K Number
K181469
Date Cleared
2018-11-09

(158 days)

Product Code
Regulation Number
884.6160
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Cryotop®US-flash and Cryotop®US-scoop are cryopreservation storage devices that are intended for use in vitrification procedures to contain and maintain human oocytes (MII) and embryos.

Device Description

This 510(k) covers two subject devices, the Cryotop®US-flash and Cryotop®US-scoop that are modified versions of the predicate device (Cryotop®US, K153027). Both devices are cryopreservation storage devices intended for embryo/oocyte vitrification. Cryotop®US-flash and Cryotop®US-scoop are provided sterile and are for single-use only.

The Cryotop®US-flash is composed of a polypropylene straw and an acrylonitrile butadiene styrene handle with a polyethylene terephthalate fine tip where oocytes or embryos are loaded. The fine device is flat with a triangular-shaped area for oocyte/embryo loading. The Cryotop®US-scoop is composed of a polypropylene straw and a polystyrene handle and fine tip where oocytes or embryos are loaded. The fine tip of this device is U-shaped. After oocyte or embryo loading, the handle of the subject devices is inserted into the straw enclosure. A hermetic seal is created via a tapered handle with a stop location integrated into the handle. As the handle is placed into the straw enclosure it creates a closed system keeping the fine tip isolated from the liquid nitrogen. The straw component of both devices also includes a weight at its distal end to aid in maintaining correct device position during storage in liquid nitrogen.

AI/ML Overview

The provided text describes a 510(k) premarket notification for two new cryopreservation storage devices: Cryotop®US-flash and Cryotop®US-scoop. The submission aims to demonstrate substantial equivalence to a predicate device, Cryotop®US (K153027). The document focuses on non-clinical performance data to support this claim, rather than a clinical study comparing the devices.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based only on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Test CategorySpecific Test / CriterionAcceptance CriteriaReported Device Performance (Cryotop®US-flash & Cryotop®US-scoop)
PerformanceCooling Rate3,000°C/minPassed: 3,000°C/min
Warming Rate44,000°C/minPassed: 44,000°C/min
Physical/MechanicalDimensional TestingOuter diameter and length per specifications (specific values not provided in this extract)Passed (implied)
Durability TestingNo burst or liquid nitrogen ingress into straw after 30-second immersionPassed
Mechanical Tensile TestingTensile strength to withstand 5NPassed
BiocompatibilityBacterial Endotoxins<0.5 EU/device per USP <85> and ANSI/AAMI ST72:2002/(R)2010Passed: <0.5 EU/device
Mouse Embryo Assay (MEA)≥80% of 1-cell embryos developed to blastocysts at 96 hoursPassed: ≥80%
EnvironmentalTransportation Simulation TestingPer ASTM D4169-09Passed (implied)
Shelf-LifePackage Integrity Testing: Dye penetrationPer ASTM F1929-15Passed (implied)
Package Integrity Testing: Seal strengthPer ASTM F88/F88M-15Passed (implied)
Shelf-Life (Dimensional, Mechanical, Durability, Endotoxin, MEA)In accordance with methods & acceptance criteria mentioned above for each testPassed (implied)

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated for each individual test. The document mentions "non-clinical performance testing" was conducted, implying multiple tests on multiple units, but no specific numbers of devices tested are provided.
  • Data Provenance: The tests were conducted by the manufacturer, Kitazato Corporation, as "internal requirements." The text does not specify the country of origin of the data beyond "Kitazato Corporation 81 Nakajima, Fuji-city Shizuoka 416-0907 JAPAN." These are retrospective tests performed for the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not provided as the tests are non-clinical, laboratory-based performance tests, not clinical studies requiring expert interpretation of outcomes.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • This information is not applicable as the tests are non-clinical and do not involve human interpretation or adjudication in the context of clinical outcomes. The results are objective measurements against defined criteria.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC comparative effectiveness study was done. This is a 510(k) submission for a medical device (cryopreservation storage device), not an AI-powered diagnostic tool, and therefore, an MRMC study is not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • This is not applicable as the submission is for a medical device (cryopreservation storage device), not an algorithm or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the non-clinical performance tests, the "ground truth" is established by objective physical, chemical, and biological measurements against pre-defined, quantitative standards and specifications (e.g., specific temperatures, mechanical forces, endotoxin levels, embryo development rates).

