AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens is a daily wear soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or non-aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for Astigmatism is indicated daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with nondiseased eyes and whose powers are from -10.00 to +3.00 diopters and astigmatic corrections are from -0.25 to -5.00 diopters.

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for Presbyopia is indicated daily wear for the correction of ametropia (myopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -10.00 to +3.00 diopters with add powers from +0.25 to +3.50 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

Eye care practitioners may prescribe the lens for either single-use disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for single-use disposable wear, the lens is to be discarded after each removal, therefore no cleaning or disinfecting is required. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only.

Device Description

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens is in hemispherical flexible shells for myopia, hyperopia, astigmatisam and Presbyopia. The lens is made from HEMA containing UV blocker. The composition of the lens is 45% ocufilcon D and 55% water. The lens is tinted with "Reactive Blue 246" to make a light blue color for handling visibility purpose. A Benzophenone UV absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280-315nm and less than 50% in the UVA range of 316-380mm. It is supplied in a sterile package with buffered saline solution.

AI/ML Overview

This document is a 510(k) Pre-Market Notification from Largan Medical Co. Ltd. to the FDA for their Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lenses. It seeks to demonstrate substantial equivalence to a predicate device, the Biomedics 55 (ocufilcon D) Soft (hydrophilic) Contact Lens (K091339) manufactured by CooperVision, Inc.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a quantitative performance metric for the device itself (e.g., a specific visual acuity improvement percentage). Instead, the acceptance is based on demonstrating substantial equivalence to a legally marketed predicate device. This means the new device must be as safe and effective as the predicate.

The comparison table below outlines key characteristics of the new device and the predicate device, highlighting their similarities and differences. The implied acceptance criteria for most metrics are that the new device's specifications should be comparable or within acceptable ranges to the predicate.

ItemAcceptance Criteria (Implied by Predicate)Reported Device Performance (Largan 55 UV)
Product NameBiomedics Daily Wear Soft Contact LensLargan 55 UV daily wear soft contact lens
ManufacturerCooper Vision., Inc.Largan Medical Co. Ltd
Intended UseMyopia, Hyperopia, astigmatism, PresbyopiaMyopia, Hyperopia, astigmatism, Presbyopia
Lens DesignSpherical, aspherical, toric, multifocalaspherical, toric, or multifocal
Replacement ScheduleDaily WearDaily Wear
Chemical compositionOcufilcon DOcufilcon D
ClassificationGroup IV (ionic, High water)Group IV (ionic, High water)
Water Content55%55 %
Oxygen Permeability (DK, 35°C)19.619.6 (Fatt method)
Base Curve Range (mm)6.50 to 10.88.0~9.0
Diameter (mm)12.8 to 18.013.0~15.0
Center Thickness0.025-0.40 mm (Varies with power)Varies with design and power (0.084 mm at -3.00D)
Powers-20.00 to +20.00D-10.00D to +3.00D
Add Powers+0.25D to +3.00D+0.25E to +3.50D
Cylinder Power-0.25D to -10.00D-0.25D to -5.00D
Axis1° to 180°1° to 180°
Refractive Index1.411.405
Light Transmittance>95%90%
Blue handling tintEntrapment DyeReactive Blue 246
UV Block(Implied by predicate's safety)Less than 5% in UVB (280-315nm) and 50% in UVA (316-380nm)
Toxicology/BiocompatibilityNon-toxic and biocompatible in ocular environment (implied by predicate)Non-toxic and biocompatibility acceptable

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "no clinical data is required." This indicates that there was no test set (as in, a set of subjects on whom clinical performance was specifically tested for this 510(k)). The determination of substantial equivalence was based on a comparison of characteristics to the predicate device and non-clinical laboratory testing. Therefore, there is no sample size for a test set, nor data provenance in the context of a clinical study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Since no clinical test set was used, there were no experts used to establish ground truth in the context of clinical performance data for the Largan 55 UV contact lenses. The ground truth for the safety and effectiveness is largely established by the predicate device's prior approval and the general understanding of contact lens materials and designs.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

