K Number
K180715
Date Cleared
2018-07-04

(107 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The React™ 68 Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

Device Description

The React™ 68 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with a durable polymer outer layer. A lubricous, polytetrafluoroethylene liner is used to create a structure that has both proximal stiffness and distal flexibility, and an encapsulated radiopaque distal platinum-iridium markerband which is used for visualization under fluoroscopy. The React™ 68 Catheter is introduced into the vasculature through the split y-introducer sheath. The proximal end of the React™ 68 Catheter is designed with a thermoplastic elastomer strain relief and a clear hub. The React™ 68 Catheter is designed with a hydrophilic coating.

AI/ML Overview

This is a 510(k) premarket notification for the React™ 68 Catheter. The document focuses on demonstrating substantial equivalence to predicate devices through design comparisons and extensive bench testing. Animal and clinical studies were not deemed necessary for this submission.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes numerous functional, biocompatibility, and sterilization-related tests. The "Acceptance Criteria" for each test are implicitly that the device "met the acceptance criteria" or demonstrated results within acceptable limits for safe and effective use. The "Reported Device Performance" is the stated outcome of each test.

Test DescriptionAcceptance Criteria (Implicit)Reported Device Performance
Biocompatibility:
Cytotoxicity (Elution Method)No evidence of cell lysis/toxicity (grade < 2)Grade of 0 (no reactivity), considered non-cytotoxic
Sensitization (Guinea Pig Maximization)No evidence causing delayed dermal contact sensitizationNo evidence of causing delayed dermal contact sensitization
Irritation (Intracutaneous Reactivity)Difference between test/control scores of 0Difference of 0.0 for NaCl and sesame oil extracts, considered non-irritant
Acute Systemic Toxicity (Systemic Toxicity)No mortality or evidence of systemic toxicityNo mortality or evidence of systemic toxicity
Hemocompatibility (Hemolysis)Hemolytic index within acceptable limitsHemolytic index of 0.9% (direct), 0.0% (extract)
Hemocompatibility (Complement Activation)Test article not statistically higher than controlsTest article not statistically higher than controls, not a potential activator of complement system
Hemocompatibility (Thrombogenicity)Minimal thrombus formationMinimal thrombus formation after 4 hours
Pyrogenicity (Material Mediated)No animal temperature rise $\ge$ 0.5°C over baselineNo animal showed a temperature rise of 0.5°C or more
Performance Data – Bench (Microbial):
Ethylene Oxide ResidualMet ISO 10993-7 acceptance criteriaMet the acceptance criteria for ethylene oxide residual
Ethylene Chlorohydrin ResidualMet ISO 10993-7 acceptance criteriaMet the acceptance criteria for ethylene chlorohydrin residual
BioburdenMet ISO 11737-1 acceptance criteriaMet the acceptance criteria for bioburden
Bacterial EndotoxinMet ANSI/AAMI ST72 and USP <161> acceptance criteriaMet the acceptance criteria for bacterial endotoxin
Performance Data – Bench (Performance):
Visual InspectionMet acceptance criteriaMet the acceptance criteria for visual inspection
Dimensional MeasurementsMet acceptance criteria for all measured dimensionsMet the acceptance criteria for dimensional measurements
Tip BucklingMet acceptance criteria for maximum compressive forceMet the acceptance criteria for tip buckling
Kink ResistanceMet acceptance criteria for maximum kink diameterMet the acceptance criteria for kink resistance
ParticulateMet USP <788> acceptance criteriaMet the acceptance criteria for particulate evaluation
Coating LubricityMet acceptance criteria for average friction forcesMet the acceptance criteria for coating lubricity
Tensile StrengthMet ISO 10555-1 Annex B acceptance criteriaMet the acceptance criteria for tensile strength at hub and shaft
Liquid LeakMet ISO 10555-1 Annex C acceptance criteriaMet the acceptance criteria for liquid leak
Corrosion ResistanceMet ISO 10555-1 Annex A acceptance criteriaMet the acceptance criteria for corrosion resistance
Hub Aspiration ResistanceMet ISO 10555-1 Annex D acceptance criteriaMet the acceptance criteria for hub air aspiration
CompatibilityMet acceptance criteria for visual damage during useMet the acceptance criteria for delivering and retrieving interventional devices
RadiopacityMet acceptance criteria (markerband/wall thickness, fluoroscopy)Met the acceptance criteria for radiopacity
Luer StandardsMet ISO 594-1 and ISO 80369-7 acceptance criteriaMet the acceptance criteria for luer standards
Torque StrengthAble to withstand typical clinical torsional forcesAble to withstand torsional forces typical of clinical use
Dynamic PressureAble to withstand typical clinical pressuresAble to withstand pressures typical of clinical use
Coating IntegrityCoating remained covered and lubriciousRemained coated and lubricous
UsabilityMet acceptance criteria for maneuverability and flexibilityMet the acceptance criteria for usability

