K Number
K181127
Manufacturer
Date Cleared
2019-01-03

(248 days)

Product Code
Regulation Number
884.1690
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EndoRotor® is intended for use in endoscopic procedures by a trained gastroenterologist to resect and remove tissue, not intended for biopsy, of the gastrointestinal (GI) system including post-endoscopic mucosal resection (EMR) tissue persistence with a scarred base and residual tissue from the peripheral margins following EMR.

Device Description

The EndoRotor®, with the design modifications and expanded indications for use described in the current 510(k) premarket notification, will continue to be marketed as the EndoRotor®. The new EndoRotor® will be referred to as the "modified EndoRotor" in this 510(k) Summary.

The overall device description remains unchanged since initial FDA clearance. Both the predicate and modified EndoRotor® are powered resection tools consisting of a power console, foot control, specimen trap with pre-loaded filter, and a single-use resection catheter that is inserted into the working channel of a compatible endoscope. The design modifications to the EndoRotor® described in the current 510(k) premarket notification include enhancements to improve the ease of use of the console and catheter components as well as provide flexibility to the gastroenterologist.

AI/ML Overview

The provided text describes modifications to an existing device, the EndoRotor®, and does not present an AI/ML device or its associated performance studies. Therefore, I cannot extract information related to AI/ML specific acceptance criteria, and study details like sample size for test sets, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details, or ground truth establishment.

Instead, the document details non-clinical and clinical performance testing conducted to support the substantial equivalence of the modified EndoRotor® to its predicate device (K170120) and to support expanded indications for use.

Here's the relevant information about the acceptance criteria and the studies that prove the device meets them:

1. A table of acceptance criteria and the reported device performance:

The document doesn't explicitly list acceptance criteria in a quantitative table format with corresponding numerical performance for the modified device in the same way an AI/ML device submission might. Instead, it states that:

EndoRotor® ComponentTesting PerformedResults
EndoRotor® Catheter• Biocompatibility Testing• Functional Testing• All components biocompatible• All acceptance criteria met
EndoRotor® Console• Power-up and Set-up Testing• Functional Testing• All acceptance criteria met

2. Sample size used for the test set and the data provenance:

  • Non-clinical (Animal Testing for Predicate Device, applicable to modified):

    • Sample Size: 6 porcine animals (4 recovery, 2 acute) for mucosal resections, totaling 124 mucosal resections. Additionally, 4 animals for design validation and usability assessment.
    • Data Provenance: Porcine animal model.
  • Clinical (Investigator-led study in UK):

    • Sample Size: 19 patients.
    • Data Provenance: Retrospective, from Queens Alexandra Hospital in Portsmouth, United Kingdom.
  • Clinical (Commercial procedures in Western Europe):

    • Sample Size: 78 patients.
    • Data Provenance: Real-world commercial procedures in Western Europe (Austria, Germany, The Netherlands, Switzerland, and the United Kingdom). The document implies this data is also retrospective "compiled data provides real world clinical evidence".

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Clinical (Investigator-led study in UK):
    • Number of Experts: Not explicitly stated, but the study was conducted by "Gastroenterologists" in Queens Alexandra Hospital.
    • Qualifications: "trained gastroenterologist" as per the Indications for Use. The study authors are listed as Khandiah K, Subramanian S, Thayalasekaran S, Chedgy F, and Bhandari P, implying multiple gastroenterologists were involved.
  • Clinical (Commercial procedures in Western Europe):
    • Number of Experts: Not explicitly stated, but implies "Physicians" performing the procedures.
    • Qualifications: "trained gastroenterologist" as per the Indications for Use.

4. Adjudication method for the test set:

Not explicitly mentioned for the clinical studies. Decisions appear to have been made by the treating gastroenterologists.

5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done:

No, an MRMC comparative effectiveness study was not done. This is a medical device submission, not specifically an AI/ML device, and the studies performed focused on the safety and effectiveness of the device itself rather than human reader improvement with AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable, as this is a medical device (EndoRotor®), not an algorithm. The device is always under the control of a trained physician.

