(57 days)
The Multi-Micro Dose (MMD) Tip accessory is indicated for use in conjunction with the XTRAC laser system to filter the NB-UVB light at delivery in order to calculate the maximum non-blistering dose for a particular patient.
The Multi-Micro Dose (MMD) Tip accessory is indicated for use in conjunction with the XTRAC laser system to filter the NB-UVB light at delivery in order to calculate and individualize the maximum non-blistering dose for a particular patient.
The subject device is the Multi-Micro Dose ("MMD") Tip.
The Multi-Micro Dose ("MMD") tip provides another method, separate of the MED mode, for determining the Optimal Therapeutic Dose ("OTD") for each patient. The OTD is the power setting on the XTRAC Phototherapy system customized for each patient that enables the user to provide patient specific treatment doses. As an alternative to the trial and error method of the MED mode, using the MMD Tip can often determine the optimal XTRAC power setting with a single test application.
The MMD Tip filters and reduces the laser output, providing four different dose levels in a single application. The reaction of the target area to the different dose levels assists in determining the patient specific Optimal Therapeutic Dose ("OTD").
The MMD Tip design places four (5mm x 5mm) square filters in the laser aperture.
The MMD Tip design is constructed of similar materials and components as the predicate tip. It connects to the XTRAC hand-piece in the same manner as the standard Tips of the XTRAC system.
This document describes the MMD Tip, an accessory for the XTRAC laser system. The document is a 510(k) premarket notification and focuses on establishing substantial equivalence to a legally marketed predicate device (K073659 - Strata Skin Sciences - XTRAC).
Here's an analysis of the acceptance criteria and study information provided:
1. A table of acceptance criteria and the reported device performance:
The document primarily states that "The performance was within pre-defined acceptance criteria" for the bench tests. It does not provide specific numerical acceptance criteria or the exact reported performance values. It only lists the tests performed and a general statement of compliance.
| Acceptance Criteria (Not explicitly stated in numerical terms) | Reported Device Performance |
|---|---|
| Bench Testing: | |
| Tip Dose Accuracy and Uniformity | Met performance criteria |
| Tip Transmission | Met performance criteria |
| Tip Shelf Life | Met performance criteria |
| Tip Shipping | Met performance criteria |
| Biocompatibility Testing (ISO 10993-1): | |
| Cytotoxicity | Non-cytotoxic |
| Sensitization | Non-irritant |
| Irritation or Intracutaneous Reactivity | Non-toxic |
| Acute Systemic Toxicity | Non-toxic |
| Material Mediated Pyrogenicity | Non-toxic |
2. Sample size used for the test set and the data provenance:
- Bench Testing: The document does not specify the sample size for any of the bench tests (Tip Dose Accuracy and Uniformity, Tip Transmission, Tip Shelf Life, Tip Shipping). It merely states that "We performed testing to support the filtration performance of the MMD Tip" and "We have performed bench tests and found that the MMD Tip... met all requirements specifications."
- Biocompatibility Testing: The document does not specify the sample size used for the biocompatibility tests.
- Data Provenance: Not specified, but given the nature of the tests (bench and biocompatibility), the data would originate from laboratory or manufacturing testing. There is no mention of human subject data for proving the device's functionality.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable to this submission. The MMD Tip is an accessory to a laser system used to calculate the Optimal Therapeutic Dose (OTD) for patients. The studies described are bench and biocompatibility tests, not clinical studies involving expert interpretation of results to establish ground truth about medical conditions.
4. Adjudication method for the test set:
This information is not applicable. There is no mention of a human test set requiring adjudication in this submission.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. The MMD Tip is a physical accessory for a laser system. It is not an AI-powered diagnostic or assistive device, and therefore no MRMC study or AI-related effectiveness study was conducted or is relevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This information is not applicable. The MMD Tip is a physical device, not an algorithm.
7. The type of ground truth used:
- Bench Testing: The ground truth would be established by physical measurements and engineering specifications, e.g., using calibrated measuring equipment to verify dose accuracy and transmission, or environmental chambers for shelf-life testing.
