K Number
K181748
Device Name
Magnesium
Date Cleared
2018-09-12

(72 days)

Product Code
Regulation Number
862.1495
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine on the ARCHITECT c8000 System.

Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).

Device Description

The Magnesium reagent kit contains Reagent 1 and Reagent 2. Magnesium present in the sample is a cofactor in an enzymatic reaction with isocitrate dehydrogenase. The rate of increase in absorbance at 340 nm, due to the formation of NADPH, is directly proportional to the magnesium concentration.

AI/ML Overview

The provided document is a 510(k) summary for a medical device called "Magnesium" from Abbott Laboratories. It details the performance testing conducted to demonstrate its substantial equivalence to a predicate device. This document describes an in vitro diagnostic (IVD) device for measuring magnesium levels, not an AI/ML-based device. Therefore, many of the requested elements related to AI (e.g., ground truth establishment by experts, adjudication, MRMC studies, training set details) are not applicable to this submission.

However, I can extract information relevant to the acceptance criteria and performance of this IVD device.

Here's the breakdown of the acceptance criteria and study that proves the device meets them, based on the provided text:

Device: Abbott Laboratories Magnesium Assay (List No. 3P68)
Intended Use: Quantitation of magnesium in human serum, plasma, or urine on the ARCHITECT c8000 System for diagnosis and treatment of hypomagnesemia and hypermagnesemia.
Predicate Device: Roche Magnesium Gen.2 (K983416)

1. Acceptance Criteria and Reported Device Performance

The document describes various performance studies and their results. The implicit acceptance criteria are that the device's performance characteristics demonstrate substantial equivalence to the predicate device and are within acceptable ranges for clinical utility.

Study TypeAcceptance Criteria (Implicit)Reported Device Performance
Limit of Blank (LoB)LoB should be low, demonstrating minimal signal in the absence of analyte. (No explicit numerical criterion stated, but values are reported).Urine application: LoB of 0.04 mg/dL
Limit of Detection (LoD)LoD should be low enough to detect clinically relevant low levels. (No explicit numerical criterion stated, but values are reported).Urine application: LoD of 0.09 mg/dL
Limit of Quantitation (LoQ)LoQ should be low enough for accurate quantitation at clinically relevant low levels. (No explicit numerical criterion stated, but values are reported).Urine application: LoQ of 0.75 mg/dL
Within-Laboratory Precision (Imprecision)%CV should be clinically acceptable, demonstrating consistency of results over time. (No explicit numerical criterion stated, but values are reported as evidence of acceptable precision).Urine (within-laboratory imprecision):- Bio-Rad Level 1: 1.3 %CV- Bio-Rad Level 2: 1.3 %CV- LoQ Urine Pool -Low Mg: 2.4 %CV- Human Urine Pool - Normal Mg: 1.8 %CV- Human Urine Pool Abnormal Mg: 1.8 %CV
InterferenceAssay results should be impacted by no more than ±10% for specific interferent levels.For magnesium samples targeted to 5 mg/dL: no more than ±10% interference for listed substances (Albumin ≤ 64.0 mg/dL, Ascorbic Acid ≤ 200 mg/dL, Bilirubin (Conjugated) ≤ 59.9 mg/dL, Calcium ≤ 26.0 mg/dL, Glucose ≤ 1220 mg/dL, Hemoglobin ≤ 1200 mg/dL, Phosphorous ≤ 307 mg/dL, Boric Acid ≤ 1000 mg/dL, 6N Hydrochloric Acid ≤ 3.0 mL/dL, Copper ≤ 21.6 µg/dL, Zinc ≤ 3504 µg/L, Iron ≤ 0.6 mg/dL).For magnesium samples targeted to 14 or 15 mg/dL: similar results for the same interferents with slightly different calcium and bilirubin levels.Acetic acid, nitric acid, and sodium fluoride did not meet the ±10% criterion and are noted as limitations.
LinearityThe assay should be linear across its analytical measuring interval. (No explicit R-squared or slope criterion, but stated that it was "demonstrated to be linear").Urine application demonstrated linearity across 1.04 to 36.24 mg/dL, spanning the analytical measuring interval of 1.81 to 26.35 mg/dL.
Measuring Interval (Analytical Measuring Range)Defined by LoQ and highest linear point.1.81 to 26.35 mg/dL
Method Comparison (Correlation to Predicate)Demonstrate acceptable correlation (slope and correlation coefficient) to the predicate device across the measuring interval.Urine application showed acceptable correlation to predicate:- pH <2: slope 1.08, r-value 1.00- No acidification: slope 1.04, r-value 1.00- pH ~1: slope 1.05, r-value 1.00
Automated Dilution ProtocolResults impacted by not more than ±10% at high analyte concentrations.Impacted by not more than ±10% for analyte concentration of 75 mg/dL.
Manual DilutionResults impacted by not more than ±10% for specific analyte concentrations after manual dilution.Impacted by not more than ±10% for analyte concentrations 15, 25, and 40 mg/dL with 1:2 manual dilution.

