(140 days)
The Immunalysis Ethyl Alcohol Enzyme Assay is an in vitro diagnostic device for the quantitative analysis of ethyl alcohol (ethanol) in human urine, serum or plasma with automated clinical chemistry analyzers. The measurement of ethanol is used for the diagnosis and treatment of alcohol intoxication and poisoning. This assay is calibrated against ethyl alcohol. This device is intended for prescription use only.
The Immunalysis Ethyl Alcohol Assay is based on the oxidation of ethyl alcohol to acetaldehyde by alcohol dehydrogenase (ADH) and nicotinamide adenine dinucleotide (NAD) reduced to NADH resulting in an absorbance change measured spectrophotometrically at 340nm. The concentration of ethanol in the sample is directly proportional to the ADH activity.
The Immunalysis Ethyl Alcohol Enzyme Assay is an in vitro diagnostic device for the quantitative analysis of ethyl alcohol (ethanol) in human urine, serum, or plasma using automated clinical chemistry analyzers. The measurement of ethanol is used for the diagnosis and treatment of alcohol intoxication and poisoning.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Characteristic | Acceptance Criteria (Implicit from Study Design) | Reported Device Performance |
|---|---|---|
| Limit of Blank (LoB) | Calculation from sixty (60) blank measurements of unaltered drug-free negative human urine, serum, and plasma. | Urine: 0.481 mg/dL Serum: 0.668 mg/dL Plasma: 0.652 mg/dL |
| Limit of Detection (LoD) | Testing of four samples per matrix (urine, serum, plasma) at concentrations in the range LoB to 4xLoB, in replicates of five, on two reagent lots over three days. | Urine: 1.3 mg/dL Serum: 1.7 mg/dL Plasma: 1.7 mg/dL |
| Limit of Quantitation (LoQ) | Four independent samples of each matrix spiked at the LoQ concentration were tested on two reagent lots, demonstrating suitable accuracy. | All matrices: 2.9 mg/dL (quantitatively determined with suitable accuracy) |
| Linearity/Recovery | Serially diluted samples (from high concentration EtOH to drug-free) covering 3 mg/dL to 600 mg/dL, tested in triplicate. Slope and r² values to demonstrate linearity. Recovery percentages. | Urine: Slope: 0.996, Intercept: -0.172, r²: 0.999 (Recovery 97.5% - 102.6%) Serum: Slope: 0.978, Intercept: 2.869, r²: 0.999 (Recovery 96.6% - 104.8%) Plasma: Slope: 0.970, Intercept: 0.307, r²: 0.999 (Recovery 95.1% - 106.1%) |
| Precision | Performed for 20 days, 2 runs per day in duplicate (N=80 for each matrix) on drug-free samples spiked with EtOH at 25, 75, 125, 150, 175, 200 mg/dL and calibrators/controls at 50, 100, 300 mg/dL. Spiked concentrations confirmed by GC-FID. CV% values for repeatability, between run, and within-laboratory. | Urine: Within-Laboratory CV% range from 1.5% to 1.8% for spiked samples. Serum: Within-Laboratory CV% range from 2.0% to 3.2% for spiked samples. Plasma: Within-Laboratory CV% range from 1.5% to 4.9% for spiked samples. |
| Specificity and Cross-Reactivity | Structurally and functionally similar compounds spiked into drug-free urine, serum, and plasma at levels yielding the equivalent of 10 mg/dL and 100 mg/dL EtOH. Verified ability of the device to exclusively determine certain drugs. | Low cross-reactivity for most tested compounds (<0.3% to <3.3%). n-Propanol showed higher cross-reactivity (Urine: 16.7% at 10 mg/dL, 9.8% at 100 mg/dL; Serum/Plasma: 11.1% at both 10 mg/dL and 100 mg/dL). |
| Interference (Structurally Unrelated Compounds) | Structurally unrelated compounds spiked into drug-free urine containing EtOH at 10 mg/dL and 100 mg/dL. Assay performance to be unaffected. | Assay performance was unaffected by all tested compounds at specified concentrations (e.g., Acetaminophen 500,000 mg/dL, Caffeine 500,000 mg/dL, Ibuprofen 100,000 mg/dL, etc.). |
| Interference (Endogenous Compounds & Urine Preservatives) | Endogenous compounds and urine preservatives spiked into drug-free urine containing EtOH at 10 mg/dL and 100 mg/dL. Assay performance to be unaffected. | Assay performance was unaffected by all tested endogenous compounds (e.g., Ascorbic Acid 1.5 g/dL, Glucose 2.0 g/dL, Hemoglobin 0.3 g/dL) and urine preservatives (e.g., Boric Acid 1% w/v, Sodium Azide 1% w/v). |
| Interference (Interferents in Serum & Plasma) | Potential interfering substances and anticoagulants spiked into drug-free serum and plasma containing EtOH at 10 mg/dL and 100 mg/dL. Assay performance to be unaffected. | Assay performance was unaffected by all tested compounds in serum/plasma (e.g., Bilirubin total 60-80 mg/dL, Hemoglobin 62.5-1000 mg/dL, Triglycerides 500 mg/dL) and anticoagulants (e.g., Heparin 3 U/mL, EDTA 2 mg/mL). |
| Interference (pH) | Device performance tested over a range of urine pH values (3.0 to 11.0) with EtOH at 10 mg/dL and 100 mg/dL. No interference observed. | No interference observed at urine pH values ranging from 3.0 to 11.0. |
| Interference (Specific Gravity) | Device performance tested over a range of physiologically relevant urine specific gravity values (1.000 to 1.030) with EtOH at 10 mg/dL and 100 mg/dL. No interference observed. | No interference observed at urine specific gravity values ranging from 1.000 to 1.030. |
| Calibration Duration | Drug-free negative urine spiked with EtOH at 10, 100, and 300 mg/dL tested up to 14 days, using an initial two-point calibration. Test results met acceptance criteria. | Calibration is stable for 14 days. Recommended frequency of calibration is 14 days. |
| Specimen Stability | Urine, serum, and plasma samples spiked with EtOH at 100 mg/dL and 300 mg/dL, stored in industry standard collection containers and tested at various time points and temperatures (up to 30°C and 2°C - 8°C). | Urine: Stable up to 14 days at 30°C, up to 6 months at 2°C - 8°C. Serum: Stable up to 10 days at 30°C, up to 14 days at 2°C - 8°C. Plasma: Stable up to 3 days at 30°C, up to 6 days at 2°C - 8°C. |
| Method Comparison | Comparison against a predicate device using de-identified, unaltered leftover clinical urine, serum, and plasma samples. Correlation coefficient (r²) and slope/intercept values. | Urine: n=80, Slope: 1.0284, Intercept: -2.7848, r²: 0.9964 (Range 3-600 mg/dL) Serum: n=76, Slope: 1.0067, Intercept: -5.8802, r²: 0.9948 (Range 3-600 mg/dL) Plasma: n=64, Slope: 1.0595, Intercept: -7.3902, r²: 0.9932 (Range 3-600 mg/dL) |
2. Sample Size Used for the Test Set and the Data Provenance
- Limit of Blank (LoB): 60 blank measurements for urine, 60 for serum, 60 for plasma samples.
