K Number
K181864
Device Name
Polaris RF Ablation System
Date Cleared
2019-01-02

(174 days)

Product Code
Regulation Number
882.4400
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue.
Device Description
The Polaris RF Ablation System includes the following components: 1. Polaris Radiofrequency Generator and Desk Stand 2. Polaris Pump, Pump Cable and Desk Stand 3. Polaris Standard Connector Hub / Polaris Cooled Connector Hub 4. Polaris Footswitch 5. Polaris Cooled RF Probe Kit: i. Polaris Cooled RF Ablation Probe ii. Polaris Tube Kit iii. Polaris Introducer 6. Polaris Single-Use RF Probe 7. Polaris Reusable RF Probe 8. Polaris RF Cannula The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue. The system is designed to deliver controlled RF energy from the Polaris RF Generator to target tissues via the Polaris RF Probes for standard RF procedures or Polaris Cooled RF Probes for cooled RF procedures. RF energy is applied based on the configured RF generator settings to create the desired lesions in the target tissue. The generator also delivers low frequency stimulation pulses during procedures. During standard RF procedures, the Polaris RF Probe is used with a compatible Polaris RF Cannula to enable RF energy delivery via the active electrode to the target tissues. During cooled RF procedures with Polaris Cooled RF Probes, the system integrates a cooling mechanism by internally circulating water through the probes using the Polaris Pump Unit. This cooling results in minimal charring of the surrounding tissue and also prevents tissue from adhering to the electrodes. This allows for larger and more consistent ablation volumes to be achieved.
More Information

Unknown

No
The description focuses on the hardware components and the physical process of delivering RF energy and cooling, with no mention of AI or ML for analysis, decision-making, or control.

Yes.
The device is intended for the creation of radiofrequency lesions in nervous tissue, which is a therapeutic intervention aimed at treating medical conditions.

No
The device's intended use is for the creation of radiofrequency lesions in nervous tissue, which is a therapeutic intervention, not a diagnostic one.

No

The device description explicitly lists multiple hardware components including a generator, pump, probes, cannula, and footswitch. While software is mentioned in the performance studies, the device is clearly a system with significant hardware components.

Based on the provided information, the Polaris RF Ablation System is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • Intended Use: The intended use is "for the creation of radiofrequency lesions in nervous tissue." This describes a therapeutic procedure performed directly on a patient's body, not a test performed on a sample taken from the body.
  • Device Description: The components listed are all designed for delivering energy and managing the procedure within the body (generator, probes, pump, etc.). There are no components mentioned for analyzing biological samples.
  • Lack of IVD Characteristics: The description does not mention any analysis of biological samples (blood, urine, tissue, etc.) or the use of reagents, which are hallmarks of IVD devices.
  • Anatomical Site: The target is "Nervous tissue," indicating a direct interaction with the patient's anatomy.

IVD devices are used to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes. The Polaris RF Ablation System is a therapeutic device used to treat conditions by creating lesions in nervous tissue.

N/A

Intended Use / Indications for Use

The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue.

Product codes

GXD, GXI

Device Description

The Polaris RF Ablation System includes the following components:

  1. Polaris Radiofrequency Generator and Desk Stand
  2. Polaris Pump, Pump Cable and Desk Stand
  3. Polaris Standard Connector Hub / Polaris Cooled Connector Hub
  4. Polaris Footswitch
  5. Polaris Cooled RF Probe Kit:
    i. Polaris Cooled RF Ablation Probe
    ii. Polaris Tube Kit
    iii. Polaris Introducer
  6. Polaris Single-Use RF Probe
  7. Polaris Reusable RF Probe
  8. Polaris RF Cannula

The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue. The system is designed to deliver controlled RF energy from the Polaris RF Generator to target tissues via the Polaris RF Probes for standard RF procedures or Polaris Cooled RF Probes for cooled RF procedures. RF energy is applied based on the configured RF generator settings to create the desired lesions in the target tissue. The generator also delivers low frequency stimulation pulses during procedures.

During standard RF procedures, the Polaris RF Probe is used with a compatible Polaris RF Cannula to enable RF energy delivery via the active electrode to the target tissues. During cooled RF procedures with Polaris Cooled RF Probes, the system integrates a cooling mechanism by internally circulating water through the probes using the Polaris Pump Unit. This cooling results in minimal charring of the surrounding tissue and also prevents tissue from adhering to the electrodes. This allows for larger and more consistent ablation volumes to be achieved.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Nervous tissue

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Physicians familiar with RF lesion techniques

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance testing was completed to demonstrate substantial equivalence of the Polaris RF Ablation System and relevant system components to the identified predicate devices in Table 7.1. The system components were subjected to the verification and validation testing listed in Table 7.4.

Polaris Re-Usable RF Probe

  • Mechanical Testing:
    • Test Method Summary: Performed Temperature Test, Pull Test, Cleaning and Sterilization Reuse Test, Mechanical Reuse Test, Mechanical Test (Flexion), Continuity Test, Functional Test.
    • Results: All samples passed acceptance criteria, confirming compliance with IEC 60601-2-2:2009 and supporting substantial equivalence.
  • Electrical Testing:
    • Test Method Summary: Electrical verification by comparison with Polaris Standard Single-Use Probe to assess electrical stress resistance.
    • Results: Verified to meet requirements of IEC 60601-1:2005+A1:2012 and IEC 60601-2-2:2009, supporting substantial equivalence.

Biocompatibility

  • Test Method Summary: Demonstrated through comparisons to similar legally marketed devices per ISO 10993-1 and FDA guidance.
  • Results: Confirmed compliance with ISO 10993-1 requirements and supported substantial equivalence.

