K Number
K181585

Validate with FDA (Live)

Date Cleared
2018-07-20

(35 days)

Product Code
Regulation Number
888.3045
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The JGHI Core Decompression Procedure Kit, consisting of a bone void filler and manual surgical instruments, is intended to be used during core decompression procedures. The bone void filler component resorbs and is replaced with bone during the healing process. The bone void filler included in the JGHI Core Decompression Procedure Kit is not intended to be used as a load-bearing device.

Device Description

The JGHI Core Decompression Procedure Kit, consisting of a bone void filler and manual surgical instruments.

AI/ML Overview

This document is an FDA 510(k) clearance letter for the AgNovos Healthcare USA, LLC's JGH1 Core Decompression Procedure Kit. It confirms that the device is substantially equivalent to legally marketed predicate devices.

However, this document does not contain information about acceptance criteria, device performance studies, sample sizes, expert qualifications, adjudication methods, or AI effectiveness studies.

The letter primarily focuses on:

  • Granting FDA clearance for the JGH1 Core Decompression Procedure Kit (K181585).
  • Identifying the device, its regulation number, and product code.
  • Stating the Indications for Use for the device, which are "during core decompression procedures. The bone void filler component resorbs and is replaced with bone during the healing process. The bone void filler included in the JGHI Core Decompression Procedure Kit is not intended to be used as a load-bearing device."
  • Outlining general regulatory requirements that the manufacturer must still comply with.

Therefore, I cannot provide the requested information based on the provided text. The detail you are asking for would typically be found in the 510(k) submission summary or a separate clinical study report, neither of which are part of this FDA clearance letter.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

July 20, 2018

AgNovos Healthcare USA, LLC Felipe Aguel, Ph.D. Director, Global Regulatory Affairs 7301 Calhoun Place, Suite 100 Rockville, Maryland 20855

Re: K181585

Trade/Device Name: JGH1 Core Decompression Procedure Kit Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MOV Dated: June 6, 2018 Received: June 15, 2018

Dear Dr. Aguel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurence D. Coyne -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K181585

Device Name JGH1 Core Decompression Procedure Kit

Indications for Use (Describe)

The JGHI Core Decompression Procedure Kit, consisting of a bone void filler and manual surgical instruments, is intended to be used during core decompression procedures. The bone void filler component resorbs and is replaced with bone during the healing process. The bone void filler included in the JGHI Core Decompression Procedure Kit is not intended to be used as a load-bearing device.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.