K Number
K181233
Device Name
Elecsys FT4 III
Manufacturer
Date Cleared
2018-09-18

(132 days)

Product Code
Regulation Number
862.1695
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Assay for the in vitro quantitative determination of free thyroxine in human serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid disease. The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 411 immunoassay analyzer.
Device Description
The Elecsys FT4 III immunoassay is used for the in vitro quantitative determination of free thyroxine in human serum and plasma. It is intended for use on the cobas e 411 immunoassay analyzer. The cobas e family of analyzers uses electrochemiluminescence immunoassay "ECLIA" technology. The reagent working solutions include: Rackpack (kit placed on analyzer) M: Streptavidin-coated microparticles. R1: Anti-T4-Ab~biotin. R2: Anti-T4-Ab~Ru(bpy).
More Information

No
The summary describes a standard immunoassay and analyzer, with no mention of AI or ML in the device description, intended use, or performance studies. The modifications mentioned are related to assay buffer composition for biotin resistance.

No

Explanation: This device is an in vitro diagnostic (IVD) assay used for quantitative determination of free thyroxine. It is used to aid in the diagnosis and treatment of thyroid disease, but it does not directly treat the disease, thus it is not a therapeutic device.

Yes

Explanation: The "Intended Use / Indications for Use" section explicitly states, "Measurements obtained by this device are used in the diagnosis and treatment of thyroid disease." This directly indicates its role as a diagnostic device.

No

The device is an immunoassay reagent kit intended for use on a specific hardware analyzer (cobas e 411). The description details the chemical components of the reagent and performance studies related to the reagent's stability and interaction with biological samples, not software functionality.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use/Indications for Use: The document explicitly states "Assay for the in vitro quantitative determination of free thyroxine in human serum and plasma." and "Measurements obtained by this device are used in the diagnosis and treatment of thyroid disease." The term "in vitro" is a key indicator of an IVD, meaning it's used to test samples taken from the body, not on the body itself. The purpose of diagnosis and treatment also aligns with the definition of an IVD.
  • Device Description: The description details the components of the immunoassay and its use on an analyzer, which are typical characteristics of IVD systems.
  • Performance Studies: The document describes various performance studies conducted on human serum and plasma samples (e.g., Precision, Endogenous Interferences, Sample Matrix Comparison, Method Comparison). These studies are required for IVDs to demonstrate their analytical performance.
  • Predicate Device: The mention of a "Predicate Device" (K131244; Elecsys FT4 II) is common in regulatory submissions for IVDs, indicating a comparison to a previously cleared device.

All these elements strongly support the classification of this device as an In Vitro Diagnostic.

N/A

Intended Use / Indications for Use

Assay for the in vitro quantitative determination of free thyroxine in human serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid disease.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 411 immunoassay analyzer.

Product codes

CEC

Device Description

The Elecsys FT4 III immunoassay is used for the in vitro quantitative determination of free thyroxine in human serum and plasma. It is intended for use on the cobas e 411 immunoassay analyzer. The cobas e family of analyzers uses electrochemiluminescence immunoassay "ECLIA" technology.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Non-clinical performance evaluation for Elecsys FT4 III executed with the study briefly summarized.

Precision: Precision was evaluated on a single cobas e 411 analyzer according to CLSI guideline EP05-A3 using one reagent lot. The protocol consisted of testing 2 replicates of each of two levels of control, PreciControl Universal and five human sera per run, 2 runs per day for 21 days. The samples were run in randomized order on the cobas e 411 analyzer. Serum samples used were native single donors, native serum pools, and spiked serum pools.

Human Anti-Mouse Antibodies (HAMA): Not Applicable (no mouse antibodies used)

Endogenous Interferences - Biotin: The purpose of this study was to evaluate endogenous substances for potential interference with the parameters measured with the Elecsys FT4 III on the cobas e 411 immunoassay analyzer using human serum samples (single donors, native as well as spiked). For each interfering substance three human serum samples containing low, mid and high concentrations of fT4 were tested in accordance with CLSI EP07-A2.