Packaging

  • Test Method Summary: Ship testing performed to ensure integrity through shipping and handling.
  • Results: All packaging met relevant requirements, supporting substantial equivalence.

Cleaning and Sterilization

  • Test Method Summary: Validated recommended reprocessing methods in Instructions for Use to achieve a Sterility Assurance Level of 10-6.
  • Results: Verified to meet AAMI TIR12, AAMI TIR30, ANSI/AAMI ST81 requirements and FDA guidance, supporting substantial equivalence.

Polaris Single-Use RF Probe

  • Mechanical Testing:
    • Test Method Summary: Performed Temperature Test, Pull Test, Durability Test, Mechanical Test (Flexion), Continuity Test, Functional Test.
    • Results: All samples passed acceptance criteria, confirming compliance with IEC 60601-2-2:2009 and supporting substantial equivalence.
  • Electrical Testing:
    • Test Method Summary: Performed High Frequency Leakage Current Test - Cable, High Frequency Dielectric Strength Test – Cable, High Frequency Dielectric Strength Test - Handle, Mains Frequency Dielectric Strength Test - Cable, Mains Frequency Dielectric Strength Test - Handle, Mains Frequency Dielectric Strength Test – Connector.
    • Results: All samples passed acceptance criteria, confirming compliance with IEC 60601-1:2005+A1:2012 and IEC 60601-2-2:2009, supporting substantial equivalence.

Biocompatibility

  • Test Method Summary: Demonstrated through comparisons to similar legally marketed devices per ISO 10993-1 and FDA guidance.
  • Results: Confirmed compliance with ISO 10993-1 requirements and supported substantial equivalence.

Packaging

  • Test Method Summary: Ship testing performed to ensure integrity through shipping and handling. Seal strength and sterile barrier integrity validated per ANSI/AAMI/ISO 11607-1 and 11607-2 over shelf life.
  • Results: Met relevant requirements, confirming compliance with ANSI/AAMI/ISO 11607-1 and 11607-2, supporting substantial equivalence.

Polaris RF Cannula

  • Mechanical Testing:
    • Test Method Summary: Performed Positive Pressure Liquid Leakage Test, Sub-atmospheric Pressure Air Leakage Test, Resistance to Separation Test, Stress Cracking Test, Hub Strength Test.
    • Results: All samples passed acceptance criteria, confirming compliance with ISO 80369-7:2016, ISO 594-1:1986, and ISO 7864:2016, supporting substantial equivalence.
  • Electrical Testing:
    • Test Method Summary: Performed High Frequency Leakage Current Test, High Frequency Dielectric Strength Test, Mains Frequency Dielectric Strength Test.
    • Results: All samples passed acceptance criteria, confirming compliance with IEC 60601-1:2005+A1:2012 and IEC 60601-2-2:2009, supporting substantial equivalence.

Biocompatibility

  • Test Method Summary: Demonstrated through biological testing per ISO 10993-1 and FDA guidance.
  • Results: Confirmed compliance with ISO 10993-1 requirements and supported substantial equivalence.

Packaging

  • Test Method Summary: Ship testing performed to ensure integrity through shipping and handling. Seal strength and sterile barrier integrity validated per ANSI/AAMI/ISO 11607-1 and 11607-2 over shelf life.
  • Results: Met relevant requirements, confirming compliance with ANSI/AAMI/ISO 11607-1 and 11607-2, supporting substantial equivalence.

Polaris Cooled RF Probe Kit

  • Mechanical Testing (Polaris Cooled RF Probe):
    • Test Method Summary: Performed Leak/Pressure Test, Pull Test, Temperature Test, Tip Compression Test, Mechanical Test (Flexion), Continuity Test, Functional Test.
    • Results: All samples passed acceptance criteria, confirming compliance with IEC 60601-2-2:2009 and supporting substantial equivalence.
  • Mechanical Testing (Polaris Introducer):
    • Test Method Summary: Performed Resistance to Separation from Unscrewing Test, Resistance to Separation Test.
    • Results: All samples passed acceptance criteria, supporting substantial equivalence.
  • Electrical Testing (Polaris Cooled RF Probe):
    • Test Method Summary: Performed High Frequency Leakage Current Test - Cable, High Frequency Leakage Current Test - Shaft, High Frequency Dielectric Strength Test – Cable, High Frequency Dielectric Strength Test - Handle/Shaft, Mains Frequency Dielectric Strength Test – Cable, Mains Frequency Dielectric Strength Test - Handle/Shaft, Mains Frequency Dielectric Strength Test Connector.
    • Results: All samples passed acceptance criteria, confirming compliance with IEC 60601-1:2005+A1:2012 and IEC 60601-2-2:2009, supporting substantial equivalence.

Biocompatibility

  • Test Method Summary: Demonstrated through biological testing and comparisons to similar legally marketed devices per ISO 10993-1 and FDA guidance.
  • Results: Confirmed compliance with ISO 10993-1 requirements and supported substantial equivalence.

Packaging

  • Test Method Summary: Ship testing performed to ensure integrity through shipping and handling. Seal strength and sterile barrier integrity validated per ANSI/AAMI/ISO 11607-1 and 11607-2 over shelf life.
  • Results: Met relevant requirements, confirming compliance with ANSI/AAMI/ISO 11607-1 and 11607-2, supporting substantial equivalence.

Polaris Radiofrequency Generator

  • Verification:
    • Test Method Summary: Tested function, including electrical testing of RF board, User Interface Board, and Multiplexer board.
    • Results: Met all relevant requirements, supporting substantial equivalence.

