K Number
K182030
Device Name
Faros Mobile
Date Cleared
2018-11-08

(101 days)

Product Code
Regulation Number
870.2800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Faros Mobile system is intended for use in clinical long term ambulatory ECG monitoring, data transfer and analysis. Faros Mobile is indicated for adult and pediatric patients who require ECG monitoring inside or outside hospital or healthcare facility environments.

Faros Mobile provides the detection and reporting features appropriate for the indications below

  • · Evaluation of patients with symptoms related to rhythm disturbances or symptoms suggesting arrhythmias.
  • · Evaluation of patients for ST segment changes
  • · Evaluation patients with pacemaker
  • · Evaluating patient rest and stress ECG
  • · Reporting heart rate variability analysis
  • · Wireless transmission of patient ECG data and arrhythmia events for further analysis.

Interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information. The Faros Mobile does not provide interpretive statements.

Faros Mobile is contraindicated for

  • · Those patients requiring attended, in-hospital monitoring for life threatening arrhythmias
  • · Pediatric patients weighting less than 10 kg
Device Description

Faros Mobile is used for long term registration of a patient's heart electrocardiogram (ECG) and the wireless or wired transmission of the registered electrocardiogram arrhythmia events to a receiver or analysis system running on computer. The device is connected to a patient via electrode leads and the patient wears the device during the recording period and/or while he or she performs normal daily activities. The device is used with commercially available snap ECG electrodes.

Data is transferred to computer via USB. The ECG recording is analyzed at a medical facility using a Cardiac Navigator or Cardiac Explorer software.

In wireless transmission mode, Bluetooth communication protocol enables use of the Faros sensor as a part of a Mobile Cardiac Telemetry (MCT) or Cardiac Event Monitor (CEM) system. Data is transferred from device to companion device. For clarity, all the data is stored on sensor memory when the device operates via Bluetooth. The communication protocol is provided for 30° party integration purposes.

Device measurement configurations are managed via Faros Device Manager. The application operator can manage measurement configurations which include: ECG and motion data sampling frequencies, ECG channel count, heart rate variability, temperature, auto start and measurement auto stop features. Values for cardiac arrhythmia event detections are handled via Faros Manager application.

Faros Mobile system consists of:

  • . Faros ECG sensor
    • o Bluetooth communication protocol and documentation for implementing interface for data collection from Faros sensor
    • channel and 3-channel cable sets o
    • general micro-USB cable for recharging device O
  • Faros Manager application
  • Cardiac Explorer application
  • Cardiac Navigator application
AI/ML Overview

The provided text is a 510(k) Premarket Notification for the Faros Mobile device, and it does not contain a specific study proving the device meets acceptance criteria in the way typically associated with algorithmic performance (e.g., accuracy, sensitivity, specificity benchmarks).

Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (CARDIO SPY ECG Holter Systems by LABTECH KFT.) and a reference device (eMotion Faros ECG Mobile by Mega Electronics Ltd.) through non-clinical testing and conformance to recognized voluntary standards.

Therefore, many of the requested items regarding acceptance criteria and performance studies for an algorithm's performance cannot be directly extracted from this document, as the submission primarily relies on the established safety and efficacy of the predicate device and standardized testing for hardware and software functionality.

However, I can extract information regarding the general acceptance criteria implied by regulatory standards and the types of testing performed to demonstrate substantial equivalence.

Here's a breakdown based on the provided text, addressing what can be answered and where information is not present:


Implied Acceptance Criteria and Device Performance (Based on Substantial Equivalence and Standards Conformance)

Since a direct algorithmic performance study with specific numerical acceptance criteria is not provided, the "acceptance criteria" here are inferred from the demonstrated conformance to regulatory standards and the equivalence to predicate devices. The device performance is generally stated as meeting these standards.

