AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

B.O.L.T model (VA01) Base unit with it`s accessory devices attachment (VA01-0, VA01-1 and VA01-2) and the B.O.L.T application is a patient vitals measuring and monitoring system used for spot-checking and measuring the physiological parameters of adult patients / users of age or older). The physiological parameters measured and monitored are: · Non-Invasive Blood Pressure (NIBP), Pulse Rate (PR) · Oxygen saturation (SpO2), Pulse Rate (PR) · Body Temperature (TEMP) B.O.L.T application is downloaded and installed in a Bluetooth enabled smart devices/ computer. The operations of the base unit and the attached devices like start and stop are controlled by the B.O.L.T application and the results are displayed by the B.O.L.T application on the screen of the smart devices/ computer. This device is applicable for use by an adult and it can be used in a clinical setting like a physician office and also in a home environment for patients to keep track of the physiological parameters mentioned above. Home users are advised to contact their physician if any abnormal values are indicated. B.O.L.T. NIBP model Base unit (VA01) along with an inflatable cuff (VA01-0) that is wrapped around the upper arm, is a fully automatic, non-invasive, wireless blood pressurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual. B.O.L.T, IRT model Base unit (VA01) along with a thermometer probe (VA01-1) using infrared sensor detects body temperature from the ear canal in the adult population. B.O.L.T, Pulse Oximeter model Base unit (VA01) along with a pulse oximeter probe (VA01-2) is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) from the fingertip of adult users.

Device Description

The B.O.L.T patient monitor system base unit VA01 is elegantly designed with small form factor. The devices NIBP, IRT thermometer and Pulse Oximeter connects to the base unit. The device has two LED indicators that display power status and Bluetooth connectivity status. The device has an on/off slide switch on the side and USB power socket. The device has one accessories port with Lemo connector to connect IRT Thermometer and Pulse Oximeter, one at a time. There are two other ports to connect the twin tubes of the NIBP Cuff. The base unit has a rechargeable lithium polymer battery and device should be used only in battery mode and it should not be used while charging. The power adapter is given only for charging the device. The B.O.L.T. NIBP device model VA01-0 connects to the VA01 Base unit's twin port and measures both systolic and diastolic blood pressure and pulse rate via a standard oscillometric method. The oscillometric method senses the vibrating signal via the closed air pipe system and utilizes a microcontroller in the base unit to automatically sense the characteristics of the pulse signal. Unlike with the traditional measuring method, based o6i the Korotkov sound, with the oscillometric method the use of a stethoscope is not required. Through simple calculations, this method provides accurate blood pressure readings: the systolic pressure is defined as the blood pressure when the cuff pressure oscillating amplitude begins to increase, while the diastolic blood pressure is defined as the pressure when the cuff pressure oscillating amplitude stops decreasing. The device includes a plastic enclosure and external wraparound cuff, and it requires an external device (e.g. a smartphone) to display results and perform user interaction. The IRT Thermometer module of B.O.LT., model VA01-1, patient monitoring system is an electronic thermometer using an infrared sensor (thermopile) to detect body temperature from the ear canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces. The unit measures the infrared energy emitting from the middle ear and the surrounding tissue. This energy is absorbed by lenses and converted into temperature values. The IRT electronic thermometer accessory (VA01-1) connects to the accessory connector of the B.O.L.T Base unit (VA01) patient monitoring system and readout value is displayed on the connected Bluetooth device like a Smart Phone, tablet or personal computer. The Infrared Ear thermometer module of B.O.L.T, model VA01-1 patient monitoring system consists of the following parts: - Thermopile Sensor . - Application-Specific Integrated Circuit . - . Lens - Probe cover (Single use and probe cover to be disposed after each measurement) ● The B.O.L.T model VA01-2 fingertip pulse oximeter accessory device features a small form factor. low power consumption, a convenient operation, and portability. It is only necessary for a patient to put one of his/her fingers into the fingertip clips for measurement. The fingertip pulse oximeter accessory device connects to the accessory connector of the B.O.L.T VA01 base unit. The readout value Spo2(%) and Pulse rate values are displayed on the connected Bluetooth device like a Smart Phone, tablet or personal computer. The user manual which is provided with the device provides the detailed specifications. The system is controlled and operated via the B.O.L.T application which is downloaded and installed in a Bluetooth enabled smartphone/tablet/computer. The B.O.L.T application is supported for the following versions of Android/IOS and Windows operating system. - . The Android OS version 4.0 or higher. - . The iOS version 7 or higher. - . The Windows OS version 8 or higher The software communicates with the base unit (VA01) through Bluetooth communication to the connected smartphone/tablet/computer. The user has to pair the base unit with handheld mobile devices or computers through Bluetooth and use the Software application to operate and control the devices attached to the base unit. Functions like activating and stopping the device and display of the test value are all performed by the B.O.L.T application residing in the Bluetooth connected smartphone/tablet/computer. The device/s is connected to the accessory port of the base unit. The base unit takes signals from the connected device like NIBP Cuff, Pulse oximeter, IRT Thermometer. The software and firmware of the B.O.L.T patient monitoring system, process the data from the accessory devices, then display the parameters/measured data on the screen of a wirelessly connected Bluetooth device like a smartphone/ tablet/ personal computer. B.O.L.T application can hold any number of records and recognized by user ID. The number of records is limited by the data storage capacity of the secured cloud infrastructure. FCC grant is obtained under FCC ID: 2AFV6-AMI-BU-2. The data collected from devices can be securely uploaded/stored in a secured cloud. The secured cloud is HIPAA compliant and hosted in a HIPAA compliant data Centre. All the protected health information and the physiological data of a user/patient is encrypted at rest. The communication over the internet to the cloud utilizes the HTTPS/TLS protocol. The B.O.L.T is also packaged with a SDK (Software Development Kit) to integrate the reading from the devices to an external system like a Care Portal/Electronic Medical Record software system

AI/ML Overview

The provided document details the 510(k) summary for the B.O.L.T (Body Life Tracker) device, which includes modules for Non-Invasive Blood Pressure (NIBP), Infrared Radiation Energy Technology (IRT) Thermometer, and Pulse Oximeter (SpO2). The acceptance criteria and study designs are described for each module to demonstrate substantial equivalence to predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

1.1. B.O.L.T NIBP (Non-Invasive Blood Pressure)

Acceptance Criteria (Standard: EN ISO 81060-2:2014)Reported Device Performance
Criterion 1: Mean error (NIBP)Systolic Blood Pressure: 2.1 mmHg (satisfies ISO criterion of < 5 mmHg) Diastolic Blood Pressure: 1.2 mmHg (satisfies ISO criterion of < 5 mmHg)
Criterion 1: Standard Deviation (NIBP)Systolic Blood Pressure: 5.4 mmHg (satisfies ISO criterion of < 8 mmHg) Diastolic Blood Pressure: 5.7 mmHg (satisfies ISO criterion of < 8 mmHg)
Criterion 2: Standard Deviation (NIBP)Systolic Blood Pressure: 4.39 (satisfies ISO criterion of < 6.62) Diastolic Blood Pressure: 4.98 (satisfies ISO criterion of < 6.84)
Pulse Rate Accuracy:±5% from the reference device. 87 out of 89 readings (97.75%) fell within the acceptance criteria.