Sample Matrix Comparison: The effect on quantitation of analyte in the presence of anticoagulants with the Elecsys FT4 III immunoassay was determined by comparing values obtained from samples (native single donors and pools as well as spiked samples) drawn into Serum, K2-and K3-EDTA plasma primary tubes. 53 serum/plasma pairs per sample material were tested in singleton with one reagent lot on one cobas e 411 immunoassay analyzer.

Method Comparison to Predicate: A method comparison was performed using the Elecsys FT4 III assay (candidate device, Y) and the Elecsys FT4 II assay (predicate device, X) to assess the bias between the two assays. A total of 141 serum samples (138 native human serum samples and 3 spiked sample pools) were measured in singleton on the cobas e 411 analyzer in one run covering the entire measuring range. FT4 values ranged between 2.31 and 91.9 pmol/L for the Reference Method (X).

Reagent Stability: Four studies were executed with two studies completed:
Study 1: Reagent stability after first opening at 2-8°C (84 days). Tested on one cobas e 411 immunoassay analyzer. A fresh reagent rackpack was placed on the analyzer and calibrated. Reference values for the samples tested were determined. After initial measurement the kit was removed from the analyzer and kept at 2-8 °C for 36, 63 and 85 days. After 36, 63 and 85 days the kit was placed on the analyzer again, calibrated and the test samples were determined. Samples tested in duplicate include six human serum (HS) samples and two controls (PreciControl Universal). The human samples used were native single donor serum, native serum pools, and spiked serum pools.
Study 2: On board reagent stability (28 days). Tested on one cobas e 411 immunoassay analyzer. A fresh reagent rackpack was placed on the analyzer and calibrated. All samples were measured on day 0. On day 8, 15, 22 and 29, the same samples were measured with the same reagent kit kept at 20°C ± 3°C (on-board condition) using the calibration curves established on day 0, 8, 15, 22 and 29, respectively. Samples tested include five human serum (HS) samples and two controls (PreciControl Universal). Each sample was tested in two-fold determination. Human serum samples used were native single donor serum, native serum pools, and spiked serum pools.
Study 3: Alternate storage in the refrigerator / on the analyzer (56 days, not exceeding 120 hours onboard). Tested on one cobas e 411 immunoassay analyzer. A fresh reagent rackpack was placed on the analyzer and calibrated. All samples were measured on day 0. On day 24, 29, 36, 43, 50 and 57, the same samples were measured with the same reagent kit kept in the refrigerator at 2-8°C and at 20°C ± 3°C (on-board condition) for 57 days not exceeding 120 hours under on-board condition using the calibration curves established on day 24, 29, 36, 43, 50 and 57, respectively. Samples tested include five human serum (HS) samples and two controls (PreciControl Universal). Each sample was tested in two-fold determination. Human serum samples used were native single donor serum, native serum pools, and spiked serum pools.
Study 4: A real-time stability study is ongoing to support shelf-life stability claim. The Elecsys FT4 III reagent is stored at 2-8°C. The stored assay reagents are tested at time point T=0 (after manufacturing) and at specified intervals over the shelf life of the device up to the planned shelf life plus one month. Testing will be performed using three human serum samples (stored at -20°C). Data for the time-points at 0, 7, 10 and 13 months tested in duplicate are available. The average on-test recovery value will be calculated as absolute deviation / percent recovery compared to the reference value measured at T=0 for the human serum samples.