2. Sample Size and Data Provenance

  • Test Set Sample Sizes:

    • LoB, LoD, LoQ: 4 saline samples (zero-analyte), 2 low-analyte level samples at each of 5 target concentrations (0.390, 0.780, 1.285, 1.560, 3.130 mg/dL). Zero-analyte samples tested in 10 replicates. Low-analyte samples tested in 10 replicates.
    • Within-Laboratory Precision: Control materials (Bio-Rad Level 1 & 2, LoQ Urine Pool -Low Mg, Human Urine Pool Normal Mg, Human Urine Pool Abnormal Mg) tested in 2 replicates, 2 times per day for 20 testing days.
    • Interference (Endogenous/Preservatives): Control and test level samples tested in a minimum of 7 replicates.
    • Interference (Cations - Copper/Zinc): Control sample (urine) and test level samples tested in 12 replicates.
    • Linearity: 3 sets of linearity standards, each with 12 levels. Each level tested in a minimum of 4 replicates.
    • Method Comparison: 118 patient urine specimens. 12 of these were normal urine specimens spiked with magnesium.
    • Automated Dilution: 1 sample (75 mg/dL target) evaluated in 10 replicates.
    • Manual Dilution: 3 urine pools (15, 25, 40 mg/dL target). Each pool evaluated after 1:2 manual dilution. Samples tested in a minimum of 12 replicates.
  • Data Provenance: The document does not explicitly state the country of origin for the patient samples or if they were retrospective or prospective. Given the submitter is an American company (Abbott Laboratories, Irving, TX) and the regulatory submission is to the U.S. FDA, it is highly likely the studies were conducted in the U.S. and the samples were generally from a U.S. population. The studies described are performance studies for an IVD, which typically use banked or collected samples, so they would be considered retrospective in terms of sample collection for the specific study; however, the study itself is prospective in its design and execution.

3. Number of Experts and Qualifications for Ground Truth

  • Not Applicable. This is an IVD device measuring analyte concentration using a chemical reaction. The "ground truth" for the test set is established by the analytical reference methods and gravimetric preparations, not by human expert interpretation (like radiologists for imaging).

4. Adjudication Method for Test Set

  • Not Applicable. As this is an IVD device measuring a chemical analyte, there is no expert adjudication process for the test set results. The ground truth for performance studies is based on traceable standards, established reference methods, or gravimetric preparations.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Not Applicable. MRMC studies are relevant for devices that involve human interpretation, particularly in diagnostic imaging or pathology. This is an automated IVD device.

6. Standalone (Algorithm Only) Performance

  • The performance data presented (LoB, LoD, LoQ, Precision, Interference, Linearity, Analytical Measuring Interval, Method Comparison, Dilution studies) are all standalone performance metrics of the Magnesium assay on the ARCHITECT c8000 System. There is no human-in-the-loop component for the measurement itself; the device provides a direct quantitative result.

