed perfect agreement with GC/MS (49 Low Negative, 52 High Positive). 4 samples were Near Cutoff Negative by GC/MS but Negative by immunoassay (expected). 3 samples were Near Cutoff Positive by GC/MS but Positive by immunoassay (expected). One discordant result: Sample 51 was Immunoassay Positive but 294 ng/mL by GC/MS (which is slightly below the 300 ng/mL cutoff). This is a single false positive near the cutoff, which can occur for immunoassay methods. | ||
| Calibration Curve Stability | Calibration should remain stable for a reasonable period. | A stored calibration curve was effective up to at least 15 days, which is an acceptable stability period for a laboratory assay. |
2. Sample sizes used for the test set and the data provenance
- Precision Studies:
- Target Concentrations: 9 levels for each cutoff (0.0 to 200.0 ng/mL for 100 ng/mL cutoff; 0.0 to 600.0 ng/mL for 300 ng/mL cutoff).
- Replicates: Quaduplicate twice a day for 20 days.
- Total Sample Size per Cutoff (for precision): 9 levels * 4 replicates * 2 (times/day) * 20 days = 1440 individual readings for the qualitative precision tables. (Note: The table summary states N=160 for each concentration level, implying 160 results for each of the 9 spiked levels.)
- Data Provenance: Drug-free, negative human urine (spiked with EDDP). The country of origin is not specified but is likely the U.S. as it's an FDA submission. This is a prospective study (laboratory testing).
- Analytical Recovery:
- Target Concentrations: 12 levels (50.0 to 1000.0 ng/mL).
- Replicates: N=6 for mean concentration calculation.
- Total Sample Size: 12 levels * 6 replicates = 72 individual readings.
- Data Provenance: Drug-free, negative human urine (spiked with EDDP). Prospective laboratory testing.
- Analytical Specificity (Cross-reactivity):
- Number of Compounds Tested: 7 structurally related compounds.
- Sample Size: Each compound tested at minimum concentration for positive result (or highest tested) at both cutoff levels. Implies multiple tests per compound, but specific N is not given per compound; typically, this would be triplicate.
- Data Provenance: Drug-free, negative human urine (spiked with compounds). Prospective laboratory testing.
- Interference (Structurally Unrelated, Endogenous Substances, Specific Gravity, pH):
- Number of Compounds/Conditions Tested: Numerous (tens of compounds + SG and pH conditions).
- Sample Size: Each compound/condition tested at two EDDP levels (±25% of cutoff). Implies multiple tests per condition, but specific N is not given per condition; typically, this would be triplicate.
- Data Provenance: Drug-free, negative human urine (spiked with EDDP and interferent). Prospective laboratory testing.
- Method Comparison:
- Sample Size: 109 unaltered clinical human urine specimens.
- Data Provenance: "unaltered clinical human urine specimens that are not individually identifiable." Country of origin not specified, but usually implies local (US) clinical samples. Described as "clinical human urine specimens," suggesting these were collected from actual patients, making this a retrospective evaluation of real-world samples.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The ground truth for this diagnostic device is established by Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS), which are highly accurate analytical chemistry techniques considered the gold standard for drug confirmation in urine. These methods are performed in laboratories by trained laboratory professionals (analytical chemists, toxicologists). The document states the GC/MS confirmatory method was performed by a "licensed reference laboratory."
- Number of experts: Not explicitly stated, as GC/MS analysis is a scientific measurement, not typically an "expert reader" interpretation process like image analysis. The "experts" are the qualified laboratory personnel operating and interpreting the GC/MS results.
- Qualifications of experts: Implied to be qualified laboratory professionals with expertise in analytical chemistry and toxicology, as they work in a "licensed reference laboratory." Specific certifications or experience years are not given.
4. Adjudication method for the test set
Not applicable. The ground truth is established by a definitive analytical method (GC/MS). There is no "adjudication" in the sense of reconciling disagreements among multiple human readers for an image-based or qualitative assessment. The GC/MS provides a quantitative concentration, which is then used to determine positive/negative relative to the cutoffs. Any "discordance" (like the one noted in the method comparison for Sample 51) is a direct comparison between the immunoassay result and the GC/MS result, not a result of human reader discrepancy requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) assay designed to provide an automated analytical result (qualitative or semi-quantitative concentration of EDDP). It is not an imaging AI device that assists human readers in interpreting medical images. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, the primary performance evaluation is standalone. The ARK EDDP Assay is an automated immunoassay performed on clinical chemistry analyzers. The results (qualitative positive/negative or semi-quantitative concentration) are generated by the assay system (reagents + analyzer) without human interpretive input for the result itself. The results are then reported to a clinician for medical interpretation in conjunction with other clinical information. The precision, analytical recovery, specificity, interference, and method comparison studies all represent the "algorithm only" or "device only" performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth used is definitive analytical method (Gas Chromatography/Mass Spectrometry - GC/MS). For in vitro diagnostic assays measuring specific chemical compounds, GC/MS is considered the most accurate and reliable "truth" method.
8. The sample size for the training set
This document does not specify a separate "training set" sample size. For an immunoassay like the ARK EDDP Assay, the "training" (development and optimization) phase would involve numerous laboratory experiments to select reagents, optimize assay conditions, and establish initial performance characteristics. This is distinct from machine learning model training sets. The studies presented (precision, analytical recovery, specificity, interference, method comparison) are performance validation studies for the final device, not datasets used to train an algorithm.
9. How the ground truth for the training set was established
Not applicable in the context of an immunoassay. The "ground truth" for developing an immunoassay would be established by controlled spiking experiments using certified reference materials of the analyte (EDDP) in a known matrix (drug-free human urine), and confirming those concentrations using methods like GC/MS or LC-MS/MS, or by gravimetric/volumetric preparation. The goal is to ensure the assay accurately measures the target compound.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2018
ARK Diagnostics, Inc. Cherry Mun Manager, Quality and Regulatory Affairs 48089 Fremont Boulevard Fremont, California 94538
Re: K182779
Trade/Device Name: ARK EDDP Assay Regulation Number: 21 CFR 862.3620 Regulation Name: Methadone test system Regulatory Class: Class II Product Code: DJR Dated: September 28, 2018 Received: October 1, 2018
Dear Cherry Mun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and
Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Paula Caposino -S
for Courtnev H. Lias. Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. 0910-0120 | |
|---|---|
| Expiration Date: 06/30/2020 | |
| See PRA Statement below. |
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
| 510(k) Number (if known) | K182779 |
|---|---|
| Device Name | ARK EDDP Assay |
Indications for Use (Describe)
The ARK EDDP Assay is an immunoassay intended for the qualitative and/or semiquantitative determination of EDDP in
human urine at cutoff concentrations of 100 ng/mL and 300 ng/mL. The assay is intended for use in laboratories with
automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.