2. Sample Size Used for the Test Set and the Data Provenance

The document does not explicitly state the sample sizes for each specific test within the bench testing. However, such tests would typically involve a statistically relevant number of units to ensure reliability. The data provenance is from bench testing conducted by the manufacturer, Micro Therapeutics, Inc. d/b/a ev3 Neurovascular. The data is prospective as it was generated specifically for this submission. The country of origin of the data is not explicitly stated but can be inferred as the United States given the FDA submission and the company's address in Irvine, CA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. This device is a physical catheter, and its performance is evaluated through objective, standardized physical and chemical tests (e.g., ISO standards, USP, visual inspection, dimensional measurements). There is no "ground truth" derived from expert interpretation in the context of diagnostic performance testing of, for example, medical images.

4. Adjudication Method for the Test Set

Not applicable. As described above, the ground truth for this type of device's performance is established through objective laboratory testing against predetermined specifications and international standards, not through adjudications of expert opinions.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

Not applicable. This is a physical medical device (catheter) and not an AI or imaging diagnostic device. Therefore, MRMC studies are irrelevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Not applicable. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the React™ 68 Catheter's performance is defined by its compliance with:

  • Industry Standards: Such as ISO 10993 (Biocompatibility), ISO 11737-1 (Bioburden), ANSI/AAMI ST72 (Bacterial Endotoxin), USP <161>/<788> (Bacterial Endotoxin/Particulate), ISO 10555-1 (Catheter testing), ISO 594-1/80369-7 (Luer standards).
  • Predicate Device Performance: Demonstrating substantial equivalence to legally marketed predicate devices in terms of materials, dimensions, and functional characteristics.
  • Engineering Specifications: Internal design requirements for dimensions, material properties, and functional performance (e.g., tip buckling, kink resistance, lubricity, tensile strength).

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI or machine learning algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As above, there is no "training set" for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 4, 2018

Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Ryan Kenney Senior Regulatory Affairs Specialist 9775 Toledo Wav Irvine, California 92618

Re: K180715

Trade/Device Name: React 68 Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: June 4, 2018 Received: June 6, 2018

Dear Ryan Kenney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Xiaolin Zheng" in a large, bold font, followed by a hyphen. Below the name, there is a large, stylized FDA logo in a light gray color. To the left of the logo, there is a large letter "S".

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180715

Device Name React 68 Catheter

Indications for Use (Describe)

The React™ 68 Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: K180715

510(k) Owner:Micro Therapeutics, Inc. d/b/a ev3 Neurovascular9775 Toledo WayIrvine, CA 92618Establishment Registration No.: 2029214
Contact Person:Ryan KenneySenior Regulatory Affairs SpecialistTelephone: (949) 297-5489Email: ryan.j.kenney@medtronic.com
Date SummaryPrepared:June 25, 2018
Trade Name ofDevice:React™ 68 Catheter
Common Name ofDevice:Percutaneous Catheter
Review Panel:Neurology
Product Code:DQY
Regulation Number:21 CFR 870.1250
Regulation Name:Percutaneous Catheter
Device Classification:Class II
Predicate Device(s):AXS Catalyst™ Distal Access Catheter510(k)#: K151667ReFlex™ Guide Catheter510(k)#: K110055

Device Description:

The React™ 68 Catheter is a single lumen, flexible, variable stiffness composite catheter with a nitinol structure that is jacketed with a durable polymer outer layer. A lubricous, polytetrafluoroethylene liner is used to create a structure that has both proximal stiffness and distal flexibility, and an encapsulated radiopaque distal platinum-iridium markerband which is used for visualization under fluoroscopy. The React™ 68 Catheter is introduced into the vasculature through the split y-introducer sheath. The proximal end of the React™ 68 Catheter is designed with a thermoplastic elastomer strain relief and a clear hub. The React™ 68 Catheter is designed with a hydrophilic coating.

