(269 days)
The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10 ng/mL in neat oral fluid collected with the Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of PCP in human oral fluid with clinical analyzers. This assay is calibrated against PCP. This in vitro diagnostic device is for prescription use only.
The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.
The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.
The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay is a sensitive in vitro diagnostic test to detect the presence of PCP in human oral fluid samples collected with the Quantisal II Oral Fluid Collection Device.
Quantisal II Oral Fluid Collection Device is a collection system comprised of a dual pad collector and transport vials. The dual pad collector is separated after collection of oral fluid from a subject's mouth enabling each specimen-saturated collection pad to be placed into its own transport vial. The split specimen (referred to as "A" and "B") allows for one sample to be tested in a screening assay and confirmed by a quantitative laboratory method (such as liquid chromatography tandem mass spectrometry [LC-MS/MS] and the second sample to be stored for secondary confirmation if needed.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Device Name: Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" as a separate list with pass/fail values. Instead, it describes various performance studies and their results. The implicit acceptance criteria are that the device performs reliably and consistently, and that its results correlate well with a confirmatory method (LC-MS/MS).
However, based on the provided tables, we can infer some key performance metrics and their results:
| Criteria/Performance Metric | Description | Reported Device Performance |
|---|---|---|
| Precision | Consistency and repeatability of qualitative and semi-quantitative results across multiple determinations at various concentrations relative to the cutoff (10 ng/mL). | Qualitative (Quantisal II "A" & "B"):- 0, 2.5, 5, 7.5 ng/mL (negative concentrations): 60/60 Negative.- 12.5, 15, 17.5, 20 ng/mL (positive concentrations): 60/60 Positive.- 10 ng/mL (Cutoff): 29-32 Negative / 28-31 Positive (expected result at cutoff).Semi-Quantitative (Quantisal II "A" & "B"):- Mean concentrations for spiked samples are close to the expected values (e.g., 0.5 ng/mL for 0 ng/mL, 20.2 ng/mL for 20 ng/mL), demonstrating accurate semi-quantification. At the cutoff (10 ng/mL), results show an appropriate split between negative and positive calls, consistent with the nature of a cutoff analysis. |
| Specificity/Cross-Reactivity | Ability to exclusively determine PCP without significant interference from structurally and functionally similar compounds. Compounds spiked to yield PCP equivalent of 10 ng/mL. | Cross-Reactive (Positive at 10 ng/mL equivalence): Amitriptyline, Chlorpromazine, Clomipramine, Cyclobenzaprine, Diphenhydramine, Doxepin, 4-Hydroxyphencyclidine (PCHP - 11.76% cross-reactivity), Imipramine, Methoxetamine, Thioridazine.Non-Cross-Reactive (Negative at <10 ng/mL equivalence): Desipramine, Dextromethorphan, Doxylamine, EDDP, Nortriptyline, Protriptyline, Trimipramine, Venlafaxine.(Note: Cross-reactivity percentages indicate the concentration of the interfering compound needed to produce a response equivalent to 10 ng/mL of PCP) |
| Interference – Structurally Unrelated Compounds | Assay performance unaffected by external ingested compounds. Evaluated by spiking high concentrations of various unrelated compounds into drug-free oral fluid containing PCP at ±25% of the cutoff. | Over 70 structurally unrelated compounds (e.g., 6-Acetylcodeine, Alprazolam, Amphetamine, Caffeine, Cocaine, Fentanyl, Ibuprofen, Morphine, Zoldipem Tartrate) tested at high concentrations (e.g., 40,000 ng/mL) were found to be non-interfering in both qualitative and semi-quantitative modes. |
| Interference – Endogenous & Exogenous Compounds | Assay performance unaffected by naturally occurring bodily compounds and common exogenous substances (e.g., food, drinks, hygiene products). Evaluated by spiking or collecting samples from volunteers after use. | Non-interfering Endogenous Compounds: Ascorbic Acid, Bilirubin, Cholesterol, γ-Globulin, Hemoglobin, Human Serum Albumin, IgA, IgG, IgM, Salivary-α-amylase at relevant concentrations.Non-interfering Exogenous Compounds: Acetaminophen, Acetylsalicylic Acid, Baking Soda, Cotinine, Denture Adhesive, Ibuprofen, Alcohol, Caffeine, Coffee, Cranberry Juice, Hydrogen Peroxide, Milk, Mouthwash, Naproxen, Orange Juice, Soft Drink (Pepsi), Sodium Chloride, Sugar, Tea, Toothpaste at relevant concentrations.Non-interfering Orally Used Exogenous Products (tested via volunteer use): Teeth Whitener, Hydrogen Peroxide, Cigarette, Hard Candy, Chewing Gum, Cough Syrup. |
| Interference – pH | Assay performance unaffected by a range of oral fluid pH values. Evaluated by testing samples with PCP at ±25% of cutoff across pH 3.0-11.0. | No positive or negative interference observed at oral fluid pH values ranging from 3.0 to 11.0 for each test mode. |
| Linearity/Recovery | The semi-quantitative mode provides accurate recovery of spiked PCP concentrations within a defined range. | For both Quantisal II "A" and "B" collectors, recovery percentages were generally within 92.5% to 105.8% across an expected concentration range of 0 to 44 ng/mL, indicating good linearity and recovery. |
| Calibration Duration | The device maintains performance over a recommended calibration period. | Test results met acceptance criteria at time points up to 14 days, with calibration curves established at the initial time point remaining valid. Recommended frequency of calibration is 14 days. |