1.2. B.O.L.T IRT Thermometer

Acceptance Criteria (Standard: EN ISO 80601-2-56:2012)Reported Device Performance
Mean Difference (Temperature Fahrenheit)-0.1 (Mean bias: -0.10)
Standard Deviation (Temperature Fahrenheit)0.4
Limit of Agreement (Temperature Fahrenheit)-0.86 to 0.66
Clinical Repeatability (ASTM E1965-98)0.1
Clinical Bias (ASTM E1965-98)-0.008
Standard Deviation (ASTM E1965-98)0.01
Limit of Agreement (ASTM E1965-98)-0.26 to 0.25

1.3. B.O.L.T Pulse Oximeter

Acceptance Criteria (Standard: EN ISO 80601-2-61:2011)Reported Device Performance
SpO2 Accuracy (A_rms value, 70-100% SpO2)2.57
SpO2 Mean Difference2.49 (95% CI: 2.4, 2.6)
Pulse (in SpO2) A_rms value0.437
Pulse (in SpO2) Mean Difference0.067 (95% CI: 0.0108, 0.123)

2. Sample Size Used for the Test Set and Data Provenance

  • B.O.L.T NIBP:

    • Sample Size: 106 subjects (adult males and females between 18 and 65 years).
    • Readings for NIBP: 318 readings for statistical analysis.
    • Readings for Pulse Rate: 89 readings for statistical analysis.
    • Data Provenance: Clinical study, likely prospective, country of origin not specified but the standards are international (EN ISO).
  • B.O.L.T IRT Thermometer:

    • Sample Size (EN ISO 80601-2-56:2012): 105 subjects (adult males and females between 18 and 65 years). 37 febrile, 68 non-febrile.
    • Readings for Statistical Analysis: 105 readings.
    • Sample Size (ASTM E1965-98): 120 subjects (adult males and females between 18 and 65 years). 35 febrile, 83 afebrile.
    • Readings for Statistical Analysis: 118 readings.
    • Data Provenance: Controlled human clinical studies, likely prospective, country of origin not specified.
  • B.O.L.T Pulse Oximeter:

    • Sample Size: 12 subjects recruited, 10 subjects used for the clinical study (adult males and females between 18 and 65 years). 2 dark pigmented, 8 light pigmented.
    • Readings for SpO2: 250 readings for statistical analysis.
    • Readings for Pulse Rate: 225 pulse rate readings obtained.
    • Data Provenance: Clinical study (invasive controlled de-saturation study), likely prospective, country of origin not specified.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • The document does not explicitly state the number of experts used or their qualifications for establishing ground truth for the clinical studies. Ground truth was established by comparison against reference devices that are cleared under existing K numbers:
    • NIBP: Omron model HEM 907 (K032305)
    • IRT Thermometer: Braun PRO 4000 (K031928), and DT-K11E (K172861) as a contact reference thermometer.
    • Pulse Oximeter: Contec Medical Systems model CMS-50D (K082641) (as a reference device) and Radiometer model ABL 800 Basic with Co-oximeter (for measuring SaO2 from arterial blood sample).

4. Adjudication Method for the Test Set

  • The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the test sets. The accuracy of the B.O.L.T devices was evaluated by comparing their measurements directly to those obtained from the established reference devices and standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done. The studies described are device-to-reference device comparisons, not human reader performance studies with or without AI assistance. The B.O.L.T is a vital signs monitor, not an AI-assisted diagnostic imaging tool that would typically involve MRMC studies.

6. Standalone Performance

  • Yes, standalone performance (algorithm only without human-in-the-loop performance) was effectively done for each module. The studies evaluate the accuracy of the B.O.L.T device's measurements (NIBP, SpO2, Temperature) against established reference devices, without the direct involvement of human interpretation of the device's output for diagnostic purposes beyond reading the measurements themselves. The device's operation is controlled by an application, but the core measurement accuracy is assessed as a standalone system.

7. Type of Ground Truth Used

  • The ground truth used was established by comparison with measurements from predicate/reference medical devices and in the case of SpO2, also with outcomes data (SaO2 values from arterial blood samples) measured by a Co-oximeter which is considered a gold standard for blood oxygen saturation. This is a form of empirical validation against established clinical standards and reference instruments.

8. Sample Size for the Training Set

  • The document describes clinical validation studies (test sets) but does not provide any information regarding the sample size used for the training set for the device's development or algorithms. The device's technological characteristics are based on established methods (e.g., oscillometric method for NIBP, infrared for temperature, photoelectric for pulse oximetry), rather than deep learning algorithms requiring large training datasets.

9. How the Ground Truth for the Training Set Was Established

  • Since no specific training set for an AI/ML algorithm is described, the ground truth establishment for a training set is not applicable here. The device's underlying principles are based on known physiological measurement techniques and are validated against established medical device standards and reference instruments.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The Department of Health and Human Services logo is a stylized depiction of an eagle. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 6, 2019

AmZetta Technologies Private Limited Sridharan Mani Chief Executive Officer Kumaran Nagar, Semmanchery, Off Rajiv Gandhi Salai (OMR) Chennai, 600 119 In

Re: K182401

Trade/Device Name: B.O.L.T (Body Life Tracker) Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DXN, DQA, FLL Dated: August 29, 2018 Received: September 4, 2018

Dear Sridharan Mani:

This letter corrects our substantially equivalent letter of August 28, 2019.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S5

Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182401

Device Name B.O.L.T (Body Life Tracker)

Indications for Use (Describe)

B.O.L.T model (VA01) Base unit with it's accessory devices attachment (VA01-0, VA01-1 and VA01-2) and the B.O.L.T application is a patient vitals measuring and monitoring system used for spot-checking and measuring the physiological parameters of adult patients / users of age or older). The physiological parameters measured and monitored are: · Non-Invasive Blood Pressure (NIBP), Pulse Rate (PR)

· Oxygen saturation (SpO2), Pulse Rate (PR)

· Body Temperature (TEMP)

B.O.L.T application is downloaded and installed in a Bluetooth enabled smart devices/ computer. The operations of the base unit and the attached devices like start and stop are controlled by the B.O.L.T application and the results are displayed by the B.O.L.T application on the screen of the smart devices/ computer.

This device is applicable for use by an adult and it can be used in a clinical setting like a physician office and also in a home environment for patients to keep track of the physiological parameters mentioned above. Home users are advised to contact their physician if any abnormal values are indicated.

B.O.L.T. NIBP model Base unit (VA01) along with an inflatable cuff (VA01-0) that is wrapped around the upper arm, is a fully automatic, non-invasive, wireless blood pressurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual.

B.O.L.T, IRT model Base unit (VA01) along with a thermometer probe (VA01-1) using infrared sensor detects body temperature from the ear canal in the adult population.

B.O.L.T, Pulse Oximeter model Base unit (VA01) along with a pulse oximeter probe (VA01-2) is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) from the fingertip of adult users.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for AmZetta. The logo features the name "AmZetta" in a stylized font, with "Am" in blue and "Zetta" in green. A light blue cloud shape is above and behind the "Am" portion of the name. The letters are bold and slightly slanted to the right, giving the logo a dynamic and modern appearance.