7. Type of Ground Truth Used

The ground truth for these studies relies on:

  • Reference Standards/Materials: For LoB, LoD, LoQ, Linearity, and Interference studies, prepared samples with known concentrations (e.g., saline, diluted or spiked magnesium standards).
  • Established Methods/Predicate Device: For Method Comparison, the results are compared to a legally marketed predicate device (Roche Magnesium Gen.2), which serves as the "truth" for demonstrating substantial equivalence.
  • Control Materials: Commercially available quality control materials and in-house prepared human urine pools with expected magnesium levels are used for precision studies.

8. Sample Size for the Training Set

  • Not Applicable/Not Explicitly Stated. For an IVD like this, there isn't a "training set" in the sense of machine learning algorithms. The device's operational parameters (reagent formulation, reaction kinetics, detection methods) are developed through R&D and optimization, not typically by training on a distinct data set in the way an AI model is trained. The studies described are validation and verification studies to demonstrate performance, not a training process.

9. How Ground Truth for the Training Set Was Established

  • Not Applicable. As there's no "training set" in the typical AI sense, the establishment of ground truth for such a set is not relevant. The analytical parameters of the device are based on established chemical principles and optimized through research and development.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 12, 2018

Abbott Laboratories Mark Littlefield Assoc. Director, Regulatory Affairs 1921 Hurd Drive Irving, TX 75038

Re: K181748

Trade/Device Name: Magnesium Regulation Number: 21 CFR 862.1495 Regulation Name: Magnesium test system Regulatory Class: Class I, reserved Product Code: JGJ Dated: June 28, 2018 Received: July 2, 2018

Dear Mark Littlefield:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181748

Device Name Magnesium

Indications for Use (Describe)

The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine on the ARCHITECT c8000 System.

Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (Summary of Safety and Effectiveness)

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K181748.

1. Applicant Name

Mark Littlefield, ADD, Associate Director, Regulatory Affairs Abbott Laboratories 1921 Hurd Drive Irving, TX 75038 (972) 518-6062 Fax: (972) 518-7498 Email: mark.littlefield@abbott.com

Date Summary prepared: August 28, 2018

2. Device Name

Trade Name: Magnesium Device Classification: Class I Reserved Classification Name: Magnesium Reagent Governing Regulation: CFR 862.1495 Product Code: JGJ

3. Predicate Device

Roche Magnesium Gen.2 (510(k) number K983416)

{4}------------------------------------------------

4. Description of Device

The Magnesium reagent kit contains:

Component3P68-223P68-32
Number of Tests670* (urine)2540* (urine)
Reagent 1 (R1)5 × 39 mL10 × 71 mL
Reagent 2 (R2)5 × 11 mL10 × 18 mL
  • Calculation is based on the minimum reagent fill volume per kit and may vary depending on the mix of urine samples.
ReagentReactive IngredientsConcentration
Reagent 1Isocitrate dehydrogenase2.2 U/mL
D-Isocitrate potassium salt1.47 mg/mL
Reagent 2NADP8.37 mg/mL

Principles of the Procedure

Magnesium present in the sample is a cofactor in an enzymatic reaction with isocitrate dehydrogenase. The rate of increase in absorbance at 340 nm, due to the formation of NADPH, is directly proportional to the magnesium concentration.

$$\text{D-isocitive acid} + \text{NADP} \xrightarrow{\text{Isocificate dehydrogenase}} 2 \text{-oxoglutarate} + \text{CO}_2 + \text{NADPH}$$

Methodology: Enzymatic

{5}------------------------------------------------

5. Intended Use of the Device

The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine on the ARCHITECT c8000 System.

Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).

6. Comparison of Technological Characteristics

The Magnesium assay is used for the quantitation of magnesium in human serum, plasma, or urine on the ARCHITECT c8000 System.

A comparison of the candidate assay (Magnesium, List No. 3P68) and the predicate assay (Roche Magnesium Gen.2 REF 06407358 190) is presented in the table on page 4.

