K Number
K182703
Manufacturer
Date Cleared
2019-06-19

(265 days)

Product Code
Regulation Number
866.3990
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EntericBio" Dx assay, performed on ABI 7500 Fast Dx real-time instrument, is an in vitro multiplexed nucleic acid test for the direct, simultaneous, qualitative detection and identification of multiple enteric pathogens in Cary-Blair preserved stool specimens from individuals with signs and symptoms of infectious colitis or gastroenteritis. The test is based on detection of nucleic acids from:

  • . Salmonella enterica spp.
  • Shigella spp./ Enteroinvasive E. coli (EIEC)
  • Campylobacter spp. (jejuni, coli and lari)
  • STEC (Shiga-like toxin-producing E. coli), stx1/stx2 genes
  • Vibrio spp. (cholerae and parahaemolyticus)
  • . Giardia lamblia (also known as G. intestinalis and G. duodenalis)
  • Entamoeba histolytica

Testing is performed on Cary-Blair preserved diarrheal specimens from symptomatic patients with suspected acute gastroenteritis, enteritis or colitis of bacterial or parasitic origin. The test is performed directly on the specimen, utilizing real-time polymerase chain reaction (PCR) for the amplification of Salmonella-specific, Campylobacter-specific, Shigella/ EIEC-specific ipaH, stx1/stx2, Vibrio-specific, Entamoeba-specific and Giardia-specific gene sequences. The test utilizes fluorogenic sequence-specific hybridization probes for the detection of the amplified DNA.

This test is intended for use, in conjunction with clinical presentation, laboratory findings and epidemiological information, as an aid in the diagnosis of Salmonella, Shigella / EIEC, Shigalike toxin-producing E. coli, Campylobacter spp., Entamoeba histolytica and Giardia spp. infections in humans.

Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative EntericBio" Dx assay results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

Device Description

The EntericBio® Dx Assay provides PCR reagents to be used in conjunction with an automated pipetting system and the ABI 7500 Fast Dx instrument using standard filters. Results are interpreted using the EntericBio FastFinder plugin. The system provides automated, real-time amplification, detection and analysis and a user constructed template suitable for the EntericBio® Dx Assay.

The assay is composed of Stool Preparation Solution (SPS) tubes, PCR reagent strips containing lyophilized reagents, Resuspension Buffer (Negative Kit Control), Positive Kit Control containing DNA from all target analytes (with appropriate reconstitution buffer), and associated accessories,instruments and software for detection of bacterial and parasitic causes of gastroenteritis in humans.

The EntericBio" Dx assay detects target DNA from diarrheal Cary-Blair stool specimens from symptomatic individuals with suspected gastroenteritis or infectious colitis.

The assay works directly from a Cary-Blair preserved stool sample and does not require commercial nucleic acid extraction /purification. The PCR master mix with all the reagents required to perform each test is lyophilized into individual reaction wells on a strip. Each reaction well contains an Internal Amplification Control (IAC) to monitor for PCR inhibition.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device's performance, based on the provided FDA 510(k) summary for the Serosep, Ltd. EntericBio® Dx Assay:


Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the performance targets outlined for the analytical and clinical studies. Due to the nature of this 510(k) summary, specific numerical acceptance thresholds for clinical performance (e.g., minimum sensitivity/specificity) are not explicitly stated, but rather the observed performance metrics are presented. The reproducibility study explicitly states acceptance criteria.

Table of Acceptance Criteria and Reported Device Performance

CategoryMetric/TargetAcceptance Criteria (If explicitly stated)Reported Device Performance (EntericBio® Dx Assay)
Analytical Performance
Reproducibility (Accuracy)Agreement with expected results for target analytes (Moderate Positive)100% agreementShigella sonnei: 100% (90/90)Vibrio parahaemolyticus: 100% (90/90)Giardia lamblia: 100% (90/90)
Reproducibility (Accuracy)Agreement with expected results for target analytes (Low Positive)≥95% agreementShigella sonnei: 100% (90/90)Vibrio parahaemolyticus: 100% (90/90)Giardia lamblia: 98% (89/90)
Reproducibility (Precision)Cq Value Coefficient of Variation (%CV)Not explicitly stated as an acceptance criterion for individual Cq values, but presented as a measure of precision.Shigella sonnei: Moderate Positive (2.44% CV), Low Positive (1.97% CV)Vibrio parahaemolyticus: Moderate Positive (1.21% CV), Low Positive (1.82% CV)Giardia lamblia: Moderate Positive (3.53% CV), Low Positive (4.05% CV)
Limit of Detection (LoD)Lowest concentration of analyte consistently detected≥95% detection rateSalmonella spp.: 8 x 10^4 CFU/mLShigella spp./EIEC: 1.25 x 10^4 - 1 x 10^4 CFU/mLCampylobacter spp.: 1 x 10^4 - 4 x 10^4 CFU/mLE. coli (STEC): 5 x 10^5 - 1 x 10^6 CFU/mLVibrio spp.: 1 x 10^4 CFU/mLGiardia lamblia: 25-100 cells/mLEntamoeba histolytica: 25-100 cells/mL
Analytical ReactivityDetect all tested strains within target diversityAll organisms tested detectable over three replicates.All 101 target organisms were detectable.
Analytical ExclusivityNo cross-reactivity with non-target organisms (or limited to closely related organisms with shared sequence similarity)Not explicitly stated, but implies minimal/no false positives from unrelated organisms.Cross-reactivity observed with Vibrio campbellii, V. fluvialis, V. furnissii, V. mimicus, V. fischeri, V. natriegens (all cross-reacted with Vibrio target). Other noted "cross-reactivity" issues were attributed to potential sample contamination or expected in-silico findings for some highly related organisms (e.g., Cryptosporidium viatorum/Shigella, Cryptosporidium cuniculus/Giardia).
Microbial InterferenceNo interference from high concentrations of common microorganisms.No interference in presence of 3X LoD of each target analyte.No interference observed with 11 tested microorganisms.
Interfering SubstancesNo interference from common exogenous/endogenous substances.Detection of targets/IAC in presence of substances.Most substances (21/23) showed no interference. False negative observed for V. parahaemolyticus with hemorrhoidal cream (1/3 replicates) and C. jejuni with tetracycline (2/3 replicates). IAC failures with >1% Benzalkonium chloride and >1% hemorrhoidal cream.
Competitive InhibitionNo competitive inhibition when multiple target analytes are present.No competitive inhibition.Competitive inhibition observed for Giardia lamblia in presence of Entamoeba histolytica. (Note: Not evaluated for Well B analytes: Vibrio and STEC).
Carry-over/Cross-ContaminationNo carry-over/cross-contamination.No carryover or cross-contamination.None observed.
Specimen StabilityDetermine recommended storage conditions.Established storage conditions.5 days at 2-8°C for Cary-Blair preserved stool.
Real Time StabilityDetermine shelf life.Established shelf life.6 months at 2-8°C.
Clinical Performance
Overall Agreement (Positive)Agreement with comparator method for target detection for Fresh, Archived, and Contrived samples. (Individual values depend on analyte)Not explicitly stated, but implied to be high enough for substantial equivalence.Salmonella: Fresh (92.3%), Select (90.0%), Frozen (85.7%). Campylobacter: Fresh (91.1%), Select (90.0%), Frozen (93.8%). Shigella/EIEC: Fresh (100%), Select (100%), Frozen (90.9%). STEC: Fresh (100%), Select (100%), Frozen (94.6%). Vibrio: Fresh (0.0% - due to low prevalence), Simulated (100%). Giardia: Fresh (85.7%), Select (100%), Frozen (100%). Entamoeba: Fresh (NA - no positives), Frozen (0.0% - no positives), Simulated (98.6%).
Overall Agreement (Negative)Agreement with comparator method for non-detection for Fresh, Archived, and Contrived samples. (Individual values depend on analyte)Not explicitly stated, but implied to be high enough for substantial equivalence.Salmonella: Fresh (100%), Select (100%), Frozen (100%). Campylobacter: Fresh (99.7%), Select (100%), Frozen (100%). Shigella/EIEC: Fresh (100%), Select (100%), Frozen (100%). STEC: Fresh (99.9%), Select (100%), Frozen (100%). Vibrio: Fresh (100%), Simulated (100%). Giardia: Fresh (99.9%), Select (100%), Frozen (100%). Entamoeba: Fresh (100%), Frozen (100%), Simulated (100%).
Invalid RateRate of invalid results.Not explicitly stated, but low rate desirable.Fresh: Initial 1.9% (resolved to 0.6%). Frozen: Initial 0.9% (resolved to 0.5%). Total: Initial 1.7% (resolved to 0.6%).
Indeterminate RateRate of indeterminate results.Not explicitly stated, but low rate desirable.Fresh: Initial 0.1% (remained 0.1%). Frozen: Initial 0.9% (remained 0.9%). Total: Initial 0.2% (remained 0.2%).