Calibration Stability: Two studies were completed:
Study 1: Lot calibration stability. Elecsys FT4 III was calibrated with a fresh reagent kit on day 0 using a cobas e 411 immunoassay analyzer. After 36 days (5 weeks) a new reagent kit of the same lot was used and recovery of samples was determined using the calibration curve of day 0. Five human serum (HS) samples and two controls (PreciControl Universal) were tested; each sample was tested with two-fold determination. Human serum samples used were single donor serum, native serum pools, and spiked serum pools.
Study 2: On-board calibration stability. Tested on one cobas e 411 immunoassay analyzer. A fresh reagent rackpack was placed on the analyzer and calibrated. All samples were measured on day 0. On day 8 the same samples were measured with the same reagent kit kept at 20°C ± 3°C (on-board condition) using the calibration curve established on day 0. Samples tested include five human serum (HS) samples and two controls (PreciControl Universal). Each sample was tested in two-fold determination. Human serum samples used were single donor serum, native serum pools, and spiked serum pools.

Biotin: The Elecsys FT4 II was cleared under K131244; Elecsys FT4 III is a modification of this assay with only minor changes in the assay buffer composition to increase the Biotin resistance. For this purpose the D-Biotin concentration in the R2 reagent buffer was reduced.

Key Metrics

Not Found

Predicate Device(s)

Elecsys FT4 II (K131244)

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 862.1695 Free thyroxine test system.

(a)
Identification. A free thyroxine test system is a device intended to measure free (not protein bound) thyroxine (thyroid hormone) in serum or plasma. Levels of free thyroxine in plasma are thought to reflect the amount of thyroxine hormone available to the cells and may therefore determine the clinical metabolic status of thyroxine. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.(b)
Classification. Class II.

0

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the text "ADMINISTRATION".

September 18, 2018

Roche Diagnostics Reanna Toney Principal, Regulatory Affairs 9115 Hague Road Indianapolis, IN 46250

Re: K181233

Trade/Device Name: Elecsvs FT4 III Regulation Number: 21 CFR 862.1695 Regulation Name: Free thyroxine test system Regulatory Class: Class II Product Code: CEC Dated: August 2, 2018 Received: August 3, 2018

Dear Reanna Toney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

1

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website

(http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K181233

Device Name Elecsys FT4 III

Indications for Use (Describe)

Assay for the in vitro quantitative determination of free thyroxine in human serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid disease.

The electrochemiluminescence immunoasay "ECLIA" is intended for use on the cobas e 411 immunoassay analyzer.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

K181233 Elecsys FT4 III 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

In accordance with 21 CFR 807.87. Roche Diagnostics hereby submits official notification as required by Section 510(k) of the Federal Food, Drug and Cosmetics Act of our intention to market the device described in this Premarket Notification 510(k).

The purpose of this Traditional 510(k) Premarket Notification is to obtain FDA review and clearance for the Elecsys FT4 III.

4

Submitter NameRoche Diagnostics
Address9115 Hague Road
P.O. Box 50416
Indianapolis, IN 46250-0457
ContactReanna Toney
Phone: (317) 521-4516
FAX: (317) 521-2324
Email: reanna.toney@roche.com

Secondary Contact Name
Michael Leuther
Phone: (317) 521-3930
FAX: (317) 521-2324
Email: Michael.leuther@roche.com |
| Date Prepared | September 18, 2018 |
| Proprietary Name | Elecsys FT4 III |
| Common Name | Free thyroxine |
| Classification Name | Radioimmunoassay, Free thyroxine test system |
| Product Codes,
Regulation Numbers | CEC, 21CFR862.1695 |
| Predicate Devices | Elecsys FT4 II (K131244) |
| Establishment Registration | For the Elecsys FT4 III, the establishment registration number for Roche
Diagnostics GmbH in Mannheim, Germany is 9610126, and for Penzberg,
Germany, 9610529. The establishment registration number for Roche
Diagnostics in the United States is 1823260 |

5

1. DEVICE DESCRIPTION

The Elecsys FT4 III immunoassay is used for the in vitro quantitative determination of free thyroxine in human serum and plasma. It is intended for use on the cobas e 411 immunoassay analyzer. The cobas e family of analyzers uses electrochemiluminescence immunoassay "ECLIA" technology.

1.1. Reagents

The reagent working solutions include:

Rackpack (kit placed on analyzer)

  • M: Streptavidin-coated microparticles .
  • R1: Anti-T4-Ab~biotin .
  • R2: Anti-T4-Ab~Ru(bpy) ​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​ .