510k Summary

Sponsor Details

AmZetta Technologies Private Limited., (Formerly Known As American Megatrends India Private Limited) Kumaran Nagar, Semmanchery, Off. Old Mahabalipuram Road Chennai-600119, Tamilnadu India Phone : +914461240022 Fax : +914461240025 Contact : Sridharan Mani Title : Chief Executive Officer Email : sridharanm@amzetta.co.in Alternate Contact

Alternate Contact: G. Ketharaman
Title: Unit Head- Healthcare Group
Email: amzusfda@amzetta.co.in

Summary Preparation Date: 22nd July, 2019

Device/s Submitted for FDA 510K Premarket Notification

Trade NameB.O.L.T. (Body Life Tracker)
Product CodesB.O.L.T BT1, BT2, BT3, BT4
Device Model NumbersVA01 - B.O.L.T Base UnitVA01-0 - B.O.L.T Non-Invasive Blood Pressure(NIBP) CuffVA01-1 - B.OL.T Infrared Radiation EnergyTechnology (IRT) Thermometer probeVA01-2 - B.O.L.T Pulse Oximeter (SPO2) probe
Common NamePatient Physiological Monitor
Classification NameMonitor, Physiological, Patient (without arrhythmiadetection or alarms)
ClassificationClass II
Product codeMWI
Other Product codesDQA- Pulse OximeterDXN - Blood Pressure, non-invasiveFLL -Thermometer, electronic, clinical
Regulation Number21 CFR 870.2300
Review PanelCardiovascular, General Hospital

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the AmZetta logo. The logo features the word "AmZetta" in a combination of blue and green colors. The "Am" portion of the word is in blue, while the "Zetta" portion is in green. There is a light blue cloud graphic above and behind the "Am" portion of the logo.

Devices submitted for 510K:

The B.O.L.T Patient Physiological Monitor which is called the B.O.L.T Life tracker is a patient vitals measuring and monitoring device which can be used by home users as well as a clinical setting like physician office for spot-checking and measuring of the following patient's physiological parameters:

  • Measurement of the Systolic Blood Pressure, Diastolic Blood Pressure and Pulse Rate ● (PR), non- invasively;
  • . Spot measurement of functional blood Oxyqen saturation (SpO2) and Pulse Rate (PR), non-invasively; and
  • . Measurement of the Body Temperature through ear canal, non invasively

The B.O.L.T system is modularly designed and consists of:

  • Base unit (model VA01); ●
  • NIBP (Non-Invasive Blood Pressure) Cuff (model VA01-0); ●
  • Infrared Radiation Energy Technology-IRT Thermometer (model VA01-1); and
  • . Pulse Oximeter (VA01-2).

The system is packaged as a complete system with all the devices included or alternatively packaged separately with the base unit and individual device/s. When individually packaged, the system is supplied with a base unit and the required device/s. The different product bundle along with the product code/name is given below,

B.O.L.T.ProductName/codePackage BundleDevices model numbers
BT1Base Unit and NIBPVA01 and VA01-0
BT2Base Unit, NIBP and BodyTemperatureVA01, VA01-0 and VA01-1
BT3Base Unit, NIBP and SpO2VA01, VA01-0 and VA01-2
BT4NIBP, Body Temperature and SpO2VA01, VA01-0, VA01-1 andVA01-2

Device Description:

The B.O.L.T patient monitor system base unit VA01 is elegantly designed with small form factor. The devices NIBP, IRT thermometer and Pulse Oximeter connects to the base unit. The device has two LED indicators that display power status and Bluetooth connectivity status. The device has an on/off slide switch on the side and USB power socket. The device has one accessories port with Lemo connector to connect IRT Thermometer and Pulse Oximeter, one at a time. There are two other ports to connect the twin tubes of the NIBP Cuff. The base unit has a rechargeable lithium polymer battery and device should be used only in battery mode and it should not be used while charging. The power adapter is given only for charging the device.

The B.O.L.T. NIBP device model VA01-0 connects to the VA01 Base unit's twin port and measures both systolic and diastolic blood pressure and pulse rate via a standard oscillometric method. The oscillometric method senses the vibrating signal via the closed air pipe system and utilizes a microcontroller in the base unit to automatically sense the characteristics of the pulse

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for AmZetta. The logo features a light blue cloud outline above the company name. The "Am" portion of the name is in blue, while the "Zetta" portion is in green, with a trademark symbol next to it.

signal. Unlike with the traditional measuring method, based o6i the Korotkov sound, with the oscillometric method the use of a stethoscope is not required. Through simple calculations, this method provides accurate blood pressure readings: the systolic pressure is defined as the blood pressure when the cuff pressure oscillating amplitude begins to increase, while the diastolic blood pressure is defined as the pressure when the cuff pressure oscillating amplitude stops decreasing. The device includes a plastic enclosure and external wraparound cuff, and it requires an external device (e.g. a smartphone) to display results and perform user interaction.

The IRT Thermometer module of B.O.LT., model VA01-1, patient monitoring system is an electronic thermometer using an infrared sensor (thermopile) to detect body temperature from the ear canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces. The unit measures the infrared energy emitting from the middle ear and the surrounding tissue. This energy is absorbed by lenses and converted into temperature values. The IRT electronic thermometer accessory (VA01-1) connects to the accessory connector of the B.O.L.T Base unit (VA01) patient monitoring system and readout value is displayed on the connected Bluetooth device like a Smart Phone, tablet or personal computer.

The Infrared Ear thermometer module of B.O.L.T, model VA01-1 patient monitoring system consists of the following parts:

  • Thermopile Sensor .
  • Application-Specific Integrated Circuit .
  • . Lens
  • Probe cover (Single use and probe cover to be disposed after each measurement) ●

The B.O.L.T model VA01-2 fingertip pulse oximeter accessory device features a small form factor. low power consumption, a convenient operation, and portability. It is only necessary for a patient to put one of his/her fingers into the fingertip clips for measurement. The fingertip pulse oximeter accessory device connects to the accessory connector of the B.O.L.T VA01 base unit. The readout value Spo2(%) and Pulse rate values are displayed on the connected Bluetooth device like a Smart Phone, tablet or personal computer.

The user manual which is provided with the device provides the detailed specifications.

The system is controlled and operated via the B.O.L.T application which is downloaded and installed in a Bluetooth enabled smartphone/tablet/computer. The B.O.L.T application is supported for the following versions of Android/IOS and Windows operating system.

  • . The Android OS version 4.0 or higher.
  • . The iOS version 7 or higher.
  • . The Windows OS version 8 or higher

The software communicates with the base unit (VA01) through Bluetooth communication to the connected smartphone/tablet/computer. The user has to pair the base unit with handheld mobile devices or computers through Bluetooth and use the Software application to operate and control the devices attached to the base unit. Functions like activating and stopping the device and display of the test value are all performed by the B.O.L.T application residing in the Bluetooth connected smartphone/tablet/computer. The device/s is connected to the accessory port of the base unit. The base unit takes signals from the connected device like NIBP Cuff, Pulse oximeter, IRT Thermometer. The software and firmware of the B.O.L.T patient monitoring system, process the data from the accessory devices, then display the parameters/measured data on the screen

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for AmZetta. The logo features the name "AmZetta" in a stylized font, with "Am" in blue and "Zetta" in green. Above the name is a light blue cloud shape, and a small "TM" symbol is next to the "a" in "Zetta."

of a wirelessly connected Bluetooth device like a smartphone/ tablet/ personal computer. B.O.L.T application can hold any number of records and recognized by user ID. The number of records is limited by the data storage capacity of the secured cloud infrastructure. FCC grant is obtained under FCC ID: 2AFV6-AMI-BU-2.