Study Details:

  1. Sample Sizes Used for the Test Set and Data Provenance:

    • Reproducibility Study (Test Set): Contrived stool samples spiked with various combinations of Vibrio parahaemolyticus, Shigella sonnei, and Giardia lamblia. Each analyte evaluated at True Negative, Low Positive (1.5x LoD), and Moderate Positive (3x LoD).
      • Sample Size: For accuracy, 30 samples per concentration per site (90 total per concentration per analyte across 3 sites). For negative, 60 samples per site (180 total across 3 sites).
      • Data Provenance: In-house (simulating multi-site) from Serosep. Retrospective.
    • LoD Study (Test Set): Contrived samples with target analytes spiked into negative stool matrix.
      • Sample Size: 20 replicates for each sample type/concentration.
      • Data Provenance: In-house, retrospective.
    • Analytical Reactivity/Exclusivity/Interference/Competitive Inhibition/Carry-Over (Test Sets): Contrived samples spiked into negative stool matrix or genomic DNA/RNA.
      • Sample Size: Generally 3 replicates for analytical reactivity/exclusivity/interference/carry-over. Competitive inhibition tested in triplicate.
      • Data Provenance: In-house, retrospective.
    • Clinical Performance Study (Test Set):
      • Fresh Clinical Samples (Prospective): 1491 prospective, 32 select. Total 1523 enrolled, 1472 evaluable.
        • Data Provenance: Multi-site clinical study, including two (2) US clinical sites and one (1) non-US clinical site. Prospective collection from symptomatic patients.
      • Frozen Clinical Samples (Archived): 212 enrolled, 209 evaluable.
        • Data Provenance: Archived samples from unknown collection sites, selected based on previous positive testing. Retrospective.
      • Contrived Samples (Simulated): 310 evaluable samples.
        • Data Provenance: Prepared using residual fresh specimens that tested negative. Retrospective, in-house.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • For the EntericBio® Dx Assay, the ground truth for the clinical study was established by comparator methods, not directly by human experts classifying images/data. The summary does not specify the number or qualifications of experts involved in running or interpreting the results of the FDA-cleared comparator assays, or for establishing the initial diagnosis for the archived samples. In diagnostics studies like this, the "ground truth" is typically another diagnostic method (often a laboratory reference method or a composite of multiple methods), which itself is run by trained laboratory personnel.
  3. Adjudication Method for the Test Set:

    • For STEC and Campylobacter in fresh specimens, a composite comparator method was used: a specimen was positive if 2 out of 3 comparator assays were positive, and negative if 2 out of 3 were negative.
    • For all other target analytes and archived specimens, the comparator method consisted of one FDA-cleared assay.
    • This implies an adjudication method for the composite comparator (2/3 rule) but not for single comparator methods. There is no mention of human expert adjudication post-test to resolve discrepancies.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is a molecular diagnostic assay (PCR-based), not an AI-assisted diagnostic imaging device that involves human "readers" interpreting images. Therefore, the concept of human readers improving with AI assistance is not applicable here.
  5. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The EntericBio® Dx Assay is a standalone diagnostic test in the sense that its performance is evaluated directly against comparator methods. It does not inherently involve a "human-in-the-loop" decision-making process in the same way an AI-CAD (Computer-Aided Detection) system for radiology would. The device provides a result (positive/negative) based on its PCR analysis, and clinical interpretation occurs downstream by lab personnel and clinicians. The clinical performance tables (e.g., Table 12a-g) represent this standalone performance of the assay.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for the clinical performance study was established by comparator methods, specifically:
      • For STEC and Campylobacter (fresh specimens): A composite comparator method of three FDA-cleared assays (2/3 rule).
      • For other targets (fresh) and all archived specimens: One FDA-cleared assay.
      • For contrived samples: The known spike concentration/strain was the ground truth.
  7. The Sample Size for the Training Set:

    • This document is a 510(k) summary for an in vitro diagnostic (IVD) assay based on PCR, not a machine learning/AI model. Therefore, the concept of a "training set" for an algorithm is not directly applicable in the same way it would be for AI/ML device development. The device's design and optimization (analogous to "training") would have occurred during its development phase, using proprietary methods, and is not explicitly detailed as a distinct 'training set' of clinical samples in this regulatory submission.
  8. How the Ground Truth for the Training Set was Established:

    • As explained above, this product is not an AI/ML algorithm requiring a "training set" in the conventional sense. The "ground truth" for the analytical studies (e.g., LoD, exclusivity) was based on precisely prepared samples with known concentrations and identities of microorganisms. For the development and validation of the molecular targets (primers/probes), ground truth would have been established through extensive molecular biology techniques, including sequencing and controlled lab cultures. These methods are inherent to the development of robust PCR assays.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 19, 2019

Serosep, Ltd. % Fran White MDC Associates, LLC 180 Cabot Street Beverly, Massachusetts 01915

Re: K182703

Trade/Device Name: EntericBio Dx Assay Regulation Number: 21 CFR 866.3990 Regulation Name: Gastrointestinal microorganism multiplex nucleic acid-based assay Regulatory Class: Class II Product Code: PCH, OOI, NSU Dated: September 26, 2018 Received: September 27, 2018

Dear Fran White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190121

Device Name IDS SHBG

Indications for Use (Describe)

The IDS SHBG assay is an in vitro diagnostic device intended for the quantitative determination of SHBG in human serum or plasma on the IDS System. Results are to be used as an aid in the diagnosis of androgen disorders

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date of Summary: June 18, 2019

Sponsor

Serosep, Ltd. Annacotty Business Park Annacotty, Limerick, Ireland

Correspondent

MDC Associates, Inc 180 Cabot Street Beverly, MA 01915 Contact : Fran White Phone : (978) 705 5011 Fax : (978) 927 1308 Email : fran@mdcassoc.com

Device Trade or Proprietary Name

EntericBio® Dx Assay

Common Name

Gastrointestinal microorganism multiplex nucleic acid-based assay

Product Classification 866.3990

Classification

PCH, Class II

Substantial Equivalency

Serosep, Ltd. believes that the subject devices (new device) of this pre-submission document and subsequently a premarket notification submission (510k) is similar to other molecular devices currently marketed in the US. The device design, features and performance when compared to these devices is similar to BioFire, FilmArray Gastrointestinal (GI) Panel (K140407) in that the intended use, the targeted organism for detection, the analytes, technological principles, and specimen types are similar or the same.

There are differences between the subject new device and the predicate device. Because of these differences, the subject device features and suitability will be validated for its intended use using clinical specimens sourced from symptomatic patients or clinical library specimens and tested by the predicate device.

The similarities and differences between the subject device and the predicate device are summarized below.