- Limit of Detection (LoD): 4 samples per matrix, each tested in replicates of 5 (20 measurements per matrix).
- Limit of Quantitation (LoQ): 4 independent samples per matrix.
- Linearity/Recovery: 12 concentrations for urine, 12 for serum, 12 for plasma, each tested in triplicate.
- Precision: 80 samples per matrix (urine, serum, plasma) at each of 10 concentration levels (25, 50, 75, 100, 125, 150, 175, 200, 300 mg/dL, plus 0 mg/dL for urine and 0.4 mg/dL for serum and 0.2 mg/dL for plasma). This implies 800 data points per matrix for precision.
- Specificity and Cross-Reactivity: Compounds tested at two concentrations (10 mg/dL and 100 mg/dL equivalent EtOH) in urine, serum, and plasma.
- Interference (Structurally Unrelated Compounds): Numerous compounds tested at specific concentrations with EtOH at 10 mg/dL and 100 mg/dL in urine.
- Interference (Endogenous Compounds & Urine Preservatives): Various compounds tested with EtOH at 10 mg/dL and 100 mg/dL in urine.
- Interference (Interferents in Serum & Plasma) & Anticoagulants: Various compounds and anticoagulants tested with EtOH at 10 mg/dL and 100 mg/dL in serum and plasma.
- Interference (pH): Test samples prepared in drug-free urine containing EtOH at 10 mg/dL and 100 mg/dL, tested across a pH range of 3.0 to 11.0.
- Interference (Specific Gravity): Test samples prepared in drug-free urine containing EtOH at 10 mg/dL and 100 mg/dL, tested across specific gravity range of 1.000 to 1.030.
- Calibration Duration: Drug-free negative urine spiked with EtOH at 10, 100, and 300 mg/dL tested at various time points up to 14 days.
- Specimen Stability: Urine, serum, and plasma samples spiked with EtOH at 100 mg/dL and 300 mg/dL, tested at various time points and temperatures.
- Method Comparison: 80 de-identified, unaltered leftover clinical urine samples; 76 de-identified, unaltered leftover clinical serum samples; 64 de-identified, unaltered leftover clinical plasma samples.
Data Provenance:
The data provenance specifies "unaltered drug-free negative human urine, serum, and plasma" for many studies and "de-identified, unaltered leftover clinical urine samples," "de-identified, unaltered leftover clinical serum samples," and "de-identified, unaltered leftover clinical plasma samples obtained from clinical testing laboratories" for the method comparison. This indicates the data is retrospective and derived from human biological samples, likely from clinical settings. The country of origin is not explicitly stated, but given the FDA submission, it is typically expected to be from studies conducted under US or internationally recognized standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This type of device (in vitro diagnostic assay for chemical analysis) does not typically involve human expert consensus for establishing ground truth in the same way an imaging AI algorithm would. Instead, the ground truth for samples (e.g., spiked concentrations, known concentrations in clinical samples) is established through reference methods or highly accurate laboratory techniques.
For the precision study, spiked concentrations were confirmed by Gas Chromatography - Flame Ionization Detector (GC-FID), which is a highly accurate analytical method.
4. Adjudication Method for the Test Set
Not applicable for this type of in vitro diagnostic assay. Ground truth is established by analytical methods, not human adjudication of subjective interpretations.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an in vitro diagnostic assay, not an AI-assisted diagnostic imaging or interpretation device that would involve human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the device operates in a standalone (algorithm only) manner. It is an automated clinical chemistry analyzer which quantitatively measures ethyl alcohol. Its performance is evaluated directly through the performance studies described (LoB, LoD, LoQ, Linearity, Precision, Specificity, Interference, etc.). These studies quantify the device's accuracy and reliability in generating results, without human interpretation of the assay's direct output.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The ground truth used for this device is based on:
- Known concentrations of spiked ethyl alcohol in various matrices (urine, serum, plasma).
- Reference analytical methods, specifically Gas Chromatography - Flame Ionization Detector (GC-FID), to confirm spiked concentrations and potentially for method comparison where samples were cross-validated.
- Comparison to a legally marketed predicate device for method comparison, which itself would have established performance characteristics against accepted methods.