| PCP Stability in Oral Fluid | PCP in collected oral fluid samples remains stable under various storage conditions for specific durations. | Oral fluid samples containing PCP are stable for up to 10 days stored in Quantisal II transport tubes at ambient temperature (up to 30°C) and up to 2 months stored at 2°C - 8°C. |
| Sample Transportation Stability | PCP concentration in samples remains stable during simulated transport conditions. | PCP concentration in samples collected by Quantisal II Oral Fluid Collection Device remained within 20% of the initial value during a 4-day (96 hours) simulated transportation study with temperatures ranging from -20°C to 40°C. |
| Sample Recovery | The Quantisal II Collection Device efficiently recovers PCP from oral fluid. | The Quantisal II Collection Device recovers PCP at greater than 80% of the original concentration. |
| Quantisal II Sample Volume Consistency | The collection device consistently collects the intended volume of oral fluid. | Results confirmed consistent sample volume of 1 mL collected by each of Quantisal II collectors from both healthy volunteers (N=50) and known drug users (N=75). |
| Quantisal II Sample Collection Time | The collection device enables sample collection within a reasonable and claimed timeframe. | The sample collection time for Quantisal II Oral Fluid Collection Device is within the claimed time of 10 minutes in over 90% of subjects (tested with 50 volunteers and 75 known drug users). |
| Method Comparison (Agreement with LC-MS/MS) | The device's results (qualitative and semi-quantitative) correlate highly with a gold-standard confirmatory method (LC-MS/MS) for PCP detection. | Qualitative and Semi-Quantitative (Quantisal II "A" & "B"): 100% agreement (40/40) for both Positive and Negative call groups when compared against LC-MS/MS results for PCP concentrations across various ranges (<5 ng/mL, 5-9 ng/mL, 10-15 ng/mL, >15 ng/mL). This indicates excellent diagnostic agreement. |
2. Sample Size Used for the Test Set and Data Provenance
The document describes several test sets for different performance characteristics:
- Precision/Cutoff Characterization:
- Sample Size: 60 determinations at each of 9 concentration levels (0, 2.5, 5, 7.5, 10, 12.5, 15, 17.5, 20 ng/mL) for both "A" and "B" collectors. This totals 60 x 9 x 2 = 1080 individual tests across 15 days, two runs per day, with two collections per run.
- Data Provenance: Drug-free negative urine spiked to specific concentrations. The origin of the urine itself is not specified but it's an artificially prepared test set.
- Specificity and Cross-Reactivity:
- Sample Size: Not explicitly stated as a number of individual test runs per compound, but implies sufficient testing to determine cross-reactivity for 18 listed compounds.
- Data Provenance: Drug-free oral fluid spiked with various compounds. Origin of the oral fluid is not specified, but it's an artificially prepared test set.
- Interference – Structurally Unrelated Compounds:
- Sample Size: Not explicitly stated, but includes testing for over 70 compounds.
- Data Provenance: Drug-free oral fluid containing PCP at ±25% of the cutoff, spiked with potential interferents. Origin of the oral fluid is not specified, but it's an artificially prepared test set.
- Interference – Endogenous Compounds and Exogenous Compounds:
- Sample Size: Not explicitly stated for spiked compounds. For orally used products, likely tested with volunteers (number not specified for this specific section, but volunteer numbers are mentioned in other collection device studies).
- Data Provenance: Drug-free oral fluid containing PCP at ±25% of the cutoff (for spiked compounds) or collected from volunteers after substance use.
- Interference – pH:
- Sample Size: Not explicitly stated, but tested across 9 pH values (3.0 to 11.0).
- Data Provenance: Drug-free oral fluid containing PCP at ±25% of the cutoff, adjusted to various pH values.
- Linearity/Recovery:
- Sample Size: 13 concentration levels, each tested in triplicate for both "A" and "B" collectors. This totals 13 x 3 x 2 = 78 individual tests.
- Data Provenance: Drug-free oral fluid pooled and spiked with high concentrations of PCP, then serially diluted.
- Calibration Duration:
- Sample Size: Not explicitly stated, but tested at multiple time points up to 14 days.
- Data Provenance: Drug-free negative oral fluid spiked with PCP at ±25% of the cutoff.
- PCP Stability in Oral Fluid:
- Sample Size: Not explicitly stated, but tested by LC-MS/MS at multiple time points and storage conditions.
- Data Provenance: Drug-free negative oral fluid spiked with PCP at +50% of the cutoff.
- Sample Transportation Stability:
- Sample Size: Not explicitly stated, but tested in replicates of two against a reference sample across varying temperatures.
- Data Provenance: Drug-free negative oral fluid spiked with PCP at ±50% of the cutoff.
- Sample Recovery:
- Sample Size: Not explicitly stated.
- Data Provenance: Drug-free negative oral fluid spiked with PCP at ±25% and +50% of the cutoff.
- Quantisal II Sample Volume:
- Sample Size: 50 oral fluid samples from healthy volunteers and an additional 75 oral fluid samples from known drug users. Total = 125 unique samples.
- Data Provenance: Prospective collection from healthy volunteers and known drug users. Country of origin not specified.
- Quantisal II Sample Collection Time:
- Sample Size: 50 oral fluid samples from volunteers and 75 oral fluid samples from known drug users. Total = 125 unique samples.