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Inferred from Standards & Substantial Equivalence)Reported Device Performance and Conformance (From Section 8 & 10)
Safety and Essential Performance (General): Adherence to fundamental safety and performance requirements for medical electrical equipment.Meets IEC 60601-1:2005+AMD1:2012 (Medical Electrical Equipment - Part 1 General Requirements for Basic Safety and Essential Performance). "Performance data demonstrated that Faros Mobile is effective, secure, safe and suitable for its indication for use..."
Electromagnetic Compatibility (EMC): Device operates without being unduly affected by, or unduly affecting, electromagnetic interference.Meets IEC 60601-1-2: 2014 (Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests).
Specific Requirements for Ambulatory ECG Systems: Conformance to particular safety and performance aspects relevant to ambulatory electrocardiographic systems.Meets IEC 60601-2-47:2012 (Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems). The predicate device also meets this standard.
Biocompatibility: Materials in contact with the patient are safe and do not cause adverse biological reactions.Meets ISO 10993-1: 2015 (Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing).
Home Healthcare Environment Suitability: Safe and effective for use in a home environment.Meets IEC 60601-1-11:2015 (Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment).
Usability Engineering: Device is designed with user safety and efficiency in mind, minimizing use errors.Meets IEC 62366-1:2015 (Medical devices -- Application of usability engineering to medical devices) and IEC 60601-1-6:2010+AMD1:2013 (Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability).
Software Life Cycle Processes: Software is developed and maintained according to recognized quality and safety standards.Meets IEC 62304:2006 (Medical device software -- Software life cycle processes).
Functional Equivalence to Predicate: Device functionality for ECG recording, data transfer, and analysis, and interpretation algorithm capabilities (which require physician confirmation) are comparable to the predicate.The document extensively compares the Faros Mobile to the predicate (K140847) and reference (K143032) devices in Table 1 ("Comparison of technological characteristics") and Table 2 ("Comparison of indications for use"), stating: "The technological characteristics and principles of operation of the Faros Mobile are the same as the predicate device." It explicitly states that "Interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician..."
No New Questions of Safety/Effectiveness: The device does not introduce new safety or effectiveness concerns compared to the predicate device."It has been shown in this Abbreviated 510(k) submission that the difference between the Faros Mobile and the predicate devices do not raise any new questions regarding its safety and effectiveness."

Information Not Present in the Document for Algorithmic Performance

The following points pertain to a specific study proving device meets acceptance criteria for algorithmic performance (e.g., accuracy in arrhythmia detection), which is not detailed in this 510(k) submission. The submission relies on "performance tests" that validate compliance with general safety and functional standards, rather than detailing a specific clinical or annotated dataset study for the algorithm's diagnostic accuracy.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not provided. The document states "ECG and algorithm performance tests" were done, but does not detail the dataset used for these tests, its size, or provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not provided. Ground truth establishment for a test set is not detailed. The interpretation algorithm's analysis "must be confirmed by a physician with other relevant clinical information," suggesting the algorithm is an aid, but no specific number or qualification of experts for validating the algorithm's performance is mentioned.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not provided.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC study is not mentioned. The document states "There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not detailed. While "algorithm performance tests" are mentioned, detailed results or methodology for a standalone performance study (e.g., specific sensitivity/specificity values for arrhythmia detection on a labeled dataset) are not provided. The algorithm is presented as providing a "computer-generated analysis" that "must be confirmed by a physician," implying it's not intended for standalone diagnostic use.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not provided.

8. The sample size for the training set

  • Not provided.

9. How the ground truth for the training set was established

  • Not provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, with the letters "FDA" in a blue box. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 8, 2018

Bittium Biosignals Ltd. Taneli Vaaraniemi Quality Manager Pioneerinkatu 6 Kuopio. Finland FI-70800

Re: K182030

Trade/Device Name: Faros Mobile Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MLO, DSI, DXH Dated: October 5, 2018 Received: October 9, 2018

Dear Taneli Vaaraniemi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica E. Paulsen -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182030

Device Name Faros Mobile

Indications for Use (Describe)

The Faros Mobile system is intended for use in clinical long term ambulatory ECG monitoring, data transfer and analysis. Faros Mobile is indicated for adult and pediatric patients who require ECG monitoring inside or outside hospital or healthcare facility environments.

