K Number
K182504

Validate with FDA (Live)

Date Cleared
2018-10-12

(30 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 3.0T AIR MP M is a receive only RF coil designed for use with GE 3.0T MRI systems to produce diagnostic images on general human anatomy including extremities. The nucleus detected is hydrogen.

The 3.0T AIR MP L is a receive only RF coil designed for use with GE 3.0T MRI systems to produce diagnostic images on general human anatomy including extremities. The nucleus detected is hydrogen.

Device Description

The 3.0T AIR MP M is a receive-only coil designed to provide optimum signal-to noise and uniform coverage of general human anatomy including extremities and are designed for use with GE 3.0T MRI Systems. This is a 20 element coils tuned to image proton nuclei. Each coil element has an integrated preamplifier to improve image quality. The 3.0T AIR MP M is provided with P-connectors. The coil has a soft material to conform to the patient's anatomy and maximize patient comfort. Due to the flexibility of the coils, a coil holder may be used to assist in securing the coil in place.

The 3.0T AIR MP L is a receive-only coil designed to provide optimum signal-to noise and uniform coverage of general human anatomy including extremities and are designed for use with GE 3.0T MRI Systems. This is a 21 element coils tuned to image proton nuclei. Each coil element has an integrated preamplifier to improve image quality. The 3.0T AIR MP L is provided with P-connectors. The coil has a soft material to conform to the patient's anatomy and maximize patient comfort. Due to the flexibility of the coils, a coil holder may be used to assist in securing the coil in place.

AI/ML Overview

The provided text describes a 510(k) premarket notification for an MRI coil, not an AI/ML medical device. Therefore, much of the requested information regarding acceptance criteria, study design for performance, and ground truth for AI algorithms is not applicable.

However, I can extract the relevant information about the device itself and the non-clinical testing conducted to demonstrate substantial equivalence, which serves as the "proof" the device meets certain criteria for safety and performance.

Device Description and Intended Use:

The devices are the 3.0T AIR MP M (20 elements) and 3.0T AIR MP L (21 elements), which are receive-only RF coils designed for use with GE 3.0T MRI systems to produce diagnostic images on general human anatomy, including extremities. The nucleus detected is hydrogen.

1. Table of acceptance criteria and the reported device performance

Since this is an MRI coil, "acceptance criteria" and "device performance" are primarily related to safety, electromagnetic compatibility, and image quality as demonstrated through non-clinical testing against recognized standards. The document states the device performs "as intended" based on these tests.

Acceptance Criteria (based on standards/testing)Reported Device Performance
Safety Standards
AAMI/ANSI ES60601-1 (general medical electrical equipment safety)Complies with standard
IEC 60601-1-2 (electromagnetic compatibility)Complies with standard
IEC 60601-2-33 (MR equipment specific safety)Complies with standard
NEMA MS 6 (MR image quality standards)Complies with standard, performs as intended (implies acceptable IQ)
NEMA MS 9 (MR image quality standards)Complies with standard, performs as intended (implies acceptable IQ)
Biocompatibility (ISO 10993)Successful biocompatibility track record
Network Blocking AnalysisTesting conducted
Heat TestingTesting conducted
Quality Assurance Measures
Risk AnalysisApplied to development
Requirements ReviewsApplied to development
Design ReviewsApplied to development
Unit level testing (Module verification)Applied to development
Integration testing (System verification)Applied to development
Performance testing (Verification)Applied to development
Safety testing (Verification)Applied to development
Simulated use testing (Validation)Applied to development

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "The subjects of this premarket submission, the 3.0T AIR MP M and 3.0T AIR MP L, do not require clinical studies to support substantial equivalence. Sample clinical images have been included in this submission." This indicates that there was no formal "test set" in the context of a clinical study for performance evaluation. The "sample clinical images" would likely be for illustrative purposes rather than a statistically powered test set for performance metrics. Therefore, details about sample size (beyond "sample images"), data provenance, retrospective/prospective nature are not provided and likely not relevant for this type of device submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no formal clinical test set or ground truth establishment by experts for performance evaluation was conducted or required.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no formal clinical test set or ground truth establishment was conducted or required.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an MRI coil, not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is an MRI coil, not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable, as no formal ground truth was established for a clinical performance study. The "ground truth" for this type of device is implicitly tied to its ability to produce diagnostic quality images that are comparable to a predicate device, as verified through physical and electrical testing against recognized standards.

8. The sample size for the training set

Not applicable. This is an MRI coil, not an AI/ML device. There is no "training set" in this context. The development process involves engineering and quality assurance, not machine learning model training.