- Data Provenance: Prospective collection from healthy volunteers and known drug users. Country of origin not specified.
- Method Comparison:
- Sample Size: 80 de-identified, unaltered clinical oral fluid samples.
- Data Provenance: Retrospective and prospective. "Obtained from drug treatment facilities." Country of origin not specified, but likely within the US, given the FDA submission. The samples are clinical, not contrived.
3. Number of Experts and Qualifications for Ground Truth
- The document does not mention the use of human experts to establish ground truth for any of the test sets in the traditional sense of medical image interpretation or clinical diagnosis.
- For chemical assays, "ground truth" is typically established by reference methods or by precisely known concentrations of analytes.
4. Adjudication Method for the Test Set
- Adjudication methods like 2+1 or 3+1 (common in medical image reading studies) are not applicable here as the device is a chemical immunoassay, not an AI for image interpretation.
- The ground truth for most analytical performance studies (precision, specificity, linearity, etc.) was established by known spiked concentrations of PCP or other compounds.
- For the Method Comparison study, the ground truth was established by a confirmatory method: Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS), which is considered the gold standard for drug quantification. There was no "adjudication" between multiple experts; rather, the device's results were compared against the definitive LC-MS/MS measurements.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., radiologists reading images) and the impact of AI assistance on their performance.
- This device is an automated in vitro diagnostic immunoassay for chemical analysis, not a system intended to assist human readers in making a diagnosis from complex data patterns that require human cognitive input.
6. Standalone Performance
- Yes, the studies primarily assessed standalone (algorithm only) performance. The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay is an automated system run on clinical analyzers (specifically, the Beckman Coulter AU480 chemistry analyzer was used for these studies).
- All the precision, specificity, linearity, stability, and pH interference studies directly demonstrate the standalone performance of the assay and collection device combination.
- The "Method Comparison" study compares the device's standalone output to the LC-MS/MS reference, further confirming its standalone accuracy.
7. Type of Ground Truth Used
The ground truth varied depending on the performance characteristic being evaluated:
- Spiked Concentrations: For precision, specificity, interference, linearity, calibration duration, stability, transportation stability, and sample recovery, the ground truth was based on precisely known concentrations of PCP or other compounds spiked into drug-free oral fluid or urine.
- Confirmatory Method (LC-MS/MS): For the Method Comparison study, the ground truth for clinical samples was established by Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS), which is the preferred confirmatory method for drug analyses.
- Measured Physical Properties: For sample volume and collection time consistency, the ground truth was established by physical measurements (weighing for volume, stopwatch for time).
8. Sample Size for the Training Set
The document is a 510(k) submission for a diagnostic kit, not an AI/ML model that requires a "training set" in the computational sense. Therefore, there is no specific training set described for the device. The "training" for such devices typically involves the manufacturer's internal development and optimization processes, which are not detailed in a 510(k) summary.
9. How the Ground Truth for the Training Set was Established
As there is no "training set" for an AI/ML model, this question is not applicable. The development of the assay and its reagents would have relied on standard chemical and biological research and development practices, using analytical standards and reference materials to establish optimal performance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 24, 2019
Immunalysis Corporation Michelle Bodien Associate Director, Regulatory Affairs 829 Towne Center Drive Pomona, CA 91767
Re: K181135
Trade/Device Name: Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay Regulatory Class: Unclassified, 510(k) required Product Code: LCM Dated: December 13, 2018 Received: December 14, 2018
Dear Michelle Bodien:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kellie B. Kelm -S
for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K181135
Device Name
Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay
Indications for Use (Describe)
The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 10 ng/mL in neat oral fluid collected with the Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of PCP in human oral fluid with clinical analyzers. This assay is calibrated against PCP. This in vitro diagnostic device is for prescription use only.
The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.
The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image. The font is sans-serif and the letters are bolded.
510(k) SUMMARY K181135
A. GENERAL INFORMATION
| Applicant Name: | Immunalysis Corporation829 Towne Center DrivePomona, CA 91767Establishment # 2020952 |
|---|---|
| Company Contact: | Michelle BodienAssoc. Director, Regulatory AffairsPhone: (210)4058453Email: michelle.bodien@alere.com |
| Date Prepared: | January 22, 2019 |
B. DEVICE IDENTIFICATION
| Trade or Proprietary Names: | Immunalysis SEFRIA™ PCP Oral Fluid Enzyme Immunoassay |
|---|---|
| ----------------------------- | ------------------------------------------------------- |
C. REGULATORY INFORMATION
| Device Classification Name: | Enzyme Immunoassay, Phencyclidine |
|---|---|
| Product Codes: | LCM |
| Regulatory Class: | Class II |
| Classification Regulation: | Unclassified |
| Panel: | Toxicology (91) |
| Predicate Device: | RapidFRET Oral Fluid Assay for PCP, PCP Calibrator Set, PCP Control Set and RapidEase Oral Fluid Collector [K122703] |
D. DEVICE DESCRIPTION
The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay is a sensitive in vitro diagnostic test to detect the presence of PCP in human oral fluid samples collected with the Quantisal II Oral Fluid Collection Device.
Quantisal II Oral Fluid Collection Device is a collection system comprised of a dual pad collector and transport vials. The dual pad collector is separated after collection of oral fluid from a subject's mouth enabling each specimen-saturated collection pad to be placed into its own transport vial. The split specimen (referred to as "A" and "B") allows for one sample to be tested in a screening assay and confirmed by a quantitative laboratory method (such as liquid chromatography tandem mass spectrometry [LC-MS/MS] and the second sample to be stored for secondary confirmation if needed.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "IMMUNALYSIS" in white letters against a red background. The letters are bold and sans-serif. The background is a solid red color, and the word is centered in the image.