- Biocompatibility Testing: The ground truth is established by adherence to recognized international standards, specifically ISO 10993-1 tests, which have defined endpoints for toxicity, sensitization, etc.
8. The sample size for the training set:
This information is not applicable. The MMD Tip is a physical accessory and does not involve AI or machine learning that would require a training set.
9. How the ground truth for the training set was established:
This information is not applicable for the same reasons as above.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 1, 2018
Strata Skin Sciences, Inc. % Paul Dryden Consultant Strata Skin Sciences, Inc. % ProMedic, LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704
Re: K181480
Trade/Device Name: MMD Tip Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 2, 2018 Received: June 5, 2018
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/ofpmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
MMD Tip
Indications for Use (Describe)
The Multi-Micro Dose (MMD) Tip accessory is indicated for use in conjunction with the XTRAC laser system to filter the NB-UVB light at delivery in order to calculate the maximum non-blistering dose for a particular patient.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
510(k) Summary 5
We have prepared the 510(k) Summary according 21 CFR 807.92. This regulation has a broad definition of acceptability. We have reviewed other similar Summaries and included the applicable information and details.
{4}------------------------------------------------
| Date prepared: | June 2, 2018 |
|---|---|
| Strata Skin Sciences, Inc.100 Lakeside Drive, Suite 100Horsham, PA 19044 | Tel - 215-619-3200 |
| Official Contact: | Shmuel Gov – VP & General Manager |
| Proprietary or Trade Name: | MMD Tip |
| Common/Usual Name: | Laser surgical instrument for use in general and plastic surgery and inDermatology |
| Classification Code/Name: | GEX - Laser surgical instrument for use in general and plastic surgery and indermatology21CFR 878.4810, Class 2 |
| Device: | Multi-Micro Dose “MMD” Tip |
| Predicate Device: | K073659 - Strata Skin Sciences - XTRAC |
Device Description:
The XTRAC Ultra 2 Excimer Laser System, Model AL 10000 utilizes a hand-piece which contains tips and the subject device attaches in the same manner.
The subject device is the Multi-Micro Dose ("MMD") Tip.
The Multi-Micro Dose ("MMD") tip provides another method, separate of the MED mode, for determining the Optimal Therapeutic Dose ("OTD") for each patient. The OTD is the power setting on the XTRAC Phototherapy system customized for each patient that enables the user to provide patient specific treatment doses. As an alternative to the trial and error method of the MED mode, using the MMD Tip can often determine the optimal XTRAC power setting with a single test application.
The MMD Tip filters and reduces the laser output, providing four different dose levels in a single application. The reaction of the target area to the different dose levels assists in determining the patient specific Optimal Therapeutic Dose ("OTD").
The MMD Tip design places four (5mm x 5mm) square filters in the laser aperture. A drawing of the MMD Tip is shown below in comparison to the existing Standard Tip.
Indications for Use:
The Multi-Micro Dose (MMD) Tip accessory is indicated for use in conjunction with the XTRAC laser system to filter the NB-UVB light at delivery in order to calculate and individualize the maximum nonblistering dose for a particular patient.
We note that the XTRAC system has the following indications:
The XTRAC Excimer Laser Phototherapy System is indicated for use in the treatment of psoriasis, vitiligo, atopic dermatitis and leukoderma by the application of targeted UVB light.
{5}------------------------------------------------
Patient Population: Patients undergoing treatment of psoriasis, vitiligo, atopic dermatitis and leukoderma by the application of targeted UVB light.
Environment of Use: Anywhere a patient may undergo a treatment with the XTRAC system.