9. How the ground truth for the training set was established

Not applicable. This is an MRI coil, not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

GE Healthcare Coils (USA Instruments, Inc.) Veronica Meridith Regulatory Affairs Leader 1515 Danner Drive AURORA OH 44202

Re: K182504

Trade/Device Name: 3.0T Air MP M, 3.0T Air MP L Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: September 10, 2018 Received: September 12, 2018

Dear Veronica Meridith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

October 12, 2018

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara For

Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K182504

Device Name 3.0T AIR MP M

Indications for Use (Describe)

The 3.0T AIR MP M is a receive only RF coil designed for use with GE 3.0T MRI systems to produce diagnostic images on general human anatomy including extremities. The nucleus detected is hydrogen.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K182504

Device Name 3.0T AIR MP L

Indications for Use (Describe)

The 3.0T AIR MP L is a receive only RF coil designed for use with GE 3.0T MRI systems to produce diagnostic images on general human anatomy including extremities. The nucleus detected is hydrogen.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. The circle is surrounded by four stylized water droplets or flourishes, one at each cardinal direction (top, bottom, left, and right).

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:September 07, 2018
Submitter:GE Healthcare Coils (USA Instruments, Inc.)1515 Danner DriveAurora, OH 44202USA
Primary Contact Person:Veronica MeridithRegulatory Affairs LeaderGE HealthcarePhone: 262-955-5427Fax: 414-908-9585
Secondary Contact Person:Andrew MendenRegulatory Affairs ManagerGE HealthcarePhone: 262-521-6223Fax: 414-908-9585