PCP was first synthesized in 1926 and later tested after World War II as a surgical anesthetic. Because of its adverse side effects, such as hallucinations, mania, delirium, and disorientation, it was shelved until the 1950s. The drug is easily synthesized by anyone with a basic knowledge of chemistry and has become one of the drugs most frequently used by drug abusers. It has a variety of street names, including "angel dust," "animal tranquilizer," "PCP," "peace pill," "crystal joints," and "peace weed," with the name often reflecting the form in which it is taken. It can be smoked, "snorted" through the nose, ingested, or taken intravenously. Phencyclidine has also been shown to cause schizophrenia-like changes in Nacetylaspartate and Nacetylaspartylglutamate in the rat brain, which are detectable both in living rats and upon necropsy examination of brain tissue. It also induces symptoms in humans that mimic schizophrenia. Behavioral effects can vary by dosage. Low doses produce numbness in the extremities and intoxication, characterized by staggering, unsteady gait, slurred speech, bloodshot eyes, and loss of balance. Moderate doses (5-10 mg intranasal, or 0.01-0.02 mg/kg intramuscular or intravenous) will produce analgesia and anesthesia. High doses may lead to convulsions. Users frequently do not know how much of the drug they are taking due to the tendency of the drug to be made illegally in uncontrolled conditions.
E. INTENDED USE
Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay
The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay is a homogenous enzyme immunoassay with a cutoff of 10 ng/mL in neat oral fluid collected with the Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of PCP in human oral fluid with clinical analyzers. This assay is calibrated against PCP. This in vitro diagnostic device is for prescription use only.
The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.
The Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.
| Attribute | Predicate DeviceRapidFRET Oral Fluid Assayfor PCP [K122703] | Candidate DeviceImmunalysis SEFRIA PCPOral Fluid EnzymeImmunoassay |
|---|---|---|
| Similarities | ||
| Test Principle | Homogeneous enzymeimmunoassay | Same |
| Antibody | Antibodies to PCP | Same |
F. COMPARISON WITH PREDICATE
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a red arrow pointing to the right with the word "IMMUNALYS" written in white letters inside the arrow. The word is written in all capital letters and is centered within the arrow. The arrow is a bright red color, and the white letters stand out against the red background.
| Attribute | Predicate DeviceRapidFRET Oral Fluid Assayfor PCP [K122703] | Candidate DeviceImmunalysis SEFRIA PCPOral Fluid EnzymeImmunoassay |
|---|---|---|
| Materials | PCP specific antibody reagent,PCP drug conjugate reagent | Same |
| Cutoff Level | 10 ng/mL | Same |
| User Environment | For use in laboratories | Same |
| Sample Matrix | Human oral fluid | Same |
| Reagent Storage | 2-8°C until expiration date | Same |
| Differences | ||
| Intended Use | Qualitative determination ofphencyclidine in human oralfluid | Qualitative and semi-quantitative determination ofphencyclidine in human oralfluid. |
| Sample Collection Device | Neat oral fluid is collected withthe RapidEASE Oral FluidCollector via directexpectoration. No diluent is usedand sample is stored in a glasssample tube with inert screwcap. | Oral fluid is collected with theQuantisal II Oral FluidCollection Device. Sample isstored in a plastic tubecontaining preservative bufferwith snap cap. |
G. PERFORMANCE CHARACTERISTICS
The following laboratory performance studies were performed to determine substantial equivalence of the Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay and Quantisal II Oral Fluid Collection Devices to the predicate device. Assay performance was established using the Beckman Coulter AU480 chemistry analyzer.
1. Precision/ Cutoff Characterization
Precision/ Cutoff Characterization study was performed over 15 days, two collection device runs per day with two collections per run (N=60) on three lots of Quantisal II oral fluid collection devices. Drug free negative urine was spiked to concentrations of assay cutoff and ±25%, ±50%, ±75%, ±100% of the cutoff and transferred to the collection devices. The spiked concentrations were confirmed by mass spectrometry (LC-MS/MS). The study established the repeatability of the testing system, including assay and oral fluid collection device. Test results in qualitative and semi-quantitative modes for a representative lot are presented in Tables 1 to 6.
| Concentration(ng/mL) | % of cutoff | # ofdeterminations | Result |
|---|---|---|---|
| 0 | -100% | 60 | 60 Negative |
Table 1. Precision - Quantisal II "A" - Qualitative
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.