Packaging

  • Test Method Summary: Ship testing performed to ensure integrity through shipping and handling.
  • Results: All packaging met relevant requirements, supporting substantial equivalence.

Polaris Pump Unit

  • Verification:
    • Test Method Summary: Tested specified design requirements, including hardware. Performance assessed for inspection, configuration and Installation, operation, environment, and service life.
    • Results: Met all relevant requirements, supporting substantial equivalence.

Packaging

  • Test Method Summary: Ship testing performed to ensure integrity through shipping and handling.
  • Results: All packaging met relevant requirements, supporting substantial equivalence.

Polaris Connector Hub

  • Verification:
    • Test Method Summary: Tested specified design requirements, including hardware. Performance assessed through Cable Pull Test, Cable Impulse Test, Device Identification.
    • Results: Met all relevant requirements, confirmed compliance with Baylis self-enforced requirements and supported substantial equivalence.

Packaging

  • Test Method Summary: Ship testing performed to ensure integrity through shipping and handling.
  • Results: All packaging met relevant requirements, confirmed compliance of device packaging with Baylis self-enforced requirements and supported substantial equivalence.

Polaris Footswitch

  • Packaging:
    • Test Method Summary: Ship testing performed to ensure integrity through shipping and handling.
    • Results: All packaging met relevant requirements, supporting substantial equivalence.

Polaris RF Ablation System

  • Electrical Product Safety:
    • Test Method Summary: Tested Polaris RF Generator with accessories and detachable parts for compliance with IEC 60601-1:2005+A1:2012 (including ANSI AAMI ES 60601-1:2005(R) 2012 + A1:2012) and IEC 60601-2-2:2009.
    • Results: Met all relevant requirements, confirming compliance and supporting substantial equivalence.
  • Electromagnetic Compatibility:
    • Test Method Summary: Tested Polaris RF Generator with system components for compliance with IEC 60601-1-2:2014.
    • Results: Met all relevant requirements, confirming compliance and supporting substantial equivalence.
  • Benchtop Lesion Validation:
    • Test Method Summary: Comparative lesion validation testing in a soft tissue model to demonstrate ablation performance. Performed for single and multi-probe scenarios.
    • Results: Met all relevant requirements, supporting substantial equivalence.
  • Usability:
    • Test Method Summary: Verified and validated usability requirements. Included normal use cases, foreseeable worst-case use scenario, and testable requirements for primary operating functions.
    • Results: Met all relevant requirements, supporting substantial equivalence.
  • Software:
    • Test Method Summary: Software verification and validation completed. FDA's guidance for software in medical devices used to determine Level of Concern.
    • Results: Met all relevant requirements, supporting substantial equivalence.

All test requirements were met as specified by applicable standards and the test protocols. Results demonstrate that the Polaris RF Ablation System functions as intended and is as safe and effective as the predicate device.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K072478, K062937, K053082, K002389, K972846

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 882.4400 Radiofrequency lesion generator.

(a)
Identification. A radiofrequency lesion generator is a device used to produce lesions in the nervous system or other tissue by the direct application of radiofrequency currents to selected sites.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Baylis Medical Company Inc. May Tsai Team Leader, Regulatory Affairs 2775 Matheson Blvd. East Mississauga, L4W 4P7 Canada

January 2, 2019

Re: K181864

Trade/Device Name: Polaris RF Ablation System Regulation Number: 21 CFR 882.4400 Regulation Name: Radiofrequency Lesion Generator Regulatory Class: Class II Product Code: GXD, GXI Dated: November 30, 2018 Received: December 3, 2018

Dear May Tsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

$$\textbf{John Marker-S}_{\textbf{Date:2019.01.02}\textbf{16:14:56}}^{\textbf{Digitalby signed by John}}$$

For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K181864

Device Name Polaris RF Ablation System

Indications for Use (Describe)

The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

7. 510(k) Summary

7.1 Submitter Information

  • A. Company Name: Baylis Medical Company Inc.
  • B. Company Address: 2775 Matheson Blvd. East Mississauga, Ontario L4W 4P7 Canada
  • C. Company Phone: (905) 602-4875
  • D. Company Facsimile: (905) 602-5671
  • E. Contact Person: May Tsai, Requlatory & Scientific Affairs Team Leader
  • F. Summary Prepared on: 11-Jul-2018

7.2 Device Identification

  • A. Device Trade Name:
    • . Polaris RF Ablation System
  • B. Device Common Name:
    • Radiofrequency lesion generator ●
    • . Radiofrequency lesion probe
  • C. Classification Name:
    • CFR 882.4400 Radiofrequency lesion generator ●
    • CFR 882.4725 Radiofrequency lesion probe .
  • D. Product Code:
    • . GXD, GXI
  • E. Device Class: Class II

4

7.3 Identification of Predicate Device

Table 7.1: Predicate Devices

Predicate DeviceManufacturer510(k)
Baylis Pain Management Generator-TDK072478
Baylis TransDiscal SystemK062937
Baylis Pain Management Cooled ProbeHalyard Health*K053082
Baylis Pain Management Probe and
Connector CableK002389
BMC RF CannulaK972846

*Note: Originally submitted by Baylis Medical Company Inc. as the manufacturer, but product line has since been acquired by Halyard Health.

7.4 Indications for Use

The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue.