7. The type of ground truth used:

  • Non-clinical (Animal Testing): Histopathological assessment of tissue response (favorable and clinically acceptable tissue response).
  • Clinical (Investigator-led study in UK & Commercial procedures): Physician assessment of lesion presence/scarring and procedure outcomes (disease eradication, luminal preservation, avoidance of surgery, no incidences of perforation or delayed bleeding). This can be considered expert consensus/clinical outcomes data.

8. The sample size for the training set:

Not applicable. This is not an AI/ML device; therefore, there isn't a "training set" in that context. The device's design improvements were based on feedback and engineering enhancements.

9. How the ground truth for the training set was established:

Not applicable for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue.

January 3, 2019

Interscope, Inc. % Cynthia Nolte Director, Regulatory ICON Clinical Research LLC 2100 Pennbrook Parkway North Wales, Pennsylvania 19454

Re: K181127

Trade/Device Name: EndoRotor Console, EndoRotor Catheters, EndoRotor Specimen Trap, EndoRotor Filter Set, EndoRotor Foot Control Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: Class II Product Code: PTE Dated: December 13, 2018 Received: December 14, 2018

Dear Cynthia Nolte:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel G. Walter Jr -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181127

Device Name Endo Rotor®

Indications for Use (Describe)

The EndoRotor® is intended for use in endoscopic procedures by a trained gastroenterologist to resect and remove tissue, not intended for biopsy, of the gastrointestinal (GI) system including post-endoscopic mucosal resection (EMR) tissue persistence with a scarred base and residual tissue from the peripheral margins following EMR.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

Modification to the Interscope, Inc., EndoRotor®: Generation 2 (Gen 2) (per 21CFR 807.92)

1. SUBMITTER/510(K) HOLDER

Interscope, Inc.
100 Main Street, Suite 108
Whitinsville, MA 01588
Phone:+1 877-420-7299
Contact:Jeffery Ryan, Co-Founder, President & CEO
Contact Phone:617-360-1168
Contact Email:Jeffery.ryan@interscopemed.com
Date Prepared:December 4, 2018

2. DEVICE NAME

Proprietary Name:EndoRotor®
Classification Name:Endoscopic Morcellator Gastroenterology
Regulation Name:Hysteroscope and accessories
Classification Regulation:21 CFR 884.1690
Product code:PTE

The EndoRotor®, with the design modifications and expanded indications for use described in the current 510(k) premarket notification, will continue to be marketed as the EndoRotor®.

3. PREDICATE DEVICE

Manufacturer:Interscope, Inc.
Proprietary Name:EndoRotor®
Classification Name:Endoscopic Morcellator Gastroenterology
510(k) Number:K170120
Regulation Name:Hysteroscope and accessories
Classification Regulation:21 CFR 884.1690
Product Code:PTE

4. DEVICE DESCRIPTION

The purpose of the current 510(k) Premarket Notification is to obtain clearance for modifications to the EndoRotor® design to expand the range of catheters available to gastroenterologists and improve the flexibility and ease of use of the device. In addition, the indications for use for the EndoRotor® have been clarified to include post-endoscopic mucosal resection (EMR) persistent tissue with a scarred base.

{4}------------------------------------------------

The EndoRotor®, with the design modifications and expanded indications for use described in the current 510(k) premarket notification, will continue to be marketed as the EndoRotor® The new EndoRotor® will be referred to as the "modified EndoRotor" in this 510(k) Summary.

The overall device description remains unchanged since initial FDA clearance. Both the predicate and modified EndoRotor® are powered resection tools consisting of a power console, foot control, specimen trap with pre-loaded filter, and a single-use resection catheter that is inserted into the working channel of a compatible endoscope. The design modifications to the EndoRotor® described in the current 510(k) premarket notification include enhancements to improve the ease of use of the console and catheter components as well as provide flexibility to the gastroenterologist.

5. INDICATION FOR USE/INTENDED USE

The EndoRotor® is intended for use in endoscopic procedures by a trained gastroenterologist to resect and remove tissue, not intended for biopsy, of the gastrointestinal (GI) system including post-endoscopic mucosal resection (EMR) tissue persistence with a scarred base and residual tissue from the peripheral margins following EMR.

