K Number
K180559

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-10-29

(242 days)

Product Code
Regulation Number
866.3309
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HSV 1&2 ELITe MGB® Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of Herpes Simplex Virus 1 and 2 (HSV-2) DNA in cutaneous or mucocutaneous lesion swab specimens from patients with signs and symptoms of HSV-2 infection. This test is an aid in the differential diagnosis of HSV-1 and HSV-2 infections.

The HSV 1&2 ELITe MGB Assay is not FDA cleared for use with cerebrospinal fluid (CSF) specimens. The assay is not intended to be used for prenatal screening or for screening blood or blood products.

Device Description

The HSV 1&2 ELITe MGB Assay is a qualitative in vitro diagnostic Real-Time PCR Assay for the direct detection of Herpes Simplex Virus (HSV) DNA (glycoprotein D gene for HSV-1 and glycoprotein G gene for HSV-2) in symptomatic male and female patients using DNA purified from swab specimens collected from individuals with cutaneous or mucocutaneous herpetic lesions.

The HSV 1&2 ELITe MGB Assay system is comprised of three major processes: (1) automated preparation of unprocessed sample to extract nucleic acids from primary swab specimens using the ELITe InGenius SP 200 Extraction Cartridge, (2) PCR amplification of target DNA sequences using HSV-1 and HSV-2 specific primers, and (3) real-time detection of fluorescent-labeled HSV-1 and HSV-2 specific oligonucleotide detection probes.

An Internal Control (IC), containing unrelated randomized DNA sequence, is added to all samples prior to extraction and monitors the integrity of the reagents, equipment function, and the presence of inhibitors in the samples. A positive signal in the Internal Control channel in the absence of HSV DNA indicates that the PCR has not been inhibited.

The amplification reagents, Positive Control and Internal Control are packaged as part of the HSV 1&2 ELITe MGB Assay.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the HSV 1&2 ELITe MGB Assay, based on the provided text:

Acceptance Criteria and Reported Device Performance

The acceptance criteria for the HSV 1&2 ELITe MGB Assay are presented in terms of Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) against a composite reference method.

MetricAcceptance Criteria (Implied by achieved performance and regulatory clearance)Reported Device Performance (95% CI) - Cutaneous Lesion SamplesReported Device Performance (95% CI) - Mucocutaneous Lesion Samples
HSV-1 PPAHigh (e.g., typically >90-95%)98.7% (93.2-99.8%)99.2% (95.7-99.9%)
HSV-1 NPAHigh (e.g., typically >90-95%)98.5% (96.9-99.3%)97.6% (95.8-98.6%)
HSV-2 PPAHigh (e.g., typically >90-95%)96.2% (91.3-98.3%)97.6% (94.0-99.1%)
HSV-2 NPAHigh (e.g., typically >90-95%)98.6% (96.9-99.3%)98.2% (96.6-99.1%)

Study Information

  1. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: 1,174 prospectively collected archived swab samples were initially collected. After exclusions:
      • 1,171 samples were analyzed for HSV-1.
      • 1,170 samples were analyzed for HSV-2.
      • A separate contrived oral panel study involved 75 individual negative cheek swab samples spiked with HSV-2, plus 10 HSV-1 positive samples and 15 HSV-1/HSV-2 negative oral samples.
    • Data Provenance: The samples were "left-over prospectively collected archived swab samples" from symptomatic patients. The country of origin is not explicitly stated, but the context of an FDA submission generally implies US-based data.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • The document does not explicitly state the number of experts or their qualifications for establishing the ground truth. The ground truth was established using a composite reference method.
  3. Adjudication Method for the Test Set:

    • The adjudication method for the test set was a "composite reference method" defined as: "an FDA cleared assay and a validated HSV 1&2 PCR followed by bi-directional sequencing of gel electrophoresis-positive samples."
    • "A positive result by the composite reference method is defined as a positive by the FDA cleared PCR or the validated sequencing. Two negative results are needed to confirm a negative." This implies a type of adjudicted consensus, where the "truth" is determined by agreement or hierarchical resolution between different methods, rather than human experts per se.
  4. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    • No. This document describes the analytical and clinical performance of an in vitro diagnostic (IVD) assay, which is an algorithm-only (standalone) test. It does not involve human readers in the primary diagnostic decision-making process that would warrant an MRMC study comparing human performance with and without AI assistance.
  5. If a Standalone Performance Study Was Done:

    • Yes. The entire clinical performance section (Section 11.b) is a standalone performance study where the HSV 1&2 ELITe MGB Assay's performance is compared directly against a composite reference method. The assay operates without human intervention in the interpretation of results.
  6. The Type of Ground Truth Used:

    • Composite Reference Method: This ground truth combined an FDA-cleared PCR assay and a validated HSV 1&2 PCR followed by bi-directional sequencing. This type of ground truth integrates highly sensitive and specific molecular methods, reflecting a robust laboratory-based gold standard.
  7. The Sample Size for the Training Set:

    • The document does not explicitly mention a "training set" in the context of an AI/ML algorithm. This device is a real-time PCR assay, which is a chemical and enzymatic reaction rather than a machine learning model that would typically require a separate training set. The various analytical performance studies (Limit of Detection, Analytical Reactivity, Specificity, etc.) serve to characterize the assay's performance across different conditions, which is analogous to how a system is "trained" or refined during development, but not in the sense of a distinct machine learning training set. The "assay cut-off" was established using a separate set of 141 clinical samples, which could be considered a form of calibration/training for the threshold of positivity.
  8. How the Ground Truth for the Training Set Was Established:

    • As noted above, a traditional "training set" for an AI model is not applicable here. For the establishment of the assay cut-off (using 141 clinical samples), the ground truth was established using the same "composite reference method" (FDA-cleared real-time PCR assay combined with PCR amplification and bidirectional sequencing). For other analytical studies (like LoD, inclusivity, cross-reactivity), quantitated viral strains or characterized organisms were used as the reference.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 29, 2018

ELITechGroup Walt Mahoney Regulatory Affairs Manager 21720 23rd Drive SE Suite 150 Bothell, Washington 98021

Re: K180559

Trade/Device Name: HSV 1 & 2 ELITe MGB Kit; ELITe InGenius Regulation Number: 21 CFR 866.3309 Regulation Name: Herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel Regulatory Class: Class II Product Code: PGI, OOI Dated: January 17, 2018 Received: March 1, 2018

Dear Walt Mahoney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Steven R. Gitterman -S for

Uwe Scherf, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180559

Device Name HSV 1&2 ELITe MGB Assay

Indications for Use (Describe)

HSV 1&2 ELITe MGB Assay

The HSV 1&2 ELITe MGB® Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of Herpes Simplex Virus 1 and 2 (HSV-2) DNA in cutaneous or mucocutaneous lesion swab specimens from patients with signs and symptoms of HSV-2 infection. This test is an aid in the differential diagnosis of HSV-1 and HSV-2 infections.