{4}------------------------------------------------

ElementSubject (New) DeviceProposed Predicate DeviceDiscussion
Device NameEntericBio® Dx AssayFilmArray Gastrointestinal (GI) PanelNew vs. Predicate Device
FDA DevicePremarketNotificationNew deviceK140407New vs Predicate Device
FDA DeviceClassificationClass II, 21 CFR 866.3990 –Gastrointestinal microorganismmultiplex nucleic acid-based assay,Microbiology (83 Panel)Class II, 21 CFR 866.3990 –Gastrointestinal microorganismmultiplex nucleic acid-based assay,Microbiology (83 Panel)Same
Type of TestQualitative nucleic acid testQualitative nucleic acid testSame
UsersCLIA certified clinical laboratoriesCLIA certified clinical laboratoriesSame
Assay MethodThe EntericBio® Dx Assayprovides PCR reagents to be used inconjunction with an automatedpipetting system and the ABI 7500 FastDx instrument using standard filters.Results are interpreted using theEntericBio FastFinder plugin. The systemprovides automated, real-timeamplification, detection and analysisand a user constructed templatesuitable for the EntericBio® Dx Assay.The BioFire FilmArray GastrointestinalPanel is designed to be used with theFilmArray® instrument. The FilmArrayGl pouch contains freeze-driedreagents to perform nucleic acidpurification and nested, multiplex PCRwith DNA melt analysis.Different: New device differs fromthe predicate device in that therealtime PCR assay kits aredesigned to be used with anautomated pipetting station toaccelerate sample preparation,and a commercially available, FDA-cleared PCR instrument tomeasure the fluorescent probesignal generated duringamplification that are analyzed bythe EntericBio automated analysisand interpretation software.
Targets forDetectionSalmonella enterica spp.Shigella spp./ Enteroinvasive E. coli(EIEC)Campylobacter spp. (jejuni, coli and lari)Clostridium difficile ( C. difficile ) toxinA/B, Campylobacter spp. ( C. jejuni, C.coli and C. upsaliensis ), Plesiomonasshigelloides, Salmonella spp., Vibriospp., Yersinia enterocolitica.Similar, except the predicatedevice is additionally cleared fordetection of a range of othernucleic acid targets from bacteria,parasites and viruses.
ElementSubject (New) DeviceProposed Predicate DeviceDiscussion
STEC (Shiga toxin-producing E. coli),stx1/stx2 genesVibrio spp. (cholerae andparahaemolyticus )Giardia lamblia (also known as G. intestinalis and G. duodenalis )Entamoeba histolyticaEnteriaggregative E. coli ,Enteropathogenic E. coli ,Enterotoxigenic E. coli LT/ST toxins,Shiga-like toxin-producing E. coli ,Shigella/ Enteroinvasive E. coli ,Cryptosporidium spp. , Cyclospora cayetanensis , Entamoeba histolytica ,Giardia lamblia , Adenovirus F40/41,Astrovirus, Norovirus GI/GII, Rotavirus A, Sapovirus (GI, GII, GIV and GV).
Intended UseThe EntericBio® Dx Assay performed onABI 7500 Fast Dx real-time instrument,is an in vitro multiplexed nucleic acidtest for the direct, simultaneousqualitative detection and identificationof multiple enteric pathogens in Cary-Blair preserved stool specimens fromindividuals with signs and symptoms ofinfectious colitis or gastroenteritis. Thetest is based on detection of nucleicacids from the following organisms: (seeorganisms above).The FilmArray Gastrointestinal Panel(GI) is intendent for use with theFilmArray® instrument for thequalitative in vitro detection andidentification of multiple bacteria,viruses and parasites. The FilmArray GIPanel is performed directly from stoolspecimens in Cary-Blair transportmedia. The following pathogen types,subtypes and toxin genes areidentified using the FilmArray GI Panel:(see organisms above).Similar except for the additionalorganisms detected by thepredicate device.
AnalyteDNA/RNA from Cary-Blair preservedfecal specimensDNA/ RNA from Cary-Blair preservedfecal specimensSame
TechnologicalPrinciplesMultiplex nucleic acid PCRMultiplex nucleic acid PCRSame
SpecimenTypesHuman stool (Cary-Blair preserved)Human stool (Cary-Blair preserved)Same
ElementSubject (New) DeviceProposed Predicate DeviceDiscussion
ControlsInternal Amplification Control for eachsample. Kit positive and negativecontrols are processed with each batchof samples.Two controls are included in eachreagent pouch to control for sampleprocessing and both stages of PCR andmelt analysis.Similar: The new device IAC islyophilized within the PCR mixwhereas the predicate device hastwo controls included in eachreagent pouch.Substantial equivalence will bedemonstrated in clinical testingusing human specimens andcompared to the predicate device.
PCR SamplePreparation/ExtractionSample processed directly followingheat treatment of specimen in a SPStube.Sample processing is automated in theFilmArray® instrument. The sample islysed by a combination of mechanical(bead beating) and chemical meansand the liberated nucleic acid iscaptured, washed and eluted usingmagnetic bead technology.EntericBio® Dx kit providesreagents and procedure for DNAtesting without extraction of stoolspecimens.Substantial equivalence will bedemonstrated in clinical testingusing human specimens andcompared to the predicate device.
TechnologicalPrinciplesReal-time multiplex RT-PCR based onthe hydrolysis probe reagent chemistry.Nested multiplex RT-PCR followed byhigh resolution melting analysis toconfirm identity of the PCR product.Different: Both devices usemultiplex real-time PCR howeverthe new device uses hydrolysisprobe reagent chemistrycompared to melting analysis inthe predicate device.Substantial equivalence will bedemonstrated in clinical testingusing human specimens andcompared to the predicate device.
ElementSubject (New) DeviceProposed Predicate DeviceDiscussion
DetectionMethodology/PlatformThe Applied Biosystems 7500 Fast DxReal-Time PCR instrument is a real-timenucleic acid amplification and five colorfluorescence detection system for usewith the EntericBio® Dx Assay. Resultsare analyzed and interpreted using theEntericBio® FastFinder plugin.Detection and interpretation isautomated on the FilmArray®instrument by analysis of the specificPCR product melts (meltingtemperature).Different: The detection systemand analysis differ between thetwo devices.Substantial equivalence will bedemonstrated in clinical testingusing human specimens andcompared to the predicate device.
Device FormatThe EntericBio® Dx Assay kits provide aPCR master mix with all the reagentsrequired to perform each test which arelyophilized into individual reactionwells. Each reaction well contains anInternal Amplification Control (IAC) tomonitor for PCR inhibition. A syntheticPositive Control (containing targetsequences) is provided with each kit tomonitor the thermal cycling steps andreagent integrity during amplificationand detection process.The FilmArray Gl panel provides all thereagents lyophilized into a disposablepouch. Each pouch contains twocontrols which monitor sampleprocessing and both stages of PCR andmelt analysis.Assay specific requirements

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Intended Use

The EntericBio" Dx assay, performed on ABI 7500 Fast Dx real-time instrument, is an in vitro multiplexed nucleic acid test for the direct, simultaneous, qualitative detection and identification of multiple enteric pathogens in Cary-Blair preserved stool specimens from individuals with signs and symptoms of infectious colitis or gastroenteritis. The test is based on detection of nucleic acids from:

  • . Salmonella enterica spp.
  • Shigella spp./ Enteroinvasive E. coli (EIEC)
  • Campylobacter spp. (jejuni, coli and lari)
  • STEC (Shiga-like toxin-producing E. coli), stx1/stx2 genes
  • Vibrio spp. (cholerae and parahaemolyticus)
  • . Giardia lamblia (also known as G. intestinalis and G. duodenalis)
  • Entamoeba histolytica

Testing is performed on Cary-Blair preserved diarrheal specimens from symptomatic patients with suspected acute gastroenteritis, enteritis or colitis of bacterial or parasitic origin. The test is performed directly on the specimen, utilizing real-time polymerase chain reaction (PCR) for the amplification of Salmonella-specific, Campylobacter-specific, Shigella/ EIEC-specific ipaH, stx1/stx2, Vibrio-specific, Entamoeba-specific and Giardia-specific gene sequences. The test utilizes fluorogenic sequence-specific hybridization probes for the detection of the amplified DNA.

This test is intended for use, in conjunction with clinical presentation, laboratory findings and epidemiological information, as an aid in the diagnosis of Salmonella, Shigella / EIEC, Shigalike toxin-producing E. coli, Campylobacter spp., Entamoeba histolytica and Giardia spp. infections in humans.

Results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative EntericBio" Dx assay results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.

Methodology

The EntericBio" Dx assay is a molecular in vitro diagnostic test for direct, qualitative detection and identification of the following enteric organisms, associated with human gastroenteritis, directly, from Cary-Blair preserved fecal specimens:

  • . Salmonella enterica spp.,
  • Shigella spp./ Enteroinvasive E. coli,
  • Campylobacter spp. (jejuni, coli and lari),
  • Vibrio spp. (cholerae and parahaemolyticus), ●

{9}------------------------------------------------

  • STEC(Shiga-like toxin-producing E. coli) stx1/stx2 genes,
  • Giardia lamblia (also known as G. intestinalis and G. duodenalis),
  • Entamoeba histolytica

The assay is composed of Stool Preparation Solution (SPS) tubes, PCR reagent strips containing lyophilized reagents, Resuspension Buffer (Negative Kit Control), Positive Kit Control containing DNA from all target analytes (with appropriate reconstitution buffer), and associated accessories,instruments and software for detection of bacterial and parasitic causes of gastroenteritis in humans.

The EntericBio" Dx assay detects target DNA from diarrheal Cary-Blair stool specimens from symptomatic individuals with suspected gastroenteritis or infectious colitis.

The assay works directly from a Cary-Blair preserved stool sample and does not require commercial nucleic acid extraction /purification. The PCR master mix with all the reagents required to perform each test is lyophilized into individual reaction wells on a strip. Each reaction well contains an Internal Amplification Control (IAC) to monitor for PCR inhibition.

Performance Data:

Analytical Performance

Reproducibility

A reproducibility study was performed to determine the inter-site and overall reproducibility of the EntericBio® Dx assay. Reproducibility testing was performed in-house, simulating a multi-site study by using three instruments (hereinafter 'in-house sites') and panels of contrived stool samples, each spiked with various combinations of Vibrio parahaemolyticus, Shigella sonnei and Giardia lamblia. These three target analytes were representative of each of the three component multiplex assays and each analyte was evaluated at three different concentrations (True Negative, Low Positive and Moderate Positive) in three independently manufactured batches of EntericBio® Dx Assay. Reproducibility panels were tested at each of the in-house sites by two different operators for five non-consecutive days.

The acceptance criteria for this study were that the moderate positive target concentration (3x LoD) must show 100% agreement with the expected result for each target analyte and the low target concentration (1.5x LoD) must show ≥95% agreement with the expected result for each target analyte.

All targets showed 100% agreement with the expected result for moderate positive concentrations tested across the in-house sites. Shigella sonnei and Vibrio parahaemolyticus samples showed 100% agreement with the expected result for low positive concentrations tested the in-house sites. Giardia lamblia at the low positive concentration was observed at 98% relative to the expected result. All acceptance criteria for each target analyte at each concentration tested across the in-house sites were met, as shown in Tables 1-2 below.