The data collected from devices can be securely uploaded/stored in a secured cloud. The secured cloud is HIPAA compliant and hosted in a HIPAA compliant data Centre. All the protected health information and the physiological data of a user/patient is encrypted at rest. The communication over the internet to the cloud utilizes the HTTPS/TLS protocol. The B.O.L.T is also packaged with a SDK (Software Development Kit) to integrate the reading from the devices to an external system like a Care Portal/Electronic Medical Record software system

Indications for use:

B.O.L.T model (VA01) Base unit with it`s accessory devices attachment (VA01-0, VA01-1 and VA01-2) and the B.O.L.T application is a patient vitals measuring and monitoring system used for spot-checking and measuring the physiological parameters of adult patients / users of age or older). The physiological parameters measured and monitored are:

  • Non-Invasive Blood Pressure (NIBP), Pulse Rate (PR)
  • · Oxygen saturation (SpO2), Pulse Rate (PR)
  • Body Temperature (TEMP)

B.O.L.T application is downloaded and installed in a Bluetooth enabled smart devices/ computer. The operations of the base unit and the attached devices like start and stop are controlled by the B.O.L.T application and the results are displayed by the B.O.L.T application on the screen of the smart devices/ computer.

This device is applicable for use by an adult and it can be used in a clinical setting like a physician office and also in a home environment for patients to keep track of the physiological parameters mentioned above. Home users are advised to contact their physician if any abnormal values are indicated.

B.O.L.T, NIBP model Base unit (VA01) along with an inflatable cuff (VA01-0) that is wrapped around the upper arm, is a fully automatic, non-invasive, wireless blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual.

B.O.L.T, IRT model Base unit (VA01) along with a thermometer probe (VA01-1) using infrared sensor detects body temperature from the ear canal in the adult population.

B.O.L.T, Pulse Oximeter model Base unit (VA01) along with a pulse oximeter probe (VA01-2) is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) from the fingertip of adult users.

Contraindications:

The following contraindications are mentioned in the labelling information: The B.O.L.T device is not recommended for people with serious arrhythmia. Consult your doctor during pregnancy, arrhythmia, and arteriosclerosis. The analyzed results from the devices are not

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for AmZetta. The logo features the word "AmZetta" in a stylized font, with the "Am" in blue and the "Zetta" in green. There is a cloud graphic above the word "Am", also in blue. The letters are bold and slightly slanted to the right, giving the logo a dynamic and modern feel.

sufficient to make a correct diagnosis of the patient`s clinical condition. A detailed clinical history of the patient together with the results of any other tests suggested by a doctor is also required. This device is contraindicated for any person who is connected to a wearable or implantable electronic device or instrument, such as a pacemaker or defibrillator.

Substantial Equivalence comparison of the B.O.L.T NIBP with predicate Devices:

The B.O.L.T NIBP model VA01- 0 is compared against the following predicate devices:

K133125 Withings Blood Pressure Monitor, Upper Arm Type: BP-801 model BP01 from Withings

K120672 Blood Health BPS Fully Automatic Arm Cuff Wireless Blood Pressure dock model KD 936 from Andon Health

Comparison ItemSubmitted proposed devicePredicate device 1Predicate device 2
ApplicantAmZetta TechnologiesPrivate Limited.WithingsAndon Health Co.,Ltd
Trade NameB.O.L.TWithings BloodPressure Monitor,Upper ArmType: BP-801Blood Health BPSFully Automatic ArmCuff Wireless BloodPressure dock
Model NameB.O.L.T VA01-0BP-801KD-936
510K NumberNew ListingK133125K120672
TechnologicalcharacteristicsOscillometric methodOscillometric methodOscillometric method
Measuring methodOscillometric method, automatic inflation and measurementOscillometric method, automatic inflation and measurementOscillometric method, automatic inflation and measurement
SensorSemiconductor gauge sensorSemiconductor gauge sensorSemiconductor gauge sensor
Rapid Air releaseThrough an active electronic control valveThrough an active electronic control valveThrough an active electronic control valve
Cuff tissuesLycra, Velcro loop, Velcro hook, PU leather, biasLycra, Velcro loop, Velcro hook, PU leather, biasLycra, Velcro loop, Velcro hook, PU leather, bias
Pressure AccuracyPressure: $\pm$ 3mmHgPressure: $\pm$ 3mmHg or$\pm$ 2% of Pressurereadout valuePressure: $\pm$ 3mmHg
Pulse accuracy$\pm$ 5% of Pressurereading value$\pm$ 5% of Pressurereading value$\pm$ 5% of Pressurereading value
SystemArchitectureRequires an external device to constitute a complete blood pressure measurement systeme.g. blood pressure cuffIndependent operationRequires an external device to constitute a complete blood pressure measurement system

AmZetta Technologies Private Ltd.,

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for AmZetta. The word "AmZetta" is written in a stylized font, with "Am" in blue and "Zetta" in green. There is a light blue cloud shape behind the "Am" portion of the logo. The trademark symbol is located to the right of the word "Zetta".

Display an UserInteractionRequires an externaldevice to displayresults and performuser interaction.Devices supportedany Bluetooth enabledsmart phone, tablet orpersonal computer.LCD monitor andpersonal computerRequires an externaldevice to displayresults and performuser interaction
CommunicationWireless, based onBluetooth V4.0 + EDR(Bluetooth Lowenergy)Wireless, based onBluetooth V4.0Dual mode(V2.1 +EDR)Wireless, based onBluetooth V3.0 +EDR
Power SourceRechargeablebatteries (Li-Polymer)3.7 volt,850mAhand DC power 5V, 2A.Disposable 4xAAAalkaline batteriesRechargeablebatteries (Li-lon 400mAh)
Cuff Type and SizeUpper arm type, size:22 to 42 cmUpper arm type, size:22 to 42 cmUpper arm type, size:23 to 48 cm
OperatingTemperature andHumidity10 to 40° CAtmospheric: 15 to85% RH non-condensing10 to 40° C 15 to90% RH.Atmospheric:86Kpato 106KpaAltitude: 2000m5 to 40° CAtmospheric: <90%RH
StorageTemperature andHumidity-10 to 55° CAtmospheric: 15 to85% RH non-condensing-5 to 70° C 10 to95% RH.Atmospheric:86Kpato 106KpaAltitude: 2000m-22 to 55° CAtmospheric: <90%RH

Material Make-up of the device:

Comparison ItemApplicantPredicate Device 1Predicate Device 2
BP CuffLycra fabric on therear side, Velcroloop, Velcro hook,PU leather, biasLycra, Velcro loop,Velcro hook, PUleather, biasLycra, Velcro loop,Velcro hook, PUleather, bias
Rubber tube of BPCuffNeoprene (tubingwall)Neoprene (tubingwall)Neoprene (tubingwall)

Summary with predicate device:

The principles of operations, core technology and design specifications used in the NIBP module of the B.O.L.T model VA01-0 are similar to the predicate devices except the assembly and the implementation of the core design. A detailed document on the principles of operations, design, assembly and technology characteristics were submitted.