8. The sample size for the training set:

  • This concept is not applicable to this submission, as it's for physical medical devices and the non-clinical tests described do not involve machine learning or a training set.

9. How the ground truth for the training set was established:

  • This question is not applicable as there is no training set for the described non-clinical performance tests for this medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a blue square and the agency's name in a sans-serif font.

November 9, 2018

Kitazato Corporation % Diane Sudduth Senior Consultant, RA Emergo Global Consulting. LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, TX 78746

Re: K181469

Trade/Device Name: Cryotop®US-flash and Cryotop®US-scoop Regulation Number: 21 CFR& 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MQK Dated: October 10, 2018 Received: October 11, 2018

Dear Diane Sudduth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael T. Bailey -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181469

Device Name

Cryotop®US-flash and Cryotop®US-scoop

Indications for Use (Describe)

Cryotop@US-flash and Cryotop®US-scoop are cryopreservation storage devices that are intended for use in vitrification procedures to contain and maintain human oocytes (MII) and embryos.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary — K181469

1. Submission Sponsor

Kitazato Corporation 81 Nakajima, Fuji-city Shizuoka 416-0907 JAPAN Phone number: +(81) 546-66-2202 Contact: Futoshi Inoue Title: President

2. Submission Correspondent

Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327.9997 Contact: Diane Sudduth Title: Senior Consultant, RA

3. Date Prepared

November 8, 2018

4. Device Identification

Trade Name:Cryotop®US-flash and Cryotop®US-scoop
Common Name:Cryopreservation Storage Device
Regulation Name:Assisted Reproduction Labware
Regulation Number:21 CFR 884.6160
Product Code:MQK (Labware, Assisted Reproduction)
Regulatory Class:II

5. Predicate Device

Cryotop®US (K153027) manufactured by Kitazato Corporation (formally KITAZATO BIOPHARMA CO., LTD.) This predicate device has not been subject to any design related recalls.

6. Device Description

This 510(k) covers two subject devices, the Cryotop®US-flash and Cryotop®US-scoop that are modified versions of the predicate device (Cryotop®US, K153027). Both devices are cryopreservation storage devices intended for embryo/oocyte vitrification. Cryotop®US-flash and Cryotop®US-scoop are provided sterile and are for single-use only.

{4}------------------------------------------------

The Cryotop®US-flash is composed of a polypropylene straw and an acrylonitrile butadiene styrene handle with a polyethylene terephthalate fine tip where oocytes or embryos are loaded. The fine device is flat with a triangular-shaped area for oocyte/embryo loading. The Cryotop®US-scoop is composed of a polypropylene straw and a polystyrene handle and fine tip where oocytes or embryos are loaded. The fine tip of this device is U-shaped. After oocyte or embryo loading, the handle of the subject devices is inserted into the straw enclosure. A hermetic seal is created via a tapered handle with a stop location integrated into the handle. As the handle is placed into the straw enclosure it creates a closed system keeping the fine tip isolated from the liquid nitrogen. The straw component of both devices also includes a weight at its distal end to aid in maintaining correct device position during storage in liquid nitrogen.

7. Indication for Use

Cryotop®US-flash and Cryotop®US-scoop are cryopreservation storage devices that are intended for use in vitrification procedures to contain and maintain human oocytes (MII) and embryos.