The HSV 1&2 ELITe MGB Assay is not FDA cleared for use with cerebrospinal fluid (CSF) specimens. The assay is not intended to be used for prenatal screening or for screening blood or blood products.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

ELITe MGB HSV 1&2 Assay on the ELITe InGenius system

1.Date:September 25th, 2018
2.Submitter:ELITechGroup Inc. Molecular Diagnostics
21720 23rd Drive SE, Suite 150
Bothell, WA 98021
3.Contact Person:Walt Mahoney
21720 23rd Dr. SE, Suite 150
Bothell, WA 98021
Phone:425-482-5173
Fax:425-482-5550
Email:w.mahoney@elitechgroup.com
4.Device Description:K180559HSV 1&2 ELITE MBG AssayClass IIPGI21CFR 866.3309
5.Predicate Devices:K151906LUMINEX CorporationARIES® HSV 1&2 Assay

{4}------------------------------------------------

Intended Use 6.

HSV 1&2 ELITe MGB Assay The HSV 1&2 ELITe MGB® Assay is a real-time polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection and differentiation of Herpes Simplex Virus 1 and 2 (HSV-1 and HSV-2) DNA in cutaneous or mucocutaneous lesion swab specimens from patients with signs and symptoms of HSV-1 or HSV-2 infection. This test is an aid in the differential diagnosis of HSV-1 and HSV-2 infections.

The HSV 1&2 ELITe MGB Assay is not FDA cleared for use with cerebrospinal fluid (CSF) specimens. The assay is not intended to be used for prenatal screening or for screening blood or blood products.

Special instrument requirements: for use on the ELITe InGenius system

7. Device Descriptions

HSV 1&2 ELITe MGB Assay

The HSV 1&2 ELITe MGB Assay is a qualitative in vitro diagnostic Real-Time PCR Assay for the direct detection of Herpes Simplex Virus (HSV) DNA (glycoprotein D gene for HSV-1 and glycoprotein G gene for HSV-2) in symptomatic male and female patients using DNA purified from swab specimens collected from individuals with cutaneous or mucocutaneous herpetic lesions.

The HSV 1&2 ELITe MGB Assay system is comprised of three major processes: (1) automated preparation of unprocessed sample to extract nucleic acids from primary swab specimens using the ELITe InGenius SP 200 Extraction Cartridge, (2) PCR amplification of target DNA sequences using HSV-1 and HSV-2 specific primers, and (3) real-time detection of fluorescent-labeled HSV-1 and HSV-2 specific oligonucleotide detection probes.

An Internal Control (IC), containing unrelated randomized DNA sequence, is added to all samples prior to extraction and monitors the integrity of the reagents, equipment function, and the presence of inhibitors in the samples. A positive signal in the Internal Control channel in the absence of HSV DNA indicates that the PCR has not been inhibited.

The amplification reagents, Positive Control and Internal Control are packaged as part of the HSV 1&2 ELITe MGB Assay.

8. Substantial Equivalence Information

Device Name: HSV 1&2 ELITe MGB Assay Predicate Device Name: Aries HSV 1&2 Assay, 510k number: K151906

Comparison with predicate

{5}------------------------------------------------

Similarities
ItemNew DeviceHSV 1&2 ELITE MGB AssayPredicate DeviceAries HSV 1&2 Assay, K151906
IntendedUse/Indications forUseThe HSV 1&2 ELITE MGB® Assay is areal-time polymerase chain reaction(PCR) based qualitative in vitrodiagnostic test for the direct detectionand differentiation of Herpes SimplexVirus 1 and 2 (HSV-1 and HSV-2) DNAin cutaneous or mucocutaneous lesionswab specimens from patients withsigns and symptoms of HSV-1 or HSV-2 infection. This test is an aid in thedifferential diagnosis of HSV-1 andHSV-2 infections.The HSV 1&2 ELITE MGB Assay is notFDA Cleared for use with cerebrospinalfluid (CSF) specimens. The assay is notintended to be used for prenatalscreening or for screening blood orblood products.The ARIES® HSV 1&2 Assay is a real-time polymerase chain reaction (PCR)based qualitative in vitro diagnostic testfor the direct detection anddifferentiation of Herpes Simplex Virus1 and 2 (HSV 1 and HSV 2) DNA incutaneous or mucocutaneous lesionspecimens from symptomatic patients.The test is indicated for use as an aid indiagnosis of HSV infection insymptomatic patients. The ARIES®HSV 1&2 Assay is indicated for use onthe ARIES® System.WARNING: The ARIES® HSV 1&2Assay is not FDA cleared for use withcerebrospinal fluid (CSF). The assay isnot intended to be used for prenatalscreening.
SpecimentypesMale and female cutaneous andmucocutaneous lesion swab specimensSame
MajorTechnologyQualitative real-time PCRSame
SampleextractionmethodAutomated sample extractionSame
AssayresultsQualitative detection and differentiationof HSV-1 and HSV-2 DNASame

Differences

Differences
ParameterNew DeviceHSV 1&2 ELITe MGB AssayPredicate DeviceAries HSV 1&2 Assay, K151906
Sample extraction &AmplificationInstrumentationELITe InGenius systemARIES system
AnalytemeasuresDNA sequences from HSV-1glycoprotein D gene and HSV-2glycoprotein G gene.DNA sequences from Herpes SimplexVirus type 1 (HSV-1) and HerpesSimplex Virus type 2 (HSV-2)
DetectionMethodMultiplex assay with paired reporterand quencher fluorescence labeledprobes and different reporter dyesfor each target. Measures increasein assay fluorescence with eachPCR cycle.Pairs fluorescent-labeled primers withquencher labeled nucleotides.Measures decrease in assayfluorescence with each PCR cycle.

{6}------------------------------------------------

9. Instrumentation/Software:

The HSV 1&2 ELITe MGB Assay is used on the ELITe InGenius system. The ELITe InGenius system is a bench top instrument integrating all required hardware, reagent and software components to perform nucleic acid sample preparation and real-time PCR operations.

The ELITe InGenius system can process from 1 to 12 samples in 12 parallel tracks, and samples may be loaded in primary tubes or in secondary tubes provided. The system utilizes a universal, cassette-based process for sample extraction, and allows for multiple and independent PCRs to be performed from a single nucleic acid eluate.