INDICATIONS FOR USE 2.

Assay for the in vitro quantitative determination of free thyroxine in human serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid disease.

The electrochemiluminescence immunoassay "ECLIA" is intended for use on the cobas e 411 immunoassay analyzer.

3. TECHNOLOGICAL CHARACTERISTICS

The following tables compare the Elecsys FT4 III with its predicate device, Elecsys FT4 II assay (K131244).

6

| Feature | Predicate Device:
Elecsys FT4 II (K131244) | Candidate Device:
Elecsys FT4 III |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/
Indications for
Use | Assay for the in vitro quantitative
determination of free thyroxine in human
serum and plasma. Measurements
obtained by this device are used in the
diagnosis and treatment of thyroid
disease.
The electrochemiluminescence
immunoassay "ECLIA” is intended for
use on Elecsys and cobas e immunoassay
analyzers. | Assay for the in vitro quantitative determination of
free thyroxine in human serum and plasma.
Measurements obtained by this device are used in
the diagnosis and treatment of thyroid disease.
The electrochemiluminescence immunoassay
"ECLIA" is intended for use on the cobas e 411
immunoassay analyzer. |
| Detection
Method | Electrochemiluminescent Assay | same |
| Applications/Test
Time | 18 minute application | same |
| Instrument
Platform | Elecsys and cobas e | cobas e 411 |
| Sample/Reagent
Ratio | 15 µL | same |
| Sample
Type/Matrix | Serum and plasma (Li-heparin, K2-EDTA,
K3-EDTA | same |
| Reagents | M: Streptavidin-coated microparticles
(transparent cap), 1 bottle, 12 mL:
Streptavidin-coated microparticles
0.72 mg/mL; preservative
R1: Anti T4-AbRu(bpy) (gray cap), 1
bottle, 18 mL: Polyclonal anti-T4-
antibody (sheep) labeled with ruthenium
complex 75 ng/mL; phosphate buffer 100
mmol/L, pH 7.0; preservative
R2: T4
biotin (black cap), 1 bottle, 18
mL: Biotinylated T4 2.5 ng/mL;
phosphate buffer 100 mmol/L, pH 7.0;
preservative. | M: Streptavidin-coated microparticles (transparent
cap), 1 bottle, 12 mL: Streptavidin-coated
microparticles 0.72 mg/mL; preservative
R1: Anti T4-AbRu(bpy) (gray cap), 1 bottle, 18
mL: Polyclonal anti-T4-antibody (sheep) labeled
with ruthenium complex 75 ng/mL; phosphate
buffer 100 mmol/L, pH 7.0; preservative
R2: T4
biotin (black cap), 1 bottle, 18 mL:
Biotinylated T4 2.5 ng/mL; free D-biotin 8 ng/mL;
phosphate buffer 100 mmol/L, pH 7.0; preservative. |
| Biotin | D-Biotin concentration in the R2 reagent
buffer: 50 ng/mL | D-Biotin concentration in the R2 reagent buffer:
8 ng/mL |
| Calibrator | FT4 II CalSet | CalSet FT4 III |
| Calibration
Method | 2-point calibration | same |
| Feature | Predicate Device:
Elecsys FT4 II (K131244) | Candidate Device:
Elecsys FT4 III |
| Calibration
Interval | Calibration must be performed once per
reagent lot using fresh reagent (i.e. not
more than 24 hours since the reagent kit
was registered on the analyzer).
Calibration interval may be extended
based on acceptable verification of
calibration by the laboratory.
Renewed calibration is recommended as
follows:
after 1 month (28 days) when
using the same reagent lot
after 7 days (when using the same
reagent kit on the analyzer)
as required: e.g. quality control
findings outside the defined limits | same |
| Controls | PreciControl Universal | same |
| Traceability/
Standardization | This method has been standardized
against the Elecsys FT4 method. The
Elecsys FT4 assay is traceable to the
Enzymun-Test which was standardized
using equilibrium dialysis. | This method has been standardized against the
Elecsys FT4 II method. The Elecsys FT4 II assay is
traceable to the Enzymun Test which was
standardized using equilibrium dialysis |
| Reagent Stability | Store at 2 8 °C.
Do not freeze.
Store the Elecsys reagent kit upright in
order to ensure complete availability of
the microparticles during automatic
mixing prior to use.
Unopened at 2-8 °C. up to stated
expiration date
After opening at 2-8 ℃. 84 days (12
weeks)
On the analyzers. 28 days (4 weeks)
onboard or 56 days (8 weeks) when stored
alternatively in the refrigerator and on the
analyzer, with the total time onboard on
the analyzer not exceeding 120 hours | same |
| Measuring Range | 0.101-7.77 ng/dL
(1.3-100 pmol/L | same |
| LoB | 0.03 ng/dL (0.4 pmol/L) | 0.02 ng/dL (0.3 pmol/L) |
| LoD | 0.05 ng/dL (0.6 pmol/L) | 0.04 ng/dL (0.5 pmol/L) |
| LoO | 0.101 ng/dL (1.3 pmol/L) | same |

