K Number
K181668

Validate with FDA (Live)

Manufacturer
Date Cleared
2019-03-13

(261 days)

Product Code
Regulation Number
878.4750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FasTouch Absorbable Fixation System is intended for fixation of prosthetic material to soft tissues in various minimally invasive and open surgical procedures such as hernia repairs.

Device Description

The FasTouch Absorbable Fixation System is a disposable, sterile single-use system designed to deliver absorbable fastener into tissue and prosthesis during general surgery procedures such as hernia repair. The Via Surgical FasTouch Absorbable Fixation System, is designed to be inserted through a 5mm or larger laparoscopic port sleeve. The fasteners two ends are designed to be locked together in the tissue by the Fas Touch firing mechanism, thus forming a closed locked loop into the tissue affixing the surgical mesh to the tissue. The fasteners are absorbable and made of PURASORB PLG 8218 dye with D&C violet No. 2.

AI/ML Overview

This document describes a 510(k) premarket notification for the FasTouch Absorbable Fixation System. It details the device's indications for use, comparison to predicate devices, and performance data used to establish substantial equivalence.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document states that "All tests met the predefined acceptance criteria" for various performance evaluations. However, it does not explicitly list the quantitative acceptance criteria for each test. Instead, it reports either success or compliance with standards.

Test CategorySpecific Test / StandardReported Device Performance
Risk AnalysisISO 14971:2012Performed
BiocompatibilityCytotoxicity Study (ISO 10993-5: 2009)Completed with passing results
Irritation - ISO Intracutaneous Study (ISO 10993-10: 2010)Completed with passing results
Systemic Toxicity Study (ISO 10993:11:2010)Completed with passing results
ISO Maximization Sensitization Test (ISO 10993-10: 2010)Completed with passing results
Pyrogen Study - Material Mediated (ISO 10993-11: 2010)Completed with passing results
Implantation/Toxicity (ISO 10993-6, ISO 10993-11)Completed with passing results
Sterilization, Packaging, and Shelf LifeSterilization validation (ISO 11135-1)Performed, demonstrated compliance
Shelf life and packaging testingPerformed, successfully completed to support labeled shelf life
Bench TestingFixation Strength Evaluation (BT-212)Met predefined acceptance criteria
Mesh compatibility and Integrity (BT-213)Met predefined acceptance criteria
Performance Evaluation (BT-214)Met predefined acceptance criteria
Animal StudyFunctionality and Usability AssessmentNo adverse events, devices performed well, users satisfied
Histopathology evaluation (up to 3 months)Met predefined acceptance criteria

2. Sample size used for the test set and the data provenance

  • Sample Size: The document does not specify the exact sample sizes for the bench tests or the animal study.
  • Data Provenance:
    • Bench Testing: The origin of the data is not explicitly stated, but it is implied to be from the manufacturer's internal testing or a contract lab.
    • Animal Study: Performed at the Institute of Animal Research in Kibbutz Lahav in Israel.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Animal Study: "Several trained surgeons" evaluated the device. Specific number and qualifications (e.g., years of experience, specialization) are not provided beyond "trained surgeons."
  • Other Tests: For bench and biocompatibility tests, the ground truth is established by meeting predefined criteria or compliance with international standards, not by expert consensus in the same way clinical ground truth is established.

4. Adjudication method for the test set

  • The document does not describe any formal adjudication method (e.g., 2+1, 3+1) for the test sets. For the animal study, the assessment was based on the performance observed by "trained surgeons" and histopathology evaluation.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a multi-reader, multi-case comparative effectiveness study was not done. This device is a surgical fixation system, not an AI-powered diagnostic or assistive tool for human readers. Therefore, the concept of "human readers improve with AI" does not apply.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • No, this device is a physical surgical tool. The concept of an "algorithm only" or "standalone" performance without human interaction is not applicable. The device's performance is inherently linked to its use by a surgeon.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Biocompatibility: Ground truth is established by adherence to and passing results from recognized international standards (ISO 10993 series).
  • Sterilization, Packaging, Shelf Life: Ground truth is established by adherence to and passing results from recognized international standards (ISO 11135-1) and successful completion of internal testing.
  • Bench Testing: Ground truth is established by predefined acceptance criteria, which likely relate to mechanical properties, material integrity, and functionality.
  • Animal Study: Ground truth is based on:
    • Direct observation and assessment by "trained surgeons" (functionality and usability).
    • Histopathology evaluation (biological response to implantation).

8. The sample size for the training set

  • This information is not applicable. This device is a physical surgical product and does not involve AI/machine learning models that require a "training set" in the conventional sense.

9. How the ground truth for the training set was established

  • This information is not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in blue, followed by the words "U.S. Food & Drug Administration" in blue.