The indications for use for the EndoRotor® described in the current 510(k) premarket notification were clarified to include post EMR persistent tissue with a scarred base. The modified EndoRotor® continues to be indicated to resect and remove residual tissue from the peripheral margins following EMR.

6. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICE/S

Both the modified and predicate EndoRotor® devices are powered resection tools consisting of a power console, foot control, specimen trap with preloaded filter and a single-use resection catheter that is inserted into the working channel of a compatible endoscope. Resection of residual tissue with the catheter is achieved through reciprocating and rotational movement of the inner cannula relative to the fixed outer cannula while removal is performed via vacuum suction. The device is always under the control of a trained physician and is technique dependent with regards to depth of removal and avoiding perforation.

To improve the flexibility, performance, and ease of the system, Interscope has implemented design enhancements to the catheter and console components as described below.

Console

The console houses the control panel, drive motor, vacuum control valve, and peristaltic irrigation pump drive. The principles of operation and overall characteristics of the modified

{5}------------------------------------------------

EndoRotor® Console remain unchanged since initial clearance (K170120).

The console was modified to "Generation 2" (Gen2) to improve usability and ease of use as described in Table 5-1.

Design ChangeJustification of Design Change
New Mechanical Irrigation PumpOffers improved usability and component quality.
Addition of Lavage ButtonAllows user to enable or disable irrigation flow. Improvesusability, giving user control of irrigation to improve the field ofview of the resection site.
Reduction in Irrigation Flow RateReduction in the irrigation flow rate from 5-12mL/min to a fixed5mL/min. Improves usability, providing better visibility to theresection site.
Increase in Prime Flow RateIncrease in the prime flow rate from 4mL/min to 25 mL/min andduration from 10 to 25 seconds. Improves usability by allowingcomplete priming with one button push.
Increase in Pitch Valve Open TimingIncrease in the pinch valve open timing during load from 10 to 25seconds. Improves usability by enabling increased prime flow rate.

Table 5-1. EndoRotor Gen2 Console Design Changes

Catheter

The catheter design was updated to produce "Generation II" (Gen2) catheters; overall design principles are the same. All catheters include a debriding cutter and support functions of lavage and aspiration from the anatomical site through the endoscope working channel to the EndoRotor® Specimen Trap.

Design for all catheters was enhanced per Table 5-2 to provide more options to the clinician and procedure and improve performance. Additionally, Interscope has added two (2) new catheters. The two (2) new catheters (bolded in table below) were added to increase cutting efficiency. An overview of all available catheters and characteristics are provided in Table 5-3.

Design ChangeJustification of Design ChangeApplicability to Catheters
New cutting window1 orientationinterface, reinforced outer braidedshaft, added MobilizeⓇ lubriciousadditive to outer braided shaftEnables improved performance of orientingthe cutting window of the catheter to thetarget tissue.ER 10-01-OP2ER 10-02-O3ER 10-03-OP4ER 10-03-OP-SER 10-03-F-S
Modified catheter lengthsTo expand compatibility with the range ofendoscope on the market.
Expanded cutting window geometryTo address limitations in certain anatomicallocations (i.e., frontal face resection)ER 10-03-OP-SER 10-03-F-S
Added serrated teeth to inner andouter cuttersTo increase cutting efficiency