The semiquantitative mode is for the purpose of (1) enabling laboratories to determine an appropriate dilution of the
specimen for confirmation by a confirmatory method, such as Gas Chromatography/Mass Spectrometry (GC/MS) or
Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS), or (2) permitting laboratories to establish quality
control procedures.
The ARK EDDP Assay provides only a preliminary analytical test result. A more specific alternative chemical method
must be used in order to obtain a confirmed positive analytical result. Gas Chromatography/Mass Spectrometry (GC/MS)
or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical
consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary
test result is positive.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (7/17) | Page 1 of 1 |
|---|---|
| ---------------------- | ------------- |
PSC Publishing Services (301) 443-6740 EF
노-남 10 レ-৮ ခုနှစ် ချောင်းများ မြို့နယ်ရှိ ရွာများ BotU
{3}------------------------------------------------
510(k) SUMMARY
This 510(k) Summary of Safety and Effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act of 1990 and 21 CFR 807.92.
The assigned 510(k) number is K182779.
| 807.92 (a)(1): Name: | ARK Diagnostics, Inc. |
|---|---|
| Address: | 48089 Fremont BlvdFremont, CA 94538 USA |
| Owner Operator Number: | 10027663 |
| Establishment Registration: | 3005755244 |
| Phone: | (510) 270-6270 |
| FAX: | (510) 270-6298 |
| Contact: | Cherry Mun – (510) 270-6288Manager of Quality and Regulatory Affairs |
Date Prepared: November 19th, 2018
807.92 (a)(2): Device Name – Trade Name, Common Name, and Classification
| Trade Name: | ARK™ EDDP Assay |
|---|---|
| Common Name: | Homogeneous Enzyme Immunoassay, Methadone Test System |
Classification:
| Product Code | Classification | Regulation Section | Panel |
|---|---|---|---|
| DJR | Class II | 21 CFR 862.3620Methadone Test System | Toxicology(91) |
807.92 (a)(3): Identification of the Legally Marketed Predicate Device
Immunalysis EDDP Specific Urine Enzyme Immunoassay (K151395)
{4}------------------------------------------------
807.92 (a)(4): Device Description
The ARK EDDP Assay is a homogeneous enzyme immunoassay technique used for the analysis of EDDP in human urine. The assay is based on competition between EDDP in the specimen and EDDP labeled with recombinant glucose-6-phosphate dehydrogenase (rG6PDH) for antibody binding sites. As the latter binds antibody, enzyme activity decreases. In the presence of EDDP from the specimen, enzyme activity increases and is directly related to the EDDP concentration. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH in the presence of glucose-6-phosphate (G6P), resulting in an absorbance change that is measured spectrophotometrically. Endogenous G6PDH does not interfere because the coenzyme NAD functions only with the bacterial enzyme used in the assay.
The ARK EDDP Assay consists of reagents R1 anti-EDDP rabbit antibody with substrate and R2 EDDP derivative labeled with bacterial recombinant G6PDH enzyme.
807.92 (a)(5): Intended Use / Indications for Use
ARK EDDP Assay
The ARK EDDP Assay is an immunoassay intended for the qualitative and/or semiquantitative determination of EDDP in human urine at cutoff concentrations of 100 ng/mL and 300 ng/mL. The assay is intended for use in laboratories with automated clinical chemistry analyzers. This in vitro diagnostic device is for prescription use only.
The semiquantitative mode is for the purpose of (1) enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method, such as Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS), or (2) permitting laboratories to establish quality control procedures.
The ARK EDDP Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed positive analytical result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug test result, particularly when the preliminary test result is positive.
{5}------------------------------------------------
807.92 (a)(6): Technological Similarities and Differences to the Predicate
SUBSTANTIAL EQUIVALENCE COMPARATIVE TABLE
Comparison between the Immunalysis EDDP Specific Urine Enzyme Immunoassay and the ARK™ EDDP Assay
| Characteristic | Predicate DeviceImmunalysis EDDP Specific UrineEnzyme Immunoassay (K151395) | Candidate DeviceARKTM EDDP Assay |
|---|---|---|
| Similarities | ||
| Test System | Homogenous enzyme immunoassay(EIA) | Same |
| Intended Use | For the qualitative and semiquantitativedetermination of EDDP in human urine;For in vitro diagnostic use | Same |
| Sample Matrix | Human urine | Same |
| User Environment | Clinical laboratories; Prescription useonly | Same |
| Mass SpectrometryConfirmation | Required to confirm preliminary positiveanalytical results | Same |
| Platform Required | Automated clinical chemistry analyzer | Same |
| Reagents Form | Liquid - Ready to use | Same |
| Reagent Materials | Two (2) reagent system:Antibody/substrate reagent (antibodies toEDDP) and enzyme labeled conjugate(EDDP derivative labeled with enzyme)Sodium azide preservative | Same |
| Storage | 2-8°C until expiration date | Same |
| Measured Analyte | EDDP | Same |
| Detection | Absorbance change measuredspectrophotometrically at 340 nm | Same |
| Characteristic | Predicate DeviceImmunalysis EDDP Specific UrineEnzyme Immunoassay (K151395) | Candidate DeviceARKTM EDDP Assay |
| Differences | ||
| Cutoff Levels | 100 ng/mL, 300 ng/mL and 1000 ng/mL | 100 ng/mL and 300 ng/mL |
{6}------------------------------------------------
807.92 (b)(1) and 807.92 (b)(2): Brief Description of Nonclinical and Clinical Data
The following performance characteristics were obtained on the Beckman Coulter AU680® automated clinical chemistry analyzer.