The system can operate in three different modes: nucleic acid extraction only, PCR amplification only, or nucleic acid extraction with PCR amplification.

10 Performance Characteristics - Analytical Performance

a. Limit of Detection

The analytical sensitivity of the HSV 1&2 ELITe MGB Assay was determined using 4 HSV strains (two for each target). Quantitated viral strains were obtained and diluted with HSV neqative pooled human cheek matrix in Universal Transport Medium (UTM) to values spanning the range of approximately .05 to 1000 TCIDsg/mL (depending on the strain). All dilutions were tested, and the limit of detection (LoD) was determined using Probit (Logit) Data Analysis software (Analyse-it for Microsoft Excel v4.80.2, Logistic Function model). LoD for each strain represents the lowest viral titer in TCID50/mL at which a positive result will be obtained with at least 95% confidence. LoD for each strain was then verified by testing at least 20 replicates. Results indicate that, depending on the strain, the HSV 1&2 ELITe MGB Assay will produce a positive result with 95% confidence for a swab eluate containing 59 (HSV-1; HSV-1 MacIntyre Strain) and/or 5.4 (HSV-2; HSV-2 MS Strain) TCIDg6/mL.

OrganismIsolate/StrainCellLineQualitativeresults#detected/TotalMean CT ±SDfrom detectedreplicates1×LoDTCID50/mL
HSV-1MacIntyrestrainVero20/2037.91 ± 0.6959.0TCID50/mL
HSV-1Isolate #15(Zeptometrix)Vero20/2039.94 ± 0.951.5TCID50/mL
HSV-2MS strainVero20/2037.90 ± 0.925.4TCID50/mL
HSV-2Isolate #2(Zeptometrix)Vero20/2038.67 ± 1.030.3TCID50/mL

Limit of Detection Results

A swab elution efficiency study was also performed by using the same HSV negative pooled human cheek swab matrix and the HSV-1 MacIntyre strain used to verify the LoB. This study showed 100% elution efficiency from the Copan reqular flocked swab compared with the same volume of material directly spiked into UTM. Therefore, LoD values in TCID56/mL units will be directly proportional (with a constant = 1) to the LoD in TCIDs /swab units, depending only on the volume of the media in the collection device.

b. Assay Cut Off

{7}------------------------------------------------

The assay cut-off analysis was performed on a separate set of 141 clinical samples collected from 3 clinical sites. Each clinical sample was evaluated using HSV 1&2 ELITe MGB Assay in conjunction with the ELITe InGenius instrument and a composite reference method (FDA-cleared real-time PCR assay combined with PCR amplification and bidirectional sequencinq). Both targets in clinical samples were detected up to cycle 45. Therefore C- of 45 was established as a diagnostic assay cut-off for both HSV-1 and HSV-2 targets.

c. Analytical Reactivity (Inclusivity)

Performance of the HSV 1&2 ELITe MGB Assay was tested on 44 well characterized HSV-1 and HSV-2 isolates, which were obtained through commercial means. All strains were tested with the HSV 1&2 ELITE MGB Assay as spiked samples in UTM that had been prepared at 3×LOD level (177 TCIDsp/mL for HSV-1 and 16.2 TCID50/mL for HSV-2). For HSV-1 and HSV-2 viral isolates not detected (negative) at 3×LoD concentrations, 2× incremental concentrations were tested, and the lowest detectable level was determined, and the final test concentration is reported. All of the HSV-1 and HSV-2 tested isolates were detected by the HSV 1&2 ELITe MGB Assay at concentrations of 16.2 - 354 TCID50/mL

#IsolateEstimated1×LoD(TCID50/mL)×LoDTestedFinal TestConc.(TCID50/mL)Positivity
1HSV-1 MacIntyreStrain591773/3
2HSV-1 Isolate #1ട് കરૂપ્177ર્ડારે
3HSV-1 Isolate #2રેત્વે છે. વિત્તર પ્રદેશના પશુપાલન છે. આ ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં મુખ્યત્વે ખેત-ઉત177રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
4HSV-1 Isolate #3ട് വ1773/3
5HSV-1 Isolate #4ട്ടു177ર્ડારે
6HSV-1 Isolate #5રેત્વે છે.રૂપ્ર177ર્ડારે
7HSV-1 Isolate #659રૂપ્ર1770/3
ನಿರಿ354રાજ
8HSV-1 Isolate #7રેત્વે છે. વિત્તર પ્રદેશના પશુપાલન છે. આ ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં મુખ્યત્વે ખેત-ઉત177ર્ડારે
9HSV-1 Isolate #859177ર્ડારે
10HSV-1 Isolate #959રૂપ્ર177ર્ડારે
11HSV-1 Isolate #10591773/3
12HSV-1 Isolate #11591773/3
13HSV-1 Isolate #12રેત્વે છે.177ર્ડારે
14HSV-1 Isolate #13ട് ഒര1773/3
15HSV-1 Isolate #14591773/3
16HSV-1 Isolate #15591773/3
17HSV-1 Isolate #1659177રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
18HSV-1 Isolate #17રેત્વે છે.177ર્ડારે
19HSV-1 Isolate #1859177રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
20HSV-1 Isolate #19ട് ക177રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
21ട്ടു1770/3
HSV-1 Isolate #2059354રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
22HSV-1 Isolate #21ട് ഒരરૂપ્ર177રાજ
23HSV-2 MS Strain5.416.2રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
24HSV-2 Isolate #15.416.2રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
25HSV-2 Isolate #25.416.2ર્ડારે
26HSV-2 Isolate #35.4રજ16.23/3

Summary of Inclusivity Results

{8}------------------------------------------------

27HSV-2 Isolate #45.4રૂપ્16.23/3
28HSV-2 Isolate #55.416.2રીકે
29HSV-2 Isolate #65.416.2રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
30HSV-2 Isolate #75.4રૂપ્16.2રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
5.416.22/3
31HSV-2 Isolate #85.4રૂપ્16.23/3
32HSV-2 Isolate #95.416.2રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
33HSV-2 Isolate #105.416.2ર્ડારે
345.416.22/3
HSV-2 Isolate #115.432.4રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
355.416.2113
HSV-2 Isolate #125.432.43/3
5.4રૂપ્16.20/3
5.432.42/3
36HSV-2 Isolate #135.412×64.82/3
5.424×129.63/3
5.416.2113
37HSV-2 Isolate #145.432.4રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
38HSV-2 Isolate #155.416.20/3
5.432.4રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
39HSV-2 Isolate #165.416.2113
5.432.4રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
HSV-2 Isolate #175.416.2113
405.432.4113
5.412x64.8રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
41HSV-2 Isolate #185.4રૂપ્16.2રીકે
42HSV-2 Isolate #195.4રૂપ્16.22/3
5.432.4113
5.412×64.8રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના
5.416.20/3
43HSV-2 Isolate #205.432.4113
5.412×64.83/3
44HSV-2 Isolate #215.4રૂપ્16.2રૂડિયારે તે જેવી સાચના સાચના દિવેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના

d. Analytical Specificity (Cross-Reactivity)

To determine if there are any organisms that could be potential cross-reactive with the HSV 1&2 ELITE MGB Assay the NCBI BLAST in silico (computer) analysis of the HSV-1 and HSV-2 amplicons was performed. Found sequence homology in both the primer and probe regions was not sufficient for HSV1&2 ELITe MGB Assay to potentially cross-react with the non-HSV organisms.