{5}------------------------------------------------

3.0T AIR MP M3.0T AIR MP L
Device Trade Name:3.0T AIR MP M3.0T AIR MP L
Common/Usual Name:Coil, Magnetic Resonance, SpecialtyCoil, Magnetic Resonance, Specialty
Classification Names:Magnetic Resonance Diagnostic Device per 21CFR 892.1000Magnetic Resonance Diagnostic Device per 21CFR 892.1000
Product Code:MOSMOS
Predicate Device(s):3.0T Anterior Array (K172695)3.0T Anterior Array (K172695)
Device Description:The 3.0T AIR MP M is a receive-only coil designedto provide optimum signal-to noise and uniformcoverage of general human anatomy includingextremities and are designed for use with GE 3.0TMRI Systems. This is a 20 element coils tuned toimage proton nuclei. Each coil element has anintegrated preamplifier to improve image quality.The 3.0T AIR MP M is provided with P-connectors.The coil has a soft material to conform to thepatient's anatomy and maximize patient comfort.Due to the flexibility of the coils, a coil holder maybe used to assist in securing the coil in place.The 3.0T AIR MP L is a receive-only coil designedto provide optimum signal-to noise and uniformcoverage of general human anatomy includingextremities and are designed for use with GE 3.0TMRI Systems. This is a 21 element coils tuned toimage proton nuclei. Each coil element has anintegrated preamplifier to improve image quality.The 3.0T AIR MP L is provided with P-connectors.The coil has a soft material to conform to thepatient's anatomy and maximize patient comfort.Due to the flexibility of the coils, a coil holder maybe used to assist in securing the coil in place.
Intended Use:The 3.0T AIR MP M is a receive-only RF coildesigned for use with GE 3.0T MRI systems toproduce diagnostic images of general humananatomy including extremities. The nucleus detectedis hydrogenThe 3.0T AIR MP L is a receive-only RF coildesigned for use with GE 3.0T MRI systems toproduce diagnostic images of general humananatomy including extremities. The nucleus detectedis hydrogen.
Comparison ofBoth the 3.0T AIR MP M and the predicateBoth the 3.0T AIR MP M and the predicate
Intended Use:device are classified as coils for magneticresonance imaging devices and are intended fordiagnostic use. Both indications for usestatements are functional in nature, and do notlist specific diseases or conditions. The 3.0TAIR MP M and the predicate device areindicated for the same patient population, andfor the same clinical setting.device are classified as coils for magneticresonance imaging devices and are intended fordiagnostic use. Both indications for usestatements are functional in nature, and do notlist specific diseases or conditions. The 3.0TAIR MP M and the predicate device areindicated for the same patient population, andfor the same clinical setting.
The proposed Indications for use are simplifiedto reflect current coil intent, similar to theSiemens Contour 24 coil K173446, 510kcleared 11/17/2017. The Siemens Contour 24coil Indications for Use is: "The Contour 24 isintended for use with Siemens 3.0T MR systemsto produce diagnostic images of general humananatomy that can be interpreted by a trainedphysician."The proposed Indications for use are simplifiedto reflect current coil intent, similar to theSiemens Contour 24 coil K173446, 510kcleared 11/17/2017. The Siemens Contour 24coil Indications for Use is: "The Contour 24 isintended for use with Siemens 3.0T MR systemsto produce diagnostic images of general humananatomy that can be interpreted by a trainedphysician."
Therefore, GE Healthcare believes that the 3.0TAIR MP M has the same intended use as thepredicate device in accordance with the FDA'sguidance document “The 510(k) Program:Evaluating Substantial Equivalence in PremarketNotifications [510(k)]", dated 28 July 2014.Therefore, GE Healthcare believes that the 3.0TAIR MP M has the same intended use as thepredicate device in accordance with the FDA'sguidance document “The 510(k) Program:Evaluating Substantial Equivalence in PremarketNotifications [510(k)]", dated 28 July 2014.
Comparison of TechnologicalCharacteristics:Overall, the 3.0T AIR MP M employs the samefundamental scientific technology as its predicatedevice.Overall, the 3.0T AIR MP L employs the samefundamental scientific technology as its predicatedevice.
Coil Design: The most notable technologicaldifferences between the 3.0T AIR MP M and theCoil Design: The most notable technologicaldifferences between the 3.0T AIR MP L and the
predicate AIR Anterior Array is thatpredicate AIR Anterior Array is that
1)The 3.0T AIR MP M contains 20 elements ascompared to the predicate's 30 elements and1)The 3.0T AIR MP L contains 21 elements ascompared to the predicate's 30 elements and
2)The 3.0T AIR MP M is used on GE 3.0T MRIas compared to the predicate's GE 3.0T MRIsystems2)The 3.0T AIR MP L is used on GE 3.0T MRIas compared to the predicate's GE 3.0T MRIsystems
Operating Principles: The 3.0T MP M and 3.0T AIR MP L both operate on the same principles as thepredicate devices.
Materials: The 3.0T AIR MP M and 3.0T AIR MP L electrical elements and associated circuitry aresupported throughout the coil by flame rated V2 co-polyester injection mold enclosure and V0 equivalentflame rated enclosures. Both coils and the predicate device are certified to AAMI/ANSI ES60601-1, whichincludes flammability requirements.
Safety and Performance Testing: The 3.0T MP M and 3.0T AIR MP L and the predicate devices complywith the same safety and performance testing (see Determination of Substantial Equivalence, below).
These technological differences do not raise any different questions of safety and effectiveness
Determination of SubstantialEquivalence:Summary of Non-Clinical Tests:The predicate and modified devices have been subject to similar risk management testing to demonstratesubstantial equivalence of safety and performance. Testing included:• AAMI/ANSI ES60601-1• IEC 60601-1-2• IEC 60601-2-33• NEMA MS 6• NEMA MS 9
Network blocking analysis●Heat Testing●
Additionally, both predicate and modified devices have a successful biocompatibility track record, asdemonstrated by ISO 10993 testing and by their history of use in previously cleared devices.
The following quality assurance measures were applied to the development of the devices:
Risk Analysis●Requirements Reviews●Design Reviews●Testing on unit level (Module verification)●Integration testing (System verification)●Performance testing (Verification)●Safety testing (Verification)●Simulated use testing (Validation)●
Summary of Clinical Tests:
The subjects of this premarket submission, the 3.0T AIR MP M and 3.0T AIR MP L, do not require clinicalstudies to support substantial equivalence. Sample clinical images have been included in this submission.
Substantial Equivalence Conclusion:
The indications for use of the proposed devices are comparable to the claimed predicate devices. The 3.0TAIR MP M and 3.0T AIR MP L employs equivalent technology to the claimed predicate devices.Additionally, the results from the above non-clinical tests demonstrate that the devices perform as intended.Thus, the 3.0T AIR MP M and 3.0T AIR MP L are substantially equivalent to the predicate devices to whichthey have been compared.
Conclusion:GE Healthcare considers the 3.0T AIR MP M and 3.0T AIR MP L to be as safe, as effective, andperformance is substantially equivalent to the predicate devices.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has a design of water droplets or stylized lines surrounding the letters, giving it a distinctive and recognizable appearance.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circle. The circle is adorned with droplet-like shapes around its perimeter, giving it a dynamic and fluid appearance. The logo is presented in a light blue color, which contrasts with the white background.

  • Maximum B1 peak

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. Three water droplet shapes surround the circle, adding a dynamic element to the design.

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.