As there was no clinical test set, there was no adjudication method employed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device in question is a contact lens, not an AI-powered diagnostic tool. Therefore, no MRMC study or assessment of human reader improvement with/without AI assistance was performed or reported.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical contact lens, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" in this context is the established safety and effectiveness of the predicate device (Biomedics 55, K091339) and the scientific understanding of the materials (ocufilcon D) and design principles for contact lenses. The evaluation relies on:

  • Physical and chemical properties: Assessed according to ISO 18369.
  • Toxicology/Biocompatibility: Demonstrated through non-clinical studies.
  • Comparison of characteristics: To a legally marketed predicate device.

Essentially, the ground truth is that if the new device is sufficiently similar to an already approved, safe, and effective device in its critical characteristics, then it is also considered safe and effective.

8. The sample size for the training set

This is not applicable. The device is a contact lens and does not involve AI or machine learning models that require a training set.

9. How the ground truth for the training set was established

This is not applicable, as there is no training set for an AI/ML algorithm involved with this medical device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a stylized human figure on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in blue, with "FDA" in a larger font size than the rest of the text. The human figure is also in blue and appears to be a representation of health and well-being. The logo is commonly used to represent the FDA and its role in regulating food and drugs in the United States.

June 29, 2018

Largan Medical Co. Ltd. % Jennifer Ting Project Manager Jens Medical Consulting Ltd. 6F, No. 39, Ln 224, Jixian Rd., Luzhou Dist. 247 New Taipei City, Taiwan R.O.C.

Re: K181232

Trade/Device Name: Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens, Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for Astigmatism, Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for Presbyopia Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: May 7, 2018 Received: May 9, 2018

Dear Jennifer Ting:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

J Angelo Green -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181232

Device Name

Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for astigamtism Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for presbyopia

Indications for Use (Describe)

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens is a daily wear soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or non-aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for Astigmatism is indicated daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with nondiseased eyes and whose powers are from -10.00 to +3.00 diopters and astigmatic corrections are from -0.25 to -5.00 diopters.

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for Presbyopia is indicated daily wear for the correction of ametropia (myopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -10.00 to +3.00 diopters with add powers from +0.25 to +3.50 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

Eye care practitioners may prescribe the lens for either single-use disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for single-use disposable wear, the lens is to be discarded after each removal, therefore no cleaning or disinfecting is required. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K181232

510(k) Summary

Preparation Date: June 23, 2018

1.1 Establishment Information:

NameLargan Medical Co. Ltd.
OwnerAdam Lin
TitleCEO
Address2F., No. 14, 23rd Rd., Taichung Industrial Park, Nantun Dist., Taichung,40850, Taiwan
Phone No.886-4-3600-0203
Fax No.886-4-3601-0203
E-mailinfo@larganmed.com.tw

1.2 Contact Person:

Contact NameAmy Tien
Phone No.886-4-3600-0203
Fax No.886-4-3601-0203
E-mailamytien@larganmed.com.tw

1.3 Device Identification:

Proprietary NameLargan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic)Contact Lens
Common NameSoft (hydrophilic) Contact Lenses
Classification NameLenses, Soft Contact, Daily Wear,(21 CFR 886.5925, Product Code LPL)Lenses, Soft Contact, Daily Wear (Disposable),(21 CFR 886.5925, Product Code MVN)
ClassificationII
Regulation NumberCFR 886.5925
Review PanelOphthalmic
Product CodeLPL/MVN

1.4 Legally Marketed Equivalent Device:

Predicate Device NameBiomedics 55 (ocufilcon D) Soft (hydrophilic) Contact Lens
ManufacturerCooperVision, Inc.
510(k) NumberK091339
Product CodeLPL/MVN

1.5 Device Description

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens is in hemispherical flexible shells for myopia, hyperopia, astigmatisam and Presbyopia. The lens is made from HEMA containing UV blocker. The composition of the lens is 45% ocufilcon D and 55% water. The lens is tinted with "Reactive Blue 246" to make a light blue color for handling visibility purpose. A Benzophenone UV absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280-315nm and less than 50% in the UVA range of 316-380mm. It is supplied in a sterile package with buffered saline solution.

