K Number
K180578
Date Cleared
2018-12-12

(282 days)

Product Code
Regulation Number
872.3770
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For the fabrication of crowns, bridges and structures for implant supported provisional removable denture and appliance prosthetics.
-Provisional anterior and posterior crowns & bridges

  • Implant and abutment supported prosthetics.
  • Partial, complete and hybrid denture prosthetics (base and teeth) - Removable appliances (splint)
Device Description

Harvest Dental Polymer Blocks are industrially polymerized, pre-colored or clear dental milling discs and blocks designed for milled fabrication of temporary anterior and posterior crowns and bridges (tooth-colored variants) or bite splints and dental CAD/CAM systems.

AI/ML Overview

The Harvest Dental Polymer Blocks device is a dental milling disc and block designed for the CAD/CAM fabrication of temporary anterior and posterior crowns and bridges, as well as provisional removable denture and appliance prosthetics. The submission demonstrates substantial equivalence to its predicate device, Idodentine (K150432), based on similar technological characteristics, indications for use, and performance testing.

1. A table of acceptance criteria and the reported device performance:

RequirementRequired Value (ISO 20795-1)Required Value (ISO 10477)Value (Submission Device)Passed/Failed
Flexural Strength≥ 65 [MPa]≥ 50 [MPa]> 90 MpaPassed
Water Absorption≤ 32 [µg/mm³]≤ 40 [µg/mm³]0.024 mg/mm³Passed
Water Solubility≤ 1.6 [µg/mm³]≤ 7.5 [µg/mm³]0.0000 mg/mm³Passed
Residual Monomer Content≤ 2.2 %-< 1.1 %Passed

2. Sample size used for the test set and the data provenance:

The document does not explicitly state the sample size used for each specific test or the specific provenance (e.g., country of origin, retrospective/prospective) of the testing data beyond stating it was performed in a laboratory.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as the acceptance criteria and performance evaluation for this device are based on objective, standardized physical and chemical property testing rather than expert-derived ground truth from medical images or clinical observations.

4. Adjudication method for the test set:

Not applicable. The performance is assessed against established ISO and ADA standards through laboratory testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a dental material, not an AI-assisted diagnostic tool requiring human reader studies.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

Not applicable. This device is a dental material, not an algorithm. The performance testing is analogous to "standalone" in that it evaluates the material properties directly without human interpretation as part of the primary outcome.

7. The type of ground truth used:

The ground truth used for evaluating the device's performance against its acceptance criteria is based on:

  • ISO Standards: Established international standards for dental materials (ISO 10477, ISO 20795-1).
  • Biocompatibility Standards: ISO 10993 for biological evaluation of medical devices.
  • ADA Standards: ADA 53 for Polymer-Based crown and Bridge Resin.

These standards define the acceptable range or threshold for specific physical, chemical, and biological properties.

8. The sample size for the training set:

Not applicable. This device is a physical dental material with established chemical and physical properties, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

Harvest Dental Products, LLC Sasha Der Avanessian Owner 905 Columbia Street Brea, California 92821

December 12, 2018

Re: K180578

Trade/Device Name: Harvest Dental Polymer Blocks Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown And Bridge Resin Regulatory Class: Class II Product Code: EBG, EBI, MQC Dated: September 14, 2018 Received: September 14, 2018

Dear Sasha Der Avanessian:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Digitally signed by Mary S. Mary S. F Runner - S3 rvically -5:
Runner -S33 - Date: 2018.12.12 14:33:57
Runner -S33 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180578

Device Name Harvest Dental Polymer Blocks

Indications for Use (Describe)

Indications for Use:

For the fabrication of crowns, bridges and structures for implant supported provisional removable denture and appliance prosthetics.

-Provisional anterior and posterior crowns & bridges

  • Implant and abutment supported prosthetics.
  • Partial, complete and hybrid denture prosthetics (base and teeth) - Removable appliances (splint)

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the words "HarvestDental" in a gray sans-serif font. Below the company name are the words "LABORATORY PRODUCTS" in a smaller, gray sans-serif font.