{6}------------------------------------------------

AssayCharacteristicsMagnesiumLN 3P68Roche Magnesium Gen.2
Analyte MeasuredMagnesiumSame
Intended UseThe Magnesium assay is used forthe quantitation of magnesium inhuman serum, plasma, or urine onthe ARCHITECT c8000 System.Magnesium measurements areused in the diagnosis andtreatment of hypomagnesemia(abnormally low plasma levels ofmagnesium) andhypermagnesemia (abnormallyhigh plasma levels ofmagnesium).In vitro test for the quantitativedetermination of magnesium inhuman serum, plasma and urineon Roche/Hitachi cobas csystems.
Assay PrincipleMagnesium present in the sampleis a cofactor in an enzymaticreaction with isocitratedehydrogenase. The rate ofincrease in absorbance at 340 nm,due to the formation of NADPH,is directly proportional to themagnesium concentration.Colorimetric endpoint method.In alkaline solution, magnesiumforms a purple complex withxylidyl blue, diazonium salt. Themagnesium concentration ismeasured photometrically viathe decrease in the xylidyl blueabsorbance.
Detection ofAnalyteRate-up EnzymaticEndpoint
SamplesUrineSerum, plasma or urine
Assay Range- Urine1.81 mg/dL to 26.35 mg/dL(0.74 mmol/L to 10.85 mmol/L)1.36 to 26.7 mg/dL(0.56-11.0 mmol/L)
Reference Range- UrineRange (mg/day)72.9 to 121.5Same
Analysis MediumAqueous solutionSame
Use of CalibratorsYesYes
Use of ControlsYesYes

Comparison of Magnesium (LN 3P68) to Roche Magnesium Gen.2

{7}------------------------------------------------

7. Summary of Performance Testing

Limit of Blank, Limit of Detection, and Limit of Quantitation

The study was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP17-A2. LoB was determined using 4 obtained saline samples (zero-analyte samples). LoD and LoQ were determined using low-level analyte samples prepared from a magnesium standard. Two low-analyte level samples were gravimetrically prepared at each of the following 5 target concentration levels: 0.390, 0.780, 1.285, 1.560, 3.130 mg/dL. The zero-analyte samples were tested in replicates of 10. The low-analyte samples were tested in replicates of 10. Testing was performed over 3 days, two runs per day, using 2 lots of reagent, 1 lot of calibrators, and 1 lot of commercially available controls on 1 ARCHITECT c8000 System.

The urine application of the Magnesium assay, had an LoB of 0.04 mg/dL, an LoD of 0.09 mg/dL, and an LoQ of 0.75 mg/dL.

Within-Laboratory Precision (20-Day)

Precision was evaluated using the following control materials.

  • . Level 1: Bio-Rad Liquichek Urine Chemistry Control Level 1
  • Level 2: Bio-Rad Liquichek Urine Chemistry Control Level 2 .
  • . LoQ Urine Pool -Low Mg
  • Human Urine Pool Normal Mg# .
  • Human Urine Pool Abnormal Mg® . * LoQ Urine Pool - Low Mg was prepared by diluting Bio-Rad Liquichek Urine Chemistry Control Level 1 with normal saline. # Human Urine Pool - Normal Mg is a pool of human urine specimens. 5 Human Urine Pool - Abnormal Mg were prepared by spiking normal human urine with a MgCl2 stock solution.

The levels were tested in 2 replicates, 2 times per day (separated by a minimum of 2 hours) for a total of 20 testing days. Testing was performed using 1 lot of reagents, 1 lot of calibrators and 1 lot of commercially available controls on 1

ARCHITECT c8000 System.

The evaluation was performed by instrument based on guidance from CLSI document EP05-A2.

{8}------------------------------------------------

The within-laboratory imprecision (within-run, between-run, and between-day) for Magnesium urine was as follows:

  • . 1.3 %CV for Bio-Rad Level 1
  • . 1.3 %CV for Bio-Rad Level 2
  • . 2.4 %CV for LoQ Urine Pool -Low Mg
  • . 1.8 %CV for Human Urine Pool - Normal Mg
  • 1.8 %CV for Human Urine Pool Abnormal Mg •

Interference

The interference study for endogenous substances (albumin, ascorbic acid, bilirubin (conjugated), calcium, glucose, hemoglobin, and phosphorous) and urine preservatives (acetic acid, boric acid, 6N hydrochloric acid, nitric acid, and sodium fluoride) was performed based on guidance from CLSI document EP07-A2. Interference effects were assessed by comparing test samples containing potentially interfering albumin. ascorbic acid, bilirubin (conjugated), calcium, glucose, hemoglobin, phosphorous, acetic acid, boric acid, 6N hydrochloric acid, nitric acid, and sodium fluoride to control level samples.