Table 1: Assay Comparison

7

8

| Feature | Predicate Device:
Elecsys FT4 II (K131244) | | | Candidate Device:
Elecsys FT4 III | | |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|---------------------------|
| Analytical
Specificity | The following cross-reactivities were
found, tested with fT4 concentrations of
approximately 0.974 ng/dL (12.5 pmol/L)
and 2.66 ng/dL (34.2 pmol/L): | | | The following cross-reactivities were found, tested
with fT4 concentrations of approximately
1.01 ng/dL (13 pmol/L) and 3.03 ng/dL
(39 pmol/L): | | |
| | Cross-
reactant | Concentrat
ion tested
ng/dL | Cross-
reactivity
% | Cross-
reactant | Concentration
tested
ng/dL | Cross-
reactivity
% |
| | L-T3 | 50000 | 0.005 | L-T3 | 50000 | 0.005 |
| | D-T3 | 50000 | 0.002 | D-T3 | 50000 | 0.002 |
| | rT3 | 190000 | 0.007 | rT3 | 190000 | 0.007 |
| | 3-iodo-L-
tyrosine | 10000000 | 0.000 | 3-iodo-L-
tyrosine | 10000000 | 0.000 |
| | 3,5-diiodo-L-
tyrosine | 10000000 | 0.000 | 3,5-diiodo-L-
tyrosine | 10000000 | 0.000 |
| | 3,3',5-
triiodothyra
cetic acid | 100000 | 0.000 | 3,3',5-
triiodothyra
cetic acid | 100000 | 0.000 |
| | 3,3',5,5'-
tetraiodothyroaceti
c acid | 100000 | 0.001 | 3,3',5,5'-
tetraiodothyroacetic
acid | 100000 | 0.001 |