March 13, 2019

Via Surgical Ltd. % Mr. Leo Basta Owner Northstar Biomedical Associates 93 Benefit Street Providence. Rhode Island 02904

Re: K181668

Trade/Device Name: FasTouch Absorbable Fixation System Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: June 22, 2018 Received: June 25, 2018

Dear Mr. Basta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/5 description: The image shows the name "Joseph Nielsen -S" in a large font. To the right of the name is a digital signature from Joseph Nielsen -S. The signature includes the date 2019.03.13 13:22:02 -04'00' and other information such as DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Joseph Nielsen -S, and 0.9.2342.19200300.100.1.1=20003675.

Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181668

Device Name FasTouch Absorbable Fixation System

Indications for Use (Describe)

The FasTouch Absorbable Fixation System is intended for fixation of prosthetic material to soft tissues in various minimally invasive and open surgical procedures such as hernia repairs.

Type of Use (Select one or both, as applicable)
Research Use (Part 21 CFR 601.21; Subpart B) Over-The-Counter Use (21 CFR 201.66) Research Use (Part 21 CFR 601.21; Subpart B) Over-The-Counter Use (21 CFR 201.66)
Research Use (Part 21 CFR 601.21; Subpart B)
Over-The-Counter Use (21 CFR 201.66)

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Via Surgical. The logo features a stylized, abstract graphic to the left of the company name. The graphic is a gray triangle with a green arrow pointing upwards and to the right. The company name, "Via Surgical," is written in a sans-serif font, with "Via" in a slightly bolder font than "Surgical."

Traditional Premarket Notification Submission - 510(k) FasTouch Absorbable Fixation System

Date Prepared: June 19, 2018

I. SUBMITTER

Via Surgical Ltd. Mitzpe Kineret Street 22/1 Moshav Amirim, 20115, Israel Tel: +972-52-639-5765 Fax: +972-153-77-500-8214

Contact Person

Leo Basta Northstar Biomedical Associates for Via Surgical Ltd. 755 Westminster Street Unit 120 Providence, RI 02903 617.834.9866 (phone) 401.454.1733 lbasta@northstarbiomedical.com

and/or

Orly Maor 25A Sirkin Street Kfar Saba 44421, Israel Tel: +972-9-7453607 Fax: +972-153-9-7453607 oram.ma(@gmail.com

II. DEVICE

Name of Device: FasTouch Absorbable Fixation System

Common or Usual Name: FasTouch Absorbable Fixation System

Classification Name: Implantable staple, Regulation No. 878.4750 Regulatory Class: II

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Via Surgical. The logo features a stylized gray and green icon on the left, resembling a globe with a green arrow piercing through it. To the right of the icon, the text "Via Surgical" is written in a clean, sans-serif font, with the letters in a light gray color.

Product Code: GDW

III. PREDICATE DEVICE

Via Surgical Ltd. believes that the FasTouch Absorbable Fixation System is substantially equivalent to the following predicate device:

  • Via Surgical FasTouch Fixation System cleared under ● K162252 and (product code GDW Regulation No. 878.4750) and K132698 (original submission).
    In addition, the OptifixTM Absorbable Fixation System (C.R. BARD, Inc) cleared under K132134 (product code GDW, Regulation No. 878.4750) is being used as a reference device for its technological characteristics of the fastener used for the same clinical intended use.

IV. DEVICE DESCRIPTION

The FasTouch Absorbable Fixation System is a disposable, sterile single-use system designed to deliver absorbable fastener into tissue and prosthesis during general surgery procedures such as hernia repair.

The Via Surgical FasTouch Absorbable Fixation System, is designed to be inserted through a 5mm or larger laparoscopic port sleeve.

The fasteners two ends are designed to be locked together in the tissue by the Fas Touch firing mechanism, thus forming a closed locked loop into the tissue affixing the surgical mesh to the tissue. The fasteners are absorbable and made of PURASORB PLG 8218 dye with D&C violet No. 2.

V. INDICATIONS FOR USE

The FasTouch Absorbable Fixation System is intended for fixation of prosthetic material to soft tissues in various minimally invasive and open surgical procedures such as hernia repairs

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The FasTouch Absorbable Fixation System has the same intended use as the predicate device. In addition, the subject device has the identical indications for use to that of the reference device. The dimensions of the FasTouch Absorbable Fixation System are identical to the predicate. Similar tests and test methods performed in accordance

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Via Surgical. The logo features a stylized letter "V" with a green arrow pointing upwards and to the right. The text "Via Surgical" is written in a sans-serif font to the right of the "V" symbol. The logo is simple and modern, and the green arrow suggests growth and progress.

with the same standards were used in both FasTouch Absorbable Fixation System and the predicate device to validate the design. The testing results showed that the minor differences in device characteristics between the subject devices and predicate devices do not raise any new questions of safety or effectiveness.

The FasTouch Absorbable Fixation System have the same technological characteristics as the predicate device and as the reference device as demonstrated in the SE table below:

Based on the above analysis, Via Surgical Ltd. believes that the FasTouch Absorbable Fixation System is substantially equivalent to the legally marketed predicate device.