Potential cross-reactivity of the HSV 1&2 ELITe MGB Assay was then evaluated by testing varied species of organisms that are closely related to HSV or cause similar clinical symptoms or may be present in the anogenital and oral cutaneous and mucocutaneous sites tested by this device. 49 potential cross reactants were evaluated. For each organism, the sample to be tested was prepared from quantified stock diluted to the required concentration using UTM.

The potential cross reactants were tested, the concentrations were evaluated and the results are presented in the following table:

No.Potential Cross-ReactantsTestedConcentrationQualitative Result(#Detected/#Total)
---------------------------------------------------------------------------------------------------

{9}------------------------------------------------

HSV-1HSV-2
1Acinetobacter calcoaceticus1×106 CFU/mL0/30/3
2Acinetobacter lwoffi1×106 CFU/mL0/30/3
3Adenovirus type 21×105 TCID50/mL0/30/3
4Bacteroides fragilis1×106 CFU/mL0/30/3
5Candida albicans1×106 CFU/mL0/30/3
6Candida glabrata1×106 CFU/mL0/30/3
7Candida guilliermondii1×106 CFU/mL0/30/3
8Candida krusei1×106 CFU/mL0/30/3
9Candida lustaniae1×106 CFU/mL0/30/3
10Candida parapsilosis1×106 CFU/mL0/30/3
11Candida tropicalis1×106 CFU/mL0/30/3
12Chlamydia trachomatis1×106 CFU/mL0/30/3
13Cytomegalovirus1×105 TCID50/mL0/30/3
14Enterobacter cloacae1×106 CFU/mL0/30/3
15Enterovirus1×105 TCID50/mL0/30/3
16Epstein-Barr Virus1×105 TCID50/mL0/30/3
17Escherichia coli1×106 CFU/mL0/30/3
18Fusobacterium nucleatum1×106 CFU/mL0/30/3
19Gardnerella vaginalis1×106 CFU/mL0/30/3
20Haemophilus ducreyi1×106 CFU/mL0/30/3
21Human Genomic DNA500 ng/swab0/30/3
22Human Herpes Virus 61×105 TCID50/mL0/30/3
23Human Herpes Virus 71×105 TCID50/mL0/30/3
24Human papilloma virus 161×105 TCID50/mL0/30/3
25Human papilloma virus 181×105 TCID50/mL0/30/3
26Herpes Simplex Virus 1(HSV-1), isolate 20, ZMC1×105 TCID50/mL3/30/3
27Herpes Simplex Virus 2(HSV-2), isolate 20. ZMC1×105 TCID50/mL0/33/3
28Klebsiella pneumoniae1×106 CFU/mL0/30/3
29Lactobacillus acidophilus1×106 CFU/mL0/30/3
30Mobiluncus curtsii1×106 CFU/mL0/30/3
31Mobiluncus mulieris1×106 CFU/mL0/30/3
32Moraxella catarrhalis1×106 CFU/mL0/30/3
33Mycoplasma hominis1×106 CFU/mL0/30/3
34Neisseria gonorrhea1×106 CFU/mL0/30/3
35Neisseria meningitides1×106 CFU/mL0/30/3
36Prevotella melaninogenica1×106 CFU/mL0/30/3
37Rubella Virus1×105 TCID50/mL0/30/3
38Staphylococcus aureus(MSSA)1×106 CFU/mL0/30/3
39Staphylococcus epidermidis(MRSE)1×106 CFU/mL0/30/3
44Staphylococcussaprophyticus1×106 CFU/mL0/30/3
41Streptococcus mitis1×106 CFU/mL0/30/3
42Streptococcus mutans1×106 CFU/mL0/30/3
43Streptococcus pneumoniae1×106 CFU/mL0/30/3
44Streptococcus pyogenes1×106 CFU/mL0/30/3
45Streptococcus salivarius1×106 CFU/mL0/30/3
46Toxoplasma gondii1×106 CFU/mL0/30/3
47Trichomonas vaginalis1×106 CFU/mL0/30/3
48Varicella-Zoster Virus (VZV)1×105 TCID50/mL0/30/3
49Chlamydophila pneumoniae$1 \times 10^6$ CFU/mL0/30/3

{10}------------------------------------------------

Analytical Specificity (Microbial Interference) e.

The microbial interference was evaluated in the presence of either HSV-1 or HSV-2 spiked at 3×LoD in UTM and the 49 organisms indicated in the table above. Each microorganism was tested either at 1×10° CFU/mL or higher for bacterial isolates, or at 1×10° TCID30/mL or higher for viruses. None of the non-target organisms that are reasonably expected to be found in cutaneous and mucocutaneous swab samples interfered with the detection of HSV-1 or HSV-2 species. The only exception is that amplification of HSV-1 is completely inhibited in the presence of HSV-2 in titers of 1×103 or higher. This observed HSV-2 interference is reported as a limitation in the device package insert.

Competitive Interference of HSV-1 and HSV-2 f.

Competitive interference was studied to evaluate the effects of possible clinically relevant coinfection with both HSV-1 and HSV-2 using HSV 1&2 ELITe MGB Assay.

The study assessed whether a high concentration of one virus in the sample could potentially affect the HSV 1&2 ELITe MGB Assay performance for the other target present at low levels. A low positive sample was contrived at approximately 3×LoD for each target (HSV-1 MacIntyre strain and HSV-2 MS strain), and a baseline Ct was determined for each sample. Each potential concomitant infecting virus was spiked into the low level sample and assaved triplicate. in

Competitive interference of HSV-1 with HSV-2 was observed at 1×10°, 1×10°, 1×10° TCIDss/mLHSV-2 level. This observed interference information is included as a limitation in the device package insert.

