K Number
K180473
Manufacturer
Date Cleared
2018-03-23

(29 days)

Product Code
Regulation Number
892.1650
Panel
RA
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ANY VIEW DR SERIES (Anyviw-500DR, Anyview-240DR) is radiation medical equipment only used by professional radiologists. This product is designed to provide fluoroscopic and spot film images of the patient during diagnostic and interventional procedures.

This system can be applied in emergency room, operation room, cast room or etc. of hospital.

Device Description

The ANYVIEW DR SERIES (Anyviw-500DR, Anyview-320DR, Anyview-240DR) are mobile x-ray fluoroscopic imaging systems used by radiation experts. ANYVIEW DR SERIES are a digital fluoroscopic imaging systems with Flat Panel Detector (FPD) used in diagnostic and interventional procedures. ANYVIEW DR SERIES are composed of C-arm, x-ray generating equipment (x-ray controller, high voltage generator, x-ray tube, motor-type collimator), FPD, and workstation (console computer and monitor).

ANYVIEW imaging software is a Digital Imaging System (DIS) designed for C-arm, ANYVIEW FPD Fluoroscopic Mobile X-ray System. ANYVIEW imaging software provides useful functions to manage X-ray images obtained from ANYVIEW DR SERIES FPD Fluoroscopic Mobile X-ray System.

ANYVIEW imaging software provides various image tools. One of the most noticeable features is that the C-arm imaqes taken during an exam are stored in the database for further review. I Image data is integrated with the patient information in DICOM(OPTION) compatible format which allows compatibility with existing DICOM and PACS system.

AI/ML Overview

The ECOTRON Co., Ltd. ANYVIEW DR SERIES FPD Fluoroscopic Mobile C-Arm (ANYVIEW-240DR, ANYVIEW-320DR, ANYVIEW-500DR) is a radiation medical equipment designed to provide fluoroscopic and spot film images for diagnostic and interventional procedures. The device was found to be substantially equivalent to its predicate device, Anyview-500R Fluoroscopic Mobile X-ray System (K160279), based on non-clinical performance data.

1. Table of Acceptance Criteria and Reported Device Performance:

The provided document does not explicitly state formal "acceptance criteria" in a go/no-go format for performance metrics. Instead, it presents a comparison of technical characteristics and performance data of the subject device against its predicate and reference devices, demonstrating equivalent or improved performance without raising new safety or effectiveness concerns.

CharacteristicAcceptance Criteria (Implied by Predicate/Reference)Subject Device Performance (ANYVIEW DR SERIES)Predicate Device Performance (Anyview-500R, K160279)Reference Device Performance (D2RS AT, K150306)
Intended UseFluoroscopic and spot film images for diagnostic and interventional procedures.SameSameN/A (Reference device details less explicit for this)
Energy Source220V~230V, Single 50/60 HzSameSameN/A
X-ray Generator TypeHFG INVERTER TYPESimilar (HFG INVERTER TYPE)HFG INVERTER TYPEN/A
Fluoroscopy Continuous Mode0.5-10mASameSameN/A
Fluoroscopy Pulsed Mode0.5-20mASameSameN/A
Fluoroscopy Boost Mode30mASameSameN/A
Radiography kV Range40-125 kVSimilar (40-125 kV depending on model)40-125 kVN/A
Radiography mA Range20-100 mASimilar (15-100 mA depending on model)20-100 mAN/A
Radiography mAs Range0.8-200 mAsSimilar (0.8-200 mAs depending on model)0.8-200 mAsN/A
X-ray Tube TypeTOSHIBA XR-2551TOSHIBA XRR-2251 (Same)TOSHIBA XR-2551N/A
Detector TypeImage Intensifier (Predicate) / Flat Panel Detector (Reference)Flat Panel: TFT: a-Si w/
CsI: TI scintillatorImaging Intensifier: E5830SD-P4APixium RF 4343 (FPD)
Detector Input FOV9 inch (Predicate)12 x 12 in9 inch17x17 in
Detector Central Resolution54/62/70 lp/cm (Predicate)3.4 lp/mm54/62/70 lp/cm (for 9/6/4.5 inch modes)3.4 lp/mm
Detector MTF (1 lp/mm)N/A (Predicate, different detector type)54% (VS VIeworks), 60% (VS Thalaes)N/A62%
Detector DQE (1 lp/mm)65% (Predicate)69% (VS VIeworks), 69% (VS Thalaes)65%52%
Detector Spatial Resolution5.2 lp/mm (Predicate)3.4 lp/mm5.2 lp/mm (min)3.4 lp/mm
Pixel Size/Output Resolution1K x 1K (Predicate - implied)2K x 2K1K x 1K (implied from 12bit grayscale)N/A (implied by 2874x2840 pixels)
Electrical Safety StandardsIEC 60601-1, -1-2, -1-3, -2-28, -2-43, -2-54Meets all listed IEC standardsMeets all listed IEC standardsN/A
EPRC Performance Standard21 CFR 1020.30/1020.32Meets 21CFR 1020.30/1020.32Meets 21CFR 1020.30/1020.32N/A

2. Sample size used for the test set and the data provenance:

  • Test Set: The study primarily relies on bench testing and comparative performance data of components (specifically the Flat Panel Detector, FPD) for non-clinical evaluation. There is no mention of a traditional "test set" of patient images.
  • Data Provenance: The data provenance for the performance characteristics (MTF, DQE, Spatial Resolution) of the detectors is from the respective manufacturers (VIEWORKS for Subject Device, TOSHIBA for Predicate Device, and Thales for Reference Device). This is non-clinical, technical performance data. The document does not specify country of origin for this component-level data, nor whether it is retrospective or prospective, though it is typically collected during product development/testing by the manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

Not applicable. The study is a 510(k) submission for substantial equivalence based on technical and performance characteristics of medical hardware, not on clinical image interpretation or diagnostic accuracy by experts. No ground truth was established by human experts for a test set of images.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable, as no test set requiring expert adjudication was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No MRMC comparative effectiveness study was done. The device described is an X-ray imaging system, not an AI-powered image analysis tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

Not applicable. The device is an imaging system, not an algorithm, and its performance evaluation does not involve standalone algorithm assessment.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

Not applicable. The equivalence is established through engineering and technical performance comparisons of the device and its key components against predicate and reference devices, not against clinical ground truth. Performance metrics like MTF, DQE, and spatial resolution are derived from physical measurements and scientific testing of the imaging system and its components.

8. The sample size for the training set:

Not applicable. This is a medical device for imaging acquisition, not a machine learning or AI algorithm, so there is no concept of a "training set."

9. How the ground truth for the training set was established:

Not applicable.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.