The intended use and the indications for use of the B.O.L.T blood pressure monitor, model VA01-0 as described in its labeling are similar as the two predicate devices and all three devices are intended to be used in similar manner and environments, except B.O.L.T blood pressure system model VA01-0 can be used by the medical professional in a clinical as well as home users.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for AmZetta. The logo features the word "AmZetta" in a stylized font, with the "Am" portion in blue and the "Zetta" portion in green. Above the word "AmZetta" is a light blue cloud shape. To the right of the word "Zetta" is the trademark symbol.

Patient contact materials, performance, biocompatibility function, mechanical safety, standards met, electrical safety, and EMC are similar. The material make up of the B.O.L.T blood pressure monitor, model VA01-0 and the two predicate devices compared are similar and the exception is only in the color of the material.

B.O.L.T blood pressure monitor, model VA01-0 and the predicate devices use Bluetooth to communicate wirelessly with a smartphone, tablet or PC. While B.O.L.T blood pressure monitor uses Bluetooth V4.0 + EDR, predicate device KD-936 uses Bluetooth V3.0+EDR and the Withings Blood Pressure Monitor, Upper Arm Type: BP-801 uses Bluetooth V4.0. The user interface, visual appearance and cuff size for the new device and the predicate devices differ.

AmZetta Technologies Private Limited believe that the B.O.L.T blood pressure monitor model VA01-0 is substantially equivalent to the predicate devices and other products currently in distribution.

Substantial Equivalence comparison of the B.O.L.T IRT Thermometer with predicate Devices:

B.O.L.T IRT Thermometer model VA01-1 is compared against K170219 Microlife Digital Infrared Ear Thermometer model IR1DR1-1 from Microlife Intellectual Property GmbH, Switzerland

Comparison ItemSubmitted proposed devicePredicate Device
ApplicantAmZetta Technologies Private Limited.Microlife Intellectual Property GmbH, Switzerland
Trade NameB.O.L.T IRT ThermometerMicrolife Digital Infrared Ear Thermometer
Model NameB.O.L.T VA01-1IR1DR1-1
510K NumberNew ListingK170219
Device Measurement TechnologyInfraredInfrared
Measuring methodUses an infrared sensor (thermopile) to detect body temperature from the ear canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces.The unit measures the infrared energy emitting from the middle ear and the surrounding tissue. This energy is absorbed by lenses and converted into temperature valuesUses an infrared sensor (thermopile) to detect body temperature from the ear canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces.The unit measures the infrared energy emitting from the middle ear and the surrounding tissue. This energy is absorbed by lenses and converted into temperature values
Measurement locationEar CanalEar canal
Probe coverneeded?YesNo
Sensor typeMLX90615TPS23B sensor
Lens TypeTransparentTransparent
Measurement range35.0- 42.0°C (95 -107.6 °F)32.0- 42.2°C (89.6 -108.0 °F)
Display resolution0.1°C or 0.1°F0.1°C or 0.1°F
Temperature±0.2° C (35.0° C to 42°C),±0.3° C < 35.5° C,
measurementoutside this range:±0.2° C (35.5° C - 42° C);
accuracyHigh / Low±0.3° C > 42° C
Position indicatorYes(Application shows theusage)Yes
ElevatedTemperature AlarmNoYes (10 short beeps whenmeasured temperature is greaterthan 37.5 °C)
Display an UserInteractionRequires an external device todisplay results and perform userinteraction. Devices supportedany Bluetooth enabled smartphone, tablet or personalcomputer RUNNING on AndroidOS version 4.0 and higher OrIOS OS version 7.0 or higher orWindows OS version 8.0 orhigherLCD display built in to the device
CommunicationWireless, based on BluetoothV4.0 + EDR (Bluetooth Lowenergy)Independent
Power SourceRechargeable batteries (Li-Polymer3.7V, 850 mAh) and DCpower 5V, 2A2*AAA 1.5v alkaline battery
OperatingTemperature andHumidity10 to 40°C ((50.0°F~104.0°F)Atmospheric: 15 to 85% RH non-condensing10.0°C - 40°C (50.0°F~104.0°F)with relative humidity 95%
Storagetemperature andHumidity-10 to 55° CAtmospheric: 15 to 85% RH non-condensing-25°C to 55 °C/-13°F ~131°F withrelative humidity 95%
Probe head: TipWidth7.5mm7.3mm
Probe head:Thickness at 5mmheight8.0mm8.56mm
Probe head:Thickness at 10mmheight9.2mm10.32mm
Comparison ItemApplicantPredicate Device 1
IRT Probe coverPolypropylene materialPolypropylene material
IRT ProbeABS PlasticABS Plastic
Plastic materialABS+PC PlasticABS Plastic

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for AmZetta. The logo features the word "AmZetta" in a stylized font, with "Am" in blue and "Zetta" in green. There is a cloud graphic above and behind the "Am" portion of the logo. The trademark symbol is located to the right of the word "Zetta".

{12}------------------------------------------------

Material Make-up of the Device:

Summary with predicate device:

The principles of operations, core technology, the technological characteristics of patient contact materials, performance, biocompatibility function, mechanical safety, standards met, electrical safety, and EMC in the IRT temperature module of the B.O.L.T model VA01-1 are similar to the predicate device except the assembly and the implementation of the core design. A detailed document on the principles of operations, design, assembly and technology characteristics were submitted.

The device is similar to the predicate device in intended patient population, intended application site, working principle, contact material, resolution. Only their internal power supply. display range, communication and appearance differ and readout of the data/results where the B.O.L.T uses a wireless Bluetooth device to display the results and the predicate device has a built-in LCD display.

The material make up of the B.O.L.T IRT thermometer, model VA01-1 and the two predicate devices compared are similar and the exception is only in the color of the material.

AmZetta Technologies Private Limited believe that the B.O.L.T model VA01-1 patient monitor's IRT thermometer is substantially equivalent to the predicate devices.

Substantial Equivalence Comparison B.O.L.T Pulse Oximeter with Predicate Devices:

The B.O.L.T Pulse Oximeter model VA01-2 is compared with the following 510K approved devices:

K163135 - Fingertip Pulse Oximeter model A330, A310 from Shenzhen Fitfaith Technology Co..Ltd

K140582 – Fingertip Pulse Oximeter JPD-500A from Shenzhen Jumper Medical Equipment Co.,Ltd

Comparison ItemSubmitted proposeddevicePredicate device 1Predicate device 2
ApplicantAmZettaTechnologies PrivateLimited.Shenzhen FitfaithTechnology Co.,LtdShenzhen JumperMedical EquipmentCo.,Ltd
Trade NameB.O.L.T PulseOximeterFingertip PulseOximeterFingertip PulseOximeter
Model NameB.O.L.T VA01-2A330, A310JPD-500A
510K NumberNew ListingK163135K140582

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for AmZetta. The logo features the word "AmZetta" in a stylized font, with the "Am" in blue and the "Zetta" in green. Above the word is a light blue cloud shape. The letters are bold and slightly slanted to the right, giving the logo a dynamic and modern feel.