No competitive interference of HSV-2 with HSV-1 at any level was observed. The results of the testing are shown in the table below.

Baseline (Low Level)Competitive Interferent(High Concentration)Qualitative Results(#Detected/#Total)
StrainConcentration(TCID50/mL)StrainConcentration(TCID50/mL)HSV-1HSV-2
HSV-1 MacIntyre177HSV-2 MS1000000/33/3
HSV-1 MacIntyre177HSV-2 MS100001/33/3
HSV-1 MacIntyre177HSV-2 MS10001/33/3
HSV-1 MacIntyre177HSV-2 MS1003/33/3
HSV-1 MacIntyre177HSV-2 MS03/30/3
HSV-2 MS16.2HSV-1 MacIntyre1000003/33/3
HSV-2 MS16.2HSV-1 MacIntyre100003/33/3
HSV-2 MS16.2HSV-1 MacIntyre10003/33/3
HSV-2 MS16.2HSV-1 MacIntyre1003/33/3
HSV-2 MS16.2HSV-1 MacIntyre00/33/3

Competitive Interference of HSV-1 and HSV-2 targets at unequal concentrations

Additionally, in a separate study both strains were tested at similar or equal concentrations of 3×LoD, 1×103 and 1×105, and no competitive interference was observed.

HSV-1 ConcentrationHSV-2 ConcentrationQualitative Results(#Detected/#Total)Quantitative Results (%CV)
-----------------------------------------------------------------------------------------------------------------

{11}------------------------------------------------

StrainTCID50/mLStrainTCID50/mLHSV-1HSV-2HSV-1HSV-2
HSV-1MacIntyre$1\times10^5$HSV-2 MS$1\times10^5$5/55/53.02 %1.64 %
HSV-1MacIntyre$1\times10^3$HSV-2 MS$1\times10^3$5/55/51.09 %2.95 %
HSV-1MacIntyre177 (3×LoD)HSV-2 MS16.2 (3×LoD)5/55/51.74 %1.88 %

g. Interfering Substances

The performance of the HSV 1&2 ELITe MGB Assay was evaluated with potentially interfering substances that could be encountered in lesion swab specimens obtained from cutaneous and mucocutaneous locations. A total of 33 substances were individually spiked into negative UTM matrix containing one of each HSV-1 and HSV-2 isolates at 3X LoD level and tested in triplicate with the HSV 1&2 ELITe MGB Assay. There were no invalid result calls. No interference was observed (see table below).

Potential InterferentInterferentConcentration#Detected/#TotalHSV-1#Detected/#TotalHSV-2IC
Whole blood with EDTA5% v/v0/30/33/3
Buffy coat5% v/v0/30/33/3
Acyclovir2.5 mg/mL0/30/33/3
Albumin5 mg/mL0/30/33/3
Casein7 mg/mL0/30/33/3
Female urine10% v/v0/30/33/3
Male urine10% v/v0/30/33/3
K-Y Brand jelly5% w/v0/30/33/3
Douche10% v/v0/30/33/3
Spermicide5% w/v0/30/33/3
Yeast-Gard1% w/v0/30/33/3
Monistat 15% w/v0/30/33/3
Monistat 35% w/v0/30/33/3
Vagisil Cream1% w/v0/30/33/3
Tioconazole 15% w/v0/30/33/3
Rite Aid Feminine Wash, Sensitive Skin10% v/v0/30/33/3
Clotrimazole-7 vaginal cream1% w/v0/30/33/3
Oral Analgesic Gel5% w/v0/30/33/3
Listerine antiseptic mouthwash10% v/v0/30/33/3
Abreva10% v/v0/30/33/3
Carmex lip balm1% w/v0/30/33/3
Releev cold sore treatment1% v/v0/30/33/3
Lip Clear lysine1% w/v0/30/33/3
Toothpaste5% w/v0/30/33/3
Acetaminophen5 mg/mL0/30/33/3
Wal-Finate5 mg/mL0/30/33/3
Cold-Eeze7% w/v0/30/33/3
Non-GMO Corn Starch1.25 mg/mL0/30/33/3
Zinc Oxide Ointment7% w/v0/30/33/3
Cough DM10mg/mL0/30/33/3
Lanacane Max Strength anti-itch cream7% w/v0/30/33/3
Seminal fluid7% v/v0/30/33/3
Foscarnet sodium5% v/v0/30/33/3

{12}------------------------------------------------

h. Carry-Over Contamination

The sample-to-sample carry-over from positive samples into the negative samples for the HSV 1&2 ELITe MGB Assay was studied by performing 5 integrated checkerboard runs (high positive for HSV-1 at 2.5×10′ TCIDso/mL concentration and negative samples (UTM) interspersed), which were compared to the overall contamination level ("background noise") of 2 negative sample runs. One operator ran 7 runs (5 checkerboard runs and 2 complete neqative runs). The High positive sample concentration in this study were high enough to exceed 95% or more of the concentration levels obtained from specimens of infected patients in the intended use population. No carry-over and cross contamination was observed. Overall percent agreement was 100% for positive and negative samples.

Run descriptionPositive Samples# Neg% Neg.Negative Samples# Pos.% Pos.
Run #1, BLANK0 / 0NA0 / 100 %
Run #2, Checkerboard0 / 50 %0 / 60 %
Run #3, Checkerboard0 / 60 %0 / 60 %
Run #4, BLANK0 / 0NA0 / 100 %
Run #5, Checkerboard0 / 60 %0 / 60 %
Run #6, Checkerboard0 / 60 %0 / 60 %
Run #7, Checkerboard0 / 60 %0 / 60 %
All runs0 / 290 %0 / 500 %

Carry-Over and Cross-Contamination Results

Sample Stability i.

This study assessed both sample stability and sample freeze-thaw stability. The samples for the stability evaluation were prepared by spiking both the HSV-1 and HSV-2 vendor quantitated viral stocks (HSV-1 MacIntyre strain and HSV-2 MS strain) in UTM, M4, M4RT, M5 and M6 media.

Each stability sample set consisted of:

  • 5 replicates spiked at 3×LoD. .
  • 5 replicates spiked at 1×103 TCIDso/mL, and ●
  • 5 replicates spiked at 1×10° TCIDso/mL (15 replicates total for each sample set). .

The stability of each sample set was assessed by sample incubation at +4°C for 1 week. All HSV-1 and HSV-2 samples were confirmed to be stable in UTM, M4, M4RT, M5 and M6 media for 1 week at +4°C.