FasTouch AbsorbableFixation SystemFasTouch FixationSystemSE Justification
510(k) NumberNot yet knownK132698 and K162252
ManufacturerVia Surgical Ltd.Via Surgical Ltd.Same
Product CodeGDWGDWSame
CFR878.4750878.4750Same
Intended Use &Indications forUseThe FasTouch AbsorbableFixation System is intendedfor fixation of prostheticmaterial to soft tissues invarious minimally invasiveand open surgical proceduressuch as hernia repairs.The FasTouch FixationDevice is intended forfixation of prostheticmaterial to soft tissues invarious minimallyinvasive and opensurgical procedures suchas hernia repairsSame
Environmentsof UseHospitals, sub-acute careinstitutions and surgerycenterHospitals, sub-acute careinstitutions and surgerycenterSame
PatientPopulationIndividuals undergoingsurgical procedure in whichprosthetic mesh is beingimplantedIndividuals undergoingsurgical procedure inwhich prosthetic mesh isbeing implantedSame
Delivery DeviceHandles with trigger"piston" gripHandles with trigger"piston" gripSame
LoadingCartridge can be loaded inORCartridge can be loadedin ORSame

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Via Surgical. The logo consists of a stylized graphic on the left and the text "Via Surgical" on the right. The graphic features a gray shape with a green arrow pointing upwards. The text "Via Surgical" is in a sans-serif font and is gray in color.

FasTouch AbsorbableFixation SystemFasTouch FixationSystemSE Justification
Shaft Length35.7cm35.7cmSame
FiringMechanismSpring-loadSpring-loadSame
PenetrationDepth6.1mm6.1mmSame
Number ofFasteners2525Same
FastenerMaterialPLGA8218 dyed with D&C 2colorant 0.05% (a copolymerof L-lactide and Glycolide ina 82/18 molar ratio)Non - absorbable polymer(PCU)Same as thereference deviceOptifixAbsorbableFixation Systemcleared underK132134
Fastener DesignStrap - suture likeStrap - suture likeSame
Single PatientUse, DisposableYesYesSame
SterilizationSterile for single use EtOSterile for single use EtOSame
Prescription UseThe device should be usedonly by trained surgeon undera physician order.The device should beused only by trainedsurgeon under a physicianorder.Same

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination:

  • Risk analysis per ISO 14971:2012 -
  • Biocompatibility testing -

An evaluation of biocompatibility was performed in compliance with ISO 10993-1. Biocompatibility evaluation included:

#TestStandard
1Cytotoxicity Study Using the ISO ElutionMethodISO 10993-5: 2009

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Via Surgical. The logo features a stylized, abstract shape in shades of gray and green on the left, resembling a mountain or a stylized letter "V". To the right of the shape, the words "Via Surgical" are written in a clean, sans-serif font, with "Via" slightly larger than "Surgical".

#TestStandard
2Irritation - ISO Intracutaneous StudyISO 10993-10: 2010
3Systemic Toxicity StudyISO 10993:11:2010
4ISO Maximization Sensitization TestISO 10993-10: 2010
5Pyrogen Study - Material MediatedISO 10993-11:2010
6Implantation/ToxicityISO 10993-6ISO 10993-11

All tests were completed with passing results.

Sterilization, Packaging and Shelf Life Testing

Sterilization validation testing of the FasTouch Absorbable Fixation System is performed to demonstrate compliance with ISO 11135-1. In addition, shelf life and packaging testing were performed to support the labeled shelf life. All tests were successfully completed.

Bench Testing

Bench testing included the following: TEST # 1: Fixation Strength Evaluation (BT-212) TEST# 2: Mesh compatibility and Integrity following fixation with the FasTouch Absorbable Fixation System (BT-213) TEST# 3: Performance Evaluation (BT-214) All tests met the predefined acceptance criteria.

Animal Study

Animal Study was performed at the Institute of Animal Research in Kibbutz Lahav in Israel by several trained surgeons evaluating the subject device in comparison to the predicate and reference devices.

The purpose of the study was to assess the functionality and usability of the FasTouch Absorbable Fixation System.

No adverse events occurred, devices performed well without malfunction to users' satisfaction. Users were able to perform the required tasks. Histopathology evaluation was also conducted at the preset intervals up to 3 months. The test met the predefined acceptance criteria.

VIII. CONCLUSIONS

The FasTouch Absorbable Fixation System has the same intended use as the predicate device and the reference device. The principal features of the device that were described, as well as the testing provided, show that the minor differences in device characteristics between the subject device and predicate device do not raise any new questions of safety or effectiveness.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Via Surgical. The logo features a stylized gray and green arrow pointing upwards and to the right. To the right of the arrow is the text "Via Surgical" in a sans-serif font, with "Via" in a slightly darker shade of gray than "Surgical".

Performance data and animal study have been provided using both the predicate and reference device as appropriate, establishing that the FasTouch Absorbable Fixation System performs as intended and in a manner that is substantially equivalent to the predicate. Therefore, the device is substantially equivalent.

§ 878.4750 Implantable staple.

(a)
Identification. An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.