Precision
Precision studies were performed using CLSI EP05-A3 as a guideline. Drug-free, negative human urine was supplemented with EDDP (0.0 to 200.0 ng/mL for the 100 ng/mL cutoff and 0.0 to 600.0 for the 300 ng/mL cutoff). Each level was assayed in quadruplicate twice a day for 20 days (N=160) and evaluated qualitatively and semiquantitatively. Results are summarized in the tables below.
| Human Urine(ng/mL) | Relative %Cutoff | # of Results | QualitativePrecision Results |
|---|---|---|---|
| 0.0 | -100 | 160 | 160 Negative |
| 25.0 | -75 | 160 | 160 Negative |
| 50.0 | -50 | 160 | 160 Negative |
| 75.0 | -25 | 160 | 160 Negative |
| 100.0 | Cutoff | 160 | 123 Negative;37 Positive |
| 125.0 | +25 | 160 | 160 Positive |
| 150.0 | +50 | 160 | 160 Positive |
| 175.0 | +75 | 160 | 160 Positive |
| 200.0 | +100 | 160 | 160 Positive |
Qualitative Precision - 100 ng/mL Cutoff
Semiquantitative Precision – 100 ng/mL Cutoff
| Human Urine(ng/mL) | Relative %Cutoff | # of Results | Mean(ng/mL) | SemiquantitativePrecision Results |
|---|---|---|---|---|
| 0.0 | -100 | 160 | 0.3 | 160 Negative |
| 25.0 | -75 | 160 | 22.6 | 160 Negative |
| 50.0 | -50 | 160 | 47.7 | 160 Negative |
| 75.0 | -25 | 160 | 72.2 | 160 Negative |
| 100.0 | Cutoff | 160 | 98.1 | 114 Negative;46 Positive |
| 125.0 | +25 | 160 | 125.3 | 160 Positive |
| 150.0 | +50 | 160 | 145.1 | 160 Positive |
| 175.0 | +75 | 160 | 169.4 | 160 Positive |
| 200.0 | +100 | 160 | 190.7 | 160 Positive |
{7}------------------------------------------------
Qualitative Precision – 300 ng/mL Cutoff
| Human Urine(ng/mL) | Relative %Cutoff | # of Results | QualitativePrecision Results |
|---|---|---|---|
| 0.0 | -100 | 160 | 160 Negative |
| 75.0 | -75 | 160 | 160 Negative |
| 150.0 | -50 | 160 | 160 Negative |
| 225.0 | -25 | 160 | 160 Negative |
| 300.0 | Cutoff | 160 | 57 Negative;103 Positive |
| 375.0 | +25 | 160 | 160 Positive |
| 450.0 | +50 | 160 | 160 Positive |
| 525.0 | +75 | 160 | 160 Positive |
| 600.0 | +100 | 160 | 160 Positive |
Semiquantitative Precision – 300 ng/mL Cutoff
| Human Urine(ng/mL) | Relative %Cutoff | # of Results | Mean(ng/mL) | SemiquantitativePrecision Results |
|---|---|---|---|---|
| 0.0 | -100 | 160 | 0.3 | 160 Negative |
| 75.0 | -75 | 160 | 72.2 | 160 Negative |
| 150.0 | -50 | 160 | 145.1 | 160 Negative |
| 225.0 | -25 | 160 | 205.9 | 160 Negative |
| 300.0 | Cutoff | 160 | 298.8 | 85 Negative;75 Positive |
| 375.0 | +25 | 160 | 381.4 | 160 Positive |
| 450.0 | +50 | 160 | 461.0 | 160 Positive |
| 525.0 | +75 | 160 | 539.8 | 160 Positive |
| 600.0 | +100 | 160 | 620.0 | 160 Positive |
{8}------------------------------------------------
Analytical Recovery
Recovery across the assay range was assessed using the semiquantitative mode. Drug-free, negative human urine was supplemented with EDDP (1100.0 ng/mL) and dilutions were made proportionally with drug-free human urine. EDDP concentrations ranged from 50.0 to 1000.0 ng/mL. At each level, percentage recovery was calculated based on the mean concentration (N=6) compared to the expected concentration. Results are summarized in the table below.
| TheoreticalConcentration(ng/mL) | MeanConcentration(ng/mL) | Recovery(%) |
|---|---|---|
| 50.0 | 47.6 | 95.1 |
| 75.0 | 72.1 | 96.1 |
| 100.0 | 97.1 | 97.1 |
| 200.0 | 189.1 | 94.6 |
| 300.0 | 286.6 | 95.5 |
| 400.0 | 414.5 | 103.6 |
| 500.0 | 506.6 | 101.3 |
| 600.0 | 647.4 | 107.9 |
| 700.0 | 722.7 | 103.2 |
| 800.0 | 800.6 | 100.1 |
| 900.0 | 880.8 | 97.9 |
| 1000.0 | 955.8 | 95.6 |
{9}------------------------------------------------
Analytical Specificity
All compounds tested were added to drug-free, negative human urine and tested with the ARK EDDP Assay in both qualitative and semiquantitative modes.
The cross-reactivity of the following structurally related compounds was evaluated by spiking these compounds into drug-free, negative human urine to determine the minimum concentration that would give a positive result approximately equivalent to the 100 ng/mL and 300 ng/mL EDDP cutoffs. These concentrations were used to determine the percent cross-reactivity according to the formula:
% Cross-reactivity = (Cutoff concentration / Lowest concentration of cross-reactant causing a positive result) X 100
For compounds that did not produce a positive result, the highest concentration tested was used to calculate percent cross-reactivity.
