K Number
K181657

Validate with FDA (Live)

Date Cleared
2019-02-01

(224 days)

Product Code
Regulation Number
878.4750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The easyLCTM Linear Cutter Staplers and Loading Units for Single Use have application in gastrointestinal for transection, resection, and/or creation of anastomoses.

The easyCSTM Circular Stapler for Single Use has applications throughout the alimentary for end to side and side to side anastomoses.

Device Description

The easyLCTM Linear Cutter Staplers and Loading Units for Single Use place two double staggered rows of titanium staples and simultaneously cut and divides tissue between the two double rows. The device is available in 60mm, 80mm length with reload in 2.5mm, 3.8mm and 4.8mm staple size to accommodate various tissue thicknesses. The device may be reloaded and fired up to 8 times in a single procedure.

The easyCS™ Circular Stapler for Single Use places a double staggered, circular row of titanium staples upon activation, which was achieved by squeezing the handles firmly as far as they could go. Immediately after formation of the staples, the excess tissue will be resect by the circular knife, and then a circular anastomosis is created. The stapler is available in 21mm, 29mm, 31mm and 33mm five specifications. Two staple sizes (3.5mm and 4.8mm) are available to accommodate various tissue thicknesses.

AI/ML Overview

The provided document is a 510(k) Summary for medical stapling devices (easyLC™ Linear Cutter Staplers and easyCS™ Circular Stapler). It describes the non-clinical tests performed to demonstrate substantial equivalence to predicate devices, but does not include information about clinical studies or acceptance criteria for AI/algorithm performance.

Therefore, I cannot provide a table of acceptance criteria and reported device performance related to AI, nor can I answer questions about sample sizes for test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance.

The document indicates:

Clinical Test Conclusion (Section 7):

  • "No clinical study is included in this submission."

This means the submission relies entirely on non-clinical (bench and biocompatibility) testing to demonstrate substantial equivalence.

Non-Clinical Test Conclusion (Section 6):
Non-clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The tests included:

  • Standards Compliance:
    • ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.
    • ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.
    • ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
    • USP 38-NF 33 <85> Bacterial Endotoxins Tests
    • ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materials;
    • ISO 11137-2:2013 Sterilization of health care products -Radiation- Part 2: Establishing the sterilization dose
    • USP 39-NF 34 <151> Pyrogen
  • Bench Tests (on porcine stomach and intestine):
    • Pressure Resistance Test
    • Closed Staple Dimension Test
    • Staple Formation Test
    • Force Required to Fire Stapler Test
  • Biocompatibility Tests:
    • Cytotoxicity
    • Irritation
    • Skin sensitization
    • Pyrogenicity

Ground Truth for Non-Clinical Tests:

  • The ground truth for these non-clinical tests is inherently defined by the physical and biological properties being measured (e.g., pressure resistance in standard units, staple dimensions, visual inspection of staple formation, force measurements, and standard assays for biocompatibility). There are no "experts" establishing ground truth in the sense of medical image interpretation; rather, the tests themselves provide objective measurements against established scientific and engineering standards.

Training Set:

  • This concept is not applicable as this is a mechanical device, not an AI/ML algorithm.

In summary, the provided document does not contain the information required to answer your questions because it describes the regulatory submission for a physical medical device (staplers) based on non-clinical testing, not an AI-powered diagnostic or assistive device that would have acceptance criteria related to algorithm performance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, with the letters "FDA" in a blue box. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 1, 2019

Ezisurg Medical Co., Ltd. % Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, CN 200120

Re: K181657

Trade/Device Name: easyLCTM Linear Cutter Staplers and Loading Units for Single Use; easyCS™ Circular Stapler for Single Use

Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW, GAG Dated: June 12, 2018 Received: June 22, 2018

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

$$\textbf{David }\textbf{Krausse}\text{-}\textbf{S}{\begin{subarray}{c}\textbf{S}{\begin{subarray}{c}\textbf{Diag}\0\le\textbf{W}\end{subarray}}\text{-}\textbf{S}{\begin{subarray}{c}\textbf{S}{\begin{subarray}{c}\textbf{Diag}\0\le\textbf{W}\end{subarray}}\text{-}\textbf{M}\end{subarray}}$$

  • for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K181657

Device Name

easyLCTM Linear Cutter Staplers and Loading Units for Single Use, easyCSTM Circular Stapler for Single Use

Indications for Use (Describe)

The easyLCTM Linear Cutter Staplers and Loading Units for Single Use have application in gastrointestinal for transection, resection, and/or creation of anastomoses.

The easyCSTM Circular Stapler for Single Use has applications throughout the alimentary for end to side and side to side anastomoses.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2020

See PRA Statement below.

{3}------------------------------------------------

Exhibit #2 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K181657

    1. Date of Preparation: 12/29/2018
    1. Sponsor Identification

Ezisurg Medical Co., Ltd.

Rm.103, Bldg.2,No.1690 Cailun Road, China (Shanghai) Pilot Free Trade Zone, 201203 Shanghai, China.

Establishment Registration Number: Not yet registered.

Contact Person: Liang Xia Position: Non-active Device Senior Regulatory Affairs Supervisor Tel: +86-21-50456176 Fax: +86-21-50676156 Email: xialiang@ezisurg.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 3609253199 Email: info@mid-link.net

{4}------------------------------------------------

  • Identification of Proposed Device 4.
    Trade Name: easyLCTM Linear Cutter Staplers and Loading Units for Single Use easyCSTM Circular Stapler for Single Use Common Name: Stapler and Reload

Regulatory Information

Classification Name: Staple, Implantable Classification: II Product Code: GDW Regulation Number: 21 CFR 878.4750 Review Panel: General & Plastic Surgery

Classification Name: Stapler, Surgical; Classification: I Subsequent Product Code: GAG; Regulation Number: 21CFR 878.4800 Review Panel: General & Plastic Surgery

Intended Use Statement:

The easyLC™ Linear Cutter Staplers and Loading Units for Single Use have application in gastrointestinal for transection, resection, and/or creation of anastomoses.