| Concentration(ng/mL) | % of cutoff | # ofdeterminations | Result |
|---|---|---|---|
| 2.5 | -75% | 60 | 60 Negative |
| 5 | -50% | 60 | 60 Negative |
| 7.5 | -25% | 60 | 60 Negative |
| 10 | Cutoff | 60 | 29 Neg /31 Pos |
| 12.5 | 25% | 60 | 60 Positive |
| 15 | 50% | 60 | 60 Positive |
| 17.5 | 75% | 60 | 60 Positive |
| 20 | 100% | 60 | 60 Positive |
Table 2. Precision – Quantisal II "A" - Semi-Quantitative
| Concentration(ng/mL) | % of cutoff | # ofdeterminations | Mean Conc.(ng/mL) | Result |
|---|---|---|---|---|
| 0 | -100% | 60 | 0.5 | 60 Negative |
| 2.5 | -75% | 60 | 2.7 | 60 Negative |
| 5 | -50% | 60 | 5.5 | 60 Negative |
| 7.5 | -25% | 60 | 7.8 | 60 Negative |
| 10 | Cutoff | 60 | 10.1 | 28 Neg / 32 Pos |
| 12.5 | 25% | 60 | 12.5 | 60 Positive |
| 15 | 50% | 60 | 15.5 | 60 Positive |
| 17.5 | 75% | 60 | 17.4 | 60 Positive |
| 20 | 100% | 60 | 20.2 | 60 Positive |
Table 3. Precision – Quantisal II "B" - Qualitative
| Concentration(ng/mL) | % of cutoff | # ofdeterminations | Result |
|---|---|---|---|
| 0 | -100% | 60 | 60 Negative |
| 2.5 | -75% | 60 | 60 Negative |
| 5 | -50% | 60 | 60 Negative |
| 7.5 | -25% | 60 | 60 Negative |
| 10 | Cutoff | 60 | 32 Neg /28 Pos |
| 12.5 | 25% | 60 | 60 Positive |
| 15 | 50% | 60 | 60 Positive |
| 17.5 | 75% | 60 | 60 Positive |
| 20 | 100% | 60 | 60 Positive |
Table 4. Precision – Quantisal II "B" - Semi-Quantitative
| Concentration(ng/mL) | % of cutoff | # ofdeterminations | Mean Conc.(ng/mL) | Result |
|---|---|---|---|---|
| 0 | -100% | 60 | 0.5 | 60 Negative |
| 2.5 | -75% | 60 | 2.7 | 60 Negative |
| 5 | -50% | 60 | 5.5 | 60 Negative |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.
| Concentration(ng/mL) | % of cutoff | # ofdeterminations | Mean Conc.(ng/mL) | Result |
|---|---|---|---|---|
| 7.5 | -25% | 60 | 8.0 | 60 Negative |
| 10 | Cutoff | 60 | 10.1 | 29 Neg / 31 Pos |
| 12.5 | 25% | 60 | 12.7 | 60 Positive |
| 15 | 50% | 60 | 15.5 | 60 Positive |
| 17.5 | 75% | 60 | 17.5 | 60 Positive |
| 20 | 100% | 60 | 20.2 | 60 Positive |
2. Specificity and Cross-Reactivity
Structurally and functionally similar compounds were spiked into drug free oral fluid at levels that will yield a result that is equivalent to the cutoff. The study verified assay performance relative to the ability of the device to exclusively determine certain drugs, in both the qualitative and semi-quantitative modes. Cross-reactivity test results for qualitative and semi-quantitative modes are presented in Table 7.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is in all capital letters and is centered in the image. The red background is a solid color, and the white letters are easy to read. The image is simple and straightforward.
| Compound | CompoundConc.(ng/mL) | PCPEquivalentConc.(ng/mL) | Result | Cross-Reactivity(%) |
|---|---|---|---|---|
| Amitriptyline | 22,000 | 10 | POS | 0.05 |
| Chlorpromazine | 5,200 | 10 | POS | 0.19 |
| Clomipramine | 22,000 | 10 | POS | 0.05 |
| Cyclobenzaprine | 1,900 | 10 | POS | 0.53 |
| Desipramine | 40,000 | <10 | NEG | <0.03 |
| Dextromethorphan | 40,000 | <10 | NEG | <0.03 |
| Diphenhydramine | 37,000 | 10 | POS | 0.03 |
| Doxepin | 5,600 | 10 | POS | 0.18 |
| Doxylamine | 40,000 | <10 | NEG | <0.03 |
| EDDP | 40,000 | <10 | NEG | <0.03 |
| 4-Hydroxyphencyclidine(PCHP) | 85 | 10 | POS | 11.76 |
| Imipramine | 13,400 | 10 | POS | 0.07 |
| Methoxetamine | 34,000 | 10 | POS | 0.03 |
| Nortriptyline | 40,000 | <10 | NEG | <0.03 |
| Protriptyline | 40,000 | <10 | NEG | <0.03 |
| Thioridazine | 8,600 | 10 | POS | 0.12 |
| Trimipramine | 40,000 | <10 | NEG | <0.03 |
| Venlafaxine | 40,000 | <10 | NEG | <0.03 |
Table 5. Cross-Reactivity - Qualitative
3. Interference – Structurally Unrelated Compounds
Structurally unrelated compounds were evaluated in qualitative and semi-quantitative modes by spiking the potential interferent into drug free oral fluid containing PCP at ±25% of the cutoff. All potential interferents analyzed verified that assay performance is unaffected by externally ingested compounds. The levels of structurally unrelated compounds that did not interfere in the assay are presented in Table 8.
| Compound | Conc. Tested (ng/mL) |
|---|---|
| 4-Bromo-2,5,Dimethoxyphenethylamine | 5,000 |
| 6-Acetylcodeine | 40,000 |
| 6-Acetylmorphine | 40,000 |
| Alprazolam | 40,000 |
| 7-Aminoclonazepam | 40,000 |
| 7-Aminoflunitrazepam | 40,000 |
| 7-Aminonitrazepam | 40,000 |
| S-(+) Amphetamine | 40,000 |
| Table 6. Non-Interfering Structurally Unrelated Compounds- Qualitative and Semi-quantitative | |||
|---|---|---|---|
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.