7.5 Device Description

The Polaris RF Ablation System includes the following components:

    1. Polaris Radiofrequency Generator and Desk Stand
    1. Polaris Pump, Pump Cable and Desk Stand
    1. Polaris Standard Connector Hub / Polaris Cooled Connector Hub
    1. Polaris Footswitch
    1. Polaris Cooled RF Probe Kit:
    • i. Polaris Cooled RF Ablation Probe
    • ii. Polaris Tube Kit
    • iii. Polaris Introducer
    1. Polaris Single-Use RF Probe
    1. Polaris Reusable RF Probe
    1. Polaris RF Cannula

The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue. The system is designed to deliver controlled RF energy from the Polaris RF Generator to target tissues via the Polaris RF Probes for standard RF procedures or Polaris Cooled RF Probes for cooled RF

5

procedures. RF energy is applied based on the configured RF generator settings to create the desired lesions in the target tissue. The generator also delivers low frequency stimulation pulses during procedures.

During standard RF procedures, the Polaris RF Probe is used with a compatible Polaris RF Cannula to enable RF energy delivery via the active electrode to the target tissues. During cooled RF procedures with Polaris Cooled RF Probes, the system integrates a cooling mechanism by internally circulating water through the probes using the Polaris Pump Unit. This cooling results in minimal charring of the surrounding tissue and also prevents tissue from adhering to the electrodes. This allows for larger and more consistent ablation volumes to be achieved.

7.6 Comparison to Predicate Device

The intended use of the subject Polaris RF Ablation System is the same as the predicate devices. The indication for use of the Polaris RF Ablation System is the same, or a subset of, the indications for use of the predicate devices. The subject and predicate devices share the same fundamental scientific technology, including principles of operation and mechanism of action (Tables 7.2 and 7.3). Differences in design and technological characteristics between the proposed and predicate devices do not raise any new types of safety and effectiveness questions. Verification and validation test results provide reasonable assurance of the substantial equivalence of the Polaris RF Ablation System compared to the predicate system.

6

Table 7.2: Comparison of Subject and Predicate Devices – Overall System

| | PREDICATE DEVICES | | | | | SUBJECT
DEVICE | Identical /
Substantially
Equivalent
(SE) |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Device Name | Baylis Pain
Management
Generator-TD | TransDiscal (TD)
System | BMC RF
Cannula | BMC PM Probe
and Connector
Cable | Baylis PM
Cooled Probe | Polaris RF
Ablation System | N/A |
| 510(k) # | K072478 | K062937 | K972846 | K002389 | K053082 | K181864 | N/A |
| Manufacturer | Halyard Health* | | | | | Baylis Medical
Company Inc. | N/A |
| Class | II | II | II | II | II | II | YES/YES |
| Product Code,
Regulation | GEI, 882.4400
GXD, 882.4400 | GXI, 882.4725 | GXI, 882.4725 | GXI, 882.4725 | GXI, 882.4725 | GXD, 882.4400
GXI, 882.4725 | YES/YES |
| Indications for
Use | Baylis Pain
Management
Generator - TD;
Model PMG-115-
TD (For
Domestic Use)
and Model PMG-
230-TD (For
International
Use) is indicated
for use to create
lesions during
neurological
lesion
procedures, and
for the
coagulation and
decompression
of disc material
to treat
symptomatic
patients with | The Baylis TD
system, used in
combination with
the Baylis PM
Generator, is
indicated for the
creation of RF
heat lesions in
nervous tissue
including that
which is situated
in the
intervertebral
disc material. | The BMC RF
Cannula is
intended for
use in RF heat
lesion
procedures for
relief of pain | To create RF
lesions in
nervous tissue. | The Baylis PM
Cooled Probe
will be used in
conjunction
with a RF
Generator to
create RF
lesions in
nervous tissue. | The Polaris RF
Ablation System
is intended for
the creation of
RF lesions in
nervous tissue. | NO/YES |
| | PREDICATE DEVICES | | | | SUBJECT
DEVICE | Identical /
Substantially
Equivalent
(SE) | |
| | contained
herniated discs.
The Baylis PMG-
TD is to be used
with separately
approved probes
such as Baylis
TransDiscal
Probe, Oratec
Spinecath and
Baylis Pain
Management
Probes. | | | | | | |
| Relevant
System
Components | RF Generator | Peristaltic Pump,
Connector Cable
Tube Kit
Introducer
Y-Connecting
Cable | RF Cannula | Standard RF
Probes (non-
cooled) | Cooled RF
Probe | RF Generator
and Desk Stand
Peristaltic Pump
and Desk Stand,
Pump Connector
Cable
Connector Hub
RF Cannula
Standard RF
Probes (non-
cooled)
Cooled RF Probe
kit (Cooled RF
Probe, Tube Kit,
Introducer) | NO/YES |

Traditional 510(k)