8. The Sample Size for the Training Set
Not applicable. This device is an in vitro diagnostic assay based on an enzymatic reaction, not a machine learning or AI model that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
October 31, 2018
Immunalysis Corporation Wenying (Jessica) Zhu Regulatory Affairs Specialist III 829 Towne Center Drive Pomona, CA 91767
Re: K181553
Trade/Device Name: Immunalysis Ethyl Alcohol Enzyme Assay Regulation Number: 21 CFR 862.3040 Regulation Name: Alcohol test system Regulatory Class: Class II Product Code: DIC Dated: September 19, 2018 Received: September 20, 2018
Dear Wenying (Jessica) Zhu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K181553
Device Name Immunalysis Ethyl Alcohol Enzyme Assay
Indications for Use (Describe)
The Immunalysis Ethyl Alcohol Enzyme Assay is an in vitro diagnostic device for the quantitative analysis of ethyl alcohol (ethanol) in human urine, serum or plasma with automated clinical chemistry analyzers. The measurement of ethanol is used for the diagnosis and treatment of alcohol intoxication and poisoning. This assay is calibrated against ethyl alcohol. This device is intended for prescription use only.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K181553 510(k) SUMMARY
A. GENERAL INFORMATION
| Applicant Name: | Immunalysis Corporation829 Towne Center DrivePomona, CA 91767Establishment # 2020952 |
|---|---|
| Company Contact: | Wenying (Jessica) ZhuRegulatory Affairs Specialist IIIPhone: (909) 451-6697Email: wzhu@immunalysis.com |
| Date Prepared: | September 19, 2018 |
B. DEVICE IDENTIFICATION
| Trade or Proprietary Names: | Immunalysis Ethyl Alcohol Enzyme Assay |
|---|---|
| Common Name: | Ethyl Alcohol Enzyme Assay |
C. REGULATORY INFORMATION
| Device Classification Name: | Alcohol Test System |
|---|---|
| Product Codes: | DIC |
| Regulatory Class: | II |
| Classification Regulation: | 21 CFR 862.3040, Alcohol Test System |
| Panel: | Toxicology (91) |
| Predicate Device: | Lin-Zhi Ethyl Alcohol Enzymatic Assay [K032461] |
D. DEVICE DESCRIPTION
The Immunalysis Ethyl Alcohol Assay is based on the oxidation of ethyl alcohol to acetaldehyde by alcohol dehydrogenase (ADH) and nicotinamide adenine dinucleotide (NAD) reduced to NADH resulting in an absorbance change measured spectrophotometrically at 340nm. The concentration of ethanol in the sample is directly proportional to the ADH activity.
E. INTENDED USE
The Immunalysis Ethyl Alcohol Enzyme Assay is an in vitro diagnostic device for the quantitative analysis of ethyl alcohol (ethanol) in human urine, serum or plasma with automated clinical chemistry analyzers. The measurement of ethanol is used for the diagnosis and treatment of alcohol intoxication and poisoning. This assay is calibrated against ethyl alcohol. This device is intended for prescription use only.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The background is shaped like a horizontal arrow. The word is written in all capital letters. The image is simple and clear.
| Attribute | Predicate DeviceLin-Zhi Ethyl Alcohol EnzymaticAssay [K032461] | Candidate DeviceImmunalysis Ethyl Alcohol EnzymeAssay |
|---|---|---|
| Intended Use | For quantitative analyses of ethylalcohol in human urine, serum, orplasma. | Same |
| Sample Matrix | Urine, serum or plasma | Same |
| Measured Analytes | Ethyl Alcohol (EtOH) | Same |
| Test Principle | Enzymatic assay | Same |
| User Environment | For use in laboratories | Same |
| Test Methodology | In the presence of nicotinamideadenine dinucleotide (NAD), ADHconverts ethyl alcohol toacetaldehyde and reduces NAD toNADH. The ethyl alcoholconcentration is then directlyproportional to the ADH activity.The enzyme activity is measured at340 nm wavelength. | Same |
| Instrumentation | Performed on clinical chemistryanalyzers | Same |
| Reagent Storage | 2-8°C until expiration date | Same |
| Assay Range | 3-600 mg/dL | Same |
F. COMPARISON WITH PREDICATE
G. PERFORMANCE CHARACTERISTICS
The following laboratory performance studies were performed to determine substantial equivalence of the Immunalysis Ethyl Alcohol Enzyme Assay to the predicate device. Assay performance was established using the Olympus AU400e analyzer.
1. Limit of Blank (LoB)/Limit of Detection (LoD)
Sixty (60) blank measurements were taken from unaltered drug free negative human urine, serum and plasma to calculate Limit of blank (LoB). Four samples of each matrix containing concentrations in the range LoB to 4xLoB were tested in replicates of five on two reagent lots over three days to calculate Limit of Detection (LoD). LoB of the assay is 0.481 mg/dL for urine and 0.668 mg/dL for serum and 0.652 mg/dL for plasma. LoD of the assay is 1.3 mg/dL for urine, 1.7 mg/dL for serum and 1.7 mg/dL for plasma.
2. Limit of Quantitation (LoQ)
Limit of Quantitation (LoQ) is selected at a level equal to or higher than LoD. Four independent samples of each matrix spiked at concentration of LoQ were tested on two reagent lots to verify LoQ. Results demonstrated that drug concentration equal to or above LoQ, 2.9 mg/dL, can be quantitatively determined with suitable accuracy.
{5}------------------------------------------------
3. Linearity/Recovery
A linearity study in the quantitative mode was conducted by spiking a drug free urine, serum and plasma pool with a high concentration of EtOH at upper limit of claimed measuring range. Additional pools were made by serially diluting the high concentration specimen with drug free urine, serum and plasma to achieve concentrations ranging from 3 mg/dL to 600 mg/dL. The high value specimen was evaluated with the predicate device to determine initial high-level value. The 0 mg/dL specimen was made from drug free urine, serum and plasma. Each pool was tested in triplicate to calculate the mean concentration values that were used to calculated drug recovery. Linearity test results are presented in Table 1 to Table 3.