| Compound | Conc. Tested (ng/mL) |
|---|---|
| Benzylpiperazine | 40,000 |
| Bromazepam | 40,000 |
| Buprenorphine | 40,000 |
| Bupropion | 40,000 |
| Butabarbital | 40,000 |
| Butalbital | 40,000 |
| Caffeine | 40,000 |
| Cannabidiol | 40,000 |
| Cannabinol | 40,000 |
| Carbamazepine | 40,000 |
| Carisoprodol | 40,000 |
| Chlordiazepoxide | 40,000 |
| cis-Tramadol | 40,000 |
| Clobazam | 40,000 |
| Clonazepam | 40,000 |
| Clozapine | 40,000 |
| Cocaine | 40,000 |
| Codeine | 40,000 |
| Cotinine | 40,000 |
| Demoxepam | 40,000 |
| Desalkylflurazepam | 40,000 |
| Dihydrocodeine | 40,000 |
| Diazepam | 40,000 |
| Digoxin | 40,000 |
| Dehydronorketamine | 40,000 |
| Delta-9-THC | 40,000 |
| Ecgonine | 40,000 |
| Ecgonine Methyl Ester | 40,000 |
| EMDP | 40,000 |
| 1R,2S(-)-Ephedrine | 40,000 |
| 1S,2R(+)-Ephedrine | 40,000 |
| Ethyl-β-D-Glucuronide | 40,000 |
| Ethylmorphine | 40,000 |
| Fenfluramine | 40,000 |
| Fentanyl | 20,000 |
| Flunitrazepam | 40,000 |
| Fluoxetine | 40,000 |
| Flurazepam | 40,000 |
| Haloperidol | 40,000 |
| Heroin | 40,000 |
| Hexobarbital | 40,000 |
| Compound | Conc. Tested (ng/mL) |
| Hydrocodone | 40,000 |
| Hydromorphone | 40,000 |
| 11-hydroxy-delta-9-THC | 40,000 |
| Ibuprofen | 40,000 |
| Ketamine | 40,000 |
| Lamotrigine | 40,000 |
| Levorphanol | 40,000 |
| Lidocaine | 40,000 |
| Lorazepam | 40,000 |
| Lorazepam Glucuronide | 40,000 |
| Lormetazepam | 40,000 |
| LSD | 40,000 |
| Maprotiline | 40,000 |
| MDA | 40,000 |
| MDEA | 40,000 |
| MDMA | 40,000 |
| Meperidine | 40,000 |
| Meprobamate | 40,000 |
| S(+)-Methamphetamine | 40,000 |
| Methadone | 40,000 |
| Methaqualone | 40,000 |
| Methylone | 40,000 |
| Methylphenidate | 40,000 |
| Midazolam | 40,000 |
| Morphine | 40,000 |
| Morphine-3-Glucuronide | 40,000 |
| Morphine-6-Glucuronide | 40,000 |
| N-desmethyltapentadol | 40,000 |
| N-desmethyl tramadol | 40,000 |
| N-desmethyl venlafaxine | 40,000 |
| Nalorphine | 40,000 |
| Naloxone | 40,000 |
| Naltrexone | 40,000 |
| Naproxen | 40,000 |
| Nitrazepam | 40,000 |
| 11-nor-9 carboxy THC | 40,000 |
| Norbuprenorphine | 40,000 |
| Norcodeine | 40,000 |
| Nordiazepam | 40,000 |
| Norketamine | 40,000 |
| Normorphine | 40,000 |
| Compound | Conc. Tested (ng/mL) |
| Noroxycodone | 40,000 |
| Noroxymorphone | 40,000 |
| Norpropoxyphene | 20,000 |
| Norpseudoephedrine | 2,000 |
| O-desmethyl tramadol | 40,000 |
| O-desmethyl venlafaxine | 40,000 |
| Oxycodone | 40,000 |
| Oxymorphone | 40,000 |
| Olanzapine | 40,000 |
| Oxazepam | 40,000 |
| Pentazocine | 40,000 |
| Pentobarbital | 40,000 |
| Phenobarbital | 40,000 |
| Phentermine | 40,000 |
| Phenylephrine | 40,000 |
| Phenytoin | 40,000 |
| Phenylpropanolamine | 40,000 |
| PMA | 40,000 |
| Prazepam | 40,000 |
| Propranolol | 40,000 |
| Propoxyphene | 40,000 |
| R,R(-)-Pseudoephedrine | 40,000 |
| S,S(+)-Pseudoephedrine | 40,000 |
| Ritalinic Acid | 40,000 |
| Salicylic Acid | 40,000 |
| Secobarbital | 40,000 |
| Sertraline | 40,000 |
| Sufentanil | 40,000 |
| Tapentadol | 40,000 |
| Temazepam | 40,000 |
| Theophylline | 40,000 |
| Trazadone | 40,000 |
| Triazolam | 40,000 |
| Trifluoromethylphenyl-piperazine | 40,000 |
| Verapamil | 40,000 |
| Zolpidem Tartrate | 40,000 |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color, and the letters are a bright white color. The image is simple and easy to read.
4. Interference - Endogenous Compounds and Exogenous Compounds
Endogenous compounds and exogenous compounds were evaluated in qualitative and semiquantitative modes by spiking the potential interferent into drug free oral fluid containing PCP at ±25% of the cutoff. Additional orally used products were tested by collecting oral fluid using
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The background is a solid red color. The image is simple and easy to read.