9

| Feature | Predicate Device:
Elecsys FT4 II (K131244) | Candidate Device:
Elecsys FT4 III |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Limitations | The assay is unaffected by icterus
(bilirubin 5 mg/day) until at least 8 hours
following the last biotin administration.
No interference was observed from
rheumatoid factors up to a concentration
of 1200 IU/mL and samples from dialysis
patients.
Any influence that might affect the
binding behavior of the binding proteins
can alter the result of the fT4 tests (e.g.
drugs, NTIs (Non Thyroid Illness) or
patients suffering from FDH (Familial
Dysalbuminemic Hyperthyroxinemia) or
increased TBG in pregnancy).
The test cannot be used in patients
receiving treatment with lipid-lowering
agents containing D T4. If the thyroid
function is to be checked in such patients,
the therapy should first be discontinued
for 4-6 weeks to allow the physiological
state to become re-established.
Autoantibodies to thyroid hormones can
interfere with the assay.
In vitro tests were performed on 17
commonly used pharmaceuticals. No
interference with the assay was found. | Bilirubin ≤ 701 µmol/L or ≤ 41 mg/dL
Hemoglobin ≤ 0.621 mmol/L or ≤ 1000 mg/dL
Intralipid ≤ 2000 mg/dL
Rheumatoid factors ≤ 1200 IU/mL
IgG ≤ 7 g/dL
IgA ≤ 1.6 g/dL
IgM ≤ 1 g/dL
Criterion: Recovery of ≤ ± 0.6 pmol/L of initial
value ≤ 6 pmol/L and ± 10 % of initial value > 6
pmol/L.
Any influence that might affect the binding behavior
of the binding proteins can alter the result of the fT4
tests (e.g. drugs, NTIs (Non Thyroid Illness) or
patients suffering from FDH (Familial
Dysalbuminemic Hyperthyroxinemia)).
The test cannot be used in patients receiving
treatment with lipid-lowering agents containing D
T4. If the thyroid function is to be checked in such
patients, the therapy should first be discontinued for
4-6 weeks to allow the physiological state to
become re-established
Autoantibodies to thyroid hormones can interfere
with the assay.
Pharmaceutical substances In vitro tests were
performed on 16 commonly used pharmaceuticals.
No interference with the assay was found. |
| Feature | Predicate Device:
Elecsys FT4 II (K131244) | Candidate Device:
Elecsys FT4 III |
| Special thyroid
drugs | The following special thyroid drugs were tested with concentrations shown in the table below. No interference with the assay was found.
Criterion: Recovery within ± 10 % of initial value. | In addition, the following special thyroid drugs were tested. No interference with the assay was found. |
| | Drug | Concentration
(µg/mL) |
| | Iodide | 0.200 |
| | Carbimazole | 6 |
| | Thiamazole | 80 |
| | Propylthiouracil | 300 |
| | Perchlorate | 2000 |
| | Propranolol | 240 |
| | Amiodarone | 200 |
| | Prednisolone | 100 |
| | Hydrocortisone | 200 |
| | Flurocortolone | 100 |
| | Octreotide | 0.300 |

10

11

| Feature | Predicate Device:
Elecsys FT4 II (K131244) | Candidate Device:
Elecsys FT4 III | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|---------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Biotin | Samples should not be taken from patients receiving therapy with high biotin doses (i.e. > 5 mg/day) until at least 8 hours following the last biotin administration. | Biotin interference
% Bias for samples containing various concentrations of Biotin | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | | Sample ng/dL (pmol/L) 52 65 78 91 104 117 150 300 600 1200 0.96 (12.4) 1.2 2.1 3.6 5.7 8.5 10.9 17.6 57.1 256 * 1.66 (21.4) -1.8 -1.4 1.3 3.0 7.3 7.6 16.1 53.7 291 * 2.61 (33.6) -0.5 0.6 2.3 4.7 7.5 9.7 15.9 56.5 * * * = Values above the measuring range. % Bias can not be calculated.
** Specimens with biotin concentrations up to 104 ng/mL demonstrated ≤ 10% bias in results. Biotin concentrations greater than 104 ng/mL can lead to higher positive bias for fT4 results.
Some studies have shown that serum concentrations of biotin can reach 355 ng/mL within the first hour after biotin ingestion for subjects consuming supplements of 20 mg biotin per day and up to 1160 ng/mL for subjects after a single dose of 300 mg biotin.
Do not test samples from patients who take biotin. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |

12

4. NON-CLINICAL PERFORMANCE EVALUATION

Non-clinical performance evaluation for Elecsys FT4 III executed with the study briefly summarized.

4.1. Precision

Repeatability and Intermediate Precision 4.1.1. -

Precision was evaluated on a single cobas e 411 analyzer according to CLSI guideline EP05-A3 using one reagent lot for evaluation. The protocol consisted of testing 2 replicates of each of two levels of control, PreciControl Universal and five human sera per run, 2 runs per day for 21 days. The samples were run in randomized order on the cobas e 411 analyzer. Serum samples used were native single donors, native serum pools, and spiked serum pools.