{10}------------------------------------------------

Organism TestedConcentrationTestedFastFinderExpectedResult% Agreement with Expected ResultAll
Serosep1Serosep2Serosep3
Shigella sonneiDSM 5570ModeratePositive3x LoDPositive30/30100%30/30100%30/30100%90/90100%
Low Positive1.5x LoDPositive30/30100%30/30100%30/30100%90/90100%
True NegativeNegative60/60100%60/60100%59/59*100%179/179100%
VibrioparahaemolyticusCCUG 14474ModeratePositive3x LoDPositive30/30100%30/30100%30/30100%90/90100%
Low Positive1.5x LoDPositive30/30100%30/30100%30/30100%90/90100%
True NegativeNegative60/60100%60/60100%59/59*100%179/179100%
Giardia lambliaP101ModeratePositive3x LoDPositive30/30100%30/30100%30/30100%90/90100%
Low Positive1.5x LoDPositive30/30100%30/30100%29/3095%89/9098%
True NegativeNegative60/60100%60/60100%59/59*100%179/179100%

Table 1: Reproducibility study results

  • One True Negative sample was invalid on the EntericBio FastFinder plugin and subsequently removed from study

{11}------------------------------------------------

Organism TestedAssayConcentrationTestedTest SiteCq Reproducibility
Mean(Cq)STDEV%CV
Shigella sonneiDSM 5570Well AModeratePositiveSerosep 127.10$\pm$ 0.682.51
Serosep 226.84$\pm$ 0.471.74
Serosep 327.01$\pm$ 0.792.91
3x LoDAll Sites26.98$\pm$ 0.662.44
Low PositiveSerosep 129.07$\pm$ 0.571.97
Serosep 228.88$\pm$ 0.531.85
Serosep 328.95$\pm$ 0.571.98
1.5x LoDAll Sites28.97$\pm$ 0.571.97
VibrioparahaemolyticusCCUG 14474Well BModeratePositiveSerosep 131.73$\pm$ 0.421.34
Serosep 231.87$\pm$ 0.240.74
Serosep 331.74$\pm$ 0.441.40
3x LoDAll Sites31.79$\pm$ 0.381.21
Low PositiveSerosep 133.50$\pm$ 0.481.44
Serosep 233.34$\pm$ 0.501.51
Serosep 333.30$\pm$ 0.792.37
1.5x LoDAll Sites33.38$\pm$ 0.611.82
Giardia lambliaP101Well CModeratePositiveSerosep 132.96$\pm$ 0.852.58
Serosep 233.04$\pm$ 1.364.12
Serosep 333.30$\pm$ 1.273.81
3x LoDAll Sites33.10$\pm$ 1.173.53
Low PositiveSerosep 133.98$\pm$ 1.183.49
Serosep 234.40$\pm$ 1.384.01
Serosep 333.11$\pm$ 1.263.80
1.5x LoDAll Sites33.85$\pm$ 1.374.05

Table 2: Summary of the Reproducibility of the Cq Values

Limit of Detection (LoD)

The analytical sensitivity (limit of detection or LoD) of the EntericBio® Dx Assay was determined using contrived samples with target analytes spiked into a negative stool matrix (Cary-Blair preserved stool) at three concentrations: greater than estimated LoD (High), estimated LoD (Medium) and less than LoD (Low). A total of 20 replicate EntericBio® Stool Preparation Solution (SPS) were tested from each sample using three independently manufactured lots of EntericBio® Dx assay. The LoD is defined as the lowest concentration of analyte that can be consistently detected ≥95% of the time. A minimum of two strains were tested for each EntericBio® Dx target organism and toxin gene. Table 3 lists the LoD determined for each target organism.

{12}------------------------------------------------

OrganismStrainLoD
Salmonella spp.Salmonella enterica serovar Enteritidis DSM 17420$8 x 10^4$ CFU/mL
Salmonella enterica serovar Typhi NCTC 10787$8 x 10^4$ CFU/mL
Shigellaspp./EIECShigella sonnei DSM 55701.25 x $10^4$ CFU/mL
E. coli (ipaH) DSM 90291 x $10^4$ CFU/mL
Campylobacterspp.Campylobacter jejuni ATCC 335604 x $10^4$ CFU/mL
Campylobacter coli DSM 46894 x $10^4$ CFU/mL
Campylobacter lari DSM 113751 x $10^4$ CFU/mL
Escherichia coli(STEC)E. coli (stx1) NVRL 15x23 RE-008 (0111:H-)5 x $10^5$ CFU/mL
E. coli (stx2) NVRL 15x24 RE-006 (O26:H11)1 x $10^6$ CFU/mL
Vibrio spp.Vibrio parahaemolyticus CCUG 144741 x $10^4$ CFU/mL
Vibrio cholerae NCTC 36611 x $10^4$ CFU/mL
Giardia lambliaGiardia intestinalis (WB) ATCC 3095725 cells/mL
Giardia intestinalis (New Orleans) ATCC 50137100 cells/mL
EntamoebahistolyticaEntamoeba histolytica (HM-1: IMSS) ATCC 3045925 cells/mL
Entamoeba histolytica (HK-9) ATCC 30015100 cells/mL

Table 3: LoD determined using the EntericBio® Dx Assay

Fresh vs. Frozen

The Fresh versus Frozen Specimen Stability study was performed to support the inclusion of frozen, retrospective specimens and contrived samples in the clinical and analytical studies of the EntericBio® Dx assay; the test is not intended for use on frozen specimens.

Sixty contrived samples were prepared for each EntericBio® Dx target analyte and tested at Time 0 (TO). These samples were subsequently frozen at -20°C for 3 months before re-testing. Contrived samples were prepared in a negative stool matrix (Cary-Blair preserved stool) with target analytes spiked at three concentrations (5X, 2X and 1X LoD).

Agreement between detection of fresh and frozen samples was 100% for Salmonella, Shigella and STEC and <80% for four analytes (Campylobacter, Vibrio, Giardia and Entamoeba). Additional testing was performed using frozen clinical specimens from the prospective study to further support the inclusion of these four analytes in clinical studies.

Analytical Reactivity/Inclusivity

The analytical reactivity (Inclusivity) of the EntericBio® Dx Assay was determined using a panel of 101 target organisms (Table 4 below) which represents the diversity of the EntericBio® Dx target analytes. Fifteen of these organisms were also evaluated in the LoD determination study.