| Compound | ConcentrationTested(ng/mL) | SemiquantitativeMode Result(Positive/Negative) | Qualitative ModeResult(Positive/Negative) | Cross-reactivity(%) |
|---|---|---|---|---|
| EDDP | 100 | Positive | Positive | 100 |
| Methadone | 2,000,000 | Negative | Negative | <0.005 |
| EMDP | 400,000 | Negative | Negative | <0.025 |
| Chlorpromazine | 100,000 | Negative | Negative | <0.1 |
| Diphenhydramine | 100,000 | Negative | Negative | <0.1 |
| Methylphenidate | 100,000 | Negative | Negative | <0.1 |
| Doxylamine | 100,000 | Negative | Negative | <0.1 |
Structurally Related Compounds - 100 ng/mL Cutoff
Structurally Related Compounds - 300 ng/mL Cutoff
| Compound | ConcentrationTested(ng/mL) | SemiquantitativeMode Result(Positive/Negative) | Qualitative ModeResult(Positive/Negative) | Cross-reactivity(%) |
|---|---|---|---|---|
| EDDP | 300 | Positive | Positive | 100 |
| Methadone | 4,500,000 | Negative | Negative | <0.007 |
| EMDP | 1,000,000 | Negative | Negative | <0.03 |
| Chlorpromazine | 100,000 | Negative | Negative | <0.3 |
| Diphenhydramine | 100,000 | Negative | Negative | <0.3 |
| Methylphenidate | 100,000 | Negative | Negative | <0.3 |
| Doxylamine | 100,000 | Negative | Negative | <0.3 |
{10}------------------------------------------------
Interference
Structurally Unrelated Compounds – 100 ng/mL Cutoff
High concentrations of the following structurally unrelated compounds were added into urine spiked with EDDP (± 25% of the 100 ng/mL cutoff concentration) and tested with the ARK EDDP Assay in both qualitative and semiquantitative modes. The substances listed at the concentrations below did not yield a false result relative to the 100 ng/mL cutoff.
| Compound | ConcentrationTested(ng/mL) | Spiked EDDP Level | |
|---|---|---|---|
| 75 ng/mL(-25% Cutoff) | 125 ng/mL(+25% Cutoff) | ||
| 4-bromo 2-5, dimethoxyphenethylamine | 100,000 | Negative | Positive |
| Acetaminophen | 500,000 | Negative | Positive |
| Acetylsalicylic Acid | 500,000 | Negative | Positive |
| 6-Acetylcodeine | 100,000 | Negative | Positive |
| 6-Acetylmorphine | 100,000 | Negative | Positive |
| Alprazolam | 100,000 | Negative | Positive |
| 7-Aminoclonazepam | 100,000 | Negative | Positive |
| 7-Aminoflunitrazepam | 100,000 | Negative | Positive |
| 7-Aminonitrazepam | 100,000 | Negative | Positive |
| Amitriptyline | 100,000 | Negative | Positive |
| Amobarbital | 100,000 | Negative | Positive |
| S-(+)-Amphetamine | 100,000 | Negative | Positive |
| Benzylpiperazine | 100,000 | Negative | Positive |
| Bromazepam | 100,000 | Negative | Positive |
| Buprenorphine | 100,000 | Negative | Positive |
| Bupropion | 100,000 | Negative | Positive |
| Butabarbital | 100,000 | Negative | Positive |
| Butalbital | 100,000 | Negative | Positive |
| Caffeine | 500,000 | Negative | Positive |
| Cannabidiol | 100,000 | Negative | Positive |
| Cannabinol | 100,000 | Negative | Positive |
| Carbamazepine | 100,000 | Negative | Positive |
| Carisoprodol | 100,000 | Negative | Positive |
| Chlordiazepoxide | 100,000 | Negative | Positive |
| Cis-Tramadol | 100,000 | Negative | Positive |
| Clobazam | 100,000 | Negative | Positive |
| Clomipramine | 100,000 | Negative | Positive |
| Clonazepam | 100,000 | Negative | Positive |
| Clozapine | 100,000 | Negative | Positive |
| Codeine | 100,000 | Negative | Positive |
| Cotinine | 100,000 | Negative | Positive |
| Cyclobenzaprine | 100,000 | Negative | Positive |
| Dehydronorketamine | 100,000 | Negative | Positive |
| Demoxepam | 100,000 | Negative | Positive |
| Desipramine | 100,000 | Negative | Positive |
| Desalkylflurazepam | 100,000 | Negative | Positive |
| Dextromethorphan | 100,000 | Negative | Positive |
{11}------------------------------------------------
| Concentration | Spiked EDDP Level | ||
|---|---|---|---|
| Compound | Tested | 75 ng/mL | 125 ng/mL |
| (ng/mL) | (-25% Cutoff) | (+25% Cutoff) | |
| Diazepam | 100,000 | Negative | Positive |
| Digoxin | 100,000 | Negative | Positive |
| Dihydrocodeine | 100,000 | Negative | Positive |
| Δ9 THC | 100,000 | Negative | Positive |
| Doxepin | 100,000 | Negative | Positive |
| 1R,2S (-) Ephedrine | 100,000 | Negative | Positive |
| 1S,2R (+) Ephedrine | 100,000 | Negative | Positive |
| Ethyl-ß-D-Glucuronide | 100,000 | Negative | Positive |
| Ethylmorphine | 100,000 | Negative | Positive |
| (S-)-Fenfluramine | 100,000 | Negative | Positive |
| (R+)-Fenfluramine | 100,000 | Negative | Positive |
| Fentanyl | 100,000 | Negative | Positive |
| Flunitrazepam | 100,000 | Negative | Positive |
| Fluoxetine | 100,000 | Negative | Positive |
| Flurazepam | 100,000 | Negative | Positive |
| Haloperidol | 100,000 | Negative | Positive |
| Heroin | 100,000 | Negative | Positive |
| Hexobarbital | 100,000 | Negative | Positive |