Human Anti-Mouse Antibodies (HAMA) 4.2.

Not Applicable (no mouse antibodies used)

Endogenous Interferences 4.3.

Biotin 4.3.1.

The purpose of this study was to evaluate endogenous substances for potential interference with the parameters measured with the Elecsys FT4 III on the cobas e 411 immunoassay analyzer using human serum samples (single donors, native as well as spiked).

For each interfering substance three human serum samples containing low, mid and high concentrations of fT4 were tested in accordance with CLSI EP07-A2.

4.4. Sample Matrix Comparison

The effect on quantitation of analyte in the presence of anticoagulants with the Elecsys FT4 III immunoassay was determined by comparing values obtained from samples (native single donors and pools as well as spiked samples) drawn into Serum, K2-and K3-EDTA plasma primary tubes. 53 serum/plasma pairs per sample material were tested in singleton with one reagent lot on one cobas e 411 immunoassay analyzer.

13

4.5. Method Comparison to Predicate

A method comparison was performed using the Elecsys FT4 III assay (candidate device, Y) and the Elecsys FT4 II assay (predicate device, X) to assess the bias between the two assays.

A total of 141 serum samples (138 native human serum samples and 3 spiked sample pools) were measured in singleton on the cobas e 411 analyzer in one run covering the entire measuring range. FT4 values ranged between 2.31 and 91.9 pmol/L for the Reference Method (X).

Reagent Stability 4.6.

To test reagent stability, four studies were executed with two studies completed including:

  • Study 1: Reagent stability after first opening at 2-8°C (84 days) .
  • Study 2: On board reagent stability (28 days) .
  • Study 3: Alternate storage in the refrigerator / on the analyzer (56 days, not exceeding . 120 hours onboard)
  • Study 4: A real-time stability study is ongoing to support shelf-life stability claim. •

Reagent Stability After First Opening 4.6.1.

Elecsys FT4 III reagent kits can be used after first opening for up to 84 days when stored at 2-8°C. Reagent stability after first opening for the Elecsys FT4 III assay was tested on one cobas e 411 immunoassay analyzer.

A fresh reagent rackpack was placed on the analyzer and calibrated. Reference values for the samples tested were determined. After initial measurement the kit was removed from the analyzer and kept at 2-8 °C for 36, 63 and 85 days. After 36, 63 and 85 days the kit was placed on the analyzer again, calibrated and the test samples were determined.

Samples tested in duplicate include six human serum (HS) samples and two controls (PreciControl Universal) for run qualification. The human samples used were native single donor serum, native serum pools, and spiked serum pools.

14

Reagent On-board stability (28 days) 4.6.2.

Elecsys FT4 III reagent kits can be stored on board of the analyzers for up to 28 days. A new calibration of the kit kept on-board is recommended every 7 days.

Reagent On-board Stability for the Elecsys FT4 III assay was tested on one cobas e 411 immunoassay analyzer.

A fresh reagent rackpack was placed on the analyzer and calibrated. All samples were measured on day 0. On day 8, 15, 22 and 29, the same samples were measured with the same reagent kit kept at 20°C ± 3°C (on-board condition) using the calibration curves established on day 0, 8, 15, 22 and 29, respectively.

Samples tested include five human serum (HS) samples and two controls (PreciControl Universal) for run qualification. Each sample was tested in two-fold determination. Human serum samples used were native single donor serum, native serum pools, and spiked serum pools.

Alternate storage in the refrigerator / on the analyzer (56 days, not exceeding 4.6.3. 120 hours onboard)

Elecsys FT4 III reagent kits can be stored in total for 56 days when stored alternatively in the refrigerator and on the analyzer, with a total time onboard the analyzer not exceeding 120 hours. Reagent refrigerator / on-board Stability for the Elecsys FT4 III assay was tested on one cobas e 411 immunoassay analyzer.