K180578 510(k) Summary

This Summary of 510K safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

510K SubmitterHarvest Dental Products, LLC905 Columbia St, Brea, CA 92821Tel: 714-674-7400hello@harvestdental.com
Contact person:Sasha Der AvanessianPresident & CEOTel: 714-674-7400sasha@harvestdental.com
Date Summary was prepared12/11/18
Trade Name of device:Harvest Dental Polymer Blocks
Common name:Crown & Bridge, Temporary, Resin
Classification name:TEMPORARY CROWN AND BRIDGE RESIN. (21 CFR 872.3770 - Product code EBG)
Classification Product CodeEBG
PanelDental
ClassificationClass II
Predicate DeviceK150432Union Dental S.AIdodentine (Dental Polymer Blank)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the text "HarvestDental" in gray. Below the text is the phrase "LABORATORY PRODUCTS" in a smaller, gray font.

Indications for Use:

For the fabrication of crowns, bridges and structures for implant supported provisional removable denture and appliance prosthetics.

  • Provisional anterior and posterior crowns & bridges
  • Implant and abutment supported prosthetics.
  • Partial, complete and hybrid denture prosthetics (base and teeth)
  • Removable appliances (splint)

Device Description:

Harvest Dental Polymer Blocks are industrially polymerized, pre-colored or clear dental milling discs and blocks designed for milled fabrication of temporary anterior and posterior crowns and bridges (tooth-colored variants) or bite splints and dental CAD/CAM systems.

Technological characteristics

The technologic characteristics are highly similar as demonstrated in performance testing and in chemical composition; both devices are composed primarily of polymethylmethacrylate, while the amount and percentage of color oxides in the submission device varies from the predicate.

Comparison of Required Technology Characteristics

The following table shows a summary of the technological characteristics of DD medical polymers compared to the predicate device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the text "HarvestDental" in gray. Below "HarvestDental" is the text "LABORATORY PRODUCTS" also in gray.

FeatureSubmission devicePredicate device
Trade nameHarvest Dental Polymer BlankIdodentine Disc
510 (K)K180578K150432
Product codeEBGEBG
Regulatory ClassCLASS IIClass II
ManufacturerHarvest DentalUnion Dental S.A
Indications for useFor the fabrication of crowns, bridges and structures for implant supported provisional removable denture and appliance prosthetics. - Provisional anterior and posterior crowns & bridges. - Implant and abutment supported prosthetics. - Partial, complete and hybrid denture prosthetics (base and teeth). - Removable appliances (splint)Acrylic polymer blank particularly suitable for making removable or temporary dental structures such as crowns and bridges using milling technology using CAD/CAM. Indications - Temporary anterior and posterior crowns - Temporary anterior and posterior bridges with up to two adjacent pontics Maximum recommended usage period: 12 months - Removable structures for dentures (denture bases) - Removable structures for therapeutic restorations (bite splints or occlusal splints)
TechnologyBlank for dental CAD/CAMBlank for dental CAD/CAM
ShapeDisc or BlockDisc or Block
ShadeVITA-shades, clear, pinkVITA-shades, clear, pink
Raw MaterialPMMAPMMA
Chemical Compositions [Units]
Material BasePolymethyl methacrylatePolymethyl methacrylate
Coloring oxides< 1,1%Not specified
Physical Characteristics [Units]
Performance TestingISO 10477ISO 20795Tested according to DIN EM ISO 20795-1 and ISO 10477
BiocompatibilityISO 10993No cytotoxicE.N ISO 10933-1, -5,

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features the word "Harvest" in gray, with the word "Dental" in green next to it. Above the words is a green leaf. Below the words "HarvestDental" are the words "LABORATORY PRODUCTS" in gray.