| Expected Concentration(mg/dL) | Mean ObservedConcentration(mg/dL) | Recovery (%) | |
|---|---|---|---|
| 0.0 | 0.0 | N/A | |
| 3.0 | 3.1 | 102.2 | |
| 12.0 | 12.0 | 100.3 | |
| 72.0 | 73.9 | 102.6 | |
| 132.0 | 132.2 | 100.2 | |
| 192.0 | 195.4 | 101.8 | |
| 252.0 | 250.3 | 99.3 | |
| 312.0 | 308.0 | 98.7 | |
| 372.0 | 363.0 | 97.6 | |
| 432.0 | 421.3 | 97.5 | |
| 492.0 | 484.4 | 98.5 | |
| 552.0 | 565.8 | 102.5 | |
| Slope | 0.996 | ||
| Intercept | -0.172 | ||
| r2 | 0.999 |
Table 1. Linearity/Recovery Test Results - Urine
Table 2. Linearity/Recovery Test Results - Serum
| Expected Concentration(mg/dL) | Mean ObservedConcentration(mg/dL) | Recovery (%) |
|---|---|---|
| 0.0 | 0.0 | N/A |
| 3.0 | 3.0 | 101.1 |
| 23.2 | 23.1 | 99.4 |
| 83.2 | 87.2 | 104.8 |
| 143.2 | 146.1 | 102.0 |
| 203.2 | 199.8 | 98.3 |
| 263.2 | 262.2 | 99.6 |
| 323.2 | 317.8 | 98.3 |
| 383.2 | 385.9 | 100.7 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The background is shaped like a horizontal arrow. The word is written in all capital letters and is centered on the arrow. The arrow is pointing to the right.
| Expected Concentration(mg/dL) | Mean ObservedConcentration(mg/dL) | Recovery (%) |
|---|---|---|
| 443.2 | 441.4 | 99.6 |
| 503.2 | 493.6 | 98.1 |
| 563.2 | 544.3 | 96.6 |
| Slope | 0.978 | |
| Intercept | 2.869 | |
| r² | 0.999 |
Table 3. Linearity/Recovery Test Results - Plasma
| Expected Concentration(mg/dL) | Mean ObservedConcentration(mg/dL) | Recovery (%) |
|---|---|---|
| 0.0 | 0.0 | N/A |
| 3.0 | 3.0 | 98.9 |
| 9.9 | 10.5 | 106.1 |
| 69.9 | 69.7 | 99.8 |
| 129.9 | 127.5 | 98.2 |
| 189.9 | 188.4 | 99.2 |
| 249.9 | 242.3 | 96.9 |
| 309.9 | 299.9 | 96.8 |
| 369.9 | 351.9 | 95.1 |
| 429.9 | 410.0 | 95.4 |
| 489.9 | 470.4 | 96.0 |
| 549.9 | 547.0 | 99.5 |
| Slope | 0.970 | |
| Intercept | 0.307 | |
| r2 | 0.999 |
4. Precision
Precision study was performed for 20 days, 2 runs per day in duplicate on drug free urine, serum and plasma (N=80 for each matrix) spiked with EtOH to concentrations of 25 mg/dL, 75 mg/dL, 125 mg/dL, 150 mg/dL, 175 mg/dL and 200 mg/dL and calibrators and controls at concentrations of 50 mg/dL, 100 mg/dL and 300 mg/dL. The spiked concentrations were confirmed by Gas Chromatography - Flame Ionization Detector (GC-FID). The study established assay reproducibility and verified the precision performance in the claimed measuring range. Precision test results are presented in Table 4 to Table 6.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color, and the letters are a bright white color. The image is simple and easy to read.
| Concentration(mg/dL) | # ofSample | MeanConc.(mg/dL) | Repeatability | Between Run | Within-Laboratory | |||
|---|---|---|---|---|---|---|---|---|
| Sample | (mg/dL) | SD | CV% | SD | CV% | SD | CV% | |
| 0 | 80 | 0.0 | 0.1 | N/A | 0.0 | N/A | 0.1 | N/A |
| 25 | 80 | 24.5 | 0.3 | 1.2 | 0.3 | 1.1 | 0.5 | 1.8 |
| 50 | 80 | 49.0 | 0.4 | 0.9 | 0.6 | 1.2 | 0.8 | 1.5 |
| 75 | 80 | 74.0 | 0.7 | 0.9 | 1.0 | 1.3 | 1.1 | 1.5 |
| 100 | 80 | 99.6 | 0.8 | 0.8 | 1.6 | 1.6 | 1.6 | 1.6 |
| 125 | 80 | 121.3 | 1.2 | 1.0 | 1.5 | 1.2 | 1.8 | 1.5 |
| 150 | 80 | 146.1 | 1.3 | 0.9 | 1.9 | 1.3 | 2.4 | 1.7 |
| 175 | 80 | 169.1 | 1.4 | 0.9 | 2.0 | 1.1 | 2.7 | 1.6 |
| 200 | 80 | 193.9 | 1.8 | 0.9 | 2.6 | 1.3 | 3.4 | 1.7 |
| 300 | 80 | 292.0 | 2.2 | 0.8 | 3.8 | 1.3 | 4.6 | 1.6 |
Table 4. Precision - Urine
Table 5. Precision – Serum
| Concentration(mg/dL) | # ofSample | MeanConc.(mg/dL) | Repeatability | Between Run | Within-Laboratory | |||
|---|---|---|---|---|---|---|---|---|
| SD | CV% | SD | CV% | SD | CV% | |||
| 0 | 80 | 0.4 | 0.2 | N/A | 0.1 | N/A | 0.2 | N/A |
| 25 | 80 | 25.8 | 0.2 | 0.9 | 0.5 | 2.0 | 0.8 | 3.2 |
| 50 | 80 | 49.3 | 0.3 | 0.7 | 1.0 | 2.0 | 1.0 | 2.0 |
| 75 | 80 | 75.7 | 0.4 | 0.5 | 1.4 | 1.8 | 2.2 | 3.0 |
| 100 | 80 | 99.6 | 0.5 | 0.5 | 2.0 | 2.1 | 2.1 | 2.1 |
| 125 | 80 | 125.9 | 0.8 | 0.6 | 2.4 | 1.9 | 3.9 | 3.1 |
| 150 | 80 | 148.4 | 1.0 | 0.6 | 2.6 | 1.7 | 3.8 | 2.6 |
| 175 | 80 | 172.3 | 1.0 | 0.6 | 3.1 | 1.8 | 5.1 | 3.0 |
| 200 | 80 | 195.2 | 1.3 | 0.7 | 3.4 | 1.8 | 4.9 | 2.5 |
| 300 | 80 | 290.4 | 1.4 | 0.5 | 5.5 | 1.9 | 6.0 | 2.1 |
Table 6. Precision - Plasma
| Concentration(mg/dL) | # ofSample | MeanConc.(mg/dL) | Repeatability | Between Run | Within-Laboratory | |||
|---|---|---|---|---|---|---|---|---|
| SD | CV% | SD | CV% | SD | CV% | |||
| 0 | 80 | 0.2 | 0.1 | N/A | 0.1 | N/A | 0.1 | N/A |
| 25 | 80 | 24.3 | 0.2 | 0.7 | 0.2 | 1.0 | 1.2 | 4.9 |
| 50 | 80 | 49.9 | 0.3 | 0.6 | 0.5 | 0.9 | 0.7 | 1.5 |
| 75 | 80 | 69.8 | 0.4 | 0.5 | 0.7 | 1.0 | 1.8 | 2.6 |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The background is shaped like a horizontal hexagon. The text is centered in the hexagon. The font is sans-serif and bold.