Table 5-2. EndoRotor® Gen2 Catheter Design Changes

l New cutting window only applies to bolded catheters

2 Revised part number – originally cleared (K170120) as part no. ER 10-01

3 Revised part number – originally cleared (K170120) as part no. ER 10-02

4 Revised part number – originally cleared (K170120) as part no ER 10-03

{6}------------------------------------------------

Part No.LengthCompatible EndoscopesWindowSizeCutterSerrated teethplacement
ER 10-01-OP13.2mm x 1890mmOlympus colonoscopes 3.2mm x 1680mm;Pentax colonoscopes 3.2 mm x 1700mmInner cutter
ER 10-02-O23.2mm x 1540mmOlympus colonoscopes 3.2mm x 1330mm3.0mm2Inner cutter
ER 10-03-OP33.2mm x 1240mmOlympus gastroscopes 3.2mm x 1030mm;Pentax gastroscopes 3.2mm x 1050mmInner cutter
ER 10-03-OP-S3.2mm x 1240mmOlympus gastroscopes 3.2mm x 1030mm;Pentax gastroscopes 3.2mm x 1050mm4.4 mm2Inner and outer cutter
ER 10-03-F-S3.2 mm x 1270 mmFuji gastroscopes 3.2 mm x 1100 mmInner and outer cutter

Table 5-3. Gen2 EndoRotor® Catheters

l Revised part number – originally cleared (K170120) as part no. ER 10-01 2Revised part number – originally cleared (K170120) as part no. ER 10-02 3 Revised part number – originally cleared (K170120) as part no ER 10-03

7. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE

Interscope has completed verification testing to support the Catheter and Console design changes, as summarized in Section 6 of this 510(k) summary. Testing executed was limited to the design changes and met all acceptance criteria as summarized in Table 5-4.

EndoRotor® ComponentTesting PerformedResults
EndoRotor® Catheter• Biocompatibility Testing• Functional Testing• All components biocompatible• All acceptance criteria met
EndoRotor® Console• Power-up and Set-up Testing• Functional Testing• All acceptance criteria met

Table 5-4. Testing to Support Gen2 Design Changes

The EndoRotor® design, as cleared under K170120, was verified and validated through a series of physical and mechanical performance tests on the catheter and console as well as animal testing. Those tests that are applicable to the modified EndoRotor® are summarized in Table 5-5.

Table 5-5. Non-Clinical Performance Testing of EndoRotor® Cleared Under K170120 Applicable to the Modified EndoRotor®

EndoRotor® ComponentTesting Performed
EndoRotor® Catheter• Sterilization validation• Pyrogenicity testing• Shelf-life• Packaging validation• Transport testing
EndoRotor® Console• Design verification testing• Electrical Safety and Electromagnetic Compatibility Testing
Animal TestingThe safety and performance of the EndoRotor® was evaluated in a porcine animal model which included one hundred and twenty-four (124) mucosal resections over 6 animals (4 recovery and 2 acute) and 3 organs (colon, stomach, esophagus). All animals tolerated the treatment well and results showed the EndoRotor® system was associated with favorable and clinically acceptable tissue response.Design validation and usability assessment were completed in a porcine model with (4)

{7}------------------------------------------------

EndoRotor® ComponentTesting Performed
additional animals without complication. No difference was noted in the safety profile ofthe device.
Procedural TestingSpecimen Trap: EndoRotor® was evaluated to determine that when proper procedure isfollowed as provided in the labeling, including a post procedure flush, there is noresidual specimen in the catheter.

8. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE

Usability studies were conducted to evaluate the clinical performance of the predicate EndoRotor®. The modifications to the console and catheter were made to improve EndoRotor performance for tissue resection and removal in procedures where biopsy is not required. The results confirmed that all system requirements related to usability were met. These studies are applicable to the modified EndoRotor®.

An investigator-led study was conducted to support the safety and effectiveness of the EndoRotor® for post-endoscopic mucosal resection (EMR) persistent tissue with a scarred Gastroenterologists in Queens Alexandra Hospital, located in Portsmouth, United base. Kingdom. completed an investigator-led series utilizing the predicate EndoRotor® in management of post EMR tissue persistence with a scarred base. In a pending peer review manuscript Khandiah K, Subramanian S, Thayalasekaran S, Chedgy F and Bhandari P. Nineteen (19) patients were referred to Queens Alexandra Hospital, a tertiary center, following diagnosed tissue persistence following EMR. Scarred sites of the colon, in each of the nineteen (19) patients, were evaluated by physicians and found to have lesions determined difficult to resect and no longer amenable to EMR due to scarring from previous EMR. The EndoRotor was used to perform resection of the identified patient site and successfully demonstrated the ability to superficially resect the scarred areas. The procedure outcomes using the EndoRotor® ensured muscle and luminal preservation in addition to disease eradication in fifteen (15) patients in twenty-one (21) procedures (1.4 procedure average) with no incidences of perforation or delayed bleeding, all patients were seen at follow up n=4 months. Study outcomes showed EndoRotor® directly led to successful avoidance of surgery in 78.9% of patients

