K Number
K181749

Validate with FDA (Live)

Date Cleared
2018-08-01

(30 days)

Product Code
Regulation Number
888.3010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Kinamed SuperCable Iso-Elastic Cerclage System is intended to be used in the following: repair of long bone fractures due to trauma or reconstruction; reattachment of the greater trochanter in total hip arthroplasty, surface replacement arthroplasty, or other procedures involving trochanteric osteotomy; sternotomy closure; and sublaminar and intrafacet wiring of the spinal column.

Device Description

The Kinamed SuperCable® Iso-Elastic™ Cerclage System consists of an implantable cable and locking clasp (the locking clasp consists of a clip and a wedge). The cable is comprised of a monofilament core of nylon with a jacket of ultra-high molecular-weight-polyethylene (UHMWPE) fibers braided around the nylon core. Manual instrumentation is used for applying the cable. The subject of the present Special 510(k) submission is a line addition of a cable implant whose locking clasp is manufactured using a metal injection molding (MIM) process.

AI/ML Overview

The provided document is a 510(k) summary for the Kinamed SuperCable® Iso-Elastic™ Cerclage System. It describes a line extension to an existing device, specifically a change in the manufacturing process for the locking clasp. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive clinical efficacy studies typically found for novel devices.

Therefore, the document does not contain the detailed information requested for acceptance criteria, device performance studies, sample sizes for test/training sets, expert qualifications, adjudication methods, or MRMC studies that would be present for an AI/ML-based medical device or a device requiring new clinical efficacy data.

Based on the information provided in the document:

1. A table of acceptance criteria and the reported device performance

The document mentions "Performance testing of the SuperCable Iso-Elastic Cerclage System includes ISO 10993 testing, Bacterial endotoxin levels (LAL), Static construct tensile testing, and Fatigue construct tensile testing." However, the specific acceptance criteria and the reported results for each of these tests are not provided in this 510(k) summary. The document only states that the device "meets the specifications of ASTM F2885 (Type 1, Densified)" for the locking clasp material. This implies the acceptance criteria for the material are the ASTM F2885 specifications.

2. Sample size used for the test set and the data provenance

Not applicable. This submission is for a modification to an existing device (change in manufacturing process for a component), not an AI/ML device that uses test sets or clinical data in the typical sense for performance evaluation. The performance testing mentioned (tensile, fatigue, etc.) would be conducted on a sample of manufactured devices, but the sample sizes are not reported.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. Ground truth establishment by experts is not relevant for this type of device modification.

4. Adjudication method for the test set

Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI-based algorithm.

7. The type of ground truth used

Not applicable in the context of clinical "ground truth." The "ground truth" for this device would be defined by engineering specifications and material standards (e.g., ASTM F2885 for the material, and internal specifications for tensile strength and fatigue life).

8. The sample size for the training set

Not applicable. This is not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable.

Summary of what is presented in the document regarding acceptance criteria and study:

The submission focuses on demonstrating that a manufacturing change (Metal Injection Molding for the locking clasp) does not alter the safety or effectiveness of the device compared to the predicate. The basis of substantial equivalence is outlined by comparing properties such as:

  • Same polymer cable
  • Same intended use
  • Same indications for use
  • Same raw material alloy (Ti-6Al-4V) for locking clasp (though the manufacturing process changed)
  • Same design
  • Similar size and dimensions
  • Same type of mating components
  • Same shelf life
  • Same biocompatibility
  • Same sterilization and packaging methods

The performance testing mentioned (ISO 10993, LAL, Static construct tensile testing, Fatigue construct tensile testing) would serve as the "studies" to prove the device meets acceptance criteria, which would be the established specifications derived from the predicate device's performance and relevant standards. However, the specific data from these tests and the exact acceptance criteria are not included in this summary. The key statement regarding acceptance criteria is that the material "meets the specifications of ASTM F2885 (Type 1, Densified)."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is blue and white.

August 1. 2018

Kinamed, Incorporated Heather Neely Senior Director of Quality Assurance and Regulatory Compliance 820 Flynn Road Camarillo. California 93012

Re: K181749

Trade/Device Name: SuperCable® Iso-Elastic™ Cerclage System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ Dated: June 25, 2018 Received: July 2, 2018

Dear Heather Neely:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows a digital signature. The signature is by Vesa Vuniqi -S. The date of the signature is 2018.08.01. The time of the signature is 17:08:45 -04'00'.

For: Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181749

Device Name

SuperCable® Iso-Elastic™ Cerclage System

Indications for Use (Describe)

The Kinamed SuperCable Iso-Elastic Cerclage System is intended to be used in the following: repair of long bone fractures due to trauma or reconstruction; reattachment of the greater in total hip arthroplasty, surface replacement arthroplasty, or other procedures involving trochanteric osteotomy; sternotomy closure; and sublaminar and intrafacet wiring of the spinal column.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Kinamed Incorporated. The logo features a green square with a stylized white "K" inside, with a gradient effect on the lower part of the "K". To the right of the square, the word "KINAMED" is written in bold, black letters, with the word "INCORPORATED" written in smaller, lighter letters below it.