The storage conditions were also validated by re-testing previously analyzed clinical samples that were stored in a -80°C freezer (≤ -70°C) for minimum of 4 months. Sample concentrations covered HSV clinical range. Ten HSV-1 or HSV-2 positive samples were tested for each media (except M6 for which only 7 HSV-positive samples were available). Positivity of all samples was confirmed after 4 month storage in a -80°C freezer (≤ -70°C).

A freeze thaw study was performed using 5 sample sets prepared as in the above using UTM, M4. M4RT, M5 and M6 media. All samples were subjected to 3 freeze-thaw cycles. All the samples were tested with the HSV 1&2 ELITe MGB Assay on the ELITe InGenius. The data obtained show that HSV-1 and HSV-2 viruses are stable after 3 freeze-thaw cycles in UTM, M4, M4RT, M5 and M6 media.

Matrix Comparison Study -

Since all analytical studies were conducted in the UTM (Universal Transport Media) and clinical studies were conducted using UTM, M4, M4RT, M5 and M6 media, the matrix comparison study was performed. The matrix comparison study was conducted using contrived sample panel made by spiking either HSV-1 or HSV-2 quantitated viral organisms

{13}------------------------------------------------

into each of the recommended media: UTM, M4, M4RT, M5 and M6. Each sample set consisted of 3 replicates spiked at 3×LoD, 3 replicates spiked at 1×10° TCIDsomL, and 3 replicates spiked at 1×10 TCID50/mL (9 replicates total for each sample set). Each sample was processed on the InGenius using the HSV 1&2 ELITe MGB Assay. All replicates in all media were detected and showed comparable results.

Target/ChannelSampleTiterTCID50/mLUTM,Avg CTSample MatrixAllMediaAvg CTAllMediaStDevAllMedia%CV
M4,Avg CTM4RT,Avg CTM5,Avg CTM6,Avg CT
HSV-2CH1,FAM1.00E+0527.1526.7626.4226.8227.2326.880.331.21%
1.00E+0333.8633.5933.7633.5134.1533.770.250.74%
3×LoD36.3235.5635.9635.5436.1435.910.350.97%
HSV-1CH4,AP5931.00E+0522.0221.1320.8220.7720.6321.080.562.66%
1.00E+0328.0128.4727.9728.5826.7227.950.742.64%
3×LoD35.8436.0737.0235.4334.6935.810.862.39%

All tested media showed comparable performance.

k. Reagent Stability

The stability of the HSV 1&2 ELITe MGB Assay was evaluated using several different methods.

Real-time Stability

A real-time stability was completed over a 12 month time period and the data support a shelf-life claim of 10 months for the HSV 1&2 ELITe MGB Assay.

Freeze-thaw Stability

Three lots of HSV 1&2 ELITe MGB Assay reagents were thawed, opened and held for a period of one hour at 30°C and then re-frozen by re-capping and replacing in ≤ -20°C (nominal) freezer. This cycle was repeated 8 times (once per day). Results show that the HSV 1&2 ELITe MGB Assay is stable when subjected to 8 freeze-thaw cycles.

11. Performance Characteristics

a. Reproducibility

The reproducibility of the HSV 1&2 ELITe MGB Assay was evaluated in a multi-site investigation using contrived clinical samples. HSV test panels were prepared by spiking HSV-1 (Maclntyre strain) or HSV-2 (MS strain) virus into UTM media at the concentrations of <1× LoD, 1 × LoD and 3 × LoD. HSV-1 and HSV-2 neqative panel members were included as panel member controls. The reproducibility panel composition is shown in the table below:

NameDescription of ContentsViral LoadExpected PositivityRate
M1HSV-1 C50 (High Negative) in UTM<1× LOD20-80% positive
M2HSV-1 C95 (Low Positive) in UTM1× LOD≥95% positive
M3HSV-1 C100 (Moderate Positive) in UTM2-3 × LOD100% positive
M4HSV-2 C50 (High Negative) in UTM<1× LOD20-80% positive
M5HSV-2 C95 (Low Positive) in UTM1× LOD≥95% positive
M6HSV-2 C100 (Moderate Positive) in UTM2-3 × LOD100% positive
M7HSV Negative in UTMNegative100% negative

{14}------------------------------------------------

Panels were tested at 3 sites by 2 operators per site with 1 run per operator per day, for 10 nonconsecutive days using a single lot of HSV 1&2 ELITe MGB Assay. Testing was performed on a minimum of 90 (30 per site) replicates per panel member for a total of >630 reproducibility samples (minimum 210 per testing site). Lot-to-Lot variability was assessed only at EGI MDx using 3 lots of HSV 1&2 ELITe MGB Assay. Controls were run daily and were included in the first run of the day.

% Agreement, average Cts and %CV for each panel member and per each site are presented in the table below.