Physicians in Western Europe (Austria, Germany, The Netherlands, Switzerland and the United Kingdom) completed seventy-eight (78) commercial EndoRotor® procedures to treat the aforementioned scarred lesions. Similarly disease eradication and luminal preservation was achieved within two (2) procedures or less with seventy (70) patients requiring one (1) treatment and eight (8) patients requiring two (2) procedures without complication.

The compiled data provides real world clinical evidence to support the expanded indications for use. The clinical experience outlined in this Premarket Notification supports the clinical safety and efficacy of the EndoRotor for use in the expanded indications.

{8}------------------------------------------------

9. CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL TESTS

Interscope concludes that the modified EndoRotor® is substantially equivalent to the predicate EndoRotor® cleared by FDA (K170120). The results of bench testing to verify the performance of the Gen2 catheters and console, in addition to the thirty eight (38) procedures demonstrating successful treatment of post EMR tissue persistence, support the safety and effectiveness of the modified EndoRotor for the expanded indications for use. A comparison of the currently marketed EndoRotor® and the EndoRotor®: Gen2 is provided in Table 5-6.

Proposed DevicePredicateSame/Different
SponsorInterscope, Inc.Interscope, Inc.
Device nameEndoRotor®EndoRotor®Same
Regulatory StatusProposedK170120N/A
DeviceClassificationCode/NamePTE – Endoscopic MorcellatorGastroenterologyPTE - Endoscopic MorcellatorGastroenterologySame
RegulationNumber21CFR 884.169021CFR 884.1690Same
Regulation NameHysteroscope and accessoriesHysteroscope and accessoriesSame
Indications forUseThe EndoRotor® is intendedfor use in endoscopicprocedures by a trainedgastroenterologist to resect andremove residual tissue, notintended for biopsy, of postendoscopic mucosal resection(EMR) tissue persistence witha scarred base, and residualtissue from the peripheralmargins following EMR.The EndoRotor® is intended foruse in endoscopic proceduresby a trained gastroenterologistto resect and remove residualtissue from the peripheralmargins following EMR(Endoscopic MucosalResection).The EndoRotor® should not beused for tissue intended biopsy.Same:-Intended Use - clarification onoriginal intended use postEMR tissue persistence witha scarred baseDifferent:- Addition of EMR tissuepersistence with a scarredbase
Components• Control unit includingperistaltic pump, motor driveand pinch valve• Catheter with cutting devicemounted on distal end• Foot control to control drivemotor• Specimen trap with pre-loaded micron filter• Control unit includingperistaltic pump, motor driveand pinch valve• Catheter with cutting devicemounted on distal end• Foot control to control drivemotor• Specimen trap with pre-loaded micron filterDifferent:Three (3) new catheters willbe available; two (2) toaccommodate endoscopevariability amongmanufactures and one (1) tosupport faster tissueacquisition
Principle ofOperationMechanical resection using acutting cannula withsimultaneous aspirationMechanical resection using acutting cannula withsimultaneous aspirationSame
Reuse statusSingle useSingle useSame
Table 5-6. Side-by-Side Comparison for Determination of Substantial Equivalence

§ 884.1690 Hysteroscope and accessories.

(a)
Identification. A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for hysteroscope accessories that are not part of a specialized instrument or device delivery system; do not have adapters, connectors, channels, or do not have portals for electrosurgical, laser, or other power sources. Such hysteroscope accessory instruments include: lens cleaning brush, cannula (without trocar or valves), clamp/hemostat/grasper, curette, instrument guide, forceps, dissector, mechanical (noninflatable), and scissors. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.