{4}------------------------------------------------

1.6 Indication for Use:

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens is a daily wear soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or non-aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for Astigmatism is indicated daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -10.00 to +3.00 diopters and astigmatic corrections are from -0.25 to -5.00 diopters.

The Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for Presbyopia is indicated daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -10.00 to +3.00 diopters with add powers from +0.25 to +3.50 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

Eye care practitioners may prescribe the lens for either single-use disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for single-use disposable wear, the lens is to be discarded after each removal, therefore no cleaning or disinfecting is required. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only.

1.7 Technological characteristic

Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens characteristics:

  • Diameter Range:13.0 to 15.0 mm .
  • . Base Curve:8.0 to 9.0 mm
  • . Center Thickness:0.084 mm for -3.00D (varies with power)
  • Power:+3.00 to -10.00 D .

Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for Astigmatism characteristics:

  • Diameter Range:13.0 to 15.0 mm •
  • . Base Curve:8.0 to 9.0 mm
  • . Center Thickness:0.084 mm for -3.00D (varies with power)
  • . Power:+3.00 to -10.00 D
  • . Cylinder: -0.25D ~ -5.00 D
  • Axis: 10° to 180° (in 10° increments) .

Largan 55 UV (ocufilcon D) Daily Wear Soft (hydrophilic) Contact Lens for Presbyopia characteristics:

  • Diameter Range: 13.0 to 15.0 mm ●
  • Base Curve:8.0 to 9.0 mm
  • Center Thickness:0.130 mm for -3.00D (varies with power)
  • . Power:+3.00 to -10.00 D
  • . Additional Powers: +0.25D ~ +3.50D

{5}------------------------------------------------

1.8 Comparison table:

Similarities and differences
ItemDevicePredicate (K091339)
Product NameLargan 55 UV daily wear softcontact lensBiomedics Daily Wear SoftContact Lens
ManufacturerLargan Medical Co. LtdCooper Vision., Inc.
Intended UseMyopia, Hyperopia,astigmatism, Presbyopiamyopia, hyperopia,astigmatism, Presbyopia
Lens Designaspherical, toric, or multifocalSpherical, aspherical, toric,multifocal
Replacement ScheduleDaily WearThe same
Chemical compositionOcufilcon DThe same
ClassificationGroup IV (ionic, High water)The same
Water Content55 %55%
Oxygen Permeability(DK, 35°C )19.6(Fatt method)19.6
Base Curve Range(mm)8.0~9.06.50 to 10.8
Diameter (mm)13.0~15.012.8 to 18.0
Center ThicknessVaries with design and power(0.084 mm at -3.00D)0.025-0.40 mm (Varies withpower)
Powers-10.00D to +3.00D-20.00 to +20.00D
Add Powers+0.25E to +3.50D+0.25D to +3.00D
Cylinder Power-0.25D to -5.00D-0.25D to -10.00D
Axis1° to 180°1° to 180°
Refractive Index1.4051.41
Light Transmittance90%>95%
Blue handling tintReactive Blue 246Entrapment Dye

The characteristic comparison to predicate device is summarized in the following table.

1.9 Nonclinical Tests Performed

  • Physiochemical studies were conducted according to ISO 18369 First edition 1.9.1 2006-08-15, Ophthalmic optics - Contact lenses (Ophthalmic). The physical, optical and chemical properties of the lens are within established specifications for the lenses.
  • 1.9.2 Toxicology studies report shows that the lenses are non-toxic and biocompatibility result is acceptable in ocular environment.

1.10 Clinical Studies

The technical characteristics and formulation of the subject device are equivalent to Biomedics 55 Daily Wear Contact Lens (K091339) current marketed by CooperVision Inc., therefore no clinical data is required.

1.11 Conclusion

Comparison to the predicate device for chemical composition, physical and optical properties, it shows that "Largan 55 UV daily wear soft (hydrophilic) contact lens" is as safe, as effective and performs as well as the predicate device.

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.