DevicesK181469 (subject device)K153027 (predicate device)
Indications forUseCryotop®US-flash and Cryotop®US-scoop arecryopreservation storage devices that areintended for use in vitrification procedures tocontain and maintain human oocytes (MII) andembryos.The Cryotop®US is a cryopreservationstorage device that is intended for use invitrification procedures to contain andmaintain human 4-8 cell and blastocyststage embryos.
FundamentaldesignSame as the predicateA handle incorporating a sample-loadingtip, and a cap (straw) to close the device.
Film tip shapeCryotop®US-flash - FlatCryotop®US-scoop - U-shapeFlat
DimensionsCryotop®US-flash:• Total length - 116 mm• Film tip length - 18 mm• Handle length - 98 mm• Straw length - 90 mm• Straw diameter - 3 mmCryotop®US-scoop:• Total length - 113 mm• Film tip length - 15 mm• Handle length - 98 mm• Straw length - 90 mm• Straw diameter - 3 mm• Total length - 105 mm• Film tip length - 20 mm• Handle length - 85 mm• Straw length - 100 mm• Straw diameter - 2.9 mm
Mode of actionSame as the predicateVitrification method - The loaded, closeddevice is placed in liquid nitrogen
Cooling rateSame as the predicate3000 °C/min
Warming rateSame as the predicate44000 °C/min
MaterialsCryotop®US-flash:• Handle – acrylonitrile butadiene styrene• Fine tip – polyethylene terephthalate• Straw – polypropylene/stainless steelCryotop®US-scoop:• Handle and fine tip – polystyrene• Straw - polypropylene/stainless steel• Handle – acrylonitrile butadiene styreneFilm sheet – polyethylene terephthalate• Straw – polypropylene/stainless steel

8. Substantial Equivalence Discussion

{5}------------------------------------------------

The subject devices are indicated to contain, vitrify and maintain MII oocytes and all stages of embryo development (2 PN-blastocyst stage), whereas the predicate device is only indicated for cleavage and blastocyst stage embryos. Inclusion of MII oocytes and earlier stage embryos (2 PN) in the subject devices indications does not represent a new intended use because processing of the embryos or oocytes is applicable to same patient population with same clinical needs – infertility preservation. Therefore, the subject and predicate devices have the same intended use.

Cryotop®US-flash and the predicate device have technological characteristics including design, mode of action, materials and cooling/warming rates; however, they have differences in dimensions. Cryotop®USscoop and the predicate device have the same fundamental design, mode of action and cooling/warming rates; however, they have differences in dimensions and materials. The differences noted do not raise different questions of safety and effectiveness.

9. Summary of Non-Clinical Performance Data

Non-clinical performance testing was conducted to support substantial equivalence to the predicate device. Cryotop®US-flash and Cryotop®US-scoop passed all the tests shown below in accordance with internal requirements and applicable standards.

  • . Cooling Rate Testing - 3,000°C/min
  • Warming Rate Testing 44,000°C/min .
  • Dimensional Testing - Outer diameter and length per specifications
  • . Durability Testing: No burst or liquid nitrogen ingress into straw after 30 second immersion
  • Mechanical Tensile Testing – Tensile strength to withstand 5N
  • Bacterial Endotoxins – <0.5 EU/device per USP <85> and ANSI/AAMI ST72:2002/(R)2010
  • . Mouse Embryo Assay (MEA) – ≥80% of 1-cell embryos developed to blastocysts at 96 hours
  • Transportation simulation testing per ASTM D4169-09
  • Shelf-life Testing: ●
    • Package integrity testing:
      • Dye penetration test per ASTM F1929-15 O
      • O Seal strength test per ASTM F88/F88M-15
    • Dimensional analysis, mechanical, durability, endotoxin, and MEA testing in accordance with the methods and acceptance criteria mentioned above

In addition, information regarding sterilization validation per ISO 11137-1:2006 and ISO 11137-2:2013 provided in K153027 was leveraged to support substantial equivalence.

10. Clinical Performance Data

Clinical information provided in K171748 (Kitazato Corporation Vitrification Kit and Thawing Kit) that incorporates the predicate Cryotop®US device was leveraged in supporting the additional uses of the Cryotop®US-flash and Cryotop®US-scoop for oocyte an 2 PN embryo vitrification procedures.

{6}------------------------------------------------

11. Conclusion

The subject and predicate devices have intended use and comparable technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject devices are substantially equivalent to the predicate device.

§ 884.6160 Assisted reproduction labware.

(a)
Identification. Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.(b)
Classification. Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.