7

8

| | PREDICATE DEVICES | SUBJECT
DEVICE | Identical /
Substantially
Equivalent
(SE) |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| User | Physicians familiar with RF lesion techniques | Physicians
familiar with RF
lesion
techniques | YES/YES |
| Anatomical
Site of Use | Nervous tissue | Nervous tissue | YES/YES |
| Access Method | Percutaneous | Percutaneous | YES/YES |
| Energy
Type | Radiofrequency | Radiofrequency | YES/YES |
| Principle of
Operation | Operator controlled; RF delivered from RF generator to compatible RF probes | Operator
controlled; RF
delivered from
RF generator to
compatible RF
probes | YES/YES |
| Mechanism of
Action | Cellular necrosis through thermal coagulation | Cellular necrosis
through thermal
coagulation | YES/YES |
| System
Feedback
Mechanism | Temperature controlled | Temperature
controlled | YES/YES |
| Ability to Make
Multiple
Lesions | Yes | Yes | YES/YES |
| | PREDICATE DEVICE | SUBJECT DEVICE | Identical/
SE |
| | RF GENERATOR | | |
| Device Name | Baylis Pain Management
Generator-TD | Polaris Radiofrequency
Generator | N/A |
| 510(k) # | K072478 | K181864 | N/A |
| Available Generator Modes | Standard RF Mode
• Stimulation
(Sensory/Motor)
• Pulsed RF (Single
RF, Multi-RF,
Bipolar RF)
• RF lesion (Single
RF, Multi-RF,
Bipolar RF) | Standard RF Mode
• Stimulation
(Sensory/Motor)
• Pulsed RF (Single
RF, Multi-RF,
Bipolar RF)
• RF lesion (Single
RF, Multi-RF,
Bipolar RF) | YES/YES |
| | Cooled RF Mode
• Stimulation
(Sensory/Motor)
• RF lesion (Single
RF, Multi-RF,
Bipolar RF) | Cooled RF Mode
• Stimulation
(Sensory/Motor)
• RF lesion (Single
RF, Multi-RF,
Bipolar RF) | |
| Maximum Power Output | 50 Watts | 50 Watts | YES/YES |
| Operating Frequency
Waveform | 461 kHz Sinusoidal | 465 kHz Sinusoidal | NO/YES |
| Temperature Monitoring
Thermocouple (RF modes) | Yes | Yes | YES/YES |
| Display Parameters | Real time temperature,
impedance, power and
voltage (pulsed RF
mode) | Real time temperature,
impedance, power and
voltage (pulsed RF
mode) | YES/YES |
| Safety Features | • Automatic shut-off for:
o Out-of-range
impedance and
temperature
o Over power,
voltage, current | • Automatic shut-off for:
o Out-of-range
impedance and
temperature
o Over power,
voltage, current | YES/YES |
| | • Output of errors,
including:
o Fault indicator,
fault code display
and description
o Audible alarm | • Output of errors,
including :
o Alert icon, fault
code display and
description
o Audible alarm | |
| Touchscreen | No | Yes | NO/YES |
| Used with Desk Stand | No | Yes | NO/YES |
| Used with Optional
Footswitch | Yes | Yes | YES/YES |
| Environment | Supplied non-sterile;
Non-sterilisable | Supplied non-sterile;
Non-sterilisable | YES/YES |