Indications for Use:

The React™ 68 Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

{4}------------------------------------------------

AXS Catalyst™ DistalAccess Catheter(K151667)ReFlex™ GuideCatheter(K110055)React™ 68 Catheter
Indication for Use(IFU) StatementThe AXS Catalyst™Distal Access Catheteris indicated for theinsertion and guidanceof appropriately sizedinterventional devicesinto a selected bloodvessel in the peripheraland neurovascularsystems. The AXSCatalyst™ DistalAccess Catheter is alsoindicated for use as aconduit for retrievaldevices.The ReFlex™ GuideCatheter is indicatedfor the introduction ofinterventional devicesinto the peripheral andneuro vasculature.Same as K110055
Materials
HubNylonTrogamid®Same as K110055
Strain ReliefThermoplastic RubberSantopreneDynaFlex®
Inner LayerPTFESameSame
ReinforcementStainless Steel withNitinol and PolymerFiberNitinolSame as K110055
Outer JacketPebax®PolymericGrilamid™Pebax®
MarkerbandPlatinum/IridiumSameSame
AdhesiveCyanoacrylateSameSame
CoatingHydrophilicSameSame
Dimensions
Working Length132 cm130 cmSame as K151667
Inner Diameter (ID)0.060"0.072"0.068"
Proximal OuterDiameter (OD)0.079"0.084"0.083"
Distal OuterDiameter (OD)0.071"0.084"0.083"
Packaged Accessories
Peelable SheathYesSameSame
Rotating HemostasisValveYesNoSame as K110055
Packaging
Packaging CardPolyethyleneSameSame
Packaging HoopPolyethyleneSameSame
AXS Catalyst™ DistalAccess Catheter(K151667)ReFlex™ GuideCatheter(K110055)React™ 68 Catheter
Other
SterilizationEthylene Oxide (EO)SameSame

Device Comparison:

{5}------------------------------------------------

Biocompatibility:

Biocompatibility was conducted for the React™ 68 Catheter. The React™ 68 Catheter is categorized as a limited exposure (< 24 hours), external communicating device contacting circulating blood. The following biocompatibility was conducted for the React™ 68 Catheter:

Test DescriptionResultsConclusions
Cytotoxicity(Elution Method)The test article extract showedno evidence of causing celllysis or toxicity and had agrade of zero (no reactivity).The test article extract met therequirements of the test sincethe grade was less than a gradetwo (mild reactivity).The ReactTM 68 Catheter isconsidered non-cytotoxic.
Sensitization(Guinea Pig Maximization Test)The test article extractsshowed no evidence ofcausing delayed dermalcontact sensitization in theguinea pig.The ReactTM 68 Catheter doesnot elicit a sensitizationresponse.
Irritation(Intracutaneous Reactivity)The difference between eachtest article extract overallmean score and correspondingcontrol extract overall meanscore was 0.0 and 0.0 for thesodium chloride and sesameoil test article extracts,respectively.The ReactTM 68 Catheter isconsidered a non-irritant.
Acute Systemic Toxicity(Systemic Toxicity)There was no mortality orevidence of systemic toxicityfrom the extracts injected. Allanimals were clinically normalthroughout the study.The ReactTM 68 Catheter doesnot indicate signs of toxicity.
Hemocompatibility(Hemolysis)The hemolytic index for thetest article in direct contactwith blood was 0.9%, and thehemolytic index for the testarticle extract was 0.0%.The ReactTM 68 Catheter isconsidered non-hemolytic.
Hemocompatibility(Complement Activation)The concentration of SC5b-9in the sponsor providedcontrol sample was notThe control and test articlesamples are not considered to

{6}------------------------------------------------

Test DescriptionResultsConclusions
statistically higher than theactivated normal human serumcontrol or the negative control.be potential activators of thecomplement system.
The concentration of SC5b-9in the test article sample wasnot statistically higher than theactivated normal human serumcontrol, the negative control orthe sponsor provided control.
Hemocompatibility(Thrombogenicity)The thrombogenic potential ofthe control article wasevaluated in comparison to thetest article. Both the controland test articles weredetermined to have minimalthrombus formation after four(4) hours (±30 minutes)without anticoagulation.The React™ 68 Catheterdemonstrates minimalthrombus formation.
Pyrogenicity(Material Mediated)Not a single animal showed atemperature rise of 0.5°C ormore above its baselinetemperature. The total rise ofthe temperature during three(3) hours was 0.0°C.The React™ 68 Catheter isconsidered non-pyrogenic.

The React™ 68 Catheter has been evaluated to meet requirements specified in ISO 10993-1.