Quantisal II Oral Fluid Collection Devices from volunteers after use of the substances. Assay performance is unaffected by all of the substances tested. Endogenous compounds and exogenous compounds tested are presented in Tables 9 and 10. Orally used compounds tested are presented in Table 11.
| Compound | Concentration Tested |
|---|---|
| Ascorbic Acid | 3 mg/mL |
| Bilirubin | 0.15 mg/mL |
| Cholesterol | 0.45 mg/mL |
| γ-Globulin | 0.8 mg/mL |
| Hemoglobin | 3 mg/mL |
| Human Serum Albumin | 15 mg/mL |
| IgA | 1 mg/mL |
| IgG | 1 mg/mL |
| IgM | 0.5 mg/mL |
| Salivary-α-amylase | 1000 U/mL |
Table 7. Non-interfering Endogenous Compounds
Table 8. Non-interfering Exogenous Compounds
| Compound | Concentration Tested |
|---|---|
| Acetaminophen | 0.1 mg/mL |
| Acetylsalicylic Acid | 0.1 mg/mL |
| Baking Soda | 0.6% v/v |
| Cotinine | 0.03 mg/mL |
| Denture Adhesive | 0.6% w/v |
| Ibuprofen | 0.1 mg/mL |
| Alcohol (Ethanol) | 6% v/v |
| Caffeine | 0.1 mg/mL |
| Coffee | 6% v/v |
| Cranberry Juice | 6% v/v |
| Hydrogen Peroxide (3% OTC) | 0.5% v/v |
| Milk | 1% v/v |
| Mouthwash | 6% v/v |
| Naproxen | 0.1 mg/mL |
| Orange Juice | 6% v/v |
| Soft Drink (Pepsi) | 6% v/v |
| Sodium Chloride | 18 mg/mL |
| Sugar | 20 mg/mL |
| Tea | 6% v/v |
| Toothpaste | 6% v/v |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The font is bold and sans-serif. The red background is a solid color and there are no other objects in the image.
| Compound | Concentration Tested |
|---|---|
| Teeth Whitener | 2 strips |
| Hydrogen Peroxide (3% OTC) | Neat (2 min mouth rinse) |
| Cigarette | 1 cigarette |
| Hard Candy | 1 piece |
| Chewing Gum | 1 piece |
| Cough Syrup | 2 Teaspoons |
Table 9. Non-interfering Orally Used Exogenous Products
5. Interference - pH
To evaluate potential interference from the effect of oral fluid pH, device performance in the qualitative and semi-quantitative modes was tested using a range of oral fluid pH values (3.0, 4.0, 5.0, 6.0, 7.0, 8.0, 9.0, 10.0 and 11.0). All test samples were prepared in drug free oral fluid containing PCP at ±25% of the cutoff. No positive or negative interference was observed at oral fluid pH values ranging from 3.0 to 11.0 for each test mode.
6. Linearity/Recovery
A linearity study in the semi-quantitative mode was conducted by spiking a drug free oral fluid pool with a high concentration of PCP above the highest calibrator. Additional pools were made by serially diluting the high concentration specimen with drug free oral fluid to achieve concentrations ranging from 4 ng/mL to 44 ng/mL specimen was made from drug free oral fluid. Each pool was collected using Quantisal II oral fluid collection devices and tested in triplicate to calculate the mean concentration values that were used to calculate drug recovery. Linearity test results in semi-quantitative mode are presented in Tables 12 to 14.
| Expected Concentration(ng/mL) | Mean Concentration(ng/mL) | Recovery (%) |
|---|---|---|
| 0 | 1.5 | N/A |
| 4 | 4.2 | 105.0 |
| 8 | 8.5 | 105.8 |
| 10 | 10.0 | 100.0 |
| 12 | 11.7 | 97.2 |
| 16 | 15.8 | 98.5 |
| 20 | 21.2 | 105.8 |
| 24 | 24.6 | 102.4 |
| 28 | 28.5 | 101.9 |
| 32 | 31.5 | 98.3 |
| 36 | 35.5 | 98.6 |
| 40 | 42.1 | 105.3 |
| 44 | 46.0 | 104.5 |
Table 10. Linearity/Recovery - Quantisal II "A"
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.
| Expected Concentration(ng/mL) | Mean Concentration(ng/mL) | Recovery (%) |
|---|---|---|
| 0 | 1.3 | N/A |
| 4 | 3.7 | 92.5 |
| 8 | 8.2 | 102.1 |
| 10 | 10.3 | 102.7 |
| 12 | 12.3 | 102.5 |
| 16 | 16.0 | 99.8 |
| 20 | 19.0 | 94.8 |
| 24 | 24.1 | 100.6 |
| 28 | 27.7 | 98.9 |
| 32 | 33.9 | 105.8 |
| 36 | 36.6 | 101.8 |
| 40 | 42.0 | 105.1 |
| 44 | 46.4 | 105.5 |
Table 11. Linearity/Recovery - Quantisal II "B"
7. Calibration Duration
Drug free negative oral fluid spiked with PCP at ±25% of the cutoff were tested in qualitative mode at time points up to 14 days. At the initial time point calibration curve was established. This calibration was used through the duration of this study. The test results met acceptance criteria at each time point. The recommended frequency of calibration is 14 days.
