K Number
K182235

Validate with FDA (Live)

Date Cleared
2019-05-20

(276 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Bluestone Medical, Inc. OWTSH-I Sharps Container is a single- use, disposable, non-sterile container intended to be used for health-care purposes for safe disposal of hazardous sharps such as hypodermic needles, syringes, lancets and blood needles. The target population is for qualified personnel in health care facilities and in other facilities in which medical sharps may be used. The container is intended to be used in areas where there is no unsupervised patient access.

Device Description

The OnSite Waste Sharps Container OWTSH-I is constructed using tinplate steel base with a Nylon and polypropylene restrictive lid system. The OnSite Waste Sharps Container OWTSH-I measures 10 1/2" wide x 4 3/4" deep, and tapers to a dimension of 9" wide x 3 1/8 deep. The height of the container is 8 1/4" in height. The OnSite Waste Sharps Container OWTSH-I is equipped with a locking lid that once closed, prevents anyone from opening it either before or after processing. Labels are on a red background with printed in black text and a bio-hazzard symbol. Labels are adhered to the container at the time of manufacture with the fill line warning printed on the label There is no feature to bend, break, or shear needle, includes blunting of needle in container.

AI/ML Overview

This FDA 510(k) summary for the OnSite Waste Sharps Container (K182235) details the performance testing against established standards. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Test Method (Acceptance Criteria)Reported Device Performance
Puncture Resistance
ASTM F2132-01 (2008) "Standard Specification for the Puncture Resistance of Materials used in containers for the Discarded Medical Needles and Other Sharps"Passed
Leak Resistance
Container filled with water, no visual leakage after 24 hoursPassed
Impact Resistance
Based on ISO 23907:2012Passed
Stability
Based on ISO 23907:2012Passed

Note: The document explicitly states "There were no clinical testing required to support the medical device," meaning the acceptance criteria and performance are based solely on non-clinical (laboratory) tests.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for each non-clinical test. However, it indicates that the puncture resistance test was "Performed by an independent materials testing lab," suggesting the use of laboratory samples. The data provenance is non-clinical laboratory testing, likely conducted in the United States or at a facility accredited for these ISO/ASTM standards. The testing is retrospective in the sense that it evaluates the manufactured product's performance against pre-defined standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. For non-clinical performance tests of a physical device like a sharps container, ground truth is established by objective measurements and adherence to specified testing standards (ASTM, ISO). Expert consensus is not typically involved in determining if a physical property (like puncture resistance or leak-proofness) meets a technical specification.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods are relevant for subjective assessments, especially in clinical or image-based studies. For physical performance tests, the results are typically determined by direct measurement or observation against a predefined pass/fail criterion.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI Vs Without AI Assistance

Not applicable. This device is a physical sharps container and does not involve AI, human readers, or image interpretation. Therefore, an MRMC comparative effectiveness study is not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a physical sharps container and does not involve an algorithm.

7. The Type of Ground Truth Used

The ground truth used for this device is based on established objective technical standards and specifications. Specifically:

  • Puncture Resistance: Defined by ASTM F2132-01 (2008).
  • Leak Resistance: Defined by visual observation against a "no leakage" criterion over 24 hours.
  • Impact Resistance: Defined by ISO 23907:2012.
  • Stability: Defined by ISO 23907:2012.

8. The Sample Size for the Training Set

Not applicable. This device is a physical medical device. It does not involve machine learning algorithms that require a "training set." The performance is evaluated based on material science and engineering principles.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set is relevant for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 20, 2019

Bluestone Medical, Inc. Jon Bricken President (DBA OnSite Waste Technologies, Inc.) 9 North Main Street East Hampton, New York 11937

Re: K182235

Trade/Device Name: OnSite Waste Sharps Container Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: MMK Dated: March 26, 2019 Received: April 15, 2019

Dear Jon Bricken:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For David Krause, PhD Acting Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182235

Device Name OnSite Waste Sharps Container

Indications for Use (Describe)

Bluestone Medical, Inc. OWTSH-I Sharps Container is a single- use, disposable, non-sterile container intended to be used for health-care purposes for safe disposal of hazardous sharps such as hypodermic needles, syringes, lancets and blood needles. The target population is for qualified personnel in health care facilities in which medical sharps may be used. The container is intended to be used in areas where there is no unsupervised patient access.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a circular logo with three sections. The top left section is green, the top right section is yellow, and the bottom section is black. The sections are curved and appear to be connected, forming a continuous loop. The logo has a grainy texture.

Bluestone Medical, Inc. 9 North Main Street, suite #7 East Hampton, N.Y. 11937

510(k) Summary

The assigned 510(k) number is: K182235

    1. Date Prepared: May 14, 2019
    1. Submitter's Identification

Bluestone Medical, Inc

Address: 9 North Main Street, suite #7 East Hampton, N.Y. 11937

Contact person: Jon Bricken President

(914)-886-3958

Email: jon.bricken@onsitewaste.com

    1. Name of the Device
      Device Name: Sharps Container

Trade Name: OnSiteWaste Sharps Container

FDA identification: OWTSH-I

    1. Classification Information:
      Product Code: MMK

Device Class: Class II

Classification: assessors to hypodermic single lumen needles

Classification Panel: General Hospital

    1. Predicate Device Information:
      Trade name: Bradford Wright Sharps Container

K182235 1/5

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a circular logo with three distinct sections. The sections are colored in green, yellow, and black. The sections are separated by white lines, creating a visually appealing and modern design.