| Concentration(mg/dL) | # ofSample | MeanConc.(mg/dL) | Repeatability | Between Run | Within-Laboratory | |||
|---|---|---|---|---|---|---|---|---|
| 100 | 80 | 100.5 | 0.5 | 0.5 | 1.0 | 1.0 | 1.6 | 1.6 |
| 125 | 80 | 118.3 | 0.6 | 0.5 | 1.0 | 0.9 | 2.7 | 2.2 |
| 150 | 80 | 155.8 | 0.9 | 0.6 | 1.2 | 0.8 | 4.1 | 2.6 |
| 175 | 80 | 170.6 | 0.9 | 0.5 | 1.5 | 0.9 | 3.2 | 1.9 |
| 200 | 80 | 197.6 | 1.1 | 0.6 | 1.4 | 0.7 | 3.9 | 2.0 |
| 300 | 80 | 292.7 | 2.0 | 0.7 | 2.7 | 0.9 | 5.2 | 1.8 |
5. Specificity and Cross-Reactivity
Structurally and functionally similar compounds were spiked into drug free urine, serum and plasma at levels that will yield a result that is equivalent to the concentration of 10 mg/dL and 100 mg/dL. The study verified assay performance relative to the ability of the device to exclusively determine certain drugs, in quantitative mode. Cross-reactivity test results are presented in Table 7 to Table 9.
Table 7. Cross-Reactivity - Urine
| 10 mg/dL | 100 mg/dL | |||
|---|---|---|---|---|
| Compound | Compound Conc.(mg/dL) | Cross-Reactivity(%) | Compound Conc.(mg/dL) | Cross-Reactivity(%) |
| Lot #1 | ||||
| Acetaldehyde | 3,000 | <0.3 | 3,000 | <3.3 |
| Acetone | 3,000 | <0.3 | 3,000 | <3.3 |
| n-Butanol | 500 | 2.0 | 3,000 | <3.3 |
| Ethylene Glycol | 3,000 | <0.3 | 3,000 | <3.3 |
| Isopropanol | 3,000 | <0.3 | 3,000 | <3.3 |
| Methanol | 3,000 | <0.3 | 3,000 | <3.3 |
| n-Propanol | 60 | 16.7 | 1,025 | 9.8 |
| Propylene Glycol | 3,000 | <0.3 | 3,000 | <3.3 |
| Lot #2 | ||||
| Acetaldehyde | 3,000 | <0.3 | 3,000 | <3.3 |
| Acetone | 3,000 | <0.3 | 3,000 | <3.3 |
| n-Butanol | 500 | 2.0 | 3,000 | <3.3 |
| Ethylene Glycol | 3,000 | <0.3 | 3,000 | <3.3 |
| Isopropanol | 3,000 | <0.3 | 3,000 | <3.3 |
| Methanol | 3,000 | <0.3 | 3,000 | <3.3 |
| n-Propanol | 60 | 16.7 | 1,025 | 9.8 |
| Propylene Glycol | 3,000 | <0.3 | 3,000 | <3.3 |
| Lot #3 |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered and in all capital letters. The background is a red hexagon shape with the word inside.
| Compound | 10 mg/dL | 100 mg/dL | ||
|---|---|---|---|---|
| Acetaldehyde | 3,000 | <0.3 | 3,000 | <3.3 |
| Acetone | 3,000 | <0.3 | 3,000 | <3.3 |
| n-Butanol | 500 | 2.0 | 3,000 | <3.3 |
| Ethylene Glycol | 3,000 | <0.3 | 3,000 | <3.3 |
| Isopropanol | 3,000 | <0.3 | 3,000 | <3.3 |
| Methanol | 3,000 | <0.3 | 3,000 | <3.3 |
| n-Propanol | 60 | 16.7 | 1,025 | 9.8 |
| Propylene Glycol | 3,000 | <0.3 | 3,000 | <3.3 |
Table 8. Cross-Reactivity - Serum
| Compound | 10 mg/dL | 100 mg/dL | ||
|---|---|---|---|---|
| Compound Conc.(mg/dL) | Cross-Reactivity(%) | Compound Conc.(mg/dL) | Cross-Reactivity(%) | |
| Acetaldehyde | 3,000 | <0.3 | 3,000 | <3.3 |
| Acetone | 3,000 | <0.3 | 3,000 | <3.3 |
| n-Butanol | 500 | 2.0 | 3,000 | <3.3 |
| Ethylene Glycol | 3,000 | <0.3 | 3,000 | <3.3 |
| Isopropanol | 3,000 | <0.3 | 3,000 | <3.3 |
| Methanol | 3,000 | <0.3 | 3,000 | <3.3 |
| n-Propanol | 90 | 11.1 | 900 | 11.1 |
| Propylene Glycol | 3,000 | <0.3 | 3,000 | <3.3 |
Table 9. Cross-Reactivity - Plasma
| Compound | 10 mg/dL | 100 mg/dL | ||
|---|---|---|---|---|
| Compound Conc.(mg/dL) | Cross-Reactivity(%) | Compound Conc.(mg/dL) | Cross-Reactivity(%) | |
| Acetaldehyde | 3,000 | <0.3 | 3,000 | <3.3 |
| Acetone | 3,000 | <0.3 | 3,000 | <3.3 |
| n-Butanol | 500 | 2.0 | 3,000 | <3.3 |
| Ethylene Glycol | 3,000 | <0.3 | 3,000 | <3.3 |
| Isopropanol | 3,000 | <0.3 | 3,000 | <3.3 |
| Methanol | 3,000 | <0.3 | 3,000 | <3.3 |
| n-Propanol | 90 | 11.1 | 900 | 11.1 |
| Propylene Glycol | 3,000 | <0.3 | 3,000 | <3.3 |
6. Interference – Structurally Unrelated Compounds in Urine
Structurally unrelated compounds were evaluated in quantitative mode by spiking the potential interferent into drug free urine containing EtOH at 10 mg/dL and 100 mg/dL. All potential interferents analyzed verified that assay performance is unaffected by externally ingested
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the word "IMMUNALYSIS" in white text on a red background. The text is bold and appears to be in all capital letters. The red background is a solid color, and the text is centered on the background.