Table 4: EntericBio® target analytes used in this study

{13}------------------------------------------------

OrganismSupplierCatalogue Number
Salmonella enterica serovar EnteritidisDSMZ1DSM 17420
Salmonella enterica serovar TyphiNCTC2NCTC 10787
Salmonella enterica subsp. enterica I serovarCholerasuisATCC3ATCC 7001
Salmonella enterica subsp. enterica I serovarParatyphi BATCCATCC 8759
Salmonella enterica subsp. enterica I serovarParatyphi AATCCATCC 9281
Salmonella enterica subsp. enterica I serovarParatyphi CATCCATCC 13428
Salmonella enterica subsp. enterica I serovarTyphimuriumDSMZDSM 101475
Salmonella enterica subsp. enterica I serovarDublinDSMZDSM 102345
Salmonella enterica subsp. enterica I serovarAgonaDSMZDSM 102864
Salmonella enterica subsp. enterica I serovarHeidelbergDSMZDSM 9379
Salmonella enterica subsp. enterica I serovarInfantisNCTCNCTC 10679
Salmonella enterica subsp. enterica I serovarThompsonNCTCNCTC 2252
Salmonella enterica subsp. enterica I serovarOranienburgNCTCNCTC 5743
Salmonella enterica subsp. enterica I serovarBareillyNCTCNCTC 5745
Salmonella enterica subsp. enterica I serovarMontevideoNCTCNCTC 5747
Salmonella enterica subsp. enterica I serovarBraenderupNCTCNCTC 5750
Salmonella enterica subsp. enterica I serovarMuenchenNCTCNCTC 5755
Salmonella enterica subsp. enterica I serovarSaintpaulNCTCNCTC 6022
Salmonella enterica subsp. enterica I serovarMississippiNCTCNCTC 6487
Salmonella enterica subsp. enterica I serovarJavianaNCTCNCTC 6495
Salmonella enterica subsp. enterica I serovarNewportNCTCNCTC 6704
Salmonella enterica subsp. enterica I serovarSchwarzengrundNCTCNCTC 6756
Salmonella enterica subsp. enterica I serovarHadarNCTCNCTC 9877
Salmonella enterica subsp. enterica I serovar 4,[5] 12:i:-NSSLRL5NSSLRL Ms170397
Salmonella enterica subsp. II (salame)DSMZDSM 9220
OrganismSupplierCatalogue Number
Salmonella enterica subsp. Illa (arizonae)DSMZDSM 9386
Salmonella enterica subsp. Illb (diarizonae)DSMZDSM 14847
Salmonella enterica subsp. IV (houtenae)DSMZDSM 9221
Salmonella enterica subsp. VI (indica)DSMZDSM 14848
Shigella sonneiDSMZDSM 5570
Shigella sonneiDSMZDSM 25715
Shigella sonneiATCCATCC 11060
Shigella sonneiATCCATCC 25931
Shigella sonneiATCCATCC 9290
Shigella flexneri (serotype 2a)DSMZDSM 4782
Shigella flexneri (serotype 2a)ATCCATCC 700930
Shigella flexneri (serotype 1a)ATCCATCC 9199
Shigella flexneri (serotype 2b)ATCCATCC 12022
Shigella flexneri (serotype 6)ATCCATCC 12025
Shigella boydii (serotype 2)DSMZDSM 7532
Shigella boydii (serotype 1)ATCCATCC 9207
Shigella boydii (serotype 20)ATCCATCC BAA-1247
Shigella boydii (serotype 10)ATCCATCC 12030
Shigella boydii (serotype 4)ATCCATCC 9210
Shigella dysenteriae (serotype 1)NCTCNCTC 4837
Shigella dysenteriae (serotype 2)NCTCNCTC 5109
Shigella dysenteriae (serotype 7)NCTCNCTC 9763
Shigella dysenteriae (serotype 3)NCTCNCTC 6340
Shigella dysenteriae (serotype 9)NCTCNCTC 9347
Escherichia coli EIEC (serotype O28ac:H-)DSMZDSM 9025
Escherichia coli EIEC (serotype O29:H10)DSMZDSM 9026
Escherichia coli EIEC (serotype 0136:H-)DSMZDSM 9032
Escherichia coli EIEC (serotype 0124:H30)DSMZDSM 9031
Escherichia coli EIEC (serotype O144:H-) (ipaH)8DSMZDSM 9029
Campylobacter jejuniATCCATCC 33560
Campylobacter jejuni subsp. jejuniDSMZDSM 104743
Campylobacter jejuni subsp. jejuniDSMZDSM 27585
Campylobacter jejuni subsp. doyleiDSMZDSM 104768
Campylobacter jejuni subsp. doyleiNCTCNCTC 12208
Campylobacter coliDSMZDSM 110395
Campylobacter coliDSMZDSM 24155
Campylobacter coliDSMZDSM 24106
Campylobacter coliDSMZDSM 24206
Campylobacter coliDSMZDSM 4689
Campylobacter lariDSMZDSM 11375
Campylobacter lariNCTCNCTC 12892
Campylobacter lariNCTCNCTC 12893
Campylobacter lariNCTCNCTC 12894
Campylobacter lariNCTCNCTC 12895
Vibrio parahaemolyticusDSMDSM 101031
OrganismSupplierCatalogue Number
Vibrio parahaemolyticusDSMDSM 11058
Vibrio parahaemolyticusDSMDSM 15477
Vibrio parahaemolyticusDSMDSM 27657
Vibrio parahaemolyticusCCUG6CCUG 14474
Vibrio cholerae (0:1 Ogawa classical)NCTCNCTC 3661
Vibrio cholerae 0:1 Biotype El TorNCTCNCTC 8457
Vibrio cholerae 0:1 OgawaNCTCNCTC 8021
Vibrio cholerae non-O:1, non-0139 (0:3)NCTCNCTC 11502
Escherichia coli 0157 (stx2)NVRL4NVRL 17X01 RE-001
Escherichia coli 0157 (stx1 & stx2)NVRLNVRL17X04 RE-002
Escherichia coli 0157 (stx2)NVRLNVRL 17X09 RE-003
Escherichia coli O157 (stx1 & stx2)NVRLNVRL 17X15 RE-004
Escherichia coli 0157 (stx2)NVRLNVRL 17S110 RE-005
Escherichia coli 0103:H2 (stx1)NVRLNVRL 17X128 RE-010
Escherichia coli 0111:H8 (stx1)NVRLNVRL 13S5371 RE-009
Escherichia coli O121:H19 (stx2)NVRLNVRL 15X18 RE-007
Escherichia coli O157:NMNVRLNVRL 06-CC3 RE-011
Escherichia coli O157:H7 (stx1 & stx2)NCTCNCTC 12079 NVRLRE-012
Escherichia coli O157:H- (stx2)NCTCNCTC 12080
Escherichia coli 0111:H- (stx1)NVRLNVRL 15x23 RE-008
Escherichia coli O26:H11 (stx2)NVRLNVRL 15x24 RE-006
Escherichia coli 0113 7N/AN/A
Escherichia coli 045 7N/AN/A
Escherichia coli O104 7N/AN/A
Escherichia coli 0145 7N/AN/A
Giardia intestinalis (WB)ATCCATCC 30957
Giardia intestinalis (New-Orleans-1)ATCCATCC 50137
Giardia intestinalis GS Assemblage BATCCATCC 50581
Giardia intestinalis Portland 1ATCCATCC 30888
Giardia intestinalis MarioATCCATCC PRA-244
Entamoeba histolytica (HM-1: IMSS)ATCCATCC 30459
Entamoeba histolytica (HK-9)ATCCATCC 30015
Entamoeba histolytica HB-301:NIHATCCATCC 30190
Entamoeba histolytica HU-21:AMCATCCATCC 30457
Entamoeba histolytica IP:1182;2ATCCATCC PRA-357

{14}------------------------------------------------

{15}------------------------------------------------

1 DSMZ - Deutsche Sammlung von Mikroorganismen und Zenllkulturen

2 NCTC - National Collection of Type Cultures, a Culture Collection of Public Health England

3 ATCC – American Type Culture Collection

4 NVRL - National VTEC Reference Laboratory, Ireland

  • 5 NSSLRL- National Salmonella, Shigella and Listeria Reference Lab, Galway, Ireland
  • 6 CCUG Culture Collection University of Gothenburg
  • 7 Inclusivity predicted based on in-silico analysis

8 EIEC strain which also generates positive result for Shigella/EIEC (ipaH)

{16}------------------------------------------------

Testing was performed using contrived samples with target analytes spiked into a negative stool matrix (Cary-Blair preserved stool) at a concentration of approximately three times the respective LoD, where possible.

All organisms tested were detectable over three replicates demonstrating the inclusivity of the EntericBio® Dx assay.

Analytical Exclusivity

The analytical reactivity (Exclusivity) of the EntericBio® Dx Assay was determined using a comprehensive panel of 133 bacteria, viruses and parasites (Table 5 below). The panel of organisms tested consisted of organisms closely related to the EntericBio® Dx assay targets and organisms likely to be found in human feces.

OrganismOrganismOrganismOrganism
Yersinia aldovaeCryptosporidiummeleagridisCitrobacter koseriPseudomonas putida1
Yersinia bercovieriCryptosporidium Skunkgenotype1Clostridium difficileRahnella aquatilis
YersiniaentomophagaCryptosporidiumubiquitum1Clostridium perfringensRotavirus A1
Yersinia frederikseniiCryptosporidiumviatorum1, 2Clostridium sordelliRuminococcus gauvreauii
Yersinia intermediaEscherichia vulnerisCronobacter sakazakiiSaccharomyces cerevisiae
Yersinia kristenseniiCryptosporidiumcuniculus1,3Dientamoeba fragilisSapovirus1
Yersinia massiliensisCryptosporidium Horsegenotype1Edwardsiella tardaSerratia liquefaciens
Yersinia mollarettiCryptosporidium felis1Encephalitozoon cuniculi1Serratia marcescens
Yersinia nurmiiCryptosporidium canis1Enterobacter aerogenesSerratia odoriferae
Yersinia pekkaneneiiCryptosporidium xiaoi1, 4Enterobacter cloacaeSerratia rubidaea
YersiniapseudotuberculosisCryptosporidiumandersoni1Enterococcus faecalisStaphylococcus aureus
Yersinia rodheiCryptosporidium baileyi1Enterococcus faeciumStaphylococcus epidermidis
Yersinia ruckeriEntamoeba dispar5Eubacterium rectale1, 6Stenotrophomonasmaltophilia
Yersinia similisEntamoebamoshkovoskiiEwingella americanaStreptococcus agalactiae
Vibrio alginolyticusEntamoeba invadensFusobacteriumgonidiaformansStreptococcus bovis
Vibrio fluvialisAeromonas hydrophiliaFusobacterium nucleatumStreptococcus equinus
OrganismOrganismOrganismOrganism
Vibrio furnissiiAdenovirus F401Fusobacterium variumToxoplasma gondii
Vibrio mimicusAlcaligenes faecalisHafnia alveiCampylobacter concisus1
Vibrio harveiAnaerococcushydrogenalisKlebsiella oxytocaCampylobacter curvus1
Vibrio fischeri1Anaerostipes hadrus7Klebsiella pneumoniaeCampylobacter fetus subsp.Fetus
Vibrio damselaArcobacter butzleriLactobacillus acidophilusCampylobacter fetus subsp.Venerealis
Griomontia hollisaeAstrovirus1Lactobacillus lactisCampylobacter gracilis
Vibrio diazotrophicusBacillus cereusListeria monocytogenesCampylobacter helveticus1
Vibrio proteolyticusBacillus subtilis subspsubtilisMorganella morganiiCampylobacter hominis
Vibrio natrigensBacillus subtilis subspspizizeniiNeisseria gonorrhoeaeCampylobacterhyointestinalis
Vibrio pelagius1Bacteroides fragilisNorovirus GGI1Campylobacter mucosalis1
Vibrio campbelliiBacteroidesthetaiotaomicronNorovirus GGII1Campylobacter rectus1
Vibrio vulnificus8Bifidobacterium brevePlesiomonas shigelloidesCampylobacter showae
Escherichia coli nontoxigenicBifidobacterium longumPrevotella melaninogenicaCampylobacter sputorum1,9
Escherichia coli(Enteropathogenic)Blastocystis hominisProteus mirabilisCampylobacter ureolyticus
Escherichia coli(Enterotoxigenic)Candida albicansProteus vulgarisSalmonella bongori
EscherichiahermanniiCitrobacteramalonaticus1Providencia stuartiiSalmonella subterranea
Escherichia blattaeCitrobacter freundiiPseudomonas aeruginosaCampylobacter upsaliensis

Table 5: Microorganisms used in this study

{17}------------------------------------------------

Escherichia fergusonii

1 Organisms for which testing was performed using genomic DNA

3 Cryptosporidium viatorum: Shiqella detected in 3/3 replicates. Bi-directional sequencing demonstrated the presence of Shigella DNA suggesting possible sample contamination with Shigella. In silico analysis indicated no cross-reactivity should occur.