TechnologicalcharacteristicsPhotoelectricOxyhemoglobinInspectionTechnologyPhotoelectricOxyhemoglobinInspectionTechnologyPhotoelectricOxyhemoglobinInspectionTechnology
Measuring methodA mathematicalformula is establishedmaking use ofLambert Beer Lawaccording toSpectrum AbsorptionCharacteristics ofReductivehemoglobin (RHb)andOxyhemoglobin(HbO2) in red andnear-infraredzones. Operationprinciple of theinstrument:PhotoelectricOxyhemoglobinInspectionTechnology isadopted inaccordance withCapacity PulseScanning andRecordingTechnology, so thattwo beams ofdifferent wavelengthof lights can befocused onto ahuman nail tipthrough a clampingfinger-type sensor. Ameasured signalobtained by aphotosensitiveelement, is shown onthe connectedBluetooth device likea smartphone, tabletor personal computerdisplay throughprocess in electroniccircuits andmicrocontroller of theA mathematicalformula isestablished makinguse ofLambert Beer Lawaccording toSpectrum AbsorptionCharacteristics ofReductivehemoglobin (RHb)andOxyhemoglobin(HbO2) in red andnear-infraredzones. Operationprinciple of theinstrument:PhotoelectricOxyhemoglobinInspectionTechnology isadopted inaccordance withCapacity PulseScanning andRecordingTechnology, so thattwo beams ofdifferent wavelengthof lights can befocusedonto a human nail tipthrough a clampingfinger-type sensor. Ameasured signalobtained by aphotosensitiveelement, is shown onthe Oximeter'sdisplay throughprocess in electroniccircuits andmicroprocessor.A mathematicalformula isestablished makinguse ofLambert Beer Lawaccording toSpectrum AbsorptionCharacteristics ofReductivehemoglobin (RHb)andOxyhemoglobin(HbO2) in red andnear-infraredzones. Operationprinciple of theinstrument:PhotoelectricOxyhemoglobinInspectionTechnology isadopted inaccordance withCapacity PulseScanning andRecordingTechnology, so thattwo beams ofdifferent wavelengthof lights can befocusedonto a human nail tipthrough a clampingfinger-type sensor. Ameasured signalobtained by aphotosensitiveelement, is shown onthe Oximeter'sdisplay throughprocess in electroniccircuits andmicroprocessor.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for AmZetta. The logo features a light blue cloud in the background, with the word "AmZetta" in front of it. The "Am" portion of the word is in blue, while the "Zetta" portion is in green. There is a trademark symbol to the right of the word.

Semiconductor gauge sensorSemiconductor gauge sensorSemiconductor gauge sensor
SensorSemiconductor gauge sensorSemiconductor gauge sensorSemiconductor gauge sensor
Light Specification
Wavelength660nm±3nm for red light,905nm±5nm for infrared light (IR)660nm±3nm for red light, 905nm±5nm for infrared light (IR)660nm±3nm for red light, 905nm±5nm for infrared light (IR)
Maximum optical power1.5 mW for red light (660nm)1.2 mW for IR (905nm)1.5 mW for red light (660nm)1.2 mW for IR (905nm)1.5 mW for red light (660nm)1.2 mW for IR (905nm)
Contact materialABS+PC for enclosure, Silicone Rubber clipABS for enclosure, silica gel for clipABS for enclosure, silica gel for clip
Resolution
SpO21%1%1%
Pulse rate1 bpm1 bpm1 bpm
Measurement Range
SpO20%~100%0%~100%0%~100%
Pulse rate30~250 bpm25~250 bpm25~250 bpm
Measurement Accuracy
SpO2Arms ≤2.6%:(70%~100%)Unspecified:(0%~69%)±2%: (70%~100%)Unspecified:(0%~69%)±2%: (70%~100%)No definition:(0%~69%)
Pulse rate±2 bpm±2 bpm±2 bpm
System ArchitectureRequires the base unit B.O.L.T to connect to constitute a complete measurement systemIndependent operationIndependent operation
Display an User InteractionRequires an external device to display results and perform user interaction.Devices supported any Bluetooth enabled smart phone, tablet or personal computer running Android OS version 4.0 and higher Or iOS OS version 7.0 or higher or Windows OS version 8.0 or higherLED display built in to the deviceLED display built in to the device
CommunicationWireless, based on Bluetooth V4.0 + EDR (Bluetooth Low energy)IndependentIndependent
Power SourceRechargeablebatteries (Li-Polymer3.7V, 850 mAh) andDC power 5V, 2A2*AAA 1.5V alkalinebattery2*AAA 1.5V alkalinebattery
OperatingTemperature andHumidity10 to 40°CAtmospheric: 15 to85% RH non-condensing5°C~40°CAtmospheric:15%~85%noncondensing5°C~40°CAtmospheric:15%~85%noncondensing
StorageTemperature andHumidity-10 to 55°CAtmospheric: 15 to85% RH non-condensing-20°C~+55°CAtmospheric:10%~95%-10°C~+50°CAtmospheric:15%~93%

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for AmZetta. The logo features a light blue cloud in the background, with the word "AmZetta" in front of it. The "Am" portion of the word is in blue, while the "Zetta" portion is in green. There is a trademark symbol to the right of the word.

Material make-up of the device:

Comparison ItemApplicantPredicate Device 1Predicate Device 2
SPO2 Finger clipMaterialSilicon rubberSilicon GelSilicon Gel
Plastic Material forenclosureABS+PC PlasticABS PlasticABS Plastic

Summary with predicate device:

The principles of operations, core technology and the design used in the B.O.L.T Pulse oximeter, model VA01-2 Pulse oximeter are similar to the predicate device except the implementation design and the system architecture. A detailed design document on the principles of operations. design, assembly and technology characteristics were submitted.

The B.O.L.T pulse oximeter, model VA01-2 is similar to the predicate device in intended use. intended patient population, intended application site, working principle display range, resolution. Only their contact material, internal power supply, appearance differ and readout of the data/results where B.O.L.T uses a wireless Bluetooth connected devices to display the results and the predicate devices has a built-in display.

The B.O.L.T pulse oximeter, model VA01, VA01-2 and the predicate devices are substantially equivalent in the technological characteristics of patient contact materials, performance, biocompatibility function, mechanical safety, standards met, electrical safety, and EMC. The material make up of the B.O.L.T pulse oximeter, model VA01, VA01-2 and the two predicate devices compared are similar and the exception is only in the color of the material and B.O.L.T pulse oximeter uses Silicon rubber and the predicate devices use silicon gel.

AmZetta Technologies Private Limited believe that the B.O.L.T pulse oximeter model VA01, VA01-2 is substantially equivalent to the predicate devices.

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for AmZetta. The logo features the company name in a stylized font, with the "Am" portion in blue and the "Zetta" portion in green. A light blue cloud-like shape is above the "Am" portion of the name. The letters are bold and slightly slanted to the right.