8. PCP Stability in Oral Fluid
Drug free negative oral fluid spiked with PCP at +50% of the cutoff were collected and stored in Quantisal II Oral Fluid Collection Device and tested by LC-MS/MS at each time point at 30°C and at 2℃ - 8℃. Test results indicated that oral fluid samples containing PCP are stable for up to 10 days stored in Quantisal II transport tubes at ambient temperature up to 30°C and up to 2 months stored in Quantisal II transport tubes at 2℃ - 8℃.
9. Sample Transportation Stability
Drug free negative oral fluid spiked with PCP at ±50% of the cutoff were collected and stored in Quantisal II Oral Fluid Collection Device and packed in standard boxes used by common carrier (FedEx). During the 4-day (96 hours) simulated transportation study, the samples were stored in oven/freezer at temperatures ranged from -20°C to 40°C to encompass the temperatures likely to occur during shipment of products. The device used as the reference (unstressed) condition was stored continuously at the recommended storage condition at 2℃ - 8℃. LC-MS/MS testing was performed in replicates of two and compared to the reference sample. The studies demonstrated the PCP concentration of the sample collected by Quantisal II Oral Fluid Collection Device is within 20% of initial value during transportation.
10. Sample Recovery
Drug free negative oral fluid spiked with PCP at ±25%, +50% of the cutoff were collected and stored in Quantisal II Oral Fluid Collection Devices overnight at room temperature. LC-MS/MS testing was performed the next day to determine percentage recovery. The studies demonstrated
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.
that the Quantisal II Collection Device recovers PCP at greater than 80% of the original concentration.
11. Quantisal II Sample Volume
Fifty oral fluid samples were collected using Quantisal II collectors (collection pad with plastic stem) from fifty volunteers. Prior to collection, each collector (A and B) was weighed. After the volume adequacy indicator turned blue on both A and B collector stems, each collector was weighed again. The difference in weight was noted. Specific gravity of saliva was rounded to 1.000 to compute the volume collection. The results confirmed consistency of sample volume of 1 mL collected by each of Quantisal II collectors in this population.
Additional seventy-five oral fluid samples from known drug users were collected using Quantisal II collector. After the volume adequacy indicator turned blue on both A and B collector stems, each collector was weighed and compared to the average weight of collector before collection. The difference in weight was noted. Specific gravity of saliva was rounded to 1.000 to compute the volume collection. The results confirmed consistency of sample volume of 1 mL collected by each of Quantisal II collectors in this population.
12. Quantisal II Sample Collection Time
Fifty oral fluid samples from volunteers and seventy-five oral fluid samples from known drug users were collected using Quantisal II collection pad with plastic stem). The collection time was documented. The results verified the sample collection time for Quantisal II Oral Fluid Collection Device is within the claimed time of 10 minutes in over 90% of subjects.
13. Method Comparison
Eighty deidentified, unaltered clinical oral fluid samples collected by Quantisal II Oral Fluid Collection Devices were obtained from drug treatment facilities, analyzed for PCP at assay cutoff with the Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay in both qualitative and semi-quantitative modes and compared to Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) results with neat oral fluid collected by expectoration (spitting). The instruments used were the Beckman Coulter AU480 chemistry analyzer and an Agilent 6430 Liquid Chromatography-Tandem Mass Spectrometry. Method comparison test results in qualitative and semi-quantitative modes are presented from Tables 15 to 17.
| Immunalysis SEFRIAPCP Oral FluidEnzyme ImmunoassayResult | LC-MS/MS PCP Neat Oral Fluid Concentration | Agreement(%) | ||||
|---|---|---|---|---|---|---|
| < 5ng/mL(less than-50%cutoff) | 5 - 9 ng/mL(between -50% cutoffand cutoff) | 10 - 15ng/mL(betweencutoff and+50% cutoff) | > 15 ng/mL(greaterthan +50%cutoff) | |||
| Qual. | Positive | 0 | 0 | 5 | 35 | 100% (40/40) |
| Negative | 36 | 4 | 0 | 0 | 100% (40/40) | |
| Semi-Quant. | Positive | 0 | 0 | 5 | 35 | 100% (40/40) |
| Negative | 36 | 4 | 0 | 0 | 100% (40/40) |
Table 12. Method Comparison - Quantisal II "A"
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered and in all caps. The background is a solid red color.
| Immunalysis SEFRIAPCP Oral FluidEnzyme ImmunoassayResult | LC-MS/MS PCP Neat Oral Fluid Concentration | |||||
|---|---|---|---|---|---|---|
| < 5 ng/mL(less than-50%cutoff) | 5 – 9 ng/mL(between -50% cutoffand cutoff) | 10 - 15 ng/mL(betweencutoff and+50% cutoff) | > 15 ng/mL(greaterthan +50%cutoff) | Agreement(%) | ||
| Qual. | Positive | 0 | 0 | 5 | 35 | 100% (40/40) |
| Negative | 36 | 4 | 0 | 0 | 100% (40/40) | |
| Semi-Quant. | Positive | 0 | 0 | 5 | 35 | 100% (40/40) |
| Negative | 36 | 4 | 0 | 0 | 100% (40/40) |
H. CONCLUSION
The information provided in this pre-market notification demonstrates that the Immunalysis SEFRIA PCP Oral Fluid Enzyme Immunoassay and Immunalysis Quantisal II Oral Fluid Collection Device are substantially equivalent to the legally marketed predicate device for their intended uses.
N/A