A fresh reagent rackpack was placed on the analyzer and calibrated. All samples were measured on day 0. On day 24, 29, 36, 43, 50 and 57, the same samples were measured with the same reagent kit kept in the refrigerator at 2-8°C and at 20°C ± 3°C (on-board condition) for 57 days not exceeding 120 hours under on-board condition using the calibration curves established on day 24, 29, 36, 43, 50 and 57, respectively.

Samples tested include five human serum (HS) samples and two controls (PreciControl Universal) for run qualification. Each sample was tested in two-fold determination. Human

15

serum samples used were native single donor serum, native serum pools, and spiked serum pools.

Reagent Real-time Stability 4.6.4.

In the ongoing real-time stability study, the Elecsys FT4 III reagent is stored at 2-8°C. The stored assay reagents are tested at time point T=0 (after manufacturing) and at specified intervals over the shelf life of the device up to the planned shelf life plus one month. Testing will be performed using three human serum samples (stored at -20°C).

Data for the time-points at 0, 7, 10 and 13 months tested in duplicate are available. The average on-test recovery value will be calculated as absolute deviation / percent recovery compared to the reference value measured at T=0 for the human serum samples.

4.7. Calibration Stability

To test calibration stability, two studies were completed, including:

  • Study 1. Lot calibration stability .
  • Study 2. On-board calibration stability .
  • Lot calibration study 4.7.1.

Calibration of an Elecsys FT4 III reagent lot is recommended every 28 days (4 weeks). During that time period fresh reagent kits of the same lot can be used without calibration using the calibration curve of the day 0 reagent kit.

Elecsys FT4 III was calibrated with a fresh reagent kit on day 0 using a cobas e 411 immunoassay analyzer. After 36 days (5 weeks) a new reagent kit of the same lot was used and recovery of samples was determined using the calibration curve of day 0.

Five human serum (HS) samples and two controls (PreciControl Universal) for run qualification were tested; each sample was tested with two-fold determination. Human serum samples used were single donor serum, native serum pools, and spiked serum pools.

16

On-board Calibration Stability 4.7.2.

Elecsys FT4 III reagent kits can be stored on board of the analyzers for up to 7 days without a new calibration.

Reagent On-board Calibration stability for the Elecsys FT4 III assay was tested on one cobas e 411 immunoassay analyzer.

A fresh reagent rackpack was placed on the analyzer and calibrated. All samples were measured on day 0. On day 8 the same samples were measured with the same reagent kit kept at 20°C ± 3°C (on-board condition) using the calibration curve established on day 0.

Samples tested include five human serum (HS) samples and two controls (PreciControl Universal) for run qualification. Each sample was tested in two-fold determination. Human serum samples used were single donor serum. native serum pools, and spiked serum pools.

Biotin 4.8.

The Elecsys FT4 II was cleared under K131244; Elecsys FT4 III is a modification of this assay with only minor changes in the assay buffer composition to increase the Biotin resistance. For this purpose the D-Biotin concentration in the R2 reagent buffer was reduced.

EXTERNAL (CLINICAL) TESTING 5.

Not Applicable

CLINICAL PERFORMANCE EVALUATION 6.

Not Applicable

ADDITIONAL INFORMATION 7.

The Elecsys FT4 III is intended to be used with the following calibrators and controls:

  • FT4 III CalSet .
  • Elecsys FT4 CalCheck •
  • PreciControl Universal .

17

FT4 III CalSet, product code JIS, is a Class II 510(k) Exempt device and therefore, is not included with this submission.

The Elecsys FT4 CalCheck, product code JJX, is a Class I 510(k) Exempt device and therefore, is not included with this submission.

PreciControl Universal, product code JJY, is a Class I 510(k) Exempt device and therefore, is not included with this submission.

8. CONCLUSIONS

The information provided in this 510(k) Premarket Notification will support a determination of substantial equivalence for the Elecsys FT4 III. The data supports a safe, effective device which performs as well as or better than the predicate device.