3 Cryptosporidium cuniculus: Giardia detected in 2/3 replicates. Bi-directional sequencing of amplicon failed and thus it could not be confirmed empirically if the sample was contaminated with Giardia or if cross reactivity-occurred. In silicated no crossreactivity should occur.

4 Cryptosporidium xiaoi: Campylobacter detected in 3/3 replicates. Bi-directional sequencing demonstrated the presence of Campylobacter DNA suggesting possible sample contamination with Compylobacter. In silico analysis indicated no cross-reactivity should occur.

5 Entamoeba dispar: IAC failures were observed for 5/6 replicates after repeat testing. In silico analysis demology for the primers and predicts amplification to occur and competitively inhibit the IAC, but the probe has several to produce a detected signal.

6 Eubacterium rectale: Campylobacter detected in 3/3 replicates. Bi-directional sequencing demonstrated the presence of Campylobacter and Vibrio DNA suggesting possible sample contamination with Vibrio and Campylobacter. In silico analysis indicated no cross-reactivity should occur.

7 Anaerostipes hadrus: Campylobacter detected in 2/3 replicates. Bi-directional sequencing demonstrated the presence of Campylobacter DNA suggesting possible sample contamination with Campylobocter. In silico analysis indicated no cross-reactivity should occur. 8 Vibrio vulnificus: Evaluated by in silico analysis indicated no cross-reactivity should occur.

9 Campylobacter sputorum: Vibrio detected in 1/3 replicates. Bi-directional sequencing demonstrated the presence of Vibrio DNA suggesting possible sample contamination with Vibrio. In silico analysis indicated no cross-reactivity should occur.

{18}------------------------------------------------

Testing was performed using contrived samples with target analytes spiked into a negative stool matrix (Cary-Blair preserved stool) at a concentration of 10° CFU or cells/mL. Genomic DNA/RNA was tested at a concentration of 106 genomic equivalents per reaction where possible.

The non-target organisms which were shown to cross react were all closely related to their respective target organism and shared significant sequence similarity and are presented below (Table 6).

Table 6: Non-target organisms for which cross reactivity was observed with the EntericBio® Dx assay

Cross Reacting OrganismsTarget Analyte
Vibrio campbelliiVibrio
Vibrio fluvialisVibrio
Vibrio furnissiiVibrio
Vibrio mimicusVibrio
Vibrio fischeriVibrio
Vibrio natriegensVibrio

Microbial Interference

A study was performed to evaluate the performance of the EntericBio® Dx Assay in the presence of high concentrations of eleven microorganisms that are commonly found in fecal specimens (Table 7). Each potentially interfering microorganism was tested in the presence of 3X LoD of each target analyte in the EntericBio® Dx assay. No interference was observed between the potentially interfering organisms and the IAC or any of the EntericBio® Dx target analytes tested.

SubstanceSource/IDConcentration tested
Aeromonas hydrophilaDSM 17695106 CFU/mL
Bacteroides fragilisDSM 2151106 CFU/mL
Staphylococcus aureusDSM 20231106 CFU/mL
Escherichia coliDSM 30083106 CFU/mL
Enterococcus faecalisDSM 20478106 CFU/mL
Clostridium perfringensDSM 798106 CFU/mL
Saccharomyces cerevisiaeDSM 1848106 CFU/mL
Blastocystis hominisClinical specimenNot quantifiable
Pseudomonas aeruginosaDSM 50071106 CFU/mL
Klebsiella oxytocaDSM 5175106 CFU/mL
Candida albicansDSM 1577106 CFU/mL

Table 7: Potentially interfering microorganisms tested in this study

DSMZ - Deutsche Sammlung von Mikroorganismen und Zenllkulturen

{19}------------------------------------------------

Potentially Interfering Substances

A study was performed to evaluate the performance of the EntericBio® Dx Assay in the presence of 23 potentially interfering/cross-reactive substances, at high, but clinically relevant levels, that might be present in fecal specimens. Each substance was tested in the presence of each target analyte detected by the assay at low positive concentrations. Each substance was also evaluated in negative stool samples (without target organisms) where results demonstrated IAC failure in the presence of Benzalkonium chloride at a concentration of >1% v/v and Hemorrhoidal cream at a concentration of >1% w/v.

Overall, the presence of 21/23 potentially interfering substances had no effect on the detection of the target analytes or the EntericBio® Dx assay internal control at the concentrations listed in Table 8 below. A false negative result for V. parahaemolyticus was observed for 1/3 sample replicates containing hemorrhoidal cream (1% w/v) and false negative results for C. jejuni were observed for 2/3 sample replicates containing tetracycline (1.6% w/v).

SubstancePassing Concentration
Amoxicillin1% w/v3
Benzalkonium Chloride0.15% v/v3
Ceftriaxone1% w/v
Cholesterol7% w/v
Ciprofloxacin5.4% w/v
Erythromycin1.5% w/v
Hemorrhoidal cream1% w/v1,3
Human DNA0.1% v/v
Hydrocortisone50% w/v
Laxative5% v/v
Loperamide Hydrochloride0.5% w/v3
Lubricant50% w/v
Magnesium Hydroxide0.5% w/v3
Metronidazole6% w/v
Mucin10% w/v
Naproxen sodium (NSAID)10% w/v
Nystatin Cream50% w/v
Sudocrem50% w/v
Sulfamethoxazole4% w/v3
Tetracycline0.8% w/v2,3
Trimethoprim1.6% w/v
Vagisil50% w/v
Whole Human Blood5% v/v3

Table 8: Highest concentration of each substance for which most EntericBio® Dx target analytes were detected (Substances for which Interference was observed are bolded)

1 V. parahaemolyticus failed in 1/3 replicates at this concentration

² C. jejuni was only tested at 1.6% w/v and 2/3 replicates failed at this concentration

3 Interference for detection of one or more targeted organisms was observed for samples with higher than the listed concentration

Competitive Inhibition

This study was performed to evaluate the performance of the EntericBio® Dx Assay when challenged with combinations of target analytes in order to determine the potential for

{20}------------------------------------------------

competitive interference in patient specimens with mixed infections. The combinations of analytes tested were selected based on the frequency of co-infections reported in the literature. This study was performed using Cary Blair preserved negative stool specimens spiked with dual EntericBio® Dx target analyte combinations. Binary combinations of target analytes were spiked at both high and low concentrations and tested in triplicate (Table 9).

The potential for competitive inhibition was not evaluated for analytes detected in Well B (Vibrio cholerae/parahaemolyticus and STEC)

Based on the results data, competitive inhibition was observed for Giardia lamblia in the presence of both low or high concentrations of Entamoeba histolytica.

Table 9: Combinations of Entericbio " Dx Target Analytes Evaluated For Competitive Inhibition

EntericBio® Dx Target Analyte Combinations
Salmonella High (300X LoD)& Campylobacter Low (3X LoD)
Salmonella Low (18X LoD)& Campylobacter High (50X LoD)
Shigella High (200X LoD)& Giardia Low (3X LoD)
Shigella Low (12X LoD)& Giardia High (50X LoD)
Entamoeba High (50X LoD)& Cryptosporidium Low (10X LoD)
Entamoeba Low (3X LoD)& Cryptosporidium High (50X LoD)
Cryptosporidium High (50X LoD)& Giardia Low (3X LoD)
Cryptosporidium Low (10X LoD)& Giardia High (50X LoD)
Campylobacter High (50X LoD)& Giardia Low (3X LoD)
Campylobacter Low (3X LoD)& Giardia High (50X LoD)
Campylobacter High (50X LoD)& Cryptosporidium Low (10X LoD)
Campylobacter Low (3X LoD)& Cryptosporidium High (50X LoD)
Salmonella High (300X LoD)& Cryptosporidium Low (10X LoD)
Salmonella Low (12X LoD)& Cryptosporidium High (50X LoD)
Entamoeba High (50X LoD)& Giardia Low (3X LoD)
Entamoeba Low (3X LoD)& Giardia High (50X LoD)

Carry-Over and Cross Contamination

The Cross-Contamination (Carryover) study was performed to investigate the potential for carryover and cross-contamination of the EntericBio® Dx assay on the EntericBio® Workstation between and within experiments. Samples with a high concentration of target organism(s) were processed in alternating sequence with negative samples. Contrived samples were prepared in a negative stool matrix (Cary-Blair preserved stool) with two representative target analytes (one bacterial (Shigella sonnei) and one parasitic target (Giardia lamblia)) spiked at a high concentration (800x and 1000x LoD respectively). Two representative negative samples were prepared from an uninoculated negative stool matrix. The study consisted of three separate experiments for each target analyte and each experiment contained 30 samples (15 positive and 15 negative).