9

Table 7.3: Comparison of Subject and Predicate Devices – Relevant Components

10

PREDICATE DEVICESUBJECT DEVICEIdentical/SE
COOLED RF PROBE
Device NamePM Cooled ProbePolaris Cooled RF
Ablation ProbeN/A
510(k) #K053082K181864N/A
Probe ConfigurationMonopolarMonopolarYES/YES
Tissue Temperature
MeasurementIncludes Thermocouple forYesYesYES/YES
Location of ThermocoupleExposed at probe distal
tipEmbedded within probe
distal tipNO/YES
Patient-
ContactShaft,
ThermocoupleStainless steelStainless steelYES/YES
MaterialsInsulationPolyimidePolyimide
Key Non-HandleAcetalABSNO/YES
patientCableSiliconeSilicone
Contact
MaterialsTubingPolyvinyl chloridePolyvinyl chloride
Luer lockPolycarbonatePolycarbonate
Key
DimensionsShaft
Diameter18 Gauge18 GaugeNO/YES
Active Tip
Lengths2, 4, 5.5, 6 mm4, 5.5 mm
Compatible
Introducer
Lengths50, 75, 100, 150 mm50, 100, 150 mm
Cable48" (length);
3.80 mm (OD)48" (length);
3.80 mm (OD)
Tubing48" (length);
3.18 mm (OD)48" (length);
3.18 mm (OD)
EnvironmentProvided sterile;
Single useProvided sterile;
Single useYES/YES
Sterilization Method:
Sterility Assurance Level
(SAL)Ethylene oxide;
10-6Ethylene oxide;
10-6YES/YES
TUBE KIT
Device NameTransDiscal Tube KitPolaris Tube KitN/A
510(k) #K062937K181864N/A
Key non-
patient
Contact
MaterialsTubing
Burette,
Female Luer
Cap, Luer
LocksTygon
PolycarbonateTygon
PolycarbonateNO/YES
CrimpN/AStainless steel
KeyTubing Length140"140"YES/YES
DimensionsBurette
Capacity70 mL70 mL
EnvironmentProvided sterile;
Single useProvided sterile;
Single useYES/YES
Sterilization Method;
(SAL)Sterility Assurance LevelEthylene oxide;
10-6Ethylene oxide;
10-6YES/YES
PREDICATE DEVICESUBJECT DEVICEIdentical/
SE
INTRODUCER
Device NameTransDiscal IntroducerPolaris IntroducerN/A
510(k) #K062937K181864N/A
Comprises Cannula and StyletYesYesYES/YES
Depth MarkersYesYesYES/YES
Patient-
contact
MaterialsShaft
Introducer, StyletStainless steelStainless steelNO/YES
InsulationPolyimidePolyimide
HubUnknownCopolyester
Key
DimensionsShaft Outer
Diameter17 Gauge17 GaugeNO/YES
Usable
Length(s)50, 75, 100, 150 mm50, 100, 150mm
EnvironmentProvided sterile;
Single useProvided sterile;
Single useYES/YES
Sterilization Method;
Sterility Assurance Level
(SAL)Ethylene oxide;
10-6Ethylene oxide;
10-6YES/YES
SINGLE-USE RF PROBE
Device NamePain Management ProbePolaris Single-Use RF
ProbeN/A
510(k) #K002389K181864N/A
ConfigurationMonopolarMonopolarYES/YES
Used with Compatible
CannulaYesYesYES/YES
Thermocouple for Tissue
Temperature MeasurementYesYesYES/YES
Patient-
contact
MaterialsShaft, StyletStainless steelStainless steelYES/YES
Key
DimensionsShaft
Diameter24, 28, 31 Ga28 GaNO/YES
Compatible
Cannula
Length(s)50, 54, 60, 100, 145,
200 mm50, 100, and 150 mm
EnvironmentProvided non-sterile;
Re-sterilisable (steam)Provided sterile;
Single useNO/YES
Sterilization Method;
Sterility Assurance Level
(SAL)N/AEthylene oxide;
10-6NO/YES
REUSABLE RF PROBE
Device NamePain Management ProbePolaris Reusable RF
ProbeN/A
510(k) #K002389K181864N/A
ConfigurationMonopolarMonopolarYES/YES
Used with CannulaYesYesYES/YES
Thermocouple for Tissue
Temperature MeasurementYesYesYES/YES
PREDICATE DEVICESUBJECT DEVICEIdentical/
SE
Patient-
contact
MaterialsShaft, StyletStainless steel or nitinolStainless steel or nitinolYES/YES
Key
DimensionsShaft
Diameter24, 28, 31 Ga28 GaNO/YES
Compatible
Cannula
Length(s)50, 54, 60, 100, 145,
200 mm50, 100, and 150 mmNO/YES
EnvironmentProvided non-sterile;
Re-sterilisable (steam)Provided non-sterile;
Re-sterilisable (steam)YES/YES
RF CANNULA
Device NameBMC RF CannulaPolaris RF CannulaN/A
510(k) #K972846K181864N/A
Patient-
contact
MaterialsShaft, StyletStainless steelStainless steelYES/YES
Shaft
InsulationPolyethylene
terephthalatePolyethylene
terephthalateYES/YES
HubCopolyesterCopolyesterYES/YES
Key
DimensionsShaft
Diameter16, 18, 20, 21, 22 Ga16, 18, 20, 22 GaNO/YES
Usable
Length50, 54, 60, 100, 145,
200 mm50, 100, 150 mmNO/YES
Bare Active
Tip2, 4, 5, 10, 15 mm5, 10 mmNO/YES
Tip ConfigurationsStraight sharp
Straight blunt
Curved sharp
Curved bluntStraight sharp
Curved sharp
Curved bluntNO/YES
EnvironmentProvided sterile;
Single useProvided sterile;
Single useYES/YES
Sterilization Method;
Sterility Assurance Level
(SAL)Ethylene oxide;
10-6Ethylene oxide;
10-6YES/YES
PUMP UNIT
Device NameTransDiscal Pump UnitPolaris Pump UnitN/A
510(k) #K062937K181864N/A
Number of Pump HeadsTwoTwoYES/YES
Pump Controlled by
GeneratorYesYesYES/YES
Used with Tube Kit(s)YesYesYES/YES
Safety SwitchYesYesYES/YES
Used with Desk StandNoYesNO/YES
EnvironmentSupplied non-sterile;
Non-sterilisableSupplied non-sterile;
Non-sterilisableYES/YES
CONNECTOR HUB
Device NameTransDiscal Y-
Connecting CablePolaris Connector HubN/A
510(k) #K062937K181864N/A
Number of Probe
Connections PossibleTwoFourNO/YES
PREDICATE DEVICESUBJECT DEVICEIdentical/
SE
ConnectorProbe5-pin female7-pin femaleNO/YES
Generator19-pin male33-pin HG4 maleNO/YES
Key Cable MaterialsSilicone, Thermoplastic elastomerThermoplastic elastomer, injection molded plasticNO/YES
Cable Length9 feet8 feetNO/YES
EnvironmentSupplied non-sterile; Non-sterilisableSupplied non-sterile; Non-sterilisableYES/YES

11

12

13

7.7 Performance Testing

Performance testing was completed to demonstrate substantial equivalence of the Polaris RF Ablation System and relevant system components to the identified predicate devices in Table 7.1. The system components were subjected to the verification and validation testing listed in Table 7.4.

Table 7.4: Performance Testing of Subject Device

TestsTest Method SummaryResults
MechanicalPolaris Re-Usable RF Probe
MechanicalThe following tests were conducted
to verify whether the subject device
was capable of withstanding the
relevant mechanical stresses without
failure:
• Temperature Test
• Pull Test
• Cleaning and Sterilization
Reuse Test
• Mechanical Reuse Test
• Mechanical Test (Flexion)
• Continuity Test
• Functional TestAll samples passed the
acceptance criteria. Results
confirmed compliance of the
device with IEC 60601-2-
2:2009 and support
substantial equivalence with
the predicate device.
ElectricalElectrical verification was conducted
through comparison with the Polaris
Standard Single-Use Probe to verify
whether the device was capable of
withstanding the relevant electrical
stresses without failure.The device was verified to
meet the requirements of IEC
60601-1:2005+A1:2012 and
IEC 60601-2-2:2009. Results
support substantial
equivalence with the predicate
device.