Faros Mobile provides the detection and reporting features appropriate for the indications below

  • · Evaluation of patients with symptoms related to rhythm disturbances or symptoms suggesting arrhythmias.
  • · Evaluation of patients for ST segment changes
  • · Evaluation patients with pacemaker
  • · Evaluating patient rest and stress ECG
  • · Reporting heart rate variability analysis
  • · Wireless transmission of patient ECG data and arrhythmia events for further analysis.

Interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information. The Faros Mobile does not provide interpretive statements.

Faros Mobile is contraindicated for

  • · Those patients requiring attended, in-hospital monitoring for life threatening arrhythmias
  • · Pediatric patients weighting less than 10 kg
Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 005 - Abbreviated 510k Summary

1. Submitter

Bittium Biosignals Ltd.Bittium Biosignals Ltd.Pioneerinkatu 6Kuopio, Finland FI-70800
Phone+358 17 581 7700
Fax:+358 17 580 0978
Registration number1000425315
ContactTaneli Vääräniemi (taneli.vaaraniemi@bittium.com)Quality Manager

2. Date

November 7, 2018

3. Device identification

Device nameFaros Mobile
Trade/Proprietary Name:Faros ambulatory ECG recorder, telemetric
Common/Usual Name:Digital Ambulatory ECG recorder, transmitter
Classification Product Code:MLO - Electrocardiograph, Ambulatory, with Analysis Algorithm
Regulation Classification870.2800
Subsequent Product CodeDSI - Arrhythmia detector and alarm
Regulation Classification870.1025
Subsequent Product CodeDXH - Transmitters and Receivers, Electrocardiograph, Telephone
Regulation Classification870.2920
Class of Device:Class II
Panel:Cardiovascular
Device Classification EU:Class II
Medical Device Directive:93/42EEC as an amended by 2007/47/EC
Certificate issued by VTT Expert services Ov. NB0537

{4}------------------------------------------------

Abbreviated 510(k) Premarket Submission

4. Substantial equivalence, predicate device

K140847, CARDIO SPY ECG Holter Systems by LABTECH KFT.

Reference device: technology, design functionality

K143032, eMotion Faros ECG Mobile by Mega Electronics Ltd.

5. Indications for Use Statement

The Faros Mobile system is intended for use in clinical long term ambulatory ECG monitoring, data transfer and analysis. Faros Mobile is indicated for adult and pediatric patients who require ECG monitoring inside or outside hospital or healthcare facility environments.

Faros Mobile provides the detection and reporting features appropriate for the indications below

  • . Evaluation of patients with symptoms related to rhythm disturbances or symptoms suggesting arrhythmias.
  • Evaluation of patients for ST segment changes
  • Evaluation patients with pacemaker
  • Evaluating patient rest and stress ECG
  • Reporting heart rate variability analysis
  • Wireless transmission of patient ECG data and arrhythmia events for further analysis.

Interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information. The Faros Mobile does not provide interpretive statements.

Faros Mobile is contraindicated for

  • . Those patients requiring attended, in-hospital monitoring for life threatening arrhythmias
  • Pediatric patients weighting less than 10 kg

{5}------------------------------------------------

6. Device Description

Faros Mobile is used for long term registration of a patient's heart electrocardiogram (ECG) and the wireless or wired transmission of the registered electrocardiogram arrhythmia events to a receiver or analysis system running on computer. The device is connected to a patient via electrode leads and the patient wears the device during the recording period and/or while he or she performs normal daily activities. The device is used with commercially available snap ECG electrodes.