Preclinical testing and standards Clinical and Non clinical performance

The biocompatibility test for "Harvest Dental Polymer Blocks" was made in the laboratory and all the results was:

StandardConclusionTraceability
ISO 10993-10, 2010.In the test conditions, the testsubstance Harvest DentalPolymer Blocks was classifiedas no sensitizing.Report Number:RL15885/2018LLNA-BStudy Number:15885/2018LLNA
ISO 10993-10, annex B32010.Under the study conditions, thetest substance Harvest DentalPolymer Blocks was classified asnon-irritant to oral mucose ofhamsters, accordingtoISO10993: 10, 2010.Report Number:RL15886/2018IMO-BStudy Number:15886/2018IMO
ISO 10993-5, 2009 annex CUnder the conditions of thisstudy, the test substanceHarvest Dental Polymer Blocksdid not promote reduction of cellviability higher than 30%.Therefore the test substanceHarvest Dental Polymer Blockswas not cytotoxic in this assay.Report Number:RL16016/2018CT-BStudy Number:16016/2018CT

This PMMA has been tested for biocompatibility and physical properties. This PMMA used has been rated as biocompatible and non-toxic and this PMMA material also met the applicable requirements of ISO 10447(crown & bridges material), and ISO 22112(polymer teeth) standards. The subject device has passed testing according to ISO 10477/2004: DENTISTRY - POLYMERBASED CROWN AND BRIDGE MATERIALS. This international standard specification, requirements and test methods for the polymer-based crown and bridge materials. The subject device also has passed testing according to ISO 10993-5 and has been rated as Non cytotoxic.

Standards:

ISO 10477:2004 – Dentistry Polymer-Based crown and Bridge Materials ISO 22112:2005 - Polymer teeth ADA - 53 - Polymer-Based crown and Bridge Resin

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the text "HarvestDental" in gray. Below "HarvestDental" is the text "LABORATORY PRODUCTS" in a smaller, gray font.

Requirements acc. to DIN EN ISO 20795-1 / ISO 10477:
RequirementRequired Value(ISO20795-1)Required Value(ISO10477)ValueSubm.DeviceValuePredicateDevicePassed/Failed
Flexuralstrength≥ 65[MPa]≥ 50[MPa]> 90Mpa90 [MPa]Passed
Waterabsorption≤ 32[µg/mm³]≤ 40[µg/mm³]0.024mg/mm³23[µg/mm³]Passed
Watersolubility≤ 1.6[µg/mm³]≤ 7.5[µg/mm³]0.0000mg/mm³0.0[µg/mm³]Passed
Residualmonomercontent [%]≤ 2.2 %-< 1,1%1.4 %Passed

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the text "HarvestDental" in gray. Below "HarvestDental" is the text "LABORATORY PRODUCTS" in a smaller, gray font.

Substantial Equivalence:

The differences in indications for use are limited to the following wording differences as follows:-

Submission DevicePredicate deviceDifferencesDiscussion
Fabrication of crowns,bridgesSame
Same
Structures for implantsupported provisionalremovable denture andappliance prosthetics.TemporaryThe use of the wordprovisional versesTemporaryNote 1
Provisional anterior andposterior crowns & bridges.Samewith up to twoadjacentponticsNote 2
Implant and abutmentsupported prosthetics.Samebite splintsor occlusal splintsNote 3
- Partial, complete andhybrid denture prosthetics(base and teeth).Same
Maximumrecommended usageperiod: 12 monthsSubmission doesn'tuse these wordsNote 1
Acrylic polymer blankSubmissiondoesn'tuse these wordsNote 4
using millingtechnology usingCAD/CAM.Submission doesn'tuse these wordsNote 5

Note 1

These two words can often be used interchangeably, however not necessarily meaning the same thing in true context. For example: -

The Predicate device uses "Temporary" meaning the device is expected to come to an end, in their case at 12 months.

The Submission device uses the word "Provisional" meaning that it will come to an end but doesn't mean that it has finished it's useful life.

The reason we chose the word "Provisional" is to establish that our device can actually last a lot longer than 12 months as indicated in our "Shelf life report" and our description of device, which shows the device can actually remain in place for 5 years. This is supported by the test data submitted in this review. Again this use of the word has no adverse impact on the safety or effective performance of the submitted device.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the words "HarvestDental" in gray. Below "HarvestDental" are the words "LABORATORY PRODUCTS" in a smaller, gray font.