| Hydrocodone | 100,000 | Negative | Positive |
| Hydromorphone | 100,000 | Negative | Positive |
| 11-hydroxy-Δ9 THC | 100,000 | Negative | Positive |
| Ibuprofen | 500,000 | Negative | Positive |
| Imipramine | 100,000 | Negative | Positive |
| Ketamine | 100,000 | Negative | Positive |
| Lamotrigine | 100,000 | Negative | Positive |
| Levorphanol Tartrate | 100,000 | Negative | Positive |
| Lidocaine | 100,000 | Negative | Positive |
| Lorazepam | 100,000 | Negative | Positive |
| Lorazepam Glucuronide | 50,000 | Negative | Positive |
| Lormetrazepam | 100,000 | Negative | Positive |
| LSD | 100,000 | Negative | Positive |
| Maprotiline | 100,000 | Negative | Positive |
| (+)-MDA | 100,000 | Negative | Positive |
| MDEA | 100,000 | Negative | Positive |
| MDMA | 100,000 | Negative | Positive |
| Meperidine | 100,000 | Negative | Positive |
| Meprobamate | 100,000 | Negative | Positive |
| S(+)-Methamphetamine | 100,000 | Negative | Positive |
| Methaquolone | 100,000 | Negative | Positive |
| Methoxetamine | 100,000 | Negative | Positive |
| Methylone | 100,000 | Negative | Positive |
| Midazolam | 100,000 | Negative | Positive |
| Morphine | 100,000 | Negative | Positive |
| Morphine-3B-D-Glucuronide | 100,000 | Negative | Positive |
| Morphine-6ß-D-Glucuronide | 50,000 | Negative | Positive |
| N-Desmethyltapentadol | 100,000 | Negative | Positive |
| Nalorphine | 100,000 | Negative | Positive |
ARK Diagnostics, Inc. – 510(k) Summary ARK EDDP Assay
Page 5-9 of 5-19
{12}------------------------------------------------
| Compound | ConcentrationTested(ng/mL) | Spiked EDDP Level | |
|---|---|---|---|
| 75 ng/mL(-25% Cutoff) | 125 ng/mL(+25% Cutoff) | ||
| Naloxone | 100,000 | Negative | Positive |
| Naltrexone | 100,000 | Negative | Positive |
| Naproxen | 100,000 | Negative | Positive |
| Nitrazepam | 100,000 | Negative | Positive |
| 11-nor-9-carboxy-Δ9-THC | 100,000 | Negative | Positive |
| Norbuprenorphine | 50,000 | Negative | Positive |
| Norcodeine | 100,000 | Negative | Positive |
| Nordiazepam | 100,000 | Negative | Positive |
| Norketamine | 100,000 | Negative | Positive |
| Normorphine | 100,000 | Negative | Positive |
| Norpropoxyphene | 100,000 | Negative | Positive |
| Norpseudoephedrine | 100,000 | Negative | Positive |
| Nortriptyline | 100,000 | Negative | Positive |
| Olanzapine | 100,000 | Negative | Positive |
| Oxazepam | 100,000 | Negative | Positive |
| Oxycodone | 100,000 | Negative | Positive |
| Oxymorphone | 100,000 | Negative | Positive |
| PCP | 100,000 | Negative | Positive |
| Pentazocine | 100,000 | Negative | Positive |
| Pentobarbital | 100,000 | Negative | Positive |
| Phenobarbital | 100,000 | Negative | Positive |
| Phentermine | 100,000 | Negative | Positive |
| Phenylephedrine | 100,000 | Negative | Positive |
| Phenylpropanolamine | 100,000 | Negative | Positive |
| Phenytoin | 100,000 | Negative | Positive |
| PMA | 100,000 | Negative | Positive |
| Prazepam | 100,000 | Negative | Positive |
| Propoxyphene | 100,000 | Negative | Positive |
| Propranolol | 100,000 | Negative | Positive |
| Protriptyline | 100,000 | Negative | Positive |
| R,R (+)- Pseudoephedrine | 100,000 | Negative | Positive |
| S,S (-)- Pseudoephedrine | 100,000 | Negative | Positive |
| Ranitidine | 100,000 | Negative | Positive |
| Ritalinic Acid | 100,000 | Negative | Positive |
| Salicylic Acid | 100,000 | Negative | Positive |
| Secobarbital | 100,000 | Negative | Positive |
| Sertraline | 100,000 | Negative | Positive |
| Sufentanil Citrate | 50,000 | Negative | Positive |
| Tapentadol | 100,000 | Negative | Positive |
| Temazepam | 100,000 | Negative | Positive |
| Theophylline | 100,000 | Negative | Positive |
| Thioridazine | 100,000 | Negative | Positive |
| Trazodone | 100,000 | Negative | Positive |
| Triazolam | 100,000 | Negative | Positive |
| Trifluoromethylphenylpiperazine | 100,000 | Negative | Positive |
| Trimipramine | 100,000 | Negative | Positive |
| Venlafaxine | 100,000 | Negative | Positive |
ARK Diagnostics, Inc. – 510(k) Summary ARK EDDP Assay
Page 5-10 of 5-19
{13}------------------------------------------------
| Concentration | Spiked EDDP Level | ||
|---|---|---|---|
| Compound | Tested(ng/mL) | 75 ng/mL(-25% Cutoff) | 125 ng/mL(+25% Cutoff) |
| Verapamil | 100,000 | Negative | Positive |
| Zolpidem Tartrate | 100,000 | Negative | Positive |
Structurally Unrelated Compounds – 300 ng/mL Cutoff
High concentrations of the following structurally unrelated compounds were added into urine spiked with EDDP (± 25% of the 300 ng/mL cutoff concentration) and tested with the ARK EDDP Assay in both qualitative and semiquantitative modes. The substances listed at the concentrations below did not yield a false result relative to the 300 ng/mL cutoff.