Discussion of Clinical Tests performed:

B.O.L.T NIBP:

The B.O.L.T blood pressure monitor, model VA01-0 is compliant to the EN ISO 81060-2: 2014 Non-invasive sphygmommsnometers – Part 2: Clinical investigation of automated measurement type Standard for Manual, electronic, or automated sphygmomanometers. All the relevant activities were performed and the results demonstrated that the predetermined acceptance criteria were fully met.

106 subjects were recruited for the clinical study which included adult both male and female between 18 years and 65 years for Blood pressure and a total of 318 reading were taken for statistical analysis. For pulse rate statistical analysis 89 pulse rate reading was obtained. Omron model HEM 907 was used as the reference device. Omron IntellisenseTM Digital Blood Pressure Monitor, Model HEM-907 is approved under K032305.

subject distribution was as per the EN ISO 81060-2:2014 Non-invasive The sphygmommsnometers – Part 2: Clinical investigation of automated measurement type Standard for Manual, electronic, or automated sphygmomanometers standards.

The clinical study met the accuracy criteria as specified by EN ISO 81060-2:2014 standards. As per the Criterion I, mean error is 2.1 and 1.2 for Systolic Blood Pressure and Diastolic Blood Pressure respectively (falling within specified ISO criteria of less than 5). The standard deviation is 5.4 and 5.7 for Systolic Blood Pressure and Diastolic Blood Pressure respectively (falling within specified ISO criteria of less than 8). As per the Criterion 2, standard deviation for Systolic Blood Pressure is 4.39; and Diastolic Blood Pressure is 4.98 (Satisfied for both Systolic Blood Pressure (4.39 < 6.62) and Diastolic Blood Pressure (4.98 < 6.84).

The clinical study for Pulse rate accuracy based on the standard protocol specification. It is defined as ±5% from the reference device. 87 out of the 89 (97.75%) reading obtained for the study falls within the acceptance criteria. The detailed clinical study report was submitted.

B.O.L.T IRT Thermometer:

Controlled human clinical studies were conducted on the subject device, B.O.L.T IRT Thermometer, model VA01-1. Clinical protocol used for controlled human clinical studies and Clinical validation data is presented below.

IRT Temperature module of the B.O.L.T model VA01-1 patient monitoring system is compliant to the EN ISO 80601-2-56:2012 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. All the relevant activities were performed and the, results demonstrated that the predetermined acceptance criteria were fully met.

105 subjects were recruited for the clinical study which included adult both male and female between 18 years and 65 years. Of the 105 subjects, 37 were febrile subjects and 68 non-febrile subjects. A total of 105 reading were taken for statistical analysis. Braun PRO 4000 was used as the reference device. Braun ear thermometer, Model Pro 4000 is approved under K031928.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for AmZetta. The logo features the company name in a stylized font, with the "Am" portion in blue and the "Zetta" portion in green. A light blue cloud shape is positioned above and behind the "Am" part of the name, adding a visual element of cloud computing or data storage.

The clinical study for Body temperature met the accuracy criteria as specified by EN ISO 80601-2-56:2012 standards. As per the Criterion the Parameter Temperature Fahrenheit, the mean difference is -0.1 with mean bias of -0.10. standard deviation of 0.4 and limit of agreement of -0.86 to 0.66.

In addition, further clinical study were conducted as per ASTM E1965-98.

IRT Temperature module of the B.O.L.T model VA01-1 patient monitoring system is compliant to the ASTM E1965 – 98 (Reapproved 2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature. All the relevant activities were performed and the, results demonstrated that the predetermined acceptance criteria were fully met.

120 subjects were recruited for the clinical study which included adult both male and female between 18 years and 65 years. Of the 118 subjects, 35 were febrile subjects and 83 afebrile subjects. A total of 118 reading were taken for statistical analysis. Braun PRO 4000 was used as the reference device. Braun ear thermometer, Model Pro 4000 is approved under K031928. The DT-K11E was used as the contact reference thermometer is approved under K172861.

The clinical study for Body temperature met the accuracy criteria as specified by ASTM E1965-98 (reapproved 2016) standards. As per the Criterion the Parameter Temperature Fahrenheit, the clinical repeatability 0.1 with clinical bias of -0.008, standard deviation of 0.01 and limit of agreement of -0.26 to 0.25.

A detailed Clinical Study Report performed under EN ISO 80601-2-56:2012 standards and ASTM E1965-98 were submitted.

B.O.L.T Pulse Oximeter:

The B.O.L.T Pulse Oximeter model VA01-2 is compliant to the EN ISO 80601-2-61: 2011 -Medical electrical equipment - Part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment.

The purpose of the clinical trial was to evaluate the SpO2 accuracy performance of the B.O.L.T Fingertip Pulse Oximeter model VA01-2 during stationary (non-motion) conditions over a wide range. The Blood oxygen saturation (SpO2) accuracy are measured as per ISO 80601-2-61:2011. section EE.2.2: Invasive controlled de-saturation study on Patients. The SpO2 Accuracy of B.O.L.T BT4 are measured by comparing SpO2 readings to values obtained with that is traceable to Co-Oximeter SaO2 Values.

All the relevant activities were performed and the, results demonstrated that the predetermined acceptance criteria were fully met.

12 subjects were recruited and 10 Subjects were used for the clinical study which included adult both male and female between 18 years and 65 years. In the study there were 2 dark pigmented and 8 light pigmented subjects. For SpO2 functional oxygen saturation level clinical study, a total of 250 readings were taken for statistical analysis. For SpO2 pulse rate, statistical analysis 225 pulse rate readings were obtained using Electronic Pulse Simulator. Contec Medical Systems model CMS-50D (510K approval – K082641) was used as the reference device and Radiometer model ABL 800 Basic with Co-oximeter was used as the measuring SaO2 from arterial blood sample.

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the logo for AmZetta. The logo features the name "AmZetta" in a stylized font, with the "Am" portion in blue and the "Zetta" portion in green. A light blue cloud-like shape is positioned above and behind the "Am" part of the name, suggesting a cloud computing or technology-related theme. The letters are bold and slightly italicized, giving the logo a dynamic and modern appearance.

B.O.L.T Pulse Oximeter model VA01-2 controlled desaturation clinical study for functional oxygen saturation level (Sp02) met both the accuracy criteria as specified by EN ISO 80601-2-61:2011 -Medical electrical equipment - Part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment non-invasive method - induced controlled desaturation environment in healthy adults during non-motion conditions over the range of 70-100% SpO%

As per the standards, for the parameter Oxygen saturation (SpO2 %) Ams value is 2.57. The mean difference is 2.49 with 95% confidence interval of (2.4, 2.6). For the parameter Pulse (In SpO2) Ams value is 0.437. The mean difference is 0.067 with 95% confidence interval of (0.0108, 0.123).