{21}------------------------------------------------

No carryover or cross-contamination occurred with the EntericBio® Dx assay on the EntericBio® Workstation between or within each of the assay experiments.

Specimen Stability

The specimen stability study was performed to determine the recommended specimen storage conditions for use with the EntericBio® Dx assay.

The recommended storage conditions for Cary-Blair preserved stool is 5 days at 2-8°C.

Real Time Stability

The objective of this study is to determine the shelf life in real-time of the EntericBio® Dx assay and its constituent components.

The shelf life was determined to be six (6) months at 2-8°C.

Clinical Performance

Performance characteristics of the EntericBio" Dx assay were established in a multi-site clinical study, including three (3) distinct study sites - two (2) US clinical sites and one (1) non-US clinical site.

The performance of the EntericBio" Dx assay was assessed by testing 4 specimen types:

    1. Fresh, prospectively-collected, Cary-Blair preserved fecal specimens from patients presenting with symptoms of gastrointestinal infection;
    1. Fresh, selected Cary-Blair preserved fecal specimens from patients presenting with symptoms of gastrointestinal infection. Specimens were selected based on the results obtained with Standard of Care methods in use at the study sites;
    1. Frozen, well characterized Cary-Blair preserved fecal specimens from patients presenting with symptoms of gastrointestinal infection, known to be positive for selected target analytes;
    1. Spiked/Contrived Cary-Blair preserved fecal samples for lower prevalence targets

The performance of the EntericBio" Dx assay in fresh and archived specimens was evaluated by comparing the test result for each target analyte with the appropriate comparator method(s).

For STEC and Campylobacter target analytes, a composite comparator method of three FDAcleared assays was used for fresh specimens. A specimen was characterised as positive if 2 out of 3 comparator assays were positive and a specimen was characterized as negative if 2 out of 3 comparator assays were negative. For the remaining target analytes and archived specimens, the comparator method consisted of one, FDA-cleared assay.

{22}------------------------------------------------

A total of 1523 fresh samples (1491 prospective, 32 select) were enrolled during the clinical trial. Of the 1491 prospective samples, 19 were excluded from the study, giving 1472 evaluable samples. The reasons for exclusion included invalid test result for the EntericBio Dx assay (n=9) or by the comparator method (n=6), Indeterminate result by the EntericBio Dx assay (n=1), specimen not tested with comparator method within sample stability (n=2), duplicate specimen from previously enrolled patient (n=1).

Table 10 provides a summary of demographic information for the fresh specimens enrolled into the clinical study.

GenderNumber of specimens (%)
Male636 (41.8%)
Female880 (57.8%)
N/A7 (0.4%)
Age GroupNumber of specimens (%)
<1 year11 (0.7%)
1-5 years65 (4.3%)
6-12 years27 (1.8%)
13-21 years62 (4.1%)
22-65 years863 (56.7%)
+ 65 years489 (32.1%)
N/A6 (0.4%)
Patient StatusNumber of specimens (%)
Outpatient692 (45.4%)
In-patient700 (46.0%)
Emergency Care94 (6.2%)
Long term care22 (1.4%)
N/A15 (1.0%)
Total1523

Table 10: Demographic Summary for the fresh specimens (n=1523) enrolled into the clinical study

Several target analytes had a low prevalence in the fresh clinical samples. To supplement the results of the fresh testing, 212 frozen, retrospective specimens were included in the

{23}------------------------------------------------

study. These specimens were selected and archived based on previously testing positive for the desired target analytes by routine diagnostic methods used by the collection sites.

Of the 212 archived specimens enrolled into the frozen clinical study, 3 samples were excluded from the study, giving 209 evaluable samples. The reason for exclusion was invalid test result for the EntericBio Dx assay (n=1) and Indeterminate result by the EntericBio "Dx assay (n=2).

The archived specimens were distributed across the three testing sites and were randomized such that the users performing the testing were blinded as to the expected test result.

Table 11 provides a summary of demographic information for the archived specimens enrolled into the clinical study.

Table 11: Demographic Summary for the archived specimens (n=212) enrolled into the
clinical study
GenderNumber of specimens (%)
Male105 (49.5%)
Female103 (48.6%)
N/A4 (1.9%)
Age GroupNumber of specimens (%)
<1 year5 (2.4%)
1-5 years51 (24.1%)
6-12 years21 (9.9%)
13-21 years14 (6.6%)
22-65 years96 (45.3%)
+ 65 years21 (9.9%)
N/A4 (1.9%)
Patient StatusNumber of specimens (%)
Outpatient136 (64.2%)
In-patient42 (19.8%)
Emergency Care27 (12.7%)
Long term care1 (0.5%)
N/A6 (2.8%)
Total212

{24}------------------------------------------------

Since prevalence of some target analytes (Vibrio and Entamoeba histolytica) was very low, both prospective and retrospective specimens did not yield adequate specimen numbers to demonstrate sufficient performance with the EntericBio" Dx assay. To supplement the fresh and frozen specimen data, contrived samples (n=310) were prepared and tested with the EntericBio" Dx assay. Contrived samples were prepared for each target analyte using unique fecal matrix from residual fresh specimens which had previously tested negative for all target analytes. Positive specimens were spiked at various levels, using multiple strains for each organism. Contrived samples were randomized with negative specimens so that the users were blinded to the expected result.

Of the 310 contrived samples prepared for the contrived sample study, there were no samples that were excluded from the study, giving 310 evaluable samples.

The performance of the EntericBio Dx assay in contrived samples was evaluated by comparing the test result for each target analyte with the expected sample result, based on the organism/ strain used for spiking.

The results of the EntericBio® Dx assay testing are presented in Tables 12a-12g below. Clinical study results have been stratified by the specimen type.

Table 12a: Summary of the Clinical Performance for Salmonella
Specimen Typen=% Agreement (95% CI)
PositiveNegative
SalmonellaClinical SpecimensFreshAll-Comers147292.3%24/261(75.9-97.9)100%1446/1446(99.7-100)
Select3290.0%9/102(59.6-98.2)100%22/22(85.1-100)
FrozenArchived20985.7%12/143(60.1-96.0)100%195/195(98.1-100)
Simulated310NA99.7%309/310(98.2-99.9)

12/2 Salmonella FN observed were negative for Salmonella when tested with an alternative FDA cleared PCR assay

21/1 Salmonella FN observed was positive for Salmonella when tested with an alternative FDA cleared PCR assay

31/2 Salmonella FN observed was negative for Salmonella when tested with an alternative FDA cleared PCR assay. 1/2 Salmonella FN observed was positive for Salmonella when tested with an alternative FDA cleared PCR assay.

{25}------------------------------------------------

Specimen Typen=% Agreement (95% CI)
CampylobacterClinical SpecimensPositiveNegative
FreshAll-Comers1472
Select32
FrozenArchived20993.8%15/164(71.7-98.9)
Simulated310NA100%310/310(98.8-100)

41/1 Campylobacter FN observed were negative for Campylobacter when tested with an alternative FDA cleared PCR assay

Table 12c: Summary of the Clinical Performance for Shigella/EIEC
Shigella/EIECClinical SpecimensSpecimen Typen=% Agreement (95% CI)
PositiveNegative
FreshAll-Comers1472100%14/14(78.5-100)100%1458/1458(99.7-100)
Select32100%3/3(43.9-100)100%29/29(88.3-100)
Archived20990.9%10/115(62.3-98.4)100%198/198(98.1-100)
Frozen

³1/1 Shigella/EIEC FN observed was negative for Shigella/EIEC when tested with an alternative FDA cleared PCR assay

{26}------------------------------------------------

Table 12d: Summary of the Clinical Performance for STEC
Specimen Typen=% Agreement (95% CI)
PositiveNegative
STECClinical Specimens FreshAll-Comers1472100%8/8(67.6-100)99.9%1462/1464(99.5-99.9)
Select32100%3/3(43.9-100)100%29/29(88.3-100)
FrozenArchived20994.6%70-746(86.9-97.9)100%135/135(92.7-100)
Simulated310NA100%310/310(98.8-100)

َ4/4 STEC FN observed were negative for STEC when tested with an alternative FDA cleared PCR assay

Table 12e: Summary of the Clinical Performance for Vibrio
VibrioClinical SpecimensSpecimen Typen=% Agreement (95% CI)
PositiveNegative
FreshAll-Comers14720.0%0/37(0.0-56.1)100%1469/1469(99.7-100)
Select32NA100%32/32(89.3-100)
FrozenArchived209NA100%209/209(98.2-100)
Simulated310100%100/100(96.3-100)100%210/210(98.2-100)