K181749

Page 1 of 3

Special 510(k) Summary

Manufacturer:KINAMED® Incorporated820 Flynn RoadCamarillo, CA 93012-8701USA
Phone (805) 384-2748FAX (805) 384-2792
Contact Person:Heather Neely, RAC

Senior Director of Quality Assurance and Regulatory Compliance Kinamed®, Inc. Phone (805) 886-1069 and (805) 384-2748 x210 FAX (805) 384-2792 Email: hneely@kinamed.com

Date Prepared: June 25, 2018

DEVICE INFORMATION

Trade/Proprietary Name: SuperCable® Iso-Elastic™ Cerclage System

Common Name:Bone fixation cerclage
Classification Name:Cerclage, Fixation

21 CFR 888.3010 Class II Device Product Code: JDQ

K030256 (SuperCable® Iso-Elastic™ Cerclage Predicate Device: System)

Product Description:

The Kinamed SuperCable® Iso-Elastic™ Cerclage System consists of an implantable cable and locking clasp (the locking clasp consists of a clip and a wedge). The cable is comprised of a monofilament core of nylon with a jacket of ultra-high molecular-weight-polyethylene (UHMWPE) fibers braided around the nylon core. Manual instrumentation is used for applying the cable.

{4}------------------------------------------------

K181749

Page 2 of 3

The SuperCable Iso-Elastic Cerclage System that is the subject of this submission is a line extension to the predicate SuperCable Iso-Elastic Cerclage System (K030256) that was cleared on October 21, 2003.

Additional sizes of the SuperCable implant were previously added to the System via a Special 510(k) line extension (K102834) that was cleared on January 12, 2011.

The subject of the present Special 510(k) submission is a line addition of a cable implant whose locking clasp is manufactured using a metal injection molding (MIM) process. The cable portion of the implant is the exact same as in the predicate system. This line addition is intended to improve manufacturing efficiency and is not motivated by any clinical or performance factors associated with the SuperCable Iso-Elastic Cerclage System. An overview of the changes being proposed is described in Figure 5-1.

Figure 5-1
Description of Proposed Changes for the present submission
510(k) DesignationLocking ClaspMaterial of ConstructionLocking ClaspManufacturing Process
K030256(predicate)ASTM F1361Titanium(implant-grade)Machined from Wrought
K181749(this submission)ASTM F28852Titanium(implant-grade)Metal Injection Molded(MIM)

There are no changes to the polymer cable itself or to the product's intended use, indications for use, functionality, shelf life, biocompatibility, packaging, and sterilization method.

The SuperCable Iso-Elastic System locking clasp that is the subject of this submission meets the specifications of ASTM F2885 (Type 1, Densified).

Indications for Use (for this submission):

The Kinamed SuperCable Iso-Elastic Cerclage System is intended to be used in the following: repair of long bone fractures due to trauma or reconstruction; reattachment of the greater trochanter in total hip arthroplasty, surface

1 ASTM F136. Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401)

2 ASTM F2885, Standard Specification for Metal Injection Molded Titanium-6Aluminum-4Vanadium Components for Surgical Implant Applications

{5}------------------------------------------------

Page 3 of 3

replacement arthroplasty, or other procedures involving trochanteric osteotomy; sternotomy closure; and sublaminar and intrafacet wiring of the spinal column.

With the exception of the addition of Kinamed's "SuperCable" brand name, there are no changes to the indications for use or intended use from the previously cleared SuperCable Iso-Elastic Cerclage System predicate (K030256). Form FDA 3881, Indications for Use, can be found in Section 4 of this Special 510(k) submission. The addition of the "SuperCable" brand name to the Indications for Use statement is not considered a change because the product line has been branded "SuperCable" since it was launched in the USA in 2004.

Performance Testing

Performance testing of the SuperCable Iso-Elastic Cerclage System includes ISO 10993 testing, Bacterial endotoxin levels (LAL), Static construct tensile testing, and Fatigue construct tensile testing.

Basis of Substantial Equivalence

The SuperCable Iso-Elastic Cerclage System that is the subject of this submission shares the following similarities with the predicate SuperCable Iso-Elastic Cerclage System cleared under K030256:

  • same polymer cable
  • same intended use .
  • same indications for use ●
  • same raw material alloy (Ti-6Al-4V) for locking clasp ●
  • . same design
  • similar size and dimensions
  • . same type of mating components
  • same shelf life .
  • same biocompatibility .
  • same sterilization and packaging methods .

Conclusion:

The data and information provided in this submission support the conclusion that the SuperCable Iso-Elastic Cerclage System that is the subject of this 510(k) submission is substantially equivalent to its predicate device (K030256), and the proposed change does not significantly affect the safety or effectiveness of the device. This line addition is being proposed to improve manufacturing efficiency and is not motivated by any clinical or performance factors.

§ 888.3010 Bone fixation cerclage.

(a)
Identification. A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.(b)
Classification. Class II.