A detailed Clinical Study Report performed were submitted

Non-Clinical Tests Performed:

Non-clinical tests were conducted to verify that the proposed device met all design specifications and to evaluate the safety and effectiveness. All the non-clinical test reports were submitted. The test results demonstrated that the proposed device complies with the standards are listed below

  • IEC 60601-1:2005(Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 Medical . devices Part1: General requirements for basic safety and essential performance Amendment 1, General requirements: Electrical Safety
  • IEC 60601-1-11: 2015(Second Edition) Medical electrical equipment part 1-11: General . requirements for basic safety and essential performance - Collateral Standard Requirement for medical equipment and medical electrical systems used in home healthcare environment - General Requirements: Electrical Safety
  • ANSI/AAMI/IEC 60601-1-2: 2014 Medical electrical equipment part1-2: General . requirements for basic safety and essential performance - Collateral standards: Electromagnetic compatibility - Requirements and tests, General requirements: EMC
  • . IEC 62304:2006(First Edition) + A1:2015 Medical device software: Software life-cycle processes - General requirements: Software
  • IEC 60601-1-6:2010. AMD1:2013 Medical electrical equipment Part 1 6: General . Requirement for basic safety- Collateral Standard: Usability - General requirements: Usability
  • . ISO 10993-1:2009 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process - Biological evaluation
  • . ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5 Tests for In Vitro Cytotoxicity - General Requirements: Biological estimation
  • . ISO 10993-10:2010 Biological Evaluation of Medical Devices – Part 10: Tests for Irritation and Delayed-Type Hypersensitivity - General Requirements: Biological estimation
  • . ISO 13485 Medical devices- Quality management systems-Requirements for regulatory purposes - Quality Management Systems

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the logo for AmZetta. The logo features the name "AmZetta" in a stylized font, with the "Am" portion in blue and the "Zetta" portion in green. A light blue cloud shape is positioned above and behind the "Am" part of the name. The letters are bold and slightly slanted, giving the logo a dynamic and modern appearance.

  • IEC 80601-2-30:2009/AMD1:2013 Medical electrical equipment Part 2-30: Particular ● requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers - General requirements: Essential performance
  • . EN ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment -General requirements: Essential performance
  • ISO 80601-2-56:2009 Medical electrical equipment -- Part 2-56: Requirements for basic . safety and essential performance of clinical thermometers for body temperature Measurement - Essential Performance

Biocompatibility Assessment:

Biocompatibility assessment was carried out for components in the devices where there is a surface contact to the body. Components tested were:

  • . Plastics used in the devices. Plastic used in the device is ABS + Polycorbonate. There is a surface contact to the body and the duration is less than 1 minutes.
  • NIBP Cuff Rubber Tube. There is a surface contact to the body and the duration is . less than 5 minutes.
  • NIBP Cuff. There is a surface contact to the body and the duration is less than 5 ● minutes.
  • . Pulse Oximeter (Spo2) Finqer rubber. There is a surface contact to the body and the duration is less than 1 minutes
  • . IRT Probe cover. There is a surface contact to the body and the duration is less than 1 minutes.

Following standards were used for the Biocompatibility assessment. The Biocompatibility assessment reports were submitted.

  • . Invitro Cytotoxicity tests were carried as per the ISO 10993 – Part 5 standards.
  • Intracutaneous reactivity test were carried as per the ISO 10993 Part 10 standards. .
  • . Skin Sensitization tests were performed as per the ISO 10993 - Part 10 standards.

Shelf-life:

As there are no degradable components in the subject devices, shelf-life does not apply to the subject devices except the rechargeable battery. There are no time dependent parts used. The base unit has a rechargeable battery (Li-Polymer3.7V, 850 mAh) and DC power 5V, 2A. The rechargeable batteries usually has a shelf-life of 2-3 years and the batteries can be easily replaced.

{20}------------------------------------------------

Discussion on sterilization requirements:

There is no sterilization required. If required, the cuff can be wiped with a soft damp cloth.

Clean the rubber touching the finger inside the SpO2 probe gently by a soft cloth dampened with 70% isopropyl alcohol before measurement before each use.

There is no sterilization required for the B.O.L.T IRT Thermometer model VA01-1.

A new probe cover must be used for each measurement. Probe covers need to be disposed after the single use. Only the AmZetta provided/recommended Soft Touch cover must be used.

The user manual which is included in the device provides a detailed instruction on the care and maintenance of the device and the usage of the device.

Label and Labelling:

We have followed the standards of 21 CFR Part 801 as suggested by FDA and the following describes the proposed labelling.

The labelling is provided in the packaging. A user manual which is provided along with the system gives clear instructions on how to use the devices, the indications of use and any contraindications. The user manuals also provide the proper usage of the devices and the maintenance of the devices.

Proposed labelling information and user manual were submitted.

Software Development Methodology and Software Level of concern classification:

American Megatrends India Pvt Ltd is an ISO 13485 certified organization. The submission has undergone American Megatrends India Pvt Ltd's, design control, verification and validation testing. The validation testing of all executable code and functionality and confirmation that all identified risks have been adequately addressed by software functionality. the user interface, documentation and the User Guide.

We assign a software safety class (A. B. or C) to each software system according to the possible effects on the patient, operator, or other people resulting from a HAZARD (being a potential source of Harm) to which the software system can contribute. We assign the software safety classes to the medical device software application based on the severity as follows:

  • . Class A: No injury or damage to health is possible
  • Class B: Non-SERIOUS INJURY is possible ●
  • Class C: Death or SERIOUS INJURY is possible

The following flowchart provides the how software level of classification is derived at:

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image shows the logo for AmZetta. The logo features the company name in a stylized font, with the "Am" portion in blue and the "Zetta" portion in green. A light blue cloud graphic is positioned above and to the left of the company name. The letters are bold and slanted to the right.

Image /page/21/Figure/2 description: The image is a flowchart that describes the software safety classification process. The flowchart starts with "Class C (by default)" and asks the question "Can a HAZARDOUS SITUATION arise from a failure of the software?". If the answer is no, the flowchart goes to "Class A", otherwise it goes to "Evaluate effectiveness of RISK CONTROL measures external to the software". The flowchart then asks "Does failure of the software result in unacceptable RISK?" and if the answer is no, the flowchart goes to "Class A", otherwise it asks "What severity of injury is possible?" and goes to "Class B" if the answer is "Non SERIOUS INJURY" and "Class C" if the answer is "SERIOUS INJURY/death".

Software level of concern classification was done for the all the devices and the software level of concern was deemed Class B (Moderate).

The System verification and validation activities as part of the design control process include testing of all Design Specifications based on risk analysis and verification plans. American Megatrends India Pyt Ltd's test plan execution ensures each type of medical devices (NIBP, Spo2 and IRT) works with the base unit connectors for devices; and the base unit communication to the connected blue tooth device/s; and the SDK (Software Development Kit) to integrate the reading from the devices to an external system like a Care Portal/Electronic Medical Record software system. The output of these design control verification analysis documents meets its requirements and design specifications as intended.

Software verification and validation were performed internally and audited by an external agency for the company's software verification and validation protocol, procedures and policies. The audit reports from external audit were submitted.

510K Summary Conclusions:

We believe that technological characteristics, the operational characteristics, mode of operations, the performance and labelling information of the submitted B.O.L.T devices are similar to the predicate devices and systems currently in the market. The design of the submitted device specifications are substantially similar to the predicate devices.

The standards used for clinical and non-clinical tests are according to the standards recommended for the industry by FDA.

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification AmZetta Technologies Private Limited. concludes that the B.O.L.T patient monitoring system, model VA01, VA01-0, VA01-1 and VA01-2 are substantially equivalent to predicate devices as described herein.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).