The easyCS™ Circular Stapler for Single Use has applications throughout the alimentary for end to end, end to side and side to side anastomoses.

Device Description

The easyLCTM Linear Cutter Staplers and Loading Units for Single Use place two double staggered rows of titanium staples and simultaneously cut and divides tissue between the two double rows. The device is available in 60mm, 80mm length with reload in 2.5mm, 3.8mm and 4.8mm staple size to accommodate various tissue thicknesses. The device may be reloaded and fired up to 8 times in a single procedure.

The easyCS™ Circular Stapler for Single Use places a double staggered, circular row of titanium staples upon activation, which was achieved by squeezing the handles firmly as far as they could go. Immediately after formation of the staples, the excess tissue will be resect by the circular knife, and then a circular anastomosis is created. The stapler is available in 21mm, 29mm, 31mm and 33mm

{5}------------------------------------------------

five specifications. Two staple sizes (3.5mm and 4.8mm) are available to accommodate various tissue thicknesses.

  • ર . Identification of Predicate Devices
    Predicate Device 1 510(k) Number: K020779 Product Name: ENDOPATH™ and PROXIMATE™ Linear Cutters and Staplers

Predicate Device 2 510(k) Number: K983536 Device Name: PROXIMATE Curved and Straight Intraluminal Staplers

Predicate Device 3 510(k) Number: K100723 Device Name: PROXIMATE Curved and Straight Intraluminal Staplers

  • Non-Clinical Test Conclusion 6.
    Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.

  • ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.

  • ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity

  • USP 38-NF 33 <85> Bacterial Endotoxins Tests

  • ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materials;

  • ISO 11137-2:2013 Sterilization of health care products -Radiation- Part 2: Establishing the sterilization dose

  • USP 39-NF 34 <151> Pyrogen

Bench test was conducted on porcine stomach and intestine for both proposed device and predicate device to determine substantially equivalence. The bench tests include following tests

  • Pressure Resistance Test

  • Closed Staple Dimension Test

  • Staple Formation Test

{6}------------------------------------------------

Force Required to Fire Stapler Test

Biocompatibility test was conducted on the proposed device, the test include cytotoxicity, irritation, skin sensitization and pyrogenicity.

    1. Clinical Test Conclusion
      No clinical study is included in this submission.
    1. Substantially Equivalent (SE) Comparison
ItemProposed DevicePredicate Device 1
Product CodeGDWGDW
Regulation Number21 CFR 878.475021 CFR 878.4750
Intended UseThe easyLCTM Linear Cutter Staplersand Loading Units for Single Use haveapplication in gastrointestinal fortransection, resection, and/or creation ofanastomoses.The PROXIMATE Linear and VascularLinear Cutters with Safety Lockouthave application in gastrointestinal,gynecologic, thoracic and pediatricsurgery for transection, resection, and/orcreation of anastomoses and can be usedwith staple line or tissue buttressingmaterials, such as bovine pericardium
Cutting MechanismLinearLinear
Operation PrincipleManualManual
Safety MechanismSafety release for prevent frommis-firingSafety release for prevent frommis-firing
Suture Length60mm, 80mm, 100mm55, 75mm
Closed Staple Height1.0mm, 1.5mm, 2.0mm1.0mm, 1.5mm, 1.8mm, 2.0mm
Closed staple formImage: staple formImage: staple form
Endotoxin Limit20 EUSame
LabelingConforms with 21 CFR 801Conforms with 21 CFR 801

Table 1 Comparison for easyLC™ Linear Cutter Staplers and Loading Units for Single Use

{7}------------------------------------------------

ItemProposed DevicePredicate Device 2K983536Predicate Device 3K100723
Product CodeGDWGDWGDW
Regulation Number21 CFR 878.475021 CFR 878.475021 CFR 878.4750
Intended UseThe easyCSTM Circular Stapler forSingle Use has applications throughoutthe alimentary for end to end, end toside and side to side anastomoses.The PROXIMATER Curved andStraight Intraluminal Staplers haveapplication throughout the alimentaryfor end to end, end to side and side toside anastomoses.The Disposable Circular Stapler hasapplication throughout the alimentarytract for end-to-end, end-to-side ancside-to-side anastomoses.
Cutting MechanismCircular KnifeCircular KnifeCircular Knife
Operation PrincipleManualManualManual
Safety MechanismSafety Release is used for preventingfrom mis-firing.Safety Release is used for preventingfrom mis-firing.Safety Release is used for preventingfrom mis-firing.
Diameter21mm, 25mm, 29mm, 31mm, 33mm21mm, 25mm, 29mm, 33mm25.5mm, 29.5mm
Closed Staple Height1.0~2.5mm1.0~2.5mm1.5~2.0mm
Closed staple formImage: staple formImage: staple formImage: staple form
Endotoxin Limit20 EU20EU20EU
LabelingConforms with 21 CFR 801Conforms with 21 CFR 801Conforms with 21 CFR 801
Table 2 Comparison for easy CSTM Circular Stapler for Single Use
--------------------------------------------------------------------

{8}------------------------------------------------

    1. Substantially Equivalent (SE) Conclusion
      Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

§ 878.4750 Implantable staple.

(a)
Identification. An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.