The control and test level samples were tested in a minimum of 7 replicates using 1 lot of reagents,1 lot of Multiconstituent Calibrator (LN 1E65) and 1 lot commercially available controls on 1 ARCHITECT c8000 System. The control and test level samples for a given potential interferent/magnesium level combination were tested in the same run.

Additionally, a cation interference study was performed for copper and zinc.

For the copper and zinc study, solutions of each potential interferent were prepared by spiking urine with stock solutions in order to generate copper and zinc samples with concentrations of approximately 21.6 ug/dL and 3504 ug/L respectively. A control sample (urine) and test level samples were tested in 12 replicates using 1 lot of reagents, 1 lot of Multiconstituent Calibrator (LN 1E65) and 1 lot commercially available controls on 1 ARCHITECT c8000 System.

{9}------------------------------------------------

For magnesium samples targeted to 5 mg/dL, the assay showed no more than ± 10% interference for the listed substances at the interferent levels indicated in the following table.

InterferentInterferentLevel
Albumin≤ 64.0 mg/dL
Ascorbic Acid≤ 200 mg/dL
Bilirubin (Conjugated)≤ 59.9 mg/dL
Calcium≤ 26.0 mg/dL
Glucose≤ 1220 mg/dL
Hemoglobin≤ 1200 mg/dL
Phosphorous≤ 307 mg/dL
Boric Acid≤ 1000 mg/dL
6N Hydrochloric Acid≤ 3.0 mL/dL
Copper≤ 21.6 µg/dL
Zinc≤ 3504 µg/L
Iron≤ 0.6 mg/dL

For magnesium samples targeted to 14 or 15 mg/dL, the assay showed no more than ± 10% interference for the listed substances at the interferent levels indicated in the following table.

InterferentInterferent Level
Albumin≤ 64.0 mg/dL
Ascorbic Acid≤ 200 mg/dL
Bilirubin (Conjugated)≤ 59.5 mg/dL
Calcium≤ 27.0 mg/dL
Glucose≤ 1237 mg/dL
Hemoglobin≤ 1200 mg/dL
Phosphorous≤ 313 mg/dL

{10}------------------------------------------------

Boric Acid≤ 1000 mg/dL
6N Hydrochloric Acid≤ 3.0 mL/dL
Copper≤ 21.6 µg/dL
Zinc≤ 3504 µg/L
Iron≤ 0.6 mg/dL

Acetic acid, nitric acid, and sodium fluoride did not meet the evaluation criteria of having a difference within or equal to ± 10% at a target of 5 mg/dL or 15 mg/dL levels of magnesium and will be included in the limitations of the procedure section of the package insert.

Linearity

Linearity was determined based on guidance from Clinical and Laboratory Standards Institute (CLSI) document EP06-A. Three sets of linearity standards were prepared using a magnesium standard and saline (diluent). For each sample set, a low and high sample pool was prepared.

  • . the low sample pool (Level 1) had a concentration greater than 0.0 mg/dL but below the LoQ and
  • . the high sample pool (Level 12) was 20 to 30% beyond the highest expected measurement concentration.

A sample set was prepared for each combined magnesium pool. Each sample set consisted of 12 levels at the following magnesium target concentrations: 0.90, 1.43, 1.97, 3.03, 5.16, 9.43, 13.69, 17.95, 22.21, 26.48, 30.74 and 35.00 mg/dL. Levels 1 through 12 for each sample set were tested in a random order in a minimum of 4 replicates using 2 lots reagents and 1 lot each of Multiconstituent Calibrator (LN 1E65) and commercially available controls on 1 ARCHITECT c8000 System. All levels in a sample set were tested in the same run.