compounds. The levels of structurally unrelated compounds that did not interfere in the assay are presented in Table 10.
| Compound | Compound Conc. (mg/dL) |
|---|---|
| Acetaminophen | 500,000 |
| 6-Acetylcodeine | 100,000 |
| 6-Acetylmorphine | 100,000 |
| Acetylsalicylic Acid | 500,000 |
| Alphenal | 100,000 |
| Alprazolam | 100,000 |
| 7-Aminoclonazepam | 40,000 |
| 7-Aminoflunitrazepam | 100,000 |
| 7-Aminonitrazepam | 100,000 |
| Amitriptyline | 100,000 |
| Amobarbital | 100,000 |
| S-(+)-Amphetamine | 100,000 |
| Aprobarbital | 100,000 |
| Barbital | 100,000 |
| Benzoylecgonine | 100,000 |
| Benzylpiperazine | 100,000 |
| Bromazepam | 100,000 |
| 4-bromo 2-5, dimethoxyphenethylamine | 100,000 |
| Buprenorphine | 50,000 |
| Bupropion | 100,000 |
| Butabarbital | 100,000 |
| Butalbital | 100,000 |
| Caffeine | 500,000 |
| Cannabidiol | 100,000 |
| Cannabinol | 100,000 |
| Carbamazepine | 100,000 |
| Carisoprodol | 100,000 |
| Chlordiazepoxide | 100,000 |
| Chlorpromazine | 100,000 |
| Clobazam | 100,000 |
| Clomipramine | 100,000 |
| Clonazepam | 100,000 |
| Clozapine | 100,000 |
| Cocaine | 100,000 |
| Codeine | 100,000 |
| Compound | Compound Conc. (mg/dL) |
| Cotinine | 100,000 |
| Cyclobenzaprine | 100,000 |
| Cyclopentobarbital | 100,000 |
| Demoxepam | 100,000 |
| Desalkylflurazepam | 100,000 |
| Desipramine | 100,000 |
| Dextromethorphan | 100,000 |
| Diazepam | 100,000 |
| Digoxin | 100,000 |
| Dihydrocodeine | 100,000 |
| Diphenhydramine | 500,000 |
| Dehydronorketamine | 40,000 |
| delta9 THC | 100,000 |
| Doxepin | 100,000 |
| Doxylamine | 100,000 |
| ecgonine | 100,000 |
| ecgonine methyl ester | 100,000 |
| EDDP | 100,000 |
| EMDP | 100,000 |
| 1R,2S Ephedrine | 100,000 |
| 1S,2R Ephedrine | 100,000 |
| Ethyl-β-D-Glucuronide | 50,000 |
| Ethylmorphine | 100,000 |
| Fenfluramine | 100,000 |
| Fentanyl | 100,000 |
| Flunitrazepam | 100,000 |
| Fluoxetine | 100,000 |
| Flurazepam | 100,000 |
| Haloperidol | 100,000 |
| Heroin | 100,000 |
| Hexobarbital | 100,000 |
| hydrocodone | 100,000 |
| hydromorphone | 100,000 |
| 11-hydroxy-delta9 THC | 100,000 |
| Ibuprofen | 100,000 |
| Imipramine | 100,000 |
| Ketamine | 100,000 |
| Lamotrigine | 100,000 |
| Levorphanol Tartrate | 100,000 |
| Compound | Compound Conc. (mg/dL) |
| Lidocaine | 100,000 |
| Lorazepam | 100,000 |
| Lorazepam Glucuronide | 50,000 |
| Lormetazepam | 100,000 |
| LSD | 100,000 |
| Maprotiline | 100,000 |
| MDA | 100,000 |
| MDEA | 100,000 |
| MDMA | 100,000 |
| Meperidine | 100,000 |
| Meprobamate | 100,000 |
| Methamphetamine | 100,000 |
| Methadone | 500,000 |
| Methaqualone | 100,000 |
| Methoxetamine | 100,000 |
| Methylone | 100,000 |
| Methylphenidate | 100,000 |
| Midazolam | 100,000 |
| Morphine | 100,000 |
| Morphine-3-glucuronide | 100,000 |
| Morphine-6-glucuronide | 100,000 |
| N-Desmethyltapentadol | 100,000 |
| Nalorphine | 100,000 |
| Naloxone | 100,000 |
| Naltrexone | 100,000 |
| Naproxen | 100,000 |
| Nitrazepam | 100,000 |
| 11-nor-9-carboxy-delta9-THC | 100,000 |
| Norbuprenorphine | 50,000 |
| Norcodeine | 100,000 |
| Nordiazepam | 100,000 |
| Norketamine | 100,000 |
| Normorphine | 100,000 |
| Norproxyphene | 100,000 |
| Norpseudoephedrine | 50,000 |
| Nortriptyline | 100,000 |
| N-desmethyltramadol | 100,000 |
| N-desmethylvenlafaxine | 100,000 |
| O-desmethyltramadol | 100,000 |
| Compound | Compound Conc. (mg/dL) |
| O-desmethylvenlafaxine | 100,000 |
| Olanzapine | 100,000 |
| Oxycodone | 100,000 |
| Oxymorphone | 100,000 |
| Oxazepam | 100,000 |
| PCP | 100,000 |
| Pentazocine | 100,000 |
| Pentobarbital | 100,000 |
| Phenazepam | 100,000 |
| Phenobarbital | 100,000 |
| Phentermine | 100,000 |
| Phenylephrine | 100,000 |
| Phenylpropanolamine | 100,000 |
| Phenytoin | 100,000 |
| PMA | 100,000 |
| Prazepam | 100,000 |
| Propranolol | 100,000 |
| Propoxyphene | 100,000 |
| Protriptyline | 100,000 |
| R,R Pseudoephedrine | 100,000 |
| S,S Pseudoephedrine | 100,000 |
| Ritalinic Acid | 100,000 |
| Salicylic Acid | 100,000 |
| Sertraline | 100,000 |
| Sufentanil Citrate | 50,000 |
| Talbutal | 50,000 |
| Tapentadol | 100,000 |
| Temazepam | 100,000 |
| Theophylline | 100,000 |
| Thiopental | 100,000 |
| Thioridazine | 100,000 |
| Tramadol | 100,000 |
| Trazadone | 100,000 |
| Triazolam | 100,000 |
| Trifluoromethylphenyl-piperazine | 100,000 |
| Trimipramine | 100,000 |
| Venlafaxine | 100,000 |