'3/3 Vibrio FN observed were negative for Vibrio when tested with an alternative FDA cleared PCR assay

{27}------------------------------------------------

Table 12f: Summary of the Clinical Performance for Giardia lamblia
Giardia lambliaClinical SpecimensSpecimen Typen=% Agreement (95% CI)
PositiveNegative
FreshAll-Comers147285.7%12/148(60.1-96.0)99.9%1457/1458(99.6-99.9)
Select32100%3/3(43.9-100)100%29/29(88.3-100)
FrozenArchived209100%29/29(88.3-100)100%180/180(97.9-100)
Simulated310NA100%310/310(98.8-100)

82/2 Giardia FN observed were negative for Giardia when tested with an alternative FDA cleared PCR assay

Table 12g: Summary of the Clinical Performance for Entamoeba histolytica
Entamoeba histolyticaClinical SpecimensSpecimen Typen=% Agreement (95% CI)
PositiveNegative
FreshAll-Comers1472NA100%1472/1472(99.7-100)
Select32NA100%32/32(89.3-100)
FrozenArchived2090.0%0/29(0.0-65.8)100%207/207(98.2-100)
Simulated31098.6%74/75(92.3-99.8)100%235/235(98.4-100)

92/2 Entamoeba FN observed were positive for Entamoeba when tested with an alternative FDA cleared PCR

The EntericBio" Dx assay reported multiple organism detections (co-infections) for a total of 3 specimens. This represents 0.20% of all fresh specimens tested (3/1504). All multiple detections contained two target analytes and all were concordant with the comparator method(s) used for the respective target analytes. The summary of the multi-detections reported by the EntericBio Dx assay is presented in Table 13.

The comparator assay(s) reported multiple organism detections (co-infections) for a total of 5 specimens. This represents 0.33% of all fresh specimens tested (5/1504). All multiple detections contained two target analytes. From the 5 samples with reported co-detections, the EntericBio® Dx assay did not detect a second target analyte in 2 specimens and none of the discordant specimens reported as co-infections were confirmed with an alternative FDA-

{28}------------------------------------------------

cleared assay. The summary of the multi-detections reported by the comparator assay is presented in Table 14.

Analyte_1Analyte_2PrevalenceNumber of Discrepantspecimens (FP byEntericBio)DiscrepantAnalyte(s)
CampylobacterGiardia10.070N/A
ShigellaGiardia10.070N/A
CampylobacterSTEC10.070N/A
Total30.20

Table 13: Distinct Multi-detections detected in fresh clinical specimens (n=1,504) by the EntericBio" Dx assay

Table 14: Distinct Multi-detections detected in fresh clinical specimens (n=1,504) by the
comparator methods
Analyte_1Analyte_2PrevalenceNumber ofDiscrepantspecimens (FN byEntericBio)DiscrepantAnalyte(s)
CampylobacterGiardia20.131aGiardiaa
ShigellaGiardia20.131bGiardiab
CampylobacterSTEC10.070N/A
Total50.33

a Sample was negative with EntericBio® Dx and an alternative FDA-cleared assay

b Sample was negative with EntericBio® Dx and an alternative FDA-cleared assay

Of the 1482 prospective (fresh) specimens initially evaluated with EntericBio" Dx assay, 28 specimens (1.9%) were initially reported as Invalid. Following a repeat test, 19 out of 28 invalid specimens generated valid results. Repeat testing was not performed for 3 of the 28 invalid specimens and these specimens remained as Invalid. None of the 32 select (fresh) specimens were initially reported as invalid. Of the 212 retrospective specimens initially evaluated with EntericBio" Dx assay, 2 specimens (0.9%) were initially reported as Invalid. Following a repeat test, 1 out of 2 invalid specimens was resolved. Repeat testing was not performed for the other specimen which remained invalid. The total numbers provided in Table 15 are based on compliant specimens and EntericBio® Dx assay results.

{29}------------------------------------------------

Table 15: Summary of Invalid results observed during clinical trial with the EntericBio Dx assay

Initial InvalidFinal Invalid
CountPercent95% CICountPercent95% CI
Prospective (Fresh)28/14821.9%1.3-2.7%9*/14820.6%0.3-1.2%
Select (Fresh)0/320.0%0.0-10.7%0/320.0%0.0-10.7%
Total (Fresh)28/15141.9%1.3-2.7%9*/15140.6%0.3-1.1%
Retrospective (Frozen)2/2120.9%0.3-3.4%1**/2120.5%0.1-2.6%
Total (All)30/17261.7%1.2-2.5%10/17260.6%0.3-1.2%

*19/28 initial invalids for fresh specimens were resolved upon repeat; 6/28 were invalid upon repeat and 3/28 were not repeated **1/2 initial Invalid results was resolved upon repeat and 1/2 was not repeated

Of the 1482 prospective (fresh) specimens initially evaluated with EntericBio® Dx assay, 1 specimen (0.1%) was initially reported as Indeterminate. Repeat testing was not performed and the specimen remained as Indeterminate.

None of the 32 select (fresh) specimens initially reported as Indeterminate. Of the 212 retrospective specimens initially evaluated with EntericBio® Dx assay, 2 specimens (0.9%) were initially reported as Indeterminate. Repeat testing for these 2 specimens was not performed and the specimens remained as Indeterminate. The total numbers provided in Table 16 are based on compliant specimens and EntericBio® Dx assay results.

Table 16: Summary of Indeterminate results observed during clinical trial with the EntericBio® Dx assay

Initial IndeterminateFinal Indeterminate
CountPercent95% CICountPercent95% CI
Prospective (Fresh)1/14820.1%0.0-0.4%1*/14820.1%0.0-0.4%
Select (Fresh)0/320.0%0.0-10.7%0/320.0%0.0-10.7%
Total (Fresh)1/15140.1%0.0-0.4%1*/15140.1%0.0-0.4%
Retrospective (Frozen)2/2120.9%0.3-3.4%2**/2120.9%0.3-3.4%
Total (All)3/17260.2%0.1-0.5%3/17260.2%0.1-0.5%

*1/1 initial indeterminate result was not repeated

**2/2 initial indeterminate results were not repeated

Of the 1482 prospective (fresh) specimens initially evaluated with EntericBio® Dx assay, 29 specimens (2.0%) were initially Non-reportable (Invalid and Indeterminate combined). Following a repeat test, 19 out of 29 Non-reportable results were resolved. Repeat testing was not performed for 4 of the 29 non-reportable results and the specimens remained as Non-reportable. None of the 32 select (fresh) specimens were Non-reportable.

Of the 212 retrospective specimens initially evaluated with EntericBio® Dx assay, 4 specimens (1.9%) were initially Non-reportable. Following a repeat test, 1 out of 4 Non-reportable results was resolved. Repeat testing was not performed for 3 of the 4 specimens with non-reportable

{30}------------------------------------------------

results, and the specimens remained as Non-reportable.

The total numbers provided in Table 17 are based on compliant specimens and EntericBio® Dx assay results.

Table 17: Summary of all Non-reportable results observed during clinical trial with the
EntericBio® Dx assay
Initial All Non-reportableFinal All Non-reportable
CountPercent95% CICountPercent95% CI
Prospective (Fresh)29/14822.0%1.4-2.8%10/14820.7%0.4-1.2%
Select (Fresh)0/320.0%0.0-10.7%0/320.0%0.0-10.7%
Total (Fresh)29/15141.9%1.3-2.7%10*/15140.7%0.4-1.2%
Retrospective (Frozen)4/2121.9%0.7-4.8%3**/2121.4%0.5-4.1%
Total (All)33/17261.9%1.4-2.7%13/17260.8%0.4-1.3%

*19/29 initial non-reportable results for fresh specimens were resolved upon reportable upon repeat and 4/29 were not repeated

**1/4 initial non-reportable results was resolved upon repeat and 3/4 were not repeated

Statement of Safety and Effectiveness

The data presented clearly demonstrates the safety and efficacy of the EntericBio" Dx assay as compared to the reference method when the product's Instructions for Use are followed.

§ 866.3990 Gastrointestinal microorganism multiplex nucleic acid-based assay.

(a)
Identification. A gastrointestinal microorganism multiplex nucleic acid-based assay is a qualitativein vitro diagnostic device intended to simultaneously detect and identify multiple gastrointestinal microbial nucleic acids extracted from human stool specimens. The device detects specific nucleic acid sequences for organism identification as well as for determining the presence of toxin genes. The detection and identification of a specific gastrointestinal microbial nucleic acid from individuals exhibiting signs and symptoms of gastrointestinal infection aids in the diagnosis of gastrointestinal infection when used in conjunction with clinical evaluation and other laboratory findings. A gastrointestinal microorganism multiplex nucleic acid-based assay also aids in the detection and identification of acute gastroenteritis in the context of outbreaks.(b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Gastrointestinal Microorganism Multiplex Nucleic Acid-Based Assays for Detection and Identification of Microorganisms and Toxin Genes from Human Stool Specimens.” For availability of the guideline document, see § 866.1(e).