Site - 1Site - 2Site - 3
%%%%
Agree-Agree-Agree-Agree-
ment withment withment withment with
ExpectedAvg.TotalExpectedAvg.TotalExpectedAvg.TotalExpected
TargetSampleResultsCt%CVResultsCt%CVResultsCt%CVResults95% Cl
HSV-1100.0%100.0%100.0%100.0%95.9 to
Low Pos(30/30)38.91.70%(30/30)38.32.10%(30/30)382.00%(90/90)100.0%
HSV-1100.0%100.0%100.0%100.0%95.9 to
Mod Pos(30/30)36.41.30%(30/30)35.55.20%(30/30)35.61.50%(90/90)100.0%
HSV-2100.0%100.0%100.0%100.0%95.6 to
Low Pos(30/30) aNANA(29/29) aNANA(30/30) aNANA(89/89)100.0%
HSV-1HSV-2100.0%100.0%100.0%100.0%95.9 to
ResultMod Pos(30/30) ªNANA(30/30) ªNANA(30/30) aNANA(90/90)100.0%
HSV100.0%100.0%100.0%100.0%97.3 to
(60/60)NANA(38/38) a41.42.50%(40/40)NANA(138/138)100.0%
Neg100.0%
Pos100.0%27.51.30%100.0%(515)27.51.20%100.0%(515)270.80%91.2 to
Control(30/30)100.0%(40/40)100.0%
Total(210/210)100.0%100.0%100.0%98.2 to
Agreement(162/162)(165/165)5371537)100.0%
4 Expected Results of HSV-2 Low Positive, HSV-2 Moderate Positive samples are "Negative" for HSV-1.
Site - 1Site - 2Site - 3
%%%%
Agree-Agree-Agree-Agree-
ment withment withment withment with
ExpectedAvg.TotalExpectedAvg.TotalExpectedAvg.TotalExpected
TargetSampleResultsCt%CVResultsCt%CVResultsCt%CVResults95% Cl
HSV-1100.0%NANA100.0%NANA100.0%NANA100.0%95.9 to
Low Pos(30/30) b(30/30) 6(30/30) b(90/90)100.0%
HSV-1100.0%NANA100.0%NANA100.0%NANA100.0%95.9 to
Mod Pos(30/30) P(30/30) P(30/30) P(90/90)100.0%
HSV-2100.0%36.83.10%100.0%37.82.30%100.0%36.61.90%100.0%95.9 to
Low Pos(30/30)(29/29)(30/30)(89/89)100.0%
HSV-2HSV-2100.0%35.21.30%100.0%35.951.60%100.0%34.62.30%100.0%95.9 to
ResultMod Pos(30/30)(30/30)(30/30)(90/90)100.0%
HSV100.0%100.0%100.0%100.0%95.9 to
Neg(60/60) 6NANA(38/38) 6NANA(40/40) 6NANA(138/138)100.0%
Pos100.0%100.0%100.0%100.0%95.9 to
Control(30/30)271.30%(5/5)27.41.50%(5/5)26.81.40%(40/40)100.0%
Total100.0%100.0%100.0%100.0%95.9 to
Agreement(210/210)(162/162)(165/165)5371537100.0%
P Expected Results of HSV-1 Low Positive, HSV-1 Moderate Positive samples are "Negative" for HSV-2.
Site - 1Site - 2Site - 3
%%%%
Agree-Agree-Agree-Agree-
ment withment withment withment with
ExpectedAvgTotalExpectedAvg.TotalExpectedAvg.TotalExpected
TargetSampleResultsCt%CVResultsCt%CVResultsCt%CVResults95% Cl
HSV-1100.0%100.0%100.0%100.0%95.9 to
Low Pos(30/30)30.43.80%(30/30)30.32.00%(30/30)30.21.70%(90/90)100.0%
HSV-1100.0%100.0%100.0%100.0%95.9 to
ICMod Pos(30/30)30.22.30%(30/30)30.42.80%(30/30)30.10.90%(90/90)100.0%
ResultHSV-2100.0%100.0%100.0%100.0%95.9 to
Low Pos(30/30)29.90.50%(29/29)30.42.20%(30/30)30.20.60%(89/89)100.0%
HSV-2100.0%100.0%100.0%100.0%95.9 to
Mod Pos(30/30)29.70.80%(30/30)30.41.20%(30/30)30.10.60%(90/90)100.0%
HSVNeg100.0%(60/60)30.21.10%100.0%(40/40)30.21.90%100.0%(38/38)30.10.90%100.0%(138/138)97.3 to100.0%
PosControl100.0%(30/30)29.31.40%100.0%(5/5)30.22.10%100.0%(5/5)29.40.90%100.0%(40/40)91.2 to100.0%
TotalAgreement100.0%(210/210)100.0%(164/164)100.0%(163/163)100.0%(537/537)98.2 to100.0%

{15}------------------------------------------------

The highest HSV 1&2 ELITe MGB Assay Site-to-Site variability (as measured by %CV based on Ct values) is 2.19%; the highest Lot is 0.23%, and the highest Operator variability is 0.93% for Moderate Positive panel members.

b. Clinical Performance

To evaluate the clinical performance of the HSV 1&2 ELITe MGB Assay, device performance was compared to a composite reference method. It consisted of an FDA cleared assay and a validated HSV 1&2 PCR followed by bi-directional sequencing of gel electrophoresis-positive samples). Validated HSV 1&2 PCR targeted genomic regions distinct from the HSV 1&2 ELITe MGB Assay. A positive result by the composite reference method is defined as a positive by the FDA cleared PCR or the validated sequencing. Two negative results are needed to confirm a negative)

A total of 1,174 left-over prospectively collected archived swab samples from cutaneous (546) and mucocutaneous (628) lesions from symptomatic patients were collected and evaluated in the study.

The samples were tested with HSV 1&2 ELITe MGB Assay and the Composite Reference Method. Out of the 1,174 tested samples were found invalid by the ELITe MGB Assay and were excluded from the performance analysis tables.

Out of the 1172 remaining samples 1 additional invalid sample result for HSV1 and 2 additional invalid sample results for HSV2 by the composite reference method were removed from the performance analysis tables.

Therefore for HSV1, 1171 samples analyzed and for HSV2 1170 samples were analyzed.

HSV-1 Positive/Neqative Percent Agreements (PPA/NPA) - Summary of the Results:

The PPA/NPA performance of HSV 1&2 ELITe MGB Assay when compared to the Composite Reference Method in detection of HSV-1 DNA in cutaneous and mucocutaneous lesions is summarized in the table below:

Summary of HSV-1 Results for Valid Cutaneous Lesion Samples (N=545)
HSV 1&2 ELITe MGB KitComposite reference method
PositiveNegativeTotal
Positive78785
Negative1459460
Total79466545
95% CI
PPA98.7%93.2-99.8%
NPA98.5%96.9-99.3%
Summary of HSV-1 Results for Valid Mucocutaneous Lesion Samples (N=626)
HSV 1&2 ELITE MGB KitComposite reference method
PositiveNegativeTotal
Positive12612138
Negative1487488
Total127499626
95% CI
PPA99.2%95.7-99.9%
NPA97.6%95.8-98.6%

{16}------------------------------------------------

HSV-2 Positive/Negative Percent Agreements (PPA/NPA) - Summary of the Results:

The PPA/NPA performance of HSV 1&2 ELITe MGB Assay when compared to the Composite Reference Method in detection of HSV-2 DNA in cutaneous and mucocutaneous lesions is summarized in the table below:

Summary of HSV-2 Results for Valid Cutaneous Lesion Samples (N=545)
HSV 1&2 ELITe MGB KitComposite reference method
PositiveNegativeTotal
Positive1256131
Negative5409414
Total130415545
95% CI
PPA96.2%91.3-98.3%
NPA98.6%96.9-99.3%
Summary of HSV-2 Results for Valid Mucocutaneous Lesion Samples (N=625)
HSV 1&2 ELITe MGB KitComposite reference method
PositiveNegativeTotal
Positive1648172
Negative4449453
Total168457625
95% CI
PPA97.6%94.0-99.1%
NPA98.2%96.6-99.1%

Results: Expected values/Reference Range ﻦ

The observed expected values for HSV-1 and HSV-2 in the study population using the HSV 1&2 ELITe MGB Assay were calculated for cutaneous and mucocutaneous specimens and is summarized for the combined sample set per age group, by gender and by specimen source in the tables below. A total number of 6 dual positives for HSV1 and HSV2 detected by the ELITe MGB Assay and one of the samples was confirmed positive by the composite reference method.