| Verapamil | 100,000 |
| Zolpidem Tartrate | 100,000 |
Table 10. Non-Interfering Structurally Unrelated Compounds in Urine
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is in all capital letters and is centered in the image. The background is a solid red color. The image is simple and easy to read.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image is a red arrow pointing to the right with the word "IMMUNALYSIS" written in white letters inside the arrow. The word is written in all capital letters and is centered in the arrow. The arrow is a bright red color, and the letters are a bright white color. The image is simple and easy to read.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The background is shaped like a rounded rectangle with pointed ends, similar to a road sign. The text is bold and centered within the shape.
{14}------------------------------------------------
7. Interference - Endogenous Compounds and Urine Preservatives
Endogenous compounds and urine preservatives were evaluated in quantitative mode by spiking the potential interferent into drug free urine containing EtOH at concentration of 10 mg/dL and 100 mg/dL. Assay performance is unaffected by all the internally existing physiological conditions tested. The levels of endogenous compounds and urine preservatives that did not interfere in the assay are presented in Table 11 to Table 12.
| Compound | 10 mg/dL and 100 mg/dLCompound Conc. |
|---|---|
| Ascorbic Acid | 1.5 g/dL |
| Bilirubin | 0.02 g/dL |
| Creatinine | 0.5 g/dL |
| Galactose | 0.01 g/dL |
| γ-Globulin | 0.5 g/dL |
| Glucose | 2.0 g/dL |
| Hemoglobin | 0.3 g/dL |
| Human Serum Albumin | 0.5 g/dL |
| Oxalic Acid | 0.1 g/dL |
| Riboflavin | 0.0075 g/dL |
| Sodium Chloride | 6.0 g/dL |
| Urea | 6.0 g/dL |
Table 11. Non-interfering Endogenous Compounds - Urine
Table 12. Non-interfering Urine Preservatives - Urine
| Compound | 10 mg/dL and 100 mg/dL | |
|---|---|---|
| Compound Conc. | ||
| Boric Acid | 1% w/v | |
| Sodium Azide | 1% w/v | |
| Sodium Fluoride | 1% w/v | |
| Compound | 10 mg/dL | 100 mg/dL |
| Compound Conc. | Compound Conc. | |
| Acetaminophen | 20 mg/dL | 20 mg/dL |
| Amikacin | 8 mg/dL | 8 mg/dL |
| Ampicillin | 5.3 mg/dL | 5.3 mg/dL |
| Ascorbic Acid | 6 mg/dL | 6 mg/dL |
| Bilirubin total | 60 mg/dL | 80 mg/dL |
| Bilirubin direct | 60 mg/dL | 80 mg/dL |
| Caffeine | 6 mg/dL | 6 mg/dL |
| Carbamazepine | 3 mg/dL | 3 mg/dL |
| Chloramphenicol | 5 mg/dL | 5 mg/dL |
| Chlordiazepoxide | 1 mg/dL | 1 mg/dL |
| Chlorpromazine | 0.2 mg/dL | 0.2 mg/dL |
| Cholesterol | 30 mg/dL | 503 mg/dL |
| Cimetidine | 2 mg/dL | 2 mg/dL |
| Creatinine | 30 mg/dL | 30 mg/dL |
| Dextran | 6000 mg/dL | 6000 mg/dL |
| Diazepam | 0.51 mg/dL | 0.51 mg/dL |
| Digoxin | 6.1 mg/dL | 6.1 mg/dL |
| Erythromycin | 6 mg/dL | 6 mg/dL |
| Ethosuximide | 25 mg/dL | 25 mg/dL |
| Furosemide | 6 mg/dL | 6 mg/dL |
| Gentamicin | 1 mg/dL | 1 mg/dL |
| Hemoglobin | 62.5 mg/dL | 1,000 mg/dL |
| Heparin | 3 U/mL | 3 U/mL |
| Ibuprofen | 50 mg/dL | 50 mg/dL |
| IgG | 5,000 mg/dL | 5,000 mg/dL |
| Intralipid | 187.5 mg/dL | 750 mg/dL |
| Lactate Dehydrogenase | 1,000 U/L | 237,500 U/L |
| Lactate | 150 mg/dL | 901 mg/dL |
| Lidocaine | 1.2 mg/dL | 1.2 mg/dL |
| Lithium | 2.2 mg/dL | 2.2 mg/dL |
| Mannitol | 500 mg/dL | 500 mg/dL |
| Penicillin | 25 U/mL | 25 U/mL |
| Pentobarbital | 8 mg/dL | 8 mg/dL |
| Phenobarbital | 10 mg/dL | 10 mg/dL |
| Phenytoin | 5 mg/dL | 5 mg/dL |
| Primidone | 4 mg/dL | 4 mg/dL |
| Propoxyphene | 0.16 mg/dL | 0.16 mg/dL |
| HSA (albumin) | 37.5 mg/dL | 1,250 mg/dL |
| Compound | 10 mg/dL | 100 mg/dL |
| Compound Conc. | Compound Conc. | |
| Protein (total) | 50 mg/dL | 1,500 mg/dL |
| Salicylic Acid | 60 mg/dL | 60 mg/dL |
| Theophylline | 4 mg/dL | 4 mg/dL |
| Triglycerides | 500 mg/dL | 500 mg/dL |
| Urea | 500 mg/dL | 500 mg/dL |
| Uric Acid | 20 mg/dL | 20 mg/dL |
| Valproic Acid | 50 mg/dL | 50 mg/dL |
8. Interference - Interferents in Serum and Plasma
Potential interfering substances in serum and plasma matrices were evaluated in quantitative mode by spiking the compound into drug free serum and plasma containing EtOH at concentration of 10 mg/dL and 100 mg/dL, respectively. Additional anticoagulants were evaluated by spiking into drug free plasma containing EtOH at concentration of 10 mg/dL and 100 mg/dL. Assay performance is unaffected by all the internally existing physiological conditions tested. The levels of substances that did not interfere in the assay are presented in Table 13 and Table 14.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color, and the letters are a bright white color. The image is simple and easy to read.