Performance Data – Bench:

Non-clinical bench testing was conducted to evaluate the performance of the React™ 68 Catheter. The following non-clinical bench testing was conducted for the React™ 68 Catheter:

TestTest Method SummaryResults
Microbial
Ethylene Oxide ResidualThe React™ 68 Catheter was evaluated per ISO 10993-7.The React™ 68 Catheter met the acceptance criteria for ethylene oxide residual.
Ethylene Chlorohydrin ResidualThe React™ 68 Catheter was evaluated per ISO 10993-7.The React™ 68 Catheter met the acceptance criteria for ethylene chlorohydrin residual.
BioburdenThe React™ 68 Catheter was evaluated per ISO 11737-1.The React™ 68 Catheter met the acceptance criteria for bioburden.
TestTest Method SummaryResults
Bacterial EndotoxinThe ReactTM 68 Catheter wasevaluated per ANSI/AAMIST72 and USP <161>.The ReactTM 68 Catheter metthe acceptance criteria forbacterial endotoxin.
Performance
Visual InspectionThe ReactTM 68 Catheter wasinspected under x2.5magnification.The ReactTM 68 Catheter metthe acceptance criteria forvisual inspection.
Dimensional MeasurementsThe proximal ID, distal ID,proximal OD, distal OD, usablelength, total length, coatinglength, and distal tip length ofthe ReactTM 68 Catheter weremeasured.The ReactTM 68 Catheter metthe acceptance criteria fordimensional measurements.
Tip BucklingThe ReactTM 68 Catheter wasevaluated for the maximumcompressive force.The ReactTM 68 Catheter metthe acceptance criteria for tipbuckling.
Kink ResistanceThe ReactTM 68 Catheter wasevaluated for the maximumkink diameter.The ReactTM 68 Catheter metthe acceptance criteria for kinkresistance.
ParticulateThe ReactTM 68 Catheter wasevaluated per USP <788>.The ReactTM 68 Catheter metthe acceptance criteria forparticulate evaluation.
Coating LubricityThe ReactTM 68 Catheter wasevaluated for the averagefriction forces.The ReactTM 68 Catheter metthe acceptance criteria forcoating lubricity.
Tensile StrengthThe ReactTM 68 Catheter wasevaluated per ISO 10555-1.Annex B.The ReactTM 68 Catheter metthe acceptance criteria fortensile strength at the hub andshaft.
Liquid LeakThe ReactTM 68 Catheter wasevaluated per ISO 10555-1.Annex C.The ReactTM 68 Catheter metthe acceptance criteria forliquid leak.
Corrosion ResistanceThe ReactTM 68 Catheter wasevaluated per ISO 10555-1.Annex A.The ReactTM 68 Catheter metthe acceptance criteria forcorrosion resistance.
Hub Aspiration ResistanceThe ReactTM 68 Catheter wasevaluated per ISO 10555-1.Annex D.The ReactTM 68 Catheter metthe acceptance criteria for hubair aspiration.
CompatibilityThe ReactTM 68 Catheter wasinspected for visual damage ofthe catheter when deliveringand retrieving interventionaldevicesThe ReactTM 68 Catheter metthe acceptance criteria fordelivering and retrievinginterventional devices.
TestTest Method SummaryResults
RadiopacityThe markerband length andwall thickness of the React™ 68 Catheter were measured. Inaddition, radiopacity wasconfirmed via fluoroscopy.The React™ 68 Catheter metthe acceptance criteria forradiopacity.
Luer StandardsThe React™ 68 Catheter wasevaluated per ISO 594-1 andISO 80369-7.The React™ 68 Catheter metthe acceptance criteria for luerstandards.
Torque StrengthThe React™ 68 Catheter wasevaluated for transmission ofproximal torque to the distaltip.The React™ 68 Catheter wasable to withstand torsionalforces that are typical ofclinical use.
Dynamic PressureThe React™ 68 Catheter wasevaluated for the amount ofpressure it can withstand.The React™ 68 Catheter wasable to withstand pressures thatare typical of clinical use.
Coating IntegrityThe React™ 68 Catheter wasevaluated for coating coverageand lubricity.The React™ 68 Catheterremained coated and lubricous.
UsabilityThe React™ 68 Catheter andpredicate device were evaluatedfor maneuverability andflexibility.The React™ 68 Catheter metthe acceptance criteria forusability.

{7}------------------------------------------------

{8}------------------------------------------------

Performance Data - Animal:

Not Applicable. A determination of substantial equivalence is based upon successful completion of non-clinical bench testing as there is no change to the intended use or fundamental scientific technology.

Performance Data - Clinical:

Not Applicable. A determination of substantial equivalence is based upon successful completion of non-clinical bench testing as there is no change to the intended use or fundamental scientific technology.

Conclusion:

The design modifications incorporated do not alter the intended use or fundamental scientific technology.

Non-clinical bench testing supports a determination that the subject React™ 68 Catheter is substantially equivalent to the predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).