GenderAge GroupTotalHSV 1&2 ELITe MGBAssay HSV-1 resultsHSV 1&2 ELITe MGBAssay HSV-2 results
PositivePrevalencePositivePrevalence
Female<20421842.9%1228.6%
20-292446827.9%7028.7%
30-391432416.8%4531.5%
40-49971414.4%2525.8%
50-59881820.5%2427.3%
≥601232117.1%3024.4%
All73716322.1%20628.0%
Male<2020420.0%210.0%
20-291442517.4%3322.9%
30-391171512.8%2521.4%

{17}------------------------------------------------

40-4948510.4%1531.3%
50-5944613.6%920.5%
≥606158.2%1321.3%
All4346013.8%9722.4%
Gender is not identified100%00%
ALL117222319.0%30325.9%

Cutaneous HSV 1&2 Prevalence by Lesion Source

HSV 1&2 ELITe MGBAssayHSV-1 resultsHSV 1&2 ELITe MGBAssayHSV-2 results
Lesion SourceTotalPositivePrevalencePositivePrevalence
Genital/Anogenital2483815.3%7831.5%
Skin lesion2974715.8%5317.8%
Overall5458515.6%13124.0%

Mucocutaneous HSV 1&2 Prevalence by Lesion Source

HSV 1&2 ELITe MGBAssay HSV-1 resultsHSV 1&2 ELITe MGBAssay HSV-2 results
Lesion SourceTotalPositivePrevalencePositivePrevalence
Genital/Vaginal/Cervical50110921.8%16332.5%
Oral742128.4%22.7%
Other27518.5%27.4 %
Anorectal12216.7%541.7%
Urethral600 %00 %
Ocular500 %00 %
Nasal2150.0 %00 %
Overall62713822.0%17227.4%

d. HSV-2 Contrived Oral Panel Study

Due to the difficulty in obtaining sufficient HSV-2 positive oral samples, testing for HSV-2 was supplemented by using a contrived panel. The panel consisted of 75 individual negative cheek swab samples collected in UTM and spiked with HSV-2 at concentrations of 3×LoD, 8×LoD, 40×LoD, 200×LoD and 1000×LoD (10 of each), 10 HSV-1 Positive samples spiked at 10×LoD and 15 HSV-1/HSV-2 Negative Oral Samples

All panel members were randomized, blinded to the tested with HSV 1&2 ELITe MGB Assay on the ELITe InGenius instrument according to the clinical study protocol.

The HSV-2 Oral Contrived Panel Study revealed that 49 out of 50 oral HSV-2 contrived samples were positive using HSV 1&2 ELITe MGB Assay (98% detection). All 10 HSV-1 Positive samples confirmed 100% positivity.

{18}------------------------------------------------

Guidance Documents Referenced

FDA Guidance Documents:

    1. Evaluating Substantial Equivalence in Premarket Notifications 510(k)
    1. Statistical Guidance on Reporting Results from studies evaluating diagnostic tests
  • Guidance Content of Premarket Submissions for Management of Cybersecurity in 3. Medical Devices (DRAFT 10-2-14)
  • Guidance for Clinical Investigators, Sponsors, and IRBs Adverse Event Reporting to 4. IRBs - Improving Human Subject Protection
  • Guidance for Evaluation and Reporting of Age-, Race-, and Ethnicity-Specific Data in 5. Medical Device Clinical Studies (1500626) (09-12-17)
  • Guidance for Informed Consent for In Vitro Diagnostic Device Studies Using Leftover 6. Human Specimens that are Not Individually Identifiable
  • Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in 7. Medical Devices
  • Guidance for Refuse to Accept Policy for 510(k)s 8.

CLSI Guidance Documents:

  • EP05-A3 Vol. 34 No. 13 Evaluation of Precision Performance of Quantitative 1. Measurement Methods; Approved Guideline-Third Edition
    1. EP12-A2 Vol. 28 No. 3 - User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition
  • EP17-A2 Vol. 32 No. 8 Evaluation of Detection Capability for Clinical Laboratory 3. Measurement Procedures; Approved Guideline - Second Edition
  • EP24-A2 Vol. 31 No. 23 Assessment of the Diagnostic Accuracy of Laboratory Tests 4. Using Receiver Operating Characteristic Curves; Approved Guideline - Second Edition
  • EP25-A Vol. 29 No. 20 Evaluation of Stability of in Vitro Diagnostic Reagents; Approved 5. Guideline
    1. M29-A4 Vol. 34 No. 8 - Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline - 4th Edition
  • MM03 Molecular Diagnostic Methods for Infectious Diseases; Approved Guideline, 3rd 7. Edition
  • MM06-A2 Vol. 30 No. 22 Quantitative Molecular Methods for Infectious Diseases; 8. Approved Guideline Second Edition
    1. MM09-A2 Vol. 34 No.4 - Nucleic Acid Sequencing Methods in Diaqnostic Laboratory Medicine; Approved Guideline
    1. MM13-A Vol. 25 No. 31 Collection, Transport, Preparation, and Storage of Specimens for Molecular Methods; Approved Guideline
    1. MM17 Ed 2 Vol. 38 No. 9 Verification and Validation of Multiplex Nucleic Acid Assays (NATS); Approved Guideline

§ 866.3309 Herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel.

Link to an amendment published at 90 FR 55982, Dec. 4, 2025. (a)
Identification. A herpes virus nucleic acid-based cutaneous and mucocutaneous lesion panel is a qualitative in vitro diagnostic device intended for the simultaneous detection and differentiation of different herpes viruses in cutaneous and mucocutaneous lesion samples from symptomatic patients suspected of Herpetic infections. Negative results do not preclude infection and should not be used as the sole basis for treatment or other patient management decisions. The assay is not intended for use in cerebrospinal fluid samples.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed documentation for the device description, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer design and selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination.
(3) Premarket notification submissions must include detailed documentation of a clinical study using lesion samples in which Herpes Simplex Virus 1, Herpes Simplex Virus 2, or Varicella Zoster Virus DNA detection was requested. The study must compare the device performance to an appropriate well established reference method.
(4) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(5) The device labeling must include a limitation statement that reads: “The device is not intended for use with cerebrospinal fluid or to aid in the diagnosis of HSV or VZV infections of the central nervous system (CNS).”
(6) Premarket notification submissions must include quality assurance protocols and a detailed documentation for device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software.
(7) The risk management activities performed as part of the manufacturer's 21 CFR 820.30 design controls must document an appropriate end user device training program that will be offered as part of efforts to mitigate the risk of failure to correctly operate the instrument.