Table 13. Non-interfering Substances – Serum and Plasma
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered in the image and is written in all capital letters. The background is a bright red color, and the letters are a slightly lighter shade of white. The image is simple and easy to read.
Table 14. Non-interfering Anticoagulants -Plasma
| Compound | 10 mg/dLCompound Conc. | 100 mg/dLCompound Conc. |
|---|---|---|
| Heparin | 3 U/mL | 3 U/mL |
| EDTA | 2 mg/mL | 2 mg/mL |
| Oxalic Acid | 2 mg/mL | 2 mg/mL |
| Sodium Citrate | 3.8 g/dL | 3.8 g/dL |
| Sodium Fluoride | 2 mg/mL | 2 mg/mL |
9. Interference - pH
To evaluate potential interference from the effect of urine pH, device performance in the quantitative mode was tested using a range of urine pH values. All test samples were prepared in drug-free urine containing EtOH at concentration of 10 mg/dL and 100 mg/dL. No interference was observed at urine pH values ranging from 3.0 to 11.0.
10. Interference - Specific Gravity
To evaluate potential interference from the specific gravity of urine, device performance in the quantitative mode was tested using a range of physiologically relevant urine specific gravity values. All test samples were prepared in drug free urine containing EtOH at concentration of 10 mg/dL and 100 mg/dL. No interference was observed at urine specific gravity values ranging from 1.000 to 1.030.
11. Calibration Duration
Drug free negative urine spiked with EtOH at concentration of 10 mg/dL, 100 mg/dL and 300 mg/dL were tested in quantitative mode at time points up to 14 days. At the initial time point, a two-point calibration curve was established. This calibration was used through the duration of this study. The test results met acceptance criteria at each time point. The recommended frequency of calibration is 14 days.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color. The image is simple and easy to read.
12. Specimen Stability
Drug free negative urine, serum and plasma spiked with EtOH at concentration of 100 mg/dL and 300 mg/dL were stored in industry standard collection containers and tested by assay kit at each time point at 30°C and at 2°C - 8°C. Test results indicated that urine samples containing EtOH are stable for up to 14 days stored at ambient temperature up to 30℃, up to 6 months stored at 2°C - 8°C; serum samples containing EtOH are stable for up to 10 days stored at ambient temperature up to 30°C, up to 14 days stored at 2°C - 8°C; plasma samples containing EtOH are stable for up to 3 day stored at ambient temperature up to 30°C, up to 6 days stored at 2°C - 8°C.
13. Method Comparison
Eighty (80) de-identified, unaltered leftover clinical urine samples, seventy-six (76) de-identified, unaltered leftover clinical serum samples and sixty-four (64) de-identified, unaltered leftover clinical plasma samples obtained from clinical testing laboratories were analyzed for EtOH with the Immunalysis Ethyl Alcohol Enzyme Assay in quantitative mode and compared to results of predicate device. The instruments used were an Olympus AU 400e. Method comparison test results are presented in Table 15.
Table 15. Method Comparison Test Results
| Matrix | n | Slope | Intercept | r² | CandidateDevice Range(mg/dL) | PredicateDeviceRange(mg/dL) |
|---|---|---|---|---|---|---|
| Urine | 80 | 1.0284 | -2.7848 | 0.9964 | 3-600 | 3-600 |
| Serum | 76 | 1.0067 | -5.8802 | 0.9948 | 3-600 | 3-600 |
| Plasma | 64 | 1.0595 | -7.3902 | 0.9932 | 3-600 | 3-600 |
CONCLUSION
The information provided in this pre-market notification demonstrates that the Immunalysis Ethyl Alcohol Enzyme Assay is substantially equivalent to the legally marketed predicate device for its intended use.
§ 862.3040 Alcohol test system.
(a)
Identification. An alcohol test system is a device intented to measure alcohol (e.g., ethanol, methanol, isopropanol, etc.) in human body fluids (e.g., serum, whole blood, and urine). Measurements obtained by this device are used in the diagnosis and treatment of alcohol intoxication and poisoning.(b)
Classification. Class II.