Data is transferred to computer via USB. The ECG recording is analyzed at a medical facility using a Cardiac Navigator or Cardiac Explorer software.

In wireless transmission mode, Bluetooth communication protocol enables use of the Faros sensor as a part of a Mobile Cardiac Telemetry (MCT) or Cardiac Event Monitor (CEM) system. Data is transferred from device to companion device. For clarity, all the data is stored on sensor memory when the device operates via Bluetooth. The communication protocol is provided for 30° party integration purposes.

Device measurement configurations are managed via Faros Device Manager. The application operator can manage measurement configurations which include: ECG and motion data sampling frequencies, ECG channel count, heart rate variability, temperature, auto start and measurement auto stop features. Values for cardiac arrhythmia event detections are handled via Faros Manager application.

Faros Mobile system consists of:

  • . Faros ECG sensor
    • o Bluetooth communication protocol and documentation for implementing interface for data collection from Faros sensor
    • channel and 3-channel cable sets o
    • general micro-USB cable for recharging device O
  • Faros Manager application
  • Cardiac Explorer application
  • Cardiac Navigator application

{6}------------------------------------------------

7. Comparison of technological characteristics

The following table compares the Faros Mobile to the predicate device with respect to indications for use, technological characteristics and principles of operation providing more information regarding the basis for the determination of substantial equivalence.

The technological characteristics and principles of operation of the Faros Mobile are the same as the predicate device.

ManufacturerBittium Biosignals LtdLABTECH KFT.Mega Electronics Ltd
Trade nameFaros MobileCARDIOSPY ECG Holter SystemseMotion Faros ECG Mobile
Subject devicePredicate deviceReference device
510(k) numberK182030K140847K143032
Product code and regulation numberDSI - 21 CFR 870.1025MLO - 21 CFR 870.2800DXH - 21 CFR 870.2920
Subsequent Product Code and regulation numberDXH - 21 CFR 870.2920MLO 21 CFR 870.2800n/an/a
ECG channels1- or 3-channels1, 2, 3, 12 -channels1- or 3-channels
ECG performance standardIEC 60601-2-47IEC 60601-2-47IEC 60601-2-25
RR AlgorithmYesYesYes
Cardiac arrhythmia detectionYesYesNo
Event markerYesYesYes
Data transferYesYesYes
Battery typeLi Ion battery1x1.2 V AAA NiMH battery(or 1x1.5 V AAA alkaline battery)Li Ion battery
SoftwareCardiac Navigator, Cardiac ExplorerCardiospy softwareLiveECG and Virtual Clinic
Simple, user friendly software with multiple functionsYesYesYes
Full disclosure ECG review toolYesYesYes
Precise QRS classification and rhythm analysisYesYesNo
Arrhythmia analysis, arrhythmia overviewYesYesNo
Atrial Fibrillation analysisYesYesNo
ManufacturerBittium Biosignals LtdLABTECH KFT.Mega Electronics Ltd
Trade nameFaros MobileCARDIOSPY ECG Holter SystemseMotion Faros ECG Mobile
Indications for useThe Faros Mobile system isintended for use in clinicallong term ambulatory ECGmonitoring, data transferand analysis. Faros Mobile isindicated for adult andpediatric patients whorequire ECG monitoringinside or outside hospital orhealthcare facilityenvironments.Faros Mobile provides thedetection and reportingfeatures appropriate for theindications below• Evaluation of patients withsymptoms related to rhythmdisturbances or symptomssuggesting arrhythmias.• Evaluation of patients forST segment changes• Evaluation patients withpacemaker• Evaluating patient rest andstress ECG• Reporting heart ratevariability analysis• Wireless transmission ofpatient ECG data andarrhythmia events forfurther analysis.Interpretation algorithmprovides a computer-generated analysis ofpotential patient cardiacabnormalities, which mustbe confirmed by a physicianwith other relevant clinicalinformation. The FarosMobile does not provideinterpretive statements.Faros Mobile isThe Cardiospy ECG HolterSystem is intended for use ina clinical, by qualifiedprofessionals, for patientsrequiring ambulatory(Holter) monitoring 24, 48,72, 168 h hours. Suchmonitoring is mostfrequently used for purposeof prospective andretrospective cardiac dataand arrhythmia analysis. TheSystem, among others,provides the detection andreporting featuresappropriate to theindications below forchildren and adults of allages (up to 2 years):• Evaluation of patients withsymptoms related to rhythmdisturbances or symptomssuggesting arrhythmia.• Evaluation of patients forST segment changes• Evaluation of patients withpacemakers• Reporting of time domainheart rate variability• Evaluation of a patient'sresponse after resumingoccupational or recreationalactivities (e.g., after MI. orcardiac surgery.)• Evaluation of ECGdocumenting therapeuticinterventions in individualpatients orgroups of patients• Clinical andepidemiological researchstudiesThe eMotion Faros ECGsensor is a wearable,portable, externally applied,electrocardiographrecorder and transmitter forthe purpose of healthmonitoring, biofeedback andscientific research.The eMotion Faros ECGMobile is intended for use inclinical and non-clinicalsettings to collect andtransmit health parametersto healthcare professionalsfor monitoring andevaluation.Health parameters arecollected from a variety ofcommercially available,external plug-in devices suchas ECG sensors, WeightScales, Blood PressureMeters and Pulse OximetersIndicated for adult patientswho require clinical or non-clinical ECG monitoring inhealthcare facilityenvironment undersupervision of a physician orprescript by the supervisingphysician to supplementdata acquisition in homeenvironment.The eMotion Faros ECGMobile does not provide anyautomatic analysis ordiagnosis.
contraindicated for
• Those patients requiringattended, in-hospitalmonitoring for lifethreatening arrhythmias
• Pediatric patientsweighing less than 10 kg