14

TestsTest Method SummaryResults
BiocompatibilityBiocompatibility of the subject device
was demonstrated through
comparisons to similar legally
marketed devices as per current ISO
10993-1 requirements and the FDA
guidance, "Use of International
Standard ISO 10993-1, "Biological
evaluation of medical devices - Part
1: Evaluation and testing within a
risk management process".Results confirmed compliance
of the device with the current
ISO 10993-1 requirements
and support substantial
equivalence with the predicate
device.
PackagingShip testing was performed to ensure
the integrity of the subject device
packaging through the rigors of
shipping and handling.All packaging met relevant
requirements. Results support
substantial equivalence with
the predicate device.
Cleaning and SterilizationThe recommended reprocessing
methods provided in the Instructions
for use were validated to ensure the
recommended parameters and
method could achieve a Sterility
Assurance Level of 10-6.The device was verified to
meet the requirements of
AAMI TIR12, AAMI TIR30 and
ANSI/AAMI ST81
requirements and the FDA
guidance, "Reprocessing
Medical Devices in Health Care
Settings: Validation Methods
and Labeling". Results support
substantial equivalence with
the predicate device.
MechanicalPolaris Single-Use RF Probe
The following tests were conducted
to verify whether the subject device
was capable of withstanding the
relevant mechanical stresses without
failure:
● Temperature Test
● Pull Test
● Durability Test
● Mechanical Test (Flexion)
● Continuity Test
● Functional TestAll samples passed the
acceptance criteria. Results
confirmed compliance of the
device with IEC 60601-2-
2:2009 and support
substantial equivalence with
the predicate device.
TestsTest Method SummaryResults
ElectricalThe following tests were conducted
to verify whether the subject device
was capable of withstanding the
relevant electrical stresses without
failure:
High Frequency Leakage
Current Test - Cable High Frequency Dielectric
Strength Test – Cable High Frequency Dielectric
Strength Test - Handle Mains Frequency Dielectric
Strength Test - Cable Mains Frequency Dielectric
Strength Test - Handle Mains Frequency Dielectric
Strength Test – ConnectorAll samples passed the
acceptance criteria. Results
confirmed compliance of the
device with IEC 60601-
1:2005+A1:2012 and IEC
60601-2-2:2009 and support
substantial equivalence with
the predicate device.
BiocompatibilityBiocompatibility of the subject device
was demonstrated through
comparisons to similar legally
marketed devices as per current ISO
10993-1 requirements and the FDA
guidance, "Use of International
Standard ISO 10993-1, "Biological
evaluation of medical devices - Part
1: Evaluation and testing within a
risk management process".Results confirmed compliance
of the device with the current
ISO 10993-1 requirements
and support substantial
equivalence with the predicate
device.
PackagingShip testing was performed to ensure
the integrity of the subject device
packaging through the rigors of
shipping and handling. The seal
strength and sterile barrier integrity
was validated per ANSI/AAMI/ISO
11607-1 and 11607-2 over the shelf
life of the device.The device packaging met
relevant requirements. Results
confirmed compliance of
device packaging with
ANSI/AAMI/ISO 11607-1 and
11607-2 and support
substantial equivalence with
the predicate device.
TestsTest Method SummaryResults
Polaris RF Cannula
MechanicalThe following tests were conducted
to verify whether the subject device
was capable of withstanding the
relevant mechanical stresses without
failure:
Positive Pressure Liquid
Leakage Test Sub-atmospheric Pressure Air
Leakage Test Resistance to Separation Test Stress Cracking Test Hub Strength TestAll samples passed the
acceptance criteria. Results
confirmed the device met with
the mechanical requirements
of ISO 80369-7:2016, ISO
594-1:1986 and ISO
7864:2016 and support
substantial equivalence with
the predicate device.
ElectricalThe following tests were conducted
to verify whether the subject device
was capable of withstanding the
relevant electrical stresses without
failure:
High Frequency Leakage
Current Test High Frequency Dielectric
Strength Test Mains Frequency Dielectric
Strength TestAll samples passed the
acceptance criteria. Results
confirmed compliance of the
device with IEC 60601-
1:2005+A1:2012 and IEC
60601-2-2:2009 and support
substantial equivalence with
the predicate device.
BiocompatibilityBiocompatibility of the subject device
was demonstrated through biological
testing as per current ISO 10993-1
requirements and the FDA guidance,
"Use of International Standard ISO
10993-1, "Biological evaluation of
medical devices - Part 1: Evaluation
and testing within a risk
management process".Results confirmed compliance
of the device with the current
ISO 10993-1 requirements
and support substantial
equivalence with the predicate
device.
PackagingShip testing was performed to ensure
the integrity of the subject device
packaging through the rigors of
shipping and handling. The seal
strength and sterile barrier integrity
was validated per ANSI/AAMI/ISO
11607-1 and 11607-2 over the shelf
life of the device.The device packaging met
relevant requirements. Results
confirmed compliance of
device packaging with
ANSI/AAMI/ISO 11607-1 and
11607-2 and support
substantial equivalence with
the predicate device.
TestsTest Method SummaryResults
Polaris Cooled RF Probe Kit
MechanicalThe following tests were conducted to verify whether the Polaris Cooled RF Probe was capable of withstanding the relevant mechanical stresses without failure:
Leak/Pressure Test Pull Test Temperature Test Tip Compression Test Mechanical Test (Flexion) Continuity Test Functional TestAll samples passed the acceptance criteria. Results confirmed compliance of the device with IEC 60601-2-2:2009 and support substantial equivalence with the predicate device.
The following tests were conducted to verify whether the Polaris Introducer was capable of withstanding the relevant mechanical stresses without failure:
Resistance to Separation from Unscrewing Test Resistance to Separation TestAll samples passed the acceptance criteria. Results support substantial equivalence with the predicate device.
ElectricalThe following tests were conducted to verify whether the Polaris Cooled RF Probe was capable of withstanding the relevant electrical stresses without failure:
High Frequency Leakage Current Test - Cable High Frequency Leakage Current Test - Shaft High Frequency Dielectric Strength Test – Cable High Frequency Dielectric Strength Test - Handle/Shaft Mains Frequency Dielectric Strength Test – Cable Mains Frequency Dielectric Strength Test - Handle/Shaft Mains Frequency Dielectric Strength Test ConnectorAll samples passed the acceptance criteria. Results confirmed compliance of the device with IEC 60601-1:2005+A1:2012 and IEC 60601-2-2:2009 and support substantial equivalence with the predicate device.
