K Number
K180856
Date Cleared
2018-06-19

(78 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Neutrogena Light Therapy Aging Mask+ is an over the counter device that is indicated for the treatment of full face wrinkles.

Device Description

The Neutrogena Light Therapy Aging Mask+ uses known LED light therapy technology for the treatment of facial wrinkles. The device emits a combination of red light and infrared light. Users place the lightweight mask over the face and press the "On" button on the controller module to start treatment. The device will automatically turn off after each 10-minute treatment.

AI/ML Overview

The provided document focuses on the FDA's 510(k) clearance for the Neutrogena Light Therapy Aging Mask+ and primarily uses substantial equivalence to predicate devices, rather than establishing new acceptance criteria and proving the device meets them through clinical studies with specific performance metrics.

Therefore, the information required to answer your questions regarding acceptance criteria, study details, expert involvement, and ground truth establishment (as typically found in clinical trials for new efficacy claims) is not present in this document.

The document states: "Performance testing performed on the Neutrogena Light Therapy Aging Mask+ is sufficient to demonstrate that the subject device is as safe and effective as the legally marketed predicate devices." This indicates that the primary method for demonstrating effectiveness was through comparison to existing devices with established safety and efficacy profiles, rather than establishing new, quantitative performance criteria for wrinkle reduction itself.

However, I can extract the following information that is available:

1. Table of acceptance criteria and the reported device performance:

This document does not specify quantitative acceptance criteria for wrinkle reduction performance or report the device's performance against such criteria. Instead, it compares the technological characteristics and indications for use of the subject device to predicate devices to demonstrate substantial equivalence.

Device FeatureNeutrogena Light Therapy Aging Mask+ (Subject Device)
Intended UseLight-based treatment to reduce wrinkles on the body in general or specific anatomical area
Indications for UseThe Neutrogena Light Therapy Aging Mask+ is an over the counter device that is indicated for the treatment of full face wrinkles.
Wavelength (nm)Red 620-640nm, IR 820-880nm
Intensity (mW/cm²)1.32 mW/cm²
Treatment Regimen10 minutes/day for 60 sessions
Total Energy Dose (J/cm²)47.58 J/cm²

2. Sample size used for the test set and the data provenance: Not provided in the document for efficacy testing. The performance and safety testing mentioned are:
* Performance verification
* Human Factors Evaluation
* Biocompatibility testing
* Electrical Safety and Electromagnetic Compatibility testing

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided. This type of information would typically be relevant for studies involving qualitative assessments (e.g., by dermatologists for wrinkle severity), which are not detailed here.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a light therapy device, not an algorithm.

7. The type of ground truth used: Not explicitly stated for performance/efficacy, as the primary basis for clearance is substantial equivalence. For safety assessments, recognized standards (biocompatibility, electrical safety) serve as the "ground truth" or reference.

8. The sample size for the training set: Not applicable, as this refers to a medical device's clearance based on technological characteristics and comparison to predicates, not an AI model's training.

9. How the ground truth for the training set was established: Not applicable.

In summary, the FDA 510(k) clearance process for this device relies heavily on demonstrating substantial equivalence to pre-existing, legally marketed predicate devices, rather than presenting a novel clinical study with new, specific performance acceptance criteria for wrinkle reduction. The provided document details the technological characteristics and safety testing (bench, human factors, biocompatibility, electrical safety) that support this substantial equivalence claim.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 19, 2018

Johnson & Johnson Consumer, Inc. Cindy Abraham Associate Director, Regulatory Affairs 7050 Camp Hill Road Fort Washington, Pennsylvania 19034

Re: K180856

Trade/Device Name: Neutrogena Light Therapy Aging Mask+ Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: March 21, 2018 Received: April 2, 2018

Dear Cindy Abraham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director

Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180856

Device Name

Neutrogena Light Therapy Aging Mask+

Indications for Use (Describe)

The Neutrogena Light Therapy Aging Mask+ is an over the counter device that is indicated for the treatment of full face wrinkles.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

ട്. 510(k) SUMMARY

510(k) NUMBER:K180856
510(k) SUBMITTER:Johnson & Johnson Consumer, Inc.7050 Camp Hill RoadFort Washington, PA 19034
CONTACT:Cindy R. AbrahamAssociate Director, Regulatory Affairscabraha3@its.jnj.com
DATE PREPARED:March 21, 2018
PROPRIETARY NAME:Neutrogena® Light Therapy Aging Mask+
PANEL:General & Plastic Surgery
REGULATION NUMBER:CFR Title 21, 878.4810
CLASSIFICATION:Class II
PRODUCT CODES:OHS
COMMON NAME:Light-based Over The Counter Wrinkle Reduction

Predicate Devices [807.92(a)(3)] 5.1

5.2 Device Description [807.92(a)(4)]

The Neutrogena Light Therapy Aging Mask+ uses known LED light therapy technology for the treatment of facial wrinkles. The device emits a combination of red light and infrared light. Users place the lightweight mask over the face and press the "On" button on the controller module to start treatment. The device will automatically turn off after each 10-minute treatment.