| Compound | ConcentrationTested(ng/mL) | Spiked EDDP Level | |
|---|---|---|---|
| 225 ng/mL(-25% Cutoff) | 375 ng/mL(+25% Cutoff) | ||
| 4-bromo 2-5, dimethoxyphenethylamine | 100,000 | Negative | Positive |
| Acetaminophen | 500,000 | Negative | Positive |
| Acetylsalicylic Acid | 500,000 | Negative | Positive |
| 6-Acetylcodeine | 100,000 | Negative | Positive |
| 6-Acetylmorphine | 100,000 | Negative | Positive |
| Alprazolam | 100,000 | Negative | Positive |
| 7-Aminoclonazepam | 100,000 | Negative | Positive |
| 7-Aminoflunitrazepam | 100,000 | Negative | Positive |
| 7-Aminonitrazepam | 100,000 | Negative | Positive |
| Amitriptyline | 100,000 | Negative | Positive |
| Amobarbital | 100,000 | Negative | Positive |
| S-(+)-Amphetamine | 100,000 | Negative | Positive |
| Benzylpiperazine | 100,000 | Negative | Positive |
| Bromazepam | 100,000 | Negative | Positive |
| Buprenorphine | 100,000 | Negative | Positive |
| Bupropion | 100,000 | Negative | Positive |
| Butabarbital | 100,000 | Negative | Positive |
| Butalbital | 100,000 | Negative | Positive |
| Caffeine | 500,000 | Negative | Positive |
| Cannabidiol | 100,000 | Negative | Positive |
| Cannabinol | 100,000 | Negative | Positive |
| Carbamazepine | 100,000 | Negative | Positive |
| Carisoprodol | 100,000 | Negative | Positive |
| Chlordiazepoxide | 100,000 | Negative | Positive |
| Cis-Tramadol | 100,000 | Negative | Positive |
| Clobazam | 100,000 | Negative | Positive |
| Clomipramine | 100,000 | Negative | Positive |
| Clonazepam | 100,000 | Negative | Positive |
| Clozapine | 100,000 | Negative | Positive |
| Codeine | 100,000 | Negative | Positive |
| Cotinine | 100,000 | Negative | Positive |
| Cyclobenzaprine | 100,000 | Negative | Positive |
| Dehydronorketamine | 100,000 | Negative | Positive |
| Concentration | Spiked EDDP Level | ||
| Compound | Tested (ng/mL) | 225 ng/mL (-25% Cutoff) | 375 ng/mL (+25% Cutoff) |
| Demoxepam | 100,000 | Negative | Positive |
| Desipramine | 100,000 | Negative | Positive |
| Desalkylflurazepam | 100,000 | Negative | Positive |
| Dextromethorphan | 100,000 | Negative | Positive |
| Diazepam | 100,000 | Negative | Positive |
| Digoxin | 100,000 | Negative | Positive |
| Dihydrocodeine | 100,000 | Negative | Positive |
| Δ9 THC | 100,000 | Negative | Positive |
| Doxepin | 100,000 | Negative | Positive |
| 1R,2S (-) Ephedrine | 100,000 | Negative | Positive |
| 1S,2R (+) Ephedrine | 100,000 | Negative | Positive |
| Ethyl-β-D-Glucuronide | 100,000 | Negative | Positive |
| Ethylmorphine | 100,000 | Negative | Positive |
| (S-)-Fenfluramine | 100.000 | Negative | Positive |
| (R+)-Fenfluramine | 100,000 | Negative | Positive |
| Fentanyl | 100,000 | Negative | Positive |
| Flunitrazepam | 100,000 | Negative | Positive |
| Fluoxetine | 100,000 | Negative | Positive |
| Flurazepam | 100,000 | Negative | Positive |
| Haloperidol | 100,000 | Negative | Positive |
| Heroin | 100,000 | Negative | Positive |
| Hexobarbital | 100,000 | Negative | Positive |
| Hydrocodone | 100,000 | Negative | Positive |
| Hydromorphone | 100,000 | Negative | Positive |
| 11-hydroxy-Δ9 THC | 100,000 | Negative | Positive |
| Ibuprofen | 500,000 | Negative | Positive |
| Imipramine | 100,000 | Negative | Positive |
| Ketamine | 100,000 | Negative | Positive |
| Lamotrigine | 100,000 | Negative | Positive |
| Levorphanol Tartrate | 100,000 | Negative | Positive |
| Lidocaine | 100,000 | Negative | Positive |
| Lorazepam | 100.000 | Negative | Positive |
| Lorazepam Glucuronide | 50,000 | Negative | Positive |
| Lormetrazepam | 100,000 | Negative | Positive |
| LSD | 100,000 | Negative | Positive |
| Maprotiline | 100,000 | Negative | Positive |
| (+)-MDA | 100,000 | Negative | Positive |
| MDEA | 100,000 | Negative | Positive |
| MDMA | 100,000 | Negative | Positive |
| Meperidine | 100,000 | Negative | Positive |
| Meprobamate | 100,000 | Negative | Positive |
| S(+)-Methamphetamine | 100,000 | Negative | Positive |
| Methaquolone | 100,000 | Negative | Positive |
| Methoxetamine | 100,000 | Negative | Positive |
| Methylone | 100,000 | Negative | Positive |
| Midazolam | 100,000 | Negative | Positive |
| Morphine | 100,000 | Negative | Positive |
| Compound | ConcentrationTested(ng/mL) | Spiked EDDP Level | |
| 225 ng/mL(-25% Cutoff) | 375 ng/mL(+25% Cutoff) | ||
| Morphine-3β-D-Glucuronide | 100,000 | Negative | Positive |
| Morphine-6β-D-Glucuronide | 50,000 | Negative | Positive |
| N-Desmethyltapentadol | 100,000 | Negative | Positive |
| Nalorphine | 100,000 | Negative | Positive |
| Naloxone | 100,000 | Negative | Positive |
| Naltrexone | 100,000 | Negative | Positive |
| Naproxen | 100,000 | Negative | Positive |
| Nitrazepam | 100,000 | Negative | Positive |
| 11-nor-9-carboxy-Δ9-THC | 100,000 | Negative | Positive |
| Norbuprenorphine | 50,000 | Negative | Positive |
| Norcodeine | 100,000 | Negative | Positive |
| Nordiazepam | 100,000 | Negative | Positive |
| Norketamine | 100,000 | Negative | Positive |
| Normorphine | 100,000 | Negative | Positive |
| Norpropoxyphene | 75,000 | Negative | Positive |
| Norpseudoephedrine | 100,000 | Negative | Positive |
| Nortriptyline | 100,000 | Negative | Positive |
| Olanzapine | 100,000 | Negative | Positive |
| Oxazepam | 100,000 | Negative | Positive |