Note 2

Pontics are used to replace missing teeth, however the primary causes of complications in implant dentistry are related to biomechanics. For example, early loading failures especially in soft bone, when forces are greater than usual and/or implant sizes are shorter than 10mm.

The number of pontics for a prosthesis is related to strength of the abutments and the number of pontics with implants is often increased compared to natural teeth which can cause a problem.

In fixed prosthesis design, three adjacent pontics in the posterior reqions of the mouth are contraindicated with natural. The adjacent abutments are subjected to considerable additional force when they must support three missing teeth, especially in the posterior regions of the mouth.

Also the occlusal force to the five posterior teeth cannot be supported by two implants. The posterior reqions have greater bite forces than the anterior reqions. This is because typically the teeth in the posterior have less bone density, which puts an increased load on the anterior teeth

The additional forces distributed to the pontic abutments will cause them to flex under load. The greater the span the greater the flex on the pontic causing potential mechanical shear and tensile load failure on the abutments.

The chart below shows the additional flex placed on a pontic assuming a load of 25lbs and all other variables beinq equal.

# of PonticsFlexAccept/Reject
18 μ - 0.01 mmAccept
264 μ - 0.08 mmAccept
3216 μ - 0.216 mmReject

Flex is not so much an issue on natural teeth as they have a small degree of natural flex, apically and laterally in the area of the root. However an implant is more ridged due to its increased modulus of elasticity (Doesn't flex as much as a natural tooth).

Further more angled forces increase the stress and strain on an implant restoration especially in the maxillary anterior teeth. This can lead to porcelain fractures, abutment screws working loose and uncementing of the restoration causing even more stress on the restoration. More stress forces can cause the pontic to cantilever, creating an increased risk of crestal bone loss, failure of the implant or fracture of an abutment screw or implant body, especially in the posterior region due to greater bite force.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the words "HarvestDental" in gray. Below "HarvestDental" are the words "LABORATORY PRODUCTS" in a smaller, gray font.

Pontics should be limited to 2 in proportion to the size of the premolars, typically between 13.5mm - 16mm. There are instances when a molar is one of the missing teeth to be replaced, which is between existing teeth creating a space between 10-15mm. In the instance when the space is greater than 15mm it is better for the doctor to use two pontics to replace the molar. Therefore when a 200 premolar and 136 molar are missing it would require the doctor to plan the replacement of 3 teeth rather than two as the span is related to the number of roots in the mandible and buccal roots in the maxilla.

By using 3 pontics or more in either a fixed prosthesis or implant restoration is contraindicative and a skilled restoration specialized would always design a restoration using a maximum of 2 pontics where possible or use more implants.

Therefore to either state or not state 2 pontic on the Substantial Equivalence indications has no impact on safety or effectiveness of the submitted device, as this is a decision of the specialist doctor when designing the Provisional anterior and posterior crowns & bridges.

Note 3

We have deliberately left out the need to use the Polymer block for the use of bite splints or occlusal splints, even though this material can actually be used for both we will market a separate material to cover this need. This has no impact on the submitted device for safety or effectiveness.

Note 4

The use of the word "Acrylic Polymer Blank" is not required as this is already a given. This has no impact on the submitted device for safety or effectiveness.

Note 5

The use of the words "using milling technology using CAD/CAM is not required as this is also a qiven. This has no impact on the submitted device for safety or effectiveness.

Conclusion

Information provided in this 510(k) submission shows that the product is substantially equivalent to the Harvest Dental Polymer Block as predicate device Idodentine Blank.

Based on comparison of technology, including composition and performance testing, biocompatibility testing, and highly similar indications for use, Harvest Dental Polymer Blocks are equivalent to Idodentine (K150432).

§ 872.3770 Temporary crown and bridge resin.

(a)
Identification. A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.(b)
Classification. Class II.