Bluestone Medical, Inc. 9 North Main Street, suite #7 East Hampton, N.Y. 11937

Common name: Sharps Container

Product code: MMK

Classification: Accessory to hypodermic single lumen needles

CFR Reference: 21 CFR880.5570-Class II

Classification: panel: General Hospital

Legally Marketed Predicate Device:

CompanyProduct description510(k) #
Bradford Wright Companies, Inc.Demolizer II Sharps ContainerK131708 (Primary Predicate)

6. Intended use / Indication for Use:

Bluestone Medical, Inc. OWTSH-I Sharps Container is a single- use, disposable, non-sterile container intended to be used for health-care purposes for safe disposal of hazardous sharps such as hypodermic needles, syringes, lancets and blood needles. The target population is for qualified personnel in health care facilities and in other facilities in which medical sharps may be used. The container is intended to be used in areas where there is no unsupervised patient access.

7. Device Description:

The OnSite Waste Sharps Container OWTSH-I is constructed using tinplate steel base with a Nylon and polypropylene restrictive lid system. The OnSite Waste Sharps Container OWTSH-I measures 10 1/2" wide x 4 3/4" deep, and tapers to a dimension of 9" wide x 3 1/8 deep. The height of the container is 8 1/4" in height. The OnSite Waste Sharps Container OWTSH-I is equipped with a locking lid that once closed, prevents anyone from opening it either before or after processing.

2/5 K182235

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a circular logo with three teardrop-shaped segments. The segments are arranged in a clockwise direction and are colored differently. The top-left segment is a dark green, the top-right segment is a golden yellow, and the bottom segment is black. The segments are arranged in a way that they appear to be rotating around a central point.

Bluestone Medical, Inc. 9 North Main Street, suite #7 East Hampton, N.Y. 11937

Labels are on a red background with printed in black text and a bio-hazzard symbol. Labels are adhered to the container at the time of manufacture with the fill line warning printed on the label There is no feature to bend, break, or shear needle, includes blunting of needle in container.

ComponentMaterialDepth(inches)Width(inches)Height(inches)
Container at TopTinplate Steel4"10"8"
Container at BottomTinplate Steel3"9"N.A.
Sharps Chute closedMolded Polypropylene4 3/4"10 1/2"2 1/4"
Sharps Chute aloneMolded Polypropylene3 3/4"9 1/2"2"
Top RimMolded Polypropylene4 3/4"10 1/2"1/4"
Lid PlugMolded Polypropylene3 3/4"9 1/4"3/4"
Sharps ContainerMetal & Plastic4 3/4"10 1/2"8 1/4"

OnSite Waste Sharps Container OWTSH-I

Technological Characteristics 8.

Element of ComparisonSubject DevicePredicate Device
CompanyBluestone Medical, Inc.Bradford Wright Companies, Inc.
Device NameOWTSH-I Sharps ContainerBWCSH-i Sharps Container
FDA 510(k) NumberK182235K131708
Intended UsesContainer is a single- use,disposable, non-sterilecontainer intended to be usedfor health-care purposes forsafe disposal of hazardoussharps such as hypodermicneedles, syringes, lancetsand blood needles.sharps container has a one-gallon fillline capacity that is intended to beused for the safe disposal of hazardoussharps.
Target PopulationHealthcare ProfessionalsHealthcare Professionals
MaterialsTin Plate & PlasticTin Plate & Plastic

{6}------------------------------------------------

Element of ComparisonSubject DevicePredicate Device
Sharps AccessSharps inserted through thetopSharps inserted through the top
Sharps ClosureLid locked in placeLid locked in place
Impact ResistanceYesYes
Puncture ResistanceYesYes
Leak ProofYesYes
Non-SterileYesYes
Single UseYesYes
No features to bend, breakor shear needleNo Features PresentNo Features Present
Fill indicationsLabeled "Do Not Fill AboveThis Line"Labeled "Do Not Fill Above This Line"
ConstructionInjection molded lid and TinPlate baseInjection molded lid and Tin Plate base
Dimensions10.5"x4.75"x8.25"10.5"x4.75"x8.25"
Fill CapacityOne US GallonOne Us Gallon
  • Summary of Non-Clinical Testing 9.
    Test Methods:

Puncture Resistance (Performed by and independent materials testing lab) - Passed ASTM F 2132-01 (2008) · "Standard Specification for the Puncture Resistance of Materials used in containers for the Discarded Medical Needles and Other Sharps".

Leak Resistance of bottom and sides - Container is filled with water. No Leakage visually observed after 24 hour period - Passed.

Impact Resistance - Passed Based on ISO 23907:2012

Stability: - Passed Based on ISO 23907:2012

{7}------------------------------------------------

10. Discussion of Clinical Tests Performed: There were no clinical testing required to support the medical device.

Conclusions: 11.

The results of the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the predicate device

1

ﻳﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ

្រព្យ

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).