| Oxycodone | 100,000 | Negative | Positive |
| Oxymorphone | 100,000 | Negative | Positive |
| PCP | 50,000 | Negative | Positive |
| Pentazocine | 100,000 | Negative | Positive |
| Pentobarbital | 100,000 | Negative | Positive |
| Phenobarbital | 100,000 | Negative | Positive |
| Phentermine | 100,000 | Negative | Positive |
| Phenylephedrine | 100,000 | Negative | Positive |
| Phenylpropanolamine | 100,000 | Negative | Positive |
| Phenytoin | 100,000 | Negative | Positive |
| PMA | 100,000 | Negative | Positive |
| Prazepam | 100,000 | Negative | Positive |
| Propoxyphene | 100,000 | Negative | Positive |
| Propranolol | 100,000 | Negative | Positive |
| Protriptyline | 100,000 | Negative | Positive |
| R,R (+)- Pseudoephedrine | 100,000 | Negative | Positive |
| S,S (-)- Pseudoephedrine | 100,000 | Negative | Positive |
| Ranitidine | 100,000 | Negative | Positive |
| Ritalinic Acid | 100,000 | Negative | Positive |
| Salicylic Acid | 100,000 | Negative | Positive |
| Secobarbital | 100,000 | Negative | Positive |
| Sertraline | 100,000 | Negative | Positive |
| Sufentanil Citrate | 50,000 | Negative | Positive |
| Tapentadol | 100,000 | Negative | Positive |
| Temazepam | 100,000 | Negative | Positive |
| Theophylline | 100,000 | Negative | Positive |
| Thioridazine | 100,000 | Negative | Positive |
| Trazodone | 100,000 | Negative | Positive |
| Compound | ConcentrationTested(ng/mL) | Spiked EDDP Level | |
| 225 ng/mL(-25% Cutoff) | 375 ng/mL(+25% Cutoff) | ||
| Triazolam | 100,000 | Negative | Positive |
| Trifluoromethylphenylpiperazine | 100,000 | Negative | Positive |
| Trimipramine | 100,000 | Negative | Positive |
| Venlafaxine | 100,000 | Negative | Positive |
| Verapamil | 100,000 | Negative | Positive |
| Zolpidem Tartrate | 100,000 | Negative | Positive |
{14}------------------------------------------------
ARK Diagnostics, Inc. – 510(k) Summary ARK EDDP Assay
Page 5-12 of 5-19
{15}------------------------------------------------
ARK Diagnostics, Inc. – 510(k) Summary ARK EDDP Assay
Page 5-13 of 5-19
{16}------------------------------------------------
Endogenous Substances – 100 ng/mL Cutoff
Interference studies were performed using CLSI EP07-A2 as a guideline. High concentrations of the following endogenous substances were added into urine spiked with EDDP (± 25% of the 100 ng/mL cutoff concentration). No interference was observed when tested with the ARK EDDP Assay in both qualitative and semiquantitative modes.
| Spiked EDDP Level | ||||
|---|---|---|---|---|
| Compound | ConcentrationTested | 75 ng/mL(-25% Cutoff) | 125 ng/mL(+25% Cutoff) | |
| Acetone | 1000 mg/dL | Negative | Positive | |
| Ascorbic Acid | 1500 mg/dL | Negative | Positive | |
| Bilirubin – Conjugated | 2 mg/dL | Negative | Positive | |
| Bilirubin – Unconjugated | 2 mg/dL | Negative | Positive | |
| Boric Acid | 1% w/v | Negative | Positive | |
| Creatinine | 500 mg/dL | Negative | Positive | |
| Ethanol | 1000 mg/dL | Negative | Positive | |
| Galactose | 10 mg/dL | Negative | Positive | |
| Gamma Globulin | 500 mg/dL | Negative | Positive | |
| Glucose | 2000 mg/dL | Negative | Positive | |
| Hemoglobin | 300 mg/dL | Negative | Positive | |
| Human Albumin | 500 mg/dL | Negative | Positive | |
| Oxalic Acid | 100 mg/dL | Negative | Positive | |
| Riboflavin | 7.5 mg/dL | Negative | Positive | |
| Sodium Azide | 1% w/v | Negative | Positive | |
| Sodium Chloride | 6000 mg/dL | Negative | Positive | |
| Sodium Fluoride | 1% w/v | Negative | Positive | |
| Urea | 6000 mg/dL | Negative | Positive |
{17}------------------------------------------------
Endogenous Substances – 300 ng/mL Cutoff
Interference studies were performed using CLSI EP07-A2 as a guideline. High concentrations of the following endogenous substances were added into urine spiked with EDDP (± 25% of the 300 ng/mL cutoff concentration). No interference was observed when tested with the ARK EDDP Assay in both qualitative and semiquantitative modes.
| Compound | ConcentrationTested | Spiked EDDP Level | |
|---|---|---|---|
| 225 ng/mL(-25% Cutoff) | 375 ng/mL(+25% Cutoff) | ||
| Acetone | 1000 mg/dL | Negative | Positive |
| Ascorbic Acid | 1500 mg/dL | Negative | Positive |
| Bilirubin – Conjugated | 2 mg/dL | Negative | Positive |
| Bilirubin – Unconjugated | 2 mg/dL | Negative | Positive |
| Boric Acid | 1% w/v | Negative | Positive |
| Creatinine | 500 mg/dL | Negative | Positive |
| Ethanol | 1000 mg/dL | Negative | Positive |
| Galactose | 10 mg/dL | Negative | Positive |
| Gamma Globulin | 500 mg/dL | Negative | Positive |
| Glucose | 2000 mg/dL | Negative | Positive |
| Hemoglobin | 300 mg/dL | Negative | Positive |
| Human Albumin | 500 mg/dL | Negative | Positive |
| Oxalic Acid | 100 mg/dL | Negative | Positive |
| Riboflavin | 7.5 mg/dL | Negative | Positive |
| Sodium Azide | 1% w/v | Negative | Positive |
| Sodium Chloride | 6000 mg/dL | Negative | Positive |
| Sodium Fluoride | 1% w/v | Negative | Positive |
| Urea | 6000 mg/dL | Negative | Positive |
Specific Gravity - 100 ng/mL Cutoff
Urine samples with specific gravity values ranging from 1.002 to 1.030 were tested in the presence of the two levels of EDDP at ± 25% of the 100 ng/mL cutoff concentration. No interference was observed when tested with the ARK EDDP Assay in both qualitative and semiquantitative modes.