Substantial Equivalence Discussion: Table 1 below compares the key features of the proposed device with the predicate.
| Feature | PredicateK073659XTRAC system | Subject DeviceMMD Tip for use with XTRAC system |
|---|---|---|
| Indications for Use | The XTRAC Excimer LaserPhototherapy System (Model AL10000) is indicated for use in the treatment of psoriasis, vitiligo, atopic dermatitis and leukoderma by the application of targeted UVB ultraviolet light. | The XTRAC Excimer Laser Phototherapy System (Model AL10000) is indicated for use in the treatment of psoriasis, vitiligo, atopic dermatitis and leukoderma by the application of targeted UVB ultraviolet light.The Multi-Micro Dose (MMD) Tip accessory is indicated for use in conjunction with the XTRAC laser system to filter the NB-UVB light at delivery in order to calculate and individualize the maximum non-blistering dose for a particular patient. |
| Classification | Laser surgical instrument for use in general and plastic surgery and in dermatology21CFR 878.4810Product Code: GEX | Laser surgical instrument for use in general and plastic surgery and in dermatologyAn Accessory to this device21CFR 878.4810Product Code: GEX |
| Patient Population | Patients needing treatment of psoriasis, vitiligo, atopic dermatitis and leukoderma by the application of targeted UVB ultraviolet light | Patients needing treatment of psoriasis, vitiligo, atopic dermatitis and leukoderma by the application of targeted UVB ultraviolet light |
| Environment of use | Clinical settings | Clinical settings |
| Single Use | Equipment is reusableTips are reusable | Tips are Single patient use, disposable |
| Sterile | No | No |
| Mechanism of action | Tip directs the UVB light to the target area.Dose – 100% | MMD Tip provides four micro-dose outputsDose - 94%, 60%, 45% and 30% |
| Biocompatibility | Surface ContactSkin and BreachedLimited duration (<24 hours) | Surface ContactSkin and BreachedLimited duration (<24 hours) |
| Testing | CytotoxicitySensitizationIrritation or Intracutaneous ReactivityAcute Systemic ToxicityMaterial Mediated Pyrogenicity | |
| Bench testing | Tip Dose Accuracy and UniformityTip Transmission VerificationTip Shelf LifeTip Shipping |
{6}------------------------------------------------
Indications for Use / Patient Population / Environment of Use:
As in comparison of Indications For Use above, the indications for the proposed MMD Tip is as an accessory to the predicate K073659 XTRAC system with its standard tips.
Discussion: There are no differences in proposed indications for use related to this accessory tip. It allows different dose levels to be delivered at one time for evaluation as to the dose which is non-blistering.
This difference does not raise different questions of risk or safety concerns compared to the predicate.
Prescriptive:
The devices are prescription devices. Discussion: There are no differences.
Design and Technology:
The MMD Tip design is constructed of similar materials and components as the predicate tip. It connects to the XTRAC hand-piece in the same manner as the standard Tips of the XTRAC system. Discussion: There are no differences in design, technology, or principle of operation which would raise different safety or effectiveness concerns compared to the predicate.
Performance and Specifications:
We performed testing to support the filtration performance of the MMD Tip when used with the predicate XTRAC system. The performance was within pre-defined acceptance criteria. Discussion: There are no differences in the performance which would raise different safety or effectiveness concerns compared to the predicate.
Performance Testing:
Nonclinical / Bench:
We have performed bench tests and found that the MMD Tip when used with the XTRAC System met all requirements specifications and can be found to be equivalent in comparison to the predicate. Testing included
- Tip Dose Accuracy and Uniformity
- Tip Transmission ●
- Tip Shelf Life ●
- Tip Shipping .
In all cases the subject device met its performance criteria.
Biocompatibility:
We assessed the patient contact type and performed the applicable ISO 10993-1 tests which included:
- Cytotoxicity ●
- Sensitization
- Irritation or Intracutaneous Reactivity .
- . Acute Systemic Toxicity
- Material Mediated Pyrogenicity ●
The test results supported the materials as non-cytotoxic, non-irritants, and non-toxic.
Discussion of Differences
The only difference is that the subject MMD Tip has 4 filters and allows the user to deliver difference doses in one procedure vs. multiple doses.
{7}------------------------------------------------
This difference does not raise different safety or effectiveness concerns.
Substantial Equivalence Conclusion
Based upon the performance testing and comparison to the legally marketed predicate device for indications for use, technology, and performance we have demonstrated that the MMD Tip when used with the XTRAC system, K073659, is substantially equivalent in safety and effectiveness to the predicate device.
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.