{4}------------------------------------------------

Intended Use and Indications for Use [807.92(a)(5)] 5.3

The Neutrogena Light Therapy Aging Mask+ is an over the counter device that is indicated for the treatment of full face wrinkles.

5.4 Comparison of Technological Characteristics with the Predicate Devices [807.92(a)(6)]

The Neutrogena Light Therapy Aging Mask+ is substantially equivalent to the predicate devices in that it has the same intended use and technological characteristics, as demonstrated in Table 5.1. Performance testing performed on the Neutrogena Light Therapy Aging Mask+ is sufficient to demonstrate that the subject device is as safe and effective as the legally marketed predicate devices. The technological and labeling differences do not raise new or different questions about safety or effectiveness. The Neutrogena Light Therapy Aging Mask+ is substantially equivalent to the predicate devices.

{5}------------------------------------------------

Device FeatureJohnson & JohnsonNeutrogena LightTherapy AgingMask+Subject DeviceLG ElectronicsLG Beauty LEDMaskPredicate 1TrophyRejuvalite MDPredicate 2La Lumiere, LLCPro X OTC 5 LightTherapyPredicate 3
510(k) NumberNot yet assignedK170984K133896K140471
Classification21 CFR 878.4810
Common NameLight-based over-the-counter wrinkle reduction
Intended UseLight-based treatment to reduce wrinkles on the body in general or specific anatomical area
Product CodeOHS
Rx and/or OTCOTC
Use EnvironmentHome-Use
AnatomicalLocationFull face
Indications forUseThe NeutrogenaLight Therapy AgingMask+ is an over thecounter device that isindicated for thetreatment of full facewrinkles.The LG BEAUTYLED MASK is anover the counterdevice that isintended for the usein the treatment offull face wrinkles.The Rejuvalite MD isan Over the Counterdevice that isintended for the usein the treatment offull face wrinkles.The Pro X OTC 5 isan Over-The-Counterdevice intended foruse in the treatmentof facial wrinkles. Itis for people withwrinkles on their faceand who haveFitzpatrick skin typesI, II and/or III.
Power SourceNickel-MetalHydride rechargeablebatteriesLi-Ion rechargeablebatteriesAC to DCAA alkaline batteries
Energy SourceLight-emitting diodes
Wavelength (nm)Red 620-640nmIR 820-880nmRed 637nmIR 854nmRed 600, 622, 660nmIR 860nmRed 620-640nmIR 820-880nm
Intensity(mW/cm²)1.32 mW/cm²4.08 mW/cm²2.35 mW/cm²0.76 mW/cm²
TreatmentRegimen10 minutes/day for 60sessions9 minutes/day, 5days a week for 8weeks3 minutes/day, 5 daysa week for 8 weeks15 minutes/day for 60sessions
Total EnergyDose (J/cm²)47.58 J/cm²88.21 J/cm²16.94 J/cm²41.23 J/cm²

Table 5.1 Technological Characteristics

{6}------------------------------------------------

ર.5 Performance Data [807.92(b)(1), (b)(2)]

Performance bench and safety testing include

  • . Performance verification
  • Human Factors Evaluation .
  • Biocompatibility testing in accordance with requirements put forth in . ISO 10993-1
  • Electrical Safety and Electromagnetic Compatibility testing .

The verification, electrical safety testing, electromagnetic compatibility testing, and human factors data presented in this 510(k) submission demonstrate the Neutrogena Light Therapy Aging Mask+ device met the established specifications for its intended use and environment of use In addition, the testing demonstrated that the subject device does not raise new or different questions of safety or effectiveness when compared to the legally-marketed predicate devices.

5.6 Conclusion [807.92(b)(3)]

The basis for substantial equivalence for the Neutrogena Light Therapy Aging Mask+ device and the predicate devices is performance data and conformity with recognized standards. Performance bench verification and validation demonstrate that the subject device performs comparably to the predicate devices that are marketed for the same intended use. Based on the performance testing and the similarities of the indications for use and the technological characteristics, it can be concluded that the Neutrogena Light Therapy Aging Mask+ device is substantially equivalent to the predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.