| Spiked EDDP Level | ||
|---|---|---|
| Compound | 75 ng/mL(-25% Cutoff) | 125 ng/mL(+25% Cutoff) |
| Specific Gravity 1.002 | Negative | Positive |
| Specific Gravity 1.004 | Negative | Positive |
| Specific Gravity 1.012 | Negative | Positive |
| Specific Gravity 1.018 | Negative | Positive |
| Specific Gravity 1.019 | Negative | Positive |
| Specific Gravity 1.026 | Negative | Positive |
| Specific Gravity 1.030 | Negative | Positive |
{18}------------------------------------------------
Specific Gravity - 300 ng/mL Cutoff
Urine samples with specific gravity values ranging from 1.002 to 1.030 were tested in the presence of the two levels of EDDP at ± 25% of the 300 ng/mL cutoff concentration. No interference was observed when tested with the ARK EDDP Assay in both qualitative and semiquantitative modes.
| Spiked EDDP Level | ||
|---|---|---|
| Compound | 225 ng/mL(-25% Cutoff) | 375 ng/mL(+25% Cutoff) |
| Specific Gravity 1.002 | Negative | Positive |
| Specific Gravity 1.004 | Negative | Positive |
| Specific Gravity 1.012 | Negative | Positive |
| Specific Gravity 1.018 | Negative | Positive |
| Specific Gravity 1.019 | Negative | Positive |
| Specific Gravity 1.026 | Negative | Positive |
| Specific Gravity 1.030 | Negative | Positive |
pH – 100 ng/mL Cutoff
Urine samples with pH values from 3.0 to 11.0 were tested in the presence of the two levels of EDDP at ± 25% of the 100 ng/mL cutoff concentration. No interference was observed when tested with the ARK EDDP Assay in both qualitative and semiquantitative modes.
| Spiked EDDP Level | ||
|---|---|---|
| Compound | 75 ng/mL(-25% Cutoff) | 125 ng/mL(+25% Cutoff) |
| Urine pH 3.0 | Negative | Positive |
| Urine pH 4.0 | Negative | Positive |
| Urine pH 5.0 | Negative | Positive |
| Urine pH 6.0 | Negative | Positive |
| Urine pH 7.0 | Negative | Positive |
| Urine pH 8.0 | Negative | Positive |
| Urine pH 9.0 | Negative | Positive |
| Urine pH 10.0 | Negative | Positive |
| Urine pH 11.0 | Negative | Positive |
{19}------------------------------------------------
pH – 300 ng/mL Cutoff
Urine samples with pH values from 3.0 to 11.0 were tested in the presence of the two levels of EDDP at ± 25% of the 300 ng/mL cutoff concentration. No interference was observed when tested with the ARK EDDP Assay in both qualitative and semiquantitative modes.
| Spiked EDDP Level | ||
|---|---|---|
| Compound | 225 ng/mL(-25% Cutoff) | 375 ng/mL(+25% Cutoff) |
| Urine pH 3.0 | Negative | Positive |
| Urine pH 4.0 | Negative | Positive |
| Urine pH 5.0 | Negative | Positive |
| Urine pH 6.0 | Negative | Positive |
| Urine pH 7.0 | Negative | Positive |
| Urine pH 8.0 | Negative | Positive |
| Urine pH 9.0 | Negative | Positive |
| Urine pH 10.0 | Negative | Positive |
| Urine pH 11.0 | Negative | Positive |
{20}------------------------------------------------
Method Comparison
A total of one hundred nine (109) unaltered clinical human urine specimens that are not individually identifiable were analyzed for EDDP at the two cutoff levels with the ARK EDDP Assay in both qualitative and semiquantitative modes and the results were compared to GC/MS. The GC/MS confirmatory method was performed by a licensed reference laboratory. Results are summarized in the tables below.
Method Comparison - 100 ng/mL Cutoff
| ARKImmunoassayResult | Low NegativeLess than50% belowthe Cutoff(< 50 ng/mLby GC/MS) | Near CutoffNegativeBetween 50%below theCutoff and theCutoff(50 — 99ng/mL byGC/MS) | Near CutoffPositiveBetween theCutoff and50% abovethe Cutoff(100 - 150ng/mL byGC/MS) | High PositiveGreater than50% abovethe Cutoff(> 150 ng/mLby GC/MS) |
|---|---|---|---|---|
| Negative | 40 | 5 | 0 | 0 |
| Positive | 0 | 0 | 4 | 60 |
Method Comparison – 300 ng/mL Cutoff
| ARKImmunoassayResult | Low NegativeLess than50% belowthe Cutoff(< 150 ng/mLby GC/MS) | Near CutoffNegativeBetween 50%below theCutoff and theCutoff(150 – 299ng/mL byGC/MS) | Near CutoffPositiveBetween theCutoff and50% abovethe Cutoff(300 - 450ng/mL byGC/MS) | High PositiveGreater than50% abovethe Cutoff(> 450 ng/mLby GC/MS) |
|---|---|---|---|---|
| Negative | 49 | 4 | 0 | 0 |
| Positive | 0 | 1* | 3 | 52 |
*Discordant Result
| Sample ID Number | ARK ImmunoassayResult | EDDP(ng/mL by GC/MS) |
|---|---|---|
| 51 | Positive | 294 |
{21}------------------------------------------------
Traceability and Value Assignment
ARK EDDP Calibrators and Controls are prepared by volumetric dilution of high purity EDDP (certified solution traceable to HPLC) into non-sterile, processed human urine free of EDDP. Testing is performed with the ARK EDDP Assay on the Beckman Coulter AU680 automated clinical chemistry analyzer, calibrated with the ARK EDDP Calibrator.
Calibration Curve Stability
A stored calibration curve was effective up to at least 15 days based on supporting data. Calibration curve stability may depend on individual laboratory performance.
807.92 (b)(3): Conclusions from Nonclinical Testing
As summarized above, the ARK EDDP Assay is substantially equivalent to the legally marketed predicate device, Immunalysis EDDP Specific Urine Enzyme Immunoassay (K151395).
§ 862.3620 Methadone test system.
(a)
Identification. A methadone test system is a device intended to measure methadone, an addictive narcotic pain-relieving drug, in serum and urine. Measurements obtained by this device are used in the diagnosis and treatment of methadone use or overdose and to determine compliance with regulations in methadone maintenance treatment.(b)
Classification. Class II (special controls). A methadone test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).