Table 1: Comparison of characteristics

{7}------------------------------------------------

Table 2: Comparison of indications for use

{8}------------------------------------------------

Abbreviated 510(k) Premarket Submission

{9}------------------------------------------------

Abbreviated 510(k) Premarket Submission

8. Performance Data Summary

Faros Mobile submission was written according to and in conformance with FDA Guidance "Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm" released on October 2003. The test reports in the submission demonstrate the Faros Mobile meets its intended use and design requirements.

The verification, validation, and testing activities establish the performance and functionality characteristics of the Faros Mobile. Testing comprised electrical and mechanical safety tests, environmental tests, ECG and algorithm performance tests, usability and functional tests.

Performance data demonstrated that Faros Mobile is effective, secure, safe and suitable for its indication for use, and data has shown conformance to following voluntary FDA recognized standards:

  • . IEC 60601-1:2005+AMD1:2012, Medical Electrical Equipment - Part 1 General Requirements for Basic Safety and Essential Performance
  • IEC 60601-1-2: 2014, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests
  • . IEC 60601-2-47:2012, Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems
  • ISO 10993-1: 2015, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing
  • IEC 60601-1-11:2015, Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • IEC 62366-1:2015, Medical devices -- Application of usability engineering to medical devices
  • IEC 62304:2006, Medical device software -- Software life cycle processes
  • IEC 60601-1-6:2010+AMD1:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

{10}------------------------------------------------

9. Clinical Testing

There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The nonclinical testing performed supports the substantial equivalence of the device.

10. Statement of Substantial Equivalence

Faros Mobile device is substantially equivalent to a predicate device when the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the modified device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.

It has been shown in this Abbreviated 510(k) submission that the difference between the Faros Mobile and the predicate devices do not raise any new questions regarding its safety and effectiveness. Performance testing and compliance with voluntary standards, demonstrate that the Faros Mobile is substantially equivalent to the relevant aspects of the predicate devices in terms of design, principle of operation, performance characteristics, and intended use.

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).