The urine application of the Magnesium assay was demonstrated to be linear across the range of 1.04 to 36.24 mg/dL, which spans the analytical measuring interval of 1.81 to 26.35 mg/dL.

{11}------------------------------------------------

Measuring Interval

The measuring interval was determined based on the results from 3 studies: Within Laboratory Precision (20-Day); Linearity; and Limit of Blank, Limit of Detection, and Limit of Quantitation.

The analytical measuring interval for the Magnesium assay was determined to be from 1.81 to 26.35 mg/dL.

Method Comparison

The study was performed based on guidance from CLSI document EP09-A3. A total of 118 patient urine specimens were evaluated with the Magnesium (LN 3P68) and Roche Magnesium Gen.2 (REF 06407358 190) assays. Of these samples, 12 of the samples were normal urine specimens spiked with magnesium stock to achieve samples with magnesium concentrations in the range of 15.5 to 26 mg/dL. Three aliquots were prepared and tested:

  • a) No acidification
  • b) Acidified using 6N HCl to pH <2 (1.8 to 1.9)
  • c) Acidified using 6N HCl to pH ~1 (0.9 to 1.1)

Two replicates of each aliquot (a), (b) and (c) were run using the Magnesium (LN 3P68) reagent kit where the first valid replicate was used in the analysis. One replicate of each aliquot (c) was run at UT Southwestern Medical Center (UTSW, Dallas) using Roche Magnesium Gen.2 (REF 06407358 190) reagent.

The urine application of the Magnesium (LN 3P68) assay, which had a regression slope of 1.08 and correlation coefficient (r-value) of 1.00 for urine samples at pH <2 across the measuring interval of the assay, a regression slope of 1.04 and correlation coefficient (r-value) of 1.00 for urine samples with no acidification across the measuring interval of the assay, and a regression slope of 1.05 and correlation coefficient (r-value) of 1.00 for urine samples at pH ~1 across the measuring interval of the assay demonstrated acceptable correlation to the predicate device.

{12}------------------------------------------------

Automated Dilution Protocol

One sample was prepared targeting magnesium concentration of 75 mg/dL and evaluated in replicates of 10 using the autodilution protocol. All samples were tested in the same run using 1 lot of reagents and 1 lot each of commercially available calibrators and controls on one ARCHITECT c8000 System.

Magnesium assay results were impacted by not more than ±10% for analyte concentration of 75 mg/dL.

Manual Dilution

Three urine pools were prepared targeting magnesium concentrations of 15, 25, and 40 mg/dL. Endogenous magnesium concentrations were measured in random urine specimens which were then spiked using a concentrated MgCl2 stock to the target concentrations. Each analyte pool was evaluated after 1:2 manual dilution. Dilutions were performed using 0.85% and 0.90% (NaCl) saline. Samples were tested in a minimum of 12 replicates using 1 lot of reagent, 1 lot of Multiconstituent Calibrator (LN 1E65), and 1 lot of commercially available controls on 1 ARCHITECT c8000 System. Diluted samples at a given magnesium level were tested in the same run.

Magnesium assay results were impacted by not more than ±10% for analyte concentrations 15, 25, and 40mg/dL when evaluated with a 1:2 manual dilution (using 0.85% or 0.90% saline).

{13}------------------------------------------------

8. Conclusion Drawn from Performance Testing

The results presented in this 510(k) premarket notification demonstrate that the candidate assay (Magnesium, List No. 3P68) performance is substantially equivalent to the predicate assay (Roche Magnesium Gen.2 (REF 06407358 190)).

The similarities and differences between the candidate assay and the predicate assay are presented in the table on page 4. The results presented in this 510(k) provide reasonable assurance that the Magnesium assay is safe and effective for the stated intended use. Any differences between the candidate assay and the predicate assay shown in the tables do not affect the safety and effectiveness of the candidate assay.

There is no known potential adverse effect to the operator when using this device according to the Magnesium package insert instructions.

§ 862.1495 Magnesium test system.

(a)
Identification. A magnesium test system is a device intended to measure magnesium levels in serum and plasma. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).(b)
Classification. Class I.