TestsTest Method SummaryResults
BiocompatibilityBiocompatibility of the Polaris Cooled
RF Probe and Polaris Introducer was
demonstrated through biological
testing and comparisons to similar
legally marketed devices as per
current ISO 10993-1 requirements
and the FDA guidance, "Use of
International Standard ISO 10993-1,
"Biological evaluation of medical
devices - Part 1: Evaluation and
testing within a risk management
process".Results confirmed compliance
of the devices with the current
ISO 10993-1 requirements
and support substantial
equivalence with the predicate
devices.
PackagingShip testing was performed to ensure
the integrity of the subject device
packaging through the rigors of
shipping and handling. The seal
strength and sterile barrier integrity
was validated per ANSI/AAMI/ISO
11607-1 and 11607-2 over the shelf
life of the device.The subject device packaging
met relevant requirements.
Results confirmed compliance
of device packaging with
ANSI/AAMI/ISO 11607-1 and
11607-2 and support
substantial equivalence with
the predicate device.
VerificationPolaris Radiofrequency Generator
Testing was conducted to verify the
function of the subject device,
including electrical testing of the
following hardware components:
• RF board
• User Interface Board
• Multiplexer boardThe subject device met all
relevant requirements. Results
support substantial
equivalence with the predicate
device.
PackagingShip testing was performed to ensure
the integrity of the subject device
packaging through the rigors of
shipping and handling.All packaging met relevant
requirements. Results support
substantial equivalence with
the predicate device.
TestsTest Method SummaryResults
Polaris Pump Unit
VerificationTesting was conducted to verify specified design requirements of the subject device, including hardware. Performance was assessed in test cases covering the following aspects:
● Inspection
● Configuration and Installation
● Operation
● Environment and Service LifeThe subject device met all relevant requirements. Results support substantial equivalence with the predicate device.
PackagingShip testing was performed to ensure the integrity of the subject device packaging through the rigors of shipping and handling.All packaging met relevant requirements. Results support substantial equivalence with the predicate device.
Polaris Connector Hub
VerificationTesting was conducted to verify specified design requirements of the subject device, including hardware. Performance was assessed through various test cases, including:
● Cable Pull Test
● Cable Impulse Test
● Device IdentificationThe subject device met all relevant requirements. Results confirmed compliance of the device with Baylis self-enforced requirements and support substantial equivalence with the predicate device.
PackagingShip testing was performed to ensure the integrity of the device packaging through the rigors of shipping and handling.All packaging met relevant requirements. Results confirmed compliance of device packaging with Baylis self-enforced requirements and support substantial equivalence with the predicate device.
Polaris Footswitch
PackagingShip testing was performed to ensure the integrity of the subject device packaging through the rigors of shipping and handling.All packaging met relevant requirements. Results support substantial equivalence with the predicate device.
TestsTest Method SummaryResults
Polaris RF Ablation System
Electrical Product
SafetyThe Polaris RF Generator, together
with relevant accessories and
detachable parts, was tested to
demonstrate compliance with the
applicable requirements of IEC
60601-1:2005+A1:2012 (including
national deviations as per ANSI AAMI
ES 60601-1:2005(R) 2012 +
A1:2012) and IEC 60601-2-2: 2009.The subject device met all
relevant requirements. Results
confirmed compliance of the
device with
IEC 60601-1:2005+A1:2012
and IEC 60601-2-2:2009 and
support substantial
equivalence with the predicate
device.
Electromagnetic
CompatibilityThe Polaris RF Generator, together
with applicable system components,
was tested to demonstrate
compliance with the applicable
requirements of IEC 60601-1-
2:2014.The subject device met all
relevant requirements. Results
confirmed compliance of the
device with
IEC 60601-1-2:2014 and
support substantial
equivalence with the predicate
device.
Benchtop Lesion
ValidationComparative lesion validation testing
was performed using a soft tissue
model to demonstrate the
substantially equivalent ablation
performance of the subject and
predicate systems. Testing was
performed for both single and multi-
probe scenarios.The subject device met all
relevant requirements. Results
support substantial
equivalence with the predicate
device.
UsabilityTesting was performed to verify and
validate the usability requirements of
the subject Polaris RF Ablation
System. Elements captured included
normal use cases, a foreseeable
worst-case use scenario, and
testable requirements for primary
operating functions.The device met all relevant
requirements. Results support
substantial equivalence with
the predicate device.
TestsTest Method SummaryResults
SoftwareSoftware verification and validation
was completed for the subject
device. FDA's current "Guidance for
the Content of Premarket
Submissions for Software Contained
in Medical Devices" was used to
determine the Level of Concern for
the software in the subject Polaris RF
Ablation System.The device met all relevant
requirements. Results support
substantial equivalence with
the predicate device.

15

16

17

18

19

20

21

All test requirements were met as specified by applicable standards and the test protocols. Results demonstrate that the Polaris RF Ablation System functions as intended and is as safe and effective as the predicate device.

7.8 Conclusions

The intended use of the subject Polaris RF Ablation System is the same as the predicate devices. The indications for use of the Polaris RF Ablation System is the same, or a subset of, the indications for use of the predicate devices. The fundamental scientific technology of the Polaris RF Ablation System, including principles of operation and mechanism of action, is the same as the predicate devices. Differences in design and technological characteristics between the subject and predicate devices do not raise any new types of questions of safety and effectiveness. Verification and validation test results support the substantial equivalence of the Polaris RF Ablation System to the predicate system.